2 User’s Guide
TABLE OF CONTENTS
Product Description and Intended Use......................................................
Warnings and Cautions................................................................................
Symbol Explanations....................................................................................
EU Declaration of Conformity for Medical Applications...........................
Please check the following items are present when you unpack the box, and save the
packing materials in case you will need to ship or transport the monitor in future.
•AMM240WTD LCD Monitor and two video cables (1) HD15 VGA cable* (1) DVI-D
cable*
*
•AC Power cord**
•AC-Adapter
•Composite Video BNC Jack Cable and Super Video Cable
CAUTION
Manufacturer : JEC Korea
Model No :
JMW1 150KA2400F04
*
•User Manual and 4pcs VESA screws
*Might be optional item, check with local representative
**Might vary pending on region stardard
4 User’s Guide
Warnings and Cautions
Please read this manual and follow its instructions carefully . The words warning,
caution, and note carry special meanings and should be carefully reviewed:
Warranty is void if any of these warnings are disregarded.
ADV AN Int’l Corp accepts full responsibility for the ef fects on safety, reliability , and
performance of the equipment only if:
• Re-adjustments, modifications, and/or repairs are carried out exclusively
by ADVAN Int’l Corp.
• The electrical installation of the relevant operating room complies with the
applicable IEC and CE requirements.
Warning Federal law (United States of America) restricts this
device to use by, or on order of a physician.
The ADV AN Int’l Corp AMM240WTD monitor has been tested under UL 60601-1
standard and UL listed for Medical application.
ADV AN Int’l Corp reserves the right to make improvements in the product(s)
described herein. Product(s), therefore, may not agree in detail to the published
design or specifications. All specifications are subject to change without notice.
Please contact ADV AN Int’l Corp directly or phone your local ADVAN Int’l Corp sales
representative or agent for information on changes and new products.
AMM240WTD 5
Warnings
1. Read the operating manual thoroughly and be familiar with its contents prior to using
this equipment.
2. Carefully unpack the unit and check if any damage occurred during shipment.
3. Should any solid object or liquid fall into the panel, unplug the unit and have it checked
by qualified personnel before operating it any further.
4. Uplug the unit if it is not to be used for an extended period of time. To disconnect the
cord, pull it out by the plug. Never pull the cord itself.
5. Be a qualified physician, having complete knowledge of the use of this equipment.
6. Test this equipment prior to a medical procedure. This monitor was fully tested at the
factory before shipment.
7. Avoid removing covers on control unit to avoid electric shock.
8. Attempt no internal repairs or adjustments not specifically detailed in this operating
manual.
9. Pay close attention to the care, cleaning instructions in this manual. A deviation may
cause damage (refer to the Cleaning section).
10. DO NOT STERILIZE MONITOR.
11. Read the entire instruction manual before assembling or connecting the camera.
12. Do not place the monitor or any other heavy object on the power cord. Damage to the
cable can cause fire or electirc shock.
13. Monitor with power supply is suitable for use in patient environment.
14. DO NOT stack more than 8 boxes high.
15. When the monitor is equipped with the base stand, please install the base stand
at horizontally flat place for normal use.
This equipment has been tested and found to comply with the limits for medical devices
in IEC 601-1-2:2003. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to
other devices in the vicinity. However, there is no guarantee that interference will not occur
in a particular installation. If this equipment does cause harmful interference to other
devices, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
- Consult the manufacturer or field service technician for help.
NOTICES TO USER
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
FCC WARNING
This equipement generates or uses radio frequency energy. Changes or modifications not
expressly approved by the party responsible for compliance could void the user’s authority
to operate the equipment. The user could lose the authority to operate this equipment if an
unauthorized change or modification is made.
6 User’s Guide
Cautions
1. The AC Adapter must be plugged into a Grounded power outlet.
2. Use only the proprietary AMM240WTD power supply for the AMM240WTD monitor.
Make a proper connection by ensuring that the shrink tubing completely secures the
connection between the DC power cord and the extension cord.
3. Turn power off when unit is not in use.
4. Never operate the unit right after having transported from a cold location directly to a
warm location.
5. Do not expose the monitor to moisture or directly apply liquid cleaners directly to the
screen. Spray the cleaning solution into a soft cloth and clean gently.
6. Handle the monitor with care. Do not strike or scratch the screen.
7. Do not block the monitor cooling vents. The monitor is cooled by natural convection
and has no fan.
8. Do not force the monitor past 28 degrees of vertical when adjusting the screen
position. (For monitors equipped with stands only.)
9. Remove the power module and connection when transporting the unit.
10. Save the original carton and associated packing material. They will be useful should
you have to transport or ship the unit.
11. Allow adequate air circulation to prevent internal heat buildup.
12. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains,
draperies) that may block the ventilation slots.
13. Do not install the unit near sunlight, excessive dust, mechanical vibration or shock.
14. The unit is designed for operation in a horizontal position. Never operate the unit in
a vertical position.
15. Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker.)
16. Do not expose the monitor to moisture or excessive dust.
17. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or
IEC 60601-1-1 harmonized national standard or the combination should be evaluated.
Do not touch the patient with signal input or output connectors.
18. Use only a hospital grade power supply cord.
19. This equipment has been tested and found to comply with the limit for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide resonable generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference
will not occur in a particular installation, which can be detemined by turing the
equipment off and on, the user is encourage to try to correct the interference by one
or more of the following measures:
20. Grounding reliability can only be achieved when the equipment is connected to an
equipment receptacle labeled “Hospital Only” or “Hospital Grade.”
NoteTo connect to an international power supply, use a an attachment plug
appropriate for the power outlet.
NoteRefer to the “Electromagnetic Compatibility” (EMC) section of this manual to
ensure EMC. The AMM240WTD must be installed and operated according to the
EMC information provided in this manual.
• Reorient or relocate the receiving device.
• Increase the separation distance between the equipment.
• Connect the equipment to an outlet on a circuit different from that to which the
other device(s) are connected.
• Consult the manufacturer or field service technican for help.
Symbol Definitions
AMM240WTD 7
8 User’s Guide
EU Declaration of Conformity for Medical Applications
A Declaration of Conformity has been filed for this product. A sample of
this document may be found in the addendum which accompanied this
manual. For a copy of the Declaration of Conformity document, please
contact ADVAN Int’l Corp. and request for AMM240WTD DOC.
Prepare for Unpack
Before you unpack your monitor, prepare a suitable workspace. You
need a stable and level surface near a grounded wall outlet in an area
which is relatively free of glare from sunlight or other sources of bright
light. The monitor is cooled by natural convection (it has no fan).
For optimum performance, do not block the cooling vents.
While unpacking the monitor, inspect it and other package contents for
shipping damage that could cause a fire or shock hazard. Immediately
report any shipping damage to the carrier or transportation company
and contact customer service for monitor in the future or in case of
return.
After you unpack the monitor, make sure the following items are included
• Monitor with video cable
• AC adapter with cable
CAUTION: AC Adapter must be plugged into Grounded a power outlet
CAUTION : AC adapter
Manufacturer: JEC Korea
Model No: JMW1150KA2400F04
• This operations manual
Note: Your system provider may offer alternative cords or cables
depending on the installation requirement and local geography issues.
AMM240WTD 9
SAFETY PRECAUTION
·Avoid placing the monitor , or any other heavy object, on the power cord to prevent
fire or electrical shock from damage to the power cord.
·Do not expose the monitor to rain, excessive moisture, or dust to avoid fire or shock
hazard.
·Do not cover the slots or openings of the monitor for proper heat dissipation. Always
put the monitor in a place where there is adequate ventilation.
·Avoid placing the monitor against a bright background or where sunlight or other
light sources may reflect on the area of the monitor. Place the monitor just below
eye level.
·Handle with care when transporting the monitor.
·Refrain from giving the shock or scratch to the screen, as screen is fragile.
CLEANING YOUR MONITOR
No specific liquid or chemical necessary when cleaning this LCD monitor
However, we suggest to clean the monitor with non-abrasive cloths and cleaning
solutions used in hospitals to clean similar equipment. We recommend using 70%
Isopropyl alcohol for the screen surface and warm water and a mild detergent for all
other surfaces. Other acceptable cleaning agents are listed below:
• 70% isopropyl alcohol
• Cidex (2.4% glutaraldehyde solution)
• 0.5% Chlorhexidine in 70% isopropyl alcohol
To clean the screen, do not spray liquid cleaners directly on to the unit. St and away
Form the monitor and spray cleaning solution onto a cloth. Without applying excessive pressure, clean the screen with the slightly dampened rag.
10 User’s Guide
POWER MANAGEMENT FUNCTION
The monitor is equipped with the power management function which automatically reduce
the power consumption when not in use in three power level modes.
•Stand-by Mode
The monitor goes into stand-by mode when the horizontal sync signal is off for
about 10 seconds. In this mode, the screen goes off and the power LED blinks for 1
seconds On and 1 second Off. The screen is displayed after the horizontal sync
signal is restored.
•Suspend Mode
The monitor goes into suspend mode when the vertical sync signal is off for about
10 seconds. The power consumption during this is less than 8 W . In this mode, the
screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The
screen is displayed after the vertical sync signal is restored.
•Off Mode
The monitor goes into power-off mode when the vertical and horizontal sync signals
are off for about 10 seconds. In this mode, the screen goes off and the power LED
blinks for 1 seconds On and 1 second Off. The screen is displayed after the vertical
and horizontal sync signals are restored.
Power Management System
The AMM240WTD Medical Monitor power management proposal. Provides four
phases of power-saving modes by detecting the horizontal sync signal as shown in
the table below .