BARCO AMM190WTDSW Users Manual

WiSe™ 19" HDTV
Surgical Display
0240030990
Contents
1 Warnings and Cautions
2 Warnings 2 Cautions
4 About Your Device
10 Setup
10 Connections 12 Basic Video Setup
13 Operation
13 On-Screen Display (OSD) 19 Troubleshooting
21 Periodic Maintenance
21 Cleaning 22 Disposal
23 Technical Specications
27 Symbols
Warnings and Cautions
Please read this manual and follow its instructions carefully. e words warning, caution, and note carry special meanings and should be carefully reviewed:
Warning e personal safety of the patient may be involved. Disregarding this
information could result in personal injury.
Caution Special service procedures or precautions must be followed to avoid
damaging the device.
Note Special information to make maintenance easier or important
information more clear. An exclamation mark within a triangle is intended to alert the
user to the presence of important operating and maintenance instructions in this manual.
A lightning bolt within a triangle is intended to warn of the presence of hazardous voltage. Refer all service to authorized personnel.
e WiSe 19" HDTV Surgical Display has been tested under the UL 60601-1 standard and is UL listed for medical application.
e warranty is void if any of the following warnings or cautions are disregarded.
1
Warnings
To avoid potential serious injury to the user and the patient and/or damage to this device, please note the following warnings:
• Read this manual thoroughly and be familiar with its contents prior to using this device.
• Federal law (United States of America) restricts this device to sale by, or on the order of, a physician.
• Carefully unpack the device and check if any damage occurred during shipment.
• is device is non-sterile and therefore should not be placed in the sterile eld.
• Do not place the device or any other heavy object on the power cord. Damage to the cable can cause re or electric shock.
• To avoid electric shock, avoid removing the bezel.
• is device should not be used adjacent to or stacked with other devices. If adjacent or stacked use is necessary, the device should be observed to verify normal operation in the conguration in which it will be used.
• Test this device prior to a surgical procedure. is device was fully tested at the factory before shipment.
• Do not attempt internal repairs or adjustments not specically detailed in this manual. Ensure that readjustments, modications, and/or repairs are carried out by persons authorized by Stryker Endoscopy.
• Do not put any liquid or solid object into the panel. If this occurs, unplug the device and have it checked by qualied personnel before operating it any further.
• Use appropriate caution to prevent contact with uids if the device is being used with a power supply in patient environments.
• e use of cables and/or other accessories with this device, other than those specied, may result in increased emissions or decreased immunity of this device.
Cautions
• Connect the device to an AC adapter connected to a hospital grade power cord ensuring the power cord is plugged into a grounded power outlet to achieve grounding reliability.
• Do not sterilize the device, as the delicate electronics cannot withstand this procedure.
2
• Use only the proprietary surgical display power supply for the display. Completely secure the connection between the DC power cord and the extension cord.
• Never operate the device immediately aer transportation from a cold location to a warm location.
• To connect to an international power supply, use an attachment plug appropriate for the power outlet, as outlined in the "Technical Specications" section of this manual.
• Unplug the device if it is not to be used for an extended period of time. To disconnect the cord, unscrew the plug rst, then pull the cord out by the plug. Never pull the cord itself.
• Do not expose the device to moisture or apply liquid cleaners directly to the screen. Spray the cleaning solution into a so cloth and clean gently. For further detail, refer to the "Periodic Maintenance" section of this manual.
• Allow adequate air circulation to prevent internal heat buildup. Do not place the device on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the ventilation slots. e device is cooled by natural convection and has no fan.
• Do not touch the patient with signal input or output connectors. Equipment with SIP/SOP connectors should either comply with IEC60601-1 and/or IEC 60601-1-1 harmonized national standards or the combination should be evaluated for safety.
• To ensure electromagnetic compatibility, refer to the “Electromagnetic Compatibility” section of this manual. e WiSe 19" HDTV Surgical Display (0240030990) must be installed and operated according to the EMC information provided in this manual.
• Pay close attention to the cleaning instructions in this manual. A deviation may cause damage.
• Do not install the device near sunlight, excessive dust, mechanical vibration, or shock.
• Do not operate with the glass device screen facing downward.
• Handle the device with care. Do not strike or scratch the screen.
• Changes or modications not expressly approved by the party responsible for compliance could void the user’s authority to operate the device.
Note: is device has been tested and found to comply with the limit for a Class B digital device, pursuant to Part 15 of the FCC Rules. ese limits are designed to provide reasonable protection against harmful interference in a residential installation. is device generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. ere is no guarantee that interference will not occur in a particular installation, which can be determined by turning the device o and on. e user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation distance between
the device.
Connect the device to an outlet on a
circuit dierent from that to which the other device(s) are connected.
Consult the manufacturer or eld service
technician for help.
3
About Your Device
e WiSe 19" HDTV Surgical Display (display) is a wide screen LCD surgical display that can support a maximum resolution of 1080p. e display supports various video inputs, including digital RGB, analog RGB, serial digital interface (SDI), component video (YPbPr/RGB), S-video, C-video, and wireless RGB.
e display features an optional WiSe HDTV Transmitter, which allows it to receive a high-denition video signal over a radio-frequency link (USA and Canada only).
Intended Use
e WiSe 19" HDTV Surgical Display is intended for video display during surgical procedures.
4
Indications
is device is indicated for the following surgical procedures:
• General surgery
• General laparoscopy
• Nasopharynogoscopy
• Ear endoscopy
• Sinusocopy
• Plastic surgery wherever a laparoscope/endoscope/ arthroscope is indicated for use
• Laparoscopic cholecystectomy
• Laparoscopic hernia repair
• Laparoscopic appendectomy
• Laparoscopic pelvic lymph node dissection
• Laparoscopically assisted hysterectomy
• Laparoscopic & thorascopic anterior spinal fusion
• Anterior cruciate ligament reconstruction
e indicated users of this device are as follows:
• Knee arthroscopy
• Shoulder arthroscopy
• Small joint arthroscopy
• Decompression xation
• Wedge resection
• Flexible endoscopy
• Urology
• Gynecology
• Lung biopsy
• pleural biopsy
• Dorsal sympathectomy
• Pleurodesis
• Internal mammary artery dissection for coronary artery bypass graing where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement
• General surgeons
• Gynecologists
• Cardiac surgeons
• oracic surgeons
e display is a non-sterile, reusable device not intended for use in the sterile eld. e display is intended for use by qualied physicians having complete knowledge of these surgical procedures.
• Plastic surgeons
• Orthopedic surgeons
• ENT surgeons
• Urologists
Contraindications
ere are no known contraindications for this device.
5
Package Content
1
4
3
5
Package Content Part Number
1. WiSe 19" HDTV Surgical Display 240-030-990
2. (4) M4 × 10 mm VESA screws
3. Hospital-grade AC power cord (USA and Canada only)
4. WiSe 19'' Surgical Power Supply 240-030-992
5. Cable Cover
2
Optional Accessories Part Number
WiSe HDTV Transmitter (USA and Canada only) 240-030-971 15-. (5 pin) DC extension cable 240-030-951 75-. (5 pin) DC extension cable 240-030-952 WiSe 19" Display Cover 240-030-991 Hospital-grade AC power cord (USA and Canada only) 105-033-001
6