BARCO AMM190WTD User Manual

AUTO - SCANNING WITH DIGITAL CONTROL LCD COLOR MEDICAL MONITOR
AMM190WTD
Operation Manual
For future reference, record the serial number of your display monitor in the space below :
Serial Number The serial number is located on the back of the monitor
back of the monitor
The title «WARNING» is used to inform the users of pos­sible causes that could infl ict the injury, death, or property damage to the patients.
The title «CAUTION» is used to inform the users of pos­sible causes that could infl ict the patients although it might not severe enough to cause deaths.
AMM190WTD 1
ENGLISH
WARNING
CAUTION
NOTE
The title «NOTE» is used to inform the users of items that are of importance in terms of installation, operation, or maintenance of the Equipment although the failure does not infl ict the bodily harm to the patients.
ADVAN INT’L CORP.2009
All other trademarks are the property of their reference owners. This document is subject to change without notice. Advan provides this information as reference only. Reference to other vendor’s products does not imply any recommendation or endorsement.
Revision Control
Date Description
REV.03 111809 Document number
2 User’s Guide TABLE OF CONTENTS
Product Description and Intended Use ......................................................
Warnings and Cautions .............................................................................
Symbol Explanations .................................................................................
EU Declaration of Conformity for Medical Applications .............................
Safety Precaution ......................................................................................
Cleaning Your Monitor ...............................................................................
Power Management Function....................................................................
Preset Modes..............................................................................................
Video Signals .............................................................................................
PIP / POP / PBP function...........................................................................
DDC ...........................................................................................................
Installation .................................................................................................
Connecting the Power Cord ......................................................................
User Function / Messages.........................................................................
OSD Section ..............................................................................................
Troubleshooting .........................................................................................
Specifi cation of (AMM190WTD) ................................................................
Classifi cation .............................................................................................
Electromagnetic Compatibility ...................................................................
Dimension Drawing (mm) of AMM190WTD...............................................
Connectors ................................................................................................
Description of Warranty .............................................................................
3 4 7 8 9 9 10
11 12 13 14 14 15 16 17 25 26 27 28 31 32 33
AMM190WTD 3
Product Description and Intended Use
Please check the following items are present when you unpack the box, and save the packing materials in case you will need to ship or transport the monitor in future.
AMM190WTD LCD Monitor and two video cables (1) HD15 VGA cable* (1) DVI-D cable*
ENGLISH
*
AC Power cord**
AC-Adapter
Composite Video BNC Jack Cable and Super Video Cable
CAUTION Model No : JMW1100KA1800F01
*
User Manual and 4pcs VESA screws
*Might be optional item, check with local representative **Might vary pending on region stardard
4 User’s Guide
Warnings and Cautions
Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed:
Warning The personal safety of the patient or physician may be involved. Disregarding this information could result in injury to the patient or physician.
Caution Special service procedures or precautions must be followed to avoid damaging the instrument.
Note Special information to make maintenance easier or important
information more clear.
An exclamation mark within a triangle is intended to alert the
user to the presence of important operating and maintenance instructions in the literature accompanying the product.
A lightning bolt within a triangle is intended to warn of the presence of hazardous voltage. Refer all service to authorized
personnel.
Warning TO AVOID POTENTIAL SERIOUS INJURY TO THE USER
AND THE PATIENT AND/OR DAMAGE TO THIS DEVICE, THE USER MUST :
Warranty is void if any of these warnings are disregarded.
ADVAN Int’l Corp accepts full responsibility for the effects on safety, reliability, and performance of the equipment only if:
• Re-adjustments, modifi cations, and/or repairs are carried out exclusively by ADVAN Int’l Corp.
• The electrical installation of the relevant operating room complies with the applicable IEC and CE requirements.
Warning Federal law (United States of America) restricts this device to use by, or on order of a physician.
The ADVAN Int’l Corp AMM190WTD monitor has been tested under UL 60601-1 standard and UL listed for Medical application.
ADVAN Int’l Corp reserves the right to make improvements in the product(s) described herein. Product(s), therefore, may not agree in detail to the published design or specifi cations. All specifi cations are subject to change without notice. Please contact ADVAN Int’l Corp directly or phone your local ADVAN Int’l Corp sales representative or agent for information on changes and new products.
AMM190WTD 5
Warnings
1. Read the operating manual thoroughly and be familiar with its contents prior to using this equipment.
2. Carefully unpack the unit and check if any damage occurred during shipment.
3. Should any solid object or liquid fall into the panel, unplug the unit and have it checked by qualifi ed personnel before operating it any further.
4. Uplug the unit if it is not to be used for an extended period of time. To disconnect the cord, pull it out by the plug. Never pull the cord itself.
5. Be a qualifi ed physician, having complete knowledge of the use of this equipment.
6. Test this equipment prior to a medical procedure. This monitor was fully tested at the factory before shipment.
7. Avoid removing covers on control unit to avoid electric shock.
8. Attempt no internal repairs or adjustments not specifi cally detailed in this operating manual.
9. Pay close attention to the care, cleaning instructions in this manual. A deviation may cause damage (refer to the Cleaning section).
10. DO NOT STERILIZE MONITOR.
11. Read the entire instruction manual before assembling or connecting the camera.
12. Do not place the monitor or any other heavy object on the power cord. Damage to the cable can cause fi re or electirc shock.
13. Monitor with power supply is suitable for use in patient environment.
14. DO NOT stack more than 8 boxes high
This equipment has been tested and found to comply with the limits for medical devices in IEC 601-1-2:2003. These limits are designed to provide reasonable protection against harm­ful interference in a typical medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devic­es in the vicinity . However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
- Consult the manufacturer or fi eld service technician for help.
NOTICES TO USER
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any inter-
ference received, including interference that may cause undesired operation.
ENGLISH
FCC WARNING
This equipement generates or uses radio frequency energy. Changes or modifi cations not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. The user could lose the authority to operate this equipment if an
unauthorized change or modifi cation is made.
6 User’s Guide
Cautions
1. The AC Adapter must be plugged into a Grounded power outlet.
2. Use only the proprietary AMM190WTD power supply for the AMM190WTD monitor. Make a proper connection by ensuring that the shrink tubing completely secures the connection between the DC power cord and the extension cord.
3. Turn power off when unit is not in use.
4. Never operate the unit right after having transported from a cold location directly to a warm location.
5. Do not expose the monitor to moisture or directly apply liquid cleaners directly to the screen. Spray the cleaning solution into a soft cloth and clean gently.
6. Handle the monitor with care. Do not strike or scratch the screen.
7. Do not block the monitor cooling vents. The monitor is cooled by natural convection and has no fan.
8. Do not force the monitor past 28 degrees of vertical when adjusting the screen position. (For monitors equipped with stands only.)
9. Remove the power module and connection when transporting the unit.
10. Save the original carton and associated packing material. They will be useful should you have to transport or ship the unit.
11. Allow adequate air circulation to prevent internal heat buildup.
12. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the ventilation slots.
13. Do not install the unit near sunlight, excessive dust, mechanical vibration or shock.
14. The unit is designed for operation in a horizontal position. Never operate the unit in a vertical position.
15. Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker.)
16. Do not expose the monitor to moisture or excessive dust.
17. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard or the combination should be evaluated. Do not touch the patient with signal input or output connectors.
18. Use only a hospital grade power supply cord.
19. This equipment has been tested and found to comply with the limit for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable generate uses and can radiate radio frequency energy, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation, which can be detemined by turing the equipment off and on, the user is encourage to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation distance between the equipment.
• Connect the equipment to an outlet on a circuit different from that to which the
other device(s) are connected.
• Consult the manufacturer or fi eld service technican for help.
20. Grounding reliability can only be achieved when the equipment is connected to an
equipment receptacle labeled “Hospital Only” or “Hospital Grade.” Note To connect to an international power supply, use a an attachment plug
appropriate for the power outlet. Note Refer to the “Electromagnetic Compatibility” (EMC) section of this manual to
ensure EMC. The AMM190WTD must be installed and operated according to the EMC information provided in this manual.
Symbol Defi nitions
Dangerous: High Voltage
Direct Current
Serial Number
Top – Bottom
Fragile
Do not get wet
AMM190WTD 7
ENGLISH
51LJ Medical Equipment E215822
Consult accompanying documents. Indicates proof of conformity to applicable European Economic
Community Council directives and to harmonized standards published in the offi cial journal of the European Communities.
Authorized Representative in Europe
Indicates protective earth ground.
For indoor use only.
Medical Equipment is in accordance with UL 60601-1 and CAN/ CSA C22.2 No. 601.1 in regards to electric shock, fi re hazards, and mechanical hazards.
Tested to comply with FCC Class B standards.
DC power control switch
8 User’s Guide
EU Declaration of Conformity for Medical Applications
A Declaration of Conformity has been fi led for this product. A sample of this document may be found in the addendum which accompanied this manual. For a copy of the Declaration of Conformity document, please contact ADVAN Int’l Corp. and request for AMM190WTD DOC.
Prepare for Unpack
Before you unpack your monitor, prepare a suitable workspace. You need a stable and level surface near a grounded wall outlet in an area which is relatively free of glare from sunlight or other sources of bright light. The monitor is cooled by natural convection (it has no fan). For optimum performance, do not block the cooling vents.
While unpacking the monitor, inspect it and other package contents for shipping damage that could cause a fi re or shock hazard. Imme- diately report any shipping damage to the carrier or transportation company and contact customer service for monitor in the future or in case of return.
After you unpack the monitor, make sure the following items are in­cluded
• Monitor with video cable
AC adapter with cable
CAUTION: AC Adapter must be plugged into Grounded a power outlet
CAUTION : AC adapter Model No: JMW1100KA1800F01
This operations manual
Note: Your system provider may offer alternative cords or cables depending on the installation requirement and local geography issues.
AMM190WTD 9
SAFETY PRECAUTION
· Avoid placing the monitor, or any other heavy object, on the power cord to prevent fi re or electrical shock from damage to the power cord.
· Do not expose the monitor to rain, excessive moisture, or dust to avoid re or shock hazard.
· Do not cover the slots or openings of the monitor for proper heat dissipation. Always put the monitor in a place where there is adequate ventilation.
· Avoid placing the monitor against a bright background or where sunlight or other light sources may refl ect on the area of the monitor. Place the monitor just below eye level.
· Handle with care when transporting the monitor.
· Refrain from giving the shock or scratch to the screen, as screen is fragile.
CLEANING YOUR MONITOR
No specifi c liquid or chemical necessary when cleaning this LCD monitor However, we suggest to clean the monitor with non-abrasive cloths and cleaning solutions used in hospitals to clean similar equipment. We recommend using 70% Isopropyl alcohol for the screen surface and warm water and a mild detergent for all other surfaces. Other acceptable cleaning agents are listed below:
ENGLISH
• 70% isopropyl alcohol
• Cidex (2.4% glutaraldehyde solution)
• 0.5% Chlorhexidine in 70% isopropyl alcohol To clean the screen, do not spray liquid cleaners directly on to the unit. Stand away
Form the monitor and spray cleaning solution onto a cloth. Without applying exces­sive pressure, clean the screen with the slightly dampened rag.
Recomendation to Use more than One Unit
As unforeseen situation can occationally occur for the monitor, when the display is used for safety control of human, equipments or unstable picture, or for emagencies, we strongly recommend you use more than one unit or prepare a spare unit.
10 User’s Guide
POWER MANAGEMENT FUNCTION
The monitor is equipped with the power management function which automatically reduce the power consumption when not in use in three power level modes.
Stand-by Mode
The monitor goes into stand-by mode when the horizontal sync signal is off for about 10 seconds. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The screen is displayed after the horizontal sync signal is restored.
Suspend Mode
The monitor goes into suspend mode when the vertical sync signal is off for about 10 seconds. The power consumption during this is W. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The screen is displayed after the vertical sync signal is restored.
Off Mode
The monitor goes into power-off mode when the vertical and horizontal sync signals are off for about 10 seconds. In this mode, the screen goes off and the power LED blinks for 1 seconds On and 1 second Off. The screen is displayed after the vertical and horizontal sync signals are restored.
Power Management System
The AMM190WTD Medical Monitor power management proposal. Provides four phases of power-saving modes by detecting the horizontal sync signal as shown in the table below.
State
Horizontal Sync Active Inactive Active Inactive
Vertical Sync Active Active Inactive Inactive
Video Active Blanked Blanked Blanked
Power Indicator Green
Power Consumption
Normal Operation
75W 27W 27W 27W
DPMS Standby
Green Flashing (1 sec. Interval)
DPMS Suspend
Green Flashing (1 sec. Interval)
DPMS Off
Green Flashing (1 sec. Interval)
When the monitor is power saving mode or detects an incorrect timing, the screen will be blank and power LED indicator will blink.