Thank you for choosing this CORONIS FUSION 4MP/6MP display!
Coronis Fusion is an ingenious PACS display system designed to further
enhance flexibility and productivity in diagnostic imaging.
Featuring the industry’s first 30-inch color LCD that can be used as two
seamless heads or one wide-screen display, Coronis Fusion offers you
the freedom to organize your workspace just the way you want it. It
allows you to read CT, MR, cath and echo cardiogram images, or any
other combination, side by side on a single diagnostic screen.
Coronis Fusion features the latest, breakthrough In-Plane Switching (IPSPro) LCD technology. This technology brings the most advanced LCD
viewing characteristics on the market, setting new standards for
brightness and contrast, even from a wide viewing angle.
Use the instructions in this guide to install your C
ORONIS FUSION 4MP/6MP
display and discover all these interesting features yourself!
Following symbols may be used throughout this guide:
Warning: Risk of
injury to human
beings
Important notice
Caution: Risk of
damage to the
product
Note
or remark
Hint, tipAdditional
information
Read all the safety information starting on page 51 before
operating your C
If you ordered a Barco display controller, it’s also in the box together
with its accessories. A dedicated user guide is available on the system
CD.
ORONIS FUSION 4MP/6MP display guide
Keep your original packaging. It is designed for this display and
is the ideal protection during transport.
Coronis Fusion 4MP/6MP7
Parts, controls and connectors
182 3 4576
1
2
3
4
5
6
7
8
Front view
USB downstream connectorsLeft/Down key
Right/Up keyEnter key
Stand-by keyAmbient light sensor
Power status indicator lightControl wheel
8Coronis Fusion 4MP/6MP
Rear view
132
1
2
3
Connector compartment
cover
Display stand cover
Neck cover
Coronis Fusion 4MP/6MP9
CORONIS FUSION 4MP
6
1
2
3
4
5
7
1
2
3
4
567
+24 VDC power inputUSB upstream connector
USB downstream connector+5 VDC, 0.25 A power output
for accessories
+3.3 VDC, 0.25A auxiliary
power output
DVI 2 video input
DVI 1 video input
10 Coronis Fusion 4MP/6MP
CORONIS FUSION 6MP
6
1
2
3
4
5
7
1
2
3
4
567
+24 VDC power inputUSB upstream connector
USB downstream connector+5 VDC, 0.25 A power output
for accessories
+3.3 VDC, 0.25A auxiliary
DVI 1 video input
power output
DVI 2 video input
Coronis Fusion 4MP/6MP 11
Display installation
Prior to installing your CORONIS FUSION 4MP/6MP display and
connecting all necessary cables, make sure to have a suitable
display controller physically installed in your computer. If you are
using a Barco display controller, please consult the dedicated user
guide available on the system CD.
Removing the covers
CONNECTORCOMPARTMENTCOVER
The connector compartment cover should be removed to get access to
the connectors. To remove the cover, proceed as follows:
1 Gently lift the clips at one of the handles of the connector
compartment cover to release that side of the cover.
2 Do the same at the other side of the cover.
3 Remove the cover.
12 Coronis Fusion 4MP/6MP
DISPLAYSTANDCOVER
To get access to the cable routing channel, the display stand cover
should be removed. This can be done as follows:
1 Gently press and hold the clips at the inside top of the cover.
2 Slide the cover upwards while holding the clips pressed.
Coronis Fusion 4MP/6MP 13
Unlocking the height mechanism
In the factory, the height-positioning system in the display stand is
locked with a red hook to prevent damage during transportation. You’ll
have to remove this hook before adjusting your display height position.
1 While holding the display panel pushed down, pull out the red
hook in the display stand.
To retain the hook for possible future transportation, insert the
short, red end of the hook back into the stand of your display.
14 Coronis Fusion 4MP/6MP
Adjusting the display position
Now that the height-positioning system of your display is unlocked, you
can tilt, swivel, raise and lower the display as desired.
Do not try to pivot your display when attached to the stand.
Trying to do so could cause serious damage to your display and
its stand.
Coronis Fusion 4MP/6MP 15
Connecting the signal cables
1 Connect one head of the display controller to the DVI 1 connector
with one of the supplied DVI cables.
2 Connect another head of the display controller to the DVI 2
connector with one of the supplied DVI cables.
3 If you want to make use of your display’s USB downstream
connectors, connect a PC USB downstream connector to the
display’s USB upstream connector by means of the supplied USB
2.0 cable.
16 Coronis Fusion 4MP/6MP
Connecting the power cables
1 Connect the supplied external DC power supply to the +24 VDC
power input of your C
2 Plug the other end of the external DC power supply into a
grounded power outlet by means of the proper power cord
delivered in the packaging.
ORONIS FUSION 4MP/6MP display.
Routing the cables
Route all connected cables trough the cable routing channel in the stand
of your display.
The cable straps at the inside of the connector compartment
allow you to fix the cables for better shielding of the cables.
Coronis Fusion 4MP/6MP 17
Re-attaching the covers
1 Re-attach the display stand cover by sliding it downward. You’ll
hear a “click” sound of the cover’s clips when the display stand
cover is in position.
Pay attention that all cables stay in the cable channel
while replacing the cover.
2 Re-attach the connector compartment cover by sliding the cover’s
top in position and then push the cover’s bottom. You’ll hear a
“click” sound of the cover’s clips when the connector compartment
cover is in position.
18 Coronis Fusion 4MP/6MP
VESA-mount installation
The panel, standard attached to a stand, is compatible with the VESA
100 mm standard. Thus, it can be used with an arm that is approved by
VESA. This chapter shows you how to release the panel from the stand
and how to attach it to an arm. If you’re not using an arm, you can skip
this chapter.
1 Gently pull open both lower sides of the cover.
2 Slide the cover upwards while holding the lower sides pulled open.
Coronis Fusion 4MP/6MP 19
3 Lift the plastic frame that covers the fixation of the panel to the
stand. Turn it for 45 degrees to uncover the fixation screws.
4 Unscrew the four fixation screws fixing the panel to the stand.
5 Attach the panel firmly to the arm using 4 screws M4 x 10 mm.
You should mount the panel in landscape position.
Portrait position is possible but not supported.
Use an arm that can support a weight of at least 21.6 kg
(47.62 lbs). Failure to do so could make the panel fall,
causing serious injury to a child or adult, and serious
damage to the equipment.
Never move a display attached to an arm by pulling or
pushing the display itself. Instead, make sure that the
arm is equipped with a VESA approved handle and use
this to move the display.
Please refer to the instruction manual of the arm for more
information and instructions.
20 Coronis Fusion 4MP/6MP
First time starting up
You are now ready to start up your CORONIS FUSION 4MP/6MP display
system for the first time.
1 Switch on your C
“Stand-by switching” on page 24.
2 Turn on the computer connected to your display.
If you have properly installed your display and display controller, the
Windows start-up messages will appear once the boot procedure is
finished.
Your CORONIS FUSION 4MP/6MP display will be running in a basic
video mode at a default refresh rate when first time starting
up.
If you are using a Barco display controller, please consult the
dedicated user guide available on the system CD to install the
drivers, software and documentation. When this is done, your
display will automatically detect the connected video input
signal(s) and apply the correct video mode and refresh rate.
ORONIS FUSION 4MP/6MP display as described in
Coronis Fusion 4MP/6MP 21
Daily operation
Recommendations for daily operation
OPTIMIZETHELIFETIMEOFYOURDISPLAY
Enabling the Display Power Management System (DPMS) of your display
will optimize its diagnostic lifetime by automatically switching off the
backlight when the display is not used for a specified period of time. By
default, DPMS is enabled on your display, but it also needs to be
activated on your workstation. To do this, go to “Power Options
Properties” in the “Control Panel”.
Barco recommends setting DPMS activation after 20 minutes of
non-usage.
USEASCREENSAVERTOAVOIDIMAGERETENTION
Prolonged operation of an LCD with the same content on the same
screen area may result in a form of image retention.
You can avoid or significantly reduce the occurrence of this phenomenon
by using a screen saver. You can activate a screen saver in the “Display
properties” window of your workstation.
Barco recommends setting screen saver activation after 5
minutes of non-usage. A good screen saver displays moving
content.
In case you are working with the same image or an application with
static image elements for several hours continuously (so that the screen
saver is not activated), change the image content regularly to avoid
image retention of the static elements.
UNDERSTANDPIXELTECHNOLOGY
LCD displays use technology based on pixels. As a normal tolerance in
the manufacturing of the LCD, a limited number of these pixels may
remain either dark or permanently lit, without affecting the diagnostic
22 Coronis Fusion 4MP/6MP
performance of the product. To ensure optimal product quality, Barco
applies strict selection criteria for its LCD panels.
To learn more about LCD technology and missing pixels, consult
the dedicated white papers available at www.barco.com/
medical.
ENHANCEUSERCOMFORT
Every Barco multi-head display system is color matched with the highest
specification in the market.
Barco recommends keeping color-matched displays together.
Furthermore, it is important to use all displays of a multi-head
configuration at the same rate to preserve color matching
throughout the economic lifetime of the system.
MAXIMIZEQUALITYASSURANCE
The ‘MediCal QAWeb’ system offers online service for high-grade Quality
Assurance, providing maximum diagnostic confidence and uptime.
Barco recommends to install MediCal QAWeb Agent and apply
the default QAWeb policy at least. This policy includes
calibration on regular intervals. Upgrading to MediCal QAWeb
Server offers even more possibilities.
Learn more and sign up for the free MediCal QAWeb Essential
level at www.barco.com/medical/qa
Coronis Fusion 4MP/6MP 23
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.