This manual contains information concerning the installation, operation and maintenance of the Badger Meter®
M-Series M7600 electromagnetic flow meter. Read and understand the instructions given in this manual. Retain this
manual in a readily accessible location for future reference.
SAFETY PRECAUTIONS AND INSTRUCTIONS
Some procedures in this manual require special safety considerations. In such cases, the text is emphasized with the
following symbols:
SymbolExplanation
Warning indicates the potential for severe personal injury, death or substantial property damage.
Comply with the instructions and proceed with care.
Caution indicates the potential for minor personal injury or property damage. Comply with the
instructions and proceed with care.
SYSTEM DESCRIPTION
The Badger Meter M7600 electromagnetic flow meter successfully combines the most advanced electromagnetic
flow metering technology with the simplicity and ruggedness of the Badger Meter proven Batching Systems for
Industrial Applications.
The M7600 meter is intended for fluid metering in most industries including water, wastewater, food and beverage,
pharmaceutical, chemical and concrete.
The basic components of an electromagnetic flow meter are:
• The detector, which includes the flow tube, isolating liner and measuring electrodes.
• The amplifier, which is the electronic device responsible for the signal processing, flow calculation, display and
output signals.
The construction materials of the wetted parts (liner and electrodes) should be appropriate for the specifications on the
intended type of service. We recommend that you review all of the compatibilities consistent with the specifications.
Each meter is factory tested and calibrated. A calibration certificate is included with each meter.
Amplifier
®
Figure 1: Amplifier and Detector
Page 5 February 2018MAG-UM-00507-EN-02
Unpacking and Inspection
UNPACKING AND INSPECTION
Follow these guidelines when unpacking the M-Series equipment.
• If a shipping container shows any sign of damage, have the shipper present when you unpack the meter.
• Follow all unpacking, lifting and moving instructions associated with the shipping container.
• Open the container and remove all packing materials. Store the shipping container and packing materials in the event the
unit needs to be shipped for service.
• Verify that the shipment matches the packing list and your order form.
• Inspect the meter for any signs of shipping damage, scratches, or loose or broken parts.
OTE:NIf the unit was damaged in transit, it is your responsibility to request an inspection report from the carrier within 48
hours. You must then file a claim with the carrier and contact Badger Meter for appropriate repairs or replacement.
• All detectors are shipped with a liner protector on each end to maintain proper form of the polytetrafluoroethylene (PTFE)
material during shipping and storage.
OTE:NDo not remove the liner protectors until you are ready to install.
• Storage: If the meter is to be stored, place it in its original container in a dry, sheltered location. Storage temperature ranges
are: – 22…158° F (– 30…70° C).
Rigging, Lifting and Moving Large Units
WHEN RIGGING, LIFTING OR MOVING LARGE UNITS, FOLLOW THESE GUIDELINES:
• DO NOT lift or move a meter by its amplifier, junction box, detector neck, or cables.
• Use a crane rigged with soft straps to lift and move meters with flow tubes that are between two inches and four inches
(50 mm and 100 mm). Place the straps around the detector body, between the flanges, on each side of the detector.
®
Place straps between flanges.
Figure 2: Rigging Large UnitsFigure 3: Sling-Rigged Lifting Methods
• Use the sling-rigged method to lift large detectors into a vertical position while they are still crated. Use this method to
position while they are still crated. Use this method to position large detectors vertically into pipelines.
• Do not lift a detector with a forklift by positioning the detector body on the forks, with the flanges extending beyond the
lift. This could dent the housing or damage the internal coil assemblies.
• Never place forklift forks, rigging chains, straps, slings, hooks or other lifting devices inside or through the detector's flow
tube to hoist the unit. This could damage the isolating liner.
Do not lift detector with forklift.Do not lift or rig lifting devices through detector.
Page 6 February 2018MAG-UM-00507-EN-02
Figure 4: Lifting and Rigging Cautions
Meter Location, Orientation and Applications
METER LOCATION, ORIENTATION AND APPLICATIONS
Meter Mount Conguration
The meter mount configuration has the amplifier mounted directly on the detector. This compact, self-contained
configuration minimizes installation wiring.
Temperature Ranges
TO PREVENT DAMAGE TO THE METER, STRICTLY OBSERVE THE AMPLIFIER’S AND DETECTOR’S MAXIMUM
TEMPERATURE RANGES.
In regions with extremely high ambient temperatures, protect the detector.
AmplifierAmbient temperature– 4…140° F (–20…60° C)
Detector (PTFE)Fluid temperature– 40…212° F (– 40…100° C)
Pipelines and Fluid Flow
Take the following precautions during installation:
• Do not install the meter on pipes with extreme pipe vibrations. If pipes are vibrating, secure the piping with appropriate
pipe supports in front of and behind the meter.
• Do not install the meter close to pipeline valves, fittings or impediments that can cause flow disturbances.
• Do not install the meter on suction sides of pumps.
• Do not install the meter on outlet sides of piston or diaphragm pumps. Pulsating flow can affect meter performance.
• Avoid installing the meter near equipment that produces electrical interference such as electric motors, transformers, variable
frequency, and power cables.
• Verify that both ends of the signal cables are securely fastened.
• Place power cables and signal cables in separate conduits.
• Place the meter where there is enough access for installation and maintenance tasks.
Page 7 February 2018MAG-UM-00507-EN-02
RIGHT
Electrode
WRONG
BUTTERFL
VE
FLOWMETER
Meter Location, Orientation and Applications
Meter Orientation
Mag meters can operate accurately in any pipeline orientation and can measure volumetric flow in forward and reverse
directions.
OTE:NA "Forward Flow" direction arrow is printed on the detector label.
Vertical Placement
Mag meters perform best when placed vertically, with liquid flowing upward and meter electrodes in a closed, full pipe.
Vertical placement allows the pipe to remain completely full, even in low flow, low pressure applications, and it prevents
solids build-up, sediment deposit and accumulation on the liner and electrodes.
Figure 5: Vertical placement
Horizontal Placement
The M7600 meters are equipped with an Empty Pipe Detection feature. If an electrode mounted in the pipe is not covered by
fluid for five seconds, the meter will display an Empty Pipe Detection condition. The meter will send out an error message
and stop measuring flow. When the electrode is again covered with fluid, the error message disappears and the meter will
begin measuring.
Plane
Electrode
Plane
Figure 6: Horizontal placement
When installing the meter on a horizontal pipe, mount the detector to the pipe with the flow-measuring electrode axis in a
horizontal plane (three and nine o’clock). This placement helps prevent solids build-up, sediment deposit and accumulation
on the electrodes.
Straight Pipe Requirements
Sufficient straight-pipe runs are required at the detector inlet and outlet for optimum meter accuracy and performance. An
equivalent of three diameters of straight pipe is required on the inlet (upstream) side. Two diameters are required on the outlet
(downstream) side.
D (Pipe Size) D (Pipe Size)
FORWARD FLOW
CHECK VALVE
GLOBE VALVE
Y VALVE
PUMP
Page 8 February 2018MAG-UM-00507-EN-02
ELBOW
TEE
GATE VALVE
(FULLY OPEN)
MINIMUM STRAIGHT PIPE
3 x D
MINIMUM STRAIGHT PIPE
7 x D
MINIMUM PIPING REQUIREMENT
STANDARD CONCENTRIC
REDUCERS
(NO DISTANCE REQUIRED)
Figure 7: Straight pipe requirements
2 x D
MINIMUM STRAIGHT PIPE
ELBOW
TEE
ANY VAL
Meter Location, Orientation and Applications
Pipe Reducer Requirements
With pipe reducers, a smaller meter can be mounted in larger pipelines. This arrangement may increase low-flow accuracy.
There are no special requirements for standard, concentric, pipe reducers.
Custom fabricated pipe reducers must have an approximate slope angle of 15 degrees to minimize flow disturbances and
excessive loss of head. If this is not possible, install the custom pipe reducers as if they were fittings and install the required
amount of straight pipe
Figure 8: Pipe reducer requirements
Chemical Injection Applications
For water line applications with a chemical injection point, install the meter upstream of the injection point. This eliminates
any meter performance issues.
Figure 9: Chemical injection point upstream of meter
If a meter must be installed downstream of a chemical injection connection, the distance between the meter and the
injection point should be between 50 and 100 feet (15 and 30 meters). The distance must be long enough to allow the water
or chemical solution to reach the meter in a complete, homogeneous mixture.
If the injection point is too close, the meter senses the two different conductivities for each liquid. This will likely result in
inaccurate measurements. The injection method—spaced bursts, continuous stream of drips or liquid or gas—can also affect
downstream readings by the meter.
Page 9 February 2018MAG-UM-00507-EN-02
WRONG
RIGHT
Meter Location, Orientation and Applications
Partially-Filled Pipe Situations
In some locations, the process pipe may be momentarily only partially filled. Examples include: lack of back pressure,
insufficient line pressure and gravity flow applications.
To eliminate these situations:
• Do not install the meter at the highest point of the pipeline.
• Do not install the meter in a vertical, downward flow section of pipe.
• Always position the ON/OFF valves on the downstream side of the meter.
RIGHT
FLOW
Figure 10: Incorrect meter placement
WRONG
FLOW
FLOW
Do not install in a vertical, downward position.Position "On/Off" valves on downstream side.
Figure 11: Position valves on downstream side
FLOW
To minimize the possibility of partially-full pipe flows in horizontal, gravity or low pressure applications, create a pipe
arrangement that ensures the detector remains full of liquid at all times.
Figure 12: Pipe positioned to keep water in detector
Page 10 February 2018MAG-UM-00507-EN-02
GASKETS RECOMMENDED
Meter Gaskets and Grounding
METER GASKETS AND GROUNDING
Gasket and grounding requirements must be considered when determining the meter location, orientation and application.
Grounding rings are provided with the M7600 meter.
Meter/Pipeline Connection Gaskets
You must install gaskets (provided) between the detector's isolating liner and the pipeline flange to ensure a proper
and secure hydraulic seal. Use gaskets that are compatible with the fluid. Center each gasket on the flange to avoid flow
restrictions or turbulence in the line.
During installation, do not use graphite or any electrically conductive sealing compound to hold the gaskets. This could
compromise the accuracy of the measuring signal.
Figure 13: Meter/pipeline connection gaskets
Meter Grounding
Process pipeline material can be either electrically conductive (metal) or not electrically conductive (made of or lined with
PVC, fiberglass or concrete).
It is essential that the mag meter amplifier’s input ground (zero voltage reference) be electrically connected to the liquid media and
to a good, solid earth ground reference.
Conductive Pipe Grounding
To achieve an adequate ground, the meter body MUST be electrically connected to the liquid media. The mag meter flanges
are provided with grounding bolts for this purpose.
If the pipe material is electrically conductive, simply install grounding straps between these grounding bolts and the
mating flanges.
To ensure a good electrical connection at the mating flanges, we recommend that you drill and tap the flanges and install a
grounding screw (not provided).
These grounding straps must be copper wire, at least 12 AWG size. They must be connected on both sides (inlet and outlet) of
the detector and to a local, earth ground.
Pipelines with Cathodic Protection
As for pipelines with cathodic protection, install meter potential-free. No electric connection from the meter to the pipeline
system may exist and power supply is to be provided via isolating transformer.
GROUNDING RINGS ALSO NEED TO BE INSTALLED ISOLATED FROM THE PIPELINE SYSTEM.
Observe national rules for potential-free installations.
Electrically isolated
"X"
6 mm² Cu
Figure 14: Cathodic protection
Electrically isolated
M4:1
"X"
Page 11 February 2018MAG-UM-00507-EN-02
Power Connections
POWER CONNECTIONS
Wiring Safety
AT INSTALLATION, BE SURE TO COMPLY WITH THE FOLLOWING REQUIREMENTS:
• Disconnect power to the unit before attempting any connection or service to the unit.
• Do not bundle or route signal lines with power lines.
• Keep all lines as short as possible.
• Use twisted pair shielded wire for all output wiring.
• Observe all applicable local electrical codes.
Opening the Cover
The M7600 amplifier's design lets you open the cover
without completely removing it.
Follow these steps:
1. Completely remove the top two screws from the
amplier using a blade/slotted screwdriver.
2. Loosen both of the bottom screws so that the round
head of each screw clears the top face of the cover.
3. Pull down the cover to the open position.
• For the 2 x M20 cable inlets, use only flexible electric
cables.
• Use separate cable inlets to separate power from signal
and input/output cables.
Figure 15: Remove two screws
Figure 16: Open the cover
Page 12 February 2018MAG-UM-00507-EN-02
Power Connections
Power Supply Connections
External Disconnect
• Install an external disconnect switch or circuit breaker that meets local standards.
• Position the M7600 meter in an accessible location.
• Position and identify the disconnect device so as to provide safe and easy operation.
• Label the disconnect device as being for the mag meter.
AC Power Wiring
For the AC power connections, use three wire-sheathed cable with an overall cable diameter of 0.2…0.45 inch (5…12 mm).
For signal output, use 18…22 gauge (0.25…0.75 mm2) shielded wire. Overall cable diameter between 0.12…0.35 inch
(3…9 mm).
TO PREVENT ACCIDENTS, CONNECT MAIN POWER ONLY AFTER ALL OTHER WIRING HAS BEEN COMPLETED.
The amplifier is a microprocessor device. It is important that the power supply be as “clean” as possible. Avoid using power
lines that feed heavy loads: pumps, motors, etc. If dedicated lines are not available, a filtering or isolation system may
be required.
Take national applicable rules into account.
• Observe type plate (mains voltage and frequency)
• Equipment shall be provided with a external means for disconnecting it from each operating energy supply source. The
disconnecting means shall disconnect all current-carrying conductors.
1. Open the cover (see "Opening the Cover" on page 12).
2. Push the power cable through the upper cable inlet.
Pulse output, open collector (1, 2)
Alarm output (3, 4)
1 and 2
3 and 4
• USE SEPARATE CABLE INLETS FOR CABLES CONNECTED TO THE SOLID-STATE RELAY OUTPUT AND CABLES CONNECTED
TO THE OTHER INPUT/OUTPUTS.
• IN MULTIPHASE NETWORKS, SOLID-STATE RELAY SHOULD HANDLE ONLY THE SAME PHASE THAT IS USED FOR
POWERING THEMETER.
Connecting the M7600 Meter to 110V AC from Batch Control Panel Power Supply
Typical concrete batch panel.
Meter
Batch Panel
Ground
Neutral
Hot
Meter Pulse Input
Solid-State
Relay Output
Limit 230V AC
*
Neutral
Hot
S1
S2
Figure 19: Batch panel power connections
• The connection shown in Figure 19 is for batch panels that require a 115V AC hot pulse for meter signals.
• Consult the batch panel manufacturer to confirm the required pulse signal.
• For a 115V AC neutral pulse signal to the batch panel, take the S2 jumper to the 115 neutral power supply.
DO NOT PASS MORE THAN 230V AC THROUGH THE SOLIDSTATE RELAY.
Page 14 February 2018MAG-UM-00507-EN-02
Common
Power Connections
Wiring to a PC-200 Controller
To connect the open collector scaled pulse output from the Model M7600 meter to the PC-200 controller, follow Figure 20.
Open Collector
Transistor Output
1
DC-Switched
2
3
4
5
6
Figure 20: Wiring to a PC-200 controller
Wiring to an ER-10 Industrial Register
Common
Open Collector
Transistor Output
TerminalFunctionOperation
1Ground—
2Input B
Count input
3—Not used
4ResetConnect to ground to reset totalizer. This is a maintained or level-sensitive reset.
5Program enableConnect to ground to enter program mode.
6Backlight common—
7Backlight powerConnect to power to light display.
1
DC-Switched
2
3
4
5
6
7 +10-30VDC7 +10-30VDC
6 DC Common
5 Enable/R
5 Enable/R
Figure 21: Wiring to an ER-10 register
Count input
Contact closure of NPN 100 Hz max
Table 2: ER-10 wiring terminals
Rear View
RST 4
IN B 2
GND 1
Reset
Count Input
GroundProgram enable
Page 15 February 2018MAG-UM-00507-EN-02
Menu Programming Options
MENU PROGRAMMING OPTIONS
Screen Layout
The following M7600 meter programming options are available from the M7600 Menu:
• Scale factor
• Pulse/unit
• Flow unit
• Totalizer unit
Function Buttons
All M7600 programming is accomplished using the three function buttons located on the front of the amplifier. Screen
navigation, digit, and parameter selection is performed by pressing a combination of these three buttons.
Display
Push-buttons
Figure 22: Function buttons
Exit
Save
Press the Exit/Save button to access the M7600 programming options.
Press the left button to scroll through the programming options.
Press the center button to edit a programmable option.
For selection list options (for instance, Totalizer Unit or Flow Unit), press the left button to change the value. Press the right
button to confirm the new value.
For numerical options (for instance, Scale Factor or Pulse/Unit), press the left button to change the numerical digit value.
Press the center button to move to the next numerical digit. When all digits are entered, press the right button to confirm
the new value.
Status Icons
Communication interface is activated
Meter is unlocked
Error message
Empty pipe detection
Page 16 February 2018MAG-UM-00507-EN-02
Menu Programming Options
M7600 Menu
Scale FactorChanging the scale factor lets you adjust the meter’s accuracy without disturbing parameters set
by the factory. You can tune the meter to meet changing application requirements in a range of
±10% (0.90 to 1.10).
If it is necessary to recalibrate the meter, follow these steps:
1. Determine the exact quantity of uid that actually passed through the meter by using a
calibrated volumetric container or by weighing the container.
2. Note the volume of uid indicated by meter.
3. Note the current scale factor on the meter LCD (scale factor menu).
4. Use the following formula to calculate the new scale factor for recalibration:
Qty delivered
Qty on meter
X
Old scale
factor
New scale
factor
5. Use the meter push-buttons to enter the new scale factor.
6. Rerun test to verify that the recalibration is correct.
Example:
You have a 3" size meter in your installation. You run a test batch quantity of 210 gallons. The
totalizer on the meter indicates 203 gallons. Your present scale factor on the meter is 1.00.
Using the formula:
210
1.0=1.03
X
203
Your new scale factor is 1.03.
Pulse/UnitThe Pulses/Unit parameter lets you set how many pulses per unit of measure will be transmitted.
The maximum output frequency of 10,000 pulses/sec.
10 kHz must not be exceeded for the Open Collector.
1 Hz must not be exceeded for the Solid-State Relay.
If both are used, 1 Hz must not be exceeded.
For example, assuming the unit of measure is gallons:
• Setting the Pulses/Unit to 1 will transmit 1 pulse every gallon
• Setting the Pulses/Unit to 0.01 will transmit 1 pulse every 100 gallons
You must configure pulses/unit if the function of the selected output is to be open collector or
solid-state.
Flow UnitFlow Units let you select among the Flow Units listed below. Flow units are automatically
Totalizer UnitThis parameter establishes the units of measure for the totalizers.
DisplayTotalizer UnitDisplayTotalizer Unit
LLitersMGMillion Gallons
hLHectoliterIGImperial Gallons
3
m
3
ft
galU.S. GallonsAftAcre Foot
Cubic MetersbblBarrel
Cubic FeetOzFluid Ounces
Page 17 February 2018MAG-UM-00507-EN-02
Maintenance
MAINTENANCE
Mandatory, routine or scheduled maintenance should not be required for the M7600 Mag Meter electronics or flow tube after
proper installation. However, some occurrences may require personnel to perform the following:
• Flow tube and electrode cleaning
• Circuit board replacement
DO NOT CLEAN COMPONENTS INSIDE THE AMPLIFIER OR JUNCTION BOX.
Cleaning the Flow Tube and Electrode
At times flow tube, electrodes, amplifier/junction box housings and the amplifier window may need periodic cleaning,
depending on process fluid properties, fluid flow rate and surrounding environment.
Clean the flow tube and electrodes by following the material handling and cleaning procedures documented in the
Material Safety Data Sheet (MSDS) guidelines for the products(s) that were in contact with the flow tube and electrodes.
Should flow tube and/or electrode cleaning become necessary:
1. Disconnect detector from pipeline.
2. Clean electrodes according to MSDS guidelines.
3. Reconnect detector to pipeline.
Page 18 February 2018MAG-UM-00507-EN-02
Troubleshooting
TROUBLESHOOTING
The M7600 mag meter is designed for many years of optimal performance. However, should it malfunction, there are certain
things that we recommend you check before contacting our Technical Support department or your local Badger Meter
Representative.
Errors & Warnings
OTE:NThe M7600 display flashes whenever an error is detected.
DescriptionPossible CauseRecommended Action
Coil disconnected • Meter is not connected.
• Connection to meter is interrupted.
Coil shortedCoil cables shorted.Check coil cables.
Empty pipePipe may not be full.Make sure that pipe is always filled at the
Medium with low conductivity.Calibrate meter, see "M7600 Menu" on page 17.
Cable is broken or disconnected.Check the cable for the empty pipe signal.
RangeActual flow rate is exceeding the programmed full
±0.5% of rate for velocities greater than 1.64 ft/s, ±0.008 ft/s less than 1.64 ft/s
Accuracy
Repeatability ±0.1%
Fluid Temperature Maximum fluid temperature: 212° F (100° C)
Storage Temperature–40…140° F (–40…60° C)
Ambient Temperature –4…140° F (–20…60° C)
Fluid ConductivityMinimum liquid conductivity: 5 micromhos/cm
Flow Direction Unidirectional
Pressure Limits Working pressure: 232 psi (16 bar)
Cable Entries Two 1/2 in. NPT cord grip
Grounding RingsStandard (pre-installed)
Optional Grounding Electrodes2
Inputs
Power Supply 92…275 VAC
Coil ExcitationPulsed DC
Programming PC user interface
Outputs
Digital Output
Frequency Output Maximum output frequency: 10 kHz
Pulse Width 50% duty cycle
Approvals
Meter Enclosure NEMA 4
NIST Handbook B44Complies for batching applications
±0.5% of rate for velocities greater than 0.50 m/s, ±2.5 mm/s less than 0.50 m/s
Note: The maximum measuring error depends on the installation conditions.
Powder-coated cast aluminum, NEMA 4
Meter mounted only
Output 1: solid-state relay up to 230V, 500 mA
Output 2: Opto-isolated open collector, 50 mA at 24V DC