ORDERING INFORMATION
PART NO. Description
28-8002 Tru Poi nte®
1100Ultrasonic Leak
DetectorKit
28 -8012 Tru Po inte®
1100Ultrasonic
Leak DetectorKit w/
SoundBlaster
28-8003 Tru Pointe®
2100Ultrasonic Leak
DetectorKit
28 -8 013 Tru Pointe®
2100Ultrasonic
Leak DetectorKit w/
SoundBlaster
28-8000 Tru Pointe® Ultra
LeakDetector Kit w/
foldingheadset
28-8010 Tru Pointe® Ultra
LeakDetector Kit w/
foldingheadset &
SoundBlaster
28-8001 Tru Pointe® Ultra
HDLeak Detector Kit w/
stereo headphones
28 -8 011 Tr u Poin te® Ultra
HDLeak Detector Kit
w/stereo headphones
&SoundBlaster®
®
®
®
FEATURES BENEFITS
Ultrasonic to detec t any gas
leaks
Sensitivity Adjustments Unaec ted by other gases, even at high-concentrations or
Visual indication Ten-step LED Bar Graphs shows signal strength.
Unique Volume control True sound reproduction ensures fast recognition of leaks.
Pocket Sized Easy to carry
Dual Audio Intensity Displays Easily see bar graph or alpha/numeric displays
Capable of detecting a 5 PSI leak out of a 0.005” hole, 30 to
40 feet away. Ideal for stream trap testing and bearing wear
detection. Unaected by wind or temperature uctuations.
Find a leak as small as 2.5-3 oz./yr (84 g/yr)
sounds in the test area.
mybacharach.com
PRODUCT ATTRIBUTES DESCRIPTION
MECHANICAL
Dimensions 8.185” (20.78cm) X 2.4” (61mm) X .875” (22.22mm)
Weight 0.5lb (233g) Includes Batter y
Body Materials Durable ABS & Anodized Aluminum
Audio Connector 3.5mm Stereo
ELECTRICAL
Airborne Sensor Sensitivity 80db/ V-μbar
Ultrasound Converter Type Digitally Controlled Heterodyne
Frequency response, Airborne Tru Pointe® 1100: 34kHz-42kHz Tru Pointe® 2100: AirHi: 34kHz to 42kHz
Frequency response, Airborne, Low Tru Pointe® 2100: AirLow: 26kHz to 34kHz
Frequency response, Touch Probe Tru Poin te® 1100 6kHz-24kHz Tru Pointe® 2100: High, PbHi: 34kHz to 42kHz
Tru Pointe® 2100: Low, PbLo 16kHz to 24kHz
Tru Pointe® 2100: Sonic, PbS: 0 to 10kHz
Controls 4 Buttons (with AudioZoomTM)
Output, Audio 0Hz to 4kHz
Output, Visual 20 Element Bar Graph with Peak-Hold and Separate Alphanumeric Digital Display
Batte ry Type 9 Volt
Run Time 55-80 hrs
PRODUCT ATTRIBUTES DESCRIPTION
MECHANICAL
Dimensions 5.5” (14cm) X 2.4” (61mm) X .875 (22.22mm)
Weight 0.3lb (138g) Includes Battery
Body Materials Durable ABS
Audio Connector 3.5mm Stereo
ELECTRICAL
Airborne Sensor Sensitivity 80db/ V-μbar
Ultrasound Converter Type Analog Controlled Heterodyne
Frequency response, Airborne 34kHz to 42kHz
Heterodyne Filter 4kHz
Heterodyne Oscillator Analog
Controls 1 Button
Sensitivity Control Continually Variable Slide
Output, Audio 0Hz to 4kHz
Output, Visual 10 Segment Bar Graph
Batte ry Type 9 Volt
Run Time 80 -120 hrs