Product Leadership • Training • Service • Reliability
Page 2
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in material and
manufacture and will conform substantially to Bacharach Inc.'s applicable specifications. Bacharach's liability and Buyer's
remedy under this warranty are limited to the repair or replacement, at Bacharach's option, of this Product or parts thereof
returned to Seller at the factory of manufacture and shown to Bacharach Inc.'s reasonable satisfaction to have been
defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within two (2) years
after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach's liability and Buyer's remedy
under this warranty of title are limited to the removal of any title defects or, at the election of Bacharach, to the replacement of
this Product or parts thereof that are defective in title.
The warranty set forth in paragraph 1 does not apply to parts the Operating Instructions designate as having a limited shelflife or as being expended in normal use (e.g., filters).
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGATION, LIABILITY, RIGHT,
CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARISING FROM BACHARACH'S NEGLIGENCE,
ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all
other remedies including, without limitation incidental or consequential damages. No agreement varying or extending the
foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly
authorized officer of Bacharach.
Register Your Warranty by Visiting
www.mybacharach.com
Notice:
Product improvements and enhancements are continuous; therefore the specifications and
information contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language
without the prior written consent of Bacharach, Inc.
BACHARACH® is a registered trademark of Bacharach, Inc. All other trademarks, trade names,
service marks and logos referenced herein belong to their respective owners.
A
Manufacturer’s name: Bacharach, Inc.
Manufacturer’s address: 621 Hunt Valley Circle
Product Name: CO2MSZ (Carbon Dioxide Gas Monitor Single Zone)
Declaration of Conformity
New Kensington, PA 15068
conforms to the following harmonized European EMC
standards:
Thank you for investing in a
This manual provides important information on how to install, operate, and service the CO2MSZ
monitoring unit.
To assure operator safety and the proper use of the monitor, please read, understand, and follow the
contents of this manual.
If you have a working knowledge of gas monitors, you will find this manual useful as a reference tool. If
you are new to the use of gas monitors, you can educate yourself about the principles of infrared gas
detection and the proper operation of this device by reading this manual thoroughly.
Warning Statements
The use of the word
equipment. It calls attention to a procedure, practice, or condition, or the like, which if not correctly
performed or adhered to, could result in personal injury or death.
WAR
Caution Statements
The use of the word
equipment. It calls attention to a procedure, practice, condition, or the like, which if not correctly
performed or adhered to, could result in damage to the equipment.
CAUTION
Hazard Symbols on Monitor
This symbol indicates the need to consult this operating
instruction manual when opening the enclosure.
WARNING: A potential risk exists if the operating instructions
are not followed.
This symbol indicates the presence of electric shock hazards when
the enclosure is opened.
WARNING: To avoid risk of injury from electric shock, do not open
the enclosure without first disconnecting AC power.
Instruction 3015-4603
1
BACHARACH
NING
in this manual denotes a potential hazard associated with the use of this
in this manual denotes a potential hazard associated with the use of this
CO2MSZ (Carbon Dioxide Gas Monitor Single Zone).
Page 6
Safety Precautions
AC Power Supply
The CO2MSZ uses a universal power supply that accepts AC inputs of between 100 and 240 Vac, 50/60 Hz,
and has a power consumption of 15 Watts. It is highly suggested that the monitor be connected directly to
the AC power source, preferably on its own circuit (with UPS or surge protection).
A switch or circuit breaker rated 1.0 A, 250 Vac, with a minimum terminal spacing of 3.0 mm must be
attached to the monitor’s AC power leads. This switch must also be located in close proximity to the
monitor, and be in easy reach of the operator. This switch should also be clearly marked as the monitor’s
main AC disconnect device.
Protective Grounding
Under no circumstances should the CO2MSZ be operated without connection to a protective ground. Doing
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type
of equipment.
Explosive Atmosphere
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any
electrical equipment in such an environment constitutes a safety hazard.
Proper Exhaust Venting
It is imperative that the monitor’s exhaust port be vented as described in this manual. Failure to do so
may constitute a safety hazard.
Working Inside the Monitor
Extreme care should be exercised when accessing the interior of the monitor. Only qualified electrical
maintenance personnel should perform connections and adjustments. Always remove AC power before
working inside the monitor.
Misuse and Modifications to Monitor
The protection provided by the monitor may be impaired if the monitor is used in a manner not specified by
Bacharach, Inc. Changes or modifications to this monitor, not expressly approved, will void the warranty.
In Case of Malfunction
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of
such occurrence, de-energize the power supply and contact a qualified repair technician or the nearest
Bacharach Service Center. Use ONLY the provided knockouts for electrical and communication wiring.
Drilling into the box will void the warranty.
Fusing
F1, F2: 1.0 A, 250 V, Type “F”
Installation Category
Installation Category II, Pollution Degree II, as defined by UL.
Altitude Limit
6,562 ft (2,000 m)
Cleaning
Clean the outside of the monitor’s case using a dry cloth —
2
DO NOT
use soap and water.
Instruction 3015-4603
Page 7
Functional Overview
General Description
Carbon dioxide (CO2) monitors are specified to support compliance to federal, state and local safety codes
governing gas emissions. Avoiding significant CO
maintains equipment efficiency, promotes safety, and protects the environment.
The Bacharach CO2MSZ (Carbon Dioxide Gas Monitor Single Zone) is designed to continuously test for
2
the presence of CO
gas in an area that is located up to 500 ft (152.4 m) away from the monitor. The
CO2MSZ displays that CO
peak gas level detected in that area on its front panel LCD. The monitor retains a log of previous readings
that can be easily accessed for analysis.
An audible alarm and front panel indicators are provided to signal alarm and fault conditions, and relay
contacts are provided that can be used to trigger external alarm devices in the event of a system fault, or if
a leak (small), spill (medium), or evacuation (large) level of gas is detected. The system also includes a
4
−
20 mA current loop interface that can be connected to remote monitoring equipment.
The CO2MSZ requires only minor periodic maintenance such as the occasional replacement of filters. The
monitor incorporates active diagnostics that continuously check the system for proper operation. A front
panel indicator is provided to alert an operator of system malfunctions, and fault codes are generated that
enable the operator to identify the cause of the fault.
Understanding Monitoring Levels
Effective use of this monitor requires an understanding of what constitutes reasonable alarm set points for
the type of gas being monitored. Most systems leak some gas. In a good “tight” installation these
background levels will be acceptably low and often do not require corrective action. You can reduce nuisance
alarms and needless service calls if the alarm levels are set at practical limits.
The CO2MSZ is shipped with its leak, spill, and evacuate alarm levels set at 100, 300, and 500 ppm,
respectively. Setting the monitor at these recommended alarm levels will satisfy the needs of most users.
However, the ppm levels generated by system leaks into the environment are greatly influenced by the
volume of air in the sampling area, air circulation, size of the leak, distance to the monitoring point, and a
host of other variables. In some cases the alarm set points may need to be adjusted either up or down to
achieve effective monitoring.
Instruction 3015-4603
3
2
gas loss reduces equipment replacement costs,
2
gas being monitored, along with displaying both the current gas level and the
Page 8
Suggested Location of Sampling Point
At the point of a CO
2
gas leak the gas is nearly pure. As CO
2
gas is dispersed into the air, the gas
molecules diffuse causing a dilution of the original concentration. The CO2MSZ measures the gas
concentration at the point the sample is collected. Therefore, if the termination of the monitor’s gassample line is not at the exact point of the gas leak, then the monitor will read a diluted mixture of
CO
2
gas and air.
It should also be noted that CO
2
gas is heavier than air and tends to collect below the point of a leak.
Therefore a sample taken near the floor will have a greater concentration of gas than that collected above
the source of a leak. Consequently, the sampling point should ideally be located as close as possible to the
source of a potential leak. If this is impractical, then the alarm set points should be adjusted to compensate
for the dilution of the gas.
The CO2MSZ should be mounted outside of the mechanical room or at least just inside of a door to the
room. This is the “split architecture design” for safety of the operator. The monitor should be readily
accessible for easy visual monitoring and servicing. The total length of the sample line + exhaust line can
be up to 500 ft (152.4 m). If the area around the monitor is not well ventilated, then an optional exhaust
line and purge line can be run to an outside location. The total length of the exhaust line + purge line can
not exceed 200 ft (61.0 m).
It may be necessary to perform a “smoke” test of the mechanical room to determine the best monitoring
point. The smoke test would provide the pattern of air currents present in the room.
CO2MSZ Gas Leak Monitor Mechanical Room Placement
4
Instruction 3015-4603
Page 9
Installation
Installation Considerations
Warnings and Cautions
WARNING: Explosion ha
contain flammable liquids, v
in such an en
viro
WARNING: Shock hazar
the monitor.
CAUTION:
void the war
Drillin
ranty. Please use knockouts provided
CAUTION:
damaged. Be careful not to touch or disturb an
Inspection
The CO2MSZ has been thoroughly inspected and tested prior to shipment from the factory. Nevertheless, it
is recommended that the monitor be re-checked prior to installation. Inspect the outside of the enclosure to
make sure there are no obvious signs of shipping damage. Open the enclosure and inspect the interior of the
monitor for loose components that may have become dislodged during shipment. If damage is discovered,
please contact the nearest Bacharach Service Center for assistance.
T
he CO2MSZ contains sensitive electroni
Locating the Monitor
Place the monitor in a location that will require no more than 500 ft (152.4 m) of tubing (total of sample
line + exhaust line) to reach the sampling area. Since the monitor uses ambient air to zero its gas sensor,
the monitor should be located in an area that normally does not contain refrigerant gas. If an exhaust line
is not used, then locate the monitor in a well ventilated area so that the exhaust gas does not accumulate
around the monitor’s ventilation holes.
The CO2MSZ should be operated in an environment that is between 32 and 122°F (0 and 50°C); has a
relative humidity of between 5 and 90% non-condensing; and is at an altitude of no more than 6,562 ft
(2,000 m). The area should also be relatively free of dirt, grease, and oils that could adversely affect the
operation of the monitor.
The location should allow the monitor to be easily accessible for visual monitoring and servicing.
Instruction 3015-4603
5
zard! Do not mount the CO2MSZ in a
apors
or aerosols. Operation
nment constitutes a safety hazard.
d! Alwa
g holes in the CO2MSZ enclosure may damage the unit an
ys disconnect AC
power be
for electrical connections.
c components that can be easily
y of these compon
n area that may
of an
y electric
fore workin
ents.
al equipment
g inside
d will
Mounting Instructions
Page 10
The CO2MSZ should be installed plumb and level, and securely fastened to a rigid vertical mounting surface.
The enclosure utilizes keyhole mounting brackets designed for ¼ inch fasteners. Install the four
screws as shown in the diagram below or by using the provided mounting template. Allow the screw heads to
protrude approximately 1/4".
Enclosure Mounting Specs
Hold the monitor flat against the mounting surface and allow it to slide down engaging the screw heads in
the keyhole slots of the mounting brackets. Adjust the screws as necessary to hold the monitor securely
against the mounting surface.
Connecting Gas Sample, Purge, and Exhaust Lines
Overview
A single gas-sample line needs to be run from the CO2MSZ to the area of the facility to be monitored. An
optional purge line can be run to bring fresh air into the monitor for detector zeroing purposes. And an
optional exhaust line can be installed to vent the sample gas away from the monitor. All tubing
connections are made on the bottom of the enclosure as shown in the illustration below.
½" / ¾" Service
Knockout*
* There is an additional Service Knockout on top of the case
6
Purge Air
Inlet Port
½" / ¾" Service
Knockout*
CO2MSZ Bottom View
Gas Sample
Inlet Port
Gas Sample
Exhaust Port
Instruction 3015-4603
Page 11
Tubing Considerations
Use ¼" outside diameter (0.040" wall) flex tubing for all air lines (P/N 304-2742 or equivalent). The tubing
should be clean and free of moisture or other contaminants. The tubing should be cut cleanly with a sharp
knife and care should be taken not to distort the tubing end.
All tubing bends should have a radius of no less than 5" to ensure proper airflow. If kinks or obstructions
occur in the line the monitor may not function properly.
Connecting the Gas-Sample Line
Please refer to Section
a gas sample.
The gas-sample line can be up to
exhaust line (if one is used) should not exceed 500 ft. The end of this line should be placed near the potential
leak source and positioned to reduce the possibility of mists, aerosols, oil, water, dust, or other contaminates
from being drawn into the monitor. For added protection, a termination filter (P/N 3015-3420) supplied with
the monitor must be attached to the end of this line.
CAUTION:
in serious an
Suggested Lo
T
he intr
d permanent damage to th
cation of Sampling Po
500 ft (152.4 m)
oduction of contaminan
Connecting the Purge Line
The purge line functions to bring fresh air into the monitor for the purpose of purging the detector and
setting its baseline reading. This purge line is required if the monitor is located in an area that does not
contain fresh air.
The purge line can be up to
to contain fresh air. To connect the purge line to the monitor, simply push the tubing into the Purge Air
Inlet Port. If the purge line terminates outside the building, position the tubing so that no water or
moisture can enter the line. For added protection from contaminates entering the monitor, a termination
filter (P/N 3015-3420) must be attached to the end of this line.
100
ft (3
0.5 m)
int
(Page 4) to learn more about where to take
in length. The combined length of the sample line and
ts th
roug
h the gas-sample line can result
e monitor.
in length. Ideally, this line should terminate in an area known
Connecting the Exhaust Line
The exhaust line functions to carry the exhausted gas sample away from the monitor, and is required if
the monitor is located in a confined, poorly ventilated area. If no purge line is used, it is a good idea to vent
the exhausted gas sample away from the monitor to prevent erroneous readings due to improper zeroing of
the detector.
The exhaust line can be up to
location that is not exposed to the elements. Connect the exhaust line to the monitor by firmly pushing the
tubing onto the Gas Sample Exhaust Port’s barbed fitting. Ensure that no loops or dips are present that
could trap condensate. This is likely as outside air can be cooler than instrument. Note that the exhaust
line does not require a termination filter.
100 ft
(30.5
m)
in length. Ideally this line should terminate outdoors in a
Termination Filter (P/N 3015-3420)
Instruction 3015-4603
7
Page 12
Interior Schematic
LINE FUSES
1.0A, 250V, F
PURGE AIR IN
8
AC INPUT
GROUND
STUD
*NOTE: The plastic cable tie surrounding the air pump is to
ensure safe handling during shipping. Please remove
before operating.
HYDROPHOBIC
FILTER
AC INPUT
TERMINALS
Interior Schematic
FACTORY
DEFAULT
SWITCH
EXTERNAL
ALARM RELAY
CONNECTOR
IR BENCH
GAS IN
IR BENCH
GAS OUT
4–20 mA
(Output only DO NOT apply
power!)
HYDROPHOBIC
FILTER
*AIR SAMPLE
PUMP
Instruction 3015-4603
Page 13
Electrical Wiring
The CO2MSZ uses a universal power supply that accepts AC inputs of between 100 and 240 Vac, 50/60 Hz,
and has a power consumption of 15 Watts. It is highly recommended that the monitor be connected
directly to the AC power source, preferably on its own circuit. The connection should be completed with UL
−
approved multi-conductor wire (14
Locate a convenient service knockout and install electrical conduit in the typical manner.
Locate the AC Input Terminals and Ground Stud on the inside of the monitor (Page 8).
Secure the incoming AC power neutral (white) and live (black) wires to the LINE 1 and LINE 2 terminals.
Using the supplied crimp-on ring terminals, washers, and nuts, connect the incoming AC power ground
wire (green) to the monitor’s AC Input Ground Stud, and then install a separate wire between the ground
stud and the GND terminal.
A switch or circuit breaker rated 1.0 A, 250 Vac, with a minimum terminal spacing of 3.0 mm must be
attached to the monitor’s AC power leads. This switch must also be located in close proximity to the
monitor, and be in easy reach of the operator. This switch should also be clearly marked as the monitor’s
main AC disconnect device.
WARNING:
shou
ld comp
Elec
trical installation should be p
ly with all ap
WARNING:
stud. Under no circums
grou
nd. Doin
safety standards applicab
The AC powe
g so poses a pote
CAUTION:
void the war
Drillin
ranty. Please use knockouts provided
Instruction 3015-4603
9
18 AWG), rated 300 Vac at 105°C.
AC Power Connections
erformed by a certified electrician, and
plic
able N
EC/CEC and local electrical safety codes.
r groun
tances shou
d wire must first be
ld this monitor be operate
ntial sh
ock haza
rd, and is also a violation of electrical
connected to the monitor
d withou
t a protective
le to this type of equipment.
g holes in the CO2MSZ enclosure may damage the unit an
for electrical connections.
’s ground
d will
Page 14
Connecting External Alarms
Overview
The CO2MSZ includes four SPDT relays whose contacts are rated 2 A at 250 VAC (inductive) 5 A at 250
VAC (resistive). These relays are used for the connection of external alarm devices that are activated
when the relay is energized. The relays are factory assigned to energize under the following conditions:
The alarm points of relays #1, #2, and #3 are set using the monitor’s LEAKLVL, SP
EV
ACLVL function screens (Page 16).
Connection
Use any of the service knockouts to gain cable access to the interior of the monitor. Locate the External
Alarm Relay Connector (Page 8). Secure the leads from the external alarm device and its power source to
the connector as shown in the diagram below.
The diagram on Page 11 shows a typical alarm device being connected to Relay 1. Note how power to this
device is being tapped off the monitor’s AC input.
Each relay may be connected as normally open (NO), or normally closed
(NC). The relay contacts are rated 2 A at 250 VAC (inductive) 5 A at 250
VAC (resistive).
Power for the external alarm devices may be supplied from an external power source, or
from the monitor’s AC input terminals.
10
Relay #1 Leak Alarm Point 500 ppm
Relay #2 Spill Alarm Point 700 ppm
Relay #3 Evacuate Alarm Point 1000 ppm
Relay #4 Fault System Fault Event
External Alarm Relay Connector
ILLLVL
Instruction 3015-4603
, and
Page 15
Typical External Alarm Relay 1 Wiring
Jumper the ‘Neutral’ line of an external power source or the monitor’s AC input to the ‘Common’
terminals on the relay connector.
Connect one end of the strobe or horn to the ‘NO’ terminal of whichever level of alarm is appropriate for
the application.
The other end of the strobe or horn is connected to the other leg of the external power source. For
protection, install an in-line fuse of the appropriate size and design for the external alarm device being
used.
4–20 mA Current Loop Interface
An external 4–20 mA monitoring device (e.g., chart recorder) can be connected to the CO2MSZ using a
shielded-twisted-pair cable. Use any of the service knockouts to gain access to the interior of the monitor.
Locate the 4
shorting wire from its terminals.
IMPORTANT!
isolator as an accessor
NOTE:
device is used. Failure to insert th
the wire leads from the external monitoring device to the 4–20 mA connector as shown in the diagram
below, making sure that the polarity at this connector matches the wiring at the monitoring device.
The default current-to-ppm factor is set to 0.016 mA = 1 ppm, providing a measurement range 0 ppm
(4 mA) to 1,000 ppm (20 mA). The current-to-ppm factor can be changed using the monitor’s
function (Page 16).
To facilitate loop zero and span checking, note that when the
set to 20 mA, and upon exiting this function the loop output is set to 4 mA.
−
20 mA Connector (Page 8), remove it from the circuit board, and then remove the resistor or
The 4-20mA ou
y
such as a Datafor
A 100 ohm, ¼ W resistor must be inserted into th
tput
is a "floating" gr
th par
ound ty
t n
umber DSCL
pe and must be used with a loop-powered
21-01.
e 4–20 mA connector if no ex
is re
sistor may cause a loop fault code <0010> to occur (Page
tern
al monitorin
19). Secure
LOOP
LOOP
function is selected the loop output is
4–20 mA Current Loop Connecto r
g
CAUTION: N
extern
device.
Instruction 3015-4603
ever apply power to the 4–20 mA Curre
al power supply. Connect
on
ly a load resistor and/or a
nt Loop Connector from an
floating
measur
ement
11
Page 16
12
Notes:
Instruction 3015-4603
Page 17
Operation
Front Panel Display and Control
Keypad
Use these buttons to:
1) Enter function mode and
then move arrow (>) on screen
to the desired function
2) Scroll through data
3) Change a function’s value
MONITOR ON
Light (Green)
1) Flashes during
warm-up
2) Lights steady
during normal
operation
SYSTEM FAULT
Light (Yellow)
Flashes when any
one or more of the
faults
listed on Page
19 occur
ALARM Light
(Red)
Flashes when the
detected gas level
exceeds any of
three preset alarm
points
General Operation
Once the CO2MSZ has been installed, set up, and powered ON, the monitor will make measurements in
the area being sampled for CO
measurements are shown on the front panel display.
MEASURE 00955pk
340ppm CO2
In the example shown above,
and that currently a concentration of 340 ppm is being detected. This display also indicates that a peak
measurement of 955 ppm has been made.
An operator can reset the peak value to zero by pressing the ENTER
A log of up to 200 previous measurements can be viewed using the
2
gas without further operator intervention. The results of those
M
EASURE
indicates that the CO2MSZ is actively monitoring for CO2 gas,
button.
PPM LOG function (Page 15).
Display
Screen
ENTER
Press this button to:
1) Clear peak reading
2) Save a function
screen’s displayed value
SILENCE / QUIT
Press this button to:
1) Acknowledge a
system fault or gas
alarm by turning OFF
the internal audible
alarm and de-energizing
the corresponding
fault/alarm relay
2) Return to the previous
screen without saving
data when in the
function mode
If the detected gas level exceeds the preset Leak, Spill, or Evacuate alarm point, then the monitor
responds by turning ON the front panel
ALARM
If the internal audible alarm is turned ON, it too will activate (
(red) light and energizing the corresponding alarm relay.
AUDALRM
Page 16). Optional external
alarm devices can be connected to the alarm relays to alert personnel that a Leak, Spill, or Evacuate alarm
condition has occurred (Page 10). Pressing the front panel
and turn OFF all alarm indicators with the exception of the front panel
reactivate, however, if the alarm condition is not cleared within the time period set by the
SILENCE
button will acknowledge an alarm
ALARM
light. The alarm circuit will
SILENCE
function (Page 17). The ALARM light will turn OFF after the detected gas level goes below the lowest alarm
point and after the
SILENCE
button is pressed.
Instruction 3015-4603
13
Page 18
If a system fault occurs (see Fault Code list on Page 19), the monitor responds by turning ON the front
panel
SYSTEM FAULT
turned ON, it too will activate (
(yellow) light and energizing the fault relay. If the internal audible alarm is
AUDALRM
Page 16). An optional external alarm device can be connected
to the fault relay to alert personnel that a system fault has occurred (Page 10). Pressing the front panel
SILENCE
front panel
cleared within the time period set by the
button will acknowledge the fault and turn OFF all alarm indicators with the exception of the
SYSTEM
FAULT light. The alarm circuit will reactivate, however, if the fault condition is not
SILENCE
function (Page 17). The SYSTEM FAULT light will
turn OFF only after the cause of the fault has been eliminated.
A log of the alarm and fault events can be viewed using the monitor’s
ALARMS and FAULTS
function
(Page 15).
Display Screens
Initial Power Up
When the CO2MSZ is first powered up all front panel lights are turned ON, and a splash screen appears
showing the monitor’s firmware revision level. After a brief moment the
with the front panel
BACHARACH
VERSION x.xx
The monitor takes 15 minutes to warm up; after which, the MONITOR ON
Display
Data
Data Display Screen
MEASURE 00955pk PURGE 00955pk
340ppm CO2
During normal operation, the
following three functions:
ME
ASURE
example screens shown above, the monitor is currently detecting 340 ppm of CO
measurement of 955 ppm has occurred.
To reset the peak value to zero, press the ENTER button.
PURGE is displayed when the monitor is resetting its infrared detector to a baseline ppm value using
fresh air that is being drawn in through the Purge Air Inlet Port. This purging process is performed on an
“as needed” basis, which is normally once every 6 to 8 minutes.
PRES CHK
ensure the accuracy of the gas measurement under varying atmospheric conditions. This pressure check is
performed every 30 minutes.
14
screen is displayed.
indicates that the monitor is actively measuring the CO2 gas level at the sample point. In the
is displayed when the monitor is performing an atmospheric pressure check, which is done to
MO
NITOR ON
light (green) blinking.
WARM UP
340ppm CO2
Data Display
Wa
rm Up
screen is display along
light glows steady and the
PRES CHK 00955pk
340ppm CO2
screen shows when the monitor is performing the
2
gas, and that a peak
Instruction 3015-4603
Page 19
Function Screens
The Function screens are used to display stored data and to set up the monitor.
From the Data Display
Next, use the
press the
>PPM LOG FAULTS
ALARMS DIAGNOS
>GASTYPE SPILLVL
LEAKLVL EVACLVL
EN
Keypad
TER
button to select that function.
screen, press any
buttons to move the arrow (>) on the display next to the desired function, and then
>LOG INT AUDALRM
LOOP CLOCK
>SILENCE ZONETMP
SQUELCH CAL
Keypad
button to display the first
>P-CHK
Function Menu
screen.
Once a function has been selected, use the
parameter associated with that function. Press
QUIT
button to return to the previous screen without saving.
If no buttons are pressed within 90 seconds after selecting a function, the monitor will return to the Data
Display
screen.
PPM LOG – Contains records of the last 200 measurements. Each
record shows the measurement’s date, time, and ppm level. The interval at
which the measurements are logged is determined by the
function.
Use the Keypad Up and
a factor of 1. Use the
number by a factor of 10. Press
Ri
ght
Down
and
QUIT
buttons to change the record number by
Left
the example on the right, record #025 shows that a level of 995 ppm was
measured on 07/05/07 at 15:35.
The PPM Log can be cleared as described in Section
PPM Log, Alarm & Fault Data (Page 20).
ALARMS – Contains records of the last 30 alarm events, with the
monitor’s most recent alarm being shown when the
displayed. After 30 events have been recorded, the newest record
overwrites the oldest. Each record displays either LEAK, SPILL, or EVAC
including the date and time the alarm occurred. Use any of the
buttons to scroll through the other alarm records. Press
the previous screen. Refer to
(Page 18).
Work
ing with Gas
The ppm levels at which alarms occur are set using the LEAKLVL,
SPILLLVL
, and EVACLVL
functions.
In the example on the right, record #03 shows that a spill event occurred
on 07/10/07 at 15:06.
FAULTS – Contains records of the last 30 fault events, with the monitor’s
current fault status being shown when the
After 30 events have been recorded, the newest record overwrites the
oldest. Each record lists an event’s numeric fault code plus the date and
time the fault occurred. Use any of the
the other fault records. Press
to
Workin
g with System Faults
In the example on the right, record #15 shows that a Purge Flow Fault
<1000> occurred on 07/10/07 at 12:37.
QUIT
to return to the previous screen. Refer
(Page 18).
Keypad
to scroll through the displayed data or to change a
EN
TER
to save newly entered parameters. Press the
LO
G INT
buttons to change the record
to return to the previous screen. In
Clearing
th
e Stored
Alarm screen is first
Keypad
QUIT
to return to
Alarms
Fault
screen is first displayed.
Keypad
buttons to scroll through
#025 995PPM @
07/05/07 15:35
#03 SPILL @
07/10/07 15:06
#15 <1000> @
07/10/07 12:37
Instruction 3015-4603
15
Page 20
DIAGNOS – Enters the diagnostic function. Press the
Keypad
Refer to
button to toggle between the monitor’s two diagnostic screens.
Working with th
e DIA
GNO
S Function (Page 21).
Up Arrow
GASTYPE – CO
choose from, therefore CO2 is the only gas type listed. Press
return to the previous screen.
2
gas monitors do not have a library of other gases to
QUIT to
LEAKLVL – Sets the Leak Alarm level, adjustable from 1 to 1999 ppm.
Factory default is 500 ppm. The Leak Alarm level cannot be set higher
than either the Spill or Evacuate Alarm levels. Use the
the desired Leak Alarm level, and then press
ENTER
and return to the previous screen.
Keypad
to set
to save that level
SPILLLVL – Sets the Spill Alarm level, adjustable from no lower than the
Leak Alarm level to a maximum value of 1999 ppm, but not higher than the
Evacuate Alarm level. Factory default is 700 ppm. Use the
the desired Spill Alarm level, and then press
ENTER
and return to the previous screen.
Keypad
to set
to save that level
EVACLVL – Sets the Evacuate Alarm level, adjustable from no lower
than the Spill Alarm level to 1999 ppm. Factory default is 1000 ppm. Use
the
Keypad
ENTER
to set the desired spill level alarm level, and then press
to save that level and return to the previous screen.
LOG INT – Sets the interval at which measurements are logged to
memory from 1 to 9999 minutes. Factory default is 10 minutes. The logged
measurements can be viewed using the
Keypad
to enter the desired value, and then press
value and return to the previous screen.
PPM LOG
function. Use the
ENTER
to save that
LOOP – Adjusts the loop factor of the 4–20 mA current loop. Factory
default is 0.016 mA per ppm, which provides an output range of 0 ppm
(4 mA) to 1,000 ppm (20 mA). Use the
factor, and then press
previous screen.
ENTER
to save that value and return to the
Keypad
To facilitate loop zero and span checking, note that when the
function is selected the loop output is set to 20 mA, and upon exiting this
function the loop output is set to 4 mA.
to set the desired loop
LOOP
AUDALRM – Allows the monitor’s internal audible alarm to be
associated with any function of the monitoring system. Factory default is
OFF. Use the
then press
Audible Alarm Settings: OFF, ANY ALARM, SYSTEM FAULT, LEAK ALARM,
SPILL ALARM, EVAC ALARM, MONITOR STOPPED
NOTE: MONI
monitor is
16
Keypad
ENTER
TOR STOPPE
no lo
nge
to select the desired audible alarm function, and
to save that value and return to the previous screen.
r fu
nctioning
D indicate
s there
correctly.
is a
cr
itical
system fault, and th
e
4.26500v <0000>
24.5cD 14.00psi
0.4ppm 0.00
0.00002au 4.260v
SELECT GAS TYPE
CO2
EDIT LEAK LEVEL
500 PPM
EDIT SPILL LEVEL
700 PPM
EDIT EVAC LEVEL
1000 PPM
LOG INTERVAL IS
0010 min
EDIT LOOP FACTOR
0.016 ma/PPM
AUDIBLE ALARM IS
OFF
Instruction 3015-4603
Page 21
CLOCK – Sets the monitor’s date and time. Use the
the correct date and time, and then press
ENTER
Keypad
to enter
to save those values
and return to the previous screen. Time is displayed in a 24 hour format,
while the date is displayed as mm/dd/yy.
SILENCE – Used to enter a length of time for which the internal
audible alarm and the external alarm are turned OFF when the front
panel
SILENCE
(5 minutes). If the cause of the alarm/fault has not been cleared at the
end of this time period, the internal audible alarm and the external alarm
device are reactivated. Use the
and then press
screen.
button is pressed. The factory default is 300 seconds
Keypad to enter the desired time period,
ENTER
to save that value and return to the previous
SQUELCH – Sets a value of between 0.0 and 100 ppm that prevents
the display of measurements below that value. Factory default is
0 ppm. For example, if the squelch setting is set to 5 ppm, then the monitor
will not display measurements that are below that value. Use the
to enter the desired value, and then press
ENTER
to save that value and
return to the previous screen.
Keypad
ZONETEMP – Used to enter the temperature of the area being
monitored in °C, thus giving a more accurate ppm reading. The factory
default is 25°C. Use the
then press
ENTER
K
eypad
to enter the desired temperature, and
to save that value and return to the previous screen.
CAL – Used to change the monitor’s calibration factor. This function is to
be used ONLY with instructions from a Bacharach Service Center. Factory
default is 1.000.
P-CHK – This
pressure and the stored ambient pressure in psia, along with the difference
between these two pressures and the current fault code. Refer to
with the P-CHK Function
Instruction 3015-4603
Pressu
re Ch
eck
function displays the current manifold
Workin
g
(Page 20).
17
SET DATE & TIME
07/10/07 15:30
SILENCE TIMEOUT
0300 sec
SQUELCH BELOW
5 ppm
TEMP AT ZONE =
25 degC
R717 CALFACTOR
1.000
14.72 * AMB14.81
0.09dif <0000>
Page 22
Working with Gas Alarms
Overview
If the CO2 ppm level in the area being monitored exceeds its preset Leak, Spill, or Evacuate Alarm level
(Page 16), the CO2MSZ will detect this alarm condition and turn ON the front panel
Additionally, an external alarm device may activate and the monitor’s internal audible alarm may sound if
those features have been enabled (Pages 10 & 16).
Clearing / Silencing a Gas Alarm
Once the gas-alarm circuit has been triggered all alarm indicators remain turned ON, even after the
detected gas level returns to normal. Press the
cause of the alarm has been cleared and the detected gas level has dropped below all alarm levels.
Pressing the
and all external alarm devices that are connected to the alarm relays to turn OFF for a period of time as
set by the
indication that an alarm condition still exists, but will automatically turn OFF once the gas level returns to
normal. The alarm circuit will reactivate at the end of the silence period if the detected gas level is still
above an alarm level.
Viewing the Gas Alarm Log
From the Data Display
ALARMS
The alarm log shows the type of alarm (LEAK, SPILL, or EVAC), plus the date and time it occurred. If
CLEAR is displayed, this indicates that an alarm was acknowledged at the date and time shown.
Immediately after selecting the
event is displayed. In the example on the right, record #03 shows that a
Spill Alarm occurred on 07/10/07 at 15:06. If more than 30 alarm events
have occurred, then the newest event overwrites the oldest.
The Alarm Log can be cleared as described in Section
PPM Log, Alarm & Fault Data (Page 20).
Working with System Faults
Overview
If a system malfunction occurs, the CO2MSZ will detect the problem and turn ON the front panel
SYSTEM FAULT
internal audible alarm may sound if those features have been enabled (Pages 10 & 16).
Clearing / Silencing a Fault Alarm
The
SYSTEM FAULT
the fault has been eliminated.
Pressing the
external alarm device that is connected to the fault relay to turn OFF for a period of time as set by the
SILENCE
reminder that a fault condition still exists. The alarm circuit will reactivate at the end of the silence period
if the cause of the fault has not been corrected. The
once the fault has been cleared.
18
SILENCE
SILENCE
SILENCE
button while a gas-alarm condition still exists causes the internal audible alarm
function (Page 17). The front panel ALARM
button to clear all gas-alarm indicators after the
screen, use the
function. Then press ENTER to display the alarm log.
ALARM function, the most recent alarm
Keypad
buttons to place the arrow (>) on the display next to the
Clearing the Stored
(yellow) light. Additionally, an external alarm device may activate and the monitor’s
light and all other fault indicators will automatically turn OFF after the cause of
SILENCE
function (Page 17). The front panel
button while a fault condition still exists causes the internal audible alarm and the
SYSTEM FAULT
SYSTEM FAULT
ALARM
light remains ON, however, as an
#03 SPILL @
07/10/07 15:06
light remains ON, however, as a
light will automatically turn OFF
light (red).
Instruction 3015-4603
Page 23
Viewing the Fault Log
From the Data Display
FAU
LTS
function. Then press ENTER to display the fault log.
screen, use the
Keypad
buttons to place the arrow (>) on the display next to the
The fault log shows the monitor’s current fault status. If the fault is still
present when the
ULTS
function is selected, then the current cause of the
FA
fault is displayed along with the date and time it occurred. If the cause of the
fault has been cleared, then the fault log will show <0000> along with the
ke
ypad
date and time the fault was cleared. Use the
buttons to scroll
through the fault log. In the examples shown on the right, record #03 shows
that a Zone Flow Fault (fault code <0800>) occurred on 07/12/07 at 08:17,
while record #04 shows that the fault was cleared on 07/12/07 at 8:30.
The cause of the fault is identified by a numeric fault code. To convert the
fault code into a text description of the fault, first press the
and then use the
Keypad
buttons to scroll through the display until the text
ENTER
button
description of the fault appears. If the fault code is a combination of two or
more faults, then continue to use the
Keypad
buttons until all fault text
descriptions have been displayed. For example, the fault code <1800>
represents the combination of both a Zone Flow <0800> and a Purge Flow
<1000> fault as shown on the right.
The Fault Log can be cleared as described in Section
Clearing
th
e Stored
PPM
Log, Alarm & Fault Data (Page 20).
Fault Codes
FAU
LT CO
A fault code of <1800> indicates that both a Purge
Flow Fault <1000> and a Zone Flow Fault <0800>
have occurred.
<0001> Box Temp
temperature is outside normal range (or IR detector
has failed). Check that the monitor is not being
subjected to extreme temperatures. Verify that the
ventilation holes are not obstructed. Use the
DIAGNOS function (Page 21) to check the Box
Temperature.
<0002> Bench
bench is outside normal operating range (or IR
detector has failed). Check that the monitor is not
being subjected to extreme temperatures.
<0004> - Ma
manifold pressure is outside normal operating range
(or IR detector has failed). Enter the
function(Page 21) and record ALL data. Call the
with this information for further instructions.
DES ARE
Temperature Fault
nifold Pressure Fault:
ADDITIVE. For example:
erature Fault:
DIAG
Enclosure
: Optical
The
NOS
<0200> Gain Set Fault:
sequence has failed. This fault will only occur on
first boot up or after a firmware upgrade. Call the
factory for further instructions.
<0400> A/D Fault:
analog-to-digital circuitry. Contact the factory with
this information for further instructions.
<0800> Zon
in the gas-sample line or exhaust line; a blocked
external termination filter; a blocked internal
hydrophobic filter (Page 8); a failed pump.
<1000> Purge Flow Fa
in the purge line; a blocked external termination
filter; a blocked internal hydrophobic filter
(Page 8); a failed pump. Once the blockage has
been cleared, the monitor will return to normal
operation after the monitor completes a purge
cycle.
<4000> Zero Fault:
e Fl
voltage is out of tolerance. Enter the
<0010>
open, or there is a high resistance in the circuit. Check
the wiring to the load/monitoring circuit. If this feature
is not being used, a 100 ohm resistor must be
connected to the 4–20 mA connector (Page 11).
Loop
Fault:
The 4–20 mA current loop is
function (Page 21) and record ALL data. Call the
factory with this information for further
instructions.
<8000> Clipping Fa
may be out of tolerance. Use the
function
<0100> Zero Fi
lter Fa
ult:
There is CO2 gas present
in the purge air. Locate the monitor in an area
containing fresh air, or install a purge air line.
Refer to Loca
ting the Monitor
(Page 5).
(Page 21) to check the IR detector voltage. Call the
factory with this information for further instructions.
Instruction 3015-4603
19
#03 <0800> @
07/12/07 08:17
#04 <0000> @
07/12/07 08:30
FAULT CODE<1800>
12 ZONE FLOW
FAULT CODE<1800>
13 PURGE FLOW
The digipot autotune
A fault has occurred in the
ow Fault:
The IR detector’s output
ult:
Check for: A kink
ult:
Check for: A kink
D
IAGNOS
The detector voltage
DIAGNOS
Page 24
Clearing the Stored PPM Log, Alarm & Fault Data
Up to 200 gas measurements, and 30 alarm and fault events are stored by the monitor.
To clear stored data, first display the data to be cleared by using the PPM LOG, ALARMS
function (Page 15). Next, press the
time. A single, long tone should be heard when the data has been successfully cleared.
Right Arrow Keypad
and
ENTER
buttons at the same
or FAULTS
Working with the P-CHK Function
Overview
The
P-CHK
troubleshooting a flow fault problem. The monitor will trigger a flow fault if the pressure drop from
ambient is less than 0.2 psi during a purge cycle, and 0.5 psi during a measurement cycle.
Keypad Functions
From the Data Display
P-CHK function. Then press
The
Left Arrow
when the purge valve is open (purging).
The
Down Arrow Keypad
Pressing the
pressure shown on the right (must be done with the pump OFF).
Screen Display
14.72 * AMB14.81
0.09dif <0000>
Man
ifold
Stored Ambient Pressure
Pressure Difference
stored ambient pressure.
Fault Code
Purge Valve As
and closed by pressing the
purge valve is
20
function (Pressure Check Function) (Page 17) is useful to a service technician for
screen, use the
ENTER
Keypad button toggles the purge valve open and closed. An asterisk () appears
button toggles the pump ON and OFF.
ENTER
button stores the current manifold pressure shown on the left to the ambient
Keypad
to display the Pressure screen.
buttons to place the arrow (>) on the display next to the
Manifold Pressure
Pressure Difference Fault Code
Pressure – Current manifold pressure in psia.
– Stored ambient pressure in psia.
– The difference between the current manifold pressure and the
– Current fault code (Page 19).
terisk (*) – When viewing the
Le
open
ft Arro
.
w Keypad button. An asterisk appears on the display when the
*
Stored Ambient Pressure
Pressure
screen, the purge valve can be opened
Instruction 3015-4603
Page 25
Working with the DIAGNOS Function
Overview
The
DIAGNOS function displays sensor data and status information useful to a service technician for
troubleshooting various fault conditions. Explanations of the data shown in these screens are given below.
Keypad Functions
From the Data Display
DIAGNOS
Press the Up
First Diagnostic Screen
4.26500v <0000>
24.5cD* 14.72psi
0.00250n <0000>
29.5cB* 14.72psi
In the first diagnostic screen, the user can toggle between displaying the Bench Voltage / Noise and
Detector Temperature / Box Temperature by pressing the
Bench Voltage
gas this value can range from 3.90000V to 4.50000V.
Noise –
reading is valuable mainly when CO
Detector Temperature
Box Temper
Fault Code
Manifo
Purge
Keypad
Second Diagnostic Screen
0.4ppm 0.00
0.00002au*4.260v
PP
M L
referenced to standard temperature and pressure.
Average Ab
0.00000 au. When sampling CO
µMoles
Detector Voltage
Diagnostic
Purge
closed by pressing the
purge valve is
Instruction 3015-4603
screen, use the
function. Then press ENTER to display the first of two
Arrow Key
pad
button to toggle between the
Keypad
buttons to place the arrow (>) on the display next to the
First
Diagnostic
and
Second Diagnostic
screens.
screen.
Bench Voltage Fault Code
Detector Temperature °C * Manifold Pressure
Noise Fault Code
Box Temperature °C
*
Manifold Pressure
Right Arrow K
eypad
button.
– This is the current peak-to-peak output of the IR detector. In the absence of CO2
The Noise value is a 16 point running average of the noise portion of the IR detector’s output. This
2
gas is NOT present.
– This is the current detector temperature in °C.
ature
– This is the current internal enclosure temperature in °C.
– Current fault code. A value of <0000> indicates that no faults are being detected.
ld Pressure – Current manifold pressure in psia.
Valve As
button. An asterisk appears on the display when the purge valve is
ter
isk (*)
– The purge valve can be opened and closed by pressing the Le
open
.
ft A
rrow
PPM Level µMole/Liter
Avg. Absorption Unit * Detector Voltage
evel – Parts Per Million Level is the current detected gas level, and is the volume concentration
sorption Unit
/Liter
– This is the absolute concentration in micro-moles per liter of CO2 gas.
screen.
Valve As
open
21
– This is a running average of the IR detector’s bench voltage as displayed in the
ter
isk (*)
Lef
.
– This is the optical absorbency. In the absence of CO2 gas the absorbency is
2
gas, its value varies proportionally with the gas concentration.
– When viewing the diagnostic screen, the purge valve can be opened and
t Arrow Ke
ypad
button. An asterisk appears on the display when the
First
Page 26
Working with the Calibration Function
Overview
If greater than standard accuracy is desired, the factory’s default calibration factor of 1.000 may be
adjusted by performing the calibration procedure as described below, and then selecting the monitor’s
CAL
function to enter the new calibration factor.
IMPORTANT!
remain within the factory-calibrate
algorithms adjust for temperature drift, IR source agin
factory accurac
Calibration Procedure
The calibration factor is determined by sampling a known dilution of CO2 gas. The sample must be
prepared to less than half the desired accuracy and the concentration must be corrected for ambient
temperature and pressure at the time of measurement.
Calibration is best performed at or near full scale, it can however, be done at any concentration and ideally
in the range where maximum accuracy is desired down to, but not below, 340 ppm.
A cylinder of a certified calibration gas must be used to ensure sampling occurs at ambient conditions.
A minimum sample size of 5 liters is required.
The CO2MSZ should be operating for at least one hour prior to performing a calibration.
Prepare the CO2MSZ for sampling by using the
Also, use the
With the CO2MSZ operating normally, connect the gas-sample bag directly to the gas-inlet port and allow
the monitor to sample the entire bag. When sampling is complete, view the logged ppm values using the
PPM LOG
new calibration factor is computed by dividing the known gas concentration value by the measured value.
Typically this number will be between 0.95 and 1.05. Use the
new calculated calibration factor
Adjusting Calibration Factor
From the Data Display
CAL
function. Then press
CO2 CALFACTOR
1.000
Changi
y
ng t
he cali
specifications.
bration fact
d ac
curacy indefin
or will
VOI
D the fa
itely an
g, and pre
ctory
cali
bratio
n. Typically, the monito
d no calibration is required. Comple
ssure changes in order to keep the unit
r wi
x software
with
CAL
function to set the calibration factor to 1.000.
LOG INT function to set the log interval to 1 minute.
function. If the bag was large enough for multiple samples, average the most stable ones. The
CAL function as described below to enter the
screen, use the
ENTER to display the
Keypad
buttons to place the arrow (>) on the display next to the
Calibration
screen.
ll
in
With the Calibration screen displayed, use the Keypad
factor. Press
22
ENTER
to save this value.
buttons to enter the new calibration
Instruction 3015-4603
Page 27
Maintenance
Warnings and Cautions
WARNING: Always disc
CAUTION:
electric
components that can be easily
components.
When working inside the monitor, be very careful not to dislodge any
al wiring
or pneumatic tu
onnect AC power before working inside
bing. Th
damaged. Be careful not to touch or
e CO2MSZ contains se
the monitor.
nsitive electr
disturb any
onic
of these
Hydrophobic Filters
Hydrophobic Filters (Page 8) located in both the gas-sample and purge lines prevent water from entering
the IR detector. A zone flow fault will occur (fault code <0800>) if the gas-sample filter becomes blocked,
while a purge flow fault will occur (fault code <1000>) if the purge filter becomes blocked. Replace the
appropriate filter as required (P/N 07-1650).
Servicing Air Lines & Termination Filters
The gas-sample line and the optional purge and exhaust lines should be periodically checked for obvious
signs of kinks, damage, and contamination. Replace the tubing as required (P/N 304-2742).
The gas-sample line and purge-line termination filters prevent dust and dirt from entering the monitor.
Both of these filters should be periodically checked and replaced when there are obvious signs of
contamination.
A zone flow fault will occur (fault code <0800>) if the gas-sample line filter becomes blocked, while a purge
flow fault will occur (fault code <1000>) if the purge-line filter becomes blocked.
To service the filter, simply remove it from the line and replace it with a new one (P/N 3015-3420).
Fuses
The CO2MSZ is protected from electrical damage by two, 1 A, 250 V, type “F” fuses (Page 8). Carefully
remove the fuses from their fuse clips and visually inspect each fuse for damage. Replace the fuses as
required (P/N 04-2620).
Clock Battery
The Clock Battery (Page 8) maintains the correct date and time when AC power is not applied to the
monitor. Replace this battery about every 5 years (P/N 204-0020).
Sample Pump
The Sample Pump (Page 8) draws the gas sample into the monitor, through the IR detector, and
discharges the sample out the monitor’s exhaust port. When replacing the pump, remove its inlet and
outlet tubing, disconnect the AC power wires from the pump itself, and remove pump from monitor. Install
a new sample pump (P/N 3015-5176).
Instruction 3015-4603
23
Page 28
Replacement Parts & Optional Accessories
Replacement Parts
Item Description Part Number
CO2MSZ, complete assembly 3015-4601
Battery, Panasonic BR2032, 3 V (for Main PCB Board) 0204-0020*
Coverage ............................... Single zone
Monitoring Distance ............ Up to 500 ft (152.4 m) maximum gas-sample line length
Front Panel .......................... 3 Indicator lights:
MONI
TOR ON
or flashing during warm-up
SYSTEM FAULT
ALARM
– Red light ON when the gas level is above an alarm setting
Alarms .................................. Four SPDT relays are provided (rated 2 A, 250 VAC (inductive) 5 A 250 VAC
(resistive)):
three assigned to ppm level alarms – one assigned to system faults
Internal audible alarm programmable for any of the following conditions:
OFF, Any Alarm, System Fault, Leak Alarm, Spill Alarm, Evacuation Alarm,
Monitor Stopped