Bacharach Snifit 50 Operating Manual

Page 1
Instruction 19-9201
Rev. 3 – June 1999
®
Snif it Model 50
Carbon Monoxide Analyzer
Introduction
Designed for HVAC professionals and utility person­nel, the Snifit Model 50 Analyzer is ideal for measur­ing low levels of CO in ambient air such as in rooms and garages, or around registers, furnaces, stoves, water heaters, and other types of combustion appli­ances. The Snifit samples the surrounding air and shows the detected concentration of CO on its Liquid Crystal Display. The Snifit is not intended to be used in flue gases or in temperatures exceeding 104 °F (40 °C).
Features
• Measures and displays 0 to 1999 ppm CO in room air
• Sensitive CO sensor will last up to 2 years
• Backlight for viewing in dark areas
• Compact pocket size
• Low battery indication
• Manual zero adjust
• Factory calibrated on 100 ppm CO
• Simple field calibration
• Auto power-off after 35 minutes
• Single 9V battery (included) provides at least 1500 hours of operation
621 Hunt Valley Circle, New Kensington, PA 15068
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 800-736-4666
Website: www.bacharach-inc.com • E-mail: help@bacharach-inc.com
Printed in U.S.A. ® Registered trademarks
Bacharach, Inc.
Page 2
Specifications
Gas Monitored ............Carbon Monoxide (CO)
Range ..........................0 to 1999 ppm
Resolution ...................1 ppm
Accuracy......................± (5% of reading + 5 ppm)
Response .....................30 sec. to 90% of final value
Power...........................9V battery
Battery Life ................1500 hours minimum
Temperature Range ....32 to 104 °F (0 to 40 °C)
Case Material .............High impact plastic
Weight.........................0.25 lb (0.11 kg) without battery
Dimensions .................5.5" L x 2.0" W x 1.6" H
(140 x 51 x 41 mm)
Operation
Turning the Snifit On/Off
If not already done, install a 9V battery as described under Battery Installation.
Turn on the instru­ment by pressing its Power On button.
ZER O
ADJUST
LO W
BATTE RY
IN D IC A T IO N
BACKLIGH T
ON/OFF
CARBON
S n if it
M O NO XIDE ANALYZER
m odel 50
Span Adjust
LO BAT
-Power-
Backlight
Off
ppm CO
On
SPAN ADJUST POT.
DISPLAY
POW ER ON
POW ER OFF
Turn it off by pressing Power Off (or the instrument will automatically shut itself off after approxi­mately 35 minutes).
Figure 1. Snifit 50
Zeroing the Sensor
Turn on the Snifit and allow the displayed reading to stabilize before proceeding – approximately 30 sec­onds. Then with the instrument sampling fresh air
- 2 -
Page 3
(air that is free of CO), adjust the Zero Adjust knob at the top of the instrument for a reading of 0 ±1 ppm.
If you’re not sure about the quality of the surrounding air, you can apply a blend of Oxygen/Nitrogen gas to the sensor as described under Calibration.
Checking for CO
Important! Ensure that the sensor grille at the rear of the instrument is unobstructed and open to the atmo­sphere. A quick instrument check can be performed by allowing the smoke of a blown-out match to enter into the sensor grille. This should cause the indicated CO level to increase.
After zeroing the Snifit, simply hold the instrument in the area that you suspect the presence of CO gas. If CO is present, the instrument will indicate the concentration of CO in ppm on its display.
Backlight
Pressing the Backlight button will illuminate the display for a period of 8 minutes, or until it is manu­ally turned off by again pressing the Backlight button.
Overrange
If the CO level exceeds 1999 ppm, the displayed reading is replaced by the number “1”.
Low Battery
When the Snifit’s 9V battery is nearing the end of its useful life,
LO BAT will appear in the upper left
hand corner of the display. Although you can continue using the instrument under this condition, you should replace the battery as soon as possible to ensure accurate CO readings.
- 3 -
LO B AT
Page 4
234
1
Parts Shown:
1. C O Span G as Cylinder, 51-1994
2. R egulator, 03-4318*
3. Tubing, 03-6351*
4. Flow m eter, 06-6163*
5. C alibration C up Assem bly, 23-2156**
* P a rt o f C a lib ra tio n K it 24 -7 05 9 ** Part of Tool / C up Kit 19-3242
PUSH C ALIBRATION CU P OVER SENSOR HOUSING
5
Figure 2. Calibration Equipment Setup
Calibration
Important! For accurate Snifit operation, periodic calibration of its CO sensor is required.
To calibrate the CO sensor, you will need the equip­ment listed under Accessories. Note that the Zero Gas Cylinder is needed only if you’re unsure about the quality of the surrounding air for zeroing purposes.
Calibrate the sensor to a known concentration of CO gas as follows:
1. Assemble the calibration equipment per Figure 2.
2. Zero the instrument as previously described under
Zeroing the Sensor. If necessary, you can use the calibration cup to apply
a blend of Oxygen/Nitrogen gas directly over the sensor by attaching a zero gas cylinder to the regulator and adjusting the regulator knob for a flow rate of 2 SCFH.
3. Push the Calibration Cup over the sensor housing.
4. Attach a CO Span Gas Cylinder to the regulator. Then apply span gas to the sensor by adjusting the regulator knob for a flow rate of 2 SCFH.
- 4 -
Page 5
5. Allow span gas to flow until the displayed reading stabilizes – approx. 3 minutes. Then, using Calibra­tion Tool 06-9453 (supplied in Calibration Tool / Cup Kit 19-3242), turn the Span Adjust potentiom- eter until the displayed reading matches the concentration stamped on the Span Gas Cylinder.
6. Calibration is now complete. Turn off the Regulator and remove the calibration equipment.
Maintenance
Battery Installation
Remove the slotted screw from the rear of instrument; then lift off the front case.
Install a 9V Alkaline battery (Duracell MN1604 or equiv.) in the location shown in Figure 3; then reinstall the front case.
Sensor Replacement
The sensor needs replaced when it can no longer be calibrated to the Span Gas Cylinder value using the Span Adjust potentiometer.
Figure 3. Battery
Location
1. See Figure 4. Remove the screw from the rear of the instrument; then lift off the front case and remove the printed circuit board assembly.
2. Unplug the old sensor and dispose of it properly.
3. Inspect the sensor gasket and filter. Replace an item if is torn, deformed, or contaminated with dirt.
4. Remove the wire-jumper from the pins of the new sensor; plug in the sensor; then reassemble the instrument.
5. Calibrate the new sensor.
- 5 -
Page 6
CO SENSO R
W ARNING!
Remove ju m p e r-w ire fro m p in s b e fo re in s ta lla tio n
Figure 4. Sensor Installation
FILTER
GASKET
Replacement Parts
Item Part No.
CO Sensor .......................................................... 19-7061
Gasket ................................................................ 19-3234
Filter .................................................................. 19-3244
Accessories
Item Part No.
Calibration Kit .................................................. 24-7059
Calibration Tool / Cup Kit................................ 19-3242
Span Gas Cylinder, 100 ppm CO in air........... 51-1994
Zero Gas Cylinder, 20.9% O2 in Nitrogen ....... 51-7131
Bacharach Service Center
Replacement parts and accessories can be obtained by contacting the following Bacharach Service Center:
Bacharach, Inc. 625 Alpha Drive Pittsburgh, PA 15238 Phone: (412) 963-2214 FAX: (412) 963-2606
- 6 -
Loading...