Bacharach IEQ Operating Manual

Indoor Environment Quality Monitor
Instruction 1509-9001
Operation Manual
Rev. 1 - July 2011
IEQ Chek Operation Manual
ii 1509-9001 Rev 1
INSTRUMENT SERIAL NUMBER:
PURCHASE DATE:
PURCHASED FROM:
IMPORTANT NOTE
READ AND UNDERSTAND THIS MANUAL PRIOR TO USING THIS INSTRUMENT. CAREFULLY READ THE WARRANTY POLICY, SERVICE POLICY, NOTICES, DISCLAIMERS AND REVISIONS ON THE FOLLOWING PAGES.
THIS INSTRUMENT SHOULD BE INSPECTED AND CALIBRATED REGULARLY BY A QUALIFIED AND TRAINED TECHNICIAN.
THIS INSTRUMENT HAS NOT BEEN DESIGNED TO BE INTRINSICALLY SAFE IN HAZARDOUS OR EXPLOSION-RATED ENVIRONMENTS. FOR YOUR SAFETY,
DO NOT
USE IT IN AREAS CLASSIFIED AS HAZARDOUS AREAS (E.G.,
EXPLOSION-RATED ENVIRONMENTS).
IEQ Chek Operation Manual
WARRANTY POLICY
BACHARACH, INC. WARRANTS THIS INSTRUMENT, EXCLUDING SENSORS, TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP FOR A PERIOD OF TWO YEARS FROM THE DATE OF PURCHASE BY THE ORIGINAL OWNER. THE SENSORS HAVE A WARRANTY PERIOD OF ONE YEAR FROM THE DATE OF PURCHASE. IF THE PRODUCT SHOULD BECOME DEFECTIVE WITHIN THIS WARRANTY PERIOD, WE WILL REPAIR OR REPLACE IT AT OUR DISCRETION.
THE WARRANTY STATUS MAY BE AFFECTED IF THE INSTRUMENT HAS NOT BEEN USED AND MAINTAINED PER THE INSTRUCTIONS IN THIS MANUAL OR HAS BEEN ABUSED, DAMAGED, OR MODIFIED IN ANY WAY. THIS INSTRUMENT IS ONLY TO BE USED FOR PURPOSES STATED HEREIN. THE MANUFACTURER IS NOT LIABLE FOR AUXILIARY INTERFACED EQUIPMENT OR CONSEQUENTIAL DAMAGE.
DUE TO ONGOING RESEARCH, DEVELOPMENT, AND PRODUCT TESTING, THE MANUFACTURER RESERVES THE RIGHT TO CHANGE SPECIFICATIONS WITHOUT NOTICE. THE INFORMATION CONTAINED HEREIN IS BASED ON DATA CONSIDERED ACCURATE. HOWEVER, NO WARRANTY IS EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THIS DATA.
ALL GOODS MUST BE SHIPPED TO THE MANUFACTURER BY PREPAID FREIGHT. ALL RETURNED GOODS MUST BE PRE-AUTHORIZED BY OBTAINING A RETURN MERCHANDISE AUTHORIZATION (RMA) NUMBER. CONTACT THE MANUFACTURER FOR A NUMBER AND PROCEDURES REQUIRED FOR PRODUCT TRANSPORT.
1509-9001 Rev 1 iii
IEQ Chek Operation Manual
BACHARACH, INC. MAINTAINS AN INSTRUMENT SERVICE FACILITY AT THE FACTORY. SOME BACHARACH DISTRIBUTORS / AGENTS MAY ALSO HAVE REPAIR FACILITIES, HOWEVER, BACHARACH ASSUMES NO LIABILITY FOR SERVICE PERFORMED BY ANYONE OTHER THAN BACHARACH PERSONNEL. REPAIRS ARE WARRANTED FOR 90 DAYS AFTER DATE OF SHIPMENT (SENSORS, PUMPS, FILTERS AND BATTERIES HAVE INDIVIDUAL WARRANTIES). SHOULD YOUR INSTRUMENT REQUIRE NON-WARRANTY REPAIR, YOU MAY CONTACT THE DISTRIBUTOR FROM WHOM IT WAS PURCHASED OR YOU MAY CONTACT BACHARACH DIRECTLY.
IF BACHARACH IS TO DO THE REPAIR WORK, SEND THE INSTRUMENT, PREPAID, TO BACHARACH, INC. AT THE FOLLOWING ADDRESS.
BACHARACH, INC. 621 HUNT VALLEY CIRCLE NEW KENSINGTON, PA 15068 ATTENTION: SERVICE DEPARTMENT
ALWAYS INCLUDE YOUR RMA #, ADDRESS, TELEPHONE NUMBER, CONTACT NAME, SHIPPING/BILLING INFORMATION AND A DESCRIPTION OF THE DEFECT AS YOU PERCEIVE IT. YOU WILL BE CONTACTED WITH A COST ESTIMATE FOR EXPECTED REPAIRS PRIOR TO THE PERFORMANCE OF ANY SERVICE WORK. FOR LIABILITY REASONS, BACHARACH HAS A POLICY OF PERFORMING ALL NEEDED REPAIRS TO RESTORE THE INSTRUMENT TO FULL OPERATING CONDITION.
PRIOR TO SHIPPING EQUIPMENT TO BACHARACH, CONTACT OUR OFFICE FOR AN RMA # (RETURNED MERCHANDISE AUTHORIZATION). ALL RETURNED GOODS MUST BE ACCOMPANIED WITH AN RMA NUMBER.
PACK THE EQUIPMENT WELL (IN ITS ORIGINAL PACKING IF POSSIBLE), AS BACHARACH CANNOT BE HELD RESPONSIBLE FOR ANY DAMAGE INCURRED DURING SHIPPING TO OUR FACILITY.
COPYRIGHTS: THIS MANUAL IS SUBJECT TO COPYRIGHT PROTECTION; ALL RIGHTS ARE RESERVED UNDER INTERNATIONAL AND DOMESTIC COPYRIGHT LAWS. THIS MANUAL MAY NOT BE COPIED OR TRANSLATED, IN WHOLE OR IN PART, IN ANY MANNER OR FORMAT, WITHOUT THE WRITTEN PERMISSION OF BACHARACH, INC.
ALL SOFTWARE WHICH BACHARACH UTILIZES AND/OR DISTRIBUTES, HOLDS A PROPRIETARY INTEREST AND IS ALSO SUBJECT TO COPYRIGHT PROTECTION AND ALL RIGHTS ARE RESERVED. NO PARTY MAY USE OR COPY SUCH SOFTWARE IN ANY MANNER OR FORMAT, EXCEPT TO THE EXTENT THAT BACHARACH GRANTS THEM A LICENSE TO DO SO. IF THIS SOFTWARE IS BEING LOADED ONTO MORE THAN ONE COMPUTER, EXTRA SOFTWARE LICENSES MUST BE PURCHASED.
SERVICE POLICY
NOTICES
iv 1509-9001 Rev 1
DISCLAIMER
IEQ Chek Operation Manual
UNDER NO CIRCUMSTANCES WILL BACHARACH, INC. BE LIABLE FOR ANY CLAIMS, LOSSES, OR DAMAGES RESULTING FROM OR ARISING OUT OF THE REPAIR OR MODIFICATION OF THIS EQUIPMENT BY A PARTY OTHER THAN BACHARACH SERVICE TECHNICIANS, OR BY OPERATION OR USE OF THE EQUIPMENT OTHER THAN IN ACCORDANCE WITH THE PRINTED INSTRUCTIONS CONTAINED WITHIN THIS MANUAL OR IF THE EQUIPMENT HAS BEEN IMPROPERLY MAINTAINED OR SUBJECTED TO NEGLECT OR ACCIDENT. ANY OF THE FORGOING WILL VOID THE WARRANTY.
REVISIONS
BACHARACH, INC. MAKES NO WARRANTY OR REPRESENTATION, EXPRESSED OR IMPLIED INCLUDING ANY W ARRANTY OF MERCHANTABILITY OR FITNESS FOR PURPOSE, WITH RESPECT TO THIS MANUAL. ALL INFORMATION CONTAINED IN THIS MANUAL IS BELIEVED TO BE TRUE AND ACCURATE AT THE TIME OF PRINTING. HOWEVER, BACHARACH RESERVES THE RIGHT TO MAKE CHANGES AT ANY TIME WITHOUT NOTICE. REVISED COPIES OF THIS MANUAL CAN BE OBTAINED BY CONTACTING BACHARACH, INC.
SHOULD YOU DETECT ANY ERRORS OR OMISSIONS IN THIS MANUAL, PLEASE CONTACT THE COMPANY AT THE FOLLOWING ADDRESS.
TOLL FREE: 1-800-736-4666 FAX: 724-334-5001 E-MAIL: help@MyBacharach.com WEBSITE: www.MyBacharach.com
IN NO EVENT WILL BACHARACH, INC., OR ITS OFFICERS OR EMPLOYEES BE LIABLE FOR ANY DIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECT IN ANY MANUAL, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
BACHARACH, INC. 621 HUNT VALLEY CIRCLE NEW KENSINGTON, PA 15068-7074 USA
1509-9001 Rev 1 v
IEQ Chek Operation Manual
TABLE OF CONTENTS
1. OVERVIEW ................................................................................................................ 1
1.1. General Description ............................................................................................ 1
1.2. IEQ Chek™ Shipping Checklist .......................................................................... 2
1.3. Key Features ...................................................................................................... 2
1.4. Sensor Configuration Notes ................................................................................ 2
2 SPECIFICATIONS ..................................................................................................... 3
2.1. General ............................................................................................................... 3
2.2. IQ Chek™ DataLogging Software (Optional) ....................................................... 5
2.3. Available “Plug-and-Play” Sensors ...................................................................... 5
3. KEY COMPONENTS.................................................................................................. 6
3.1. Key Components: Front and Bottom Views ......................................................... 6
3.2. Key Components: Left Side View ........................................................................ 7
3.3. Key Components: Top and Back Views .............................................................. 8
4. INSTRUMENT OPERATION ...................................................................................... 9
4.1. General ............................................................................................................... 9
4.2. Switching the Instrument On/Off ........................................................................10
4.3. Battery and Battery Warnings ............................................................................10
4.4. Main Menu Functional Overview ........................................................................11
5. MENU FUNCTION DETAILS ................................................................................... 12
5.1. Calibrate Menu ..................................................................................................12
5.2. Settings Menu ....................................................................................................13
5.3. Datalog Menu ....................................................................................................15
5.4. Alarm Menu .......................................................................................................17
5.5. Info Menu ..........................................................................................................18
5.6. USB Menu .........................................................................................................19
6. CALIBR ATION ......................................................................................................... 19
6.1. Overview ...........................................................................................................19
6.2. Requirements and Cautions...............................................................................19
6.3. Calibration Process: Zero .................................................................................20
6.4. Calibration Process: 3-Point Span ....................................................................20
6.5. Calibration Process: 1-Point Span ....................................................................21
7. ADDING OR CHANGING PLUG-AND -PLAY SENSORS......................................... 22
8. MAINTENANCE ....................................................................................................... 25
9. IEQ CHEK TROUBLESHOOTING ........................................................................ 25
10. ACCESSORIES ....................................................................................................... 26
10.1. Hard Shell, Foam-Lined Carrying/Transport Case ............................................26
10.2. Extended Operation (Short Term) External Battery ..........................................26
10.3. Extended Operation (Long Term) External Battery ...........................................26
10.4. Standard Gas Sample Probe ...........................................................................26
10.5. IQ ChekDataLogging Kit...............................................................................26
vi 1509-9001 Rev 1
11. DUAL PROBE VERSION ......................................................................................... 27
11.1. Applications Overview ......................................................................................27
11.2. Probes .............................................................................................................28
11.3. Instrument Kits for Dual-Probe Applications .....................................................29
11.4. Application Setup and Configuration ................................................................30
12. MENU ARCHITECTURE .......................................................................................... 31
IEQ Chek Operation Manual
1509-9001 Rev 1 vii
IEQ Chek Operation Manual
viii 1509-9001 Rev 1
the testing of multiple gases. The
wide array of supported sensors. These
up/stabilization periods
the instrument. Any computer running
available USB communications port) can
the optional IQ Chek™
questions, please contact our service department for technical support.
1. OVERVIEW
1.1. General Description
Thank you for purchasing the IEQ Chek™ indoor environmental quality monitor. The IEQ Chek™ is an easy-to-use, portable, information-recording device for monitoring trends that pertain to indoor air quality.
The IEQ Chek™ is a battery-powered, portable, indoor air quality monitor and information recording (data logging) instrument designed for both intermittent and continuous operation. Its intended use is for indoor environments. A basic instrument includes a multi-line LCD alpha numeric display, audible alarm, rechargeable batteries, temperature and relative humidity sensors, and operation manual.
The IEQ Chek™ is equipped with top-mounted temperature and relative humidity sensors to aid in verifying readings providing a comprehensive indication of air quality. Information readings can be written to a plug-in (and removable) flash card to protect your recorded information. The sample rate can be adjusted using the supplied software or through the instrument menu. The instrument may be hand held, stood upright on a flat surface, or fastened to a wall for permanent or semi-permanent use.
The IEQ Chek™ series of instruments are manufactured with plug-and-play, interchangeable sensor modules to allow for
specifications listed in this manual indicate a
sensors may be changed in the field and operated within minutes (accurate readings will require more time as each sensor type has different warm­before they are able to perform to published specifications).
NOTE: Calibration and repair are available at our manufacturing facility and through some of our authorized and trained distributors.
The IEQ Chek™ can reliably record time­based information, which can be downloaded via the USB port located on the right side of
IEQ Chek Operation Manual
Microsoft Windows XP or higher (with an
accommodate software.
If after reading the manual you have any
The user can add up to five additional plug-and-play sensors (electrochemical toxic
internally for a maximum of seven internal sensors. Currently, a large selection of
gas and/or oxygen, infrared CO
, PID or catalytic combustible/flammable gas)
2
sensors is available to choose from. Refer to page 5.
1509-9001 Rev 0 1
Figure 1-1. IEQ Chek™
IEQ Chek Operation Manual
1.2. IEQ Chek Shipping Checklist
This checklist ensures that you have received everything required to run your IEQ Chek. If you do not receive any of the items listed below, contact the factory immediately.
IEQ Chek™ instrument
AC wall adapter (6V @ 850 mA)
Nickel-metal Hydride rechargeable batteries (3 × “AA”)
Calibration adapter (pump versions only)
Operation manual
IQ Chek™ datalogging kit (optional)
Other specified accessories (see original sales order)
1.3. Key Features
Features of the IEQ Chek™ include the following.
Light weight, contoured, and comfortable-to-hold enclosure
Flame rated ABS/Polycarbonate enclosure
Displays all installed sensors simultaneously on backlit LCD display
Multi-function, easy-to-use menu
Operates from nickel metal hydride (NiMH) rechargeable batteries (standard),
alkaline batteries, or continuously from the plug-in power adapter
Hangs on a wall (rear slot), mounts on a camera tripod, or sits flat on a desk
Accommodates a standard cable lock for securing the instrument in place
Samples air with internal sample draw pump (pump version only)
Accommodates electrochemical, catalytic, PID, or infrared sensor types
16 different plug-and-play sensor choices
Logs information and events to flash card (with optional datalogging kit)
USB data downloading and auxiliary uploading from remote probes
Optional IQ Check software compatible with Windows XP, Vista, and 7.
1.4. Sensor Configuration Notes
It is important to note that although the IEQ Chek™ will accommodate seven sensors (and five of those sensor locations are plug and play for gas sensors), there are sensor placement considerations.
Either of the two front sensor locations will accommodate an infrared CO2 sensor, PID sensor, or catalytic combustible (flammable) gas sensor.
Of the five plug-and-play sensor locations, three are for electrochemical sensors. If the user selects an oxygen (O (rear center position, determined by looking at the front of the instrument). If the user selects a nitric oxide (NO), requiring electronic bias within the circuit, it must be fitted to one specific sensor location (left rear position, determined by looking at the front of the instrument). Two sensors requiring bias cannot be fitted in the same instrument at the same time. Reference the photos on pages 23 and 24 for specific sensor locations.
) sensor, it must be fitted to a specific sensor location
2
2 1509-9001 Rev 1
2 SPECIFICATIONS
2.1. General
Categories/Subcategories Descriptions
Dimensions
Weight
Covers
L x W x D
7.75” x 3.87” x 3.06” (19.7 cm x 9.8 cm x 7.8 cm)
20 ounces (567 grams) including batteries
Sensor cover (removable; top of enclosure) Battery cover (removable; bottom of enclosure)
Threaded insert (bottom of enclosure) to
Enclosure
Mounting Options
accommodate a camera tripod Teardrop slot at the upper rear to allow instrument to
hang on a wall Flat base for sitting on a flat table or desktop
Power
User Interface
Monitoring Mode
Environ­mental
CE Certification Standards
Security
Construction
Flammability
Standard
Option
Protection
Continuous
Display
Buttons
Audio
Internal
Option
Temperature
Relative Humidity
Safety
Emissions
Immunity
Slot for cable lock (lower right side)
Custom molded, rugged, contoured, ABS/Polycarbonate
UL94 rating
Three rechargeable NiMH batteries
Three user-supplied AA alkaline batteries
In-circuit protection diode (if the batteries are accidently inserted with reversed polarity)
Plug-in, 6V DC, class-III, 2A DC max. wall adapter
Back-lit, multi-line LCD alphanumeric display
3 tactile push-buttons for user access
Openings at front top left for audible sounder
Automatic sample pump for “active” sampling of target environment
Diffusion monitoring version
Operating: 5° to 50° C (41° to 122° F) Storage: -20° to 60° C (-4° to 140° F)
Operating: 0 to 99% RH non-condensing Storage: 0 to 99% RH non-condensing
I.S. EN 61010-1:2001 (Ed.2)
EN 50270:2006
EN 50270:2006
IEQ Chek Operation Manual
1509-9001 Rev 1 3
IEQ Chek Operation Manual
7.75
3.87″ (9.8 cm)
Figure 2-1. IEQ Chek™ Dimensions
4 1509-9001 Rev 1