Product Leadership • Training • Service • Reliability
A
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in
material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications.
Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at
Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and
shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of
the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery
of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and
Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the
election of Bacharach, to the replacement of this Product or parts thereof that are defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY
AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II)
ANY OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT
ARISING FROM BACHARACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be
limited to those provided herein to the exclusion of any and all other remedies including, without limitation incidental or consequential damages. No agreement varying or extending the foregoing warranties,
remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly
authorized officer of Bacharach.
Register Your Warranty by Visiting
www.mybacharach.com
Declaration of Conformity
Manufacturer’s name: Bacharach, Inc.
Manufacturer’s address: 621 Hunt Valley Circle
New Kensington, PA 15068-7074
European operations: Bacharach Europe
Sovereign House, Queensway
Leamington Spa
Warwickshire CV31 3JR
United Kingdom
Product name: Hi Flow Sampler
conforms to the following specifications:
EMC Directive 61326:1998
IEC 61000-4-3: Radiated Immunity
IEC 61000-4-2
Notice:
Product improvements and enhancements are continuous; therefore the specifications and
information contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language
without the prior written consent of Bacharach, Inc.
7.3 Service Centers......................................................................................................66
ii
Instruction 0055-9017
Hi Flow Sampler Introduction
1 Introduction
1.1 General Description
The Hi Flow Sampler is portable, intrinsically safe, battery-powered
instrument designed to determine the rate of gas leakage around various
pipefittings, valve packings, and compressor seals found in natural gas
transmission, storage, and compressor facilities.
A component’s leak rate is measured by sampling at a high flow rate so
as to capture all the gas leaking from the component along with a certain
amount of surrounding air. By accurately measuring the flow rate of the
sampling stream and the natural gas concentration within that stream,
the gas leak rate can be calculated using Equation 1. The instrument
automatically compensates for the different specific gravity values of air
and natural gas, thus assuring accurate flow rate calculations.
Instruction 0055-9017
Leak = Flow x (Gas
sample
– Gas
background
where:
Leak = rate of gas leakage from source (cfm)
Flow = sample flow rate (cfm)
Gas
Gas
= concentration of gas from leak source (%)
sample
background
= background gas concentration (%)
To ensure that the instrument is capturing all the gas that is escaping
from the component, two measurements are performed at two different
flow rates. The first measurement is taken at the highest possible flow
rate, followed by a second measurement at a flow rate that is
approximately 70–80% of the first. If the two calculated leak rates are
within 10% of each other, then it can be assumed that all gas has been
captured during the test.
For those mathematicians who are wondering how both calculations can
be the same with different flow rates, note that the measured CH
the lower flow rate will be higher due to less air being drawn in with the
sample, thus balancing out the lower flow rate.
The instrument is packaged inside a backpack, thus leaving the
operator’s hands free for climbing ladders or descending into manholes.
The instrument is controlled by a handheld unit consisting of an LCD
and a 4-key control pad, which is attached to the main unit via a 6 foot
coiled cord.
The gas sample is drawn into the unit through a flexible 1.5 inch I.D.
hose. Various attachments connected to the end of the sampling hose
provide the means of capturing all the gas that is leaking from the
component under test.
The main unit consists of an intrinsically safe, high-flow blower that
pulls air from around the component being tested through a flexible hose
and into a gas manifold located inside the unit. The sample is first
passed through a venturi restrictor where the measured pressure
differential is used to calculate the sample’s actual flow rate. Next, a
portion of the sample is drawn from the manifold and directed to a
combustibles sensor that measures the sample’s CH
) x 10–2 Eq. 1
level at
4
concentration in
4
1
Introduction Hi Flow Sampler
the range of 0.05 to 100% gas by volume. A second identical
combustibles sensor channel measures the background CH
level within
4
the vicinity of the leaking component. The final element in the sampling
system is a blower that exhausts the gas sample back into the
atmosphere away from the sampling area.
The measured flow rate and the measured CH
levels (both leak and
4
background levels) are used to calculate the leak rate of the component
being tested, with all measured and calculated values being displayed on
the hand-held control unit.
1.2 Material Supplied
High Flow Sampler Unit
Backpack
6 Foot, 1.5 Inch Hose Assembly
Battery Pack (Qty. 2)
Battery Charger w/ Power Supply
Control Unit with LCD and 4-button keypad
Assortment of Attachments (for complete capture of the gas leak)
Flange Strap 34", 80", and 137" Capture Bag 36" x 36" Beveled Nozzle 6.5" and 24" Bag Nozzle w/ box of disposable bags Bellows Tool Claw Tool 3-Port Background Gas Sampler
1.3 Why Use the Hi Flow Sampler?
Allows ranking of gas leaks in leak-rate order
Allows users to determine if the potential savings in lost gas by
repairing a leak
Allows users to determine payback periods for particularly
expensive repairs
Allows more cost-effective use of your maintenance budget
Identifies problem areas
Determines baseline leakage conditions
Provides permanent record of leakage and repair
Greenhouse gas credits
Pays for itself in recovered gas
2
Instruction 0055-9017
Hi Flow Sampler Introduction
1.4 Control Unit Description
The control unit consists of an 8 line by 20 character LCD and the four
pushbuttons.
Figure 1-1. Control Unit
1.5 Top Panel Connections & Controls
The top panel contains the following items:
SCANNER – Bar code scanner input
COMPUTER – RS232 communications port for downloading stored
test data to an external computer
Leak Gas Inlet – Main hose connection used to collect gas from
leak source
BACKGROUND Gas Inlet – Gas hose connection used to sample
background levels of CH
GAS Inlet – Calibration gas connection
ON/OFF Switch – Turns unit ON and OFF
Figure 1-2. Top Panel Connections
4
Instruction 0055-9017
3
Introduction Hi Flow Sampler
Notes:
4
Instruction 0055-9017
Hi Flow Sampler Technical Data
2 Technical Data
Information Displayed .......... - Date and Time
- Battery voltage
- Sampling flow rate in cfm
- Background gas concentration in ppm or % by volume
- Leak concentration in ppm or % by volume
- Leak rate in cfm
- Percent difference between leak rate measurements #1 and #2
- Latitude, Longitude, Altitude, and Time of Fix, when the
optional GPS receiver is attached
Display................................. 8 line by 20 character LCD
Pushbutton Controls ............ - I/O
-
-
- ESC (Escape)
Communication.................... Three DB9 connectors providing serial data transfer at 115200
baud to a personal computer, bar code reader, and GPS receiver
Catalytic oxidation......... 0 to 5% by volume methane
Thermal conductivity ..... 5 to 100% by volume methane
Accuracy............................±5% of reading or 0.02 % methane, whichever is greater
Battery:
Type ..................................Intrinsically Safe NiMH rechargeable pack
Voltage ..............................4.8 V, max.
Recharge Time ...................8 to 10 hours
Run Time........................... >4.5 hours continuous operation @ 20 °C (68 °F)
Memory ................................Stores up to 1000 individual test parameters
Dimensions ..........................18L x 12W x 7H inches (457 x 305 x178 mm)
Weight ..................................20 lbs (0.9 kg)
Agency Approvals .................Designed to be intrinsically safe for use in hazardous locations
Class I, Division 1, Groups A, B, C & D in North America
CAN/CSA-C22.2 No. 157 – (June 1992)
ANSI (June 27, 2002)/UL913-2002
Intrinsically Safe Apparatus and Associated Apparatus for use in
Class I, Division 1, Hazardous (Classified) Locations
CE Mark – See this manual’s inside front cover for Declaration of
Conformity statement
6
Instruction 0055-9017
Hi Flow Sampler Operation
3 Operation
3.1 Operating Precautions
(Please Read)
To ensure that the gas sensors are properly zeroed at startup, the Hi
Flow Sampler must be turned ON in clean air (free of combustible gases
or vapors).
Calibrate the instrument about every 30 days to assure its accuracy. The
frequency of calibration, however, may change depending on how often
the instrument is used and the amount of gas that was sampled. Create
a maintenance log to keep track of when an instrument was calibrated,
and then use this log to help develop a calibration schedule. For
example, an instrument that fails its calibration after only being
calibrated a single time should be calibrated more often.
Important: Avoid sampling leaded gasoline vapors, or gases or vapors that
contain silicones or sulfur compounds. Tetraethyl lead, silicones, and
sulfur compounds can form contaminating compounds on the sensor
element (poison the sensor), with resulting loss in sensitivity.
Always purge the instrument with clean air after testing. This removes
combustibles from the sensor chambers and prolongs the life of the
sensors.
3.2 Battery Connection
WARNING: Explosion hazard! DO NOT connect or disconnect the battery in
an unsafe atmosphere.
The instrument’s power cord is located inside the backpack’s right-side
pocket. Insert a battery pack into this pocket and connect it to the power
cord.
Figure 3-1. Battery Pack Connection
Instruction 0055-9017
7
Operation Hi Flow Sampler
3.3 Sampling Hose & Attachment Connections
Connect the sampling hose to the instrument’s leak-gas inlet by
clamping the hose to the inlet as shown in Figure 3-2. Also connect the
background gas sampling hose to the unit’s BACKGROUND inlet.
Figure 3-2. Hose and Attachment Connections
The standard sampling hose is 6 foot long. If a longer length is needed,
the optional 12 foot hose (P/N 0055-0287) can be ordered, or additional
sections of hose can be added by simply clamping them together along
with connecting together the background-gas hose.
Choose an attachment (refer to Section 3.5 Attachments) that will capture
the entire leak and attach it to the end of the hose.
8
Instruction 0055-9017
Hi Flow Sampler Operation
3.4 Grounding
WARNING: To prevent the possibility of a static discharge, the Hi Flow
Sampler should be grounded while conducting a leak test.
Attach the unit’s grounding clamp to the nearest earth ground.
Tip: When not in use, the grounding clamp can be stored by attaching it to
one of the shoulder straps.
Figure 3-3. Grounding Clamp
Instruction 0055-9017
9
Operation Hi Flow Sampler
3.5 Attachments
A variety of attachments are available for the Hi Flow Sampler that
simplifies the task of capturing the entire gas leak. Choose the
appropriate attachment for the type of component being tested from the
ones listed below, and then connect it to the end of the instrument’s
main sampling hose.
3.5.1 Flange Strap (34", 80" & 137")
Flange leaks can be captured by wrapping this attachment around the
edge of the flange, and securing in place using the two Velcro strips.
Attach the Hi Flow Sampler’s main sampling hose to the flange strap by
simply pushing the end of the hose into the attachment’s hose coupling.
A click should be heard when the hose has been successfully connected.
Release the hose by pushing in the coupling’s release mechanism.
3.5.2 Beveled Nozzle Tool (6.5" & 24")
For simple pin-hole type leaks, position this attachment directly over the
leak source.
10
Instruction 0055-9017
Hi Flow Sampler Operation
3.5.3 Capture Bag
This reusable capture bag measures 36" x 36" and can be used to
completely surround a component that may have multiple leak sources,
or contains a leak source that is difficult to pinpoint or reach.
Attach the Hi Flow Sampler’s main sampling hose to the bag by simply
pushing the end of the hose into the bag’s hose coupling. A click should
be heard when the hose has been successfully connected. Release the
hose by pushing in the coupling’s release mechanism.
Use the draw strings to close up the end of the bag. DO NOT, however,
completely close off the bag – allow air to flow into the bag so as to
replace the volume of gas and air being removed during the test.
3.5.4 Disposable Bags with Nozzle
These disposable capture bags measure 38" x 60" and can be used to
completely surround a component that may have multiple leak sources,
or contains a leak source that is difficult to pinpoint or reach.
When using a disposable bag to surround the leak source, insert the bag
nozzle into the bag to take a gas sample.
DO NOT completely close off the bag – allow air to flow into the bag so as
to replace the volume of gas and air being removed during the test.
3.5.5 Bellows Tool
Instruction 0055-9017
11
Operation Hi Flow Sampler
This tool is useful in capturing leaks in valve stems and small fittings.
3.5.6 Claw Tool
Use this tool when the leak source is spread out over a narrow area,
such as a crack in a pipe, square flanges, and cylinder heads.
3.5.7 3-Port Background Gas Sampler
Attaches to the end of the background gas sample hose to give better
average background gas readings.
3.6 Turning ON the Hi Flow Sampler
12
Instruction 0055-9017
Hi Flow Sampler Operation
With the Hi Flow Sampler located in an area containing clean air (no
combustible gases or vapors), turn ON the instrument by first pulling up
the ON/OFF switch and then moving it to its ON position.
WARNING: Turning ON the instrument in air contaminated with
combustible gas will cause false readings to occur. Use one of Bacharach’s
portable gas detection instruments to sample the surrounding area for
levels of combustible gas before turning ON the instrument.
An initial banner screen is first displayed for 3 seconds that shows the
instrument’s name, software version, and the software’s Month, Day,
Year, and Time of creation.
B a c h a r a c h , I n c .
H i F l o w S a m p l e r
v e r 1 . 4 0
N o v 1 3 2 0 0 4 1 1 : 2 0
After the banner screen is displayed, the sensors are automatically
zeroed to ambient conditions. Zeroing time is dependent on the
temperature of the sensors, and whether there is residual gas present
inside the sensor chambers. If a sensor fails to zero, a message appears
at the bottom of the LCD describing the cause of the failure. Refer to
Section 6 Troubleshooting for possible causes and remedies for this
failure.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
Z e r o i n g S e n s o r s
P l e a s e W a i t . . .
[ 0 0 0 1 ] ( M )S - -
After the sensors are zeroed, either the instrument’s Basic or Expanded
Main Screen appears (refer to Section 3.8). The example below shows the
Basic Main Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S t a r t
[ 0 0 0 1 ] ( M )S - -
T h e a r r o w s y m b o l s o n t h e l e f t i n d i c a t e t h a t
t h e r e a r e m o r e i t e m s t o v i e w a b o v e a n d / o r
b e l o w t h e c u r r e n t d i s p l a y .
P r e s s k e y s t o v i e w a d d i t i o n a l i t e m s .
3.7 Turning OFF the Hi Flow Sampler
Turn OFF the unit by simply pulling up the ON/OFF switch and moving
it to its OFF position.
3.8 Selecting the Basic or Expanded Menu Mode
Instruction 0055-9017
13
Operation Hi Flow Sampler
Select the Basic or Expanded Menu Mode as follows:
1. From either the Basic or Expanded Main Screen, select Menu >
Menu Mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
[ 0 0 0 1 ] ( M )S - -
2. Highlight the desired menu mode, and then press the I/O key to
select that mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
B a s i c M e n u
E x p a n d e d M e n u
[ 0 0 0 1 ] ( M )S - -
3. Press ESC to return to the Main Screen.
Refer to Sections 3.8.1 and 3.8.2 for detailed descriptions of the
information contained in the Basic and Expanded Main Screens.
14
Instruction 0055-9017
Hi Flow Sampler Operation
3.8.1 Basic Main Screen
Date and Time: Date displayed as
mm/dd/yy, while time is displayed in a 12hour format. The date and time are set per
Section 5.8.
Btry(V): Battery voltage
Flow(lpm): Sample flow rate in either
liters/min or cu.ft./min.
: GPS (Global Positioning System) icon
appears when instrument detects that a GPS
receiver is connected to the unit’s GPS port
: Barcode icon appears when the instrument
detects that a barcode reader wand is
connected to the unit’s SCANNER port
: Battery status icon (refer to Section 3.10)
Back(%): Measured background gas level in
[0001]: Current Test ID number
either percent-by-volume or ppm
(A)or (M): Automatic or Manual mode (refer to
Leak(%): Measured natural gas leak
Section 3.14)
concentration at the current sample flow rate
in either percent-by-volume or ppm
Leak(lpm): Calculated leak rate in either
liters/min or cu.ft./min.*
Speed(LO<––––|>HI): Blower speed indicator.
Start / Stop: Select to Start and Stop a test**
S--: Standby
-1-: Measurement #1 (maximum flow rate)
--2: Measurement #2 (reduced flow rate)
Message Line: Various messages appear
during operation describing the current
function being performed, or errors that have
occurred (refer to Section 3.19)
Menu: Select to display the first of several
submenus (refer to Section 3.9)**
* “Not Avail.” is displayed when a value cannot be calculated from available information.
** Selections are made by using the keys to highlight the desired function, and then
pressing the I/Okey to activate that function.
Instruction 0055-9017
15
Operation Hi Flow Sampler
3.8.2 Expanded Main Screen
Date and Time: Date displayed as
mm/dd/yy, while time is displayed in a 12hour format. The date and time are set per
Section 5.8.
Save: Select to save all current measurement
parameters in memory**
Start / Stop: Select to Start and Stop a test**
Lat(°): Latitude in degrees*
Lon(°): Longitude in degrees*
Alt(ft): Altitude in feet*
Fix(sec): Time Of Fix between –1 and
−10 seconds (time lag between the received
GPS signal and the displayed values of
latitude, longitude, and altitude)*
Btry(V): Battery voltage
#1 – #2(%): Difference between test measure-
ments #1 and #2 in percent*
Flow(lpm): Sample flow rate in either
liters/min or cu.ft./min.
Back(%): Measured background gas level in
either percent-by-volume or ppm
Leak(%): Measured natural gas leak
concentration at the current sample flow rate
in either percent-by-volume or ppm
Leak(lpm): Calculated leak rate in either
liters/min or cu.ft./min.*
Menu: Select to display the first of several
submenus (refer to Section 3.9)**
: GPS (Global Positioning System) icon
appears when instrument detects that a GPS
receiver is connected to the unit’s GPS port
: Barcode icon appears when the instrument
detects that a barcode reader wand is
connected to the unit’s SCANNER port
: Battery status icon (refer to Section 3.10)
[0001]: Current Test ID number
(A)or (M): Automatic or Manual mode (refer to
Section 3.14)
S--: Standby
-1-: Measurement #1 (maximum flow rate)
--2: Measurement #2 (reduced flow rate)
Message Line: Various messages appear
during operation describing the current
function being performed, or errors that have
occurred (refer to Section 3.19)
Speed(LO<––––|>HI): Blower speed indicator.
* “Not Avail.” is displayed when a value cannot be calculated from available information.
** Selections are made by using the keys to highlight the desired function, and then
pressing the I/Okey to activate that function.
16
Instruction 0055-9017
Hi Flow Sampler Operation
3.9 Menu System
Selecting Menu from either the Basic or Expanded Main Screen provides
the user access to a system of submenus as shown in Figures 3-4 and 3-
5.
The items that are listed in the submenus depend on whether the
instrument is in its Basic or Expanded menu mode.
The functions listed are described in their associated sections of this
manual.
Figure 3-4. Basic Menu System
Figure 3-5. Expanded Menu System
Instruction 0055-9017
17
Operation Hi Flow Sampler
3.10 Battery Status
Battery status is displayed as a single battery icon shown in the lowerleft hand portion of the LCD. Since the voltage curve of the battery pack
drops off rapidly towards the end of its charge, the Battery status icon is
useful only in showing when the battery pack is nearing depletion. It
does not show the pack’s remaining percentage-of-charge.
4.0 volts and above
(Shutdown imminent, less than 5 minutes of operating time remain)
3.11 Calibration Menu
Selecting Calibration from the Menu System allows the user to either:
1) calibrate the sensors, 2) verify calibration, 3) view dates on which
calibrations were performed, 3) view amplifier gain values (in
hexadecimal) of each sensor circuit, or 4) erase calibration.
A detailed description of each calibration function is provided in
Section 4 Calibration.
3.12 Set Flow Units – lpm or cfm
Selecting Set Flow Units from the Menu System allows the user to select
whether the flow rate is displayed in either lpm (liters per minute), or cfm
(cubic feet per minute).
Select the desired flow rate unit of measure as follows:
1. From the Main Screen, select Menu > Set Flow Units.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
[ 0 0 0 1 ] ( M )S - -
2. Highlight the desired units of measure and then press the I/O key.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
c f m ( c u . f t . / m i n . )
l p m ( l i t e r s / m i n . )
[ 0 0 0 1 ] ( M )S - -
3.Press ESC to return to the Main Screen.
18
Instruction 0055-9017
Hi Flow Sampler Operation
3.13 Menu Mode – Basic or Expanded
Selecting Menu Mode from the Menu System allows the user to select
whether only the most basic amount of information is displayed on the
Main Screen (Basic Menu mode), or if all available information is
displayed (Expanded Menu mode).
Refer to Section 3.8 Selecting the Basic or Expanded Menu Mode for
detailed information on how to select the desired menu mode, and learn
about what type of information is displayed in each mode.
3.14 Operating Mode – Automatic or Manual
Note: The Operating Mode can only be selected if operating in the Expanded
Menu Mode. Refer to Section 3.8.
Selecting Operating Mode from the Expanded Menu System allows the
user to set the instrument’s operating mode to either Automatic or
Manual.
3.14.1 Automatic 2-Stage Measurement Mode
In the Automatic 2-Stage Mode the unit performs a leak rate
measurement at first a high flow rate for 1 minute, and then
automatically shifts to a lower flow rate and performs another
measurement for another minute. The final measured and calculated
values are automatically saved in memory. This mode is used to measure
leaks that are easily identifiable and captured.
Select the automatic measurement mode as follows:
1. From the Expanded Main Screen, select Menu > Operating Mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
O p e r a t i n g M o d e
A c c e s s R e c o r d s
[ 0 0 0 1 ] ( M )S - -
2. Highlight Automatic 2-Stage, and then press the I/O key.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
A u t o m a t i c 2 - S t a g e
M a n u a l 2 - S t a g e
M a n u a l 1 - S t a g e
3.Press ESC to return to the Main Screen. Observe that (A) should
Instruction 0055-9017
[ 0 0 0 5 ] ( M )S - -
appear at the bottom of the screen
19
Operation Hi Flow Sampler
3.14.2 Manual Measurement Mode
3.14.2.1 Manual 2-Stage Mode
In the Manual 2-Stage mode the user starts a measurement at a high
flow rate, waits for a stable leak rate reading, and then manually saves
the readings using the Save function. The unit then automatically lowers
the flow rate, at which time the user makes a second leak rate
measurement. After waiting for the leak rate reading to once again
stabilize, the user then manually saves all final measured and calculated
measurements in memory. This mode is used when more control of the
leak rate measurement process is required.
Select Manual 2-Stage operation as follows:
1. From the Expanded Main Screen, select Menu > Operating Mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
O p e r a t i n g M o d e
A c c e s s R e c o r d s
[ 0 0 0 1 ] ( M )S - -
2. Highlight Manual 2-Stage, and then press the I/O key.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
A u t o m a t i c 2 - S t a g e
M a n u a l 2 - S t a g e
M a n u a l 1 - S t a g e
[ 0 0 0 5 ] ( M )S - -
3. Press ESC to return to the Main Screen. Observe that (M) should
appear at the bottom of the screen
20
Instruction 0055-9017
Hi Flow Sampler Operation
3.14.2.2 Manual 1-Stage Mode
In the Manual 1-Stage mode the user starts a measurement at a high
flow rate and then waits for a stable leak rate reading to occur. During
this time the flow rate can be changed using the Speed function on the
Main Screen. After a stable leak rate reading is obtained, the user can
manually save all final measured and calculated values in memory using
the Save function on the Expanded Main Screen. This mode is used
when more control of the leak rate measurement process is required.
Select Manual 1-Stage operation as follows:
1. From the Expanded Main Screen, select Menu > Operating Mode.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
C a l i b r a t i o n
S e t F l o w U n i t s
M e n u M o d e
O p e r a t i n g M o d e
A c c e s s R e c o r d s
[ 0 0 0 1 ] ( M )S - -
2. Highlight Manual 1-Stage, and then press the I/O key.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
A u t o m a t i c 2 - S t a g e
M a n u a l 2 - S t a g e
M a n u a l 1 - S t a g e
[ 0 0 0 5 ] ( M )S - -
3. Press ESC to return to the Main Screen. Observe that (M) should
appear at the bottom of the screen
Instruction 0055-9017
21
Operation Hi Flow Sampler
3.15 Access Records
Note: Send Records can only be selected if operating in the Expanded Menu
Mode. Refer to Section 3.8.
Selecting AccessRecords from the Expanded Menu System allows a
user to either: 1) send all saved records to a personal computer, 2) view
the records on the unit’s LCD screen, or 3) erase all saved records.
3.15.1 Send Records
All the measured and calculated values that were saved in memory can
be downloaded to a personal computer as comma-delimited ASCII text,
which can later be imported into most spreadsheet programs for
analysis.
The following describes how to download saved records to a computer
that is running Windows
program. If a different operating system and/or communications
program is being used, then please consult the appropriate instruction
manuals for those products.
Note: Send Records can only be selected if operating in the Expanded Menu
Mode. Refer to Section 3.8.
1. Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9
male to DB9 female) between the computer’s COM port and the Hi
Flow’s COMPUTER connector. (For computers having a 25-pin or
USB COM port, the user must supply the appropriate adapter, or use
a serial cable with appropriate connectors on each end.)
2. On the computer, start HyperTerminal by doing one of the following:
- If HyperTerminal has already been set up to communicate with the
Hi Flow Sampler, then select Start > Programs > Accessories >
Communications > HyperTerminal, and double-click the filename
or icon that has been associated with the instrument.
- If HyperTerminal has never been run in association with the Hi
Flow Sampler, then a new HyperTerminal connection needs to be
made as follows:
a. Select Start > Run. In the Run box type “hypertrm.exe” and
click OK. The New Connection dialog box should appear.
b. Type in a New Connection Name such as “HiFlow”. Click OK to
select the default icon (if desired, select a different icon before
clicking OK). The Connect To dialog box should appear.
c. Click the Connect using drop-down menu and choose the COM
port to which the Hi Flow is connected. Then click OK to display
the COMProperties dialog box for that COM port.
®
HyperTerminal as the communications
22
Instruction 0055-9017
Hi Flow Sampler Operation
d. Set Port Settings to:
- Bits per second: 115200
- Data bits: 8
- Parity: None
- Stop bits: 1
- Flow control: None
Then click OK to display HyperTerminal’s main window.
e. Click File > Save. This creates a file with an .ht extension using
the name that was entered in Step b.
3. After HyperTerminal is running: click Transfer > Capture Text >
Start to accept the default drive\directory\filename to which all
received data will be stored as an ASCII text file.
Note: The captured text file will be stored on the hard drive in the folder
shown in the File dialog box. If desired, click Browse to select a different
drive, directory and filename. The next time data is captured, it will be
stored under the same location and filename.
Tip: Changing the filename extension to “CSV” (comma separated
variable) will allow the file to be directly opened in the spreadsheet
program. For example, double clicking the filename CAPTURE.CSV in
Windows Explorer will automatically start Microsoft Excel and load the file.
4. From the High Flow Sampler’s Expanded Main Screen, start the
download process by selecting Menu > Access Records > Send Records.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e n d R e c o r d s
V i e w R e c o r d s
E r a s e R e c o r d s
Instruction 0055-9017
[ 0 0 0 1 ] ( M )
S - -
Observe that the Hi Flow’s LCD displays the message “Sending record
x of y @ 115.2kbps,” where “x” is the current record being sent and
“y” is the number of records in memory. Also observe that
HyperTerminal displays the data as its being received.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e n d i n g r e c o r d
x o f y
@ 1 1 5 . 2 k b p s
[ 0 0 0 1 ] ( M )
S - -
5. After all data has been transmitted, stop HyperTerminal from
capturing data by clicking Transfer > Capture Text > Stop.
6. The saved data can now be viewed in a spreadsheet for analysis as
described in Section 3.16 Importing Saved Data into a Spreadsheet.
23
Operation Hi Flow Sampler
3.15.2 View Records
View the records stored in memory on the unit’s LCD screen as follows:
1. Access the Review Record Screen from the Expanded Main Screen by
selecting Menu > Access Records > View Records. This screen
allows the user to view a short summary of each record stored in
memory. If no records are in memory, then the message “No Records
Saved” briefly appears in the message line.
The first two lines in this screen contain the Test ID information as
entered per Section 3.17Access Test IDs.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
T e s t I D L i n e 1
T e s t I D L i n e 2
L e a k ( l p m ) x . x x
# 1 - # 2 ( % ) y y . y
[ 0 0 0 1 ] ( M )
Where: x.xx is the recorded leak rate in ether lpm or cfm
yy.y is the percent difference between measurements #1 and
#2
2. Use the keys to scroll through the saved records.
3. Press ESC three times to return to the Main Screen.
3.15.3 Erase Records
Erase all records from memory as follows:
1. From the Expanded Main Screen, select Menu > Access Records >
Erase Records.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e n d R e c o r d s
V i e w R e c o r d s
E r a s e R e c o r d s
[ 0 0 0 1 ] ( M )
2. Confirm or cancel the erase process by selecting Yes or No.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
S - -
S - -
E r a s e a l l r e c o r d s ?
Y e s
N o
[ 0 0 0 1 ] ( A )
3. Observe the message “Erasing all records” appears at the bottom of
the display during the erase process.
24
S - -
Instruction 0055-9017
Hi Flow Sampler Operation
3.16 Importing Saved Data into a Spreadsheet
Test records that were downloaded to a personal computer per Section
3.15.1 can be imported into spreadsheet programs that are capable of
importing comma-delimited files. Note that each data record consists of
24 fields, some of which may be blank because of options not being
installed.
TABLE 3-1. COMMA-DELIMITED FIELDS
Field
1 Record# Test Record Number
2 Inst.Serial# Instrument’s Serial Number
3 Date#1(MM/DD/YY) Date (Measurement #1)
4 Time#1(HH:MM:SS) Time (Measurement #1)
5 Btry#1(V) Battery Voltage (Measurement #1)
6 Flow#1(cfm) Sample Flow Rate (Measurement #1) in cfm
7 Back#1(%) Background Gas Level (Measurement #1) in %
8 Leak#1(%) Sample Leak Rate (Measurement #1) in %
9 Leak#1(cfm) Leak Rate of Component Under Test (Measurement #1) in
10 Date#2(MM/DD/YY) Date (Measurement #2)
11 Time #2(HH:MM:SS) Time (Measurement #2)
12 Btry#2(V) Battery Voltage (Measurement #2)
13 Flow#2(cfm) Sample Flow Rate (Measurement #2) in cfm
14 Back#2(%) Background Gas Level (Measurement #2) in %
15 Leak#2(%) Sample Leak Rate (Measurement #2) in %
16 Leak#2(cfm) Leak Rate of Component Under Test (Measurement #2) in
17 Leak#1-#2(%) Percent Difference Between Leak Measurements #1 and #2
18 Error Codes Error Codes identifying problems that occurred during the
19 Barcode# Barcode Type (UPC, EAN, Code 128, Cadabar)
20 Barcode Symbology Barcode Number Scanned
21 GPS Latitude(deg) Latitude of Instrument in Degrees
22 GPS Longitude(deg) Longitude of Instrument in Degrees
23 GPS Altitude(ft) Altitude of Instrument in Meters
24 Test Description Test ID Information (refer to Section 3.15 Access Test IDs)
Instruction 0055-9017
Label in Column
Heading
Description
cfm
cfm
test (refer to Section 6.3 Error Codes in Saved Test Records)
–3 Test Location 1 Operator John Doe
–6 Test Location 2 Operator John Doe
* Line 1: Column Headings
Line 2: Record 1, Typical Test Automatic Mode
Line 3: Record 2, Typical Test Manual Mode
The following describes how to generate a spreadsheet from a comma-
®
delimited text file using Microsoft
Excel 2000. If a different spreadsheet
program is being used, then please refer to its instruction manual for
information on how to import comma-delimited text files.
Tip: If the text file was saved with the filename extension “CSV” as
described in Section 3.13.1, a spreadsheet will automatically be created
when that file is opened in Microsoft Excel.
1. Start Microsoft Excel.
2. Click File, then click Open to display the Open dialog box.
3. Change the Files of type: to TextFiles. Then navigate to the
directory containing the text file to be imported.
4. Double-click the desired filename to display the Text Import Wizard –
Step 1 of 3 dialog box.
5. Select the Delimited radio button; then click Next to display the Text
Import Wizard – Step 2 of 3 dialog box.
6. Under Delimiters, select the Comma check box. Then click Next to
display the Text Import Wizard – Step 3 of 3 dialog box.
7. Click Finish to create the spreadsheet.
26
Instruction 0055-9017
Hi Flow Sampler Operation
3.17 Access Test IDs
Note: Access Test IDs can only be selected if operating in the Expanded
Menu Mode. Refer to Section 3.8.
Saved test records can be identified (e.g., location of test, type of test,
name of operator, etc.) by manually entering up to two lines of text, with
each line containing a maximum of 20 alphanumerical characters. This
information will be associated with all succeeding tests, and will appear
when the test record is either viewed or printed. The chosen Test ID
remains in effect until a new Test ID is selected.
Selecting Access Test IDs from the Expanded Menu System allows the
user to select a previously entered Test ID, create a new Test ID or edit
an old one, send the Test IDs to a computer, or erase all existing Test
IDs.
3.17.1 Select Test IDs
Select a Test ID as follows:
1. From the Expanded Main Screen, select Menu > Access Test IDs >
Select Test IDs.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e l e c t T e s t I D s
E d i t T e s t I D s
S e n d T e s t I D s
E r a s e T e s t I D s
[ 0 0 0 4 ] ( A )S - -
2. Use the keys to scroll through any previously stored Test IDs until
the desired ID is displayed. In the example below, Test ID 0005 has been
chosen.
Tip: To quickly scroll through the Test IDs, press and hold down
the or key.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
T e s t L o c a t i o n 1
O p e r a t o r J o h n D o e
[ 0 0 0 5 ] ( A )S - -
Instruction 0055-9017
27
Operation Hi Flow Sampler
3. Select the displayed Test ID by pressing the I/Okey, and then
return to the Main Screen by pressing ESC twice. Observe that [0005]
is now displayed at the bottom of the screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
M e n u
[ 0 0 0 5 ] ( A )S - -
3.17.2 Edit Test IDs
Edit a previously entered Test ID or to create a new one as follows:
1. First select the Test ID to edit per Section 3.17.1 Select Test IDs. To
create a new Test ID, select the last Test ID number that can be
displayed – the last Test ID number will always be blank.
2. From the Expanded Main Screen, select Menu > Access Test IDs >
Edit Test IDs.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e l e c t T e s t I D s
E d i t T e s t I D s
S e n d T e s t I D s
E r a s e T e s t I D s
[ 0 0 0 5 ] ( A )S - -
3. Use the following keys to select and then enter the desired character.
– displays desired alphanumerical character
I/O – selects displayed character and moves cursor one space
to the right
ESC – moves cursor one space to the left
Tip: To quickly scroll through the characters, press and hold down
the or key.
Tip: The “space” character is between the number “9” and the capital
“A”.
In the example below, the Test Location has been changed from 1 to
2 with the cursor positioned to the immediate right of the number 2.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
T e s t L o c a t i o n 2 |
O p e r a t o r J o h n D o e
[ 0 0 0 5 ] ( A )S - -
4. After all characters have been selected, press and hold down the
I/Okey to move the cursor to the end of the screen; after which, the
screen shown in Step 2 should reappear.
28
Instruction 0055-9017
Hi Flow Sampler Operation
5.Press the ESC key twice to return to the Main Screen.
3.17.3 Send Test IDs
The following describes how to download the current list of Test IDs to a
computer that is running Windows
communications program. If a different operating system and/or
communications program is being used, then please consult the
appropriate instruction manuals for those products.
1. Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9
male to DB9 female) between the computer’s COM port and the Hi
Flow’s COMPUTER connector. (For computers having a 25-pin or
USB COM port, the user must supply the appropriate adapter, or use
a serial cable with appropriate connectors on each end.)
2. On the computer, start HyperTerminal by doing one of the following:
- If HyperTerminal has already been set up to communicate with the
Hi Flow Sampler, then select Start > Programs > Accessories >
Communications > HyperTerminal, and double-click the filename
or icon that has been associated with the instrument.
- If HyperTerminal has never been run in association with the Hi Flow
Sampl-er, then a new HyperTerminal connection needs to be made as
follows:
a. Select Start > Run. In the Run box type “hypertrm.exe” and
click OK. The New Connection dialog box should appear.
b. Type in a New Connection Name such as “HiFlow”. Click OK to
select the default icon (if desired, select a different icon before
clicking OK). The Connect To dialog box should appear.
c. Click the Connect using drop-down menu and choose the COM
port to which the Hi Flow is connected. Then click OK to display
the COMProperties dialog box for that COM port.
d. Set Port Settings to:
- Bits per second: 115200
- Data bits: 8
- Parity: None
- Stop bits: 1
- Flow control: None
Then click OK to display HyperTerminal’s main window.
e. Click File > Save. This creates a file with an .ht extension using
the name that was entered in Step b.
3. After HyperTerminal is running: click Transfer > Capture Text >
Start to accept the default drive\directory\filename to which all
received data will be stored as an ASCII text file.
Note: The captured text file will be stored on the hard drive in the folder
shown in the File dialog box. If desired, click Browse to select a different
drive, directory and filename. The next time data is captured, it will be
stored under the same location and filename.
®
HyperTerminal as the
Instruction 0055-9017
29
Operation Hi Flow Sampler
4. From the High Flow Sampler’s Expanded Main Screen, start the
download process by selecting Menu > Access Test IDs > Send Test IDs.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e l e c t T e s t I D s
E d i t T e s t I D s
S e n d T e s t I D s
E r a s e T e s t I D s
Observe that the Hi Flow’s LCD displays the message “Sending record
x of y @ 115.2kbps,” where “x” is the current record being sent and
“y” is the number of records in memory. Also observe that
HyperTerminal displays the data as its being received.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e n d i n g r e c o r d
x o f y
@ 1 1 5 . 2 k b p s
5. After all data has been transmitted, stop HyperTerminal from
capturing data by clicking Transfer > Capture Text > Stop.
6. The saved data can now be viewed on the computer using any
program capable of displaying “txt” files.
3.17.4 Erase Test IDs
Erase all Test ID information and reset the Test ID index back to 0001 as
follows:
1. From the Expanded Main Screen, select Menu > Access Test IDs >
Erase Test IDs.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
S e l e c t T e s t I D s
E d i t T e s t I D s
S e n d T e s t I D s
E r a s e T e s t I D s
[ 0 0 0 4 ] ( A )S - -
[ 0 0 0 4 ] ( A )
S - -
[ 0 0 0 4 ] ( A )S - -
30
Instruction 0055-9017
Hi Flow Sampler Operation
2. Select Yes to erase all Test ID information, or select No to return to
the previous screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
E r a s e A l l T e s t I D s ?
Y e s
N o
[ 0 0 0 4 ] ( A )S - -
3.Press ESC twice to return to the Main Screen.
Instruction 0055-9017
31
Operation Hi Flow Sampler
3.18 Making a Leak Rate Measurement
Before making a leak rate measurement, the user must decide whether
to use . . .
the Basic or Expanded Menu Mode (Section 3.8)
lpm or cfm for the flow rate (Section 3.12)
the Automatic 2-Stage, Manual 2-Stage, or Manual 1-Stage
method of making a measurement when operating in the
Expanded Menu Mode (Section 3.14)
a Test ID when operating in the Expanded Menu Mode (Section
3.17)
3.18.1 Leak Rate Measurement – Basic Menu Mode
While operating in the Basic Menu Mode, the user starts a measurement
at a high flow rate and then waits for a stable leak rate reading to occur.
During this time the flow rate can be changed using the Speed function
on the Main Screen. This mode is used when more control of the leak
rate measurement process is required, and when saving data is not
desired.
1. If not already done, place the instrument into its Basic Menu Mode
per
Section 3.8 Selecting the Basic or Expanded Menu Mode.
2. Choose an attachment that will ensure the complete capture of the
gas leak. Connect this attachment to the end of the Hi Flow
Sampler’s main sampling hose and position the attachment’s inlet
over the leak source.
3. Position the inlet of the instrument’s background gas sampling hose
opposite the leak source, such that the leak source will not
contribute to the background measurement.
4. Select Start from the Main Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S t a r t
[ 0 0 0 1 ] ( M )S - -
5. Begin measuring until a stable leak measurement is achieved—this
may take several minutes.
If desired, the flow rate can be manually controlled by using the
Speed function. Each press of the I/O key will lower the flow rate as
indicated by the speed bar moving to the left.
32
Instruction 0055-9017
Hi Flow Sampler Operation
1 1 / 2 2 / 0 4 0 9 : 4 7 : 3 0
F l o w ( l p m ) 2 1 6 . 5
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
[ 0 0 0 1 ] ( M )- 1 -
S a m p l e T i m e - > 0 2 : 0 0
6. In the example below, the current flow rate is 216.5 lpm, the
background gas concentration is 0.8%, the leak concentration is 15%
gas, and the calculated leak rate is 42.5 lpm.
1 1 / 2 2 / 0 4 0 9 : 4 7 : 3 0
F l o w ( l p m ) 2 1 6 . 5
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
[ 0 0 0 1 ] ( M )- 1 -
S a m p l e T i m e - > 0 2 : 0 0
7. The measurement process continues until Stop is selected.
1 1 / 2 2 / 0 4 0 9 : 5 0 : 0 0
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
Note: The Automatic 2-Stage Mode can only be selected if operating in the
Expanded Menu Mode. Refer to Section 3.8.
1. If not already done, place the instrument into its Automatic 2-Stage
measurement mode (denoted by the letter (A) appearing at the bottom
of the screen). Refer to Section 3.14.1 Automatic 2-Stage Measurement Mode.
2. Choose an attachment that will ensure the complete capture of the
gas leak. Connect this attachment to the end of the Hi Flow
Sampler’s main sampling hose and position the attachment’s inlet
over the leak source.
3. Position the inlet of the instrument’s background gas sampling hose
opposite the leak source, such that the leak source will not
contribute to the background measurement.
4. Begin the measurement process by selecting Start from the Main
Screen.
Instruction 0055-9017
33
Operation Hi Flow Sampler
ar
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
[ 0 0 0 1 ] ( A )S - -
5. The user is now asked to choose a new or different Test ID.
Selecting Yes causes the Access Records menu to appear (refer to
Section 3.15). After selecting or creating a new Test ID, start the
measurement by pressing ESC until the Main Screen appears.
Selecting No returns the display to the Main Screen and starts the
measurement using the current Test ID.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
A s s i g n n e w t e s t I D ?
Y e s
N o
[ 0 0 0 1 ] ( A )S - -
6. (Optional) The instrument normally controls the flow rate for
measurements #1 and #2; however, the user can manually set the
flow rate using the Speed function. Each press of the I/O key will
lower the flow rate as indicated by the speed bar moving to the left.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
F l o w ( l p m ) 0 . 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 0 . 0 0
L e a k ( l p m ) 0 . 0
S p e e d ( L O < > H I )
[ 0 0 0 1 ] ( A )- 1 -
Speed B
34
Instruction 0055-9017
Hi Flow Sampler Operation
7. After both measurements are complete, the values in the Main Screen
are put on hold (as indicated by “H--” appearing at the bottom of the
screen), allowing the user to make note of these values before turning off
the instrument or proceeding to the next measurement.
In the examples below, the measured leak concentration was 15% gas
during the first measurement and 18% gas in the second, while the
calculated leak rates were 42.5 and 40.8 lpm, respectively. A difference
of 4%.
Important: For a test to be considered valid, the percent difference
between measurements #1 and #2 must be less than 10%. Scroll up to
the #1–#2 function on the Main Screen to see the percent difference
between the last two measurements.
1 1 / 2 2 / 0 4 1 0 : 4 6 : 3 0
F l o w ( l p m ) 2 1 6 . 5
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
[ 0 0 0 5 ] ( A )- 1 -
S a m p l e T i m e - > 0 1 : 0 0
0 2 / 2 1 / 0 3 1 0 : 4 7 : 3 0
F l o w ( l p m ) 1 8 0 . 5
B a c k ( % ) 0 . 0 7
L e a k ( % ) 1 8 . 0 0
L e a k ( l p m ) 4 0 . 8
S p e e d ( L O < > H I )
[ 0 0 0 5 ] ( A )- - 2
S a m p l e T i m e - > 0 2 : 0 0
8. When in the automatic mode, measurements are automatically
saved. These saved measurements along with their associated Test
IDs can be downloaded to a personal computer at a later time for
analysis. Refer to Section 3.16 Importing Saved Data into a Spreadsheet.
Note: The Manual 2-Stage Mode can only be selected if operating in the
Expanded Menu Mode. Refer to Section 3.8.
1. If not already done, place the instrument into its Manual 2-Stage
mode. Refer to Section 3.14.2.1 Manual 2-Stage Mode.
2. Choose an attachment that will ensure the complete capture of the
gas leak. Connect this attachment to the end of the Hi Flow
Sampler’s main sampling hose and position the attachment’s inlet
over the leak source.
3. Position the inlet of the instrument’s background gas sampling hose
opposite the leak source, such that the leak source will not
contribute to the background measurement.
4. Begin the measurement process by selecting Start from the Main
Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
[ 0 0 0 1 ] ( M )S - -
5. The user is now asked to choose a new or different Test ID.
Selecting Yes causes the Access Records menu to appear (refer to
Section 3.15). After selecting or creating a new Test ID, start the
measurement by pressing ESC until the Main Screen appears.
Selecting No returns the display to the Main Screen and starts the
measurement using the current Test ID.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
A s s i g n n e w t e s t I D ?
Y e s
N o
[ 0 0 0 1 ] ( M )S - -
6. Begin measuring at the first flow rate until a stable leak
measurement is achieved—this may take several minutes.
If desired, the flow rate can be manually controlled by using the
Speed function. Each press of the I/O key will lower the flow rate as
indicated by the speed bar moving to the left.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
F l o w ( l p m ) 0 . 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 0 . 0 0
L e a k ( l p m ) 0 . 0
S p e e d ( L O < > H I )
[ 0 0 0 1 ] ( M )- 1 -
36
Speed B
Instruction 0055-9017
Hi Flow Sampler Operation
7. Once a stable leak measurement is obtained, highlight Save and
press the I/Okey. Observe that the measurement is saved in
memory, and that the flow rate is automatically lowered.
In the example below, the measured leak concentration for
measurement #1 was 15% gas, and the calculated leak rate was 42.5
lpm.
1 1 / 2 2 / 0 4 0 9 : 4 7 : 1 5
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
S a v e
[ 0 0 0 1 ] ( M )- 1 -
S a m p l e T i m e - > 0 1 : 4 5
8. Continue sampling at the reduced flow rate until once again a stable
leak measurement is achieved. If desired, the flow rate can again be
manually controlled by using the Speed function.
With Save highlighted, press the I/Okey to save measurement #2 in
memory. The blower then stops and the values in the Main Screen are
put on hold (as indicated by “H--” appearing at the bottom of the screen),
allowing the user to make note of these values before turning off the
instrument or proceeding to the next measurement.
In the example below, the leak concentration was 18% gas, the leak
rate was 40.8 lpm, and the percent difference between measurements
#1 and #2 was 4%.
Important: For a test to be considered valid, the percent difference
between measurements #1 and #2 must be less than 10%. Scroll up to
the #1–#2 function on the Main Screen to see the percent difference
between the last two measurements.
1 1 / 2 2 / 0 4 0 9 : 4 9 : 0 0
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 8 . 0 0
L e a k ( l p m ) 4 0 . 8
S p e e d ( L O < > H I )
S a v e
[ 0 0 0 1 ] ( M )- - 2
S a v e
S a m p l e T i m e - > 0 3 : 3 0
9. Saved measurements can downloaded to a personal computer at a
later time for analysis. Refer to Section 3.16 Importing Saved Data into a Spreadsheet.
Note: The Manual 1-Stage Mode can only be selected if operating in the
Expanded Menu Mode. Refer to Section 3.8.
1. If not already done, place the instrument into its Manual 1-Stage
mode. Refer to Section 3.14.2.2 Manual 1-Stage Mode.
2. Choose an attachment that will ensure the complete capture of the
gas leak. Connect this attachment to the end of the Hi Flow
Sampler’s main sampling hose and position the attachment’s inlet
over the leak source.
3. Position the inlet of the instrument’s background gas sampling hose
opposite the leak source, such that the leak source will not
contribute to the background measurement.
4. Begin the measurement process by selecting Start from the Main
Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
[ 0 0 0 1 ] ( M )S - -
5. The user is now asked to choose a new or different Test ID.
Selecting Yes causes the Access Records menu to appear (refer to
Section 3.15). After selecting or creating a new Test ID, start the
measurement by pressing ESC until the Main Screen appears.
Selecting No returns the display to the Main Screen and starts the
measurement using the current Test ID.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
A s s i g n n e w t e s t I D ?
Y e s
N o
[ 0 0 0 1 ] ( M )S - -
6. Begin measuring until a stable leak measurement is achieved—this
may take several minutes.
If desired, the flow rate can be manually controlled by using the
Speed function. Each press of the I/O key will lower the flow rate as
indicated by the speed bar moving to the left.
38
Instruction 0055-9017
Hi Flow Sampler Operation
ar
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
F l o w ( l p m ) 0 . 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 0 . 0 0
L e a k ( l p m ) 0 . 0
S p e e d ( L O < > H I )
[ 0 0 0 1 ] ( M )- 1 -
Speed B
7. Once a stable leak measurement is obtained, all current readings can
be saved in memory by highlighting Save and pressing the I/Okey.
In the example below, the current measured leak concentration is
15% gas, and the calculated leak rate is 42.5 lpm.
1 1 / 2 2 / 0 4 0 9 : 4 7 : 1 5
B a c k ( % ) 0 . 8 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
S a v e
[ 0 0 0 1 ] ( M )- 1 -
S a m p l e T i m e - > 0 1 : 4 5
8. The measurement process continues until Stop is selected.
1 1 / 2 2 / 0 4 0 9 : 4 9 : 0 0
L e a k ( % ) 1 5 . 0 0
L e a k ( l p m ) 4 2 . 5
S p e e d ( L O < > H I )
S a v e
S t o p
S a v e
[ 0 0 0 1 ] ( M )- 1 -
S a m p l e T i m e - > 0 3 : 3 0
9. Measurements that were saved can be downloaded to a personal
computer at a later time for analysis. Refer to Section 3.16 Importing Saved Data into a Spreadsheet.
Instruction 0055-9017
39
Operation Hi Flow Sampler
3.19 Message Line
Various messages will appear at the bottom of the LCD that describe
what functions are currently being performed or errors that have been
detected. The following list describes the meaning of these messages.
Note that detailed explanations of the error messages can be found in
Section 6.2 Troubleshooting Guide.
Calibration Passed
Calibration procedure was successful.
Calibration Failed
Error – The calibration procedure was not successful. Refer to
Section 6.2.
Saving Record
A test record is currently being saved to memory.
Memory Full
Error – The instrument tried to save a test record with 1000 records
already stored in memory. Refer to Section 6.2.
No Records Saved
An attempt was made to send records to a computer with no test
records stored in memory.
Flow Rate Lowered
The flow rate has just been lowered signaling the start of
measurement #2 during an automatic or manual leak test. Refer to
Section 3.14.
Erasing All Records
All test records are being erased after selecting “Erase records” per
Section 3.15.3.
Check Background Pump
Error – The background-gas pump’s flow rate has been severely
reduced. Refer to Section 6.2.
Check Leak Pump
Error–The leak-gas pump’s flow rate has been severely reduced. Refer
to Section 6.2.
Check Background Calibration
Error – The background-gas sensor needs calibrated. Refer to
Section 6.2.
Check Leak Calibration
Error – The leak-gas sensor needs calibrated. Refer to Section 6.2.
Check Background Offset
Error – The background-gas sensor did not properly zero during
startup. Refer to Section 6.2
Check Leak Offset
Error – The leak-gas sensor did not properly zero during startup. Refer
to Section 6.2.
40
Instruction 0055-9017
Hi Flow Sampler Operation
Check Zeroing Bypassed
Error – The ESC key was pressed during start up, thus bypassing the
normal sensor-zeroing process. Refer to Section 6.2.
Check Leak Rate Difference
Error–The calculated leak rate difference between measurements #1
and #2 during an automatic or manual leak test was greater than
100%. Refer to Section 6.2.
Check Background Greater Than Leak
Error–The measured background gas level was greater than the
measured leak gas concentration. Refer to Section 6.2
Sample Time
The amount of time that a leak rate measurement has been in
progress.
Barcode Scanned
A barcode has been successfully scanned and its data stored in
memory.
Purging Sensors
The instrument automatically turns on both sensor pumps to purge
the background-gas and leak-gas sensor chambers. This action
occurred because during an idle period (blower not running) or during
shutdown the instrument has detected that more than 50 ppm of
natural gas is present in the sensor chamber(s).
Low Battery
Error–Battery voltage has fallen below 4.0 volts. Refer to Section 6.2.
3.20 GPS Data
Note: GPS data is only displayed ifoperating in the Expanded Menu Mode.
Refer to Section 3.8.
When a GPS receiver is connected to the instrument’s GPS port, the
instrument acknowledges this connection by displaying
left corner of the LCD.
At this time the instrument’s Expanded Main Screen displays the
receiver’s Lat (latitude) and Lon (longitude) in degrees; it’s Alt (altitude) in
feet; and the Fix (Time of Fix) in seconds.
If the Fix time is greater than –10 seconds, the Lat, Lon, and Alt values
are considered invalid, causing their values to be displayed as "Not
Avail."
at the bottom
Instruction 0055-9017
41
Operation Hi Flow Sampler
The GPS data is saved along with all other measured and calculated
values at the conclusion of an automatic test, or when Save is selected
during a manual test.
3.21 Barcode Scanner Data
Note: Barcode scanning can only be performed if operating in the
Expanded Menu Mode. Refer to Section 3.8.
When a barcode scanner is connected to the instrument’s SCANNER
port, the instrument acknowledges this connection by displaying
bottom left corner of the LCD.
When the barcode scanner is swiped across a barcode, the message
“Barcode Scanned” briefly appears at the bottom of the LCD. The barcode
data is stored and will be associated with all following tests until another
barcode is scanned, or until the instrument is turned OFF. The barcode
data is also saved along with all other measured and calculated values at
the conclusion of an automatic test, or when Save is selected during a
manual test.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
L e a k ( % ) 0 . 0 0
L e a k ( N o t A v a i l . )
S p e e d ( L O < > H I )
S a v e
S t a r t
[ 0 0 0 1 ] ( A )
B a r c o d e S c a n n e d
S - -
at the
42
Instruction 0055-9017
Hi Flow Sampler Calibration
4 Calibration
It is recommended that the Hi Flow Sampler be calibrated by your
nearest Bacharach Service Center every 30 days. Calibration, however,
can be performed in the field if your facility has the necessary equipment
and qualified personnel to perform the procedures described in the
following sections.
The following procedures describe how to:
View the dates on which calibrations were performed
Verify calibration
Calibrate the background and leak gas sensors
Display the gain values of the calibration amplifier circuits
Erase calibration
4.1 View Calibration Dates
Do the following to view the dates on which calibrations were performed
on the background-gas and leak-gas sensors at both low (2.5% CH
high (100% CH
) calibration gas levels:
4
1. From the Main Screen select Menu > Calibration > Calibration
Dates.
2. The following Calibration Dates Screen is displayed where mm/dd/yy
is the month, day, and year calibration was performed:
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
B a c k L o w m m / d d / y y
B a c k H i g h m m / d d / y y
L e a k L o w m m / d d / y y
L e a k H i g h m m / d d / y y
) and
4
[ 0 0 0 5 ] ( A )S - -
3.Press the ESC key twice to return to the Main Screen.
4.2 Gas Calibration Equipment Setup
The following equipment is recommended to both check the calibration
and perform a calibration procedure on the Hi Flow Sampler:
Connect the components of the calibration kit, calibration gas cylinder,
and Hi Flow Sampler together as shown in Figure 4-1. DO NOT, however,
attach the calibration equipment to the instrument until instructed to do
so in the calibration procedure.
Instruction 0055-9017
in air, P/N 0051-1121
4
, P/N00 55-0060
4
43
Calibration Hi Flow Sampler
Figure 4-1. Calibration Equipment Setup
4.3 Check Calibration Procedure
This procedure checks the calibration of both the leak-gas sensor and
the background-gas sensor using 2.5% CH
calibration equipment as described in Section 4.2, proceed to check the
instrument’s sensor calibration as follows:
1. Before turning ON the instrument, be sure that the calibration
equipment is not connected to the instrument’s GAS or
BACKGROUND inlet port, and that the instrument is in an area of
clean air.
2. Turn ON the instrument and wait for its warm-up period to complete
before proceeding with Step 3.
3. Apply 2.5% CH
BACKGROUND inlet port.
4. From the Main Screen select Menu > Calibration > Verify
Calibration to begin the calibration verification process. Observe that
the gas-sampling pump motors should start.
Wait several minutes for the gas reading to stabilize. At this time the
gas concentration shown on the display should match the
concentration stamped on the calibration gas cylinder.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
from the calibration equipment to the instrument’s
4
. After setting up the
4
B a c k ( % ) 2 . 5 0
L e a k ( % ) 0 . 0 0
[ 0 0 0 5 ] ( M )S - -
44
Instruction 0055-9017
Hi Flow Sampler Calibration
5. Disconnect the hose from the BACKGROUND inlet port and connect
it to the GAS inlet port.
6. Again wait several minutes for the gas reading to stabilize. At this
time the gas concentration shown on the display should match the
concentration stamped on the calibration gas cylinder.
1 1 / 2 2 / 0 4 1 0 : 4 5 : 3 0
B a c k ( % ) 0 . 0 0
L e a k ( % ) 2 . 5 0
[ 0 0 0 5 ] ( M )S - -
7. Disconnect the hose from the GAS port, and then allow the pumps to
run until both the Back and Leak readings fall to zero percent.
8. Press the ESC key three times to return to the Main Screen.
Note: If gas is still present in either sensor chamber when ESC is pressed,
the instrument will automatically begin to purge the sensors until all gas is
removed from the instrument.
4.4 Gas Calibration Procedure
This procedure calibrates both the leak-gas sensor and the backgroundgas sensor at 2.5% and 100% CH
equipment as described in Section 4.2, proceed to calibrate the sensors
as follows:
1. Before turning ON the instrument, be sure that the calibration
equipment is not connected to the instrument’s GAS or
BACKGROUND inlet port, and that the instrument is in an area of
clean air.
2. Turn ON the instrument and wait its warm-up period to complete
before proceeding to Step 3.
3. From the Main Screen select Menu > Calibration > Calibrate
Sensors to display the Sensor Calibration Screen.
0 2 / 2 1 / 0 3 0 9 : 4 5 : 3 0
B a c k g n d @ 2 . 5 % C H 4
B a c k g n d @ 1 0 0 % C H 4
L e a k @ 2 . 5 % C H 4
L e a k @ 1 0 0 % C H 4
. After setting up the calibration
4
4. Using the keys, highlight the sensor to be calibrated along with its
Instruction 0055-9017
[ 0 0 0 5 ] ( A )S - -
gas level. (In the example above, the background sensor will be
calibrated using 2.5% CH
.)
4
45
Calibration Hi Flow Sampler
5. Apply the appropriate level of calibration gas from the calibration
equipment to the instrument’s GAS or BACKGROUND inlet port. (In
the example above, apply 2.5% CH
to the BACKGROUND port.) Then
4
press the I/Okey to start the calibration process. Observe that the
gas-sampling-pump motors start and the calibration screen appears.
1 1 / 2 2 / 0 4 1 0 : 4 5 : 3 0
B a c k ( % ) 0 . 0 0
A p p l ( % ) 2 . 5 0
, = A d j . A p p l i e d
E N T E R = C a l i b r a t e
E S C = E x i t
[ 0 0 0 5 ] ( M )S - -
6. If necessary, use the keys to adjust the Appl(%) reading to match
the gas concentration stamped on the calibration gas cylinder.
7. Wait until the measured gas reading stabilizes, and then press the
I/Okey to calibrate the actual gas reading to that of the applied
reading. The message “Calibration Passed!” will appear at the bottom
of the screen if the calibration was successful. If, however, the
calibration was not successful, the message “Calibration Failed!” will
appear. Refer to Section 6 Troubleshooting for possible causes and
remedies for this failure.
1 1 / 2 2 / 0 4 1 1 : 4 5 : 3 0
B a c k ( % ) 2 . 5 0
A p p l ( % ) 2 . 5 0
, = A d j . A p p l i e d
E N T E R = C a l i b r a t e
E S C = E x i t
[ 0 0 0 5 ] ( M )S - -
C a l i b r a t i o n P a s s e d !
8. Disconnect the gas hose, and then wait until the measured gas
reading falls to zero percent.
9. Press the ESC key to return to the Calibration Menu Screen.
10. Repeat this procedure as necessary to calibrate both sensors at 2.5%
and 100% CH
.
4
46
Instruction 0055-9017
Hi Flow Sampler Calibration
4.5 Calibration Amplifier Gain Check
Use this procedure to display the gain values of the calibration amplifier
circuits in hexadecimal. This procedure is useful to service technicians
for troubleshooting purposes.
1. From the Main Screen select Menu > Calibration > Calibration
Gains to display the Calibration Gains Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
B a c k L o w # # # #
B a c k H i g h # # # #
L e a k L o w # # # #
L e a k H i g h # # # #
[ 0 0 0 1 ] ( M )
S - -
Where: #### is the gain value in hexadecimal.
2. Press ESCthree times to return to the Main Screen.
4.6 Erase Calibration Procedure
The following procedure resets the gain of each calibration amplifier
circuit to the factory default hexadecimal value of “FFFF”.
1. From the Main Screen select Menu > Calibration > Erase
Calibration to display the Erase Calibration Screen.
1 1 / 2 2 / 0 4 0 9 : 4 5 : 3 0
E r a s e c a l i b r a t i o n ?
Y e s
N o
[ 0 0 0 1 ] ( M )
2. Select Yes to erase the calibration values. Observe that the message
“Erasing Calibration” appears at the bottom of the screen. Select No
to abort this procedure and return to the Calibration Menu Screen.
3. Press ESCtwice to return to the Main Screen.
S - -
Instruction 0055-9017
47
Calibration Hi Flow Sampler
Notes:
48
Instruction 0055-9017
Hi Flow Sampler Maintenance
5 Maintenance
Routine maintenance of the Hi Flow Sampler consists of the following:
Charging the battery pack
Replacing the internal filters
Replacing the gas sensors
Verifying flow rate
Setting the time and date
Resetting the microprocessor
5.1 Electrostatic Discharge Precautions
CAUTION: Electrostatic discharge (ESD) can destroy sensitive electronic
components inside the Hi Flow Sampler when its cover is removed.
When performing any maintenance procedure that requires the Hi Flow
Sampler’s cover to be removed, take the following precautions to prevent
ESD from possibly destroying sensitive electronic components on the
main circuit board:
Make sure you have a reliable ground point available near the work
site
Connect your body to the ground point using a wrist strap
Always stand on an approved conductive floor mat when touching or
handling equipment
Ground all equipment you are working on with ground straps
Handle components only on a grounded anti-static work surface
Do not wear clothing that generates static electric charges every
time you move
Do not handle static generating objects while working
Maintain relative humidity in the work area between 40 and 50%
Instruction 0055-9017
49
Maintenance Hi Flow Sampler
5.2 Charging the Battery Pack
WARNING: Explosion hazard!DO NOT charge the battery pack in an
unsafe atmosphere.
1. Unplug battery pack from the instrument’s power cord; then remove
battery pack from backpack.
2. Interconnect the supplied power supply, power cord, and charger as
shown in Figure 5-1. Then plug the output of the charger into the
battery pack’s charger jack.
Note: Connecting the charger to the Battery Pack disables the battery’s
output, making it impossible to both charge the battery and run the
instrument.
3. Allow a completely discharged battery pack to charge from 8 to 10
hours.
Figure 5-1. Charging the Battery Pack
50
Instruction 0055-9017
Hi Flow Sampler Maintenance
5.3 Removing the Cover
When instructed to do maintenance inside the instrument, remove its
protective cover as follows:
1. Do the following before removing the instrument from the backpack:
- Disconnect battery pack
- Unscrew metal-cage diffuser from the instrument’s gas exhaust
port
- Disconnect ground clamp wire
2. Remove instrument from backpack.
3. Remove screws from cover as shown in Figure 5-2; then pull cover
away from unit.
Figure 5-2. Removing the Cover
Instruction 0055-9017
51
Maintenance Hi Flow Sampler
5.4 Replacing the Internal Filters
Three internal filters protect the instrument from being contaminated by
dust and dirt. Replace these filters when they look contaminated. A good
indication that a filter needs replaced is when the instrument’s gasresponse time increases.
5.4.1 Leak-Gas Filter
The leak-gas filter is located between the sampling port on the flowmeasurement tube and its associated solenoid valve on the printed
circuit board. See Figure 5-3.
Equipment Required:
Medium Phillips Head Screwdriver
Replacement Filter (P/N 07-1563)
Procedure:
1. Remove cover per Section 5.3.
2. Remove tubing from old filter’s inlet and outlet.
3. Install tubing on new filter, making sure that the filter’s flow arrow
points away from the flow-measurement tube.
4. Replace cover that was removed in Step 1.
Figure 5-3. Leak-Gas Filter
52
Instruction 0055-9017
Hi Flow Sampler Maintenance
5.4.2 Leak-Gas Sensor Filter
The leak-gas sensor filter is located inside the flow block of the leak-gas
sensor assembly. See Figure 5-4.
Equipment Required:
Medium Phillips Head Screwdriver
Replacement Filter (P/N 0055-0045)
Procedure:
1. Remove cover per Section 5.3.
2. Remove tubing from inlet of sensor assembly.
3. Unscrew the filter chamber’s gas inlet fitting; then pull out old filter
and discard.
4. Insert new filter; reinstall the inlet fitting; then reinstall tubing.
5. Replace cover that was removed in Step 1.
5.4.3 Background-Gas Sensor Filter
The background-gas sensor filter is located inside the flow block of the
background-gas sensor assembly. See Figure 5-4.
Replace filter as previously described for the leak-gas sensor filter.
Figure 5-4. Leak Gas & Background Gas Sensor Filter Replacement
Instruction 0055-9017
53
Maintenance Hi Flow Sampler
5.5 Gas Sensor Replacement
Replace a gas sensor when it fails to calibrate per Section 4 Calibration.
The gas sensors are located inside their associated flow blocks of the
leak-gas and background-gas sensor assemblies. See Figure 5-5.
Figure 5-5. Leak-Gas & Background Gas Sensor Assemblies
Equipment Required:
Medium Phillips Head Screwdriver
Replacement Sensor (P/N 0055-0040)
Procedure:
1. Remove cover per Section 5.3.
2. Remove the pump connector and multi-pin connector from printed
circuit board, noting orientation of connectors. See Figure 5-6.
3. Remove three screws from printed circuit board; then carefully pry
the board straight up from the flow block. The gas sensor is the silver
cylindrical piece in the center of the circuit board.
4. Pull sensor from circuit board and replace with a new sensor, making
sure that the new sensor is firmly seated into its receptacles and
bottomed on circuit board.
5. Reinstall printed circuit board into flow block; taking care to properly
position the board as shown in Figure 5-6.
6. Replace three screws and tighten to secure board. Reattach the pump
and multi-pin connectors to the circuit board.
7. Replace cover that was removed in Step 1.
8. Calibrate the instrument per Section 4 Calibration. Note that if a new
sensor is installed without performing a calibration, either the
message “Check Background Calibration” or “Check Leak
Calibration” will appear at the bottom of the LCD when the
instrument is turned ON.
54
Instruction 0055-9017
Hi Flow Sampler Maintenance
5.6 Pump Replacement
The pump located in either the leak-gas or background-gas sensor
assembly (see Figure 5-5) may become inoperative due to water or solid
particles getting into the valves. This will require replacement of the
pump to correct.
Equipment Required:
5
/64" Allen Wrench
Replacement Pump Assembly (P/N 0055-0061)
Procedure:
1. Remove cover per Section 5.3.
2. Back off the pump-bracket screw using a
rotate bracket arm to free pump. See Figure 5-6.
3. Unplug pump connector from printed circuit board and lift pump
straight up.
4. Install the new pump by lining up the pump’s inlet and outlet tubes
with the two flow block holes and pushing into place. Swing bracket
arm over pump and tighten—realigning pump as necessary to make a
proper fit.
5. Replace cover that was removed in Step 1.
5
/64" Allen wrench; then
Figure 5-6. Sensor or Pump Replacement
Instruction 0055-9017
55
Maintenance Hi Flow Sampler
5.7 Flow Rate Verification
This procedure verifies the instrument’s ability to provide sufficient
airflow to capture the entire gas leak, and to verify the accuracy of the
instrument’s calculated flow rate by using the optional external flow
meter (P/N 0055-0270).
Important: To assure accurate flow readings, conduct the following
procedure in clean air, free of any combustible gases.
1. Install the flow meter onto the Hi Flow Sampler’s gas intake as shown
in Figure 5-7.
2. Place the Hi Flow Sampler into its Manual 1-Stage Mode per
Section 3.14.2.2. Then start the blower by selecting Start from the
Main Screen.
3. Make note of the Flow(cfm) value as displayed on the Hi Flow
Sampler’s LCD, and the fpm flow rate as displayed on the external
flow meter.
4. Both Hi Flow Sampler's flow rate and the externally measured flow
rate should intersect within the two lines of the graph shown in
Figure 5-8. If the intersect point falls outside the two graph lines,
then return the unit to your nearest Bacharach Service Center for
evaluation.
Figure 5-7. Flow Meter Installation
56
Instruction 0055-9017
Hi Flow Sampler Maintenance
Figure 5-8. Flow Rate Validation Graph
5. With a fully charged battery pack, the measured flow rate should be
greater than 9 cfm. If not, perform the following (see Figures 5-9
and 5-10):
Clean the gas sample inlet screen.
Ensure that the exhaust port is not obstructed.
Remove the unit’s cover and ensure that the high- and low-
sampling point hoses are properly connected between the flowmeasurement tube and their corresponding solenoid valves, and
that these hoses are not pinched or kinked.
Ensure that no obstruction is present inside the flow-
measurement tube.
If none of the above solves the problem, return the unit to your
nearest Bacharach Service Center for evaluation.
Instruction 0055-9017
57
Maintenance Hi Flow Sampler
Figure 5-9. Gas Flow System
58
Instruction 0055-9017
Hi Flow Sampler Maintenance
Figure 5-10. Plumbing Diagram
5.8 Setting Date and Time
The following describes how to set the date and time as displayed on the
Hi Flow Sampler’s LCD using a personal computer that is running
Windows
operating system and/or communications program is being used, then
please consult the appropriate instruction manuals for those products.
1. Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9
2. On the computer, start HyperTerminal by doing one of the following:
Instruction 0055-9017
®
HyperTerminal as the communications program. If a different
male to DB9 female) between the computer’s COM port and the Hi
Flow Sampler’s COMPUTER connector. (For computers having a 25pin or USB COM port, the user must supply the appropriate adapter,
or use a serial cable with appropriate connectors on each end.)
- If HyperTerminal has already been set up to communicate with
the Hi Flow Sampler, then select Start > Programs > Accessories
> Communications > HyperTerminal, and double-click the file
name or icon that has been associated with the instrument.
59
Maintenance Hi Flow Sampler
- If HyperTerminal has never been run in association with the Hi
Flow Sampler, then a new HyperTerminal connection needs to be
made as follows:
a. Select Start > Run. In the Run box type “hypertrm.exe” and
click OK. The New Connection dialog box should appear.
b. Type in a New Connection Name such as “HiFlow”. Click OK
to select the default icon (if desired, select a different icon
before clicking OK). The Connect To dialog box should
appear.
c. Click the Connect using drop-down menu and choose the
COM port to which the Hi Flow Sampler is connected. Then
click OK to display the COMProperties dialog box for that
COM port.
d. Set Port Settings to:
- Bits per second: 115200
- Data bits: 8
- Parity: None
- Stop bits: 1
- Flow control: None
Then click OK to display HyperTerminal’s main window.
e. Set up the terminal program to ‘echo’ typed characters by
selecting File > Properties > Settings > ASCII Setup and
placing a checkmarks in the “Send line ends with line feeds”
box and the “Echo typed characters locally” box. Click OK.
Tip: If the typed characters do not echo on the screen, then your
version of HyperTerminal needs to be upgraded. Download and
install HyperTerminal Private Edition v4.0 or greater from
Hilgraeve’s website http://www.hilgraeve.com.
f. Select File > Save. This creates a file with an .ht extension
using the name that was entered in Step b.
3. Set the date by typing “DATWmm/dd/yy” where mm/dd/yy is the
current month, day and year (e.g., DATW02/24/03 for February 24,
2003). Press Enter on the computer’s keyboard to update the Hi Flow
Sampler’s date display.
4. Set the time in 24 hour format by typing “TIMWhh:mm:ss” where
hh:mm:ss is the current hour, minute and second (e.g.
TIMW18:00:00 for exactly 6 o'clock PM). Press Enter on the
computer’s keyboard to update the Hi Flow Sampler’s time display.
5.9 Resetting the Microprocessor
Should the instrument ‘lock up’ or fail to turn OFF, reset the
microprocessor by disconnecting the battery pack for several seconds.
60
Instruction 0055-9017
Hi Flow Sampler Troubleshooting
6 Troubleshooting
6.1 Instrument Repair
It is recommended that field repair of the Hi Flow Sampler be limited to:
All other repairs should be performed by an authorized Bacharach
Service Center. Any repairs performed by an unauthorized service
organization will void the instrument’s warranty and release Bacharach,
Inc. of any implied or written product liability.
Before returning an instrument for repair, you may be able to determine
and resolve a problem using the Troubleshooting Guide in Section 6.2.
6.2 Troubleshooting Guide
The following table lists the causes and remedies for most of the
problems that may arise with the instrument.
If none of the suggested remedies resolves the problem, or for help with
any problem that is not listed, contact one of Bacharach Service Centers
listed in Section 7.3.
TABLE 6-1. TROUBLESHOOTING GUIDE
Symptom Probable Cause & Remedy
Instrument completely nonfunctional; won’t
turn ON when the I/Okey is pressed.
a. Battery pack dead. Recharge battery pack
per Section 5-2.
b. Loose battery connector. Remove the
instrument’s cover and ensure that all wires
are attached to the battery connector, and
that the connector is securely attached to the
printed circuit board.
c. Microprocessor needs reset. Disconnect
battery pack for several seconds.
Display freezes during start up. Battery pack voltage low. Recharge battery
Instrument turned itself OFF during
operation.
“Low Battery” message appears at bottom
of display.
Instruction 0055-9017
pack per Section 5-2.
Battery pack dead. Recharge battery pack per
Section 5-2.
Battery voltage is below 4.0 volts. Recharge
battery pack per Section 5.2.
61
Troubleshooting Hi Flow Sampler
TABLE 6-1. TROUBLESHOOTING GUIDE (Cont.)
Instrument won’t respond when a control
unit key is pressed.
Microprocessor needs reset. Disconnect
battery pack for several seconds.
Low flow rate reading with fully charged
battery.
Flow path obstructed. Check flow path and
verify flow rate per Section 5.7.
Slow response time. Internal filters are dirty. Replace filters per
Section 5.2.
Battery power does not last at least
Battery pack defective. Replace.
4.5 hours.
Erratic gas readings. a. The entire gas leak is not being captured.
Use the proper attachment to capture leak.
b. Faulty sensor(s). Calibrate both CH
4
sensors per Section 4. Replace any sensor
that fails to calibrate.
Sensor does not calibrate properly. a. Wrong calibration gas or insufficient flow
being applied to sensor. Ensure that the
calibration setup is correct.
b. Depleted sensor. Replace sensor and
recalibrate.
“Calibration Failed” message appears at
bottom of display.
Calibration parameters not within
programmed limits due to either insufficient
calibration gas flow, or new calibration value
was greater than 50% of the old value. Check
calibration setup and re-perform the
calibration procedure. Replace sensor if
necessary.
“Memory Full” message appears at bottom
of display.
“No Records Saved” message appears at
bottom of display.
“Check Background Pump” message
appears at bottom of display.
“Check Leak Pump” message appears at
bottom of display.
“Check Background Calibration” message
appears at bottom of display.
“Check Leak Calibration” message appears
at bottom of display.
1000 test records have already been stored in
memory. Save all test records per Section
3.15.1 and/or erase memory per Section
3.15.3.
A send records command was given with no
records stored in memory.
Flow to the background-gas pump has been
severely reduced. Check that the background
hose is not kinked or blocked, or the internal
tubing has not become blocked (see Figure
5-10). Also check the condition of
background-gas sensor filter, replace as
necessary per Section 5.4.3.
Flow to the leak-gas pump has been severely
reduced. Check that the internal tubing has
not become blocked (see Figure 5-10). Also
check the condition of leak-gas sensor filter,
replace as necessary per Section 5.4.2.
The background-gas sensor needs calibrated
per Section 4.
The leak-gas sensor needs calibrated per
Section 4.
62
Instruction 0055-9017
Hi Flow Sampler Troubleshooting
TABLE 6-1. TROUBLESHOOTING GUIDE (Cont.)
“Check Background Offset” message
appears at bottom of display.
The background-gas sensor did not zero
properly during start-up. Switch OFF
instrument and restart in an area with clean
air. Replace sensor if necessary.
“Check Leak Offset” message appears at
bottom of display.
The leak-gas sensor did not zero properly
during start-up. Switch OFF instrument and
restart in an area with clean air. Replace
sensor if necessary.
“Check Leak Rate Difference” message
appears at bottom of display.
“Check Zeroing Bypass” message appears
at bottom of display.
The calculated difference between leak rates 1
and 2 was greater than 10%.
The ESC key was pressed during startup,
thus bypassing the normal sensor zeroing
process. Although the instrument will operate
under this condition, no leak rate tests
should be made because of the potential
inaccuracy of the sensor channels.
“Check Background Greater Than Leak”
message appears at bottom of display.
The measured background gas level is greater
than the measured leak gas concentration.
Make sure the background gas sample hose
is not positioned too close the leak source.
“Low Battery” message appears at bottom
of display.
Battery voltage is below 4.0 volts. Recharge
battery pack per Section 5.2.
6.3 Error Codes in Saved Test Records
When saved test records are sent to a personal computer and later
viewed in a spreadsheet program (refer to Section 3.16 Importing Saved Data into a Spreadsheet), Field 18 “Error Codes” in the saved records
contains letter-codes that identify problems that have occurred during
that particular test. Refer to Table 6-2 for a listing and description of
these codes:
TABLE 6-2. FIELD 18 ERROR CODES
Error
Code
A Background gas sample pump blocked.
B Leak gas sample pump blocked.
C The background gas sensor needs calibrated at 2.5% and/or 100% CH4.
D The leak gas sensor needs calibrated at 2.5% and/or 100% CH4.
E The instrument failed to zero the background gas sensor during start up.
F The instrument failed to zero the leak gas sensor during startup.
G The user pressed the ESC key during start up before the sensors had a chance to
completely zero.
H The leak rate measurement #1 – #2 calculation failed.
I The measured background gas level was greater than the measured leak source gas
level. (This is an indication of improper equipment setup.)
Description
Instruction 0055-9017
63
Troubleshooting Hi Flow Sampler
Notes:
64
Instruction 0055-9017
Hi Flow Sampler Parts & Service
7 Parts & Service
7.1 Replacement Parts
Complete Hi Flow Sampler Assembly ...................................... 0055-8020
Standard Attachments:
3-Port Background Gas Sampler ....................................... 0055-0271