Bacharach H-10PM Owner's Manual

Page 1
H-10PM
Congratulations! You have just purchased the most versatile, high performance service leak detector available in the industry today. It will detect refrigerants with a sen­sitivity to effectively pinpoint both small and large leaks quickly and efficiently. We encourage your review of this manual to assure satisfac­tory performance and a long service life.
Refrigerant Leak Detector
Page 2
WARNING: This deviceis not to be used in any application that is beyond its intended purpose or beyond the scope of its specifications. For details on appropriate use, refer to the rest of this manual. Before risking equipment damage or personal injury, contact Bacharach if you are unsure of the validity of a particular combustion analyzer application.
How the H-10PM Works
The sensor in the H-10PM uses positive ion emission technology, commonly known as a heated diode. It is very sensitive to only halon substances (refrigerants) making this product highly resistive to false alarming, while retaining superior sensitivity for pin­pointing the most difficult to find refriger­ant leaks. A pump inside the unit draws air through the sensor. Any presence of halogen gases (such as refrigerants) causes an ionized current to flow that sounds a speaker and illuminates a neon light in the probe. Sensitivity to pinpoint both large and small leaks can be controlled by setting the three position Leak Size switch.
Battery Charge
Light
Calibration
Reference
Bottle
2
Charger
Plug
Power
Switch
Low Battery
Indicator
Leak Size
Switch
Adjustment (turn
to increase heat)
Manual Balance
Panel Operating Controls
Heater
clockwise
Mode Switch
MANUAL /
AUTO
Sensor
Heater Adjust
LEDs
Control
Page 3
Before you use the unit the first time . . .
1. Remove the sticky-label disc from the cap of the Calibration Reference Bottle. This disc can be reused to help seal the vial when the leak detector is not in use, or it can be discarded. DO NOT attempt to unscrew the cap, you may damage the bottle! It is filled with enough refrigerant to last approximately six months.
2. With the Power switch in the OFF posi­tion, charge the internal battery using the furnished wall adapter. Charge time is 8–10 hours or, until the red Full Charge LED turns green.
Adjusting the Sensitivity
Set the Leak Size switch as follows: SMALL switch position is the highest sen­sitivity. The unit will indicate 0.1 ounce or greater leak rates and is used for fluorine based gases like R134a and SF6. This posi­tion also assures highest repeatability for locating 0.5 ounce or greater R134a leaks (per SAE Std. J1627 moving probe test con­ditions). It must always be used to verify performance and calibration when using the Calibration Reference Bottle. MEDIUM switch position is used for chlorine and chlorine based gases like R12, R22, R123, etc. and will indicate a 0.1 ounce per year or greater leak rate. This position supports finding a 0.5 ounce or greater leak during the moving probe test as defined in SAE J1627. It is also useful for locating larger R134a leaks of one ounce or greater. LARGE switch position is used to zero in on large leaks of any refrigerant. If a large leak is suspected, switch unit to its Manual mode. Adjust the Manual Balance control for 2-3 ticks per second and slowly
3
Page 4
approach the test equipment/area. Adjust the Manual Balance control counterclock­wise as necessary to maintain 2-3 ticks per second. As you approach the vehicle or equipment, the gas concentration will increase, causing an alarm condition. Each time an alarm occurs, readjust the Manual Balance control. Continue this process until the leak is located. Blowing out the test site with shop air may enable you to locate the leak more quickly.
NOTE: After the large leak is located and repaired, blow out the area again with shop air, set the unit on the small leak sensitivity and double check equipment for small leaks.
Operating Instructions
1. Slide mode switch to AUTO position.
2. Slide power switch to ON position and sensitivity switch to SMALL position.
3. Check Low Battery LED. If it glows red, the battery needs charging, or you may operate the unit from a 12VDC source, or with optional AC Adapter of the correct voltage and current.
4. Check for sufficient airflow by pointing the probe tip toward the floor, covering it with your finger, then releasing your finger. If proper flow exists, the red ball should noticeably rise up into the probe when you uncover the probe tip. Note that the actual height and final resting position of the red ball are not important.
If the airflow ball does not rise:
- Tap the probe lightly to ensure the
ball is not sticking.
- Check the filter in the probe tip, as
described in the Maintenance section. If the flow is still insufficient, then the
unit should be forwarded for repair to the nearest Authorized Service Center.
4
Page 5
5. Allow two minutes for the sensor to warm up, after which the flashing probe light and sound indicator will idle at 2-3 clicks per second.
6. Test operation by quickly touching the top of the calibration reference bottle (make sure sticky label is removed). The unit should respond with a rapid flash rate and sound verifying correct opera­tion and optimum sensitivity for detecting all refrigerant leaks. During product life, if unit does not respond correctly, see “Heater Adjustment” section.
5
Page 6
WARNING: Personal Injury Hazard Do not use the H-10PM in an explosive or com­bustible atmosphere. The ambient atmosphere is drawn through the probe and through the sensor, which operates at approximately 1472°F (800°C). The resulting hot mixture of air and combustible gas could explode.
YOU’RE READY TO FIND THE LEAK !
(You have a choice of two modes)
Auto Mode
This mode is preferred because the unit effectively auto zeros to block out back­ground levels of refrigerants and changes in background contamination. This greatly reduces and/or eliminates false alarms while retaining sensitivity to quickly locate small or medium size leaks. Pinpointing leaks in this mode requires continuous probe movement. (We recommend 2 inches per second as specified in SAE J1627 and J1628). If the probe is held station­ary over the leak the unit will zero out the leak, going into the idle 2-3 ticks/second condition. Briefly moving the probe away from the leak, for 1-2 seconds, permits the unit to re-establish sensitivity. Returning to the leak site, the unit will alarm again. Continuing this procedure will reliably and repeatedly pinpoint the leak with each pass over the leak site. If a large leak is present, the auto zero cir­cuit may reduce sensitivity to an unaccept­able level to find small and medium sized leaks. If this condition exists, the manual mode will enable you to pinpoint the large leak. The manual mode is also an effec­tive means to determine if a large leak of any refrigerant exists prior to searching for leaks (see “Adjusting the Sensitivity”).
6
Page 7
Manual Mode
In the Manual Mode the auto-zero circuit is disabled. The unit will not zero out the leak if the probe is held over the leak site. This mode may require frequent read­justment of the Manual Balance control to maintain the required 2-3 ticks per second that indicates proper adjustment and calibra­tion for all three sensitivity switch positions. In this mode you may notice sensitivity is slightly greater (10-20%) than the auto-zero mode. This is normal.
Heater Adjustment
(for optimum performance and calibration)
The Calibration Reference Bottle sensor heater circuit controls the sensitivity of the unit. A heat setting that is too high causes instability due to excessive sensitivity and shortens sensor life. The LOW / OK / HIGH LEDs in conjunction with the Heater Adjustment and Calibration Reference Bottle is a unique system for setting the correct sensor heat (sensitivity) for optimum perfor­mance and long sensor life. The unit is pre-set for optimum initial operation. To check the setting, slide the mode switch to AUTO, turn ON the unit and allow it to stabilize (approximately 2 min­utes). When stabilized, at 2-3 clicks per sec­ond, briefly touch the Calibration Reference Bottle with the probe (after removing the sticky label disc from the top of the bottle). If adjusted properly the red LOW LED will go out and the green OK LED will briefly glow. This indicates the sensor’s heat/sensi­tivity is adjusted for optimum performance. If the red LOW LED remains on when you briefly touch the bottle, the sensor heat is set too low and the Heater Adjustment must be turned slightly clockwise. Allow unit to stabilize about one minute and retest. Repeat
7
Page 8
this procedure until the green OK LED briefly glows. If the red HIGH LED glows, the heat is set too high and the Heater Adjustment should be turned slightly counterclockwise. Allow the unit to stabilize and then repeat the test. Continue this procedure until the proper green OK LED indication has been achieved.
NOTES:
A. After initial check for correct adjustment, disregard the LOW / OK / HIGH LEDs. Their indication is meaningless during subsequent leak testing activity.
B. Check for proper heat adjustment the first time the unit is turned on and repeat on a daily basis. This assures the product is calibrated for the correct sensitivity in all leak size switch positions for your daily test activity.
C. Frequency of sensor heat adjustment is a function of how much exposure the sensor has to refrigerant. Adjustment may be required every couple of weeks for heavy duty service and once every few months for light duty service. The sensor LED status indicators will accurately tell you when adjustment is necessary during the daily check.
D. Over the usable sensor life, when heater adjustment is fully clockwise and the green OK LED will not come on, it is time to replace the sensor. Before plugging in a new one, turn the Heater Adjustment fully counterclockwise then plug in the new sensor and test and adjust as described above to calibrate the instrument.
Application Notes
1. Start leak checking anywhere and con­tinue in a logical progression through the entire system to locate all leaks. If sur­faces are particularly dirty, or wet, wipe them off with a clean, dry cloth to reduce
8
Page 9
filter clogging and extend sensor life. DO NOT allow the unit to ingest moisture. Use of the supplied rubber probe tip helps prevent moisture from being drawn into the unit. Check ports for moisture before inserting the probe.
2. After locating and repairing any leaks requiring the use of the LARGE switch setting, switch to the MEDIUM setting and verify the system is free of leaks. For HFC refrigerants only (such as R134a), you also must verify the system is leak free using the SMALL setting. For CFC and HCFC systems (such as R11, R12, R22, R123), the MEDIUM setting is suf­ficient to verify that the system is free of leaks that require repairing.
Caution: Equipment Damage Hazard A. Submerging the probe in liquid will damage the air pump. B. Exposing the probe to a stream of pure refrig­erant coming out of a tank will severely reduce the life or destroy the sensor. Life of this rare­earth sensor technology is directly proportional to the amount of refrigerant that passes through it. With normal intermittent service, the sensor can last for one year or longer. C. Exposure to high concentrations of refrigerant may require adjustment of sensor heat. To check, repeat the test procedure in “Heater Adjustment” section.
9
Page 10
Maintenance
(replacing the filter and airflow indicator)
To protect the air pump from damage due
to foreign particles and moisture, replace the filter as it becomes dirty. With mod­erate use (15 to 30 minutes a day), it is recommended that the filter be replaced once a month. In dirty environments or with heavy use, replace the filter more frequently. Always replace the filter when it is visibly dirty or wet.
Replacing the Filter:
1. Remove the black rubber probe tip.
2. Pick out the filter with a pin or tweezers. A fine screen will remain in the tip of the probe.
3. Insert the new filter in the probe tip. Make sure that the filter is firmly seated against the fine screen.
4. Replace the rubber probe tip.
Filter
Probe Tip
Airflow
Indicator Ball
Rubber
Probe Tip
Replacing Airflow Ball Indicator:
1. Remove the clear plastic section of the probe tip by gently pulling and twisting the white grip.
2. Turn the probe tip upside-down and tap on it to remove the old airflow indicator ball.
3. Insert the new airflow indicator ball into the tip.
4. Reattach the probe tip to the probe assembly.
10
Page 11
Replacing the Sensor:
The sensor is a plug-in unit that can be thrown away when expended. It needs replaced when the H-10PM no longer responds to the Calibration Reference Bottle, even with the Heater Adjustment turned fully clockwise (make sure the reference bottle contains some refrigerant).
Caution: Personal Injury Hazard Sensor surface temperature may reach 390°F (200°C) during operation, and may cause a burn if not allowed to cool.
1. Turn the leak detector OFF.
2. Unplug the power cord if the unit is plugged in and open the sensor cover.
3. Allow the sensor to cool before touching it.
4. Unplug and discard the sensor.
5. Insert a new sensor and close the sensor cover.
6. Turn the Heater Adjustment to its full counterclockwise position.
7. Apply power to the unit and adjust the sensor’s heat setting (see “Heater Adjustment” section).
Troubleshooting
Problem Responds to reference leak in
LARGE leak switch position
Cause Heat Adjustment set too high
Solution Readjust heater (see “Heater
Adjustment” section)
continued on next page
11
Page 12
** Solution # corresponds to Cause # **
Problem No response to Calibration
Reference bottle
Cause
1
Heat Adjustment set too low or
bottle is empty
2
No air flow (indicator ball in
probe doesn’t float)
3
Sensor exposed to excessive
amounts of halogen gas
4
Water in probe
Solution
1
Readjust heater (see “Heater Adjustment” section) or replace bottle (3015-0864)
2
Replace filter in probe tip (3015-0784). Check air pump operation.
3
Move probe to clean atmo­sphere for several minutes while sensor purges itself.
4
Turn unit off and disconnect probe from chassis. Remove
screws and take out chassis.
Look at the underside of the leak detector and follow the probe to the air pump. Remove this hose from the pump. Blow clean air (5 psi) into the probe tip for one or two minutes. Reassemble unit and replace the filter (3015-0784).
Problem Probe tip lamp does
not flash
Cause Burned out or damaged probe
lamp
Solution Return for repair
12
Page 13
** Solution # corresponds to Cause # **
Problem Erratic response in all leak posi-
tions
Cause
1
Filter is clogged
2
Dirt in sensor
3
Short circuit in sensor
4
Atmosphere contaminated with excessive refrigerant gas.
Solution1Replace the filter (3015-0784)
2
Remove sensor and blow it out with clean air (not over 10psi). If unsuccessful, replace the sen­sor (3015-0486). Replace the filter (3015-0784).
3
Replace sensor (3015-0486)
4
Ventilate area
Problem Responds continuously only in
SMALL leak switch position
Cause Excessive sensitivity
Solution Readjust heater (see “Heater
Adjustment” section)
13
Page 14
Ordering Information
3015-0627 H-10PM Refrigerant Leak Detector
w/ Battery & Charger
3015-0326 14 inch Flexible Probe Extension
3015-0119 100-240VAC Charge/Run AC
Adapter
3015-0895 Battery Clamp / Cigarette Adapter
assembly
3015-0641 Clear Probe Tip
3015-0680 Rubber Probe Tip
3015-0486 Replacement Sensor
3015-0781 Tune-up Kit: Sensor, 100 Filters,
3 Airflow Balls, 3 Rubber Probe Tips, and Calibration Bottle
3015-0784 Maintenance Kit: 100 Filters,
3 Airflow Balls, 3 Rubber Probe Tips, and Calibration Bottle
3015-0737 Maintenance Kit:
12 Filters, 3 Airflow Balls
3015-0864 Calibration Reference Bottle
3015-0103 Battery
14
Page 15
Repair Information
Should it become necessary to repair your H-10PM, please contact an Authorized Service Center. Units should be carefully packed to prevent shipping damage and shipped prepaid.
Warranty Send units under warranty and a copy of the purchase receipt to an Authorized Service Center for repairs. Please include your com­pany name, telephone number including area code, and a contact name.
Non-Warranty Units that are no longer under warranty will be repaired to factory specifications for a nomi­nal fee. Please include your company name, telephone number including area code, and a contact name.
Limited Warranty The purchaser is warranted that this leak detec­tor will be free of defects in material and work­manship for 3 years from date of purchase. This warranty does not cover sensors, reference leaks, filters, airflow balls, lamps, batteries, or probe tips. Damages caused by the user will not be covered. If any defects are discovered dur­ing the warranty period, an Authorized Service Center will repair or replace the unit at their option.
The foregoing limited warranty is exclusive and in lieu of all other warranties, whether written or implied, and no warranty of mer­chantability or fitness for purpose will apply.
15
Page 16
Authorized Service Centers
United States
Bacharach Sales/Service Center 621 Hunt Valley Circle New Kensington, PA 15068 Phone: 724-334-5051 Fax: 724-334-5723 E-mail: help@mybacharach.com
Canada
Bacharach of Canada, Inc. 20 Amber St. Unit #7 Markham, Ontario L3R 5P4 Canada Phone: 905-470-8985 Fax: 905-470-8963 E-mail: bachcan@idirect.com
16
Page 17
Notes:
17
Page 18
Notes:
18
Page 19
Notes:
19
Page 20
New Kensington, PA 15068
Ph: 724-334-5000  Fax: 724-334-5001  Toll Free: 1-800-736-4666
Website: http://www.mybacharach.com  E-mail: help@mybacharach.com
621 Hunt Valley Circle
3015-0925
Rev 9 9/2012
Loading...