Bacharach Fyrite Classic Owner's Manual

INSTRUCTION 0011-9026
®
20
16
12
8
4
%CO
2
0
®
FYRITE
CO
Operation/Maintenance
Rev. 12 – December 2012
Gas Analyzer
and O2 Indicators
Printed in U.S.A.
Instruction 0011-9026 Page 1
® Registered Trademarks
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects in mater ial and manufac ture and will conform substan tially to Bacharach Inc.’s ap plicable specifi cations. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacha­rach’s option, of this Product or parts thereof returned to Seller at the factory of manufac ture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been given by Buyer to Bacharach Inc. within one (1) year after the date of delivery of this Product by Bacharach, Inc.
Bacharach, Inc. warrants to Buyer that it will convey good title to this Product. Bacharach’s liability and Buyer’s remedy under this warranty of title are limited to the removal of any title defects or, at the election of Bachara ch, to the replace­ment of this Product or parts thereof that are defective in title.
The warranty set forth in paragraph 1 does not apply to parts the Operating Instructions designate as having a limited shelf-life or as being expended in normal use.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED WARAN­TIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY OBLIGA TION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTACT OR TORT, WHETHER OR NOT ARISING FROM BACH A RACH’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies includ ing, without limita tion incidental or consequen tial damages. No agree ment varying or extending the foregoing warranties, remedies or this limitation will be binding upon Bacharach, Inc. unless in writing, signed by a duly authorized offi cer of Bacharach.
NOTE: Gaskets are considered consum able items and are excluded from the terms of this warranty.
Register Your W arranty by V isiting
www.mybacharach.com
CAUTION
Corrosive fl uid is contained inside this unit. These instructions should be thoroughly read before operation is attempted.
Copyright © 1998-2012 Bacharach, Inc., all rights reserved.
FYRITE SAFETY PRECAUTIONS
®
20 16
12
8
%CO
2
20
®
16
12
4
%CO
2
20
16
12
8
4
%CO
2
0
®
20
®
16
12
8
4
%CO
2
0
The operator (s) of this instrument should thoroughly familiarize themselves with the applicable safety precautions before handling or using the FYRITE for gas analysis.
Review Figures 1a,1b, 1c and 1d on this page. Make certain to follow the steps out­lined below and read the fl uid handling precautions for your personal safety.
1) Always use protective equipment such as safety goggles, gloves, and protective clothing as illustrated in Figure 1a.
2) Read the instruction manual thoroughly. When fi lling the FYRITE with the appli- cable fl uid charge, use safety equipment and perform the operation in the vicinity of running tap water (Figure 1b).
3) The FYRITE fl uid should always be in the bottom reservoir before and during sample introduction to the FYRITE. When venting the FYRITE to atmosphere, hold the instrument in the vertical position and at a location away from your face (Figure 1c).
4) NEVER vent FYRITE in the inverted position (with the plunger facing down­ward). This will cause fl uid to spill, which is corrosive and contains harmful ele- ments (Figure 1d).
NOTE: In the event of uid spill - refer to Section 2.2 (FYRITE Handling
Precautions). For Maintenance - refer to Section 6.0.
Figure 1a. Protective equipment required before attempting gas analysis.
Figure 1c. Venting FYRITE to at­omsphere at 45° angle.
Figure 1b. Preparing to fi ll FY- RITE using recommended protec­tive safety equipment.
Figure 1d. NEVER vent FYRITE in the inverted position, which will cause fl uid to spill.
TABLE OF CONTENTS
1.0 DESCRIPTION................................................................................................................................................1
2.0 FYRITE INSPECTION BEFORE AND DURING TEST ..............................................................................1
2.1 Pre-Operational Check .......................................................................................................................1
2.2 FYRITE FLUID HANDLING PRECAUTIONS ...............................................................................2
3.0 BASIC FYRITE OPERATION PERCENT CO
3.1 Determining CO2 and O2 FYRITE Fluid Strength .............................................................................7
AND 02 ...............................................................................3
2
3.2 Operating Precautions to Ensure Maximum Accuracy ......................................................................7
3.2.1 Draining Fluid Droplets .....................................................................................................8
3.3 FYRITE Temperature .........................................................................................................................8
3.3.1 Temperature of Gas Sample ...............................................................................................8
4.0 PRINCIPLE OF OPERATION ........................................................................................................................9
5.0 APPLICATION INFORMATION .................................................................................................................10
5.1 FYRITE Absorbing Fluid .................................................................................................................10
5.1.1 FYRITE Refi ll Fluid (Storage Life, etc.) .........................................................................10
5.2 FYRITE Operating Temperature Range ...........................................................................................11
5.3 Typical FYRITE Applications ..........................................................................................................11
5.4 CO2 FYRITE Only Combustion Testing ..........................................................................................12
5.4.1 Percent CO2 Shows Volume of Excess Air ......................................................................12
5.4.2 Too Much Excess Air Means Ineffi cient Combustion .....................................................13
5.4.3 What is Proper CO2 Value? ..............................................................................................13
5.4.4 How To Calculate Combustion Effi ciency .......................................................................13
5.5 O2 FYRITE Only Combustion Testing .............................................................................................13
5.5.1 Percent O2 Shows Volume of Excess Combustion Air ....................................................13
5.5.2 Too Much Excess Air Means Ineffi cient Combustion .....................................................13
5.5.3 What is Proper O2 Value? .................................................................................................14
5.5.4 Measuring O2 in Addition to CO2 ...................................................................................14
5.5.5 Where is Percent CO2 or O2 Measured? ..........................................................................14
5.6 Background Gases Affecting FYRITE Readings .............................................................................16
5.7 General FYRITE Applications .........................................................................................................17
5.8 Altitude Correction Table .................................................................................................................18
6.0 FYRITE MAINTENANCE ..........................................................................................................................18
6.1 Increasing FYRITE Fluid To Proper Level ......................................................................................18
6.1.1 Decreasing FYRITE Fluid to Proper Level .....................................................................19
6.2 Checking FYRITE Fluid Strength ....................................................................................................19
6.2.1 CO2 FYRITE Fluid Only .................................................................................................19
6.2.2 O2 FYRITE Fluid Only ....................................................................................................21
6.3 Replacing FYRITE Fluid ................................................................................................................21
6.4 Inspection of FYRITE for Fluid Leakage ........................................................................................22
6.5 Cleaning FYRITE .............................................................................................................................24
6.6 Replacing FYRITE Plastic and Rubber Parts ...................................................................................24
6.7 Aspirator Bulb - Sampling Assembly ...............................................................................................26
6.8 Storing FYRITE and Refi ll Fluid .....................................................................................................26
6.9 Replacing Saturator Filter Packing Part #11-0121 ..........................................................................26
7.0 ILLUSTRATED PARTS LIST FYRITE CO2/O2 ..........................................................................................29
7.1 FYRITE Illustrated Parts ..................................................................................................................30
7.2 FYRITE Sampling Assemblies ........................................................................................................31
INSTRUCTIONS 0011-9026
OPERATION/MAINTENANCE
BACHARACH FYRITE MODELS
CO2 and O2 INDICATORS
1.0 DESCRIPTION (Refer to Figure 1)
The FYRITE employs the well-known “Orsat” method of volumetric analysis involving chemical absorption of a sample gas, such as carbon dioxide or oxygen. The reagent used to absorb carbon dioxide (CO chloride (blue) is the absorbent for oxygen (O that the absorbing fl uid is also used as the indicating fl uid so that one vessel takes the place of both measuring burette and absorption pipette.
The body of the FYRITE is molded of clear high-strength plastic, comprises top and bottom reservoirs and a center tube connecting the two reservoirs. The bottom of the lower reservoir is sealed off by a synthetic rubber diaphragm that rests on a perfo­rated metal plate. The upper reservoir is covered by a molded plastic cap that contains a double -seated plunger valve. A spring holds this valve against a carefully fi nished seat in the top cap providing a perfect seal which makes the instrument spill-proof in any position. When the valve is partially depressed, the entire instrument is open to the atmosphere and the instrument is, of course, is no longer spill-proof.
The bottom reservoir is fi lled with the absorbing fl uid, which extends about 1/4 inch into the bore of the center tube when the instrument is held upright. The scale posi­tion, mounted to one side of the center tube, is adjustable.
) is potassium hydroxide (dyed red), and chromous
2
). The unique feature of the FYRITE is
2
2.0 FYRITE INSPECTION BEFORE AND DURING TEST
CAUTION
Corrosive fl uid is contained inside unit, instructions should be thoroughly read before operation is attempted.
2.1 Pre-Operational Check
To ensure proper FYRITE operation and reliable results, check the following:
(a) With FYRITE vented and in a vertical position, it should be possible to adjust
scale zero to the top of the fl uid column. Refer to FYRITE Operation for proper setup (Section 3.0 Steps 1 through 5). If this is not possible, add or remove a small amount of fl uid per Section 6.1.
Instruction 0011-9026 Page 1
(b) FYRITE Fluid Strength
To check fl uid strength.
NOTE: When repeating procedures as outlined in Section 3.0 Steps 7 through 9 (absorbing and reading percent O sorbing and reading percent CO2) and before venting FYRITE
and ab-
2
to atmosphere for next sample, if the reading increases by more than 1/2 percent for either CO
or O2, replace the fl uid.
2
It is often desirable to check fl uid strength before taking the FYRITE to a location where it will be used. It can be tested on sample atmospheric air (which contains
20.9% O
mum accuracy. (See Section 6.9)
O
samples containing 10% O
CO
proximately 350 samples containing 10% CO
). Be certain to wet Filter Saturator when checking atmospheric air for maxi-
2
(Oxygen) Fluid - Fresh FYRITE uid will absorb all O2 from approxi mately 100
2
(Carbon Dioxide) Fluid - Fresh FYRITE Fluid will absorb all CO2 from ap-
2
.
2
.
2
(c) FYRITE Sampling Assembly - Always check sampling assembly for proper (leak-
free) operation as outlined in Section 6.7 before pro ceeding with test.
(d) Make certain wool in Filter Saturator Tube is moistened with water if using
either O is not fully saturated with water vapor. Examples where fi lter-saturator must
or CO2 FYRITE in a noncombustible application where the gas sample
2
be wetted include measuring CO2 or O2 in ambient air or from compressed gas samples. Refer to Section 6.9 when replacing or moistening saturator (wool) fi lter is required.
FYRITE OPERATION PRECAUTION: To prevent fl uid spill, never depress Plunger Valve to vent FYRITE when unit is in the inverted or any other non-upright position. When de­pressing Plunger Valve in the upright position to vent FYRITE, hold at a slight angle away from the operator’s face.
2.2 FYRITE FLUID HANDLING PRECAUTIONS
CAUTION
FYRITE fl uids used in the CO and contain poisonous elements which must not be taken in­ternally. In the event of a spill or accidental body contact with FYRITE fl uid, read the following carefully.
Instruction 0011-9026Page 2
and O2 analyzers are corrosive
2
HAZARDS
Corrosive liquid causes burns. May cause blindness if splashed in eyes. Vapors are irritating and may be harmful.
PRECAUTIONS
Prevent contact with eyes, skin and clothing. Wear eye protection and gloves. Do not vent instrument until fl uid has drained from top well. Do not vent instrument (FY- RITE) in inverted position.
FIRST AID
For contact with eyes: Immediately fl ush eyes with water 20 minutes. Get immediate medical attention.
For contact with skin: Immediately fl ush skin with water 20 minutes. Get immediate medical attention.
For inhalation overexposure: If irritation develops, move victim to fresh air.
For swallowing: Do not induce vomiting. Give two glasses of water or milk if conscious and not convulsing. Get immediate medical attention.
Refer to the material safety data sheet for further information.
3.0 BASIC FYRITE OPERATION PERCENT CO2 AND 0
2
Four basic steps are required when making an analysis, many being common to both the CO
FYRITE MODEL CO (Red Fluid) (Blue Fluid)
and O2 FYRITE.
2
FYRITE MODEL O
2
2
(a) Vent and adjust scale zero. (a) Vent and adjust scale zero.
(b) Pump sample into FYRITE. (b) Pump sample into FYRITE.
(c) Absorb CO2 from sample. (c) Absorb O2 from sample.
(d) Read % CO2 on scale. (d) Read % O2 on scale.
NOTE: To prevent warming of FYRITE fl uid during analysis, hold FYRITE by the fi ns only.
Instruction 0011-9026 Page 3
CO2 FYRITE OPERATION O2 FYRITE OPERATION (RED FLUID) (BLUE FLUID)
1. Hold upright (Fig. 1) and
away from face. Depress Plunger Valve (momen­tarily) to vent FYRITE, and release.
2. Invert FYRITE (Fig. 2).
Hold at slight angle to drain fl uid into top reser- voir.
3. Turn upright. Hold FY-
RITE at 45° angle (Fig. 3) momentarily to allow uid droplet drainage into bottom reservoir.
Figure 1
Figure 2
1. Hold upright (Fig. 1) and away from face. Depress Plunger Valve (momen­tarily) to vent FYRITE, and release.
2. Invert FYRITE (Fig. 2) to absorb O RITE whenever Plunger
drawn into FY-
2
Valve is depressed). Hold at slight angle to drain uid into top reservoir.
3. Turn upright. Hold FYRITE at 45° angle to drain fl uid into bottom reservoir. Turn upright. Repeat sequence of Steps 1, 2, and 3 twice until uid level does not drop more that 1/2% when Plunger Valve is de­pressed. Turn upright. Hold FYRITE at 45° an­gle (Fig. 3) momentarily until excess fl uid droplets have been drained from inside surfaces.
4. Hold FYRITE in upright (Fig. 4) position and away from face. Depress Plunger Valve (momen­tarily) and release.
Instruction 0011-9026Page 4
Figure 3
4. Hold FYRITE in upright (Fig. 4) position and away from face. Note fl uid level in column. Depress Plunger Valve (momen­tarily), and release.
Figure 4
CO2 FYRITE OPERATION O2 FYRITE OPERATION (RED FLUID) (BLUE FLUID)
5. Holding FYRITE upright (Fig. 5), loosen locknut at rear of scale. Slide scale (Fig. 5a) until top of fl uid column lines up with zero line on scale (Fig. 5b). Tighten scale locknut.
NOTE: When setting
scale zero, hold FYRITE vertically as shown and level with eyes while sighting across scale to top of fl uid column.
6. To pump gas sample into FYRITE (Fig. 6), insert open end of metal sampling tube into area bearing gas for analysis. Hold FYRITE in upright position and place sam­pling assembly rubber connector tip over the Plunger Valve.
Depress Plunger Valve
rmly with connector tip.
Figure 5
Figure 5a
Figure 5b
5. Holding FYRITE upright (Fig. 5), loosen locknut at rear of scale. Slide scale (Fig. 5a) until top of fl uid column lines up with zero line on scale (Fig. 5b). Tighten scale locknut.
NOTE: When setting
scale zero, hold FYRITE vertically as shown and level with eyes while sighting across scale to top of fl uid column.
6. To pump gas sample into FYRITE (Fig. 6), insert open end of metal sampling tube into area bearing gas for analysis. Hold FYRITE in upright position and place sam­pling assembly rubber connector tip over the Plunger Valve.
Depress Plunger Valve
rmly with connector tip.
Pump sample by squeez-
ing and releasing as­pirator bulb 18 times. During 18th bulb squeeze (with bulb held defl ated) release connector tip and Plunger Valve.
Pump sample by squeez-
ing and releasing as­pirator bulb 18 times. During 18th bulb squeeze (with bulb held defl ated) release connector tip and Plunger Valve.
Figure 6
Instruction 0011-9026 Page 5
FYRITE OPERATION O2 FYRITE OPERATION
CO
2
(RED FLUID) (BLUE FLUID)
7. Absorb sample gas into FYRITE by inverting until fl uid drains into top reservoir (Fig. 7). Then turn upright (Fig. 8) to drain fl uid into bottom reservoir. Repeat this step once.
8. Hold FYRITE at 45° angle (Fig. 9) momentari­ly to allow fl uid droplets to drain into the bottom reservoir.
9. With FYRITE held upright (Fig. 10), per­mit fl uid in column to stabilize a few seconds, then immediately read % carbon dioxide on scale at the point corresponding to top of the fl uid column.
7. Absorb sample gas into FYRITE by inverting until fl uid drains into top reservoir (Fig. 7). Then turn upright (Fig. 8) to drain fl uid into bottom reservoir. Repeat this step three (3) more times (four complete inversions total).
Figure 7
8. Hold FYRITE at 45° angle (Fig. 9) momentari­ly to allow fl uid droplets to drain into the bottom reservoir.
Figure 8
9. With FYRITE held upright (Fig. 10), per­mit fl uid in column to stabilize a few seconds, then immediately read % oxygen on scale at the point corresponding to top of the fl uid column.
Instruction 0011-9026Page 6
Figure 9
Figure 10
10. This step completes CO2 or O2 FYRITE gas sample analysis. A few points to re­member when reading the FYRITE: (a) FYRITE accuracy is within ±1/2% CO2 or O2 compared to actual value.
(b) Always handle FYRITE by fi ns to ensure body heat is not absorbed by fl uid.
(c) A delay in reading of 5 or 10 seconds may decrease accuracy of reading
slightly but longer delays may cause substantial error.
(d) The FYRITE is calibrated to indicate on a “dry” basis for fl ue gas samples,
which are normally fully saturated with moisture. Failure to artifi cially moisten wool packing in Filter Saturator Tube when actual sample is not fully saturated will cause FYRITE to read slightly low.
Therefore, make certain wool packing in Filter Saturator Tube is suffi ciently
moistened with water for accurate results on non-fl ue gases. Avoid excessive moisture which can be drawn into the Aspirator Bulb and forced into the FYRITE during sampling.
CAUTION
Never depress plunger valve to vent FYRITE in the inverted position. This will cause fl uid to spill, which is corrosive and contains poisonous elements. In the event of a spill, read In­structions in Section 2.2 FYRITE fl uid handling precautions.
3.1 Determining CO2 and O2 FYRITE Fluid Strength
FYRITE Fluid strength can be conveniently checked immediately after the fi rst sample reading. Without venting, repeat the absorbing operations (Steps 7 through
9) by inverting FYRITE again and positioning upright until all fl uid drains to the bot-
tom reservoir. Observe scale reading. An increase of more than 1/2% CO second reading as compared to the fi rst indicates a need for fl uid replacement.
or O2 in the
2
3.2 Operating Precautions to Ensure Maximum Accuracy
Locate top of fl uid column (refer to Fig. 11). The surface at the top of the fl uid column (meniscus) in the small, center bore will be dish-shaped as shown in Figure 11. Either the high or low point of this dish- shaped surface may be used to locate top of fl uid col- umn providing the same point is always used both for setting scale zero and reading percent CO other will cause an error.
Best practice is to use high point of this fl uid surface just at the small center bore wall. In setting scale zero or reading percent CO level with eyes and sight across scale to top of fl uid column.
Instruction 0011-9026 Page 7
or O2. Obviously, using high point for one operation and low point for the
2
or 02, hold FYRITE vertically and
2
3.2.1 Draining Fluid Droplets
®
20
16
12
8
4
%CO2
0
For maximum accuracy, it is important to form the habit of follow ing a standard procedure in this operation and to use the same procedure both before adjusting scale zero and before reading percent CO
or O2 .
2
3.3 FYRITE Temperature
The FYRITE temperature should be at or close to the temperature of the working environment where the analysis is being made and should not be subject to sudden temperature changes. If FYRITE is carried from cooler to warmer location or vice versa, expose FYRITE to new temperature condition to permit uniform temperature stabilization. Usually 15 or 20 minutes will suffi ce, except when extreme temperature variations are involved. Temperature stabilization can also be accelerated by invert­ing FYRITE frequently.
3.3.1 Temperature of Gas Sample
The gas sample as it enters FYRITE should be cooled (or warmed if sampling cool gases) to same (ambient) temperature as the FYRITE.
Capacity of Standard Aspirator Bulb and Sampling Assembly (Part # 11-7029) to cool the gas sample is suffi cient if the temperature of gas being sampled does not exceed 850°F (454°C), even where a large number of consecutive samples are required if there is an interval of not less than 5 minutes between samples.
FYRITE indicators operate over a wide temperature range. They may be exposed to ambient temperatures from –30° to 150°F (–34° to 66° C). Gases up to 850° F (454° C)
may be tested with Standard Aspirator Sampling Assembly (11-7029).
Instruction 0011-9026Page 8
Top of Fluid Column
Figure 11. Locating Top of Fluid Column.
Loading...
+ 28 hidden pages