Axis 4BAPA, 4BA, 4BAN, 4BANAN, 4BAFN User Manual

...
Page 1
USER MANUAL
BD2P balance
File: 2016-04-19 BD_P bW109 GB
AXIS, 80-125 Gdańsk, ul.Kartuska 375B, POLAND www.axis.pl
Page 2
2 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Table of contents:
1. General description .................................................................................................................................................................................... 3
2. Set .............................................................................................................................................................................................................. 3
3. Technical data............................................................................................................................................................................................ 4
4. Keys and indicators ................................................................................................................................................................................... 5
5. Security rules ............................................................................................................................................................................................. 6
6. Environment protection .............................................................................................................................................................................. 6
7. Preparing working environment ................................................................................................................................................................. 7
8. Preparing scale to work ............................................................................................................................................................................. 7
9. General operation principles ...................................................................................................................................................................... 8
10. Operation rules during work with accumulators (batteries) ....................................................................................................................... 9
11. Scale checking ........................................................................................................................................................................................... 9
12. Scale calibration (CAL) ............................................................................................................................................................................ 10
13. Connecting a computer, printer or label printer ....................................................................................................................................... 11
13.1 Detailed LonG protocol description ..................................................................................................................................................... 12
13.2 Detailed EPL protocol description ....................................................................................................................................................... 14
14. Start-up .................................................................................................................................................................................................... 15
15. Weighing with tare ................................................................................................................................................................................... 16
16. Scale menu .............................................................................................................................................................................................. 17
17. Menu navigation rules .............................................................................................................................................................................. 18
18. Scale setup (SEtUP) ................................................................................................................................................................................ 23
18.1 Scale calibration (CALIb) .................................................................................................................................................................... 25
18.2 Autozeroing function (AutoZEr) ......................................................................................................................................................... 26
18.3 Weight unit selection (UnIt) ................................................................................................................................................................. 27
18.4 Serial port parameters setting (SErIAL) ............................................................................................................................................. 28
18.5 Printout configuration (PrInt) ............................................................................................................................................................. 29
18.6 Setting backlight function (b_LIGHt) ................................................................................................................................................... 31
18.7 Analog out configuration (AnALoG) .................................................................................................................................................... 32
18.8 Entering reference zero value (ZErO) ................................................................................................................................................. 33
19. Special functions description .............................................................................................................................................................. 34
19.1 Products and users database (Prod i USEr) ....................................................................................................................................... 35
19.2 Pieces counting function (PCS) ......................................................................................................................................................... 39
19.3 Percentage weighing function (PErC) ................................................................................................................................................ 40
19.4 Label choosing function (LAbEL) ........................................................................................................................................................ 41
19.5 Weighing animals function (LOC) ....................................................................................................................................................... 42
19.6 Maximum value indication function (UP) ............................................................................................................................................ 43
19.7 Force measuring function (nEWton) .................................................................................................................................................. 44
19.8 Total weight function (totAL) ............................................................................................................................................................... 45
19.9 Checkweighing function (thr) .............................................................................................................................................................. 47
19.10 Setting date and time function (dAtE) ............................................................................................................................................ 50
19.11 Radio communication channel choice function (rF CHn)............................................................................................................... 51
19.12 Charging accumulators function (bAttErY)- option ........................................................................................................................ 52
19.13 Automatic switching off the scale function (AutoOFF) ................................................................................................................... 53
19.14 Statistical calculations function (StAt) ............................................................................................................................................ 54
19.15 Paperweight calculation (PAP) ...................................................................................................................................................... 57
Maintenance and repairs of small defects .............................................................................................................................................................. 58
Declaration of Conformity ....................................................................................................................................................................................... 59
Page 3
USER MANUAL 3
_____________________________________________________________________________________________________________________
1. General description
BD2P type scales are equipped with additional display for customer and are designed to weight parcels.
All scales are metrological tested - calibration or legal verification on demand.
Scales have following verification features:
a seal protecting scale casing against opening,
- notified body stamps and green metrological marking placed on the balance name plate.
Renewing of scale legal verification is required when protective seal is violated or after period of 3 years starting from 1st December of year when first legal verification was performed.
In order to renew legal verification please contact authorized service of AXIS.
NACE classification: : 29.24.23.
Certificates:
EC Type
Approval Certificate
No. PL 13 001
AXIS management
system certificate
DIN EN ISO 9001:2009
No. 90927/C/2
2. Set
Standard set consists of:
1. Scale
2. Feeder
3. User manual (CD)
4. Guarantee
Page 4
4 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
3. Technical data
Scale type
BD2P
Max load
2kg
Min. load
5g
Readout (d)
1g
Verification plot(e)
1g
Accuracy class
III
Working temperature
-10÷40°C
Tare range
-2kg
Weighing time
<3s
Pan size
225x165mm
Scale size
235x250x65mm
Supply
~230V 50Hz 6VA / =12V 0,16A
Scale weight
3,5kg
Page 5
USER MANUAL 5
_____________________________________________________________________________________________________________________
4. Keys and indicators
key
I/
- switch-on / switch-off (standby),
key
T
- tare (subtract package weight from weighed mass)
key
- special function/weighing switch, scrolling menu
key
0
- zero, scrolling menu
key
MENU
- special functions menu,
key
- print-out,
key
HR
- increased indication resolution,
indicator
0
- zero indicator
indicator
- result stabilisation indicator
indicator
NET
- net weight indicator (indication with subtracted tare)
indicator
MODE
- special function setting,
bar indicator
- total load indicator (graduated 0-100%)
indicator
OFF
- standby,
indicator
B/G
- gross mass (key B/G),
indicator indicator
Pcs
, lb, t,
oz, kg
- pieces counting
- weight unit
indicator
%
- % masy odniesienia
Max, Min, d, e, III
- metrological parameters and accuracy class.
OFF AUTMODE
0
50
100
SUM
Max Min e d
Page 6
6 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
5. Security rules
To avoid electrical shock or damage of the scale or
connected peripheral devices, it is necessary to follow the
security rules below.
All repairs and necessary regulations can be made by authorised personnel only. To avoid fire risk use a feeder of an appropriate type (supplied with the scale).
Pay attention that supply voltage is compatible with specified technical data.
Do not use the scale when its cover is opened. Do not use the scale in explosive conditions. Do not use the scale in high humidity. If the scale seems not to operate properly, unplug it from the mains and do not
use until checked by authorised service.
6. Environment protection
According to legal regulations it if forbidden to dispose
wasted electronic equipment in waste containers.
Please return wasted scale to the point of purchase or other company specialised
in recycling of wasted electronic components.
Page 7
USER MANUAL 7
_____________________________________________________________________________________________________________________
7. Preparing working environment
Location for the scale should be chosen with care in order to limit influence of the factors that can interrupt working scale. This location has to maintain proper temperature for working scale and necessary space for its operating. The scale should stay on stable table made of material that does not influence magnetically on the scale. Rapid air blows, vibrations, dust, rapid temperature changes or air humidity over 90% are not allowed in scale surrounding. The scale should be far from heat sources and devices emitting strong electromagnetic or magnetic fields.
8. Preparing scale to work
1. Take the scale out of the package removing protective
foils. It is recommended to keep the original package in order to transport the balance safely in future.
2. Take a pan off and remove protective elements from
below the pan.
3. Place the scale on a stable ground not affected by
mechanical vibrations and airflows.
4. Level the scale with the rotating rear legs so that the
air bubble in the water-level at the back of the scale is in the middle.
Correct Wrong
Page 8
8 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
9. General operation principles
1. In order to confirm correctness of the scale during its operation, before starting and after finishing every valid measurement series it is recommended to check weighing accuracy putting calibration weight or other object of exactly known mass on the scale. In the case when allowable measurement error of the scale is exceeded, it is recommended to perform calibration with external weight or contact authorised service centre.
2. Weighed mass should be placed in the middle of the pan.
3. The scale allows taring in the whole measuring range. To tare the scale press key. Taring does not extend measuring range, but only subtracts tare value from mass value of a sample placed on the pan. To make the control of a load on the pan easier and to avoid exceeding measurement range, the scales have load indicator calibrated 0÷100%.
4. Weighing result should be read when the indicator lights, which signalises result stabilisation.
5. When the scale is not used but it is necessary for it to be ready to work, it can be switched off by pressing I/ key. The scale reading system is then switched off and scale goes to standby mode signalled with OFF indicator. Switching the scale on is preformed by pressing I/ key.
6. In sales having key (zeroing) active it should be checked if zero indicator is displayed before sample is placed on the pan. If not, press key and wait until the scale is zeroed and zero indicator appears. After that load can be placed on scale pan.
7. Scale mechanism is a precise device sensitive to overweight, mechanical shocks and strokes.
Do not overload the scale more than 20% of maximum capacity. Do not press the pan with a hand.
For transportation pan, pan support and pan ring should be packed separately.
Page 9
USER MANUAL 9
_____________________________________________________________________________________________________________________
10. Operation rules during work with accumulators (batteries)
1. Scale can be powered from ~230V supply through feeder attached with scale. Moreover accumulators,
which are placed in container inside the scale, can be used for powering. Ordinary batteries can be used as well.
2. In order to make accumulators (batteries) discharging time longer, automatic switching off display
backlight and next the whole scale is possible during breaks in weighing. Configuration of these mechanisms is done using b_LIGHT and Auto OFF functions.
3. Charging accumulators is performed automatically after connecting feeder to the scale, also during
weighing. Accumulator power level can be read using bAttErY function (bat VoL option).
11. Scale checking
It is advised to check scale indication accuracy before and after series of measurement using any load with known weight. To check the scale with legal verification use a calibration weight with valid calibration certificate. In case permissible error is exceeded it is advised to contact the nearest service to calibrate the scale.
When using batteries in place of accumulators, charging during work with feeder have to be switched off. bAttErY function is used for this purpose (bAt OFF option), which is described in further part of manual. Charging batteries can cause their breaking and serious damage of the scale.
Page 10
10 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
12. Scale calibration (CAL)
If scale indications exceed permissible error it is necessary to adjust a balance. To calibrate a balance use Max mass calibration weight (see Max value in Technical Data table).
To adjust a scale it is necessary to break protective marks – please contact the nearest service for legal verification in a place of installation or the nearest Measurements Office.
Performing calibration requires changing adjustment switch position, which is placed behind protecting mark (sticker) of a notified body. An access to the switch is possible only after removing the mark. Therefore, scale calibration causes lost of legal verification and, in consequence, the necessity of renew legal verification in the nearest notified body or in place where the balance is used. Before proceeding with calibration, adjustment switch should be set to ON position using thin screwdriver. When calibration process, described on next page, is finished, adjustment switch should be set to OFF position using thin screwdriver (the balance will move to weighing).
ON
OFF
Adjustment switch
Mark
Page 11
USER MANUAL 11
_____________________________________________________________________________________________________________________
13. Connecting a computer, printer or label printer
The scale is equipped with RS232C, which can be used to connect external devices such as computer or a printer.
When cooperating with computer, the scale sends weighing result after initialize signal from computer or after pressing key on the scale.
When cooperating with a printer data is send automatically after result stabilisation, but next transmission is possible after removing previously weighted sample.
When cooperating with label printer after pressing key, the scale sends instructions set for the label printer. Label number 0001, hour, data (if the clock is installed and on) and nett weight. During transmission LabEL communicate is displayed. The way of sending data and transmission parameters is set using SErIAL special function.
Set of send data is set using special function PrInt.
The following data can be send:
- Header (scale type, Max, d, e, serial number),
- Operator identification number,
- Successive printout number (measurement),
- Identification number or product bar code,
- Number of pcs (PCS function only),
- Single detail mass (PCS function only),
- Nett weight,
- Tare (package mass),
- Gross weight,
- Total mass (Total function only).
If the scale is equipped with two serial joints Print function is set independently for both interfaces.
Computer must have a special program for cooperation with data from a scale. Dedicated programs are also offered by AXIS.
Except RS232C joint, the scale can be equipped with USB or Wi-Fi interface. Needed controllers and instruction can be found on a CD supplied with Axis scales.
Page 12
12 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
13.1 Detailed LonG protocol description
Communication parameters: 8 bits, 1 stop bit, no parity, baud rate 9600bps,
Readout of scale indication
ComputerScale: SI CR LF (53h 49h 0Dh 0Ah), ScaleComputer: scale response according to description below (16 bytes):
Byte
1 - sign „-” or space
Byte
2 - space
Byte
34
-
digit or space
Byte
5÷9 - digit, decimal point or space
Byte
10 - digit
Byte
11 - space
Byte
12 - k, l, c, p or space
Byte
13 - g, b, t, c or %
Byte
14 - space
Byte
15 - CR
Byte
16 - LF
Readout of actual indication
ComputerScale: Sx1 CR LF – initiaton signal ScaleComputer: scale sends 16 bytes (the same as SI commad)
Readout of stabilization indicator and actual indication
ComputerScale: Sx3 CR LF – initiaton signal ScaleComputer: scale send indicator S (stable) or U (unstable) + 16 bytes (the same as SI command).
Attention: Network number different than zero (SErIAL / nr function) changes scale working mode: communication with a computer is possible after logging the scale in with 02h scale number command. To log the scale out use 03h command. For example: Using a program to test RS232 interface ( program is available in www.axis.pl / programy
komputerowe ) for scale number 1 please write: $0201 to log in, then SI, and write: $03 to close
communication.
Asking about scale presence in system (testing scale connection with computer):
ComputerScale: SJ CR LF (53h 4Ah 0Dh 0Ah), ScaleComputer: MJ CR LF (4Dh 4Ah 0Dh 0Ah),
Displaying a inscription on scale’s display (text communicate from computer):
ComputerScale: SN n n X X X X X X CR LF, nn-displaying time in seconds; XXXXXX-6 signs to display
ScaleComputer: MN CR LF (4Dh 4Eh 0Dh 0Ah),
Scale tarring (calling
T
key press) :
ComputerScale: ST CR LF (53h 54h 0Dh 0Ah), ScaleComputer: without response,
Page 13
USER MANUAL 13
_____________________________________________________________________________________________________________________
Scale zeroing (calling
0
key press):
Computer Scale: SZ CR LF (53h 5Ah 0Dh 0Ah), Scale Computer: without response,
Scale turning on / off (calling I/ key press):
Computer Scale: SS CR LF (53h 53h 0Dh 0Ah), Scale Computer: without response,
Entering to special function menu (calling MENU key press):
Computer Scale: SF CR LF (53h 46h 0Dh 0Ah), Scale Computer: without response,
Setting threshold 1 value (option):
Computer Scale: SL D1...DN CR LF (53h 4Ch D1...DN 0Dh 0Ah) D1...DN – threshold value, maximum 8 characters („-” – negative value, digits, dot – decimal separator), number of digits after dot should be the same as on scale display,
Scale Computer: without response, Example:
in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order should be sent: S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah),
in order to set low threshold 100kg in scale B150 (d=50g) the following order should be sent: S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
Setting threshold 2 value (option):
Computer Scale: SH D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah), D1...DN threshold value, maximum 8 characters Scale Computer: without response.
Setting threshold 3 value (option):
KomputerWaga: SM D1...DN CR LF (53h 4Dh D1...DN 0Dh 0Ah), gdzie: D1...DN – threshold value, maximum 8 characters WagaKomputer: without response.
Connecting cable WK-1 (scale – computer / 9-pin interface):
Connection cable WD-1 (connects scale with AXIS printer):
SCALE
KAFKA PRINTER
SCALE
COMPUTER
Page 14
14 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Setting of internal switches of AXIS printer:
SW-1
SW-2
SW-3
SW-4
SW-5
SW-6
SW-7
SW-8
on
off
on
off
off
on
off
off
13.2 Detailed EPL protocol description
Transmission parameters: 8 bits, 1 stop bit, no parity, baud rate 9600bps,
After using key in scale: ScaleLabel printer : set of instruction in EPL-2 language that initialize label printing:
US -
Steering instruction
FR"0001"
- Label number define instruction
? -
Instruction that starts list of variable signs
mm:gg
- 5 signs: minutes:hour
rrrr.mm.dd
- 10 signs: year.month.day
masa
- 10 signs: scale indication+ mass unit
P1 -
Steering instruction
Attention:
1. Except variable signs constant signs can also be inscribed e.g. factory name, product name and so on.
2. In standard only one label pattern is possible to printout (number 0001). Using bigger amount of
patterns (other label numbers) is possible thanks to LAbEL special function.
3. To achieve label printout, label printer must have inscribed label pattern (label pattern is created on
computer and using computer it is saved to label printer memory). Label pattern is designed by ZEBRA
DESIGNER program which is supplied together with label printer.
4. Scales parameters and transmission protocol must corespond to label printer type.
Page 15
USER MANUAL 15
_____________________________________________________________________________________________________________________
14. Start-up
Leave the pan empty, plug the scale to the mains with a ground contact.
The scale proceeds with following start-up actions::
Display test.
Meter type displaying
Program version
The scale is ready to work.
Attention: UnLOAd communicate means that the scale is loaded or transportation security elements situated under the pan has not been removed.
0
0
kg
kg
Zasilacz
~230V
12V
Page 16
16 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
15. Weighing with tare
If the scale is not loaded and 0 indicator doesn’t indicate, press 0 key.
Zero indication and 0 indicator mean that the scale is ready to work.
After putting container (package) tare the scale using T key. NET indicator will show up.
Put on weighted object and readout net weight (NET indicator shows that scale indicates net weight).
In order to readout gross weight press key (B/G indicator shows that scale indicates gross weight). Press again key in order to come back to net indications.
00
kg
kg
kg
kg
kgkgkg
Page 17
USER MANUAL 17
_____________________________________________________________________________________________________________________
16. Scale menu
All scales except for basic metrological functions: weighing and taring, have many special functions and configuration options.
In order to ease using functions user can create his own (personalized) menu.
Creating personalized menu:
In „out of the box” scale after pressing MENU
key only SEtuP option (it contains all configuration options) is available.
One of the configuration options is Menu that is used to create personalized menu.
To add a function to personalized menu press
T
key when the function is indicating.
Chosen function is indicated with „o” sign on the
left side of display.
After adding all necessary functions press out in order to come back to weighing mode. User now after pressing MEnu key has access to selected earlier functions and to SEtuP option. dEFAULt option is used to set factory settings.
0
kg
Page 18
18 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
17. Menu navigation rules
Choosing menu options:
First position of scale menu shows up after pressing Menu key. The position is displayed for about 7 seconds and then the scale sequentially displays next menu positions.
Choosing menu position (option) is done by pressing T key when it is displayed on the screen.
After choosing position (option) usually several options show up:
on – turning on selected option,
OFF - turning off,
out – out to menu.
Accelerated working with menu:
Menu first position is displayed for about 7s. During this time user can view next positions by using  key (or 0 ).
Immediate out to previous menu level is done by using Menu key.
0
kg
0
kg
Page 19
USER MANUAL 19
_____________________________________________________________________________________________________________________
 key working method:
During standard weighing  key is used to switch between net and gross indication.
When special function e.g. PCS is turned on, using  key enables to go back to standard weighing mode.
Sign „o” on the left side signalizes that special
function is turned on and user can go back to function mode by pressing  key.
0
0
0
kg
0
0
0
kg
kg
kg
Page 20
20 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
HR key working method:
During normal weighing temporary (5s) readout resolution increase is possible.
Return to normal indication is made automatically.
Inscribing numerical values:
Inscribing numerical values is needed in some special functions e.g. tArE function requires to inscribe tare values.
Keys: 0 - increasing digit value,
- decimal point,
T
- next digit position,
MENU - end of inscribing.
0
0
kg
kg
5s.
0
kg
Page 21
USER MANUAL 21
_____________________________________________________________________________________________________________________
Menu diagram:
key
(PCS) (totAL) ... ...
out
SEtUP
Prod USEr PCS Unit PErc LOC tArE UP NEWton totAL thr StAt PAPEr out
()o()o
MEnu
CALIB
AutoZEr
UnIt
MENU
- personalized function menu created using SEtUP / MEnu
- sensitivity calibraton (only not verificated scales)
- fast calibration (without confirmation of putting weight)
- calibration with confirmation
- exit
- autozeroing
- autozeroing on
- autozeroing off (lasts 10min)
- exit
- settings
- creating menu (” “- added to menu)
- choosing product
- choosing user
- pieces counting function
- actual unit selection
- percentage conversion function
- animals weighing function
- tare memory bank
- maximum value function
- force and torque value function
- summation series of measurements function
- threshold values comparing function
- statistics function
- paper grammage function
- exit
o
CAL on CAL StP out
Aut on AUt OFF out
CArAt (ct) MGrAM (mg) HGrAm (kg) Pound (Ib) ounCE (oz) ounCE (ozt) GrAin PEnnYW GrAM (g) out
- unit choice
- carat
- miligram
- kilogram
- pound
- ounce
- aphotecary ounce
- grain
- jednostka jubilerska
- gram
- exit
Page 22
22 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
SErIAL
PrInt
(dAtE) (b_LIGHt)
AnALoG
- ustawianie daty godziny (if the scale is equipped with clock)
- setting backlight (if display is backlighted)
- backlight off
- backlight on
- turn off after 30 s without activity
- as above but only when supplied from accumulators
- exit
- analog out setting
- range value inscribing
- configuration (plus, minus, both)
- exit
- serial ports settings settings
- port - 1
- transmission speed (1200, ... ,115 200 bps)
- bits quantity (7 or 8)
- parity control
- transmission type (Stab, no StAb, Auto,Cont.,rEMoVE)
- protocol (LonG, EPL, EPL_A, EPL-d, PEn-01)
- cooperation with barcode readers MJ-4209
- port-2 (as above)
- printout data configuration / transmission
- port - 1
- header (scale type, Max, d, e, serial nr)
- scale operator id number
- user name
- succesive printout/measurement number
- product identification number
- product barcode number
- product name
- pieces quantity (PCS function)
- product unit weight
- net weight
- package weight (tare)
- gross weight
- total weight (totAL function)
- measurement number and value in one line printout
- exit
- port-2 (as above)
- exit
bAUd bItS PArItY SEndInG Prot SCAnn out
HEAdEr User Id User nA Prn no Prod Id Prod bA Prod nA Cont APW nEtto tArE Gross totAL nr LCD out
Port-1
Port-2 out
b_L oFF b_L on b_L ECO b_L bAt out
AnG rnG AnG CFG out
Port-1
Port-2 out
Page 23
USER MANUAL 23
_____________________________________________________________________________________________________________________
(bAttEry) (AUto OFF) (ZEro) dEFAULt SErVICE out
- turn on/off accumulator charging (if the scale is equipped with accumulator)
- automatic turning off - saving accumulator power (as above)
- scale start zero inscribing (factory zero)
- restore default settings for all options
- options only for service
- exit
Page 24
24 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
18. Scale setup (SEtUP)
SEtUP contains all options used for setting scale work mode:
MEnu – creating personalized user menu CALIb – scale sensitivity calibration AutoZEro(ing) – self-maintaining zero indication
(unloaded scale)
UnIt – weight unit selection SErIAL – setting serial ports Print – transmission (printout) data selection FILtEr – anti-disturbance filter b_LIGHt – backlight setting Ad420 – analogue out configuration FIrMW(are) – updating software (only for service) dEFAULt – reset to factory settings (sample of
using in chapter 15)
SErVICE – service menu (only for service)
.
0
kg
Page 25
USER MANUAL 25
_____________________________________________________________________________________________________________________
18.1 Scale calibration (CALIb)
Press MENU key.
Press T key when CALIb function appears.
The following options will be displayed:
-CAL on – calibration with external recommended standard of mass (see technical data).
-CAL StP – calibration with external weight, confirmation of succesive steps - MENU key, out – leave without changes
Press T key when CAL StP option appears (calibration in two steps).
Press T key when weight value used for
calibration is indicating or use othEr option and inscribe proper value ( keys 0 , ,
T
)
Press MENU and wait for writing zero to the
scale.
When LOAD message appears put standard of mass on the pan. Press MENU key (CAL on doesn’t need pressing MENU key).
Wait until internal calibration is finished and zero
indication is displayed.
0
0
g
g
g
1000g
Page 26
26 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
18.2 Autozeroing function (AutoZEr)
When the function is activated, the scale automatically ensures stable zero indication if the pan is empty or if zero indication was acquired by pressing T key. To turn on the function use MENU key and using
T
key choose AutoZEr and then Aut on
To leave the function press MENU key, then with T key chose AutoZEr and Aut OFF.
Note:
1. AUt sign occurs only in scales with LCD display.
2. In scales with 0 key active function changes name into AutoZE (autozeroing) and
works only when the scales is unbiased.
0
0
kg
kg
Page 27
USER MANUAL 27
_____________________________________________________________________________________________________________________
18.3 Weight unit selection (UnIt)
The function allows selecting weighing unit:
- CarAt (1 ct= 0,2 g) - carat,
- MGrAM (1mg=0,001g) milligram,
- KGrAM (1kg=1000g) kilogram,
- Pound (1 lb=453,592374g) English pound,
- OunCE (1oz=28,349523g) - ounce,
- OunCEt(1ozt=31,1034763g) pharmaceutical ounce,
- GrAIn (1gr=0,06479891g) - grain
- PennYW (1dwt=1,55517384g) jewellery mass unit,
- GrAM (1g) - gram.
The way of choosing carats as weighing unit is shown on the example.
Attention:
In scale with LED display designations of mass units: lb, kg, oz, ozt, ct are not displayed. Units are pointed by diode light.
0
kg
kg
g
Page 28
28 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
18.4 Serial port parameters setting (SErIAL)
The function allows setting independently communication parameters of both of serial ports Port-1 and Port-2 (executed in RS232C, RS485, USB or LAN standard):
- transfer protocol (Prot):
LonG – cooperation with printer or computer, EPL – cooperation with label printer in normal mode (activates LAbEL function), EPL_A – cooperation with label printer in automatic mode (activates LAbEL function), EPL_d – cooperation with special label printers, Pen-01 – cooperation with PEN-01,
- baud rate (bAud): (4800, 9600, ….115 200bps),
- number of bits in single char. (bitS): 7, 8,
- parity control (PArItY):
nonE – no control
Odd –nonparity
Even – parity control,
- scale number in network (nr):
(if the scale doesn’t work in network the number
must be 0),
- transmission through serial interface (SendInG) :
StAb – transmission after key is used and
result is stable, noStAb transmission after key is pressed
without need of stabilisation, Auto - automatic transmission after load is put on
and result is stable (Auto),
Cont - continuous transmission, about 10 results
per second (Cont.),
Remove – transmission after removing load. Default parameter values: Long, 9600 bps, 8 bits, none, StAb,
- SCAnn – cooperation with MJ-4209 barcode
readers: ON, OFF.
In order to set needed parameters choose SErIAL function, select appropriate parameter and press T key when required option or parameter value is displayed.
In scales with an additional serial port appear Port-1 and Port-2, for the independent setting of both ports.
0
kg
Page 29
USER MANUAL 29
_____________________________________________________________________________________________________________________
18.5 Printout configuration (PrInt)
Function is used for printing additional information stored in scale memory, weighed product identification data and scale operator id. That information is inscribed using scale keys or scanner. After entering selected port (scale can have two ports) user may activate printout positions:
- HEAdEr – header: name, model and scale number,
- USEr Id – scale user identification number,
- USEr nA – user name,
- Prn no – successive printout number (choose this
option to zero counter),
- Prod Id – product number,
- Prod bA – product barcode (inscribed or scanned),
- Prod nA – product name,
- Count – counting result (PCS function),
- APW – unitary mass (PCS function),
- netto – net mass
- tArE – current tare value,
- GroSS – gross mass,
- totAL – total mass (totAL function)
In printout configuration user can set if measurement (printout) number is saved after turning off the scale or not. Enter option Print and choose Prn no. Following options will appear:
- rESEt – resetting (zeroing) measurement number
counter,
- SAVE – activate saving measurement number after
the scale is turned off.
Attention:
If Prod Id or USEr Id is chosen, it is possible to inscribe quickly their new values (with omission of main menu). In order to do that hold (about 3 seconds) MENU key and release it when Prod Id or USEr Id indicates. Inscribe new value using keys:
0
- increasing digit,
- decimal point,
T
- next digit,
MENU - end.
While inscribing Prod id user can use barcode reader connected to RS232C interface.
If the scale is equipped with two serial joints Print function is set independently for both interfaces.
0
kg
Page 30
30 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Sample printout during normal weighing (all printout positions deactivated):
Sample printout during normal weighing with clock option (all printout positions deactivated):
Sample printout during normal weighing ( some printout positions activated):
20.07 kg
20.04 kg
20.04 kg
20.07 kg 2012-11-08 10:01
20.04 kg 2012-11-08 10:01
20.04 kg 2012-11-08 10:01
BA30 MAX: 30kg e=d=0.01kg S/N :
ID OPER. : 000001 DATE : 2012-11-08 TIME : 12:26 NO : 3 ID PROD. : 01 COUNT : 0 PCS APW : 0.000 g NET : 3.08 kg TARE : 0.00 kg GROSS : 3.08 kg TOTAL : 0.00 kg
Page 31
USER MANUAL 31
_____________________________________________________________________________________________________________________
18.6 Setting backlight function (b_LIGHt)
The function is used for choosing the work mode of scale display backlight:
- b_L OFF – switch backlight off,
- b_L on – switch backlight permanently on,
- b_L ECO – switch off after 30 seconds of inactivity
(no load changes and no key operation),
- b_L bAt – like above, but when powering from
accumulators only,
- out out without changes.
Switching backlight off causes decrease of energy consumption by the scale, what is important during powering from accumulators.
0
kg
Page 32
32 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
18.7 Analog out configuration (AnALoG)
This option enables to set-up analog out (4-20mA or 0-10V) working method used e.g. in PLC regulators:
- AnG rnG – inscribing Max value
- AnG CFG – working mode configuration (PLUS –
workmode for only positive values, MinuS – only for negative values, both – for both)
0
kg
Page 33
USER MANUAL 33
_____________________________________________________________________________________________________________________
18.8 Entering reference zero value (ZErO)
Note: This function is enabled in non-legalized scales only.
ZEr0 function allows entering new value of reference zero (value referred to empty pan) without need of contacting with authorised service centre.
Press MENU key. When ZErO is displayed press T key. On the display a sign ZEr Cod will show up momentary and the a dash on last digit position.
To enter code ( in new scale: 1234) use keys:
0
- increasing digit,
T
- next digit,
MENU – end of inscribing. The following options appear successively on display:
ZEr Cod – enter new secure code value, ZEr SEt – enter new zero value
Using T key, choose ZEr SEt. Direct result from A/C converter will appear on scale display. When the pan is empty press 0 key. Wait for finishing zeroing process.
In order to change access code use ZEr Cod option (as mentioned earlier).
0
kg
Page 34
34 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19. Special functions description
All scales besides basic metrological functions: weighing and taring, have a set of special functions. Depending on meter type functions set differs. Below a list of functions available in standard ME-01 type meters:
Products data base (Prod), Users data base (USEr), pieces counting function (PCS), change of mass unit (UnIt), percentage weighing function (PErC), selecting label number function (LAbEL), weighing large animals function (LOC), entering tare function (tArE), maximum value indication function (UP) force measuring function (nEWton) statistical calculations (StAt) paperweight calculation function (PAPEr)
and functions that require additional equipment to be completely functional:
option with accumulator supply:
- Setting accumulators charging (bAttErY)
- Automatic switching off scale function (AutoOFF)
options with the clock:
- setting current date and time function (dAtE)
- total weight function (totAL)
options with the transoptors connectors (WY ):
- checkweighing function (thr)
option with radio connection:
- function of choosing communication channel (rF Chn)
LabEL function is available in scales with EPL or EPL-A transmission protocol activates (go to SetuP/SErIAL).
In scales with LED display special functions don’t have additional marks on display and names of some
functions are shortened.
Page 35
USER MANUAL 35
_____________________________________________________________________________________________________________________
19.1 Products and users database (Prod i USEr)
Scale is equipped with products and users database with capacity up to 400 products and 100 users. Database consists of:
- M nr – memory number where data is saved,
- Prod Id - product identification number,
- Prod bA – product barcode,
- Prod nA – product name,
- USEr Id – user identification number,
- USEr nA – user name,
- APW - unitary weight (used when pieces counting),
- PtArE - inscribing permament tare to the product,
- thr Lo - threshold value (low),
- thr Hi - threshold value (upper),
- LAbEL – corresponding label number.
Database can be built in Excel datasheet form, where each product has one row and each column have product data. This way created database, saved in *.csv extension with semicolons can be send to scale using Scale Database software and scale’s serial interface. Scale Database is available on our webpage www.axis.pl/en. Database and possibility to cooperate with external devices: printer, label printer, barcode reader and computer enables to built product identification and product archiving systems.
Product barcode readout (during scale working) initiates searching through database and in case of finding
proper record, recalls product data (Found communicate). Barcode reader enables also to insert numerical
data conveniently (standard ME-01 meter doesn’t have numerical keys). Using alphanumeric code (for example 128 code) it can be also used to insert names of products and users.
Producent
Masa netto: xxxxxxx Data: xx-xx-xxxx
Produkt:
Page 36
36 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Inscribing data to base
Prod and USEr options enables adding and deleting product and user data. For products database available options are:
- Prod Id – searching for product in database by
inscribing (or scanning) id number or barcode,
- ProdCLr (shows up if product was selected earlier) –
turns off actual product selection,
- EdIt – product edition from database,
- Add – add product to database,
- dEL OnE – deleting single element from database,
- dEL ALL – deleting all elements from database,
- dAtAb – changing working mode with database (default
mode Stb):
Stb – searching products in database and working
with products outside the base; if product is found then Found communicate appears and all product data is recalled; if there is no product in database then no communicate appears, the scale stores id/barcode number temporarily in memory and enables to send it to the port (to printer/computer) together with actual weighing result.
LIMIt – searching through products from database;
if product is found then Found communicate appears and all product data is recalled; if there is no product in database then not Found communicate appears.
- Prn_P – sending all products database to port.
To inscribe data use EdIt option and keys: 0 - increasing digit, T - next dixit,
MENU – end of inscribing.
Barcode reader (connected to RS232C interface) can also be used to inscribe data and this way it is faster and more effective. Each database product has following data:
- M Id – memory cel number in products database,
- Prod Id – product identification number,
- Prod bA – product barcode,
- Prod nA – product name (inscribed from PC or barcode
reader),
- APW – product unitary weight (optional),
- PtArE – product package weight (optional),
- thr LO – lower threshold (MIN value),
- thr HI – upper threshold (MAX value).
Saving inscribed product data is done by using SAVE option.
0
kg
Page 37
USER MANUAL 37
_____________________________________________________________________________________________________________________
Users database is edited by similar function named USEr and consists of several options:
- USEr Id – user identification number,
- USErCLr (shows up if user was selected earlier) – turns off actual user selection,
- USEr nA – user name (inscribed from PC or scanner),
- Prn_U – sending users database to port.
Saving data is also done by SAVE option.
Recalling from database
The fastest way to recall product from database is to readout his barcode number (Prod bA) by using barcode reader (option). It can be done in any moment. After readout of proper barcode scale indicates one of communicates:
- SCAn – barcode from outside the base accepted
(Std mode),
- not Found – barcode from outside the base not
accepted and no product is selected (LIMIt mode),
- Found – product barcode found in database and
data recalled.
Attention: If the scale doesn’t indicate any
communicate, check barcode reader connections, port configuration and transmission protocol
(SErIAL function).
Other fast way is to press and hold MENU key (about 3s). Prod Id communicate will appear. Relase the key and inscribe identification number. If the number is already saved in base Found communicate appears and all the product data is recalled. To edit data choose EdIt option and use following keys:
0 - increasing digit, T - next dixit,
MENU – end of inscribing. Product recalling is also possible by using Prod and Prod Id options (previous site). If You hold MENU key for a longer time (about 6s) ProdCLr communicate will appear and actual product selection will be turned off.
0
kg
3 sek.
0
kg
Page 38
38 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Weighing results and data transmission from scale to computer or to printer
To fully use database capabilities other options must also be used: SeriaL, LabeL (for label printer) and Print. Serial option enables to select proper transmission protocol for each port. Thereby label printer can work
independently. Recalling product is equivalent with choosing corresponding label number. If database is not used, proper label can be choosed using LabeL option.
To each weighing results transmission a set of product and user identification data is added. The set is activated in Print option.
Available data from products and users base (Print / SEtuP option):
- USEr Id – user identification number,
- USEr nA – user name (inscribed from PC or scanner).
- Prod Id – product identification number,
- Prod bA – product barcode (inscribed or scan),
- Prod nA – product name (inscribed from PC or scan),
- Label – label number for proper product,
- APW – unitary mass (PCS function),
- tArE – tare,
- totAL – total mass (totAL function).
Page 39
USER MANUAL 39
_____________________________________________________________________________________________________________________
19.2 Pieces counting function (PCS)
This function enables to count identical pieces, e.g. turnbuckles or buttons.
A measurement is performed in two phases:
- first phase - single piece weight calculation on
the basis of defined pieces amount (5, 10, 20, 50, 100, 200 or 500 pieces),
- second phase – pieces counting.
First phase options:
- PCS . . recalling of a value inserted earlier (this quantity must be inscribed earlier),
-PCS SEt – set any amount of pieces in a sample,
-PCS APW – set unitary mass directly,
-PCS rS – inserting number of details in a sample and receiving of their mass from other scale connected by RS-232C.
It is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units.
To leave function press MENU key and then using T key chose PCS and PCS oFF.
Note:
1. APW too LOW communicate signalises that
a sample was not put on the pan or if single piece weight is less than one-tenth readout plot (counting is not possible).
2. APW LOW communicate signalizes that
single piece weight is more than one-tenth but less than one readout plot. (counting possible but with bigger errors, result blinks).
3. In scales equipped with LED display pcs sign
is replaced with “ ■ ”.
0
0
g
g
5
+10
0
0
Page 40
40 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.3 Percentage weighing function (PErC)
This function allows displaying weighing result in percents.
A measurement is performed in two phases:
- first phase – weighing a reference sample (100%),
- second phase – measuring specific sample as a percentage of the reference sample.
Weighing result is displayed in different format, depending on the weight value of reference sample.
The function has the following options:
- PEr oFF – disable the function,
- PEr on– set current scale indication as 100%
and activate percentage weighing,
-out- exit without changing settings.
Note:
1. PEr Err message informs that reference 100% mass is less than 0,5*Min or was not defined.
2. In scales with LCD display sign "
" is replaced with %.
100%
-5%
0
g
0
0
Page 41
USER MANUAL 41
_____________________________________________________________________________________________________________________
19.4 Label choosing function (LAbEL)
This function is used in scale with EPL (SErIAL function) data protocol. This protocol enables label printout with actual scale indication and chosen data from PrInt special function (variable data), for example date and time. Other data, for example company address, product name, barcode can appear on label as a constant text. Label patterns with number (4 digit) used by user should be saved in scale memory according to printer manual. Label pattern choice is made by inscribing label number using LAbEL function.
Press MENU button. When LAbEL is displayed press T key. Actual label number will show. To enter new label number press T key, to exit function without number change press MENU.
To inscribe label number use keys:
0
- digit increase,
T
- next digit,
MENU end.
After entering label number, putting load and pressing key will cause sending data to
label printer.
Data format sent to label printer (label nr 1,
language EPL-2):
US (55 53 0D 0A) FR"0001" (46 52 22 30 30 30 31 22 0D 0A) ? (3F 0D 0A) 00:00 (30 30 3A 30 30 0D 0A)
2000.00.00 (32 30 30 30 2E 30 30 2E 30 30 0D 0A) 10 g (20 20 20 20 20 31 30 20 20 67 0D 0A) P1 (50 31 0D 0A)
Page 42
42 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.5 Weighing animals function (LOC)
The function allows weighing animal moving on the scale.
Press MENU key.
When LOC function is displayed press T key. The following options appear on display successively:
- LOC oFF – leave the function,
- LOC on – automatic weighing after loading the
scale,
- LOC Prn – the measurement initiated manually
by pressing key.
When LOC on is displayed press T key.
Tare the scale using T key if necessary and place the animal on the pan.
Wait until the weighing result is averaged – scale display blinks. Then scale will show stable (averaged) result and will send it through serial port. The result remains on display for about 30 second.
Important notes:
1. The loads lower than Min value are not averaged.
2. In case when putting animal on scale takes more than 5s it is suggested to choose LOC PRN
option (measurement started manually by pressing key ).
0
0
g
g
0
g
0
g
Page 43
USER MANUAL 43
_____________________________________________________________________________________________________________________
19.6 Maximum value indication function (UP)
This function allows holding maximum (or minimum) value that is indicating at the moment.
Before measurement scale should be tared. Function has following options:
-UP oFF – function off,
-HIGH – holding maximum value,
-LOW – holding minimum value. Pressing T key will cause result zeroing.
Note:
Autozeroing function and the stabilisation indicator are deactivated when UP function is running.
0
0
g
kg
1kg
Page 44
44 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.7 Force measuring function (nEWton)
Function activation causes displaying result in force units (N).
Press MENU key.
Using T key choose NEWto function. Function has several options:
- nEW oFF – function off,
- nEW on – measurement in Newtons,
- ArM – torque measurement (arm length should be inscribed in meters using T, 0 and MENU keys).
Attention:
Units convertion from mass (kg) to force (N) is made for acceleration of gravity (g=9,80665m/s2)
Note: 1N 0,1019kg
0
g
0
Page 45
USER MANUAL 45
_____________________________________________________________________________________________________________________
19.8 Total weight function (totAL)
The function allows calculating total weight for series of measurements, which can be greater than scale capacity. It allows calculating total weight as well as average value.
Press MENU key. When totAL is displayed press T key.
The following options will appear successively:
- tot Prn - report printout without clearing total register,
- tot oFF - clearing total register, report printout and leaving the function,
- tot - working with receipt printout after each measurement,
- tot - working without receipt printout,
- tot CFG – saving measurement mode (using
key: Manual, after taking off the load : auto).
Press T key when tot is displayed. Perform measurement series by pressing key
for storing results into total register.
In order to print and display results enter the function by choosing totAL and tot Prn option from menu.
The results are displayed in the following sequence:
- total weight (SUM ),
- number of registered measurements (n),
- average value (=), regarding that moving to display successive result
is performed after pressing key. Attention: In scales with LED display SUM sign is replaced by “”.
In order to go back to total weighing without
zeroing total register press key several times.
0
0
0
0
g
g
g
g
g
g
Page 46
46 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
To leave the function with clearing total register, select totAL function from menu and choose tot oFF option. Scale prints the communicate informing about clearing registers.
The form of standard receipt (measurement number and weight) after each measurement:
Print configuration option (chapter 17.5) enables to extend standard receipt form.
Report form (total weight, number of measurements, average weight):
Note:
When the scale doesn’t have an internal clock, Date and Time do not appear on printout. Maximum number of measurements is 99 999. Maximum total load 99 999 000d. The weighing unit of the total value from the register (Total) is the same as the weighing unit stated on the
keypad or is 1000 times greater, what is signalled by “o” indicator at the left of the display. If the registered value is too big to be displayed, “E” communicate appears on the display. If the number of series is too high and cannot be displayed, “Err1”communicate appears on the display
1 3 g
2 5 g 3 3 g
4 4 g
TOTAL = N = AVERAGE =
Page 47
USER MANUAL 47
_____________________________________________________________________________________________________________________
19.9 Checkweighing function (thr)
This function allows comparing weighing result with two programmed reference values: lower and upper threshold. Comparison result is signalled with indicators (MIN, OK, MAX) and sound signal generated when threshold values are exceeded. If comparison result is:
- smaller than zero threshold – no signal,
- smaller than lower threshold – the scale signals MIN (yellow colour),
- between threshold values - the scale signals OK (green colour, with the short sound signal),
- greater than upper threshold - the scale signals MAX (red colour, long sound signal).
The checkweighing results can be use to control:
- optical indicator (Indication mode),
- batching devices (Batching mode).
-
Standard scale is set for cooperation with optical indicator.
On outputs P1-P3 (Relays socket) short-circuit states appear as result of comparison scale indication with threshold values. On the chart below output states are shown during increasing load on the scale for both working modes:
Indication mode: Batching mode:
P3
P1
P2
P3
zero
zero
treshold
treshold
thr I
thr I
thr I
thr II
thr II
thr II
P1
P2
In Batching mode on P1 (thr I) and P2 (thr II) outputs short-circuit impulses appears for time of 0,5s. On P3 (zero) output short-circuit state appears when indication does not exceed threshold value signalling zero load.
Page 48
48 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Operation sequence:
Press MENU key and choose thr pressing T key. The following options are displayed successively:
- thr oFF – deactivate the function,
- thr on – activate the function,
- thr Prn – check last threshold values (press key several times),
- thr CFG – choose Relays socket mode:
0 – exit to weighing 1 – Batching mode 2 – Indication mode.
Choose thr-on option using T key. The following options for entering thresholds are displayed:
- SEt-0 - go to weighing with signalling threshold
excess,
- SEt-1 - set lower threshold value,
- SEt-2 - set upper threshold value,
- SEt-3 - set zero signalisation threshold.
Using T key select SEt-1 option. Set lower threshold value using the following keys:
0 - digit increase,
- decimal point, T - move to next digit, MENU - finish.
Then select SEt-2 option and enter upper threshold value.
Choosing SEt-0 option will cause starting work with signalisation of exceeding thresholds and zero.
To change Relays socket mode use thr CFG option. Default option is Indication.
To leave the function, press MENU key and then choose thr and thr oFF options.
0
g
g
Page 49
USER MANUAL 49
_____________________________________________________________________________________________________________________
Relays connection diagram:
Relays output is the open collector transoptor output with load capacity 25mA / 24V. Transmitter inputs
must be protected with diodes, e.g. 1N4148. It is advised to use MS3K/P electronic board (sold separately), consisting of RM96P transmitters, with DC24V input voltage and AC250V, 3A output.
Important notes:
1. After switching the scale on, both thresholds are set to maximum values.
2. When setting upper threshold value, pay attention that its value is not below lower threshold value.
3. Setting lower and upper threshold value is possible after sending appropriate orders from computer,
what is described in scale user manual.
Imax < 25mA
Page 50
50 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.10 Setting date and time function (dAtE)
The function allows setting current date and time of scale internal clock and mode of its use. The function has the following options:
- dAt oFF – deactivate date and time during printout of current weighing result,
- dAt on – activate date and time during printout of current indication ( key),
- dAt SEt - change current date and time,
- dAt PIn – data and time secure password (to prevent from changing date and time by unauthorized personel),
- dAt For – data printout in different format.
The example at the left presents how to set current date and time using dAt SEt option.
After setting proper date and time activate it with dAt on option.
Date and time format:
PL: rrrr-mm-dd gg:mm
UE: dd-mm-rrrr gg:mm
USA: mm-dd-rrrr gg:mm AM/PM
(gg – hours, mm – minutes, AM – before noon, PM
after noon, mm - month, dd - day, rrrr - year).
Attention: Inscribing non-zero PIN value causes showing PIN sign during next date and time changing and inscribing 4 digit code is necessary. (using keys 0, T and MENU).
0
kg
Page 51
USER MANUAL 51
_____________________________________________________________________________________________________________________
19.11 Radio communication channel choice function (rF CHn)
Function enables choosing radio communication channel between the scale and a pilot. In scale and in pilot the same radio channels must be chosen. Function should be used when communication is disturbed by other devices that use the same communication channel.
Press MENU key and choose rF CHn by pressing
T
key.
The following communicates will appear on display:Na wyświetlaczu pojawią się kolejno:
- CHn 01 – channel 1,
- CHn 02- – channel 2 ...
- CHn 16 - – channel 16
- out – out without changing channel.
In default setting channel 01 is on.
0
0
kg
kg
Page 52
52 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.12 Charging accumulators function (bAttErY)- option
bAttErY function allows switching on or off charging accumulators during work with feeder and checking their power level. The function has the following options:
- bAt OFF – charging off (option required if ordinary
batteries are used !!!),
- bAt on – charging on, accumulators are being
charged even after switching scale off using I/ key,
- bAt VoL – reading power level of accumulators in
% (go back to mass indication pressing MENU key),
- out – leave without changes
An attempt of charging ordinary batteries can cause serious damage of the scale.
0
kg
Page 53
USER MANUAL 53
_____________________________________________________________________________________________________________________
19.13 Automatic switching off the scale function (AutoOFF)
The function is helpful in scales supplied from accumulators. The function causes scale to switch off automatically.
Press MENU key.
When AutoOFF is displayed press T key.
The following options appear successively on display:
- AOF oFF – deactivate function,
- AOF on – activate function- scale turns off after 5
minutes of not making any actions,
- AOF bAt – as above but only when supplied from accumulators.
- Out – out without changes.
0
kg
5min.
Page 54
54 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
19.14 Statistical calculations function (StAt)
This function evaluates from series of measurements (max 1000) statistical parameters of weighting process. Adding successively measurements to register is automatic and it occur after the scale is loaded and its indications stabilize. After each loading printout is made with: number of measurements, result, date and time (if clock is installed and the function is activated).
For the obtained measurements series the scale evaluates:
- n -number of samples
- sum x -sum of all samples
n
xxsum_
- x -average value (sum x)/n
- min -minimal value from n samples
- max -maximal value from n samples
- max-min -maximal value minus minima value
- S -standard deviation
2
)(
)1(
1
n
n
xx
n
S
- srel -variance factor
x
S
srel
Statistical calculations results can be printed.
Page 55
USER MANUAL 55
_____________________________________________________________________________________________________________________
Order of operations:
Press MENU key. When StAt is displayed press T key.
The following options are displayed:
- StA Prn – monitoring and printout of statistical
data,
- StA oFF – deactivate function,
- StA
– activate function, work with printout of
chosen weighting results,
- StA - – activate function, work without printout,
- StA n – maximal samples value,
- Sta nM – inscribing nominal value for statistics,
- Sta tOL – inscribing tolerance in %,
- Sta tAr – automatic tare on/off
- StA CFG – function configuration:
-Auto Automatic work (samples are confirmed after loading the scale and indication stabilization.),
-ManuAL manual work (confirmation is made by pressing key).
- out – exit from function.
Remember first to inscribe nominal weight value and tolerance (mentioned above).
After that, push T key when StA o is displayed.
Put on successive objects on the pan (remove after indication stabilization) in order to add them to measurements register.
In order to obtain printed statistical results from measurements series press MENU key and T key when StAt is displayed and later StA Prn. After printout two options are enabled:
- rESET – erasing results,
- Contin – continuation.
0
0
g
g
g
Page 56
56 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Pressing key printouts estimated values and histogram :
Nominal - nominal value,
Tolerance - accepted value in percentage.
N - number of sample
IN TOL. – number of samples in toleranc
-TOL – amount of measurements
under allowable lower value
+TOL – amount of measurements above
allowable upper value
TOTAL - sum of weights of all n samples
AVERAGE – average weight as (Total)/n
MIN – minimum weight in n samples
MAX– maximum weight in n samples
ST. DEV. – standard deviation
ST. DEV.% – standard deviation percentage
To finish work with this function and
zeroing result register press MENU
key and then when StAt. and Sta oFF is
displayed press T button.
Statistics function cooperation with computer and
Printer. Scale can be equipped with two serial ports
marked as RS232C-I (computer) and RS232C-II
(printer). After each data printout by printer identical
set of data is sent to computer. After sending by
computer initialization signal S A CR LF
(53h 49h 0Dh 0Ah) the scale sends to computer
statistic data enclosed in histogram.
Page 57
USER MANUAL 57
_____________________________________________________________________________________________________________________
19.15 Paperweight calculation (PAP)
This function enables to calculate paperweight of 1m2 of paper basing on samples of known area. For quick access, the function is accessible directly by pressing MENU key.
The balance must be tared just before the measurement. Place the specific sample quantity of the same area (possible values: 1, 2, 5, 10, 20, 50, 100).
Press MENU key to access Function Menu. To enter the function press T key when PAPEr
is displayed. Following options show on the display:
- PAP oFF – turn off the function,
- PAP on – turn on,
- PAP n – inscribing number of paper pieces on pan,
- PAP ArE – inscribing surface of single piece (in m2)
Press T key when PAP n and PAP ArE is displayed. Enter number of samples using: 0 -increasing digit,
T
- next digit,
MENU – end of inscribing.
Press T key when PAP ArE is displayed. Enter area of a single sample (as above). The result of paperweight measurement is finished with „o” mark pointing g/m2 unit. In order to finish work with function press MENU
and then using T key choose PAPEr and PAP oFF
Note:
“PAP Err” communicate marks that wrong values were inscribed in PAP n or PAP ArE..
0
g
5
Page 58
58 BD/T USER MANUAL
_____________________________________________________________________________________________________________________
Maintenance and repairs of small defects
1. The scale should be kept clean.
2. Take care that no dirt gets between the platform and the scale base. If found any,
remove the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by short-lasting power supply decay, unplug the
scale from the mains and then plug it again after few seconds.
4. If the scale is switched on with empty pan and “Err-b” communicate appears, the load
cell has been mechanically damaged.
5. It is forbidden to make any repairs by unauthorised persons.
6. To repair the scale, please contact our nearest service. Balances can be sent for repair
as messenger delivery only in original package, if not, there is a risk of damaging the balance and loosing guarantee.
Error communicates:
Communicate
Possible cause
Remedy
C-1 ... 6
(over 1 min.)
selftests failed
if displayed more than 1 minute, contact an authorised service
the scale was switched on with loaded pan
remove a load from the pan
mechanical damage of the load cell
contact an authorised service
L
pan missing
put the pan on
mechanical damage
contact an authorised service
H
overloading
remove the load from the pan
mechanical damage
contact an authorised service
indicator does
not appear
unstable ground vibrations air flows
place the scale on a stable ground not affected by mechanical vibrations and airflows
scale is damaged
contact an authorised service
- - - - - -
taring in progress
as above
- -
taring could not be finished (the load is too small or B\G key was used)
zero the scale or press B\G key again
- -
the load is too big to be zeroed
tare the scale (T)
Page 59
USER MANUAL 59
_____________________________________________________________________________________________________________________
Declaration of Conformity
We:
AXIS Spółka z o.o. 80-125 Gdańsk, ul. Kartuska 375B
confirm with all responsibility that scales:
BD2P
marked with CE mark comply with the following:
1. Directive 2004/108/EWG (electromagnetic compatibility) and harmonized norms:
- EN 61000-4-3+A1:2008+A2:2011
- EN 61000-6-3:2008+A1:2011
2. Directive 2006/95/WE (low voltage) and harmonized norm:
- EN 61010-1:2004
Moreover scales with the following markings on the name plate: 16
- the number of the Notified Body responsible for EC verification 1443
- two-digit number of the year of EC verification
- a green metrology sticker with “M” mark
- a protective seal affixed by the Notified Body
comply with the requirements on the Type-Approval Certificate WE No. T7950R0 and are verified to comply with:
- EN 45501:2015
Additional information:
- Conformity evaluation for the Directive 2006/95/WE and 2004/108/WE was carried out
by Research Laboratory of Electrotechnology Institute Division Gdańsk, accredited by PCA,
- Conformity T7950R0 evaluation was carried out by NMI Certin B.V. (Notified Body No. 0122).
Per pro Director of AXIS Sp. z o.o.: Production Manager Jan Kończak ________________ Date: 30-10-2015