Atec V7x Series User Manual

V7X Series - Hipot Testers
With color touch LCD & high speed DSP technology, the compact & rugged V7X sets
the standard for price/performance ratio. Made in the USA to meet tough
UL, CSA, TUV and IEC Hipot requirements - the V7X provides unbeatable speed,
accuracy, user safety and reliability.
Choose from six low cost models offering AC & DC Hipot to 5KV, leakage current
measurement to 100 nano-amps, Insulation Resistance to 450GΩ and Ground Bond to 30
amps. Combine all that with USB, RS232 and Digital I/O interfaces, plus a 2 year warranty.
The V7X is simply unbeatable
Vitrek V7X Series Performance Specifications
AC Hipot
Output Voltage 10 to 5000V 50 or 60 Hz
Accuracy : 1.4% of setting +0.1% Full Scale, No load to full load Resolution : 1V at all levels Max load current : 20mArms Leakage Current Min/Max Limits during dwell of none, or up to 20mArms
Accuracy : 1.5% of reading +5uA Resolution : 1uA
DC Hipot
Output Voltage 20 to 5000V test voltage
Accuracy : 1.4% of setting+0.1% Full Scale, No load to full load Resolution : 1V at all levels Max load current : 10mA Leakage Current Accuracy : 1.5%+1uA Resolution : 0.1uA
IR - Insulation Resistance
Test Voltage 20 to 5000VDC
Accuracy : 5%+5V No load to full load
Resolution : 1V at all levels Max Charge Current: 5mA automatic Max Capacitive Load: 2uF Resistance Max IR: 450Gigohm (90M per volt) Min IR: 150K Accuracy : 2% (rdg <5% of max IR), 5% (< 15% of max IR), 10% (< 30% of max IR), 20% (above
30% of max IR)
Max Resolution : 0.1% of value
Min/Max Limits : Defined for each step, max may be set to none Test Completion Stop on Time: Determination on final reading Stop on Pass: Test terminates with PASS for any reading within limits Stop on Fail: Test terminates with FAIL for any reading outside limits Stop on Steady: Test terminates with PASS for any reading within limits which is
Low Resistance
Resistance Range: 0ohm to 60Kohm
Accuracy : 1.4 %+0.02 (<10), 3%+1ohm (<1K), 10 % above 1K
Resolution : Down to 0.001 Min/Max Limits : Defined for each step, each may be set to none Test Method 2 terminal measurement , 10.5mA max, 4.15V max Resistance Offset Test leads/fixture measurement offset may be universally applied
Ground Bond
Test Current 1 to 30Arms (42A pk), 50/60Hz
Loading...
+ 2 hidden pages