AC & DC Hipot plus Insulation Resistance and Ground Bond
Everything You Need in a 4-in-1 Safety Tester
for an Exceptionally Low Price
Only Vitrek – world leaders in
electrical safety test automation – could build a feature rich, full function 4-in-1 electrical safety tester with performance this high and a price this low. The new Vitrek V4 combines AC Hipot to 5KV, DC Hipot to 6KV, insulation resistance to 10 gig-ohms and high current ground bond measurement up to 40 amps – all in one compact, automated instrument.
The performance of the V4 is uncompromising.
The V4 delivers fast, consistent, reliable results – test after test. And it does all this from a single test connection with the touch of a button or with available QuickTest software – the stroke of a keyboard. The V4’s crystal clear LCD user interface is remarkably ea sy to use, yet provides an abundance of
information. Test results for a sequence of up to 16 tests can be viewed on a single screen. The V4’s Multi-mode IR™ capability, lets you choose the operating format for insulation resistance testing and with outputs available from 50 to 1000V – you’re sure to find the test voltage you need.
The V4 suits your needs.
Whether you need a workhorse for everyday production line testing or an analytical unit for product safety engineering you’ll find that the multi-function V4 is loaded with all the features you need to get the job done right - powerful features, including programmable ramp and dwell times, automatic calibration, SFI™ safety fault interrupt, 160 position user memory, selectable arc detection, scanner control port, bi-directional RS-232 and available GPIB interface. When you put it all together there is no comparison. The V 4 is the new undisputed champion of electrical safety testers.
WORLD LEADERS IN COMPLIANCE AUTOMATION
y
y
A
A
naall
n
yzzeerr
y
Vitrek V4 Advantages
Tests to UL, CSA, IEC & TUV specifications
4 EST functions in 1 compact, feature-rich instrument
Up to 5KVAC 40mA
Up to 6KVDC 10mA
Multi-mode IR 50-1000V up
to 10,000MO
Ground Bond up to 40Amps
Leakage current
measurement down to 10µA
Fully programmable RS-232 & HV scanner control port, optional GPIB interface
Selectable arc detection
Safety Fault Interrupt (SFI)™
circuit
CE mark certified
10 position test sequence
memory
3 year extended warranty
SM
V4 Performance Specifications
AC Hipot
Voltage Range: 100 to 5000 VAC RMS Voltage Setting Resolution: 5V/step Voltage Regulation: 1% of reading + 5V Voltage Accuracy: 1% of reading + 10V Voltage Monitoring: At Output Terminals Output Current: 30mA to 40mA 180S max,
0 to 30mA continuous. Below 500V 10mA max
Min/Max Current Limits: 0.1 to 40mA Current Accuracy: 1% of reading + 50 µA
DC Hipot
Voltage Range: 100 to 6000 VDC Voltage Setting: Resolution: 5V/step Voltage Regulation: 1% of reading + 5V Voltage Accuracy: 1% of reading + 10V Voltage Monitoring: At Output Terminals Max Output Current: 10mA continuous, below
500VDC 2mA continuous
Min/Max Current Limits: 0.01 to 10mA Current Accuracy: 1% of reading + 50µA
Insulation Resistance
DC Output Voltage : 50, 100, 500 or 1000V Resistance Range: 1 to 9999M Resistance Accuracy: (% of reading)
50, 100 VDC 1-50MO 5%
51-200MO 10%
201-1999MO 20% 500, 1000 VDC
Multi-Mode IR™: Select 1 of 3 modes - test passes on good reading, test fails on bad reading, pass/fail determined at end of test time.
Ground Bond / Continuity
Test Current: 3A to 40 amps RMS Terminal Configuration: 4 wire Kelvin Resistance Range: 0 to 600m Resistance Resolution: 100 µ Resistance Accuracy: 1% reading ±3m Compliance Voltage: 8 volts at terminals Current Resolution: 0.01A Voltage Resolution: 0.01V
General
Test Database: 10 test sequences, up to 16 steps each with up to 6 steps per test sequence Front Panel Lockout: Prevents databas e edit while in lockout mode
Arc Detect: 10 level trip select Remote Interfaces: Fully interactive RS232, -
PLC interface, scanner control port – all standard talk/listen GPIB interface - option GP -V4
Timer Accuracy: 0.25% ± 50mS Power Source: AC100V, 120V, 220V,
230V (±10% 50/60Hz), 350W max
1-500MO 5%
501-2000MO 10%
2001-9999MO 20%
Operating Environment: Indoor use, altitude up to 6500ft (2000m). installation category II. Pollution Degree 2
Operating Temperature: 0°C to 40°C, 80% RH Storage Temperature: -10°C to 70°C, 70% RH Dimension: 17.6”L x 13”W x 5.9”H,
446(L) x 330(W) x 149(H)mm
Weight: 33lbs (15kg) net, 40 lbs (18 kg) ship Accessories: NIST cal cert, TL-2050 & K-2 test
leads, instruction manual and power cord Accuracy: Stated specifications apply for 1
year at 23 °C ± 5 °C Calibration: ANSI/NCSL Z540 Certificate of calibration with data provided, traceable to NIST Compliance: CE mark certified to EN61010 and EMC directives EN50081-1 & EN50082-2 Warranty: 3-Year extended warranty with annual factory calibration, 1 year parts and labor standard
QuickTestSoftware
Option QT -V4: Bundled test automation software and PC serial interface cable. Easy to load graphical test sequence development and operating environment. Provides time/date, operator, test parameters, test results and readings. Results may be viewed, printed out or exported to a database program.
Ordering Information
V4 Four -in-1 Electrical Safety Tester GP-V4 GPIB Interface Option QT-V4 Bundled Quicktest Software with Serial
Cable
TL-115-4 Hipot/GB Outlet Te st Adaptor TL-1050 High Volt Probe Lead Set TL-2050 High Volt Alligator Lead Set TL-IEC4 IEC Socket Lead Set K-2 High Current, 4-wire,2 Clip Kelvin Lead Set K-4 High Current, 4-wire, 4 Clip Kelvin Lead Set RPO-4 Rear Panel Output Terminals RM-4 Rack Mount Kit
Represented By:
For more information call: 858 689 2755 or Email Info@vitrek.com
Specifications subject to change without notice.
Vitrek Corporation 9880A Via Pasar San Diego, CA 92126
Loading...