DATA SHEET > UCS 500N5V > 20130322
UCS 500N5V
COMBINATION WAVE (SURGE) GENERATOR UP TO 5KV AS PER
IEC/EN 61000-4-5
FOR TESTS ACCORDING TO ...
> EN 300329
> EN 300340
> EN 300342-1
> EN 300386 V1.3.2
> EN 301489-1
> EN 301489-17
> EN 301489-24
> EN 301489-7
> EN 50121
> EN 55024
> EN 61000-4-5
> EN 61000-4-9
> EN 61000-6-1
> EN 61000-6-2
> IEC 61000-4-5
> IEC 61000-4-9
> IEC 61326
> IEC 61850-3
> ITU-T K.20
> ITU-T K.21
> ITU-T K.41
> ITU-T K.45
UCS 500N5V - THE STAND-ALONE COMBINATION WAVE (SURGE) GENERATOR
The UCS 500N5V is a new dedicated Surge generator based on the successful UCS 500N series of ultra-compact simulators.
Offering all the benefits of the well-known UCS 500N5 it is designed as a Surge generator only. This makes it the full
performing Surge tester as per IEC/EN 61000-4-5 with the best benefit-to-cost ratio yet upgradeable to include other test
abilities such as EFT/Burst and PowerFail.
Apart from its built-in single phase test capability it is extendable for testing three-phase DUTs by means of automatically
controlled external coupling networks up to 100A. EM TEST supplies a large range of accessories for various applications.
HIGHLIGHTS
> Fully compliant to IEC 61000-4-5
APPLICATION AREAS
INDUSTRY
COMPONENTS
> Surge voltage up to 5kV
> Surge current up to 2.5kA
MEDICAL
BROADCAST
> Built-in single phase coupler 300V/16A
RESIDENTIAL
RENEWABLE ENERGY
> Programmable current limiters
> Monitor outputs for peak voltage and peak
TELECOM
current
www.emtest.com © EM TEST > PAGE 1/4
DATA SHEET > UCS 500N5V > 20130322
TECHNICAL DETAILS
COMBINATION WAVE / SURGE
GENERATOR
UCS 500N5V
As per EN/IEC 61000-4-5 and
EN 61000-6-1, -6-2
Voltage (o.c.) 160V - 5,000V ± 10%
Pulse front time 1.2us ± 30%
Time to half value 50us ± 20%
Current (s.c.) Max. 2,500A ± 10%
Pulse front time 8us ± 20%
Time to half value 20us ± 20%
Polarity Positive/negative/alternating
Event counter 1 - 30,000 or endless, selectable
TRIGGER CIRCUIT
Release of pulses Automatic, manual, external
Synchronization 0° - 360°, resolution 1°
COMBINATION WAVE / SURGE
GENERATOR
TEST ROUTINES
Quick Start One-line adjustable parameters,
easy-to-use
Standard Test
routines
User Test routines Change polarity after n pulses
Pulsed Magnetic
Field
As per IEC 61000-4-5, Levels 1 - 4
As per EN 61000-6-1, -6-2
Manual Standard Test routine
Change coupling after n pulses
Change voltage after n pulses
Change phase angle after n pulses
as per IEC 61000-4-9
Test levels 100, 300 and 1,000A/m
Test level steplessly adjustable
under Quick Start
Repetition rate max. 1Hz (1s - 999s)
OUTPUTS
Direct Via HV connectors for external
coupling networks (Zi = 2ohm with
optional adapter IMN 2)
Coupling mode Line to line
Line(s) to ground
DUT supply AC: 300V/16A; 50/60Hz
DC: 300V/16A
CRO trigger 5V trigger signal for oscilloscope
MEASUREMENTS
CRO Û-monitor 10Vp at 5,000V
CRO Î-monitor 10Vp at 2,500A
Peak voltage 5,000V in the LCD display
Peak current 2,500A in the LCD display
www.emtest.com © EM TEST > PAGE 2/4