Atec UCS500N User Manual

DATA SHEET > UCS 500N5 > 20131118
UCS 500N5
MULTIFUNCTIONAL TESTGENERATOR FOR TRANSIENTS (EFT/BURST, SURGE & POWER FAIL) UP TO 5.5KV
FOR TESTS ACCORDING TO ...
> EN 300329 > EN 300340 > EN 300342-1 > EN 300386 V1.3.2 > EN 301489-1 > EN 301489-17 > EN 301489-24 > EN 301489-7 > EN 50121 > EN 55024 > EN 61000-4-11 > EN 61000-4-29 > EN 61000-4-4 > EN 61000-4-5 > EN 61000-4-8 > EN 61000-4-9 > EN 61000-6-1 > EN 61000-6-2 > FCC 97-270 (part 68) > IEC 60255-22-5 > IEC 61000-4-11 > ...
UCS 500N5 - COMPACT TESTER FOR EFT/BURST, SURGE AND POWER FAIL
The UCS 500N5 ultra-compact simulator is the most versatile tester to cover transient and power fail requirements according to international standards (basic and generic standards) and product/product family standards. The UCS 500N5 is the most economic solution for tests during development as well as for full-compliant immunity tests and CE Marking for single phase DUT with the ability to be extended for testing three-phase DUTs by means of an automatically controlled external coupling network up to 100A. EM TEST supplies a large range of accessories for the various applications such as magnetic field tests.
HIGHLIGHTS
> Burst module (IEC/EN 61000-4-4) up to 5.5kV
APPLICATION AREAS
INDUSTRY
COMPONENTS
> Surge module (IEC/EN 61000-4-5) up to 5kV > PowerFail module (IEC/EN 61000-4-11)
MEDICAL
BROADCAST
> Magnetic field tests with optional accessoires
RESIDENTIAL
RENEWABLE ENERGY
> Built-in single phase coupler 300V/16A > Manual operation from front panel
TELECOM
> USB and GPIB-Bus for remote control
www.emtest.com © EM TEST > PAGE 1/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
BENEFITS
ALL IN ONE - ALL WHAT YOU NEED FOR YOUR TESTS
The UCS 500N5 includes everything necessary to conduct fully compliant tests. The power mains supply for the controls and for the DUT is separate to render it more flexible to use with different DUT supply voltages. The UCS 500N5 can be operated manually from the front panel or by remote via the built-in USB or GPIB interface. Fail inputs allow to control an ongoing test sequence based on the status of the DUT. Monitoring outputs (BNC) are offered for easy signal measurement and verification. Safety features such as interlock and warning lamp control are available. Pre-programmed Standard Test routines allow highest user convenience. Still the UCS 500N5 offers the Quick Start test routine where parameters can be changed on-line during the test to evaluate the susceptibility level of an individual DUT.
OPERATION
EASY TO OPERATE
Front panel menu and function keys enable the user to program his test routines quickly and accurately. The cursor allows fast control of all test parameters of the programmed routine, thus test procedures are simplified and confidence is generated that every step is carried out correctly.
SOFTWARE
IEC.CONTROL SOFTWARE FOR CONTROL AND DOCUMENTATION
Outstanding user convenience, clearly structured windows and operation features and the EM TEST standards library along with the flexibility to generate user specific test sequences very easily are the main features of iec.control software. The software is automatically configured according to the connected EM TEST generators. Extensive reporting capabilities help the user to create test reports that meet international requirements. iec.control is supported by Windows 2000, Windows XP, Windows Vista and Windows 7. Remote control is achieved either via USB or GPIB. iec.control supports a wide range of GPIB cards of National Instruments.
OTHER MODELS
UCS 500N SERIES - COMPACT TESTERS UP TO 7KV
The UCS 500N ultra-compact testers for EFT/burst, Surge and Power Fail are available in two different models; with voltage capability up to 5.5kV or up to 7kV.
www.emtest.com © EM TEST > PAGE 2/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
AUXILIARY DEVICES
CNI 503 - 3PHASE COUPLING/DECOUPLING NETWORKS FOR BURST AND SURGE
EM TEST offers a range of fully automatic 3-phase coupling/decoupling networks for burst and surge to extend the test capability for three-phase DUTs. The networks have a rated current of up to 100A.
MV 2616 - MOTORISED VARIAC FOR VOLTAGE VARIATION
A motorised variac is offered as an alternative to the tapped autotransformers for voltage dips/interruptions and voltage variation tests as per IEC 61000-4-11. The motorised variac can also be used for automated magnetic field tests.
V 4780 - TAPPED VOLTAGE TRANSFORMER FOR VOLTAGE DIPS AND INTERRUPTIONS
The V 4780 tapped autotransformer is designed to supply the required voltages as per IEC 61000-4-11 Ed.2:2004 to perform voltage dips and interruptions.
V 4780S2 - TAPPED VOLTAGE TRANSFORMER FOR VOLTAGE DIPS AND INTERRUPTIONS
The V 4780S2 tapped autotransformer is designed to supply the required voltages as per IEC 61000-4-11 Ed.2:2004 to perform voltage dips and interruptions. Compared to the manually operated V 4780 the V 4780S2 model offers automatic change of taps according to the selected voltage level.
CNV 504N/508N - SURGE COUPLING/DECOUPLING NETWORKS FOR SIGNAL/DATA LINES
CNV 504N/508N coupling/decoupling networks are available to perform surge tests on I/O lines, signal/data lines and telecom lines as per IEC 61000-4-5.
ACCESSORIES
MS 100N - MAGNETIC FIELD COIL FOR POWER-FREQUENCY AND PULSED MAGNETIC FIELDS
The MS 100N is a 1sqm magnetic field coil as specified in IEC/EN 61000-4-8 and IEC/EN 61000-4-9. Its design allows easy moving of the coil. The field coil is adjustable in height and allows for 360degr rotation. To generate power-frequency magnetic fields in the lower range the current transformer MC 2630 is used while high-field strength above 100A/m requires the MC 26100 current transformer.
HFK - CAPACITIVE COUPLING CLAMP
The HFK is a fully compliant capacitive coupling clamp as per specification of IEC 61000-4-4.
ITP - IMMUNITY TEST PROBES
ITP is a tool being used for development test. It consists of a variety of electrical field probes. The probes allow to locate weak points within a system or on a PCB. The burst pulse is used to generate the disturbance signal.
CA EFT KIT - VERIFICATION KIT FOR EFT/BURST PULSES
As per IEC 61000-4-4 Ed.2 the characteristic of the burst generator needs to be verified with two different loads, 50ohm and 1,000ohm. EM TEST offers a calibration kit consisting of the two loads and an adapter to verify the pulses at the DUT output.
CA HFK KIT - VERIFICATION KIT FOR CAPACITIVE COUPLING CLAMP
The IEC 61000-4-4 Ed 3.0 published 2012 recomends the calibration of the capacitive coupling clamp into a 50ohm coaxial load. The capacitive coupling clamp (HFK) is connected to the 50 ohm output of the EFT generator. A flexible insulated plate inside the HFK is connected to a coaxial 50 ohm load resistor for verificate the EFT / Burst wave of the capacitive coupling clamp.
www.emtest.com © EM TEST > PAGE 3/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
ELECTRICAL FAST TRANSIENTS
BURST MODULE, EFT/N5
As per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2
Test voltage 200V - 5,500V ± 10%;
100V - 2,750V ± 10% into 50ohm Pulse shape 5/50ns into 50ohm and 1,000ohm Rise time tr 5ns ± 30% into 50ohm;
5ns ± 30% into 1,000ohm Pulse width td 50ns ± 30% into 50ohm;
50ns -15/+100ns into 1,000ohm Source impedance 50ohm Polarity Positive/negative
TRIGGER CIRCUIT
Trigger of bursts Automatic, manual, external Synchronization 0° - 360°, resolution 1° (16 - 500Hz) Burst duration td = 0.10ms - 999ms
OPTIONS
HFK Capacitive coupling clamp as per
IEC 61000-4-4 KW50 100:1 divider, 50ohm KW1000 500:1 divider, 1,000ohm CA EFT kit Kit for burst pulse verification
consisting of KW50, KW1000 and
adapter for DUT port in a plastic case
for storage A6dB 6dB attenuator, 50ohm ITP Immunity test probes (electrical field
generation) ITP/H Immunity test probe (magnetic field
generation)
Repetition rate tr = 10ms - 9,999ms Spike frequency f = 0.1kHz - 1,000kHz Test duration T = 0:01min - 99:59min
T > 99:59min --> endless
OUTPUTS
Direct Via 50ohm coaxial connector Coupling mode L, N, PE; all combinations DUT supply AC: 300V/16A; 50/60Hz
DC: 300V/16A
CRO trigger 5V trigger signal for oscilloscope
TEST ROUTINES
Quick Start On-line adjustable parameters,
easy-to-use
Standard Test routines
User Test routines Synchronous burst release
As per IEC 61000-4-4, Levels 1 - 4 As per EN 61000-6-1, -6-2 Manual Standard Test routine
Random burst release Change voltage after T Frequency sweep within one burst Frequency sweep with constant number of pulses Frequency sweep with constant burst duration Change polarity after T
www.emtest.com © EM TEST > PAGE 4/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
COMBINATION WAVE / SURGE
SURGE MODULE, VCS/N5
As per EN/IEC 61000-4-5 and
EN 61000-6-1, -6-2 Voltage (o.c.) 160V - 5,000V ± 10% Pulse front time 1.2us ± 30% Pulse time to half
value Current (s.c.) Max. 2,500A ± 10% Pulse front time 8us ± 20% Pulse time to half
value Polarity Positive/negative/alternating Event counter 1 - 30,000 or endless, selectable
TRIGGER CIRCUIT
Release of pulses Automatic, manual, external Synchronization 0° - 360°, resolution 1° Repetition rate max. 1Hz (1s - 999s)
50us ± 20%
20us ± 20%
TEST ROUTINES
Quick Start One-line adjustable parameters,
easy-to-use
Standard Test routines
User Test routines Change polarity after n pulses
Pulsed Magnetic Field
OPTIONS
CNV504Nx Coupling network for 4 signal/data
CNV508Nx Coupling network for 8 signal/data
IMN 2 Impedance matching adapter to
As per IEC 61000-4-5, Levels 1 - 4 As per EN 61000-6-1, -6-2 Manual Standard Test routine
Change coupling after n pulses Change voltage after n pulses Change phase angle after n pulses
as per IEC 61000-4-9 Test levels 100, 300 and 1,000A/m Test level steplessly adjustable under Quick Start
lines as per IEC 61000-4-5
lines as per IEC 61000-4-5
match direct output to 2ohm source impedance
OUTPUTS
Direct Via HV connectors for external
coupling networks
(Zi = 2ohm with optional adapter
IMN 2) Coupling mode Line to line
Line(s) to ground DUT supply AC: 300V/16A; 50/60Hz
DC: 300V/16A CRO trigger 5V trigger signal for oscilloscope
MEASUREMENTS
CRO Û-monitor 10Vp at 5,000V CRO Î-monitor 10Vp at 2,500A Peak voltage 5,000V in the LCD display Peak current 2,500A in the LCD display
www.emtest.com © EM TEST > PAGE 5/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
POWER FAIL, DIPS & INTER­RUPTIONS, VOLTAGE VARIATIONS
POWER FAIL MODULE, PFS/N5
As per EN/IEC 61000-4-11,
IEC/EN 61000-4-29 and
EN 61000-6-1, -6-2 Channel PF1/PF2 AC voltage: max. 300V
AC current: max. 16A
DC voltage: max. 300V
DC current: max. 16A Frequency 16Hz - 500Hz and DC Switching time < 5us into a 100ohm resistive load Inrush current > 500A Protection Both channels are protected against
short-circuit conditions.
TRIGGER CIRCUIT
Trigger of events Automatic, manual, external Synchronization 0° - 360°, resolution 1° (16 - 500Hz) Repetition rate 10ms - 9,999s Event duration 20us - 9,999s
OUTPUTS
TEST ROUTINES
Quick Start On-line adjustable parameters,
easy-to-use
Standard Test routines
User Test routines Voltage variation, control of an
50/60Hz magnetic field
OPTIONS
V4780 Tapped autotransformer as per IEC
V4780 S2 Tapped autotransformer as per IEC
As per EN/IEC 61000-4-11 for AC supplies As per EN/IEC 61000-4-29 for DC supplies As per EN 61000-6-1, -6-2 Manual Standard Test routine
external variac Change phase angle after n events Change event duration after n events Inverse mode
As per EN/IEC 61000-4-8 Test levels 1, 3, 10 and 30A/m with external current transformer MC2630 Test levels 100, 300 and 1,000A/m with external current transformer MC26100
61000-4-11 Ed.2
61000-4-11 Ed.2 with automatic
change of tap DUT terminals L, N and PE CRO trigger 5V trigger signal for oscilloscope
MEASUREMENTS
DUT voltage In the LCD display DUT current In the LCD display MON V Measurement of the DUT voltage;
built-in 100:1 divider
MON I Measurement of the DUT current;
10mV/A; max. 1,000A
MV2616 Motorised variac (0 - 250V, 16A) MS100N Magnetic field coil, 1m x 1m MC2630 Current transformer for magnetic
fields up to 30A/m
MC26100 Current transformer for magnetic
fields up to 1,000A/m
CA PFS Calibration box for inrush current
verification as per IEC 61000-4-11
www.emtest.com © EM TEST > PAGE 6/8
DATA SHEET > UCS 500N5 > 20131118
TECHNICAL DETAILS
GENERAL DATA
INTERFACES
Serial interface USB Parallel interface IEEE 488, addresses 1 - 30 Analog output 0 - 10VDC to control an external
transformer
CN interface 15pin SubD connector to control an
external coupling network
Fail inputs DUT monitoring via Fail1 and Fail2
input (one each)
DIMENSIONS
Dimensions 19", 3HU, Weight approx. 25kg
MAINS
Supply voltage 115V/230VAC +10%/-15% Power approx. 75W
SPECIAL EQUIPMENT (ON REQUEST)
AVAILABLE MODELS:
UCS 500N5.1 Ultra compact simulator with EFT/N5
up to 5.5kV, VCS/N5 up to 5kV and
PFS/5; 1ph CDN 300V AC/DC (p-n) /
32A
UCS 500N5.2 Ultra compact simulator with
EFT/N5, VCS/N5 and PFS/N5; 1ph
CDN 400V AC/DC (p-n) / 16A
UCS 500N5.3 Ultra compact simulator with EFT/5,
VCS/5 and PFS/N5;
1ph CDN 400VAC (L-N) / 32A
UCS 500N5.7 Ultra compact simulator with EFT/N5
and VCS/N5; 1ph CDN 300V (p-n) /
16A (but without PFS module)
UCS 500N5.8 Ultra compact simulator for EFT/N5
and PFS/N5; 1ph CDN 300V AC/DC
(p-n) / 16A
Frequency 50/60Hz Fuses 2 x T 2AT (230V) or 2 x T 4AT (115V)
SAFETY
Safety standard EN/IEC 61010 Security circuit Control input (24VDC) Warning lamp Floating contact (max. 230V/6A)
ACCESSORIES INCLUDED
Mains supply Plug depends on the country of use DUT supply Plug depends on the country of use DUT adapter Socket depends on the country of
use Operation manual,
Calibration certificate, iec.control remote control software
OPTIONS
CNI 503Ax 3-phase coupling/decoupling
networks as per IEC 61000-4-4 and
-4-5 up to 100A per phase
iec.control 1 Remote control and documentation
software, including standard test routines and reporting capabilities.
www.emtest.com © EM TEST > PAGE 7/8
DATA SHEET > UCS 500N5 > 20131118
COMPETENCE WHEREVER YOU ARE
CONTACT EM TEST DIRECTLY
Switzerland
EM TEST (Switzerland) GmbH > Sternenhofstraße 15 > 4153 Reinach > Switzerland Phone +41 (0)61/7179191 > Fax +41 (0)61/7179199 Internet: www.emtest.ch > E-mail: sales.emtest@ametek.com
Germany
EM TEST GmbH > Lünener Straße 211 > 59174 Kamen > Deutschland Phone +49 (0)2307/26070-0 > Fax +49 (0)2307/17050 Internet: www.emtest.com > E-mail: info.emtest@ametek.de
France
EM TEST FRANCE > Le Trident - Parc des Collines > Immeuble B1 - Etage 3 > 36, rue Paul Cézanne > 68200 Mulhouse > France Phone +33 (0)389 31 23 50 > Fax +33 (0)389 31 23 55 Internet: www.emtest.fr > E-mail: info@emtest.fr
Poland
EM TEST Polska > ul. Ogrodowa 31/35, 00-893 Warszawa > Polska Phone +48 (0)518 64 35 12 Internet: www.emtest.com/pl > E-mail: info_polska.emtest@ametek.de
Information about scope of delivery, visual design and technical data correspond with the state of development at time of release. Technical data subject to change without further notice.
USA / Canada
EM TEST USA > 9250 Brown Deer Road > San Diego > CA 92121 Phone +1 (303) 693-1778 Internet: www.emtest.com > E-mail: sales.emtest@ametek.com
P.R. China
E & S Test Technology Limited > Rm 913, Leftbank > No. 68 Bei Si Huan Xi Lu > Haidian District > Beijing 100080 > P.R. China Phone +86 (0)10 82 67 60 27 > Fax +86 (0)10 82 67 62 38 Internet: www.emtest.com > E-mail: info@emtest.com.cn
Republic of Korea
EM TEST Korea Limited > #405 > WooYeon Plaza > #986-8 > YoungDeok-dong > Giheung-gu > Yongin-si > Gyeonggi-do > Korea Phone +82 (31) 216 8616 > Fax +82 (31) 216 8616 Internet: www.emtest.co.kr > E-mail: sales@emtest.co.kr
www.emtest.com © EM TEST > PAGE 8/8