Atec TOS9201 User Manual

TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Perfect design for System Operation, introducing our top of the line of Hipot / Insulation Resistance Testers
TOS9201(AC/DC)
TOS9200(AC)
Capable of performing hipot and insulation testing in com-
ply with safety standards, including IEC, EN, VDE, BS,
UL,CSA, JIS and the Electrical Application and Material
Safety Law (Japan)
The TOS9200 Series has been developed to meet a wide diversity of customer needs. Including the refi nement and enforcement of Kikusui’ s former series, its specifi cations refl ect the results of detailed study of our large database of user’s requirements including special orders and modifying specifi cations. The TOS9200 Series consists of four products : the testers TOS9200 and TOS9201, and the high-voltage scanners TOS9221 and TOS9220. The TOS9200 is equipped with AC hipot and insulation resistance testing functions, while the TOS9201 has a DC hipot testing function in addition to these two functions. The power block, a core component, employs a high-efficiency switching power supply and a switching amplifi er based on PWM systems. These features realize high power and enhanced stability, as well as reducing the size and weight of the unit. When combined with the ground bond tester TOS6200, the TOS9200 Series integrates three or four types of tests in a single process. Furthermore, when used together with the high-voltage scanner TOS9220/9221 (equipped with a contact check function), the tester is capable of automatically checking test points for up to 16 channels, thereby facilitating a safe, reliable automatic testing system.
Rise-time control function
Offset cancel function
Measured-value hold function
Output voltage monitoring function
Memory function
Program function
Interlock function
DC discharge function
Kikusu i Elect ronics Corpor ation
4
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Basic performance
Three functions - AC hipot testing, DC hipot testing and insulation resistance testing
The TOS9200 can perform AC hipot tests and insulation resistance tests, while the TOS9201 can also conduct DC withstanding tests. Once connected to a device being tested, the TOS9201 executes an AC hipot test, DC hipot test, and insulation resistance testing in succession in one process.
Insulation resistance testing at 25 V to 1000 V and 0.01 M
The test voltage can be set to 25 V through 1000 V at a resolution of 1 V. Insulation resistance covers a wide measurement range from 0.01 MΩ to
9.99 GΩ *. A single unit of the
AC hipot testing at 5 kV and 100 mA
Equipped with a high-effi ciency switching power supply in its high­voltage power block, a PWM-based switching amplifi er and a 500 VA high-voltage transformer, the TOS9200/TOS9201 realizes a maximum output of 5 kV/100 mA (continuous output for 30
TOS9200/9201 is capable of handling all test voltages required by JIS C 1302 1994 (Insulation Resistor Meter) and fully meets the JIS requirements.
*At a maximum rated current of 1 mA to 50 nA. minutes), or 2.5 times the output of Kikusui’s former models. At a test voltage of 500 V or more and an upper current of 100 mA, or greater the tester instantaneously satisfi es the requirements of a short-circuit current of 200 mA or more which is required by the IEC standard *. In addition, the tester ensures a load effects of 30% or less and the generation of a consistent 50 Hz/60 Hz test voltage free from the affect of the supply voltage. These features eliminate the need to readjust the output voltage once the test voltage is preset. *Continuous outputs are impossible because the output is cut off if an overcurrent is detected.
Enhanced measurement accuracy
The TOS9200/9201 is provided with a digital voltmeter for hipot
testing at an accuracy of ±(1% of reading + 30 V) and another one
for insulation resistance testing at an accuracy of ±(1% of reading
+ 1 V). Measured values are displayed not only during a test, but
while a program is being executed. A digital ammeter with an
accuracy of ±(3% of reading + 20 µA) is also provided for hipot
testing. Kikusui’s predecessors had the highest measurement
resolution of about 1 mA , with an accuracy of ±5% of the upper
cutoff current when it is set to 100 mA. In contrast, the digital
ammeter allows the TOS9200/9201 to make measurements at an
DC hipot testing at 6 kV and a maximum output of 50 W
The TOS9201 permits DC hipot testing at up to 6 kV *. The tester is equipped with a stable, low-ripple DC/DC converter with a load
accuracy of ±(3% of reading + 20 µA), even if the upper current is
set to 100 mA. The ammeter displays measured values while the
program executes, as well as during an AC or DC hipot test. factor of 1% or less. *Maximum output of 50 W for up to 1 minute.
*At 1 µA< measured current ≤ 1 mA
to 9.99 G
Ω
Test voltage Resistance measurement range
25V 0.03 MΩ to 500 M
50V 0.05 MΩ to 1.00 G
100V 0.10 MΩ to 2.00 G
125V 0.13 MΩ to 2.50 G
250V 0.25 MΩ to 5.00 G
500V 0.50 MΩ to 9.99 G
1000V 1.00 MΩ to 9.99 G
Type
Voltmeter for hipot testing ± (1% of reading + 30V)
Ammeter for hipot testing ± (3% of reading + 20µA)
Voltmeter for insulation resistance testing ± (1% of reading + 1V)
Insulation resistance meter ± (2% of reading)*
Display accuracy
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Ω
Kikusu i Elect ronics Corpor ation
5
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Diverse functions
Rise-time control function
In AC hipot testing, DC hipot testing and insulation resistance testing, you can apply a voltage gradually to reach the test voltage, instead of applying the test voltage directly at the start of a test. The voltage increase time can be set to 0.1 s through 99.9 s at a resolution of 0.1 s, and to 100 s to 200 s at a resolution of 1 s. The start voltage is also adjustable between 0% and 99% at a resolution of 1%.
Fall-time control function
In AC hipot testing, you can gradually decrease the test voltage after a PASS judgment. The voltage fall time is adjustable between 0.0 s and 99.9 s at a resolution of 0.1 s, and between 100 s and 200 s at a resolution of 1 s.
Offset cancel function
In AC hipot tests that require high sensitivity and high voltages, currents flowing into the stray capacity of the test lead wire, jigs, and other components can cause measurement errors. The TOS9200/9201 features a function to cancel these offset currents.
Voltage hold function
During measurement, this function allows you to hold the value of the voltage measured at the end of an AC or DC hipot test, as long as the test results are being displayed. When combined with the rise­time control function, this function enables to observe the insulation breakdown voltage.
Maximum Leakage current and minimum resistance hold function
By selecting “MIN/MAX Mode” in the measurement mode settings, you can hold the maximum current in hipot testing and the minimum resistance after the judgment wait time in insulation resistance testing. These values are shown on the tester’s display. They can also be read back via interface (GPIB or RS-232C).
Output voltage monitoring function
When the output voltage deviates from ±(10% of setting + 50 V), the monitoring function activates to suspend the test, thus ensuring highly reliable testing.
Memory function
Up to 100 test conditions used in AC and DC hipot testing and
insulation resistance testing, such as the test voltage, judgment value
and test time, can be stored with a specifi c name. For instance, you
can store the name of an applied safety standard and the destination
of the product to be tested. If test conditions are preset, operator can
recall relevant test conditions simply by entering the memory number.
If you previously assigned a special name to each of these test
conditions, operator can check recalled test conditions by name. The
memory function allows you to recall test conditions not only through
the recall operation on the front panel, but also by remote control.
[Storable test conditions]
Test voltage
Test frequency
Lower cutoff value
ON/OFF of the lower judgment function
Upper cutoff value
ON/OFF of the upper judgment function
ON/OFF of the offset function
Test time and ON/OFF of the timer function
Start voltage
Voltage rise time
Voltage fall time
Judgment wait time
Test voltage range
SLOW/MID/FAST settings for the response fi lter
FLOAT/GND of the LOW terminal
HIGH/LOW/OPEN settings for the scanner channel
ON/OFF of the contact check function
AC withstanding voltage testing
✔✔✔
✔✔✔
✔✔✔
✔✔✔
✔✔✔
✔✔
✔✔✔
✔✔✔
✔✔✔
✔✔✔
DC withstanding voltage testing
✔✔
Insulation resistance testing
Program function
By coordinating test conditions stored in an AC hipot test, DC hipot
test, and insulation resistance test, operator can sequentially run tests
that comprise up to 100 steps. When used together with the ground
bond tester TOS6200/6210, the TOS9200 Series permits continuous
tests combining test conditions stored in the TOS6200, as well as on
the TOS9200 itself. Sequential tests are possible, for example, on AC
hipot, insulation resistance, DC hipot, and ground bond, in order. The
TOS9200 Series stores up to 500 steps and 100 programs, which
can be recalled through the recall operation on the front panel or by
remote control.
Current detection response speed adjustment function
This function switches current detection response speeds for UPPER judgment by adjusting the integrated time constant of the current detection circuit. Three modes are available for the integrated time constant: SLOW (about 40 ms),MID (about 4 ms) and FAST (about 0.4 ms). SLOW mode is used in normal operations. MID and FAST modes are more effective in detecting a discharge occurring instantaneously or containing a large number of frequency components. They are also useful for hipot tests of test devices that insulation likely be breakdown, such as small electronic components.
Kikusu i Elect ronics Corpor ation
6
[Sample program]
Step 00 Step 01 Step 02
Memory Interval Memory Interval Memory Interval
ACW01
0.2s DCW01 0.2s IR01 0.2s
END
At Step 00, Step 01 and Step 02, memory ACW01 (AC hipot test), DCW (DC hipot test: TOS9201 only) and IR01 (insulation resistance test) are performed, receptively, in succession at
0.2-second intervals.
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Interfaces
REMOTE connector & SIGNAL I/O connector
The REMOTE connector on the front panel is intended exclusively for Kikusui’s options (remote control/ test probe). It allows start and stop operations by remote control. The SIGNAL I/O connector on the rear panel permits operator to recall panel memory and program memory contents by remote control, as well as controlling start and stop operations. Seven different signals are output from the SIGNAL I/O connector through the open collector.
[SIGNAL I/O]
No. Signal name I/O Details of signal 1 PM0 I LSB, LSD *1 2 PM1 I LSD *1 3 PM2 I LSD *1 4 PM3 I LSD *1 5 PM4 I MSD *1 6 PM5 I MSD *1 7 PM6 I MSD *1 8 PM7 I MSB, MSD *1 9 STB I Input terminal for the strobe signal of the panel memory and program memory 10 MODE0 I Selects a test mode *2 11 MODE1 I Selects a test mode *2 12 NC 13 COM Circuit common (chassis potential) 14 H.V ON O ON during a test and an automatic test (AUTO) or while a voltage remains between the output terminals 15 TEST O ON during a test (except for voltage rise and voltage fall) 16 PASS O ON during the time preset in the PASS HOLD settings when a PASS judgement is made 17 U FAUL O Continuously ON in an UPPER FAIL judgement. Continuously ON in a CONTACT FAIL judgement with the scanner connected. 18 L FAUL O Continuously ON in an LOWER FAIL judgement. Continuously ON in a CONTACT FAIL judgement with the scanner connected. 19 READY O ON during the READY status
PROTECTION
20 21 START I Input terminal for the START signal 22 STOP I Input terminal for the STOP signal 23 ENABLE I Input terminal for the ENABLE signal for the START signal 24 +24V Output terminal for +24 V internal power, with a maximum output current of 100 mA 25 COM Circuit common (chassis potential)
Input signal [Low active control input High-level input voltage: 11 V to 15 V /
Low- level input voltage: 0 V to 4 V / Low-level input current: Maximum –5 mA / Input interval: Minimum 5 ms]
Output signal [Open collector output (DC4.5V to 30V) / hipot: DC 30 V / Output
saturation voltage : Approximately 1.1 V (25 °C) /Maximum output current : 400 mA (TOTAL)]
* The input signal circuit is pulled up to +12V. Therefore, opening the input terminal is
equivalent to inputting a high-level signal.
*1 2-digit BCD low active input Signal input terminal for selection between the panel memory
for ACW, DCW, and IR, and the program memory for AUTO Memory recall by latching this
selection signal at the rise of the strobe signal *2 2-bit low active input Test mode ACW DCW IR AUTO MODE0 H L H L MODE1 H H L L
O ON when the PROTECTION function is activated
GPIB/RS-232C interface
A GPIB/RS-232C interface is provided as a standard feature to facilitate the remote control of all functions of the TOS9200/9201 except the POWER switch, the KEYLOCK function, and the program execution (AUTO) function.
RS-232C [Baud rate: 9600/19200/38400 bps/TOS6200/6210 interface (AUTO mode only): START/STOP control, test condition settings, reading of TOS6200/6210 measured values, and measurement results] GPIB [Remote control of all functions except the POWER switch, the KEYLOCK function, and the program execution (AUTO) function/SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C0, E1]
[Pin Confi guration for the SIGNAL I/O Connector]
Peripheral devices
High-voltage scanner TOS9220/TOS9221
TOS9221 Front View (same for TOS9220)
TOS9221
TOS9220
The high-voltage scanner TOS9220/TOS9221 has a function that distributes the test voltage provided by the TOS9200/9201 to multiple test points. Up to four channels can be used for outputs on this scanner. Each channel can be set to one of the three electric potential modes – HIGH, LOW, or OPEN. Operator can conduct AC/DC hipot and insulation resistance tests on any of the four test points. Furthermore, up to four scanners can be connected to the tester, allowing a maximum of 16 channels. The TOS9200 is equipped with a “contact check function” to check the contact between the output of each channel and a test point. These features ensure highly reliable and labor-saving hipot and insulation resistance tests for electrical and electronic equipment with multiple test points.
Pictures below are showing rear views of the units with cable clamp of output terminal removed.
     
TOS9221 Rear View
TOS9220 Rear View
Operation of the high-voltage scanner
On the TOS9200/TOS9201, you can select an electric potential mode for each channel–HIGH(high voltage side), LOW (low voltage side), and OPEN (open mode). The high-voltage scanner permits AC/DC hipot or insulation resistance tests on any of the four test points A to D. For instance, you can set CH1 (test point A) to HIGH,CH2 (test point B) to OPEN,and CH3 (test point C)CH4 (test point D)to LOW. To specify these settings, you can use the TOS9200/9201 panel or the GPIB/RS-232C.
*
High-voltage scanner(TOS9220)
Device being tested
CH1[HIGH]
CH2[OPEN]
CH3[LOW]
CH4[LOW]
OUT HIGH VOLTAGE terminal
OUT LOW terminal
IN HIGH VOLTAGE terminal
IN LOW terminal
Test point A
Test point B
Test point C
Test point D
To TOS9200
}
/9201
Kikusu i Elect ronics Corpor ation
7
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
For Stand alone use
...
Example of system for applying voltage by Test Lead or start/stop operation by Remote Control Box.
1
Item Model cable length
1 Hipot / Insulation Resistance Tester AC/DC TOS9201 1 pc.
2 High-Voltage Test Lead TL01-TOS 1.5m *1 1 set
3 Remote Control Box RC01-TOS *2 1.5m 1 pc.
*1: Also available for 3m cable, TL02-TOS
2
3
DUT
*2: Also available for both-hands operation, RC02-TOS
Example of system for applying voltage or start/stop operation by High-Voltage Test Probe.
1
Item Model cable length
1 Hipot / Insulation Resistance Tester AC/DC TOS9201 1 pc.
2 High-Voltage Test Lead HP01A-TOS 1.5m *1 1 pc.
*1: Also available for 3m cable, HP02A-TOS
DUT
2
Reguired numbers
Reguired numbers
For Multiple Channel Testing by High Voltage Scanner
Example of system consisting TOS9201 and TOS9221
1
3 4
1
3
2
to DUT
4
to DUT
5
5
2sets (8CH)
Item Model cable length
1 High-Voltage Scanner TOS9221 2 pc.
2 Hipot / Insulation Resistance Tester AC/DC TOS9201 1 pc.
3 Interface cable 85-50-0210 0.5m *1 2 pc.
4 High-Voltage Test Lead (red) TL07-TOS 1.5m 8 pc.
5 High-Voltage Leads for Parallel connection TL06-TOS 0.5m *2 2 set
*1: Also available for 2m cable, DD2M-8P
*2: Also available for 1.5m cable, TL04-TOS
[Rack mount bracket]
TOS9200 / 9201 (JIS) KRB150-TOS
(EIA) KRB3-TOS
TOS9220 / 9221 (JIS) KRB100-TOS
(EIA) KRB2-TOS
[CAUTION] In case of using more than 2sets of High Voltage Scanner, it is required
to rack mount or locate these units to the side of Hipot / Insulation Resistance Tester,
and it should not be piled up more than 2sets of High Voltage Scanner units.
...
Reguired numbers
Kikusu i Elect ronics Corpor ation
8
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Single process to apply until ground bond test
Example of system consisting TOS9201 and TOS6210
1
Item Model cable length
1 Ground Bond Tester TOS6210 1 pc.
(from front panel)
3
2
4
(from front panel)
5
DUT
2 Hipot / Insulation Resistance Tester AC/DC TOS9201 1 pc.
3 RS-232C Cross Cable (
4 Low-Voltage Test Lead TL12-TOS 1.5m 1 set
5 High-Voltage Test Lead TL01-TOS 1.5m *1 1 set
*
1: Also available for 3m cable, TL02-TOS
[Rack mount bracket]
TOS9200 / 9201 (JIS) KRB150-TOS
(EIA) KRB3-TOS
TOS6210 / 6200 (JIS) KRB100-TOS
(EIA) KRB2-TOS
...
9pin female-9pin female
Reguired numbers
) 1 pc.
It is capable to perform for hipot / Insulation
Resistance and Ground bond testing in one
single process by controlling TOS6210 from
TOS9201.
Fully Automated System by PC
...
Example of system consisting TOS9201, TOS9221 (4CH) and TOS6210
1
Item Model cable length
1 High-Voltage Scanner TOS9221 1 pc.
2 Hipot / Insulation Resistance Tester AC/DC TOS9201 1 pc.
4
2
3
9
8
5
to DUT
8
Possible to control TOS9201 and TOS6210 and acquire the test result.
6
(from front panel)
7
to DUT
3 Ground Bond Tester TOS6210 1 pc.
4 Interface cable 85-50-0210 0.5m *1 1 pc.
5 High-Voltage Test Lead (red) TL07-TOS 1.5m 4 pc.
6 High-Voltage Leads for Parallel connection TL06-TOS 0.5m *2 1 set
7 Low-Voltage Test Lead TL12-TOS 1.5m 1 set
8 GPIB Cable 408J-102 2m *3 2 pc.
9 PC (with GPIB Interface cable) 1 pc.
*1: Also available for 2m cable, DD2M-8P *2: Also available for 1.5m cable, TL04-TOS *3: Also available for 1m cable, 408J-101 and 4m cable, 408J-104 [Rack mount bracket] TOS9200 / 9201 (JIS) KRB150-TOS (EIA) KRB3-TOS TOS9220 / 9221 / 6210 / 6200 (JIS) KRB100-TOS (EIA) KRB2-TOS
[CAUTION] In case of use for combining more than 2sets of High Voltage Scanner unit and Ground Bond Tester, it is required to rack mount or locate these units to the side of Hipot / Insulation Resistance Tester, and it should not be piled up more than 2sets of High Voltage Scanner units.
Reguired numbers
Kikusu i Elect ronics Corpor ation
9
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Hipot test mode
Item TOS9200 TOS9201
Output section
Output-voltage range 0.05 kV to 5.00 kV AC
Maximum rated load (*1) 500 VA (5 kV/100 mA)
Maximum rated current 100 mA [output voltage of 0.2 kV or more]
Transformer capacity 500 VA
Output-voltage waveform(*2) Sine wave
AC
Frequency 50 Hz/60 Hz
Voltage regulation ±3% or less [maximum rated load → no load]
Short-circuit current 200 mA or more, 350 mA or less [at output voltage of 0.5 kV or more]
Type of output PWM switching
Output-voltage range ———— 0.05 kV to 6.00 kV DC
Maximum rated load (*1) ———— 50 W (5 kV/10 mA)
Maximum rated current ———— 10 mA
DC
Ripple No load at 5 kV ———— 50 Vp-p Typ.
Voltage regulation ———— 1% or less [maximum rated load → no load]
Short-circuit current ———— 40 mA Typ.
Discharge function ———— Forced discharge at the end of test(discharge resistance: 125 kΩ)
Start voltage The voltage at the start of the test can be set as the start voltage.
Output-voltage monitoring function If the output voltage exceeds ±(10% of the setting + 50 V), output is cut off and the protection function activates.
Voltmeter
Analog
Digital
*1 Time limitation on output The tester’s hipot generator is designed to radiate half as much heat as the rated output, in consideration of the size, weight, cost, and other factors of the tester. It is therefore necessary to use the tester within the ranges specifi ed below. Operations deviating from these ranges may heat the output section excessively, thereby activating the protective circuit. In such a case, suspend the test and wait until the temperature falls to the normal level.
[Output limitation in hipot testing (Output time = voltage rise time + test time + voltage fall time)]
Ambient temperature Upper current Pause Time Output time
t ≤ 40 ºC
*2 Test-voltage waveform When an AC test voltage is applied to a capacitive load, it is possible that the voltage becomes higher even than that when in the no load state. Furthermore, waveform distortion also may occur if the capacitance of the load is voltage-dependent (such as of ceramics capacitors). When the test voltage is not higher than 1.5 kV and the capacitance is not larger than 1000 pF, such test voltage changes are only of negligible levels. As the output type of the high-voltage generator block of the tester is PWM switching, switching noise and spike noise that the test voltage includes increase when the test voltage is 500 V or less. The lower the test voltage is, the more the waveform distortion increases.
Item TOS9200
Ammeter(*3)
Measurement range
Display
Accuracy ±(3% of the reading + 20 μA) [after the offset cancel function is activated, if the scanner is mounted]
Response Mean-value responsive / root-mean-square value display (response time of 200 ms)
Hold function The measured current at the end of the test is held during the PASS judgment time period.
Offset cancel function
Calibration Performs calibration using the root-mean-square value of a sine wave with a pure resistive load
Selection of LOW/GUARD for the GND (*4)
Time
Setting range for the voltage rise time (RISE TIME)
Setting range for the voltage fall time (FALL TIME)
Setting range for the test time (TEST TIME)
Setting range for the judgement wait time (WAIT TIME)
Accuracy ± (100 ppm + 20 ms)
Resolution 10 V
Accuracy ±(1.5% of setting + 20 V) [with no load]
Distortion 2% or less [with no load or pure resistive load at output voltage of 0.5 kV or more applied]
Accuracy ±0.1%
Resolution ———— 10 V
Accuracy ———— ±(1.5% of the setting + 20 V)
Maximum rated load ———— 150 Vp-p Typ.
Setting range 0% to 99% of the test voltage (resolution of 1%)
Scale 6 kV AC/DC F.S
Accuracy ±5% F.S
Indicator Mean-value responsive/root-mean-square value scale
Measurement range 0.0 kV to 6.00 kV AC/DC
Resolution 10 V
Accuracy ±(1.0% of the reading + 30 V)
Response Mean-value responsive/root-mean-square value display (response time of 200 ms)
HOLD function The voltage measured at the end of test is held during the PASS and FAIL judgment time period.
50< i ≤ 110 mA At least as long as the output time Maximum of 30 minutes
AC
i ≤ 50 mA Not necessary Continuous output possible
5< i ≤ 11 mA At least as long as the output time Maximum of 1 minute
DC
0.00 mA to 110 mA AC 0.00 mA to 110 mA AC/0.00 mA to 11 mA DC
i ≤ 5 mA At least as long as the judgement wait time (WAIT TIME) Continuous output possible
i < 1 mA 1 mA ≤ i < 10 mA 10 mA ≤ i < 100 mA 100 mA ≤ i
LOW
GUARD
A
❑ ❑ ❑ μ
The current fl owing to the insulation resistor between the output cables and the stray capacity is cancelled up to 100
Selection permitted for current measurement between the mode for the GND point connected to the LOW terminal,and the mode using guard.
Connects the GND point to the LOW terminal. Measures the current fl owing to the LOW terminal (chassis) (for normal operation).
Sets the GND point as guard. Measures the current fl owing to the LOW terminal, but does not measure the current fl owing to the chassis (for high-sensitivity, high-accuracy measurements).
0 s to 200 s (Valid only with PASS judgement)
❑.❑ ❑
mA
————
mA
❑ ❑.❑
0.1 s to 200 s
0.3 s to 999 s With the TIMER OFF function
mA
❑ ❑ ❑
0 s to 200 s (Valid only with PASS judgementin AC hipot testing)
0.3 s to 10 s (Only for DC hipot testing)[RISE TIME + TEST TIME > WAIT TIME]
TOS9201
i = measured current
A/kV (in AC hipot testing only).
μ
Kikusu i Elect ronics Corpor ation
10
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Item TOS9200 TOS9201 Judgement function Judgement method/action
Setting range for the upper current (UPPER) Setting range for the lower current(LOWER) 0.01 mA to 110 mA AC(With the LOWER OFF function) Judgement accuracy (*3) ±(3% of setting + 20 μA) [After the offset cancel function is activated, if the scanner is mounted] Current detection method The absolute current values are integrated and compared with the reference value. Response-speed switching function
*3 In AC hipot testing, a current fl ows into the stray capacity of measurement leadwire and fi xtures. When the optional high-voltage scanner TOS9220/9221 is used, a current of approximately 22 μA/kV fl ows into the stray capacity of each scanner. The table below shows the approximate currents fl owing into such stray capacity. When the LOW terminal is set to GND, a current fl owing into the stray capacity is added for measurement purposes to the current fl owing into the DUT. In particular, for high-sensitivity, high­ accuracy judgement, it is necessary to add the current fl owing into the stray capacity to the lower/upper current.When the LOW terminal is set to FLOAT, the effect of the current fl owing into the stray capacity is negligible. If the offset cancel function is used, the current fl owing into the stray capacity can be eliminated from the measurement.
Output voltage 1kV 2kV 3kV 4kV 5kV
Hanging a 350-mm test lead wire (Typ. value) 2 μA4
Using the accessory leadwire TL01-TOS (Typ. value) 16 µA 32 µA 48 µA 64 µA 80 µA
High-voltage scanner (Typ. value, not including the test leadwire) 22 µA 44 µA 66 µA 88 µA 110 µA
*4 With the GND set to GUARD, current measurement is disabled if the part of the DUT connected to the LOW terminal is grounded, which poses extreme danger. Never ground the DUT. In ordinary operation, set the GND to LOW.
Insulation Resistance Testing Mode
Item TOS9200 TOS9201
Output section
Output-voltage range -25 V to -1000 V DC
Maximum rated load 1 W (-1000 V DC/1 mA)
Maximum rated current 1 mA
Ripple 1 kV no-load 2 Vp-p or less
Voltage regulation 1% or less [Maximum rated load → no load]
Short-circuit current 12 mA or less
Discharge function Forced discharge at the end of test (discharge resistance : 25 kΩ)
Output-voltage monitoring function If the output voltage exceeds ±(10% of the setting + 50 V), output is cut off and the protection function activates.
Voltmeter
Analog Scale 6 kV AC/DC F.S
Digital Measurement range 0 V to -1200 V
Resistance meter
Measurement range 0.01 MΩ - 9.99 GΩ (Within the maximum rated current range of 1 mA to 50 nA)
Display
Accuracy
Hold function The measured current at the end of the test is held during the PASS period. Selection of LOW/GUARD for the GND (*5)
Resolution 1 V
Setting accuracy ±(1.5 % of Setting + 2 V)
Maximum rated load 10 Vp-p or less
Accuracy ±5% F.S
Indicator Mean-value responsive / root-mean-square value scale
Resolution 1 V
Accuracy ±(1 % of reading + 1 V)
LOW GUARD Sets the GND point as guard. Measures the current fl owing to the LOW terminal, but does not measure the current
Judgement Judgement method Display Buzzer SIGNAL I/O
UPPER FAIL When the tester detects a current exceeding the upper current,
it cuts off the output and makes an UPPER FAIL judgement. In DC hipot testing, however, no judgement is made until the judgement wait time (WIT TIME) has elapsed.
LOWER FAIL When the tester detects a current below the lower current,
it cuts off the output and makes a LOWER FAIL judgement. However, no judgement is made during the voltage rise time (RISE TIME) or voltage fall time (FALL TIME) in AC hipot testing.
PASS When the preset time has elapsed without any abnormalities,
the tester cuts off the output and makes a PASS judgement.
The FAIL LED lights up. Displayed on the LCD
The FAIL LED lights up. Displayed on the LCD
The PASS LED lights up. Displayed on the LCD
ON
ON
ON
Outputs the U FAIL signal
Outputs the L FAIL signal
Outputs the PASS signal
• The PASS signal is output at the timing preset on PASS HOLD. If HOLD is set, the PASS signal is output continuously until the STOP signal is input.

• The UPPER FAIL signal and the LOWER FAIL signal are output continuously until the STOP signal is input.
• The FAIL and PASS buzzer volumes are adjustable. However, they cannot be adjusted individually, as they are set in common.
0.01 mA to 110 mA AC 0.01 mA to 110 mA AC / 0.01 mA to 11 mA DC
0.01 mA to 110 mA AC /0.01 mA to 11 mA DC (With the LOWER OFF function)
The current-detection response speed for UPPER FAIL judgement can be set to FAST/MID/SLOW (for AC hipot testing only).
A6
μ
R < 10.0 MΩ10.0MΩ ≤ R < 100.0MΩ100.0MΩ ≤ R < 1.00GΩ1.00GΩ ≤ R ≤ 9.99G
❑.❑ ❑ MΩ ❑ ❑.❑ MΩ ❑ ❑ ❑ MΩ ❑.❑ ❑ GΩ
50 nA ≤ i ≤ 100 nA 100 nA < i ≤ 200 nA 200 nA < i ≤ 1 μA1 μA < i ≤ 1 mA
± (20 % of reading) ± (10 % of reading) ± (5 % of reading) ± (2 % of reading)
A8
μ
A 10 µA
μ
Ω
R = measured insulation resistance
i = measured current
[In the humidity range of 20 %rh to 70 %rh (no condensation), with no disturbance such as swinging of the test leadwire]
Selection permitted for current measurement between the mode for the GND point connected to the LOW terminal, and the mode using guard. Connects the GND point to the LOW terminal. Measures the current fl owing to the LOW terminal (chassis) (for normal operation).
fl owing to the chassis (for high-sensitivity, high-accuracy measurements).
Kikusu i Elect ronics Corpor ation
11
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Item TOS9200 TOS9201 Judgement function Judgement method/action
Setting range for the upper resistance (UPPER) Setting range for the lower resistance (LOWER) Judgement accuracy For both UPPER and LOWER
Time Setting range for the voltage rise time (RISE TIME) Setting range for the test time (TEST TIME) Setting range for the judgement wait time (WAIT TIME) Accuracy ± (100 ppm + 20 ms)
5 When the GND is set to GUARD, current measurement is disabled if the part of the DUT connected to the LOW terminal is grounded, which poses extreme danger. Never ground the DUT.
*
In ordinary operation, set the GND to LOW.
General Specifi cations
Item
Environment
Installation location Indoors at an altitude of up to 2000 m
Warranty range Temperature 5 °C to 35 °C
Operating range Temperature 0 °C to 40 °C
Storage range Temperature -20 °C to 70 °C
Power requirements
Nominal voltage range (Allowable voltage range)
Power consumption
Allowable frequency range 47 Hz to 63 Hz
Insulation resistance 30 MΩ or more (500 V DC) [between the AC LINE and chassis]
Hipot 1390 V AC, 2 seconds, 20 mA or less [between the AC LINE and chassis]
Ground bond 25 A AC/0.1 Ω or less
Electromagnetic compatibility (EMC) (*6) Conforms to the requirements of the following directive and standard.
Safety (*6,7) Conforms to the requirements of the following directive and standard.
Dimensions (maximum) 430 (455) W x 132 (150) H x 370 (440) D mm
Weight Approx. 19 kg
Humidity 20 %rh to 80 %rh (No condensation)
Humidity 20 %rh to 80 %rh (No condensation)
Humidity 90 %rh or less (No condensation)
Using no load (READY)
Using the rated load Maximum of 800 VA
Judgement Judgement method Display Buzzer SIGNAL I/O
UPPER FAIL When the tester detects a resistance exceeding the upper cutoff resistance,
it cuts off the output and makes an UPPER FAIL judgement. However, no judgement is made during a voltage rise time (RISE TIME).
LOWER FAIL When the tester detects a resistance below the lower cutoff resistance,
it cuts off the output and makes a LOWER FAIL judgement. However, no judgement is made until the judgement wait time (WAIT TIME) has elapsed.
PASS When the preset time has elapsed without any abnormalities,
the tester cuts off the output and makes a PASS judgement.
The FAIL LED lights up. Displayed on the LCD
The FAIL LED lights up. Displayed on the LCD
The PASS LED lights up. Displayed on the LCD
ON Outputs the
ON Outputs the
ON Outputs the
• The PASS signal is output at the timing preset on PASS HOLD. If HOLD is set, the PASS signal is output continuously until
the STOP signal is input.
• The UPPER FAIL signal and the LOWER FAIL signal are output continuously until the STOP signal is input.
• The FAIL and PASS buzzer volumes are adjustable. However, they cannot be adjusted individually, as they are set in common.
0.01 MΩ to 9.99 GΩ [Below the maximum rated current]
0.01 MΩ to 9.99 GΩ [Below the maximum rated current]
Judgement current 50 nA ≤ i ≤ 100 nA 100 nA < i ≤ 200 nA 200nA < i ≤ 1 μA1 μA < i ≤ 1 mA
UPPER, LOWER
0.01 ≤ R < 10.0 M
10.0 ≤ R < 50.0 M
50.0 ≤ R < 100 M
100 MΩ ≤ R < 200 M
200 MΩ ≤ R < 500 MΩ± (20 % of setting + 5digit) ± (10 % of setting + 5digit) ± (5 % of setting + 5digit) ± (2 % of setting + 3digit)
500 MΩ ≤ R < 1.00 GΩ± (20 % of setting + 5digit) ± (10 % of setting + 5digit) ± (5 % of setting + 5digit) ± (2 % of setting + 3digit)
1.00 GΩ ≤ R < 2.00 GΩ± (20 % of setting + 10digit) ± (10 % of setting + 5digit) ± (5 % of setting + 5digit) ––
2.00 GΩ ≤ R < 5.00 GΩ± (20 % of setting + 20digit) ± (10 % of setting + 10digit) ± (5 % of setting + 5digit) ––
5.00 GΩ ≤ R < 10.0 GΩ± (20 % of setting + 20digit) ± (10 % of setting + 10digit) –– ––
Ω
Ω
Ω
Ω
–– –– –– ± (2 % of setting + 3digit)
–– –– ± (5 % of setting + 5digit) ± (2 % of setting + 3digit)
–– –– ± (5 % of setting + 5digit) ± (2 % of setting + 3digit)
–– ± (10 % of setting + 5digit) ± (5 % of setting + 5digit) ± (2 % of setting + 3digit)
Judgement current = test voltage/(UPPER,LOWER) [In the humidity range of 20 %rh to 70 %rh (no codensation), with no disturbance such as swinging of the test leadwire] [In LOWER judgement, at least 0.5 s is necessary for testing after the WAIT TIME has elapsed. In LOWER judgement for 200 nA or lower, a wait time of at least 1.0 s is necessary.]
0.1 s to 200 s
0.5 s to 999 s With the TIMER OFF function
0.3 s to 10 s [RISE TIME + TEST TIME > WAIT TIME]
TOS9200 TOS9201
100 V to 120 V AC / 200 V to 240 V AC (85 V to 130 V AC / 170 V to 250 V AC) Selectable
EMC Directive 89/336/EEC, EN61326, EN61000-3-2, EN61000-3-3 Under following conditions
1. Used test leadwire TL01-TOS which is supplied. 2. No discharge occurs at outside of the tester.
3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.
Low Voltage Directive 73/23/EEC , EN61010-1, Class I, Pollution degree 2
100 VA or less
U FAIL signal
L FAIL signal
PASS signal
Kikusu i Elect ronics Corpor ation
12
TOS9200 SERIES
Hipot Tester with Insulation Resistance Test
Accessory
Item TOS9200 TOS9201
AC Power cable 1 pc.
High-voltage test lead wire TL01-TOS (1.5 m)
1 set
Interlock jumper 1 pc.
High-Voltage Danger seal 1 sheet
Fuse 1 pc.
Operation Manual Operation Manual for Tester: 1 copy, Operation for GPIB/RS-232C Interface: 1 copy
*6Only on models that have CE marking on the panel. Not applicable to custom order models. *7 This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument. The safety of the instrument is not guaranteed unless the instrument is grounded properly.
Electrical performance
Item TOS9200 TOS9201
Maximum rating voltage
Number of channels 4 (Each channel is settable to HIGH, LOW, or OPEN.)
Maximum number of scanners connected 4 scanners,Channel numbers are determined in order of connection to the TOS9200/9201 tester.
Contact check function None (*1) Provided
Lamps and LEDs POWER Lights as it is interlocked with the POWER switch of the TOS9200/9201 tester
Power requirements
Nominal voltage range (allowable voltage range)
Power consumption
Allowable frequency range 47 Hz to 63 Hz
Insulation resistance 30 M Ω or more (500 V DC) [between the AC LINE and chassis]
Hipot 1390 V AC, 2 seconds, 10 mA or less [between the AC LINE and chassis]
Ground bond 25 A AC/0.1 Ω or less
Electromagnetic compatibility (EMC) (*2) Conforms to the requirements of the following directive and standard.
Safety (*2,3) Conforms to the requirements of the following directive and standard.
Environment
Installation location Indoors and at altitudes up to 2000 m
Warranty range Temperature 5 °C to 35 °C
Operating range Temperature 0 °C to 40 °C
Storage range Temperature -20 °C to 70 °C
Dimensions 430(435)W ✕ 88(105)H ✕ 370(415) Dmm
Weight Approx. 6.5 kg
Accessories
AC power cable 1 pc.
High-voltage test leadwires, red 4 pc. (1.5 m each) 8 pc. (1.5 m each)
High-voltage leads for parallel connection
Interface cable 1 pc.(0.5 m)
Channel-indication stickers For the panel face: 1 sheet; for the test leadwires: 1
“HIGH VOLTAGE, DANGER” stickers 2 sheets
Fuses 2 pc. (including a spare contained in the fuse holder)
Operation Manual 1 copy
*2 Only on models that have CE marking on the panel. Not applicable to custom order models. *3 This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument. The safety of the instrument is not guaranteed unless the instrument is grounded properly. [Measurement accuracy achieved when the scanner and the TOS9220/9201 tester are connected] In an AC hipot test, a current of approx. 22 µA/kV fl ows per scanner due to stray capacitance in the scanner in comparison with use of the TOS9220/9201 tester alone. Note that this current may contribute to errors in current measurements conducted by the TOS9220/9201 tester.
AC 5.0 kV
DC 6.0 kV
1 st scanner CH1 to CH4 2 nd scanner CH5 to CH8 3 rd scanner CH9 to CH12 4 th scanner CH13 to CH16
DANGER Lights as it is interlocked with the DANGER lamp of the TOS9200/9201 tester
CHANNEL Lights during a test at each channel HIGH: red; LOW: green; Under contact check: orange
100 V to 120 V AC/200 V to 240 V AC (85 V to 132 V AC/170 V to 250 V AC) Automatic switching
In READY state Approx. 12 VA
During test 40 VA maximum
EMC Directive 89/336/EEC, EN61326, EN61000-3-2, EN61000-3-3 Under following conditions
1. Used test leadwire TL07-TOS which is supplied. 2. No discharge occurs at outside of the tester.
3. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.
Low Voltage Directive 73/23/EEC, EN61010-1, Class I, Pollution degree 2
Humidity 20 %rh to 80 %rh (no condensation)
Humidity 20 %rh to 80 %rh (no condensation)
Humidity 90 %rh or less (no condensation)
1 set (0.5 m each)
External dimensional diagrams
MAX455
MAX10
MAX150
132
430
MAX 30
MAX440
370
88
MAX105
MAX435
430
MAX5
MAX415
370
Unit: mm
Kikusu i Elect ronics Corpor ation
13
TOS5101
Hipot Tester
Output block Applied Voltage 0 to 5/ 0 to 10 kV AC and DC AC Maximum Rated*1 500VA / 10 kV, 50 mA Waveform Commercial line waveform
Voltage Regulation
Switching Use of a zero turn-on switch DC Applied Voltage 50W / 10 kV, 5 mA Ripple 100 Vp-p typ. at 10 kV, no load
Maximum Rated*1 Max. 3% (for max. rated load to no load) Output Voltmeters Analog Scale 10 kV full scale , AC/DC Class JIS Class 2.5 Accuracy ±5% of full scale AC Indication Mean value response / rms value scale Digital Full Scale 5 kV/ 10 kV full scale Accuracy ±1.5% of full scale AC Response Mean value response / rms value display Ammeter Digital Accuracy ±(5% + 20μA) of upper cutoff current AC Response Mean value response / rms value display Pass/fail Judgement Function Type of Judgement
Upper cutoff current setting range AC: 0.1 to 55 mA DC: 0.1 to 5.5 mA Lower cutoff current setting range AC: 0.1 to 55 mA DC: 0.1 to 5.5 mA Judgement Accuracy ±(5% of upper cutoff current + 20μA) Current Detection Integration of current absolute value fol-
Calibration With rms value of sine wave using a pure
No-load output voltage required for detection
Test Time Setting Range 0.5 to 999 sec (±10 ms) (timer-off function
Accuracy ±20 ms Line Voltage 100V±10%, 50/60 Hz (Nominal voltages of
Power Requirements for line voltage of 100 V Max. 50 VA under no-load conditions
for line voltage of 100 V to 200 V Max. 50 VA under no-load conditions
for line voltage of 220 V to 240 V Max. 50 VA under no-load conditions
Electromagnetic compatibility (EMC) Conforms to the requirements of the
Max. 15% (for max. rated load to no load)
200 Vp-p typ. at max. rated output
Window comparator type
FAIL judgement
*When current detected above upper cutoff current *When current detected below lower cutoff current
(FAIL signal generated when FAIL judgement made)
PASS judgement
*When set time has elapsed and no abnormality is detected
lowed by comparison with reference value.
resistance load.
Approx. 970 V when set to 50 mA AC
Approx. 160 V when set to 5 mA DC
provided)
110V, 120V, 220V, 230V and 240V avail­able as factory options.)
/ Approx. 600 VA at rated load
/ Approx. 600 VA at rated load
/ Approx. 610 VA at rated load
following directive and standard.*2
EMC Directive 89/336/EEC EN61326 EN61000-3-2 EN61000-3-3
Under following conditions
1. Used HV test leadwires which is supplied.
2. No discharge in testing.
3. Used the shielded cable less than three meters when the SIGNAL I/O
is used.
which length is
Safty
Conforms to the requirements of the follow­ing directive and standard. *2,4
Low Voltage Directive 73/23/EEC EN61010-1
Class I Pollution degree 2
Insulation resistance 30 M Ω or more (500 V DC) Hipot
Environment
1390 VAC, 2 seconds [between the AC LINE and chassis]
1200 VAC, 1 second [UL-approved products only]
Specifi cation range : 5 °C to 35°C / 20 %rh to 80 %rh
Operable range : 0 °C to 40°C / 20 %rh to 80 %rh
Storage range : -20 °C to 70 °C / 80 %rh or less
Dimensions (MAX) 430W ✕ 177(195)H ✕ 370(450)Dmm Weight for line voltage of 100 V Approx. 21 kg for line voltage of 100 V to 120 V Approx. 23 kg for line voltage of 220 V to 240 V Approx. 24 kg Accessories High-voltage test lead
Others
TL01-TOS TL03-TOS
(max.allowablevoltage: 5 kV /1.5m)
(max.allowablevoltage: 10 kV /1.5m)
14-pin amphenol plug (assembled)
*1: Continuous output time may be limited depending on current high limit reference
value and ambient temperature.
*2: Only on models that have CE marking on the panel. Not applicable to custom order
models. *3: Not applicable to custom order models. *4: This instrument is a Class I equipment. Be sure to ground the protective conductor
terminal of the instrument. The safety of the instrument is not guaranteed unless the
instrument is grounded properly.
External dimensional diagrams
430
177
MAX195
MAX450
45
[Pin Confi guration for the
SIGNAL I/O Connector]
7654321
14
370
READY L FAIL UFAIL PASS TEST H.VON N.C
8910111213
PROTECTION INTERLOCK RRSTART RRSTOP RRENABLE ISOLCOM ISOLCOM
Unit: mm
Kikusu i Elect ronics Corpor ation
21
Loading...