TDS6000B/C Series digital storage
oscilloscopes provide unprecedented
performance along with a complete
feature set designed to address design
validation, debug and compliance challenges of next generation computer,
datacom and communications equipment.
High bandwidth, high sample rate and
deep memory also provide the ideal
solution for data acquisition applications.
Uncompromised Acquisition
You won’t need to need to trade-off
bandwidth, record length and sample
rate for your serial data measurement
and analysis needs. The TDS6000C
DSOs provide acquisition architecture
with 40 GS/s maximum sample rate
and 64 Msamples record length on two
channels (20 GS/s and 32 Msamples on
each of the four channels simultaneously),
for the acquisition power you need. They
provide the ultimate combination of bandwidth, sample rate and record length
for the fastest signals. The TDS6154C
provides matched 15 GHz performance
across any two channels using advanced,
Tektronix-proprietary DSP enhancement,
important for high-speed, channel-to-channel
measurements. The user-selectable DSP
filter on each channel provides magnitude
and phase correction, plus extension of
the analog bandwidth to 15 GHz for more
accurate signal fidelity on high speed
measurements – easily capturing the fifth
harmonic of 3.0 GHz embedded clocks
used in next generation 6.0 Gb/s serial data
standards, and even the third harmonic of
5 GHz clocks being developed for future
systems. The DSP filter on each channel
can also be switched off to take advantage
of true 12 GHz analog bandwidth for
applications needing the highest available
raw data capture.
Features & Benefits
Bandwidths of 15 GHz (TDS6154C),
12 GHz (TDS6124C), 8 GHz
(TDS6804B) and 6 GHz (TDS6604B)
Rise Times to 19 ps 20 to 80%
(28 ps 10 to 90%) on TDS6154C
Typical Rise Time, with Channelmatched, User-selectable DSP
40 GS/s Real-time Sample Rate
on Two Channels*1, 20 GS/s Realtime Sample Rate on All Four
Channels with 500 fs/sample
Interpolated Points
Up to 64 Msamples on Two
Channels*1, up to 32 Msamples
Record Length on All Four
Channels with MultiView Zoom
Function for Quick Navigation
MyScope®Custom Control
Windows Enhance Productivity
Right Mouse Click Menus
for Exceptional Efficiency
Pinpoint™Triggering Provides
the Most Flexible and Highest
Performance Triggering, with Over
1400 Combinations to Address
Virtually Any Triggering Situation
Serial Pattern Triggering up
to 3.125 Gb/s with 8b/10b
Protocol Triggering
Serial Data Analysis and
Compliance at Rates of
6.25 Gb/s and Above
OpenChoice®Software with
Microsoft Windows XP OS
Delivers Built-in Networking
and Analysis
Technology Specific Software
Solutions Provide Built-in
Domain Expertise for Serial
Data, Jitter, Ethernet, DVI,
USB 2.0, Communications
and Power Measurements
System Includes: Dual Processor
System (2.8 GHz Pentium 4
and 583 MHz PowerPC), High
Resolution XGA Display, Front
Panel CD-R/W, Front Panel USB
2.0 Port and 1000Base-T
Network Connection
™
Applications
Signal Integrity, Jitter
and Timing Analysis
Validation, Debug, Characterization
and Compliance of Next
Generation Digital Designs
Computer, Datacom, Storage
Area Network Equipment Designs
and High-speed Backplanes
High Energy Physics Measurements
and Data Acquisition
*1
For C Model versions only.
Page 2
Digital Storage Oscilloscope
TDS6000B/C Series
MyScope®Custom
Control Windows
MyScope control windows allow you to
build your own control windows with only
the controls, features and capabilities that
you care about and are important in your
job. For the first time you can create your
own personalized “toolbox” of oscilloscope features. No longer do you need
to search through menus for features or
re-learn how to drive the oscilloscope
after a break from the lab. MyScope
control windows are easily created in a
matter of minutes using a simple, visual,
drag-and-drop process. Once created,
these customized windows are easily
accessed through a dedicated MyScope
button and menu selection on the oscilloscope button/menu bar, just like any
other control window. You can make
an unlimited number of custom control
windows, enabling each person who
uses the oscilloscope, in a shared environment, to have their own unique control
window. Since the control windows are
stored as files on the hard drive, they can
easily be transferred to other TDS5000B or
TDS/CSA7000B Series oscilloscopes, or
they can even be e-mailed to a co-worker
around the world when the need arises.
MyScope control windows will benefit
all oscilloscope users, from eliminating
the ramp-up time that many face when
returning to the lab after not using an
oscilloscope for a while, to the power
user who can now operate far more
efficiently. Everything you need is found
in one control window rather than having
to constantly navigate through menu after
menu to repeat similar tasks.
Right Clicks
Right mouse click menus make simple
things as they should be – simple. Right
click menus are context sensitive, meaning
the choices presented in the menu depend
on where you right clicked the mouse.
This makes right click menus extremely
intuitive. Want to change the cursor type?
Right click on a cursor or the cursor readouts. Want to change the reference levels of
an automatic measurement? Right click on
the measurement. Want to change trigger
parameters? Right click on the trigger
readouts. Want to change a waveform’s
color? Right click on the waveform handle.
Virtually all objects on the oscilloscope
display have right click menus associated
with them that include all the appropriate
actions or features relative to those objects.
There are also right click menus for regions
of the display in addition to just objects.
For example, right clicking in the main
graticule brings up a menu with choices
such as Clear Data, Default Setup, Autoset,
Screen Captures, Save All Waveforms
and Add Screen Text, providing singleclick access to many of your most
commonly performed tasks.
Pinpoint
™
Triggering
The ability to trigger an oscilloscope on
events of interest is paramount in highspeed debug and validation. Whether
you’re trying to find a system error or
need to isolate a section of a complex
signal for further analysis, Tektronix’
Pinpoint triggering provides the solution.
The Pinpoint trigger system uses Silicon
Germanium (SiGe) technology to provide trigger sensitivity of up to 9 GHz
(TDS6000C models), and allows selection
of all trigger types on both A and B trigger
circuits. It can capture glitches down to
100 ps wide with 1 ps
trigger jitter
RMS
Trigger on glitches down to 100 ps wide.
Pinpoint™trigger system provides the
most advanced, highest performance
triggering available.
typical (TDS6000C models). Other trigger
systems offer multiple trigger types only
on a single event (A Event), with delayed
trigger (B Event) selection limited to edge
type triggering and often does not provide
a way to reset the trigger sequence if the
B-event doesn’t occur. But Pinpoint triggering provides the full suite of advance
trigger types on both A and B triggers
with Reset triggering to begin the trigger
sequence again after a specified time,
state or transition so that even events
in the most complex signals can be
captured. Other oscilloscopes typically
offer less than 20 trigger combinations;
Pinpoint triggering offers over 1400
combinations, all at full performance.
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
2
Page 3
Digital Storage Oscilloscope
TDS6000B/C Series
Decode 8b/10b data streams and set up to
4 data words (40 bits) to trigger on.
Protocol Triggering and Decoding
Software (Opt. PTD)
Easily decode 8b/10b and other encoded
serial data streams, and set desired
encoded words for the serial pattern
trigger to capture. 8b/10b decoding on
data rates over 10 GB/s is possible on
the TDS6000C models. Option PTD
recovers the clock signal, identifies the
transitions, decodes the characters and
other protocol data. The TDS6000C
models can trigger on up to four consecutive 10-bit words or specified error
conditions at data rates up to
3.125 Gb/s. And you can see the
captured bit sequences decoded into
their words for convenient analysis.
Measurement System. Enables over 50
parametric measurements in the amplitude,
time and statistical domain.
Unparalleled Analysis
Waveform data analysis can take many
forms. Whether it’s a simple math expression, waveform mask testing, a pass/fail
compliance test, or a custom application
that you develop, the TDS6000B/C
Series offer the industry’s most comprehensive set of analysis and compliance
tools.
Built-in Analysis Tools
Standard tools built into the TDS6000
offer a wide range of analysis capabilities
including Cursors, Measurements, Math
Equation Editor, Serial Data
Communications Mask Testing (with Opt.
SM), and Spectrum Analysis.
8.5 Gb/s TDSRT-Eye™diagram on
the TDS6154C.
Waveform Math Equation Editor. Enables
boundless analysis on waveform data.
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
3
Page 4
Digital Storage Oscilloscope
TDS6000B/C Series
TDSRT-Eye™software. Compliance and analysis
for testing high-speed serial standards. Eye
diagrams at data rates to beyond 6.25 Gb/s.
Software clock recovery to ≥10 Gb/s.
TDSJIT3 v2.0. Jitter analysis for validation
and debug of high-speed digital systems.
Technology-specific Software Solutions
Validation, debug and compliance
testing often require automated test
tools that quickly analyze waveform
data and provide the answers you need.
The TDS6000B/C models provide
complete “turn-key” solutions for the most
demanding technologies. Software
options are also available for performing
validation and compliance measurements
on emerging industry standards.
OpenChoice
®
Analysis
Designing your own custom solution?
The analysis and networking features
of OpenChoice software adds more
flexibility to Tektronix open Windows
XP oscilloscopes:
Fast, PCI bus speed communication between
the data acquisition processor and the
Microsoft Windows desktop
TDSET3 10/100/1000Base-T Compliance.
TDSDVI Digital Visual Interface Compliance.
ActiveX controls to connect the oscilloscope
to popular Windows applications – WITHOUT
leaving the application
PnP drivers to control the scope from
LabVIEW and Lab Windows/CVI running
directly on the oscilloscope, or running
on external PCs
Support for application development
environments includes Visual BASIC,
.NET, C, C++, MATLAB, LabVIEW and
LabWindows/CVI.
Differential Probing with
Performance to >12 GHz
Most high-speed signals today are differential. The P7313 12.5 GHz Differential
Probing System provides true differential
connection to the device under test for
a variety of connection requirements:
solder-in, hand held and fixtured. Typical
system performance with the TDS6154C
exceeds 12.5 GHz bandwidth. The P7313
offers the lowest loading, highest signal
fidelity and lowest cost per connection in the
Access data seamlessly for your own custom
application such as this Microsoft Excel
example.
P7313 12.5 GHz and P7380 8 GHz
Differential Probes.
industry. Versatile, inexpensive Tip-Clip
™
adapters provide the optimum solution
for virtually any connection need.
The P7380SMA Differential Probing System
provides a 50 Ω per side termination
network with a termination voltage. The
termination voltage can be applied externally
or through the TekConnect
®
interface from
the TDS6000B/C probe menu, or not
used at all. A gain switch provides two
different sensitivity settings of the probe
and an Aux Out provides an inverted
version of the signal for driving other
equipment. The P7380SMA probing
system used with the TDS6000B/C
oscilloscopes provides an ideal solution
for validation and compliance testing of
high speed serial data links.
4
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
Page 5
Digital Storage Oscilloscope
TDS6000B/C Series
Characteristics
Vertical System
TDS6604BTDS6804BTDS6124CTDS6154C
Input Channels4
Bandwidth, Rise Time (DSP)N/A8 GHz12 GHz15 GHz
50 ps (10 to 90%) (typical)35 ps (10 to 90%) (typical)28 ps (10 to 90%) (typical)
35 ps (20 to 80%) (typical)24 ps (20 to 80%) (typical)19 ps (20 to 80%) (typical)
True Analog Bandwidth 6 GHz7 GHz12 GHz12 GHz
(–3 dB), Typical Rise Time70 ps (10 to 90%) (typical)62 ps (10 to 90%) (typical)35 ps (10 to 90%) (typical)35 ps (10 to 90%) (typical)
Hardware Bandwidth LimitsFull, 250 MHz or 20 MHz
(Requires TCA-1 MEG Adapter)
Input CouplingDC, GND
Input Impedance50 Ω ±2.5%50 Ω ±2%
Input Sensitivity, 50 ΩFull-scale 50 mV to 10 V (5 mV/div*1to 1 V/div).
Vertical Resolution8-Bit (>11-Bit with averaging)
Max Input Voltage, 50 Ω<1 V
(also determined by <7 V
TekConnect®accessory)
DC Gain Accuracy±(2.5% + (2% x offset))±2%
Position Range±5 divisions
Offset Range
*2
Offset Accuracy±(0.7% x offset + 1.5 mV + 0.1 x V/div setting) ±(0.35% x offset + 1.5 mV + 0.1 x V/div setting)
Channel-to-Channel ≥80:1 at 1.5 GHz≥150:1 at 0 to 10 GHz
Isolation for Any Two ≥15:1 at rated bandwidth>80:1 at 10 GHz to 12 GHz
Channels at Equal ≥50:1 at 12 GHz to 15 GHz
Vertical Scale
53 ps (20 to 80%) (typical)43 ps (20 to 80%) (typical)24 ps (20 to 80%) (typical)24 ps (20 to 80%) (typical)
Full-scale is the peak-to-peak digitizer
range at a given sensitivity.
Volts/div = Full-scale ÷ 10
for <100 mV/div,<1 V
RMS
for ≥100 mV/div settings<5.5 V for ≥1 V full-scale
RMS
for <1 V full-scale,
RMS
Fullscale settings:Fullscale settings:
100 mV to 500 mV: ±0.5 V100 mV: ±0.45 V
505 mV to 995 mV: ±0.25 V200 mV: ±0.4 V
1 V to 5 V: ±5 V500 mV: ±0.25 V
5.05 V to 10 V: ±2.5 V1 V: ±4.5 V
2V:±4.0V
5V:±2.5V
10 V: 0
General Formula for offset range
100 mV to 995 mV: ±|0.5 V – Fullscale/2|
1 V to 10 V: ±|5 V – Fullscale/2|
for ranges <100 mV/divfor ranges <100 mV/div
±(0.8% x offset + 15 mV + 0.1 x V/div setting) ±(0.35% x offset + 15 mV + 0.1 x V/div setting)
for ranges ≥100 mV/divfor ranges ≥100 mV/div
≥
*1
5 mV/div is a software zoom with 7 bits digitizer resolution at 50 mV full scale.
*2
Offset range in addition to ±5 division position range.
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
5
Page 6
Digital Storage Oscilloscope
TDS6000B/C Series
Vertical System (Continued)
Full Scale TDS6000BTDS6154CTDS6124CBOTH
Gain SettingDSP OFFDSP ONDSP ONDSP OFF
Noise, typical100 mV950 µV690 µV570 µV670 µV
160 mV—950 µV840 µV950 µV
200 mV1.6 mV1.1 mV940 µV1.0 mV
300 mV—1.6 mV1.45 mV1.6 mV
400 mV—2.1 mV1.85 mV2.1 mV
500 mV3.55 mV2.5 mV2.3 mV2.4 mV
800 mV—4.4 mV3.8 mV4.1 mV
900 mV—4.8 mV4.3 mV4.6 mV
1 V9.5 mV6.9 mV5.7 mV6.8 mV
2 V16 mV10.5 mV9.5 mV10.0 mV
5 V35.5 mV25 mV23 mV24 mV
10V68mV56mV46mV50mV
Timebase System
TDS6000B ModelsTDS6000C Models
Timebase Range25 ps to 40 s/div; Interpolation down to 500 fs/pt.
Timebase Delay Time Range5 ns to 250 s
Channel-to-Channel Deskew Range±75 ns
Trigger Jitter (RMS)<1.5 ps
(typical)<1 ps
RMS
RMS
(typical)
Long-term Sample Rate and Delay Time Accuracy<2 ppm over any ≥100 ms interval
Clock Stability<1 ppm (typical)
<2.5 ppm (guaranteed)
Jitter Noise Floor420 fs
Delta Time Accuracy[(0.06/sample rate) + 635 fs
(2.5 ppm x [Reading])]
*3Test conditions: Sample mode, at 20 GS/s.
4
Test conditions: Sample mode, 500 mV full scale (50 mV/div) setting, input signal 350 mV with rise time (10% to 90%) <50 ps.
*
(typical) over 10 µs duration or less
RMS
*3
, typical
RMS
over <100 ns duration, typical
RMS
*4
Acquisition System
TDS6000B ModelsTDS6000C Models
Max Sample Rate20 GS/s on 4 channels40 GS/s on 2 channels, 20 GS/s on 4 channels
Equivalent Time Sample Rate (max)2 TS/s
Maximum Record Length 32 Mb (requires Opt. 4M)64 Mb on two channels; 32 Mb on all 4 channels
per Channel(requires option 4M)
Standard2 Mb on all 4 Channels4 Mb on two channels; 2 Mb on all 4 Channels
with Memory Opt. 2M8 Mb on all 4 Channels16 Mb on two channels; 8 Mb on all 4 Channels
with Memory Opt. 3M16 Mb on all 4 Channels32 Mb on two channels; 16 Mb on all 4 Channels
with Memory Opt. 4M32 Mb on all 4 Channels64 Mb on two channels; 32 Mb on all 4 Channels
6
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
Page 7
Digital Storage Oscilloscope
TDS6000B/C Series
Maximum Time Duration Captured at Highest Real-time Resolution (All Channels)
Peak DetectCaptures narrow glitches at all real-time sampling rates
Minimum Peak Detect Pulse Width50 ps
AveragingFrom 2 to 10,000 waveforms included in average
EnvelopeFrom 2 to 2x109waveforms included in min-max envelope
Hi-ResReal-time boxcar averaging reduces random noise and increases resolution
FastFrame™AcquisitionAcquisition memory divided into segments; maximum trigger rate >310,000
waveforms per second. Time of arrival recorded with each event
Roll Mode200 kS/s at 8 Mb Record length and 500 kS/s at
4 Mb Record Length on all channels
Pinpoint™Trigger System
All Models
Sensitivity
Internal DC Coupled0.5 div DC to 50 MHz increasing to 2.5 div at 7 GHz (TDS6000B models,
typical) 0.4 div DC to 50 MHz, increasing to 1 div at 3 GHz,
increasing to 3 div at 9 GHz (TDS6000C models, typical)
External (auxiliary input)250 mV DC to 50 MHz, increasing to 350 mV at
1 GHz, increasing to 500 mV at 2 GHz (typical)
Trigger Characteristics
Main Trigger ModesAuto, Normal and Single
A Event and Delayed B Event Trigger TypesEdge, Glitch, Runt, Width, Transition Time, Timeout, Pattern, State,
Setup/Hold, Window – all except Edge, Pattern and State
can be Logic State qualified by up to two channels
Trigger SequencesMain, Delayed by Time, Delayed by Events, Reset by Time, Reset by State,
Reset by Transition. All sequences can include separate horizontal delay
after the trigger event to position the acquisition window in time
Communications-related Triggers Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications
(requires Option SM)signals up to 3.125 Gb/s. Select among isolated positive or negative one,
zero pulse form or eye patterns as applicable to standard
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
7
Page 8
Digital Storage Oscilloscope
TDS6000B/C Series
Pinpoint™Trigger System (continued)
All Models
Trigger Characteristics (continued)
Serial Pattern Trigger (requires Option ST)64-Bit serial word recognizer, bits specified in binary (high, low, don’t care)
or hex format. Trigger on NRZ-encoded data up to 1.25 Gbaud. TDS6000C
models only: trigger on 8b/10b encoded data up to 3.125 Gbaud (40 bits)
Trigger Holdoff Range250 ns minimum to 12 s maximum
Trigger Types
Edge –
Positive and/or negative slope on any channel or
front panel auxiliary input. Coupling includes DC,
AC, noise reject, HF reject and LF reject.
Glitch –
Trigger on or reject glitches of positive, negative
or either polarity. Minimum glitch width is down
to 100 ps with re-arm time of 250 ps.
Width –
Trigger on width of positive or negative pulse
(down to 100 ps) either within or out of selectable
time limits.
Runt –
Trigger on a pulse that crosses one threshold but
fails to cross a second threshold before crossing
the first again. Optional time qualification.
Timeout –
Trigger on an event which remains high, low or
either for a specified time period, selectable from
360 ps to 1 s.
Transition –
Trigger on pulse edge rates that are faster or
slower than specified. Slope may be positive,
negative or either.
Setup/Hold –
Trigger on violations of both setup time and hold
time between clock and data present on any two
input channels.
Pattern –
Trigger when pattern goes false or stays true for
specified period of time. Pattern (AND, OR, NAND,
NOR) specified for four input channels defined as
HIGH, LOW or Don’t Care.
State –
Any logical pattern of channels (1, 2, 3) clocked
by edge on channel 4. Trigger on rising or falling
clock edge.
Window –
Trigger on an event that enters or exits a window
defined by two user-adjustable thresholds. Event can
be time or logic qualified.
Trigger Delay by Time – 5 ns to 250 seconds.
Trigger Delay by Events – 1 to 10,000,000 events.
Waveform count, Hits in box, Peak hits, Median,
Maximum, Minimum, Peak-to-Peak, Mean (µ),
Standard Deviation (σ), µ + 1σ,µ+ 2σ,µ+ 3σ.
Eye Pattern-related –
Extinction Ratio (absolute, %, and dB), Eye Height,
Eye Top, Eye Base, Eye Width, Crossing %, Jitter
(peak-to-peak, RMS and 6σ), Noise (peak-to-peak
and RMS), S/N ratio, Cycle Distortion, Q-factor.
Waveform Processing/Math
Algebraic Expressions –
Define extensive algebraic expressions including
waveforms, scalars and results of parametric measurements e.g., (Integral (CH.1-Mean(CH.1)) x 1.414).
Arithmetic –
Add, subtract, multiply, divide waveforms
and scalars.
Relational –
Boolean result of comparison >, <, ≥, ≤,=,≠.
Calculus – Integrate, differentiate.
Frequency Domain Functions –
Spectral magnitude and phase, real and
imaginary spectra.
Aux Trigger Input – BNC, see Ext Trigger specification.
Parallel Port – IEEE 1284, DB-25 connector.
Audio Ports –
Miniature phone jacks for stereo microphone input
and stereo line output.
USB2.0 Ports –
Four in back. Allows connection or disconnection
of USB keyboard, mouse, or storage device while
oscilloscope power is on.
Keyboard Port – PS-2 compatible.
Mouse Port – PS-2 compatible.
LAN Port –
RJ-45 connector, supports 10Base-T, 100Base-T
and 1000Base-T.
Serial Port – DB-9 COM1 port.
Windows Video Port –
15-Pin D-sub connector on the rear panel;
connect a second monitor to use dual-monitor
display mode allowing analysis results and plots
to be viewed along with the oscilloscope display.
Video is DDC2B compliant.
GPIB Port – IEEE 488.2 standard.
Scope XGA Video Port –
15-Pin D-sub connector on the rear panel, video
is IBM XGA compatible. Connect to show the oscilloscope display, including live waveforms on an
external monitor or projector.The primary Windows
desktop can also be displayed on an external
monitor using this port.
Power Source
Power –
100 to 240 V
, ±10%, 50/60 Hz; 115 V
RMS
RMS
±10%,
400 Hz; CAT II, <500 W typical (650 VA).
Option SM
156 Standards Masks Supported –
ITU-T (1.544 Mb/s to 155 Mb/s).
ANSI T1.102 (1.544 Mb/s to 155 Mb/s).
Ethernet IEEE Std 802.3, ANSI X3.263 (1.544 Mb/s
to 3.125 Gb/s XAUI).
Sonet/SDH (51.84 Mb/s to 2.4883 Gb/s).
Fibre Channel (133 Mb/s to 4.25 Gb/s*
3
).
InfiniBand (2.5 Gb/s).
USB (12 Mb/s to 480 Mb/s).
Serial ATA (1.5 Gb/s, 3.0 Gb/s).
Serial Attached SCSI (1.5 Gb/s, 3.0 Gb/s).
IEEE 1394b (491.5 Mb/s to 1.966 Gb/s).
RapidI/O (1.25 Gb/s to 3.125 Gb/s).
OIF Standards (2.488 Gb/s to 3.11 Gb/s).
PCI Express (2.5 Gb/s).
Physical Characteristics
BENCHTOP CONFIGURATION
Dimensionsmmin.
Height28211.1
Width45718.0
Depth54023.25
Weightkglb.
Net2147
Shipping32.371.2
RACKMOUNT CONFIGURATION
Dimensionsmmin.
Height26710.5
Width49119.3
Depth63825.1
Weightkglb.
Net2249
Kit5.612.25
*3
4.25 Gb/s mask supported using Glitch Trigger.
Mechanical
Cooling Required Clearancemmin.
Top0 or >760 or >3
Bottom251
Left side763
Right side763
Front00
Rear00
Environmental
Temperature
Operating –
Excluding CD-R/W drive:
TDS6000B models: +5 ºC to +40 ºC.
TDS6000C models: +10 ºC to +45 ºC.
Including CD-R/W drive: +10 ºC to +40 ºC.
Nonoperating – –22 ºC to +60 ºC.
Humidity
Operating –
20% to 80% relative humidity with a maximum wet
bulb temperature of +29 ºC at or below +50 ºC,
noncondensing. Upper limit derated to 25% relative
humidity at +50 ºC.
Nonoperating –
5% to 90% relative humidity with a maximum wet
bulb temperature of +29 ºC at or below +60 ºC,
noncondensing. Upper limit derated to 20% relative
humidity at +60 ºC.
Altitude
Operating – 10,000 ft. (3,048 m).
Nonoperating – 40,000 ft. (12,190 m).
Random Vibration
Operating –
0.000125 g
from 350 to 500 Hz, 0.0000876 g
Overall level of 0.24 g
Nonoperating –
0.0175 g
100 to 200 Hz, 0.00875 g
–3 dB/octave from 350 to 500 Hz, 0.006132 g
at 500 Hz. Overall level of 2.28 g
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
9
Page 10
Digital Storage Oscilloscope
TDS6000B/C Series
Ordering Information
TDS6154C
15 GHz Digital Storage Oscilloscope.
TDS6124C
12 GHz Digital Storage Oscilloscope.
TDS6804B
8 GHz Digital Storage Oscilloscope.
TDS6604B
6 GHz Digital Storage Oscilloscope.
TDS6000B and TDS6000C Models Include:
Accessory pouch, front cover, USB mouse,
USB mini-keyboard, user manual (071-1658-xx),
Documentation CD-ROM (063-3760-xx), product
software CD-ROM (020-2572-xx for TDS6000B
models; 020-2647-xx for TDS6000C models),
operating system restoration CD-ROM (020-2569-xx),
optional applications software CD-ROM (020-2450-xx),
performance verification procedure PDF file, GPIB
programmer’s reference (on product software
CD-ROM), calibration certificate documenting NIST
traceability, 2 540-1 compliance and ISO9000,
power cord, one year warranty. (4) TekConnect
to SMA adapters (TCA-SMA), Deskew Fixture
(067-0484-xx).Please specify power plug
when ordering.
Instrument Options
Power Plug Options
Opt. A0 – North America power.
Opt. A1 – Universal EURO power.
Opt. A2 – United Kingdom power.
Opt. A3 – Australia power.
Opt. A5 – Switzerland power.
Opt. A6 – Japan power.
Opt. A10 – China power.
Opt. A99 – No power cord or AC adapter.
Manual Language Options
Opt. L5 – Japanese manual.
Opt. L7 – Simplified Chinese manual.
Opt. L10 – Russian manual.
Disk Drive Options
Opt. FHD – Front-panel 40 GB removable hard
disk drive, replaces front panel CD-R/W which
is moved to the back.
®
Cables
GPIB Cable (1 m) – Order 012-0991-01.
GPIB Cable (2 m) – Order 012-0991-00.
RS-232 Cable – Order 012-1298-00.
Centronics Cable – Order 012-1214-00.
Mounting Options
1K – K4000 Oscilloscope Cart.
1R – Rackmount Kit.
Service Options
Opt. C3 – Calibration Service 3 Years.
Opt. C5 – Calibration Service 5 Years.
Opt. D1 – Calibration Data Report.
Opt. D3 – Calibration Data Report 3 Years
(requires Option C3).
Opt. D5 – Calibration Data Report 5 Years
(with Option C5).
Opt. R3 – Repair Service 3 Years.
Opt. R5 – Repair Service 5 Years.
Acquisition Memory Options
TDS6000B Models:
2M – 8 M samples on all channels.
3M – 16 M samples on all channels.
4M – 32 M samples on all channels.
TDS6000C Models:
2M – 16 M samples on two channels;
8 M samples on all channels.
3M – 32 M samples on two channels;
16 M samples on all channels.
4M – 64 M samples on two channels;
32 M samples on all channels.
Protocol Trigger and Decode (trigger capability TDS6BUP Opt PTD
for TDS6000C only, requires Opt. ST)
RT-Eye Serial Data Compliance and Analysis SoftwareTDS6BUP Opt. RTE
PCI ExpressTMCompliance Module for Option RTETDS6BUP Opt. PCE (Requires Opt. RTE or RJA or RSA)
InfiniBand® Compliance Module for Option RTETDS6BUP Opt. IBA (Requires Opt. RTE or RJA or RSA)
USB2.0 Compliance, software onlyTDS6BUP Opt. USB (Requires TDSUSBF test fixture)
RT-Eye Serial Data Analysis Software, TDSJIT3 v2.0 TDS6BUP Opt. RJA
Advanced Jitter Analysis, PTD Protocol Trigger
and Decode Software Bundle
RT-Eye Serial Data Analysis, TDSJIT3 v2.0 Advanced TDS6BUP Opt. RSA
Jitter Analysis, Opt. SM Serial Mask, Opt. ST Serial Trigger,
Opt. PTD Protocol Trigger and Decode Software Bundle
Digital Storage Oscilloscopes • www.tektronix.com/oscilloscopes
11
Page 12
Digital Storage Oscilloscope
TDS6000B/C Series
OpenChoice
Provides more choices for your networking
and analysis solutions.
TekConnect
Superior signal fidelity and unparalleled versatility
at your fingertips.
Oscilloscope Software
Turn your general purpose oscilloscope
into a highly specialized analysis tool.
®
OpenChoice is a collection of software libraries, utilities, samples, industry-standard
protocols and interfaces offered with many Tektronix oscilloscopes and logic
analyzers. From 60 MHz to 15 GHz, OpenChoice allows you to communicate with your
oscilloscope or logic analyzer over a network, using numerous connectivity protocols
and physical interfaces, such as GPIB, Ethernet, RS-232 and shared memory.
®
TekConnect interface takes probe intelligence to the next level whether you are measuring
high voltage, current, power, or even micro-volt level signals. The TekConnect interface
ensures superior signal fidelity with useful bandpass up to 18 GHz at the oscilloscope
input, while offering unparalleled versatility.With the TekConnect interface, you can be
assured that maximum signal integrity is maintained to meet your present and future
bandwidth needs.
Tektronix offers exceptional application solutions that instill your oscilloscope with
specific technology or procedure expertise, vastly simplifying the development and
testing of specialized designs. From serial data standards to power measurements,
Tektronix has the broadest selection of oscilloscope software to convert your
oscilloscope into a highly specialized and power analysis tool.
Contact Tektronix:
ASEAN / Australasia (65) 6356 3900
Austria +41 52 675 3777
Balkan, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 07 81 60166
Brazil & South America 55 (11) 3741-8360
Canada 1 (800) 661-5625
Central East Europe, Ukraine and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark +45 80 88 1401
Finland +41 52 675 3777
France +33 (0) 1 69 86 81 81
Germany +49 (221) 94 77 400
Hong Kong (852) 2585-6688
India (91) 80-22275577
Italy +39 (02) 25086 1
Japan 81 (3) 6714-3010
Luxembourg +44 (0) 1344 392400
Mexico, Central America & Caribbean 52 (55) 5424700
Middle East, Asia and North Africa +41 52 675 3777
The Netherlands 090 02 021797
Norway 800 16098
People’s Republic of China 86 (10) 6235 1230
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 82 (2) 528-5299
Russia & CIS +7 (495) 7484900
South Africa +27 11 254 8360
Spain (+34) 901 988 054
Sweden 020 08 80371
Switzerland +41 52 675 3777
Taiwan 886 (2) 2722-9622
United Kingdom & Eire +44 (0) 1344 392400
USA 1 (800) 426-2200
For other areas contact Tektronix, Inc. at: 1 (503) 627-7111
Updated 12 May 2006
Our most up-to-date product information is available at: