SE-1000 SERIES SPECIFICATIONS CASCADE CHAMBERS
Interior Dimensions — W x D x H
Inches / Centimeters All Models 40 x 39.25 x 38.25 / 102 x 100 x 97
Volume — Cubic Ft / Liters All Models 34.8 / 986
Exterior Dimensions — W x D x H
Inches / Centimeters All Models 49 x 83 x 87 / 125 x 211 x 221
Temperature Range All Models 180ºC to -70ºC (356ºF to -94ºF)
Temperature Control All Models ±0.3ºC (±0.5ºF)
Temperature Uniformity
1
All Models ±0.5ºC (±0.9ºF)
Airflow All Models 1,000 cfm
Window Size — W x H
Inches / Centimeters All Models 15 x 19 / 38 x 48
Compressor Size — Hp Two 3.5 Two 6 Two 10 Two 15
Noise Level
2
— dBA
Heating / Cooling 60 / 74 60 / 76 60 / 76 60 / 76
Approximate
Shipping Weight — Lbs / Kg 1,840 / 834 1,900 / 862 2,340 / 1,064 2,415 / 1,098
Cooling Performance Capacity — Minutes
Air Temperature (Empty Chamber)
180ºC to -65ºC (356ºF to -85ºF) 78 56 34 27
71ºC to -65ºC (160ºF to -85ºF) 50 36 21 16
85ºC to -40ºC (185ºF to -40ºF) 35 26 13 11
Cooling Product Temperature — Minutes
50Lb / 23Kg Aluminum
71ºC to -65ºC (160ºF to -85ºF) 68 53 32 28
85ºC to -40ºC (185ºF to -40ºF) 53 46 26 24
Heating Performance Capacity — Minutes
Air Temperature (Empty Chamber)
-65ºC to 180ºC (-85ºF to 356ºF) 43 43 21 12
-65ºC to 71ºC (-85ºF to 160ºF) 20 20 10
-40ºC to 85ºC (-40ºF to 185ºF) 19 19 9 7
Heating Product Temperature — Minutes
50Lb / 23Kg Aluminum
-65ºC to 71ºC (-85ºF to 160ºF) 44 44 27 25
-40ºC to 85ºC (-40ºF to 185ºF) 41 41 26 24
Live Load Capacity — Watts, Temp. Mode
0ºC (32ºF) 2500 3000 3000 3000
-54ºC (-65.2ºF) 1500 2000 2000 2000
Electrical Service — No Humidity / Humidity
208 / 3 / 60 (Full Load Amps) 54 / 58 71 / 71 92 / 92 150 / —
230 / 3 / 60 (Full Load Amps) 51 / 57 65 / 65 83 / 83 136 / —
460 / 3 / 60 (Full Load Amps) 26 / 29
32
/ 32
42
/ 42
68/—
400 / 3 / 50 (Running Load Amps) 24 / 27 33 / 33 45 / 45 70 / —
Inlet Water — Gal / Liters per Minute
29ºC / 85ºF Air-Cooled Air-Cooled 16 / 60 24 / 91
24ºC / 75ºF Air
-Cooled Air-Cooled
9 / 34 14 / 53
18ºC / 65ºF Air-Cooled Air-Cooled 7.5 / 28 11 / 42
13ºC / 55ºF Air
-Cooled Air-Cooled
6 / 23 8 / 31
Model SE-1000-3-3 SE-1000-6-6 SE-1000-10-10 SE-1000-15-15
SE-1000 SERIES SPECIFICATIONS SINGLE STAGE CHAMBERS
Interior Dimensions — W x D x H
Inches / Centimeters All Models 40 x 39.25 x 38.25 / 102 x 100 x 97
Volume — Cubic Ft / Liters All Models 34.8 / 986
Exterior Dimensions — W x D x H
Inches / Centimeters All Models 49 x 83 x 87 / 125 x 211 x 221
Temperature Range All Models 180ºC to -40ºC (356ºF to -40ºF)
Temperature Control All Models ±0.3ºC (±0.5ºF)
Temperature Uniformity
1
All Models ±0.5ºC (±0.9ºF)
Airflow All Models 1,000 cfm
Window Size — W x H
Inches / Centimeters All Models 15 x 19 / 38 x 48
Compressor Size — Hp 4 6 10 15
Noise Level
2
— dBA
Heating / Cooling 60 / 74 60 / 76 60 / 76 60 / 76
Approximate
Shipping Weight — Lbs / Kg 1,660 / 753 1,860 / 844 2,065 / 937 2,115 / 959
Cooling Performance Capacity — Minutes
Air T
emperature (Empty Chamber)
180ºC to -35ºC (356ºF to -31ºF) 74 52
37 34
71ºC to -35ºC (160ºF to -31ºF) 46
33 22 21
85ºC to -20ºC (185ºF to -4ºF) 32 22 12 10
Cooling Product Temperature — Minutes
50Lb / 23Kg Aluminum
71ºC to -35ºC (160ºF to -31ºF) 66 49 36 35
85ºC to -20ºC (185ºF to -4ºF) 48 39 29 28
Heating Performance Capacity — Minutes
Air Temperature (Empty Chamber)
-35ºC to 180ºC (-31ºF to 356ºF) 36 36 20 10
-35ºC to 71ºC (-31ºF to 160ºF) 15 15 8 5
-20ºC to 85ºC (-4ºF to 185ºF) 15 15 8 5
Heating Product T
emperature — Minutes
50Lb / 23Kg Aluminum
-35ºC to 71ºC (-31ºF to 160ºF) 33 33 25 23
-20ºC to 85ºC (-4ºF to 185ºF) 33
33 25 23
Live Load Capacity — Watts, Temp. Mode
0ºC (32ºF) 2500 3000 3000 3000
-20ºC (-4ºF) 1500 2000
2000 2000
Electrical Service — No Humidity / Humidity
208 / 3 / 60 (Full Load Amps) 39 / 55 45 / 61 60 / 72 95 / 106
230 / 3 / 60 (Full Load Amps) 37 / 55 42 / 60 55 / 68 86 / 99
460 / 3 / 60 (Full Load Amps) 19 / 28
21
/ 30
27
/ 33
43
/ 50
400 / 3 / 50 (Running Load Amps) 19 / 28 24 / 32 40 / 40 50 / 55
Inlet Water — Gal / Liters per Minute
29ºC / 85ºF
Air-Cooled Air-Cooled
16 / 60
24
/ 91
24ºC / 75ºF Air-Cooled Air-Cooled 9 / 34 14 / 53
18ºC / 65ºF Air-Cooled Air-Cooled 7.5 / 28 11 / 42
13ºC / 55ºF
Air-Cooled Air-Cooled
6 / 23
8
/ 31
Model SE-1000-4 SE-1000-5 SE-1000-10 SE-1000-15
THERMOTRON INDUSTRIES
291 Kollen Park Drive
Holland, Michigan, USA 49423
Mktg: (616) 393-4580
Main: (616) 392-1491
Fax: (616) 392-5643
E-mail: info@thermotron.com
THERMOTRON INDUSTRIES, U.K.
Newton House
Winch Road
Kent Science Park
Sittingbourne, Kent
ME9 8EF England
Phone: 01795 436333
Fax: 01795 436777
Email: sales@thermotron.co.uk
Visit us on the Internet
www.thermotron.com
FORM NO. T04-005 09/04 (REV. 04/06) ©THERMOTRON INDUSTRIES PRINTED IN USA
UNITED STATES PATENT NO. 6,976,340 and 5,631,429
A full range, modular humidity system can be added to SE-Series Test Chambers
either at the time of purchase or as a field retrofit. Performance specifications for
the humidity module are characterized by a wide range, precise uniformity and
tight control.
1 Standard deviation from mean, measured at -25°C (-13°F) or at +100°C (+212°F).
2 Noise Level: A-weighted sound pressure level measured at a distance of 1.0 meter
(39.4 inches) from the surface of the equipment at a height of 1.6 meters
(63 inches) from the floor in free-field conditions, using a calibrated instrument.
3
Limited by a +7°C (+45°F) minimum dew point temperature and
a maximum dry bulb temperature of +88°C (+190°F).
4
At a dry bulb temperature above +20°C (+68°F).
5 Based upon temperature uniformity specifications.
Relative humidity indication at or near the physical limits may be af
fected by sensor
accuracy and control tolerance. An optional humidity package can be added for
applications requiring humidity levels lower than those covered by the full range
humidity system.
FULL RANGE HUMIDITY SPECIFICATIONS*
Humidity Range
3
10% to 98% RH
Dry Bulb Temperature Range 7°C to 88°C (45°F to 190°F)
Dewpoint Temperature Range 7°C to 87°C (45°F to 188°F)
Extended Dewpoint Condition -10°C (14°F)
Humidity Control
4
±2.5% RH
Humidity Uniformity
5
±1.0% RH
* Except 15hp Cascade Chamber models
STANDARD HUMIDITY RANGE
-10°C (14°F) DEWPOINT
Performance is based upon 60 Hz and laboratory ambient conditions of 23.9°C (75°F), and may vary slightly on 50 Hz power or at other
temperature and humidity levels. Chambers are designed for use under normal laboratory operating conditions. For other applications, please
consult Thermotron.
It is Thermotron’s understanding that the unit will be used in a non-hazardous environment. The unit is not designed for use with or for the
purpose of processing hazardous materials. If hazardous materials are involved, please consult the factory for an alternative quote for a properly
designed unit.
The addition of accessories may impact performance.
Specifications subject to change without notice.
1000 SERIES HUMIDITY RANGE
100
80
60
40
20
0
0 20 40 60 80 100°C
32 68 104 140 176 212°F
PERCENT RH