H-FIELD PROBE HF3061
Measuring magnetic fields from 300 kHz to 30 MHz
using instruments in the NBM-500 family
broadcasting, telecoms, and industrial equipment
Isotropic (non-directional) measurement 62 dB dynamic range without changing
measurement range
The probe contains three orthogonally arranged coils with detector diodes. The three voltages, corresponding to the spatial components, are available individually at the probe output. The NBM basic unit calculates the resulting isotropic field strength.
APPLICATIONS
The probe detects magnetic fields from 300 kHz to 30 MHz, such as those caused by short and medium wave transmitters, many RF communications services, and industrial equipment. The dynamic range from 0.012 A/m to 16 A/m (62 dB) makes it ideal for measuring field exposure in both the public and the occupational environment.
PROPERTIES
The probe has mechanical and electrical properties that are ideal for field use. The sensors are effectively protected by impact resistant plastic. The electric destruction limit is above 35 A/m for continuous wave signals, which is several times more than any of the human safety limit values.
CALIBRATION
The probe is calibrated at several frequencies. The correction values are stored in an EPROM in the probe and are automatically taken into account by the NBM instrument. Calibrated accuracy is thus obtained regardless of the combination of probe and instrument.
NSTS 0510-E0237C 1 / 2 Subject to change
SPECIFICATIONS
a
Probe HF3061 Magnetic (H-)Field
Frequency range Type of frequency response Flat
Measurement range
Dynamic range 62 dB CW damage level 35 A/m 46 W/cm² Peak damage level Sensor type Diode based system Directivity Isotropic (Tri-axial) Readout mode / spatial assessment 3 separate axes
UNCERTAINTY
Flatness of frequency response
Calibration uncertainty not included
Calibration uncertainty
@ 0.59 mW/cm² (0.125 A/m)
Linearity
Referred to 0.59 mW/cm² (0.125 A/m)
Isotropic response Temperature response +0.2/ -0.8 dB (±0.025 dB/K @ 10 to 50 °C )
GENERAL SPECIFICATIONS
Calibration frequencies
Recommended calibration interval 24 months Temperature range
Operating Non-operating (transport)
Humidity 5 to 95 % RH @ 28 °C 26 g/m³ absolute humidity Size 300 mm x 120 mm Ø Weight 190 g Compatibility NBM-500 series meters Country of origin Germany
(a) Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 25% to 75%, sinusoidal signal
Cutoff frequency at approx. -3 dB
(b) (c) Pulse length 1µsec, duty cycle 1:100 (d)
Frequency response can be compensated for by the use of correction factors stored in the probe memory
(e) Accuracy of the fields generated to calibrate the probes
Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe
(f)
(b)
(c)
(d)
300 kHz to 30 MHz
0.012 to 16 A/m (CW)
0.012 to 0.7 A/m (True RMS)
5.4 µW/cm² to 10 W/cm² (CW)
5.4 µW/cm² to 18 mW/cm² (True RMS)
350 A/m 4.6 kW/cm²
0/-1 dB (500 to 800 kHz) +0.1/ -0.5 dB (800 kHz to 30 MHz)
(e)
(f)
±0.6 dB
±3 dB (0.017 to 0.033 A/m) ±1 dB (0.033 to 0.068 A/m) ±0.5 dB (0.068 to 3 A/m) ±1 dB (3 to 16 A/m)
±1 dB
±3 dB (10 to 40 µW/cm²) ±1 dB (40 to 175 µW/cm²) ±0.5 dB (175 µW/cm² to 340 mW/cm²) ±1 dB (0.34 to 10 W/cm²)
0.1/ 0.15/ 0.2/ 0.3/ 0.4/ 0.5/ 0.6/ 0.7/ 0.8/ 0.9 MHz 1/ 1.2/ 1.5/ 2/ 3/ 4/ 5/ 10/ 15/ 20/ 25/ 27.12/ 30 MHz
0 °C to +50 °C
-40 °C to +70 °C
ORDERING INFORMATION
Probe HF3061, H-field for NBM, 300 kHz - 30 MHz, isotropic 2402/05
Part number
Narda Safety Test Solutions GmbH
Sandwiesenstrasse 7 72793 Pfullingen, Germany Phone: +49 (0) 7121-97 32-777 Fax: +49 (0) 7121-97 32-790 E-Mail: support@narda-sts.de www.narda-sts.de
® Names and Logo are registered trademarks of Narda Safety Test Solutions GmbH and L3 Communications Holdings, Inc. - Trade names are trademarks of the owners.
NSTS 0510-E0237C 2 / 2 Subject to change
Narda Safety Test Solutions
435 Moreland Road Hauppauge, NY 11788, USA Phone: +1 631 231-1700 Fax: +1 631 231-1711 E-Mail: NardaSTS@L-3COM.com www.narda-sts.com
Narda Safety Test Solutions Srl
Via Leonardo da Vinci, 21/23 20090 Segrate (Milano), Italy Phone: +39 02 2699871 Fax: +39 02 26998700 E-mail: support@narda-sts.it www.narda-sts.it