AC Output Voltage: Range: 50V to 5000V AC, 1V resolution
Frequency: 50 or 60 Hz Programmable 
Waveform: Sinusoidal 
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 0.001 to 40mAAC 
Resolution: 1 or 10µA steps 
Accuracy: ±(1% + 5cnt)
High/Low Limit Test: 1µA to 40mAAC
Accuracy: ±(1% of limit + 1mA) 
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, >1mA
DC Output V
oltage: Range: 50V to 6000V DC, 1V resolution
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: ±(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range: 0.1µ to 20mA DC
Resolution: 0.1, 1 or 10µA steps 
Accuracy: ±(1% or reading + 5cnt)
High/Low Limit Test: 0.0001mA to 20mA DC
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, >1mA
Insulation Resist
ance:
Range: 100kΩ - 50GΩ 
Accuracy: ±5% to ±15% depending upon 
voltage and resistance 
Voltage Range:  50V to 1000V DC 
Volt age Accuracy: ±(1% of setting + 5V) 
High/Low Limit Test: 100kΩ - 50GΩ
Low limit can be turned OFF
IR Test Delay: 0.3 to 99.9 seconds Programmable in
0.1 second steps
Ground Bond:
Output Current: Range: 1.0 to 30.0AAC, setting 0.1A/step
Accuracy: ±(1% of setting + 0.3A)
Display: ±(1% of reading + 3 counts) 
Frequency:  50 or 60Hz Selectable  
No Load Voltage:  6 to 15 V Programmable 
Resistance: Range: 0 - 500.0mΩ, 4 digits
Accuracy: ±(1% of reading + 3 counts) 
Resolution: 1mΩ
Hi Limit: 10mV to 510mV
Offset Function: 0 to 100mV offset, user selectable 
Test Time: 0.5 - 999sec (±20ms)
Common Features:
AC/DC Test Time: Ramp: 0.1 to 999s (±20ms)
Test: 0.1 to 999s (±20ms) and Continuous
Remote Control: Inputs: Start,Stop 
Characteristics: Optically Isolated 
with Low, Pulse Width >1ms. 
Outputs: Pass/Fail/Under Test 
Characteristics: Dry Contact relay 
Electrical Characteristics: 120V 100mAmax. 
Logic: Closed if True 
Connector: Terminal Strip and 9 pin D Series
Test Setups: 99 Test Setups with 99 Steps each, Alpha-
Numeric Label
Connectors: Front and Rear Connections 
Front Panel Lockout: Password 
Safety Features: Fast Cutoff (<0.4ms) and Fast Discharge
Adjustable Discharge: .05-5.1kV DC
Miscellanious: Fail Retest
Scanner Delay: 0.1 to 99.9, 0.1s/step 
PAUSE Mode: Program pause between steps
Indication:  Pass/fail lights, audible sound 
Buzzer Level: 1,2,3 and Off  
Standard Interface: IEEE-488  
Optional Interfaces: RS-232, Printer
Data bits: 8, Parity: None 
Stop bits: 1, Default Baud Rate: 9.6k 
EOS: CR + LF, Echo: Off 
Selectable Baud Rate: 300 - 19200
Dimensions: (w x h x d):17x6.8x17.7in (430x175x450mm) 
Weight:  53 lbs (24kg) - Net, 60 lbs (27kg) Shipping 
Environmental: Operating: 0 to + 40° C,
Humidity: <75% 
Storage: - 20 to + 70° C 
Warm-up Time: 1minute
Power: • 90 - 130V AC    • 50 or 60Hz
• 200 - 250V AC   • 500W max
Guardian 6000 Electrical Safety Analyzer
Includes:
150354 Instruction Manual 
S02 HV Lead Set, 1m 
G15 Ground Continuity Lead Set 
4200-0300 AC Power Cable 
N/A Calibration Cert. Traceable to NIST
Optional Accessories
N/A Calibration Data 
6000-01 Scanner, 8 Channel 5HV,3GC 
6000-02 Scanner, 8 Channel 3HV,5GC 
6000-03 Scanner, 8 Channel HV 
S04 HV Lead Set 2m 
S05 Foot Switch 
S08 Gun Probe 
S09 HV Lead, 1M, Unterminated 
S10 HV Lead, 1M, Unterminated
G13 Corded Product Adapter (115V) 
G14 Power Entry Adapter 
G16 International Power Strip 
G24 Scanner Cable (5000 scanners) 
G25 Corded Product Adapter (240V) 
G26 RS232 Interface 
G27 Rack Mount Flanges 
G28 Printer Interface (replace IEEE) 
G41 RS232 Cable: DB9F to DB25M, 3 feet
Ordering Information
For more detailed specifications, visit www.quadtech.com • For more information about special purchase, 
rent & lease options, call: 
1-800-253-1230 • Fax 1-978-461-4295 • Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard, MA 01754
P/N: 030081/A6
Guardian 6000
Product Specifications