Atec Global-N User Manual

Temperature Cycling Chambers
Global-N Series
Faster test chambers ...
The Global-N series chambers from ESPEC provide the temperature cycling and humidity performance you need to validate quality and reliability for increasingly sophisticated electronics and other products.
The 800-liter (28 cu. ft.) model ts in a space
1.3m by 2.2m (51 by 85 inches), except -15CW models.
2
The 380-liter (12 cu. ft.) model’s width is just 0.9 meters (35 inches), tting thru a standard doorway.
Global Features
World-standard performance and features on a minimal footprint
The Global-N chambers have temperature change rates from 5 to 20°C per minute. In addtion, controlled humidity from 10 to 95% is an available feature. There are two sizes, with interior volumes of 380 and 800 liters (12 and 28 cubic feet). They are compact, requiring the least amount of oor space for this level of performance.
These chambers feature the classic ESPEC look of stain­less steel, with a new, modern style. The control console is now on the door, making the footprint of the unit even more compact. A simple, but secure, door latch is easier than ever to use.
International applications
Global-N chambers meet Mil-Std, JEDEC, IEC, and other international test performance standards. The units can be installed around the world for consistent testing at different facilities, supported by your local ESPEC service group.
Designed for serviceability and safety
For safety and compliance, Global-N series are UL 508A certied or CE-marked.
Refrigeration service panels are hinged for fast access. Analog refrigeration gauges help you monitor system perfor­mance and pinpoint service needs.
All models feature three levels of overheat protection, plus two levels of overcool protection. The system controller pro­vides alarm messages with specic troubleshooting help to quickly resume testing.
Standard features:
Stainless steel exterior and interior•
Foam-insulated door for rigidity and long life•
High performance refrigeration utilizing reliable Scroll • compressors (except -15CW models)
Hinged service panels for easy access•
Unique non-metalic thermal breaks around the doorframe • and cable ports
Specimen power relay for interlocking test samples or • external devices to chamber power for safety
One 100mm (4") cable port, one shelf & casters included•
Innovative high-speed airow suited for fast temperature cycling applications.
Advanced refrigeration design is especially compact for high performance test chambers.
3
Performance
1209030 60 150
180
Cooling Rate
EGNZ12-7.5CWL: 15°C/m EGNZ12-6CWL: 11°C/m EGNZ12-4CWL: 5.5°C/m
Heating Rate
EGNZ12-7.5CWL: 15°C/m EGNZ12-6CWL: 15°C/m EGNZ12-4CWL: 7°C/m
-70
0
180
50
100
150
-50
Time, minutes
Air Temperature, °C
Temperature cycling up to 20 degrees per minute
Global-N’s twenty different models allows you to choose the size and performance best suited for your test applications. Temperature cycling rates from 5 to 20°C/min. are possible. Two different ultimate low temperatures are available: -40°C or
-70°C.
Models EGNL & EGNX also control humidity, for expanded testing options beyond temperature cycling.
Global-N chambers are capable of meeting the performance requirements of JEDEC, IEC, Mil­Std, and other international environmental test performance standards.
Get a performance evaluation to help select your model
For better assurance of performance for your temperature cycling application, ESPEC can pro­vide a guaranteed performance calculation based on your submitted requirements.
Please use these questions (at right) as a guide to dene your test plan. You can then give this information to your local sales representative, or submit at www.espec.com, for review by an ESPEC engineer. A recommended model with appropriate refrigeration size will be returned.
SYSTEM & PERFORMANCE OPTIONS
Low and Ultra-low Humidity - Extends • the humidity control range for EGNL & EGNX models.
Attached Air-cooled Condenser - For • applications where water-cooling isn’t practical. The condenser is on top of the chamber, adding 350mm (14 inches). Suitable for installation in sites that are not air conditioned.
Product Temperature Control - See • next page for details.
4
Different Global-N models can heat and cool at different rates. Shown above are change rates for EGNZ12 models, following IEC 60068 3-5 standard, with the sensor in the supply air.
Questions for performance evaluation request:
1. Chamber type
Desired size: 12 or 28 cu. ft.?
Local power: 50Hz or 60Hz?
Humidity control: Yes or No?
Sample description and type of material(s)
Total mass per test, including any racks or cabling
Heat output of samples, in watts, if powered during test
3. Your Test Plan
Test method, if a published standard (e.g. JEDEC JESD22-A104C)
Start temperature for cycling
End temperature for cycling
Ramp rate or time allowed between start and end temperatures
Datalogging Choices - Via traditional • chart recorders, paperless recorders, Ethernet, or computer interface.
ECA Modication - Allows condition-• ing of remotely-located equipment that cannot be placed inside the chamber.
Dry Air Purge - Keeps moisture in the • chamber extremely low by purging with -40° dewpoint air. Compressed air supply required.
Liquid Nitrogen Injection - Boosts • cooling rate.
2. Your Sample
Other options:
Viewing window with LED lighting • (maximum 150°C operation)
Inner glass door with operation ports•
Additional shelves, 45 kg (100 lb.) • loading
Cable ports with exible silicone plug, • 50, 100, or 150mm diameters
Humidity water lter•
Emergency stop button•
Advanced Control
0
50
-60
-20
0
20
40
60
80
-40
Time, minutes
Part Temperature, °C
Time Saved: 34 minutes
60
403010 20
No Boost 20°C Boost
Access to the SCP-220 programmer is on the new console mounted on the left side of the door.
Chamber Network
Web
Controller
LAN/
Intranet
Compatible
with any
internet-ready
device
Touch-screen controller enhances per­formance
Imagine walking up to your test chamber and im­mediately being able to get your test running. The exclusive SCP-220 programmer makes it a breeze to set up, run, and monitor your chamber, even for the infrequent user.
For temperature cycling, the controller allows you to select step-ramping for the fastest possible transis­tion, or controlled-ramping to achieve a desired ramp rate. Guaranteed soak settings allow you to control when timed soak periods begin.
Remote operation via Ethernet
ESPEC offers the Web Controller, an embedded server/software solution that allows monitoring and programming your chamber via any PC on your network.
No local software to install, just a web browser and access to the LAN is required. Because there are only seven screens, all operational functions are easy to nd and use. See our separate brochure or website for more detail.
Remote control and datalogging is also possible via RS-232 or IEEE-488 options.
The Web Controller is compatible with all USA-built ESPEC models with touch-screen programmer/controllers, except thermal shock.
Product temperature control generates faster ramp-rates for test samples, as well as signicant time savings for soak periods.
Ethernet
Optional product temperature control
Monitors product temperature• Enables faster product change rates• Shortens testing time •
During cycling tests, product temperature can lag behind air temperatures by 20 degrees or more. Our optional product temperature control feature lets you drive faster change rates by directly monitoring product temperature and overshooting normal air­temperature setpoints until the product approaches the desired temperature. By achieving product temperature faster, soak periods can begin earlier, saving signicant time.
As shown in the example at left, setting the cham­ber for a 20°C air temperature boost (air at -60°C) achieved the -40°C product temperature in just 16 minutes, compared for nearly 50 minutes with the chamber air set at -40°C.
5
SPECIFICATION - 380 L / 12 C u . F T . MODELS (460V)
20 40 60 80 100
100
80
60
40
20
0
Temp in °C
% RH
10° /15%
RANGE -70 TO 180°C
EGNX12-4CWL
Temp/Humidity
EGNX12-4CAL
EGNZ12-4CWL
Temp Only
Temperature Range -70 to 180°C (-94 to 354°F)
Cooling Capacity 2.5 kW at -50°C 3.6 kW at -50°C 4 kW at -50°C
Refrigeration System Cascade 4 hp Scroll Cascade 6 hp Scroll Cascade 7.5 hp Scroll
Heating System 5 kW 8 kW 8 kW
Humidity Range 10 to 95% RH per chart (EGNX models only)
Measured Airow 850 cubic meters per hour (500 CFM)
EGNZ12-4CAL
Performance Example (-40°C to 85°C with 50 lbs. Aluminum)
Heating Rate 7°C/m average 15.5°C/m average 15.5°C/m average
Cooling Rate 5.5°C/m average 11°C/m average 15.5°C/m average
Dimensions
Workspace Volume 380 L (12 cu. ft.)
Workspace (WxDxH) 600 x 743 x 850 mm (23.6" x 29.25" x 33.5")
Exterior (WxDxH)
889 x 2103 x 1768 mm (35" x 83" x 70")
(‘CAL’ suffix models are 36 cm (14") taller)
Site Requirements
Power Supply 460V 3Ø 60Hz (380-415V 3Ø 50Hz also available)
Humidity Water De-ionized water, 0.2 to 10 µS/cm resistivity (EGNX models only)
Cooling Water Maximum
with 30°C Inlet
Heat to Room
34 LPM (9 GPM)
(CWL suffix models)
40,000 BTU/hr
(CAL suffix models)
EGNX12-6CWL
EGNX12-6CAL EGNX12-7.5CWL
EGNZ12-6CWL
EGNZ12-6CAL EGNZ12-7.5CWL
53 LPM (14 GPM)
(CWL suffix models)
60,000 BTU/hr
(CAL suffix models)
72 LPM (19 GPM)
EGNX12-4CWL (this page)
EGNX28-6CWL (next page)
RANGE -40 TO 180°C
Temp/Humidity
Temperature Range -40 to 180°C (-40 to 354°F)
Cooling Capacity 3.3 kW at -20°C 4.3 kW at -20°C
Refrigeration System Single-stage 4 hp Scroll Single-stage 6 hp Scroll
Heating System 8 kW 8 kW
Humidity Range 10 to 95% RH per chart
Measured Airow 850 cubic meters per hour (500 CFM)
Performance Example (0°C to 80°C with 50 lbs. Aluminum)
Heating Rate 10°C/m average 15°C/m average
Cooling Rate 10°C/m average 15°C/m average
Workspace Volume 380 L (12 cu. ft.)
Workspace (WxDxH) 600 x 743 x 850 mm (23.6" x 29.25" x 33.4")
Exterior (WxDxH)
Power Supply 460V 3Ø 60Hz (380-415V 3Ø 50Hz also available)
Humidity Water De-ionized water, 0.2 to 10 µS/cm resistivity
Cooling Water Maximum
with 30°C Inlet
Heat to Room
6
EGNL12-4CWL
EGNL12-4CAL
EGNL12-6CWL
EGNL12-6CAL
Dimensions
889 x 2103 x 1768 mm (35" x 83" x 70")
(‘CAL’ suffix models are 36 cm (14") taller)
Site Requirements
35 LPM (9 GPM)
(CWL suffix models)
52,000 BTU/hr
(CAL suffix models)
55 LPM (14 GPM)
(CWL suffix models)
75,000 BTU/hr
(CAL suffix models)
Humidity Range (EGNX, EGNL)
Blue = standard humidity range Red = optional low-humidity range Green = optional ultra-low humidity range
NOTE:
Please submit a performance evaluation request (see page 4) to ensure selection of the correct model for your test requirement.
Site requirements also include a drain for condensate water from the chamber, and ample spacing for service and proper air cooling.
These specications are for reference only. Detailed and current specications can be obtained from your sales representative or at espec.com.
SPECIFICATION - 800 L / 28 C u . F T . MODELS (460V)
RANGE -70 TO 180°C
EGNX28-4CWL
Temp/Humidity
EGNX28-4CAL
EGNZ28-4CWL
Temp Only
Temperature Range -70 to 180°C (-94 to 354°F)
Cooling Capacity 2.5 kW at -50°C 3.6 kW at -50°C 5 kW at -50°C 9.4 kW at -50°C
Refrigeration System Cascade 4 hp Scroll Cascade 6 hp Scroll Cascade 12 hp Scroll Cascade 15 hp Discus
Heating System 10 kW 10 kW 15 kW 20 kW
Humidity Range 10 to 95% RH per chart (EGNX models only)
Measured Airow 1,350 cubic meters per hour (800 CFM)
EGNZ28-4CAL
Performance Example (-40°C to 85°C with 50 lbs. Aluminum)
Heating Rate 10°C/m average 10°C/m average 13.5°C/m average 20°C/m average
Cooling Rate 3.1°C/m average 5.5°C/m average 13.5°C/m average 20°C/m average
Workspace Volume 800 L (28 cu. ft.)
Workspace (WxDxH) 1000 x 800 x 1000 mm (39.4" x 31.5" x 39.4")
Exterior (WxDxH)
Power Supply 460V 3Ø 60Hz (380-415V 3Ø 50Hz also available)
Humidity Water De-ionized water, 0.2 to 10 µS/cm resistivity (EGNX models only)
Cooling Water Maximum
with 30°C Inlet
Heat to Room
34 LPM (9 GPM)
(CWL suffix models)
40,000 BTU/hr
(CAL suffix models)
1290 x 2159 x 1920 mm (51" x 85" x 76") (CAL suffix models are 36 cm (14") taller)
EGNX28-6CWL
EGNX28-6CAL EGNX28-12CWL EGNX28-15CW
EGNZ28-6CWL
EGNZ28-6CAL EGNZ28-12CWL EGNZ28-15CW
Dimensions
1290 x 3074 x 2014 mm
(51" x 121" x 79.5")
Site Requirements
53 LPM (14 GPM)
(CWL suffix models)
60,000 BTU/hr
(CAL suffix models)
102 LPM (27 GPM) 121 LPM (32 GPM)
RANGE -40 TO 180°C
EGNL28-4CWL
Temp/Humidity
Temperature Range -40 to 180°C (-40 to 354°F)
Cooling Capacity 3.3 kW at -20°C 4.3 kW at -20°C 5.2 kW at -20°C 10.5 kW at -20°C
Refrigeration System Single-stage 4 hp Scroll Single-stage 6 hp Scroll Single-stage 7.5 hp Scroll Single-stage 12 hp Scroll
Heating System 10 kW 10 kW 15 kW 15 kW
Humidity Range 10 to 95% RH per chart
Measured Airow 1,350 cubic meters per hour (800 CFM)
EGNL28-4CAL
Performance Example (0°C to 80°C with 50 lbs. Aluminum)
Heating Rate 10°C/m average 10°C/m average 15°C/m average 20°C/m average
Cooling Rate 5°C/m average 10°C/m average 15°C/m average 20°C/m average
Workspace Volume 800 L (28 cu. ft.)
Workspace (WxDxH) 1000 x 800 x 1000 mm (39.4" x 31.5" x 39.4")
Exterior (WxDxH)
Power Supply 460V 3Ø 60Hz (380-415V 3Ø 50Hz also available)
Humidity Water De-ionized water, 0.2 to 10 µS/cm resistivity
Cooling Water Maximum
with 30°C Inlet
Heat to Room
35 LPM (9 GPM)
(CWL suffix models)
52,000 BTU/hr
(CAL suffix models)
EGNL28-6CWL
EGNL28-6CAL EGNL28-7.5CWL EGNL28-12CWL
Dimensions
1290 x 2159 x 1920 mm (51" x 85" x 76") (CAL suffix models are 36 cm (14") taller)
Site Requirements
55 LPM (14 GPM)
(CWL suffix models)
75,000 BTU/hr
(CAL suffix models)
79 LPM (19 GPM) 106 LPM (27 GPM)
7
www.espec.com
4141 Central Parkway, Hudsonville, MI 49426, U.S.A. Tel: 1-616-896-6100
ESPEC EUROPE GmbH
Dachauer Strasse 11, D-80335, Munchen, Germany Tel: 49-89-1893-9630
ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.
F5, ShenHua Financial Building, NO 1 NingBo Road, Huangpu District Shanghai, 200002, P.R. China Tel :86-21-51036677
ESPEC (MALAYSIA) SDN. BHD.
No.10 -1, Jalan Dagang SB 4/2, Taman Sungai Besi Indah Off Jalan Sungai Besi, 43300 Seri Kembangan Selangor Darul Ehsan Malaysia Tel: 60-3-8945-1377
www.espec.co.jp/english
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan Tel: 81-6-6358-4741
ENA-EGN-D 460V Edition Published April 2008, Rev. July 2013
Loading...