USES:
• Production Testing of Medical Electronic Products and Instruments in Accordance with UL, CSA, and IEC standards such as UL1950, IEC601-1, UL2601-1, UL1563 and UL544
• Electrical Safety Compliance Testing on a Variety of Products and Devices
FEATURES:
• Programmable output voltage to 5KV AC and 6KV DC
• Ground Bond Testing to 30A AC with Adjustable Limit
• Line/Earth Leakage Current Measurement(UL2601-1/UL1950)with 5 Human Body Models
• Insulation Resistance Measurements from 100k to 50G
• Monitoring of DUT Current Draw
• Programmable Ramp and Test Times
• Storage of 50 Tests Setups with 10 Steps per Setup
• Continuous Leakage Current Monitoring
• Trip Current Programmable to 40mA AC and 20mA DC
• Front Panel Lockout via Password
• Standard IEEE and Remote Control Interfaces
For Electronic Medical Devices
Introduction
The Guardian 6100 is designed for fast, easy, production testing of single phase electronic medical devices and instruments. Without changing the connection to the device, the tester is able to perform five essential electrical safety tests in one unit, including AC Hipot, DC Hipot, Insulation Resistance, Line Leakage Current and Ground Bond measurements. Performing all these electrical safety tests in one box increases production by reducing test time and eliminating multiple test stations.
Description
AC Hipot Test Performs AC dielectric testing (hipot) over the voltage range from
50 to 5000VAC RMS. Leakage current detection is programmable from 1µA to 40mA, making the 6100 ideal for testing medical products and instruments having a wide range of leakage currents.
DC Hipot Te s t DC dielectric testing is possible from 50 to 6000VDC with leakage
current detection down to 0.1µA. The maximum total output current of 20mA allows quick charging of capacitive devices and products.
Insulation Resistance The insulation resistance test (similar to DC Hipot)
calculates and displays a product’s insulation resistance value in ohms. This resistance can be measured over the range of 100kto 50Gwith test voltages programmable from 50 to 1000VDC.
Ground Bond The ground bond test, sometimes referred to as a high current
continuity test, can be programmed from 1 to 30A for verifying the integrity of a product’s ground system. Resistance measurements are displayed and a high limit programmable between 0.1mand 510m .
Line/Earth Leakage Current A line/earth leakage test can be performed directly
on the 6100 in four possible configurations; normal operating conditions, reverse line, or a single fault normal and single fault reverse mode. Five different human body models can be selected in accordance with UL, IEC, and other standards. Leakage current limits are programmable from 0.1µA to 9.999mA.
Note: This test requires input isolation transformer for proper operation
• Optional RS232 Interface
For more detailed specifications, visit
www.quadtech.com
For more information about special purchase,
rent & lease options, call
1-800-253-1230
Fax 1-508-485-0295
Intl. 1-508-229-0806
HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS • DECADES • STANDARDS
Guardian 6100
AC Output Vo l t a g e :
Range: 50V to 5000V AC, 1V re s o l u t i o n Frequency: 50 or 60 Hz Programmable Waveform: Sinusoidal Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 1µA to 40mA AC Resolution: 0.001mA Accuracy: +/-(1% + 5cnt)
High/Low Limit Test: 1µA to 40mA AC
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, 0.001mA/step
DC Output Vo l t a g e :
Range: 50V to 6000V DC, 1V re s o l u t i o n Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range: 0.1µA to 20mA DC
Resolution: 0.0001mA Accuracy: +/-(1% or reading + 5cnt)
High/Low Limit Test: 0.1µA to 20mA DC
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, 0.001mA/step
Insulation Resistance:
Range: 100k- 50G Accuracy: +/- 5% to +/- 15% depending upon volt­age and resistance
Voltage Range: 50V to 1000V DC Voltage Accuracy: +/- (1% of setting + 5V) High/Low Limit Test: 100k- 50G
Low limit can be turned OFF
IR Test Delay: 0.3 to 99.9 seconds Programmable in 0.1 second
steps
G round Bond:
Output Curre n t : Range: 1.0 to 30.0A AC, setting 0.1A/step
Accuracy: +/- (1% of setting + 0.3A)
Display: +/- (1% of reading + 3 counts) Frequency: 50 or 60Hz Selectable No Load Voltage: 6 to 15 V Programmable Resistance: Range: 0.1m- 510.0m, 4 digits
Accuracy: +/-(1% of reading + 3 counts) Resolution: 0.1m
Hi Limit: 0.1mto 510m Offset Function: 0 to 100moffset, user selectable Test Time: 0.5 - 999sec (+/-20ms)
Line Leakage Current:
Input Voltage: Range: 0V to 300V AC, 50 or 60Hz. Line Voltage Meter: 0 - 300VAC, + / -(1% of reading + 5cnts) Line Current Meter: 0 - 10A, + / -(5% of re a d i n g ) AC Current Display: R a n g e R e s o l u t i o n A c c u r a c y
0.0001mA - 0.59mA 0.0002mA +/-(2% + 5cnts)
0.6mA - 9.999mA 0.003mA +/-(2% + 5cnts)
Current Trip Limits: 0.1µA to 9.999mA, 1µA Resolution Measuring Circuit: 5 Types of Human Body Models in accordance with
UL544 NP, UL544P, UL1563, UL2601-1,IEC601-1, IEC 950, UL1950, UL3101 Standards
Measurement Modes: Normal, Reverse and Single Fault Max. DUT Current: 10A
Common Feature s :
AC/DC Test Time: Ramp: 0.1 to 99.9s (+/-20ms)
Test: 0.1 to 999s (+/-20ms) and Continuous
Remote Control: Inputs: Start,Stop
Outputs: Pass/Fail/Under Test Connector: Terminal Strip and 9 pin D Series
Test Setups: 50 Test Setups with 10 Steps each Rear Connectors: Input: Binding Posts (line & neutral)
Isolation transformer (not included) is required for proper operation (10A, 300V max)
Output: High Voltage Sockets for connection to 115V corded adapter, G30 (included)
Front Panel Lockout: Password Safety Feature s : Fast Cutoff (<0.4ms) and Fast Discharg e Miscellaneous: Continuous Voltage on Fail
Scanner Delay: 0.1 to 99.9, 0.1s/step
Indication: Pass/fail lights, audible sound Buzzer Level: 1,2,3 and Off Interface: IEEE-488 Standard, RS-232 Optional Dimensions: (w x h x d):16.875x5.625x20.50in
(421.9x140.6x512.5mm)
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping Environmental: Operating: 0 to + 40° C,
Humidity: <75% Storage: - 20 to + 70° C Warm-up Time: 1minute
Power: • 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 500W max
Ordering Information
G u a rdian 6100 Production Safety Analyzer
Includes:
Guardian 6000 Electrical Safety Analyzer 6000-04 Line Leakage Current Scan Unit 150528 Instruction Manual S02 HV Lead Set, 1m G15 Ground Continuity Lead Set
For more detailed specifications, visit www.quadtech.com • For more information about special purchase, rent & lease options, call
G30 Corded Product Power Cable
Calibration Certificate Traceable to NIST
Adapter (115V)
Optional Accessories
Calibration Data S05 Foot Switch
G26 RS232 Interface G27 Rack Mount Flanges G28 Printer Interface
G31 Isolation Transformer G32 Isolation Transformer
(replaces IEEE 488) (500VA) (1000VA)
1-800-253-1230 Fax 1-508-485-0295 Intl. 1-508-229-0806
5 Clock Tower Place, 210 East, Maynard, MA 01754