G u a rd i a n 6100 Production Safety A n a ly ze r
USES:
• Production Testing of Medical
Electronic Products and Instruments
in Accordance with UL, CSA, and IEC
standards such as UL1950, IEC601-1,
UL2601-1, UL1563 and UL544
• Electrical Safety Compliance Testing
on a Variety of Products and Devices
FEATURES:
• Programmable output voltage to 5KV
AC and 6KV DC
• Ground Bond Testing to 30A AC with
Adjustable Limit
• Line/Earth Leakage Current
Measurement(UL2601-1/UL1950)with
5 Human Body Models
• Insulation Resistance Measurements
from 100kΩ to 50GΩ
• Monitoring of DUT Current Draw
• Programmable Ramp and Test Times
• Storage of 50 Tests Setups with 10
Steps per Setup
• Continuous Leakage Current
Monitoring
• Trip Current Programmable to 40mA
AC and 20mA DC
• Front Panel Lockout via Password
• Standard IEEE and Remote Control
Interfaces
For Electronic Medical Devices
Introduction
The Guardian 6100 is designed for fast, easy, production testing of single phase
electronic medical devices and instruments. Without changing the connection to the
device, the tester is able to perform five essential electrical safety tests in one unit,
including AC Hipot, DC Hipot, Insulation Resistance, Line Leakage Current and Ground
Bond measurements. Performing all these electrical safety tests in one box increases
production by reducing test time and eliminating multiple test stations.
Description
AC Hipot Test Performs AC dielectric testing (hipot) over the voltage range from
50 to 5000VAC RMS. Leakage current detection is programmable from 1µA to 40mA,
making the 6100 ideal for testing medical products and instruments having a wide range
of leakage currents.
DC Hipot Te s t DC dielectric testing is possible from 50 to 6000VDC with leakage
current detection down to 0.1µA. The maximum total output current of 20mA allows
quick charging of capacitive devices and products.
Insulation Resistance The insulation resistance test (similar to DC Hipot)
calculates and displays a product’s insulation resistance value in ohms. This resistance
can be measured over the range of 100kΩ to 50GΩ with test voltages programmable
from 50 to 1000VDC.
Ground Bond The ground bond test, sometimes referred to as a high current
continuity test, can be programmed from 1 to 30A for verifying the integrity of a product’s
ground system. Resistance measurements are displayed and a high limit programmable
between 0.1mΩ and 510mΩ .
Line/Earth Leakage Current A line/earth leakage test can be performed directly
on the 6100 in four possible configurations; normal operating conditions, reverse line,
or a single fault normal and single fault reverse mode. Five different human body models
can be selected in accordance with UL, IEC, and other standards. Leakage current limits
are programmable from 0.1µA to 9.999mA.
Note: This test requires input isolation transformer for proper operation
• Optional RS232 Interface
For more detailed specifications, visit
www.quadtech.com
For more information about special purchase,
rent & lease options, call
1-800-253-1230
Fax 1-508-485-0295
Intl. 1-508-229-0806
HIPOT • LCR METERS • MEGOHMMETERS • MILLIOHMMETERS • DECADES • STANDARDS
Guardian 6100
AC Output Vo l t a g e :
Range: 50V to 5000V AC, 1V re s o l u t i o n
Frequency: 50 or 60 Hz Programmable
Waveform: Sinusoidal
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 1µA to 40mA AC
Resolution: 0.001mA
Accuracy: +/-(1% + 5cnt)
High/Low Limit Test: 1µA to 40mA AC
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, 0.001mA/step
DC Output Vo l t a g e :
Range: 50V to 6000V DC, 1V re s o l u t i o n
Regulation: <(1% +5V) at Rated Load
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range: 0.1µA to 20mA DC
Resolution: 0.0001mA
Accuracy: +/-(1% or reading + 5cnt)
High/Low Limit Test: 0.1µA to 20mA DC
Low limit can be turned OFF
Arc Detection: Programmable Level and OFF, 0.001mA/step
Insulation Resistance:
Range: 100kΩ - 50GΩ
Accuracy: +/- 5% to +/- 15% depending upon voltage and resistance
Voltage Range: 50V to 1000V DC
Voltage Accuracy: +/- (1% of setting + 5V)
High/Low Limit Test: 100kΩ - 50GΩ
Low limit can be turned OFF
IR Test Delay: 0.3 to 99.9 seconds Programmable in 0.1 second
steps
G round Bond:
Output Curre n t : Range: 1.0 to 30.0A AC, setting 0.1A/step
Accuracy: +/- (1% of setting + 0.3A)
Display: +/- (1% of reading + 3 counts)
Frequency: 50 or 60Hz Selectable
No Load Voltage: 6 to 15 V Programmable
Resistance: Range: 0.1mΩ - 510.0mΩ, 4 digits
Accuracy: +/-(1% of reading + 3 counts)
Resolution: 0.1mΩ
Hi Limit: 0.1mΩ to 510mΩ
Offset Function: 0 to 100mΩ offset, user selectable
Test Time: 0.5 - 999sec (+/-20ms)
Line Leakage Current:
Input Voltage: Range: 0V to 300V AC, 50 or 60Hz.
Line Voltage Meter: 0 - 300VAC, + / -(1% of reading + 5cnts)
Line Current Meter: 0 - 10A, + / -(5% of re a d i n g )
AC Current Display: R a n g e R e s o l u t i o n A c c u r a c y
0.0001mA - 0.59mA 0.0002mA +/-(2% + 5cnts)
0.6mA - 9.999mA 0.003mA +/-(2% + 5cnts)
Current Trip Limits: 0.1µA to 9.999mA, 1µA Resolution
Measuring Circuit: 5 Types of Human Body Models in accordance with
UL544 NP, UL544P, UL1563, UL2601-1,IEC601-1,
IEC 950, UL1950, UL3101 Standards
Measurement Modes: Normal, Reverse and Single Fault
Max. DUT Current: 10A
Common Feature s :
AC/DC Test Time: Ramp: 0.1 to 99.9s (+/-20ms)
Test: 0.1 to 999s (+/-20ms) and Continuous
Remote Control: Inputs: Start,Stop
Outputs: Pass/Fail/Under Test
Connector: Terminal Strip and 9 pin D Series
Test Setups: 50 Test Setups with 10 Steps each
Rear Connectors: Input: Binding Posts (line & neutral)
Isolation transformer (not included)
is required for proper operation
(10A, 300V max)
Output: High Voltage Sockets for connection
to 115V corded adapter, G30 (included)
Front Panel Lockout: Password
Safety Feature s : Fast Cutoff (<0.4ms) and Fast Discharg e
Miscellaneous: Continuous Voltage on Fail
Scanner Delay: 0.1 to 99.9, 0.1s/step
Indication: Pass/fail lights, audible sound
Buzzer Level: 1,2,3 and Off
Interface: IEEE-488 Standard, RS-232 Optional
Dimensions: (w x h x d):16.875x5.625x20.50in
(421.9x140.6x512.5mm)
Weight: 53 lbs (24kg) - Net, 60 lbs (27kg) Shipping
Environmental: Operating: 0 to + 40° C,
Humidity: <75%
Storage: - 20 to + 70° C
Warm-up Time: 1minute
Power: • 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 500W max
Ordering Information
G u a rdian 6100 Production Safety Analyzer
Includes:
Guardian 6000 Electrical Safety Analyzer
6000-04 Line Leakage Current Scan Unit
150528 Instruction Manual
S02 HV Lead Set, 1m
G15 Ground Continuity Lead Set
For more detailed specifications, visit www.quadtech.com • For more information about special purchase, rent & lease options, call
G30 Corded Product
Power Cable
Calibration Certificate Traceable to NIST
Adapter (115V)
Optional Accessories
Calibration Data
S05 Foot Switch
G26 RS232 Interface
G27 Rack Mount Flanges
G28 Printer Interface
G31 Isolation Transformer
G32 Isolation Transformer
(replaces IEEE 488)
(500VA)
(1000VA)
1-800-253-1230 • Fax 1-508-485-0295 • Intl. 1-508-229-0806
5 Clock Tower Place, 210 East, Maynard, MA 01754
PN030089/A1