Frequency ............................................................................................................................................................................................... 3
Sweep time ............................................................................................................................................................................................. 4
Inputs and outputs .................................................................................................................................................................................. 7
R&S®FSH4/R&S®FSH8 model .24/.28 without R&S®FSH-K42 option .............................................................................................. 10
R&S®FSH4/R&S®FSH8 model .24/.28 with R&S®FSH-K41 option ................................................................................................... 11
R&S®FSH-K44 3GPP WCDMA BTS/NodeB pilot channel and pilot EVM measurement application
®
R&S
FSH-K44E 3GPP WCDMA BTS/NodeB code domain power and EVM measurement application
with HSDPA/HSPA+ analyzer ............................................................................................................................................................... 12
R&S®FSH-K46 CDMA2000® BTS pilot channel and EVM measurement application
®
FSH-K46E CDMA2000® BTS code domain power measurement application ............................................................................. 14
R&S
R&S®FSH-K47 1xEV-DO® BTS pilot channel and EVM measurement application
®
FSH-K47E 1xEV-DO® BTS PN scanner and time domain power measurement application ...................................................... 16
R&S
R&S®FSH-K48 3GPP TD-SCDMA BTS power and P-CCPCH EVM measurement application
®
FSH-K48E 3GPP TD-SCDMA/HSDPA BTS code domain power and EVM measurement application ...................................... 17
R&S
R&S®FSH-K50/R&S®FSH-K51 LTE FDD/TDD downlink pilot channel and EVM measurement application
R&S®FSH-K43 Receiver Mode and Channel Scan Measurement Application ..................................................................................... 20
General data ......................................................................................................................................................................................... 21
Equivalence of specifications for different R&S®FSH part numbers ..................................................................................................... 22
R&S®FSH-Z1 and R&S®FSH-Z18 power sensors ............................................................................................................................. 23
R&S®FSH-Z14 directional power sensor ........................................................................................................................................... 23
R&S®FSH-Z44 directional power sensor ........................................................................................................................................... 25
R&S®FSH-Z114 precision frequency reference ................................................................................................................................ 26
Ordering information .................................................................................................................................................... 27
R&S®NRP-Zxx power sensors supported by the R&S®FSH4/8/13/20 .................................................................................................. 29
Service options ..................................................................................................................................................................................... 30
Specifications apply under the following conditions:
15 minutes warm-up time at ambient temperature, specified environmental conditions met, calibration cycle adhered to.
Data without tolerances: typical values only. Data designated as “nominal” applies to design parameters and is not tested. Data
without tolerance limits is not binding.
Frequency
Frequency range R&S®FSH4 model .04/.14 9 kHz to 3.6 GHz
R&S®FSH8 model .08/.18 9 kHz to 8 GHz
R&S®FSH4 model .24 100 kHz to 3.6 GHz
R&S®FSH8 model .28 100 kHz to 8 GHz
R&S®FSH13 9 kHz to 13.6 GHz
R&S®FSH20 9 kHz to 20 GHz
Frequency resolution 1 Hz
Reference frequency, internal
Aging per year 1 × 10
Temperature drift 0 °C to +30 °C 1 × 10
+30 °C to +50 °C 3 × 10
Achievable initial calibration accuracy 5 × 10
6
6
6
7
Total reference uncertainty (time since last adjustment × aging rate) +
temperature drift + calibration accuracy
Reference frequency, with R&S®HA-Z240 GPS receiver option
Frequency uncertainty GPS on, 1 minute after satellite lock ±2.5 × 10
up to 30 minutes after loosing satellite lock ±5 × 10
Reference frequency, with R&S®FSH-Z114 precision frequency reference option
Aging per year 3.6 × 10
Temperature drift 0 °C to +50 °C 4 × 10
Achievable initial calibration accuracy 1 × 10
8
8
9
10
9
Total reference uncertainty R&S®FSH-Z114 connected
30 seconds after oscillator lock (time since last adjustment × aging rate) +
temperature drift + 3 × calibration accuracy
(nominal)
2 minutes after oscillator lock (time since last adjustment × aging rate) +
temperature drift + calibration accuracy
Frequency readout
Marker resolution 0.1 Hz
Uncertainty ±(marker frequency × reference
uncertainty + 10 % × resolution bandwidth
+ ½ (span/(sweep points – 1) + 1 Hz)
Number of sweep (trace) points 631
Marker tuning frequency step size span/630
10 Hz span 600 MHz 20 ms to 1000 s
span > 600 MHz 20 ms × span/600 MHz to 1000 s
Uncertainty span = 0 Hz nominal 1 %
span 10 Hz nominal 3 %
Bandwidths
Resolution bandwidths
Range –3 dB bandwidth 1 Hz to 3 MHz in 1/3 sequence
Bandwidth accuracy 1 Hz RBW 300 kHz nominal < 5 %
RBW > 300 kHz nominal < 10 %
Selectivity 60 dB:3 dB nominal < 5 (Gaussian type filters)
Video filters
Range –3 dB bandwidth 1 Hz to 3 MHz in 1/3 sequence
Level
Display range displayed noise floor to +30 dBm
Maximum rated input level with RF attenuation ≥ 10 dB
DC voltage R&S®FSH4/ R&S®FSH 8,
CW RF power 30 dBm (= 1 W)
Peak RF power duration < 3 s 33 dBm (= 2 W)
Max. pulse voltage 150 V
Max. pulse energy pulse width 10 µs 10 mWs
Maximum rated input level with RF attenuation < 10 dB
DC voltage 50 V
CW RF power 20 dBm (= 100 mW)
Peak RF power duration < 3 s 23 dBm (= 200 mW)
Max. pulse voltage 50 V
Max. pulse energy pulse width 10 µs 1 mWs
Intermodulation
Third-order intercept (TOI),
nominal values
Second harmonic intercept (SHI),
nominal values
model .04/.08/.14/.18
R&S®FSH4/ R&S®FSH 8, model .24/.28,
®
FSH13, R&S®FSH 20
R&S
intermodulation-free dynamic range, signal level 2 × –20 dBm, RF attenuation = 0 dB,
RF preamplifier = off
Other interfering signals,
signal level – RF attenuation < –20 dBm
FSH4/R&S®FSH8, serial number < 105000
R&S
f 3.6 GHz,
spurious at f
– 2440.7 MHz
in
3.6 GHz < f 8 GHz,
spurious at f
– 4465.7 MHz
in
< –60 dBc
< –60 dBc
R&S®FSH4/R&S®FSH8, serial number 105000, R&S®FSH13, R&S®FSH20
f 3.6 GHz,
spurious at f
– 2446.4 MHz
in
3.6 GHz < f 8 GHz,
spurious at f
– 4462.4 MHz
in
< –60 dBc
< –60 dBc
8 GHz < f 20 GHz, < –60 dBc
Other interfering signals, related to local
oscillators
f 3.6 GHz
f < 300 kHz –60 dBc
f 300 kHz < –60 dBc
f > 3.6 GHz
f < 300 kHz –54 dBc
f 300 kHz < –54 dBc
f = receive frequency
Residual spurious response input matched with 50 ,
< –90 dBm
without input signal, RBW 30 kHz,
f 3 MHz, RF attenuation = 0 dB,
tracking generator off
Level display
Logarithmic level axis 1/2/5/10/20/50/100 dB, 10 divisions
Linear level axis 0 % to 100 %, 10 divisions
Number of traces 2
Trace detectors max. peak, min. peak, auto peak, sample,
RMS
Trace functions clear/write, max. hold, min. hold, average,
view
Setting range of reference level –80 dBm to +30 dBm
Units of level axis dBm, dBmV, dBµV, V, W
Level measurement uncertainty
Absolute level uncertainty at 100 MHz +20 °C to +30 °C < 0.3 dB
Frequency response (+20 °C to +30 °C) 9 kHz f < 100 kHz
nominal < 1.5 dB
(models .04/.14/.08/.18 only)
100 kHz f < 10 MHz nominal < 1.5 dB
10 MHz f 3.6 GHz < 1 dB
3.6 GHz < f 20 GHz < 1.5 dB
Attenuator uncertainty < 0.3 dB
Uncertainty of reference level setting nominal < 0.1 dB
Display nonlinearity SNR > 16 dB, 0 dB to –50 dB,
< 0.2 dB
logarithmic level display
Bandwidth switching uncertainty reference: RBW = 10 kHz nominal < 0.1 dB
Total measurement uncertainty 95 % confidence level, +20 °C to +30 °C,
SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto
10 MHz f 3.6 GHz < 1 dB, typ. 0.5 dB
3.6 GHz < f 20 GHz < 1.5 dB, typ. 1 dB
Trigger functions
Trigger
Trigger source free run, video, external
External trigger level threshold low high transition 2.4 V
high low transition 0.7 V
Gated trigger
Gate source external
Gate delay 10 µs to 100 s, min. resolution 10 µs
Impedance 50
Connector N female
VSWR 100 kHz f 1 GHz nominal < 1.5
1 GHz < f 6 GHz nominal < 2
6 GHz < f 20 GHz nominal < 3
Input attenuator RF input only 0 dB to 40 dB in 5 dB steps
AF output
AF demodulation types AM and FM
Connector 3.5 mm mini jack
Output impedance nominal 32
Voltage (open circuit) V
Power sensor
Connector 7-contact female (type Binder 712)
Power sensors supported see accessories
Tracking generator (models .14/.18/.24/.28 only)
Frequency range models .14 and .24 100 kHz to 3.6 GHz
models .18 and .28 100 kHz to 8 GHz
Connector N female, 50
VSWR 100 kHz f 1 GHz nominal < 1.5
1 GHz < f 6 GHz nominal < 2
6 GHz < f 8 GHz, models .18 and .28 nominal < 3
Output level tracking generator attenuation = 0 dB nominal 0 dBm
Tracking generator attenuator 0 dB to 40 dB in 1 dB steps
Dynamic range for isolation
measurements
measurement mode = vector voltmeter magnitude + phase, Smith chart
Return loss
Range selectable 1/2/5/10/20/50/100 dB, linear 100 %
Resolution 0.01 dB
Measurement uncertainty see figure “Uncertainty of reflection
One-port phase
Range selectable 90/180/360/1000° to 10000° in 1/2/5 steps
Resolution 0.01°
Measurement uncertainty specifications are based on a matched DUT, RBW = 100 Hz, RF attenuation = 10 dB,
nominal source power = 0 dBm, +20 °C to +30 °C
300 kHz f 3.6 GHz
0 dB return loss < 15 dB nominal < 3°
15 dB return loss < 25 dB nominal < 6°
25 dB return loss < 35 dB nominal < 20°
3.6 GHz < f 8 GHz (R&S®FSH8 only)
0 dB return loss < 15 dB nominal < 3°
15 dB return loss < 25 dB nominal < 6°
25 dB return loss < 35 dB nominal < 20°
VSWR
Range selectable 1 to 1.1, 1.5, 2, 6, 11, 21 or 71
Smith chart
Range 1, zoom × 2, × 4, × 8
Reflection coefficient
mRho range 1 to 1000 in 1, 2, 5 steps
Corrected directivity 300 kHz f 3 GHz nominal > 43 dB
3 GHz < f 6 GHz nominal > 37 dB
6 GHz < f 8 GHz nominal > 31 dB
Corrected test port match 300 kHz f 3 GHz nominal > 40 dB
3 GHz < f 6 GHz nominal > 37 dB
6 GHz < f 8 GHz nominal > 30 dB
Uncertainty of reflection measurement with R&S®FSH-K42/-K45 option
3
nominal 0 dBm to –40 dBm in 1 dB steps
VSWR, reflection coefficient, Smith chart,
cable loss, group delay, electrical length
®
measurement with the R&S
®
R&S
FSH-K45 option”
FSH-K42/
Uncertainty of reflection measurement with the R&S
Measurement range –120 dB to +80 dB
Display range selectable 1/2/5/10/20/50/100 dB, linear 100 %
Resolution 0.01 dB
Measurement uncertainty calibration method = full two port high
accuracy
see figure “Transmission magnitude
uncertainty”
Phase
Range selectable 90/180/360/1000° to 10000°
in 1/2/5 steps
Resolution 0.01°
Measurement uncertainty specifications are based on a matched DUT, RBW = 100 Hz, RF attenuation = 10 dB,
nominal source power = 0 dBm, +20 °C to +30 °C
300 kHz f 50 MHz
0 dB insertion loss < 40 dB nominal < 2°
50 MHz < f 3.6 GHz
0 dB insertion loss < 50 dB nominal < 2°
50 dB insertion loss < 70 dB nominal < 3°
3.6 GHz < f < 6 GHz (R&S®FSH8 only)
0 dB insertion loss < 50 dB nominal < 2°
50 dB insertion loss < 70 dB nominal < 3°
6 GHz f < 8 GHz (R&S®FSH8 only)
0 dB insertion loss < 50 dB nominal < 3°
50 dB insertion loss < 70 dB nominal < 5°
Dynamic range from port 1 to port 2 RF attenuation = 0 dB, tracking generator attenuation = 10 dB, RBW = 1 kHz
100 kHz f < 300 kHz typ. 70 dB
300 kHz f < 6 GHz > 70 dB, typ. 90 dB
6 GHz f < 8 GHz typ. > 50 dB
Dynamic range from port 2 to port 1 RF attenuation = 0 dB, tracking generator attenuation = 10 dB, RBW = 1 kHz
100 kHz f < 300 kHz typ. 80 dB
300 kHz f < 6 GHz > 80 dB, typ. 100 dB
6 GHz f < 8 GHz typ. > 60 dB
Test port match as specified for tracking generator
output/RF input
Transmission magnitude uncertainty
with calibration method "Full Two Port High Accuracy", f=1GHz, IF bandwidth=100Hz