Atec Fluke-9141 User Manual

Field Dry-Wells
Interchangeable Insert Options
1/4" (6.35 mm)
3/16" (4.8 mm)
1/8" (3.2 mm)
1/2" (12.7 mm)
1/16" (1.6 mm)
3/8' (9.5 mm)
3/8" (9.5 mm)
1/4" ( 6.35 mm)
3/16" (4.8 mm)
1/4" (6.35 mm)
Insert “A” Insert “B”
Insert “C”
6 mm
Insert “D”
4 mm
6 mm
3 mm
3 mm
4 mm
Lightweight and very portable
Accuracy to ±0.25 °C
RS-232 and Interface-it software included
Easy to recalibrate
If you’ve been using dry-well calibrators for field work, you know there’s a lot more to a dry-well than its temperature range and stability. Size, weight, speed, convenience, and software are also significant.
Field dry-wells need to be portable, flexible, and suitable for high-volume cal­ibrations or certifications. If they’re not, you’ll soon forget about the great stuff the sales rep told you and realize what you’ve really bought.
At Hart Scientific, we use dry-wells every day in our manufacturing and cali­bration work, and we know what makes
a dry-well easy and productive to use— which is exactly how users describe our series of field dry-wells. These dry-wells work for you instead of the other way around.
These three units beat every other comparable dry-well in the industry in performance, size, weight, convenience, ease of calibration, software, and price. In addition, the heating and cooling rate of each of these dry-wells is adjustable from the front panel, thermal switches can be checked for actuation testing, and multi­ple-hole inserts are available for a variety of probe sizes.
Hart dry-wells are easy to calibrate. You don’t even have to open the case. This means less maintenance costs and less down time when they do need calibration.
Our Interface-it software lets you ad
­just set-points and ramp rates, log dry­well readings to a file, create an elec­tronic strip chart, and perform thermal switch testing with data collection. The software is written for Windows and has a great graphical interface. Regardless of whether you want basic software or a completely automated calibration system, we’ve got what you want. Read about all our great packages starting on page 82.
Every dry-well we ship is tested at our factory, and every unit comes with a NIST-traceable calibration. There’s no ex­trachargeforthereport,becausewe consider it an essential ingredient in our quality program. You shouldn’t have to pay extra for calibration procedures we perform anyway.
9103
The 9103 covers below-ambient temper­atures as low as –25 °C. The 9103 is sta­ble to ±0.02 °C, and its display is cali­brated to an accuracy of ±0.25 °C at all temperatures within its range. In just eight minutes, 0 °C is reached, and 100 °C is reached in six minutes, so your time is spent calibrating—not waiting.
The 9103 reaches temperatures 50 °C below ambient, so –25 °C is reached un­der normal ambient conditions. Our com­petitors like to advertise their units as reaching –45 °C when they really mean –45 °C below ambient, which typically means it will go to –20 °C. Our unit does not require you to work in a walk-in freezer to achieve its full advertised range.
Choose one of three removable inserts sized for probes from 1/16 inch to 1/2 inch in diameter. Insert A handles a full range of probe sizes with a single well of each size. Insert B features two wells each of 3/8, 1/4, and 3/16 inches in
When ordering, replace the “X” in the model number with the appropriate insert letter. Order additional inserts as your applications require.
134 Industrial
Field Dry-Wells
Specifications
Range –25 °C to 140 °C (–13 °F to 284 °F)
Accuracy ±0.25 °C ±0.5 °C (holes greater than 1/4"
Stability ±0.02°Cat–25°C
Well-to-Well Uniformity
Heating Times 18 minutes from ambient to 140 °C 12 minutes from ambient to 350 °C 12 minutes from ambient to 650 °C
Cooling Times 20 minutes from ambient to –25 °C 15 minutes from 350 °C to 100 °C 25 minutes from 650 °C to 100 °C
Stabilization Time 7minutes
Immersion Depth 124 mm (4.875")
Inserts Insert A, B, C, or D included (specify when ordering)
Outside Insert Dimensions
Computer Interface RS-232 included with free Interface-it software (Model 9930)
Power 115 VAC (±10 %), 1.3 A or 230 VAC
Size (WxHxD) 143 x 261 x 245 mm
Weight 5.7 kg (12 lb.) 2.7 kg (6 lb.) 3.6 kg (8 lb.)
NIST-Traceable Certificate
±0.1 °C between similarly sized wells ±0.1 °C with similarly sized wells ±0.1 °C below 400 °C, ±0.5 °C above
(±10 %), 0.7 A, switchable, 50/60 Hz,
(5.63 x 10.25 x 9.63 in)
Data at –25 °C, 0 °C, 25 °C, 50 °C, 75 °C,
9103 9140 9141
at 23 °C ambient
±0.04°Cat140°C
31.8 mm dia. x 124 mm (1.25 x 4.88 in) 28.5 mm dia. x 124 mm (1.12 x 4.88 in)
150 W
100 °C, and 140 °
35 °C to 350 °C (95 °F to 662 °F) 50 °C to 650 °C (122 °F to 1202 °F)
[6.35 mm]: ±1 °C)
±0.03°Cat50°C
±0.05°Cat350°C
115 VAC (±10 %), 4.4 A or 230 VAC
(±10 %), 2.2 A, switchable, 50/60 Hz,
500 W
152 x 86 x 197 mm
(6 x 3.375 x 7.75 in)
Data at 50 °C, 100 °C, 150 °C, 200 °C,
250 °C, 300 °C, and 350 °C
±0.5 °C to 400 °C; ±1.0 °C to 650 °C
(holes greater than 1/4": ±2 °C)
400 °C with similarly sized wells
115 VAC (±10 %), 8.8 A or 230 VAC
(±10 %), 4.4 A, switchable, 50/60 Hz,
Data at 100 °C, 200 °C, 300 °C, 400 °C,
±0.05°Cat100°C ±0.12°Cat500°C ±0.12°Cat650°C
1000 W
109 x 236 x 185 mm
(4.3 x 9.3 x 7.3 in)
500 °C, and 600 °C
diameter for doing comparison calibra­tions. Insert C has six 1/4-inch-diameter wells for multiple probe calibrations, and Insert D has three pairs of metric sized wells.
9140
The 9140 has a temperature range of 35 °C to 350 °C, and it reaches its maxi­mum temperature in 12 minutes. At six pounds, it’s small enough to easily carry in one hand. It’s truly a unique innovation in dry-wells.
The unit has a stability of ±0.05 °C or better and a uniformity of at least 0.4 °C in the largest-diameter wells and 0.1 °C in the smaller wells. Despite its small size, this unit performs.
Use the display, calibrated to ±0.5 °C, as your reference, or use an external ther­mometer for maximum calibration accuracy. With three removable inserts to choose from, the 9140 is as versatile as it is fast.
9141
Here’s an upright unit you’re going to love. It does calibrations up to 650 °C, weighs only eight pounds, and heats up to 650 °C in only 12 minutes—12! This dry-well does everything but get legs
and walk to the job for you. (And we’re working on one that does that too.)
This four-inch-wide dry-well is amaz­ing. You can control all functions from the front panel or hook it up to your PC with its built-in RS-232 port. And just like the 9140, it works with all of our software described on page 82.
It has three removable well inserts available, an optional carrying case, a NIST-traceable calibration, and the best price in the industry.
Ordering Information
9103-X Dry-Well (specify X, X = A, B, C, or D
3103-1 Insert, 3103-2 Insert A 3103-3 Insert B 3103-4 Insert C 3103-6 Insert D 9316 Rugged Carrying Case
9140-X Dry-Well (specify X, X = A, B, C, or D
3140-1 Insert, 3140-2 Insert A 3140-3 Insert B 3140-4 Insert C 3140-6 Insert D 9308 Rugged Carrying Case
9141-X Dry-Well (specify X, X = A, B, C, or D
3141-1 Insert, 3141-2 Insert A 3141-3 Insert B 3141-4 Insert C 3141-6 Insert D 9309 Rugged Carrying Case
included insert)
blank
included
included
insert)
blank
insert)
blank
Loading...