Accuracy is given as ±([% of reading] + [number
of least significant digits]) at 18 °C to 28 °C, with
relative humidity up to 90 %, for a period of one
year after calibration.
1
For Model 87 in the 4
number of least significant digits (counts) by 10.
AC conversions are ac-coupled and valid from
3 % to 100 % of range. Model 87 is true-rms
responding. AC crest factor can be up to 3 at full
scale, 6 at half scale. For non-sinusoidal
forms add -(2 % Rdg + 2 % full scale) typical,
crest factor up to 3.
⁄2-digit mode, multiply the
wave
for a
Fluke 87V ac voltage function specifications (true-rms)
The Fluke 87V is a true-rms responding meter. When the input leads are shorted together in the ac functions, the meter may display a residual reading between 1
and 30 counts. A 30-count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce a much
larger constant error in later measurements.
3
Frequency range: 1 kHz to 2.5 kHz.
4
A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
5
Specification increases from -1 % at 200 Hz to -6 % at 440 Hz when filter is in use.
AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sine wave input.
2
AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range.
3
Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between 1 and
30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of range. Using REL to offset this reading may produce a much larger
constant error in later measurements.
4
10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A unspecified.
5
Below a reading of 200 counts, add 10 counts.
6
400 mA continuous; 600 mA for 18 hours maximum.
Burden Voltage
(typical)
Capacitance and diode function specifications
esolution
ange
Function
R
10.00 nF0.01 nF± (1 % + 2)
100.0 nF0. 1 nF± (1 % + 2)
1.000 µF0.001 µF± (1 % + 2)
10.00 µF0.01 µF± (1 % + 2)
100.0 µF0.1 µF± (1 % + 2)
9999 µF1 µF± (1 % + 2)
3.000 V0.001 V± (2 % + 1)
1
ith a film capacitor or better
W
2 Fluke Corporation Fluke 83V and 87V detailed specifications
600 mV dc70 mV (to 400 Hz)70 mV (to 400 Hz)40 mV
600 mV ac150 mV150 mV—
6 V0.3 V0.7 V1.7 V
60 V3 V7 V (≤ 140 kHz)4 V
600 V30 V70 V (≤ 14.0 kHz)40 V
1000 V100 V700 V (≤ 1.4 kHz)100 V
Duty Cycle RangeAccuracy
0.0 to 99.9 %Within ± (0.2 % per kHz + 0.1 %) for risetimes < 1 µs
1
Maximum input for specified accuracy = 10X Range or 1000 V.
pproximate Trigger L
A
Electrical characteristics of the terminals
FunctionOverloadInput ImpedanceCommon Mode Rejection Ratio
Protection
1000 V rms10 MΩΩ < 100 pF> 120 dB at dc, 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
000 V rms
1
1000 V rms10 M
1000 V rms< 7.3 V dc< 4.1 V dc< 4.5 V dc1 mA 100 µA 10 µA1µA1 µA 0.5 µA
1000 V rms< 3.9 V dc3.000 V dc0.6 mA typical
1
106V Hz maximum
1
(nominal)(1 kΩ unbalance)Normal Mode Rejection
10 M
< 100 pF>120 dB at dc, 50 Hz or 60 Hz> 60 dB at 50 Hz or 60 Hz
ΩΩ
ΩΩ < 100 pF> 60 dB, dc to 60 Hz
(ac-coupled)
Open Circuit
oltage
est V
T
Full Scale VoltageTypical Short Circuit Current
o 6.0 M
T
Ω50 MΩ or 6
0 nS
00
6
evel
Ω6 k
6
Ω 50 MΩ
00 k 6 M
6
0 k
MIN MAX recording specifications
ModelNominal ResponseAccuracy
00 ms to 8
83V
1
87V100 ms to 80 %Specified accuracy ± 12 counts for changes > 200 ms in duration
(dc functions)> 25 % of range
120 ms to 80 %Specified accuracy ± 40 counts for changes > 350 ms and inputs
(ac functions)
250
(Model 8
1
For repetitive peaks: 1 ms for single events.
0 %Specified accuracy ± 12 counts for changes > 200 ms in duration
ounts in ac with beeper on)
40 c
±
(
s (peak)
µ
7 only)
Specified accuracy ± 100 counts for changes > 250 µs in duration
1
100 counts for readings over 6000 counts)
±
(add
100 counts for readings in Low Pass mode)
±
(add
Fluke 83V and 87V detailed specifications Fluke Corporation 3
Fluke 83V and 87V General Specifications
Maximum voltage betw
Fuse protection f
een any terminal and earth ground:1000 V rms
or mA or µA inputs:
100 A, 1000 V FAST Fuse
44/
Fuse protection for A input: 11 A, 1000 V FAST Fuse
Display:
Digital: 6000 counts updates 4/sec; (Model 87V also has 19,999 counts in high-resolution mode)
Analog: 33 segments, updates 40/sec.
Frequency: 19,999 counts, updates 3/sec at > 10 Hz
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C
Altitude:
Operating: 2000 m
Storage: 10,000 m
Temperature coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C)
Electromagnetic compatibility: In an RF field of 3 V/m total accuracy = specified accuracy
Relative humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 7 0% (35 °C to 55 °C)
Batter
y type:
9 V zinc, N
EDA 1604 or 6F22 or 006P
Battery life: 400 hours typical with alkaline (with backlight off)
Vibration: Per MIL-PRF-28800 for a Class 2 instrument
Shock: 1 Meter drop per IEC 61010-1:2001
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with holster and flex-stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight: 12.5 oz (355 g)
Weight with holster and flex-stand: 22.0 oz (624 g)
Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category
III, IEC 664 to 600 V Overvoltage Category IV. UL listed to UL3111-1. Licensed by TÜV to EN61010-1.
4 Fluke Corporation Fluke 83V and 87V detailed specifications
Fluke. Keeping your world
up and running.
Fluke Corporation
P
O Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
ax (425) 446-5
F
In Europe/
1 40) 2 6
ax (3
F
In Canada (8
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or