Atec Fluke-87V User Manual

Fluke 83V and 87V
Digital Multimeters
Detailed Specifications
For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18 °C to 28 °C, with relative humidity up to 90 %, for a period of one year after calibration.
1
For Model 87 in the 4 number of least significant digits (counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range. Model 87 is true-rms responding. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal forms add -(2 % Rdg + 2 % full scale) typical, crest factor up to 3.
2-digit mode, multiply the
wave
for a
Fluke 87V ac voltage function specifications (true-rms)
Accuracy
Function Range Resolution 45 - 65 Hz 30 - 200 Hz 200 - 440 Hz
2,4
1
Below 10 % of range, add 6 counts.
2
The Fluke 87V is a true-rms responding meter. When the input leads are shorted together in the ac functions, the meter may display a residual reading between 1 and 30 counts. A 30-count residual reading will cause only a 2-digit change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error in later measurements.
3
Frequency range: 1 kHz to 2.5 kHz.
4
A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
5
Specification increases from -1 % at 200 Hz to -6 % at 440 Hz when filter is in use.
600.0 mV 0.1 mV ± (0.7 % + 4) ± (2.0 % + 4) ± (2.0 % + 20)
6.000 V 0.001 V ± (1.0 % + 4)
60.00 V 0.01 V ± (0.7 % + 2) ± (2.0 % + 4)3unspecified
600.0 V 0.1 V 1000 V 1 V unspecified unspecified
Using low pass filter ± (0.7 % + 2) ± (1.0 % + 4) + 1 % + 4 unspecified unspecified unspecified
-6 % - 4
440 Hz - 1 kHz
5
1 - 5 kHz 5 - 20 kHz
Fluke 83V ac voltage function specifications (average responding rms indicating)
y
2
ange Resolution 50 Hz - 60 Hz 30 Hz - 1 kHz 1 kHz - 5 kHz
Function
1
1
Below a reading of 200 counts, add 10 counts
2
Frequency range: 1 kHz to 2.5 kHz
R
00.0 mV
6
6.000 V
0.00 V
6
00.0 V
6
000 V
1
1 mV
0.
0.00
0.0
0. 1 V
1 V
1 V
1 V
Accurac
(0.5 % + 4) ± (1.0 % + 4) ± (2.0 % + 4)
±
(0.5 % + 2)
±
(0.5 % + 2) ± (1.0 % + 4) ± (2.0 % + 4)
±
(0.5 % + 2) ± (1.0 % + 4) ± (2.0 % + 4)
±
(0.5 % + 2)
±
± (1.0 % + 4) ± (2.0 % + 4)
± (1.0 % + 4) unspecified
1
Fluke 83V and 87V Detailed Specifications cont.
DC voltage, resistance, and conductance function specifications
Accuracy
Function Range Resolution Fluke 83V Fluke 87V
6.000 V 0.001 V ± (0.1 % + 1) ± (0.05 % + 1)
60.00 V 0.01 V ± (0.1 % + 1) ± (0.05 % + 1)
600.0 V 0.1 V ± (0.1 % + 1) ± (0.05 % + 1) 1000 V 1 V ± (0.1 % + 1) ± (0.05 % + 1)
600.0 mV 0.1 mV ± (0.3 % + 1) ± (0.1 % + 1)
600.0 0.1 ± (0.4 % + 2)1± (0.2 % + 2)
6.000 k 0.00
1 k
±
(0.4 % + 1) ± (0.2 % + 1)
Temperature specifications (87V only)
Temperature Resolution Accuracy
-200 °C to +1090 °C 0.1 °C 1 % + 10
-328 °F to +1994 °F 0.1 °F 1 % + 18
1
Does not include error of the thermocouple probe.
2
Accuracy specification assumes ambient temperature stable to For ambient temperature changes of after 1 hour.
1
± 5 °C, rated accuracy applies
1, 2
± 1 °C.
60.00 k 0.01 k ± (0.4 % + 1) ± (0.2 % + 1)
600.0 k 0.1 k ± (0.7 % + 1) ± (0.6 % + 1)
6.000 M 0.001 M ± (0.7 % + 1) ± (0.6 % + 1)
50.00 M
0.01 M ± (1.0 % + 3)2± (1.0 % + 3)
nS
1
When using the REL function to compensate for offsets
2
Add 0.5 % of reading when measuring above 30 Min the 50 Mrange and 20 counts below 33 nS in the 60 nS range
60.00 nS 0.01 nS ± (1.0 % + 10)1± (1.0 % + 10)
2
1
Current function specifications
Accuracy
Function Range Resolution Model 83
60.00 mA 0.01 mA ± (1.2 % + 2)
400.0 mA
(45 Hz to 2 kHz) 10.00 A
60.00 mA 0.01 mA ± (0.4 % + 4) ± (0.2 % + 4) 1.8 mV/mA
400.0 mA
10.00 A
6
6.000 A 0.001 A ± (1.2 % + 2)
4
6
6.000 A 0.001 A ± (0.4 % + 4) ± (0.2 % + 4) 0.03 V/A
4
0.1 mA ± (1.2 % + 2)
0.01 A ± (1.2 % + 2)
0.1 mA ± (0.4 % + 2) ± (0.2 % + 2) 1.8 mV/mA
0.01 A ± (0.4 % + 2) ± (0.2 % + 2) 0.03 V/A
600.0 µA 0.1 µA ± (1.2 % + 2) 000 µA 1 µA ± (1.2 % + 2)
6
1
5 5 5 5
5 5
Model 87
± (1.0 % + 2) 1.8 mV/mA ± (1.0 % + 2) 1.8 mV/mA ± (1.0 % + 2) 0.03 V/A ± (1.0 % + 2) 0.03 V/A
± (1.0 % + 2) 100 µV/µA ±
2, 3
(1.0 % + 2) 100 µV/µA
(45 Hz to 2 kHz)
µA 0.1 µA ± (0.4 % + 4) ± (0.2 % + 4) 100 µV/µA
00.0
6
µA 1 µA ± (0.4 % + 2) ± (0.2 % + 2) 100 µV/µA
000
6
1
AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sine wave input.
2
AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range.
3
Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error in later measurements.
4
10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum; > 10 A unspecified.
5
Below a reading of 200 counts, add 10 counts.
6
400 mA continuous; 600 mA for 18 hours maximum.
Burden Voltage
(typical)
Capacitance and diode function specifications
esolution
ange
Function
R
10.00 nF 0.01 nF ± (1 % + 2)
100.0 nF 0. 1 nF ± (1 % + 2)
1.000 µF 0.001 µF ± (1 % + 2)
10.00 µF 0.01 µF ± (1 % + 2)
100.0 µF 0.1 µF ± (1 % + 2) 9999 µF 1 µF ± (1 % + 2)
3.000 V 0.001 V ± (2 % + 1)
1
ith a film capacitor or better
W
2 Fluke Corporation Fluke 83V and 87V detailed specifications
, using Relative mode to zero residual.
R
Accurac
y
1 1
Frequency counter specifications
Function Range Resolution Accuracy
Frequency 199.99 0.01 Hz ± (0.005 % + 1) (0.5 Hz to 200 kHz, 1999.9 0.1 Hz ± (0.005 % + 1) pulse width > 2 µs) 19.999 kHz 0.001 kHz ± (0.005 % + 1)
199.99 kHz 0.01 kHz ± (0.005 % + 1) > 200 kHz 0.1 kHz unspecified
Frequency counter sensitivity and trigger levels
Minimum Sensitivity (RMS Sine wave)
Input Range
1
5 Hz - 20 kHz 0.5 Hz - 200 kHz (DC Voltage Function)
600 mV dc 70 mV (to 400 Hz) 70 mV (to 400 Hz) 40 mV 600 mV ac 150 mV 150 mV — 6 V 0.3 V 0.7 V 1.7 V 60 V 3 V 7 V (140 kHz) 4 V 600 V 30 V 70 V (14.0 kHz) 40 V 1000 V 100 V 700 V (1.4 kHz) 100 V
Duty Cycle Range Accuracy
0.0 to 99.9 % Within ± (0.2 % per kHz + 0.1 %) for risetimes < 1 µs
1
Maximum input for specified accuracy = 10X Range or 1000 V.
pproximate Trigger L
A
Electrical characteristics of the terminals
Function Overload Input Impedance Common Mode Rejection Ratio
Protection
1000 V rms 10 M< 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
000 V rms
1 1000 V rms 10 M
1000 V rms < 7.3 V dc < 4.1 V dc < 4.5 V dc 1 mA 100 µA 10 µA 1µA 1 µA 0.5 µA
1000 V rms < 3.9 V dc 3.000 V dc 0.6 mA typical
1
106V Hz maximum
1
(nominal) (1 kunbalance) Normal Mode Rejection
10 M
< 100 pF >120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz
< 100 pF > 60 dB, dc to 60 Hz
(ac-coupled)
Open Circuit
oltage
est V
T
Full Scale Voltage Typical Short Circuit Current
o 6.0 M
T
50 Mor 6
0 nS
00
6
evel
6 k
6
50 M
00 k 6 M
6
0 k
MIN MAX recording specifications
Model Nominal Response Accuracy
00 ms to 8
83V
1
87V 100 ms to 80 % Specified accuracy ± 12 counts for changes > 200 ms in duration
(dc functions) > 25 % of range 120 ms to 80 % Specified accuracy ± 40 counts for changes > 350 ms and inputs
(ac functions) 250
(Model 8
1
For repetitive peaks: 1 ms for single events.
0 % Specified accuracy ± 12 counts for changes > 200 ms in duration
ounts in ac with beeper on)
40 c
±
(
s (peak)
µ
7 only)
Specified accuracy ± 100 counts for changes > 250 µs in duration
1
100 counts for readings over 6000 counts)
±
(add
100 counts for readings in Low Pass mode)
±
(add
Fluke 83V and 87V detailed specifications Fluke Corporation 3
Fluke 83V and 87V General Specifications
Maximum voltage betw Fuse protection f
een any terminal and earth ground:1000 V rms
or mA or µA inputs:
100 A, 1000 V FAST Fuse
44/
Fuse protection for A input: 11 A, 1000 V FAST Fuse Display:
Digital: 6000 counts updates 4/sec; (Model 87V also has 19,999 counts in high-resolution mode) Analog: 33 segments, updates 40/sec. Frequency: 19,999 counts, updates 3/sec at > 10 Hz
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C Altitude:
Operating: 2000 m Storage: 10,000 m
Temperature coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C) Electromagnetic compatibility: In an RF field of 3 V/m total accuracy = specified accuracy Relative humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 7 0% (35 °C to 55 °C) Batter
y type:
9 V zinc, N
EDA 1604 or 6F22 or 006P
Battery life: 400 hours typical with alkaline (with backlight off) Vibration: Per MIL-PRF-28800 for a Class 2 instrument Shock: 1 Meter drop per IEC 61010-1:2001 Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm) Size with holster and flex-stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm) Weight: 12.5 oz (355 g) Weight with holster and flex-stand: 22.0 oz (624 g) Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category
III, IEC 664 to 600 V Overvoltage Category IV. UL listed to UL3111-1. Licensed by TÜV to EN61010-1.
4 Fluke Corporation Fluke 83V and 87V detailed specifications
Fluke. Keeping your world
up and running.
Fluke Corporation
P
O Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
ax (425) 446-5
F In Europe/
1 40) 2 6
ax (3
F In Canada (8 Fax (905) 890-6866 From other countries +1 (425) 446-5500 or
ax +1 (425) 446-5
F Web access: http://www.fluke.com/
©2005 Fluke Corporation. All rights reserved. Printed in U
6
1
1
M-East/Africa (31 40) 2 675 200 or
75 222
00)-36-
.S.A. 4/2005 2161164 D-US-N Rev B
116
E or
K
LU
F
Loading...