Controller/Calibrator
Technical Data
Features
• Pressure ranges from 0 to 5 and 0 to  
3 000 psi (0 to 35 kPa and 0 to 21 MPa)
• Model 7250xi and 7250i provide advanced 
precision of 0.005 % of reading
• Model 7250 provides 0.003 % of full scale 
precision
• Stability: 0.0075 % of reading per year
• Time to setpoint: 20 seconds with no 
overshoot
• Control stability: 10 ppm
• Active matrix color screen with enhanced 
navigation menus
• Languages: English, French, Chinese, 
German, Japanese, Spanish and Italian
The Series 7250 high speed digital pressure 
controllers represent the fifth generation of 
automatic pressure controllers from Fluke Calibration and offers unmatched performance, an 
active matrix color screen and enhanced control 
stability. Utilizing multiple sensors and ranges in 
a single instrument, the Series 7250 combines 
precision, stability, speed and affordability. 
Models 7250xi, 7250i and 7250, up to  
2 500 psi (17.2 MPa), use a unique quartz 
sensor, the most accurate pressure sensing 
technology in a pressure calibrator. Each quartz 
sensor is manufactured and tested to provide 
the ultimate performance required by a Fluke 
Calibration pressure calibrator, ensuring that 
every customer receives quality, precision and 
stability in their instrument.
Advanced precision
Models 7250xi and 7250i offer advanced percent of reading precision for increased capability 
with a single instrument, reducing the investment required to calibrate a wide variety of 
pressure devices and ranges. The Model 7250xi 
provides 0.005 % of reading precision from 
5 % to 100 % of the instrument’s range. This 
unmatched precision is achieved through unique 
quartz pressure sensing technology and multiple quartz sensors in a single instrument. The 
7250xi is available with any full scale range 
from 20 to 2 500 psi (140 kPa to 17.2 MPa). The 
Model 7250i provides 0.005 % of reading from 
25 % to 100 % of range and is available with 
any full scale range from 5 to 2 500 psi  
(35 kPa to 17.2 MPa).
For pressure below the lower threshold of 
5% for the 7250xi and 25 % for the 7250i, the 
precision becomes 0.005 % of the lower value. 
For example, a 1 000 psi (7 MPa) Model 7250xi 
provides 0.005 % of reading from 50 to 1 000 
psi (35 kPa to 7 MPa); the precision for pressures 
from 0 to 50 psi (0 to 35 kPa) is 0.005 % of  
50 psi (35 kPa).
Selecting the full scale range of the 7250xi 
and 7250i to purchase is simplified with the 
percent of reading capability. Simply determine 
the highest pressure required and then decide if 
0.005 % of reading is required down to 25 % or 
5 % to maintain the appropriate calibration ratio. 
The complex analysis associated with selecting 
the appropriate single or multirange instruments 
with percent of full scale performance is no 

longer required. And, since the performance of 
the 7250xi and 7250i is continuous throughout 
the range, the time consuming task of controlling to atmosphere and switching ranges is 
eliminated.
The 7250xi and 7250i not only provide 
unequalled precision, but also excellent long 
term stability due to the inherent properties of 
quartz in combination with several enhancements implemented over the last few years. The 
total uncertainty of the 7250xi and 7250i over 
a one year calibration interval is 0.009 % of 
reading down to 5 % and 25 %, respectively (see 
the table, above right). Simplified specifications 
eliminate guesswork and “specmanship” to 
allow you to determine the actual performance 
delivered with every instrument.
Standard precision
For applications that do not require the level of 
performance provided in the 7250xi or 7250i, 
the Model 7250 offers an economical approach 
to automatic pressure testing and calibration 
with a precision of 0.003 % of full scale for 
ranges to 2 500 psi (17.2 MPa). 
High pressure precision
For applications that require pressures up to  
3 000 psi (21 MPa), the Model 7250HP provides 
total uncertainty (one year calibration interval) 
of 0.013% of reading down to 30% of the full 
scale. The 7250HP provides the high speed 
control and performance of the standard 7250 
with a full scale pressure range of 3 000 psi  
(21 MPa) absolute.
The Series 7250 features a unique fusedquartz sensor. This rugged transducer offers 
unequalled precision and a stability of 
0.0075 % of reading per year.
High-speed pressure 
control
All Series 7250 instruments 
up to 2 500 psi (172 bar) reach 
setpoint in 20 seconds or less 
into a 15 cubic inch (245 cc) 
volume, with no overshoot, to 
allow high-speed pressure test 
and calibration.
Dual control modes
The Series 7250 also provides 
two user-selectable control 
modes: active and passive 
mode. In active mode, the 7250 is continually 
maintaining the setpoint and can compensate 
for small leaks and pressure changes due to 
temperature. In passive mode, the user defines 
a control band and the 7250 will turn off the 
controller once it achieves the setpoint within 
the control band. In a leak-free and temperature-stable system, passive mode contributes 
no additional uncertainty, providing optimum 
performance.
Automating pressure test and  
calibration
The 7250xi, 7250i and 7250 are easy to use and 
can automate your calibrations in several ways:
Step up/down: for calibrations where the 
increments are fixed intervals, enter a userdefined step value. The Series 7250 increases 
or decreases the pressure by the step amount 
with the jog dial—no more lengthy keystroke 
sequences to program.
Sweep test: for simple exercising routines, 
as with dial gauges, enter a start value, a stop 
value, and number of times to repeat the cycle. 
The Series 7250 will automatically exercise the 
device under test prior to the calibration run.
Onboard programs: For frequently used or 
lengthy calibrations, the Series 7250 can store 
up to 20 user-defined programs/profiles with up 
to 1000 steps total in internal memory.
Computer interface: Every Series 7250 is 
provided with both an RS-232 and IEEE-488 
interface, and Series 7250 syntax follows SCPI 
protocol for easy programming. COMPASS®, 
an off-the-shelf software package is available in addition to the LabVIEW® driver, a free 
download and optional MET/CAL® driver. As 
a standard feature, software written for the 
previous generation Series 7215, 7010 and 6000 
instruments is fully supported by the Series 
7250. The Series 7250 can also be set to 510 
emulation mode to use software originally written for the DPI 510. Firmware updates can also 
be performed over the RS-232 interface.
Versatility to handle any pneumatic 
pressure calibration
The Series 7250 is versatile enough to handle 
almost any type of pneumatic pressure 
calibration.
Wide pressure range: available with any full 
scale pressure between 5 and 2 500 psi (35 kPa 
and 17.2 MPa) or for higher pressures select the 
3 000 psi absolute (21 MPa) range.
2  Fluke Calibration   Series 7250 Ruska Pressure Controller/Calibrator

Pressure units/scales: Select from over 
twelve standard units of measure, including 
inHg at 0 °C and 60 °F, kPa, bar, psi, inH2O at  
4 °C, 20 °C and 60 °F, kg/cm2, mmHg at 0 °C, 
cmHg at 0 °C, and cmH2O at 4 °C, and two userdefined units.
Head pressure: Automatic correction for head 
pressure differences
Absolute mode: The 7250i and 7250 offer 
three different methods to make absolute pressure measurements. The barometric reference 
option provides the most convenient method 
and is available on ranges 15 psi (1.0 bar) and 
higher. Alternatively, the vacuum reference 
option allows the connection of an external 
vacuum pump to the reference port of the 
instrument. An onboard vacuum sensor monitors 
the reference vacuum and allows for automatic 
zeroing in absolute mode. This option provides 
the lowest overall uncertainty since it does not 
include the additional uncertainty of a secondary barometric reference sensor. For pressure to 
50 psia (350 kPa) permanent absolute models 
are also available. The 7250xi is only available 
with the barometric reference option for absolute measurement.
Autovent and autozero: With few keystrokes, 
the Series 7250 will vent the test port to atmosphere or automatically zero itself (autovent is 
not applicable to permanent absolute models).
Protection of the device under test: Set 
upper and lower pressure limits to ensure  
protection of the device under test.
The large color display allows the pressure value to be displayed even 
when viewing a submenu selection such as the units selection screen 
shown above.
Options
The Series 7250 can be provided for gauge 
mode operation, or with:
• Optional vacuum (negative gauge) mode for 
bidirectional devices
• Optional barometric reference for absolute 
mode calibrations
• Permanent absolute ranges to 50 psia (350 
kPa) full scale which include a tare feature for 
simulated gauge mode operation
The Series 7250 Pressure Controller/Calibrator 
can easily automate your test and calibration 
workload. All are easy to use, easy to maintain, and have the reliability, performance and 
features that you want. The Series 7250 features 
an easy-to-navigate menu structure with full 
text descriptions for menus and commands.
3  Fluke Calibration   Series 7250 Ruska Pressure Controller/Calibrator
All Series 7250s are fully programmable and can store up to 20 
separate programs with up to 1000 steps.

Specications
General
Temperature Operating: 18 °C to 36 °C (64 °F to 96 °F)
Storage: -20 °C to 70 °C (-4 °F to 158 °F) 
Humidity 5 % to 95 % RH, non-condensing 
Weight Model 7250/7250i: 7.7 kg (17 lb)
Model 7250xi: 9 kg (20 lb) 
Dimensions (H x W x D) All versions: 178 mm x 419 mm x 483 mm (7 in x 16.5 in x 19 in) 
Pressure medium Nitrogen or clean dry air 
Display TFT, VGA, active matrix, 162.5 mm (6.4 in) 640 x 480 resolution, 65,000 colors 
Test port and supply 
connection
1/4 in NPT female
Warm up time 2 to 3 hours; may be left on indefinitely
Pressure ranges
Model 7250xi Any Full Scale pressure range from 20 psig to 2 500 psig (140 kPa to 
Model 7250i Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa). 
Model 7250 Any Full Scale pressure range from 5 psig to 2 500 psig (35 kPa to 17.2 MPa). 
Optional modes Absolute using barometric reference sensor for ranges from 15 psig to 
17.2 MPa).
Or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa 
absolute).
Or, Permanent Absolute ranges from 15 psia to 50 psia (100 kPa to 350 kPa 
absolute). For ranges below 35 kPa (5 psi), see the 7250LP.
2 500 psig (100 kPa to 17.2 MPa)
Vacuum (negative gauge) 
Vacuum reference (requires external vacuum pump) 
Model 7250HP 3000 psi absolute (21 MPa absolute)
Performance 
1
Precision 
Model 7250xi: From 5 % to 100 % full scale (FS): 0.005 % of reading
Below 5 % FS: 0.005% of 5% FS
Model 7250i: From 25 % to 100 % FS: 0.005 % of reading
Below 25 % FS: 0.005 % of 25 % FS
Model 7250: Ranges to 2 500 psi (17.2 MPa): 0.003 % of FS
Model 7250HP: 0.012% of reading or 0.0036% FS, whichever is greater 
Stability Ranges 2 500 psi (17.2 MPa): 0.0075 % of reading per year
Control stability Active mode: 0.001 % FS (10 ppm)
Display resolution User selectable to 1:1,000,000 
Control response 20 seconds or less with zero overshoot into a 245 cm
3 000 psia (21 MPa) range: 0.005% of reading
Passive mode: no additional uncertainty
to 2 500 psi (17.2 MPa)
(15 in3) volume for ranges 
60 seconds for 3 000 psia (21 MPa) range 
Negative gauge 
precision (optional)
Model 7250xi: 0.005 % of 5 % FS or 0.00075 psi (0.05 mbar), whichever is 
greater
Model 7250i: 0.005 % of 25 % FS or 0.0075 psi (0.5 mbar), whichever is 
greater
Model 7250: 0.003 % of FS 
Model 7250HP: Not applicable 
Barometric reference 
(optional) 
Vacuum reference 
(optional)
0.002 psi (0.014 kPa) maximum error per year
0.0002 psi (0.0014 kPa) maximum error per year (Only available on 7250i 
and 7250)
Calibration
An accredited calibration report is provided standard. For information on the calibration uncertainty, 
please see the appropriate Fluke Calibration Scope of Accreditation.
Total Uncertainty (k=2)
Model 7250xi 5 % to 100 % FS
Model 7250i 25 % to 100 % FS
Model 7250 90 day: 0.003 % FS + 0.002 % reading
Model 7250HP 30 % to 100 % FS
Control parameters
Volume 82 cm to 980 cm (5 in 
Low Control (ab-
solute)
2
90 day: 0.006 % reading 
1 year: 0.009 % reading
90 day: 0.006 % reading 
1 year:0.009 % reading
1 year: 0.003 % FS + 0.0075 % reading
90 day: 0.012% reading 
1 year: 0.013% reading
3
 to 60 in3)
0.15 psia (1 kPa) 
Communications
RS-232 and IEEE-488, SCPI syntax. Series 7215, Model 7010, Series 
6000 and DPI 510 emulation are standard.
®
MET/CAL
 driver Optional
®
LabView
 driver Optional
Firmware updates are performed via RS-232 interface
Languages
The Series 7050 is capable of displaying menus and functions in:  
English, French, Chinese, German, Japanese, Spanish and Italian
Options
Barometric reference (absolute and vacuum) 
Vacuum mode (negative gauge) 
Vacuum reference (requires external vacuum pump)
Accessories
3911197 Transportation Case with Wheels and Handles 
3879677 Rack mount kit for 19 in. EIA cabinets, 24 in. deep 
3879689 Rack mount kit for 19 in. EIA cabinets, 30 in. deep 
3876298 Lines and Fittings Kit
4014713 Vacuum Pump with auto-vent, 85 liters/minute 
3876271 Liquid Trap 
3891089 MET/CAL Pseudo-Driver 
3070175 COMPASS for Pressure Enhanced, Single License
Model 7250i Uncertainty (2 sigma)
Precision 0.005 % of reading 
Stability (one year) 0.0075 % of reading 
Calibration standard 0.0010 % of reading 
Environmental temperature 
(included in precision)
0.000 % of reading
Head pressure 0.001 % of reading 
Expanded uncertainty (2 
sigma)
1
 Precision is defined as the combined effects of linearity, repeatability 
and hysteresis throughout the operating temperature range.
2 
Total uncertainty is the maximum deviation from the true value of 
pressure including precision, stability, temperature effects and the 
calibration standard. This assumes routine zeroing of the instrument.  
Expression of uncertainty conforms with the recommendations of the 
ISO Guide to the Expression of Uncertainty in Measurement.
0.009 % of reading
Fluke Calibration. Precision, performance, confidence.
Fluke Calibration 
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V. 
PO Box 1186, 5602 BD  
Eindhoven, The Netherlands
4  Fluke Calibration   Series 7250 Ruska Pressure Controller/Calibrator
For more information call: 
In the U.S.A. (877) 355-3225 or  
Fax (425) 446-5116 
In Europe/M-East/Africa +31 (0) 40  
2675 200 or Fax +31 (0) 40 2675 222 
In Canada (800)-36-FLUKE or  
Fax (905) 890-6866 
™
From other countries +1 (425) 446-5500 or  
Fax +1 (425) 446-5116 
Web access: http://www.flukecal.com
©2010 Fluke Corporation.  
Specifications subject to change without notice.  
Printed in U.S.A. 9/2011 3833773B D-EN-N