Atec Agilent-ECP, Agilent-E26A User Manual

Agilent E4412A and E4413A Power Sensor
Operating and Service Guide
Agilent Technologies

Notices

CAUTION
WARNING
No p art o f this m anual may b e rep roduc ed in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
E4412-90013
Edition
May 29, 2009
Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95052 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user cus­tomers. Agilent provides this customary commercial license in Software and techni­cal data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Software) and, for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Com­puter Software Documentation).
Safety Notices
A CAUTION notice denotes a haz- ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
ii
Certification
General Warranty
Agilent Technologies certifies that this product met its published specifications at the time of shipment. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standard and Technology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members.
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part of) other Agilent products. During the warranty period, Agilent will, at its option, either repair or replace products which prove to be defective. The warranty period begins on the date of delivery or on the date of installation if installed by Agilent.
Warranty Service
For warranty service or repair, this product must be returned to a service facility designated by Agilent. For products returned to Agilent for warranty service, the Buyer shall prepay shipping charges to Agilent and Agilent shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent from another country.
iii
Limitation of Warranty
Exclusive Remedies
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Agilent does not warrant the Buyer’s circuitry or malfunctions of Agilent products that result from the Buyer’s circuitry. In addition, Agilent does not warrant any damage that occurs as a result of the Buyer’s circuit or any defects that result from Buyer-supplied products.
To the extent allowed by local law, Agilent makes no other warranty, expressed or implied, whether written or oral with respect to this product and specifically disclaims any implied warranty or condition of merchantability, fitness for a particular purpose or satisfactory quality.
To the extent allowed by local law, the remedies provided herein are the Buyer’s sole and exclusive remedies. Agilent shall not be liable for any direct, indirect, special, incidental, or consequential damages (including lost profit or data), whether based on warranty, contract, tort, or any other legal theory.
iv

Restricted Rights Legend

The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227-7013 (Oct 1988), DFARS
252.211-7015 (May 1991), or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR 2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun 1987) (or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Agilent standard software agreement for the product involved.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
v

Safety Summary

WARNING
CAUTION
Safety Notices
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
vi

Safety Symbols

ESD
The following symbol on the instrument and in the documentation indicates precautions that must be taken to maintain safe operation of the instrument.
Caution, risk of danger. The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied docu­mentation.
This symbol indicates that a device, or part of a device, may be susceptible to electrostatic discharges (ESD) which can result in damage to the product. Observe ESD precautions given on the product, or its user documentation, when handling equipment bearing this mark.
Alternating current (AC).
Direct current (DC).
Both direct and alternating current.
Three-phase alternating current.
Earth (ground) TERMINAL.
vii
PROTECTIVE CONDUCTOR TERMINAL.
Frame or chasis TERMINAL.
Equipotentiality.
On (Supply).
Off (Supply).
Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION.
Caution, risk of electric shock.
Caution, hot surface.
In position of bi-stable push control.
viii
Out position of bi-stable push control.
ix

Regulatory Markings

ISM 1- A
N10149
ICES/NMB - 001
The CE mark shows that the product complies with all the relevant European legal Directives (if accompanied by a year, it signifies when the design was proven).
The C-tick mark is a registered trademark of the Spectrum management Agency of Australia. This signifies compliance with the Australian EMC Framework regu­lations under the terms of the Radio Communications Act of 1992.
This product complies with the WEEE Directive (2002/96/EC) marking equip­ment. The affixed product label indicates that you must not discard this electri­cal/electronic product in domestic household waste.
This ISM device complies with the Canadian ICES-001,
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
x

General Safety Information

WARNING
CAUTION
CAUTION
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
BEFORE CONNECTING THE POWER SENSOR TO OTHER INSTRUMENTS ensure that all instruments are connected to the protective (earth) ground. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in personal injury.
Use the device with the cables provided
Repair or service that is not covered in this manual should only be
This product is designed for use in Installation category II and Pollution
This instrument has been designed and tested in accordance with IEC
Notice for germany: Noise Declaration L
performed by qualified personnels.
Degree 2.
Publication 248, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
r
A < 70 dBm am Arbeitsplatz (operator position) mormaler Betrieb (normal position) nach DIN 45635 T.19 (per ISO 7779).
xi

In This Guide ...

1 Operation and Service Guide This chapter contains information about
initial inspection, performance tests, specifications, operations, troubleshooting and service of the Agilent E4412A and E4413A power sensors.
2
Appendix
and E4413A sensors (formerly EXCP-E18A and ECP-E26A, respectively)
with a serial prefix below US3848/MYxxxxxxxx/SGxxxxxxxx.
This Appendix contains specifications that applies to E4412A
xii

Contents

Notices ii
General Warranty iii Warranty Service iii Limitation of Warranty iv Exclusive Remedies iv Technology Licenses v Safety Summary vi Safety Notices vi Safety Symbols vii Regulatory Markings x General Safety Information xi In This Guide ... xii
1 Operation and Service Guide
General Information 3 Performance Test 15 Replaceable Parts 17 Service 20 Warranty 3 Instruments Covered by Manual 3 Description 3 Specifications 4 Calibration Factor (CF) and Reflection Coefficient (Rho) 8 Installation 11 Interconnections 11 Recommended Calibration Interval 11 Storage and Shipment 12 Operation 12 Power Meter Calibrations 14 Operating Instructions 14 Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)Performance Test 15 Principles of Operation 20 Troubleshooting 20 Repair of Defective Sensor 21
xiii
Cleaning 21 Disassembly Procedure 22 Reassembly Procedure 23 Agilent Sales and Service Offices 24
A Appendix
E4412A and E4413A Specifications (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-2
E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-4
E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-5
xiv
List of Figures
Figure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP-E26A,
respectively) 2
Figure 2 Relative Mode Power Measurement Linearity with Power Meter/Sensor at
25 ºC (Typical) 7
Figure 3 N8485A power sensor with adapter 14 Figure 4 Illustrated Parts Break down 18 Figure 5 Removing Power Sensor Shell 22
xv
List of Tables
Ta b l e 1 E4412A and E4413A Specifications A-5 Ta b l e 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) A-9 Ta b l e 3 E4413A Calibration Factor Uncertainty at 1 mW (0 dBm) A-10 Ta b l e 4 Power Sensor SWR and Reflection Coefficient for the E4412Aa A-15 Ta b l e 5 Power Sensor SWR and Reflection Coef ficient for the E4413Aa A-16 Ta b l e 5 Power Sensor SWR and Reflection Coef ficient for the E4413Aa A-16 Ta b l e 6 Replaceable Parts A-19 Ta b l e 7 Agilent Sales and Service Offices A-24 Ta b l e A - 1 E4412A and E4413A Specifications (Serial Prefixes Below
US3848/MYxxxxxxxx/SGxxxxxxxx) A-2
Ta b l e A - 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes
Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-4
Ta b l e A - 3 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes
Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-5
xvi
Agilent E4412A and E4413A Power Sensors Operating and Service Guide
1 Operation and Service Guide
General Information 3 Performance Test 15 Replaceable Parts 17 Service 20
This chapter contains information about initial inspection, performance tests, specifications, operations, troubleshooting and service of the Agilent E4412A and E4413A power sensors.
Agilent Technologies
1
1Introduction
Figure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP-E26A,
respectively)
2 Agilent E4412A and E4413A Power Sensors Operating and Service Guide

General Information

Warranty

The power sensors are warranted and certified as indicated on the inside cover of this manual.

Instruments Covered by Manual

These instruments have a two- part serial number: the prefix (two letters and
the first four numbers), and the suffix (the last four numbers). The two letters
identify the country in which the unit was manufactured. The four numbers
of the prefix are a code identifying the date of the last major design change
incorporated in your Agilent Technologies product. The four- digit suffix is a
sequential number and, coupled with the prefix, provides a unique
identification for each unit produced. The contents of this manual apply
directly to all serial numbers unless otherwise indicated.
Introduction 1

Description

The E4412A and E4413A power sensors are diode power sensors. They are
intended for measurement of CW microwave power levels in a wide
dynamic range from - 70 dBm to +20 dBm (100 pW to 100 mW). The
E4412A measures at frequencies from 10 MHz to 18.0 GHz. The E4413A
measures at frequencies from 50 MHz to 26.5 GHz. These are high- speed
power sensors, and do not incorporate narrow- bandwidth averaging used in
average- power sensors. Signals with digital, pulse, or other forms of
amplitude modulation may introduce measurement errors. Multi- tone signals
(containing multiple frequency components), or signals with significant
harmonic content (> - 45 dBc) may introduce measurement errors at high
power levels. (Specifications for the power sensors are in Table 1).
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 3
1Introduction
NOTE
NOTE

Specifications

These power sensors measure CW power, displayed on a compatible power
meter in logarithmic (dBm or dB) or linear (Watts or %) measurement units.
The E4413A is shipped with a 3.5- mm to Type- N adapter, part number
08485- 60005.
The E4412A and E4413A power sensors are compatible ONLY with the newer E44XX-Series power meters. They are NOT compatible with the earlier 430-Series, E1416A, or 70100A power meters.
The E4412A and E4413A power sensors are extremely static- sensitive. Do not open the power sensor unless you and the power sensor are at a static-free workstation.
The specifications listed in Table 1 are the performance standards or limits
against which the power sensor may be tested. These specifications are valid
ONLY after proper calibration of the power meter. Refer to the Calibration
Procedure Using Agilent E- Series Power Sensors in the E4418A or the
Agilent E4419A Power Meter User’s Guide.
4 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Ta b le 1 E4412A and E4413A Specifications
Rho
0.099
0.070
0.078
0.091
0.119
20.08
23.13
22.12
20.82
18.42
0.095
0.087
0.095
0.115
20.45
21.23
20.45
18.79
Limit Comments
Frequency Range E4412A: 10 MHz to 18 GHz
E4413A: 50 MHz to 26.5 GHz
Power Range –70 dBm to +20 dBm (100 pW to
100 mW)
Impedance 50 ohm nominal
Connector Type E4412A: Type-N (male)
E4413A: 3.5 mm (male)
Introduction 1
Maximum Standing Wave Ratio (SWR) and Reflection Coefficient
b
(Rho)
E4412A
10 MHz to <30 MHz
30 MHz to <2 GHz
2 GHz to <6 GHz
6 GHz to <11 GHz
11 GHz to <18 GHz
E4413A
50 MHz to <100 MHz
100 MHz to <8 GHz
8 GHz to <18 GHz
18 GHz to 26.5 GHz
Maximum Power 200 mW peak (+23 dBm)
AC Coupled Input max: 20 VDC
Zero Set ±50 pW
SWR Return Loss (dB)
1.22
1.15
1.17
1.20
1.27
1.21
1.19
1.21
1.26
200 mW average (+23 dBm)
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 5
1Introduction
0 to 55°C
±7%
±10%
–70 dBm to +10 dBm
+10 dBm to +20 dBm
Limit Comments
Power Linearity
100 pW to 10 mW
10 mW to 100 mW
Operating Temperature 0 to 55°C
Net Weight
E4412A
E4413A
Dimensions
E4412A
E4413A
a. Limits are in percent of power in Watts. See Figure 2 for relative power measurement.
b. Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A.
ab
25 ± 5°C
±3%
±4.5%
0.47 kg (1.04 lb)
0.45 kg (1.00 lb)
Length: 130 mm (5.1 in)
Length: 102 mm (4.0 in)
Width: 38 mm (1.5 in)
Height: 30 mm (1.2 in)
(After calibration at 0 dBm at ambient temperature)
Same for both models
Same for both models
6 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction 1
±6 %
±3 %
±1 %
+20 +10
+6
0
–20
–35
–70
–35
–70
–20
–25 +20
0
A
B
Power L evel Being M easured (dB m)
Power Level Used as Reference (dBm)
Figure 2 Relative Mode Power Measurement Linearity with Power Meter/Sensor at
25 ºC (Typical)
The chart in Figure 2 shows the typical uncertainty in making relative power measurements with the E44XX- Series power meter in the Rel (relative) mode. In a relative measurement, two power levels are compared using a single sensor. The typical measurement uncertainty can be found by drawing a vertical line at a power level used as a reference, and a horizontal line at the power being measured. The region where these two lines intersect shows the typical uncertainty. This assumes that the reference power and the measured power are at the same frequency, and neglects errors due to zero set, zero drift, and noise. It also assumes no change in mismatch when measuring the “Power level used as reference” and the “Power level being measured”. This chart illustrates that the best relative power measurement accuracy is obtained when the reference and the measured power levels are equal to, or less than, –20 dBm. Care in choosing the power levels cna yield improved measurement accuracy. Example A illustrates a relative gain (amplifier measurement) and example B illustrates a relative loss (insertion loss measurement).
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 7
1Introduction

Calibration Factor (CF) and Reflection Coefficient (Rho)

EXAMPLE A (Amplifier Gain Measurement):
Input power (reference) = –35 dBm; power measured = +6 dBm’ so the relative gain is +41 dB. the chart indicates a typical error of ±6 %, which corresponds to +0.25 dB/–0.27 dB uncertainty in the gain measurement.
EXAMPLE B (Insertion Loss Measurement):
Reference power = –25 dBm; power measured = –35 dBm, for the case of an insertion loss of 10 dB. the chart indicates a typical error of ±1 %, which corresponds to ±0.04 dBm uncertainty in the measurement.
Calibration factor and reflection coefficient data are given at 1 GHz increments on a data sheet included with the power sensor. This data is unique to each sensor. If you have more than one sensor, match the serial number on the data sheet with the serial number on the power sensor you are using. The CF corrects for the frequency response of the sensor. The power meter automatically reads the CF data stored in the sensor and uses it to make the corrections.
Reflection Coefficie form ula:
SWR = (1+
Typical uncertainties of the CF data are listed in Table 2 for the E4412A
power sensor, and in Table 3 for the E4413A power sensor. The uncertainty
analysis for the calibration of the sensors was done in accordance with the
ISO/TAG4 Guide. The uncertainty data reported on the calibration certificate
is the expanded uncertainty with 95% confidence level and a coverage factor of
2.
Additionally, at frequencies other than the reference frequency, for measure-
ments above 0 dBm include 0.5 %/dB high power calibration factor uncer-
tainty.
8 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
nt (Rho, or
r
)/(1-r)
r
) relates to SWR according to the following
Ta b le 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm)
Introduction 1
Frequency Uncertainty (%)
10 MHz 1.8
30 MHz 1.8
50 MHz Reference
100 MHz 1.8
1.0 GHz 1.8
2.0 GHz 2.4
4.0 GHz 2.4
6.0 GHz 2.4
8.0 GHz 2.4
10.0 GHz 2.4
11.0 GHz 2.4
12.0 GHz 2.4
14.0 GHz 2.4
16.0 GHz 2.6
18.0 GHz 2.6
a
a. Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A.
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 9
1Introduction
Ta b le 3 E4413A Calibration Factor Uncertainty at 1 mW (0 dBm)
Frequency Uncertainty (%)
50 MHz Reference
100 MHz 1.8
1.0 GHz 1.8
2.0 GHz 2.4
4.0 GHz 2.4
6.0 GHz 2.4
8.0 GHz 2.4
10.0 GHz 2.6
11.0 GHz 2.6
12.0 GHz 2.8
14.0 GHz 2.8
16.0 GHz 2.8
17.0 GHz 2.8
18.0 GHz 2.8
19.0 GHz 3.0
a
20.0 GHz 3.0
22.0 GHz 3.0
24.0 GHz 3.0
26.0 GHz 3.0
26.5GHz 3.0
a. Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A.
10 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction 1

Installation

Initial Inspection
Inspect the shipping container for damage. If the shipping container or packaging material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is mechanical damage or if the instrument does not pass the performance tests, notify the nearest Agilent Technologies office. Keep the damaged shipping materials (if any) for inspection by the carrier and a Agilent Technologies representative.

Interconnections

Connect one end of the 11730A sensor cable to the E4412A or E4413A power sensor and co nnect the other end of the cable to the power meter’s channel input. Allow a few seconds for the power meter to download the power sensor’s calibration table before making a measurement.
Measurement connector (connects to DUT)
E4412A: T ype- N (male)
E4413A: 3.5- mm (male)
A torque wrench should be used to tighten these connectors. Use a 3/4- inch open- end wrench and torque to 12 in- lb (135 Ncm) for the Type- N connector Use a 20- mm open- end wrench and torque to 8 in- lb (90 Ncm) for the 3.5- mm connector.

Recommended Calibration Interval

Agilent Technologies recommends a one- year calibration cycle for the E4412A and E4413A power sensors.
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 11
1Introduction

Storage and Shipment

Environment
The instruments should be stored in a clean, dry environment. The following limitations apply to both storage and shipment:
Temperature –55 to +75 ºC
Relative Humidity <95 % at 40 ºC
Altitude <15,240 metres (50,000 feet)
Original Packaging
Containers and materials identical to those used in factory packaging are available through Agilent Technologies offices. If the instrument is being returned to Agilent Technologies for servicing, attach a tag indicating the type of service required, return address, model number, and serial number.
Also, mark the container FRAGILE to assure careful handling. In any orrespondence, refer to the instrument by model number and serial number.

Operation

Operating Environment
The operating environment for the power sensor should be within the following limits:
Temperature 0 to 55 ºC
Relative Humidity <95 %
Altitude <4,530 metres (15,000 feet)
12 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction 1
WARNING
Operation Precautions
BEFORE CONNECTING THE POWER SENSOR TO OTHER INSTRUMENTS, ensure that all instruments are connected to the protective (earth) ground. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in personal injury and cause damage to the power sensor.
If the following energy and power levels are exceeded, the power meter system may be damaged.
a Maximum Average Power: 200 mW (+23 dBm)
b Maximum Peak Power: 200 mW (+23 dBm)
Maximum torque at the connector should not exceed 12 in- lb (135 Ncm) for the Type- N connector, or 8 in- lb (90 Ncm) for the 3.5- mm connector to avoid damage to the connector.
Connect the power sensor by turning only the hex nut portion of the connector. Damage can occur if torque is applied to the power sensor body.
The connector plastic insulator bead deteriorates when contacted by acetone, trichloroethylene, carbon tetrachloride, benzene, etc. Refer to Application Note 326, Principals of Microwave Connector Care (5954- 1566) or Microwave Connector Care (08510- 90064) for proper cleaning methods.
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 13
1Introduction
NOTE

Power Meter Calibrations

Follow the calibration procedures given in your power meter manual. The E4413A power sensor is fitted with 3.5 mm (m) connectors as standard. To convert the 3.5 mm (m) connector for calibration, an adapter (3.5 mm (f) to Type- N (m)) is included with the power sensor. See Figure 3.
Figure 3
The 3.5 mm to Type-N adapter is intended for the use of 1 mW, 50 MHz power reference of the power meter only. Its function as a calibration reference may be compromised if it is used for other purpose.
E4413A power sensor with adapter

Operating Instructions

The E4412A and E4413A power sensors are compatible ONLY with the newer E44XX- Series power meters. They are NOT compatible with the earlier 430- Series, E1416A, or 70100A power meters. To operate the power sensor, refer to the operating instructions in the Agilent E44XX- Series Power Meter User’s Guide.
14 Agilent E4412A and E4413A Power Sensors Operating and Service Guide

Performance Test

Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)Performance Test

This section does not establish preset SWR test procedures since there are several test methods and different equipm ent available for testing the SWR or reflection coefficient. Therefore, the actual accuracy of the test equipment must be accounted for when measuring against instrument specifications to determine a pass or fail condition. The test system used must not exceed the system Rho uncertainties shown in Table 4 when testing the E4412A, or in
Table 5 when testing the E4413A
Introduction 1
Ta b le 4 Power Sensor SWR and Reflection Coefficient for the E4412A
Frequency System Rho
Uncertainty
10 MHz to < 30 MHz ± 0.010 0.099
30 MHz to < 2 GHz ± 0.010 0.070
2 GHz to < 6 MHz ± 0.010 0.078
6 GHz to < 11 GHz ± 0.010 0.091
11 GHz to 18 GHZ ± 0.010 0.119
a.Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A.
Actual Measurement
a
Maximum Rho
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 15
1Introduction
Ta b le 5 Power Sensor SWR and Reflection Coef ficient for the E4413A
Frequency System Rho
Uncertainty
50 MHz to < 100 MHz ± 0.010 0.095
100 MHz to < 8GHz ± 0.010 0.087
8 GHz to < 18MHz ± 0.010 0.095
18 GHz to < 26.5 GHz ± 0.015 0.115
a.Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A
Actual Measurement
a
Maximum Rho
16 Agilent E4412A and E4413A Power Sensors Operating and Service Guide

Replaceable Parts

NOTE
Table 6 is a list of replaceable parts. Figure 4 is the illustrated parts breakdown (IPB) that identifies all of the replaceable parts. To order a part, quote the Agilent Technologies part number, specify the quantity required, and address the order to the nearest Agilent Technologies office.
Within the USA, it is better to order directly from the Agilent Parts Center in Roseville, California. Ask your nearest Agilent office for information and forms for the “Direct Mail Order System.” Also your nearest Agilent office can supply toll free telephone numbers for ordering parts and supplies.
Introduction 1
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 17
1Introduction
Figure 4 Illustrated Parts Break down
18 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Ta b le 6 Replaceable Parts
Introduction 1
Reference
Part Number Qty Description
Designation
A1/A2 E4412A E4412-60006 1 SENSOR MODULE (E4412A)
A1/A2 E4412A E4412-69006 1 RESTORED SENSOR MODULE (E4412A)
A1/A2 E4413A E4413-60003 1 SENSOR MODULE (E4413A)
A1/A2 E4413A E4413-69003 1 RESTORED SENSOR MODULE (E4413A)
08485-60005 1 ADAPTOR, 3.5-mm to Type-N
CHASSIS PARTS
MP1 5041-9160 1 SHELL PLASTIC
MP2 5041-9160 1 SHELL PLASTIC
MP3 08481-20011 1 CHASSIS
MP4 08481-20011 1 CHASSIS
MP8 08481-00002 1 SHIELD
MP9 08481-00002 1 SHIELD
MP26 E4412-80002 1 LABEL, ID E4412A
MP26 E4413-80002 1 LABEL, ID E4413A
MP27 7121-7389 1 LABEL, POWER SENSOR
MP30 7121-7388 1 LABEL, CAL/ESD
MP31 00346-80011 1 LABEL, CAUTION
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 19
1Introduction

Service

Principles of Operation

Service instructions consist of principles of operation, troubleshooting, and repairs.
The A1 Bulkhead assembly provides a 50 ohm load to the RF signal applied to the power sensor. A diode assembly in the bulkhead rectifies the applied RF to produce a dc voltage which varies with the RF power across the 50 ohm load. Thus the voltage varies with the RF power dissipated in the load. With maximum specified RF power (100 mW) the dc voltage is approximately 1V.
The low- level dc voltage from the bulkhead assembly must be amplified before it can be transferred on standard cables to the power meter. The amplification is provided by an input amplifier assembly which consists of a chopper (sampling gate) and an input amplifier. The chopper circuit converts the dc voltage to an ac voltage. To do this, the chopper uses two field effect transistors (FETs), A2Q1 and A2Q2, controlled by a 440 Hz square wave generated by the power meter. The amplitude of the sampling gate output (drain of A2Q1, sou rce of A2Q2) is a 44 0 H z square wave which varies with the RF power input. The 440 Hz ac output is applied to the input amplifier A2Q3 which provides the input to the first amplifier stage in the power meter.
The E44XX- Series power meter automatically detects when an E44XX Series power sensor is connected and downloads the correction data from the sensor’s EEPROM. This configures the power meter to operate over the +20
dBm to - 70 dBm power range with that particular sensor’s unique correction
data applied.

Troubleshooting

Troubleshooting information is intended to first isolate the power sensor, the cable, or the power meter as the defective component. When the power sensor is isolated, a “Restored Sensor Module” must be used for repair. See Table 6.
20 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction 1
CAUTION
CAUTION
CAUTION
If error message 241 or 310 is indicated on the power meter, suspect failed
when making a measurement, try replacing the cable from the power meter to the power sensor. If the problem still exists, try using a different power sensor to determine if the problem is in the power meter or in the power sensor.
Electrostatic discharge will render the power sensor inoperative. Do not, under any circumstances, open the power sensor unless you and the power sensor are in a static free environment.
power sensor. If no error message is displayed, but a problem occurs

Repair of Defective Sensor

There are no serviceable parts inside the E44XX- Series sensors. If the sensor is defective, replace the entire “module” with the appropriate “Restored sensor Module.” See Table 6.
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 21

Cleaning

Cleaning Solutions
Keeping in mind its flammable nature; a solution of pure isopropyl or ethyl alcohol can be used to clean the connector.
Connector Cleaning
The RF connector beads deteriorate when contacted by hydrocarbon compounds such as acetone, trichloroethylene, carbon tetrachloride, and benzene.
Clean the connector only at a static free workstation. Electrostatic discharge to the center pin of the connector will render the power sensor inoperative.
1Introduction
CAUTION

Disassembly Procedure

Clean the connector face using a cotton swab dipped in isopropyl alcohol. If the swab is too big use a round wooden toothpick wrapped in a lint free cotton cloth dipped in isopropyl alcohol. Refer to Agilent Application Note 326, Principals of Microwave Connector Care (5954- 1566) or Microwave Connector Care (08510- 90064) for proper cleaning methods.
22 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Figure 5 Removing Power Sensor Shell
Diassemble the power sensor only in a static free workstation. Electrostatic discharge will render the power sensor inoperative.
Disassemble the power sensor by performing the following steps:
1 At the rear of the power sensor, insert the blade of a screwdriver between
the plastic shells (Figure 5). To prevent damage to the plastic shells use a screwdriver blade as wide as the slot between the two shells.
2 Pry alternately at both sides of the connector J1 until the plastic shells are
apart.Remove the shells and the magnetic shields.
Introduction 1

Reassembly Procedure

1 Replace the magnetic shields and the plastic shells as shown in Figure 5.
Snap the plastic shells together.
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 23
1Introduction

Agilent Sales and Service Offices

Ta b le 7 Agilent Sales and Service Offices
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800 (fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Ta i w a n :
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6755 004
or visit Agilent worlwide Web at: www.agilent.com/find/assist
24 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Agilent E4412A and E4413A Power Sensors Operating and Service Guide
A Appendix
E4412A and E4413A Specifications (Serial Prefixes Below
US3848/MYxxxxxxxx/SGxxxxxxxx) 2
E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes
Below US3848/MYxxxxxxxx/SGxxxxxxxx) 4
E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes
Below US3848/MYxxxxxxxx/SGxxxxxxxx) 5
The following specifications apply to E4412A and E4413A sensors (formerly EXCP-E18A and ECP-E26A, respectively) with a serial prefix below US3848/MYxxxxxxxx/SGxxxxxxxx.
Agilent Technologies
A-1
AAppendix
Rho
0.145
0.100
0.120
16.8
20.0
18.4
0.110
0.095
0.115
19.2
20.5
18.8
0 to 55°C
±8%
±11%
–70 dBm to +10 dBm
+10 dBm to +20 dBm
Ta b le A -1 E4412A and E4413A Specifications (Serial Prefixes Below
US3848/MYxxxxxxxx/SGxxxxxxxx)
Limit Comments
Frequency Range E4412A: 10 MHz to 18 GHz
E4413A: 50 MHz to 26.5 GHz
Power Range –70 dBm to +20 dBm (100 pW to
100 mW)
Impedance 50 ohm nominal
Connector Type E4412A: Type-N (male)
E4413A: 3.5 mm (male)
Maximum Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)
E4412A
10 MHz to <30 MHz
30 MHz to <10 GHz
10 GHz to <18 GHz
E4413A
50 MHz to <2 GHz
2 GHz to <18 GHz
18 GHz to <26.5 GHz
Maximum Power 200 mW peak (+23 dBm)
Zero Set ±50 pW
Power Linearity
100 pW to 10 mW
10 mW to 100 mW
a
SWR Return Loss (dB)
1.34
1.22
1.27
1.25
1.21
1.26
200 mW average (+23 dBm)
25 ± 5°C
±4%
±5.5%
(After calibration at 0 dBm at ambient temperature)
A-2 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Appendix A
Limit Comments
Operating Temperature 0 to 55°C
Net Weight
E4412A
E4413A
Dimensions
E4412A
E4413A
a. Limits are in percent of power in Watts. See Figure 2 for relative power measurement.
0.47 kg (1.04 lb)
0.45 kg (1.00 lb)
Length: 130 mm (5.1 in)
Length: 102 mm (4.0 in)
Width: 38 mm (1.5 in)
Height: 30 mm (1.2 in)
Same for both models
Same for both models
Agilent E4412A and E4413A Power Sensors Operating and Service Guide A-3
AAppendix
Ta b le A -2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below
US3848/MYxxxxxxxx/SGxxxxxxxx)
Frequency Uncertainty
(%)
10 MHz 2.7
30 MHz 2.7
50 MHz Reference
100 MHz 2.7
1.0 GHz 2.7
2.0 GHz 3.1
4.0 GHz 3.1
6.0 GHz 3.1
8.0 GHz 3.1
10.0 GHz 3.1
11.0 GHz 3.1
12.0 GHz 3.3
14.0 GHz 3.3
16.0 GHz 3.3
18.0 GHz 3.3
A-4 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Appendix A
Ta b le A -3 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below
US3848/MYxxxxxxxx/SGxxxxxxxx)
Frequency Uncertainty
(%)
50 MHz Reference
100 MHz 2.7
1.0 GHz 2.7
2.0 GHz 3.1
4.0 GHz 3.1
6.0 GHz 3.1
8.0 GHz 3.1
10.0 GHz 3.1
12.0 GHz 3.3
14.0 GHz 3.3
16.0 GHz 3.3
17.0 GHz 3.3
18.0 GHz 3.5
20.0 GHz 3.5
22.0 GHz 3.5
24.0 GHz 3.5
26.0 GHz 3.5
26.5 GHz 3.5
Agilent E4412A and E4413A Power Sensors Operating and Service Guide A-5
AAppendix
A-6 Agilent E4412A and E4413A Power Sensors Operating and Service Guide
www.agilent.com
Contact us
T
o obtain service, warranty or technical support assistance, contact us at the following phone numbers:
United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) 31 20 547 2111 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (080) 769 0800 (fax) (080) 769 0900 Latin America: (tel) (305) 269 7500 Ta i w a n : (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6755 0042
Or visit Agilent worlwide web at: www.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 1999, 2008,
2009
Printed in Malaysia May 29, 2009
E4412-90013
Agilent Technologies
Loading...