These specifications are the performance standards or limits against which the instrument is tested. When shipped from the factory, the E5100A/B meets the specifications listed in this section.
Values followed by (SPC) are supplemental performance characteristics.
Source
Frequency characteristics
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz to 300 MHz
Accuracy (at 23 ±5°C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±20 ppm
with Option 1D5 (at 0 to 55°C, 20 minutes after power on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 ppm
–6
/day (SPC)
with Option E5100x-1D5 (48 hours after power on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2.5 x 10
–9
/8 hours (SPC)
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 mHz
Output power characteristics
(measured at RF OUT 1, RF OUT 2 is terminated with 50 termination)
Range (nominal)
with Option E5100x-001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –9 dBm to +11 dBm
with Option E5100x-002 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –15 dBm to +5 dBm
with Option E5100x-003 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –12 dBm to +8 dBm
with Option E5100x-801 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –48 dBm to +22 dBm
with Option E5100x-802 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –54 dBm to +16 dBm
with Option E5100x-803 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –51 dBm to +19 dBm
with Option E5100x-600 (at RF OUT 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –52 dBm to +18 dBm
with Option E5100x-600 (at RF OUT 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –65 dBm to +5 dBm
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 dB
Level accuracy (at 23 ±5°C, 0 dBm output level, 50 MHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
Flatness (at 23 ±5°C, relative to 0 dBm output level at 50 MHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2 dB, –4 dB
with Option E5100x-803 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2.5 dB, –4.5 dB
Agilent E5100A/B
Network Analyzer
Data Sheet
Option ordering information: To add options to a product, use the following ordering scheme:
Model: E5100x, where x = A or B Example model option: E5100x-Option 001
Note: Options E5100x-600/704/706/708 are for E5100A only.
2
with Option E5100x-801 or E5100x-802
10 kHz freq. < 50 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +1.5 dB, –6 dB (SPC)
50 kHz freq. 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2.5 dB, –4.5 dB
100 MHz < freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +3 dB, –5 dB
with Option E5100x-600
10 kHz freq. < 50 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +1.5 dB, –7 dB (SPC)
50 kHz freq. 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2.5 dB, –4.5 dB
100 MHz < freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +3 dB, –5 dB
Linearity (at 23 ±5°C, relative to 0 dBm output level at 50 MHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
with Option E5100x-801/802/803
maximum power level –70 dB power level < maximum power level –60 dB . . . . . . . . . . . . . . . . . . . . . . ±1.5 dB
maximum power level –60 dB power level maximum power level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
Power splitter
(When the analyzer is equipped with Option E5100x-001 or E5100x-003, delete this section.)
Insertion loss (When the analyzer is equipped with Option E5100x-600, delete this item.) . . . . . . . . . . . 6 dB (nominal)
Output tracking
without Option E5100x-600
10 kHz freq. 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 dB (SPC)
100 MHz < freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2 dB (SPC)
with Option E5100x-600
10 kHz freq. 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 dB ±0.3 dB (SPC)
100 MHz < freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 dB ±0.5 dB (SPC)
Equivalent output SWR
without Option E5100x-600
10 kHz freq. < 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 (SPC)
100 MHz freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 (SPC)
with Option E5100x-600
10 kHz freq. < 50 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 (SPC)
50 kHz freq. 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 (SPC)
100 MHz < freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 (SPC)
Spectral purity characteristics
Harmonics (these are supplemental performance characteristics for the E5100B)
with Option E5100B-001 (at –4 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –35 dBc
with Option E5100B-002 (at –10 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –35 dBc
with Option E5100B-003 (at –7 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –35 dBc
with Option E5100B-600 (at +15 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –20 dBc
with Option E5100B-801 (at +21 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –20 dBc
with Option E5100B-802 (at +15 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –20 dBc
with Option E5100B-803 (at +18 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –20 dBc
Non–harmonic spurious signals (at < 300 MHz)
with Option E5100x-001 (at –4 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
with Option E5100x-002 (at –10 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
with Option E5100x-003 (at –7 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
3
with Option E5100x-600 (at 0 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
with Option E5100x-801 (at +6 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
with Option E5100x-802 (at 0 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
with Option E5100x-803 (at +3 dBm output level) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –45 dBc
Phase noise (at 10 kHz offset from 0 dBm fundamental) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –90 dBc/Hz
Other source information
Reverse power protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 dBm, 25 Vdc (SPC)
Output connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BNC female
Output impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 (nominal)
Receiver
Input characteristics
Frequency range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz to 300 MHz
1 Minput for Options E5100x-703/704/707/708 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz to 5 MHz
IF bandwidth (IF BW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Hz to 30 kHz, 1, 1.5, 2, 3, 4, 5, 8 step (nominal)
Impedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 (nominal)
1 Minput for Option E5100x-703/704/707/708. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 M// 30 pF (nominal)
Return loss (at 50 input)
10 kHz freq. < 100 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 dB (SPC)
100 MHz freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 dB (SPC)
Maximum input level
50 input
Frequency RF attenuator Maximum input level
10 kHz freq. < 200 kHz 25 dB 0 dBm 10 kHz freq. < 200 kHz 0 dB – 25 dBm 200 kHz freq. 300 MHz 25 dB +5 dBm 200 kHz freq. 300 MHz 0 dB – 20 dBm
1 MInput for Options E5100x-705/706/707/708
Frequency
1
RF attenuator Maximum input level
10 kHz freq. < 200 kHz 25 dB 0.22 Vrms 10 kHz freq. < 200 kHz 0 dB 0.013 Vrms 200 kHz freq. 300 MHz 25 dB 0.40 Vrms 200 kHz freq. 300 MHz 0 dB 0.022 Vrms
1. Measurement frequency 5 MHz
4
Damage level
DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Vdc
AC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 dBm
Averaging noise level (at magnitude measurement, 23 ±5°C, RF attenuator: 0 dB, 50 input)
• E5100A
IF BW 30 kHz (at > 1 MHz). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –100 dBm
IF BW 10 kHz (at > 300 kHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –105 dBm
IF BW 3 kHz (at > 100 kHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –110 dBm
IF BW 1 kHz
30 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –95 dBm
100 kHz freq. 300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –115 dBm
IF BW 300 Hz
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –100 dBm
100 kHz freq. 300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –120 dBm
IF BW 100 Hz
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –105 dBm
100 kHz freq. 300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –125 dBm
• E5100B
IF BW 30 kHz (at > 1 MHz). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –100 dBm
IF BW 10 kHz (at > 300 kHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –105 dBm
IF BW 3 kHz (at > 100 kHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –110 dBm
IF BW 1 kHz
30 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –95 dBm
100 kHz freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –115 dBm
Note: When the analyzer frequency is identical to the transmitted interference signal frequency, refer to “EMC” in “general characteristics.”
Input crosstalk (When the analyzer is equipped with Option E5100x-100, delete this section.) Reference input (0 dBm input level at 10 kHz to 200 kHz and +5 dBm input level at 200 kHz to 300 MHz,
RF attenuator: 25 dB, 50 input) Test input (RF attenuator: 0 dB, terminated with 50 termination)
• E5100A
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –110 dB
100 kHz freq. 300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –120 dB
• E5100B
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –85 dB
100 kHz freq. < 250 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –105 dB
250 MHz freq. 300 MHz. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –95 dB
Source crosstalk
(all RF OUT and input connectors are terminated with 50 terminations)
• E5100A
without Option E5100A-801/802/803 (at +5 dBm output level, RF attenuator: 0 dB, 50 input)
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –110 dB (SPC)
100 kHz freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –125 dB (SPC)
250 MHz freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –120 dB (SPC)
with Option E5100A-801/802/803 (at +16 dBm output level, RF attenuator: 0 dB, 50 input)
10 kHz freq. < 100 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –120 dB (SPC)
100 kHz freq. < 250 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –135 dB (SPC)
250 MHz freq. 300 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –130 dB (SPC)