This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It may also refer to products/services no longer supported by Agilent. We regret any inconvenience caused by obsolete information. For the latest information on Agilent’s test and measurement products go to:
www.agilent.com/find/products
Or in the US, call Agilent Technologies at 1-800-452-4844 (8am–8pm EST)
Discontinued Product—Support Information Only
Evaluating Their PerfoArmance Characteristics
HP E4916A Crystal Impedance Meter
PRODUCT OVERVIEW
 Meets IEC Standard through the Transmission
π Network Method
 Allows Testing at up to 180 MHz  Enables Applying 0.1 nW Low Drive Level  Automates Testing and Evaluation through HP-IB and
Parallel I/O Interfaces
1
Displays Fr, CI, and Equivalent Circuit Parameters (Backlit LCD)
Drive Level Selectable between 0.1 nW and 1 mW (No Adjustments Required)
Front View
Easy Pass/Fail Judgment (No External PC Required)
RF Output
Parallel I/O Port
Mobile radios and communicators are increasingly expanding their applications into homes and offices. This makes the evaluation of crystal resonator characteristics at higher frequencies and/or lower drive levels more critical. However, conventional crystal impedance meters based on the oscillation method cannot properly meet the current measure­ment needs due to the following limitations on:  High-end measurement frequency
(60 MHz or below)
 Applicable power level
The HP E4916A is a crystal imped­ance meter based on the transmission
π network method which enables you
to:
Reference Input Test Signal Input Analog Voltage Output
Allows Testing at up to 180 MHzv
External Reference Signal Input
Drive Level Dependency Characteristic Evaluation Function
External Trigger Input
 Test at high frequencies (1 MHz to
180 MHz)
 Apply low drive level (0.1 nW to
1 mW, when testing a 25 DUT with the π circuit).
The HP E4916A yet provides a cost effective solution, comparable enough to conventional CI meters.
In addition to Fr and CI measure­ments, the HP E4916A offers a variety of testing capabilities including the equivalent circuit analysis and the drive level dependency characteristic evaluation (DLD) through easy-to­follow steps. It comes with the HP-IB and parallel I/O interfaces for easy connection to external equipment, thus allowing an automatic testing system to be configured.
Evaporation Monitoring Function
HP-IB Interface
Rear View
Key Features
 Based on the transmission π
network method recommended by the IEC444 standard.
 Wide frequency range (1 MHz to
180 MHz)
 Easy to change the drive level
(0.1 nW to 1 mW)
 Applies up to 20 mW to the DUT
when used in combination with the optional impedance probe (option
001)
 Allows you to use the same test
fixtures as those for the network analyzer.
 Provides the π calibration to correct
errors resulting from the extension of the cable.
 Saves and recalls up to 10 patterns
of the system settings together with calibration data.
2
Meets a Variety of Measurement Needs on the Production Lines
For Final Inspection:
 DLD (drive level dependency)
characteristic measurement. The HP E4916A offers the DLD characteristic measurement function for efficient determina­tion of variations in resonance frequency and impedance resulting from variations in drive level.
 Centralized data management. The
HP-IB interface allows testing to be performed automatically and measurement data to be managed collectively.
 Integration to a handler. The
handler interface allows connec­tion between an external handler and the HP E4916A.
HP 41901A SMD π Network and its Load Capacitances
For Intermediate Inspection:
 Consistent testing with final
inspection. Based on the same measurement method as that for the network analyzer, the HP E4916A provides measurement results that can properly correlate with those obtained with the network analyzer used for final inspection.
 Simple operation. The HP E4916A
simply requires trigger inputs to provide the desired parameters.
 Space saving. The HP E4916A can
fit anywhere in the clean room. Also, it operates stand alone: no need to use a PC for testing.
 Self-contained. The HP E4916A
requires no external devices for Fr (FL) and CI measurements and Pass/Fail judgment.
For Frequency Adjustments:
 EM (evaporation monitoring)
mode. The HP E4916A comes with the EM mode for high-speed monitoring.
 Frequency trapping for automated
evaporation. You can specify multiple trap frequencies to determine the transition of the resonance frequency. The HP E4916A detects in real time when the resonance frequency falls into each of the trap frequencies and outputs the results through the handler interface for proper controlling of the evaporator operation.
For Measurement with Load Capacitance:
 Consistent testing with network
analyzer. When measuring a DUT with load capacitance, the HP E4916A can make it easy to correlate your results with those obtained with network analyzer, for which same test fixtures and load capacitance can be used.
HP 41900A π
Network Test Fixture and its Load Capacitances
3