in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Acknowledgments
UNIX® is a registered trademark of the
Open Group.
Windows NT
mark of Microsoft Corporation.
Windows
istered trademarks of Microsoft Corporation.
®
is a U.S. registered trade-
®
and MS Windows are U.S. reg-
Manual Part Number
85225-90023
Edition
First edition, April 2005
Printed in USA
Agilent Technologies, Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA 95403 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, with out notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damage s in connection
with the furnishing, use, or performance of this document or of any
information contained he re in. Should
Agilent and the user ha v e a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
agency regulation or co ntract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencie s of t h e U .S . Gov ernment will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987) . U.S . Gov er nment users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Technology Licenses
The hardware and/or sof tware describe d in
this document are furnished under a
license and may be used or copied only in
accordance wi t h the terms of such license.
Restricted Rights Legend
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
Safety and Regulatory Information
Warnings, Cautions, and Not es
This installation and user’s guide utilizes the following safety notations.
Familiarize yourself with each notation and its meaning before operating
the Agilent 85225F performance modeling system.
WARNING
CAUTION
NOTE
A WARNING notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like that, if not correctly performed or
adhered to, could result in personal injury or death. Do not proce ed
beyond a WARNING notic e unt il the indicated conditions are fully
understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like that, if not correctly performed or adhered
to, could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
A NOTE calls the user’s attention to an important point or special
information within the text. It provides additional information or
instructions.
Installation and User’s Guide3
Safety Symbols and Instrument Markings
Symbols and markings in documentation and on instruments alert you to
potential risks, provide information about conditions, and comply with
international regulations. Table A defines the safety symbols and Table B
on page 5 defines the instrument markings you may find in the
documentation or on an instrument.
Table ASafety Symbols
SymbolsDefinition
Warning: risk of electric shock.
Warning: hot surface.
Caution: refer to instrument documen tat i on.
Laser radiation symbol: marked on pr oducts that have a laser
output.
Alternating current.
Both direct and alternating current.
Terminal is at earth potential. Used for measurement and
control circuits designed to be operated with one terminal at
earth potential.
Terminal for neutral conductor on permanently installed
equipment.
Terminal for line conductor on permanently installed
equipment.
4Installation and User’s Guide
Table ASafety Symbols (continued)
SymbolsDefinition
Standby (supply). Units with this s ymb ol are not completely
disconnected from AC mains w hen this s witch is in th e stand by
position. To completely disconnect the unit from AC mains,
either disconnect the power cord, or have a qu alified/licensed
electrician install an external switch.
ON (supply). A switch with this symbol closes the instrument’s
power supply circuit, connecting it to the mains supply .
OFF (supply). A switc h with th is symbo l open s the instr umen t’s
power supply circuit, disconnecting it from the mains supply.
Table BInstrument Markings
MarkingDefinition
The instruction documentation symbol appears when it is
necessary for the user to refer to the instruction in the
documentation.
The CE mark is a registered trademark of the European
Community.
This product compli es with the WEEE Directiv e (2002/96/EC)
marking requirements. The affixed la bel indicates th at you must
not discard this electrical/el ectronic product in domestic
household waste. To return unwanted products, contact your
local Agilent Technologies office, or see www.agilent.com for
more information.
The CSA mark is a registered trademark of the
CSA-International.
N10149
ISM1-AThis text indicates t hat the instrume nt is an In dustrial S cienti fic
ICES/NMB-001This text indicates product compliance wi th the Canadian
The C-tick mark is a registered trademark of the Spectrum
Management Agency of Australia. This signifies compliance
with the Australian EMC Framework regulations under the
terms of the Radio Communications Act of 1992.
and Medical Group 1 Class A product (CISPER 11, Clause 4).
Interference-Causing Equipment Standard (ICES-001).
Operator Safety Requirements
The following general safety precautions must be observed during all
phases of operation of this system. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates
Installation and User’s Guide5
safety standards of design, manufacture, and intended use of the product.
Agilent Technologies, Inc. assumes no liability for the customer’s failure to
comply with these requirements.
For additional safety precautions, including precautions for making device
measurements in a floating ground configuration, see “To ensure your
safety while using the system" on page 76.
WARNING
WARNING
WARNING
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the mains supply cord). The mains plug shall
be inserted only in a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor inside or
outside of the product is likely to make the product dangerous.
Intentional interruption is prohibited.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used only in a
normal condition (in which all means for protection are intact) only.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. Do not operate
the instrument in the presence of flammable gases or flames.
WARNING
DO NOT REMOVE THE INSTRUMENT COVER. Operating personnel
must not remove instrument covers. Component replacement and
internal adjustments must be m a de only by qualified service
personnel. Instruments that appear damaged or defective should be
made inoperative and se cured against unintended operation until
they can be repaired by qualified service personnel.
WARNING
WARNING
WARNING
6Installation and User’s Guide
Installing additional instrument s may dest abi l ize the rack cabinet.
Installing additional instruments into the ca bine t electrical system
could produce excessive leakage current. If the protective earth
conductor is interrupted or faulted, the user risks serious inj u ry or
death.
Prior to adding any additional instruments, review all wiring and
cooling capabilities to verify adeq uate design margins for nor mal and
under single fault conditions.
Mains power
CAUTION
Ground the system
WARNING
WARNING
Before applying power
CAUTION
The mains cable shall be permanently connected to the premise circuit
breaker or connected using an agency approved twist-lock connector.
To minimize shock hazard, the rack cabinet must be connected to an
electrical protective earth ground. The power distribution unit (PDU)
must be connected to the AC power mains through a grounded power
cable, with the ground wire firmly connected to an electrical ground
(safety ground) at the power outlet.
Any interruption of the protective (grounding) conductor or
disconnection of the protective earth terminal will cause a potential
shock hazard that could result in personal injury.
Verify that the product is set to match the available line voltage, the
correct fuse is installed, and all safety precautions are taken. Before
applying power, note the product’s external markings described in
Table A, “Safety Symbols,” on page4 and Table B, “Instrument
Markings,” on page5.
CAUTION
CAUTION
CAUTION
Installation and User’s Guide7
It is recommended that the premise wiring contain an adequate circuit
breaker for system protection.
To remove power from the cabinet, remove the mains supply from the
premise electrical supply.
Before switching on this system, make sure that the supply voltage is
in the specified range.
CAUTION
Fuses and breakers
The front panel LINE switch disconnects the mains circuit from the
mains supply. Ho wev e r, the mains supply to the po wer d is tr ib ut io n u n it
remains energized.
WARNING
Before cleaning the system
WARNING
Overcurrent protection
CAUTION
For continued protection against fire hazard, use only fuses with the
required rated current, voltage, and specified type (normal blow ,
time delay). Do not use repaired fuses or short-circuited fuse
holders. Replace only with an identical fuse.
There are two resettable thermal bre a ke rs located on the power
strips. These are in the “hot” and “neutral” lines.
To prevent electrical shock, disconnect the system from mains before
cleaning. Use a dry (or slightly water-dampened) cloth to clean
external case parts. Do not atte mpt to clean internally.
If the power outlet strip breaker trips once, reset the breaker. If the
breaker trips twice, call a qualified/licensed electrician to service the
test system.
Statement of Compliance and Declaration of Conformity
This product has been designed and tested in accordance with accepted
industry standards, and has been supplied in a safe condition. The
documentation contains information and warnings that must be followed
by the user to ensure safe operation and to maintain the product in a safe
condition.
The Manufacturer’s Declaration of Conformity is available upon request.
Statement of CAN/CSA Compliance
This product has been designed and tested in accordance with
CAN/CSA- C22.2 No. 61010- 1 IEC.
8Installation and User’s Guide
Compliance with German Noise Requirements
This is to declare that this instrumentis in conformance with the German
Regulation on Noise Declaration for Machines (Laermangabe nach der
Maschinenlaermrerordnung - 3.GSGV Deutschland).
Acoustic Noise Emission/Geraeuschemission
LpA <70 dBLpA <70 dB
Operator positionam Arbeitsplatz
Normal positionnormaler Betrieb
per ISO 7779nach DIN 45635 t.19
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES- 001. Cet appareil ISM est
conformé à la norme NMB du Canada.
IEC/EN 61000-4-2 Electrost atic Discharge Immunity Test
This system passes using criterion C where operator intervention may be
necessary to restart the measurement software operations.
IEC/EN 61326 Electrostat ic Di sc harg e and Surge Im muni t y Test
This system complies with the Electrostatic Discharge and Surge Immunity
requirements in the IEC/EN 61326 standard using Performance Criterion
C.
For Technical Assistance
To receive technical assistance, visit the online assistance web site, or call
the telephone number listed in Table 19 on page 107 appropriate to the
location of modeling system.
Installation and User’s Guide9
In This Guide...
1Introducing the Agilent 85225F Performance Modeling System
2Installing the System
3Verifying System Functionality
This guide provides instruction on installing, verifying, and servicing the
system, as well as an introductory system overview and reference material.
This information is presented for use by the customer or an Agilent
Technologies field engineer.
This chapter provides a description of the system, its components,
integration, and characteristics.
Here you will find instruction on preparing the installation site, receiving
and inspecting the system (including a receiving checklist), installing the
worksurface, ensuring operator safety, connecting the bias networks, and
powering- on the system.
Turn here for instruction on choosing a level of system verification and
performing a post-installation functional verification test using a system
controller running IC- CAP software.
4Servicing the System
This chapter includes instruction on troubleshooting the system, removing
and replacing system components, ordering replacement parts, and
acquiring additional assistance in solving measurement problems.
AEnhancing Measurement Accuracy
See this appendix for instruction on cleaning the system connections,
performing a system measurement calibration, and suggested intervals for
periodic component calibration.
BDC Subsystem Functional Verificat ion Tests
Turn here to find Agilent 4156C precision semiconductor parameter
analyzer and Agilent E5260A/70B high speed/precision parameteric
measurement mainframe functional verification tests that do not require
the IC- CAP software.
CRF Subsystem Functional Verification Tests
This appendix includes an Agilent E8364B PNA Series vector network
analyzer functional verification test that does not require the IC- CAP
software.
10Installation and User’s Guide
DCV Subsystem Functional Verification Tests
This appendix includes an Agilent 4284A precision LCR meter functional
verification test that does not require the IC- CAP software.
ENoise Subsystem Functional Verifi cation Tests
This appendix includes an Agilent 35670A dynamic signal analyzer
functional verification test that does not require the IC- CAP software.
FUnderstanding the Bias Networks
Here you will find features, characteristics, a schematic diagram, and
operational information on the bias networks.
See this appendix for a summary of the network analyzer’s performance
specifications.
For Additional
Information on...
Hardware
Additional information regarding instruments and accessories within the
system is provided in the individual instrument or accessory’s
documentation.
Software
IC- CAP software operating instructions and tutorials are provided in the
Agilent 85190D IC- CAP user’s guide.
Installation and User’s Guide11
Typeface
Conventions
This guide uses the following typeface conventions to describe various
aspects of a particular hardware or software user interface.
Hardware
InterfaceExamples in Body TextExamples in Procedural Text
and Tables
Front panel hardke ys
Front panel display
softkeys
Front or rear panel
connectors, instrument
markings
Data field entries
Keyboard keys
Press Preset
Press Cal
Press [MORE]
Press [Return]
RF/DC OUT connector
STIMULUS key group
Enter Calset
Enter 18
Press Ctrl+8
Press Enter
Press Preset
Press Cal
Press [MORE]
Press [Return]
RF/DC OUT connector
STIMULUS key group
Enter Calset
Enter 18
Press Ctrl+8
Press Enter
Software
InterfaceExamples in Body Te xtExamples in Procedural Text
and Tables
Screen buttons and
selections
Menu selections
Click Enter
Select Continuous
Choose Format > Small
Choose Cal > Full
Click Enter
Select Continuous
Choose Format > Small
Choose Cal > Full
Command and menu
names
Icon and window titles
The Save commands are in
the File menu.
The Model icons are in the
IC- CAP/Main window.
Program messages
Data field entries
Is the device conn ected?
Enter Calset
Enter 18
12Installation and User’s Guide
The Save commands are in the
File menu.
The Model icons are in the
IC-CAP/Main window.
Is the device connected?
Enter Calset
Enter 18
Contents
1 Introducing the Agilent 85225F Performance Modeling System
Performance Modeling System Configuration Overview18
RF and DC Measurement System Configuration19
Figure 1. System Block Diagram19
The RF Subsystem20
The DC Subsystem20
The Bias Networks21
Component Integration22
Figure 2. System Components23
Table 3.Front Panel System Connections, with Agilent 4156C24
Figure 3.Front Panel Connections with Agilent 4156C25
Table 4.Front Panel System Connections, with Agilent E5260A or E5270B26
Figure 4.Front Panel Wiring Diagram with Agilent E5260A or E5270B27
Table 5.Rear Panel System Connections with Agilent 4156C28
Figure 5.Rear Panel Wiring Diagram with Agilent 4156C29
Table 6. Rear Panel System Connections with Agilent E5260A or E5270B30
Figure 6.Rear Panel Wiring with Agilent E5260A or E5270B31
Figure 7.DC/RF Cabling Diagram - DC and RF Configuration32
CV, RF, and DC Measurement System Configuration33
Figure 8. System Block Diagram33
The CV Subsystem34
Component Integration34
Figure 9. System Components35
Table 7.Front Panel System Connections36
Figure 10. Front Panel Wiring Diagram37
Table 8.Rear Panel System Connections38
Figure 11.Rear Panel Wiring Diagram39
Figure 12. DC/RF Cabling Diagram - DC and RF Configuration40
Figure 13.DC/RF Cabling Diagram - Parametric Configuration41
The Low Leakage Switch Mainframe42
Table 9.Rear Panel Connections, including Low Leakage Switch Mainframe42
Figure 14. Rear Panel Wiring Diagram including Low Leakage Switch Mainframe43
Figure 15.DC/RF Cabling Diagram - Parametric Configuration with Low Leakage Switch
Mainframe44
Installation and User’s Guide13
1/f Noise, CV, RF, and DC Measurement System Configuration45
Figure 16.System Block Diagram46
Figure 17. 1/f Noise Measurement Block Diagram47
Component Integration48
Figure 18. System Components49
Table 10.Front Panel System Connections50
Figure 19. Front Panel Wiring Diagram51
Tab le 11. Rear Panel System Connections52
Figure 20.Rear Panel Wiring Diagram53
Figure 21. DC/RF Cabling Diagram - DC and RF Configuration54
Figure 22.DC/RF Cabling Diagram - Parametric Configuration55
Instrument Control Interface56
Tab le 12. GPIB Addresses 56
The LAN/GPIB Gateway57
Figure 23.Rear Panel Wiring Diagram for LAN/GPIB Gateway58
The System Controller59
Table 13. Personal Computer Requirements59
Table 14. UNIX Workstation Requirements59
The Rack Cabinet 60
Performance Characteristics and Specifications61
Table 15.Supplemental System Characteristics61
Interference Standards61
Performance Modeling System Performance Specifications61
RF Subsystem Performance Specifications62
DC Subsystem Specifications62
Bias Network Characteristics62
2 Installing the System
To prepare the installation site64
Ta b le 16 . Environmen t al Requ ir em e nts64
Table 17. Electrical Requirements64
To receive the system65
To unpack the shipment crate containing the rack cabinet66
To verify the shipment68
Tab le 18. Replaceable Parts69
To install the work surface74
To ensure your safety while using the system 76
Precautions for Performing Floating-Ground Measurements78
To perform floating-ground measurements 78
14Installation and User’s Guide
Precautions for Avoiding Electrostatic Discharge79
To connect the bias networks80
Agilent 4156C Systems80
Agilent 4156C Systems with Agilent 41501B Expander Box82
Agilent E5260A/70B Systems84
To switch on power to the system87
To configure the LAN/GPIB gateway for functional verification89
3 Verifying System Functionality
To choose a verification process92
Understanding the System Functional Verification Test94
Required Tools94
Performing the System Functional Verification Test95
If you encounter a problem101
4 Servicing the System
To troubleshoot the system104
To remove or replace a system component105
To order replacement parts106
To receive additional assistance107
Ta b le 19. Contacting Agil ent Technologies107
To package the system for transport108
A Enhancing Measurement Accuracy
To enhance measurement accuracy112
Understanding System Measurement Cal ibrati on114
Required Tools114
Performing a Coaxial System Measurement Calibration115
If you encounter a problem116
Periodic System Component Calibration117
B DC Subsystem Functional Verification Test
Understanding the DC Subsystem Functional Verification Test120
Required Tools120
Performing the DC Subsystem Functional Verification Test121
If you encounter a problem121
Installation and User’s Guide15
If you encounter a problem122
C RF Subsystem Functional Verification Test
Understanding the RF Subsystem Functional Verification Test124
Required Tools124
Performing the RF Subsystem Functional Verification Test125
If you encounter a problem127
D CV Subsystem Functional Verification Test
Understanding the CV Subsystem Functional Verification Test130
Required Tools130
Performing the CV Subsystem Functional Verification Test131
If you encounter a problem132
E 1/f Noise Subsystem Functional Verific ation Test
Understanding the 1/f Noise Subsystem Functional Verification Test134
Required Tools134
Performing the 1/f Noise Subsystem Functional Verification Test135
G Network Analyzer Performance Specification Summary
Network Analyzer System Performance142
Maximum Output Power142
Dynamic Range142
Measurement Port Charact eristics143
Measurement Uncertainty143
Index
16Installation and User’s Guide
Agilent 85225F Perfor mance Model ing System
Installation and User’s Guide
1
Introducing the Agilent 85225F
Performance Modeling System
Performance Modeling System Configuration Overview 18
RF and DC Measurement System Configuration 19
CV, RF, and DC Measurement System Configuration33
1/f Noise, CV, RF, and DC Measurement System Configuration 45
The System Controller 59
Performance Characteristics and Specifications 61
Use this chapter to familiarize yourself with the measurement
configurations of the performance modeling system. This chapter
introduces the system by describing its operational theory, integration, and
performance.
Agilent Technologies
17
1Introducing the Agilent 85225F Performance Modeling System
Performance Modeling System Configuration Overview
The standard Agilent 85225F performance modeling system measures the
DC and RF performance of active and passive devices. You may configure
the Agilent 85225F performance modeling system to measure CV and 1/f
noise with the addition of optional instrumentation and IC-CAP 1/f noise
measurement modules.
For RF and DC performance measurement system configurations, see “RF
and DC Measurement System Configuration" on page 19.
For CV, RF, and DC performance measurement system configurations, see
“CV, RF, and DC Measurement System Configuration" on page 29.
For 1/f noise, CV, RF, and DC performance measurement system
configurations, see “1/f Noise, CV, RF, and DC Measurement System
Configuration" on page 45.
18Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System1
RF and DC Measurement System Configuration
In conjunction with a compatible controller running 85190-Series IC- CAP
software, the Agilent 85225F performance modeling system measures the
DC and RF performance of active and passive devices. The IC- CAP
software then extracts the device parameters and displays the results.
The Agilent 85225F performance modeling system is the integration of
rack- mounted RF and DC subsystems, bias networks, and a system
controller
Figure 1System Block Diagram
*
, as shown in Figure 1†.
* The system controller is not included and must be provided.
† This block diagram shows a system with an Agile nt 4156C as the DC subsystem. Other instrumentation may
be used. See “The DC Subsystem"on page 20.
Installation and User’s Guide19
1Introducing the Agilent 85225F Performance Modeling System
The RF Subsystem
S- parameter device characterization is provided by the RF subsystem.
The RF subsystem contains the Agilent E8364B PNA Series vector network
analyzer.
Its integrated synthesizer supplies a swept or CW RF source signal from
10 MHz
The integrated test set separates the RF source signal into reference and
test signals, and provides RF connection via cables and adapters to the
external bias networks.
*
to 50 GHz.
The DC Subsystem
Precision DC characterization and bias for the S-parameter measurements
are provided by one of the following three DC subsystems.
The DC subsystem may contain one of the following three instruments.
Agilent 4156C Precision Semi conducto r Par a meter Analyzer
The Agilent 4156C precision semiconductor parameter analyzer provides
DC force (supply) and sense (measure) capability from its HRSMUs (high
resolution source/monitor units).
Optionally, the Agilent 4156C may be configured with a 41501B SMU PGU
expander is connected to and controlled by the 4156C via the expander
box interface. The 41501B provides a GNDU (active ground unit) and,
depending on option configuration, an HPSMU (high- power source/monitor
unit), two MPSMUs (medium- power source monitor units), and/or two
PGUs (pulse generator units).
The DC signals are routed through feedthrough panels via triaxial cables to
the bias networks.
Agilent E5260A 8-Slot High Speed Parametric Measurement Mainframe
The Agilent E5260A provides DC force (supply) and sense (measure)
capability from its plug- in source/monitor units.
The Agilent E5290A plug- in high speed high power source/monitor unit
provides up to 200 volts of potential and 1 amp of current to the device
under test.
The Agilent E5291A plug- in high speed medium power source/monitor unit
provides up to 100 volts of potential and 200 milliamps of current to the
device under test.
* Due to the minimum operating frequency of the bias networks, the performance modeling system low end
frequency range is 45 MHz.
20Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System1
The Agilent E5270B provides DC force (supply) and sense (measure)
capability from its plug- in source/monitor units.
The Agilent E5280A plug- in high power source/monitor unit provides up
to 200 volts of potential and 1 amp of current to the device under test.
The Agilent E5281A plug- in medium power source/monitor unit provides
up to 100 volts of potential and 200 milliamps of current to the device
under test.
CAUTION
The Bias Networks
Exposing the bias networks to currents greater than 500 milliamps or voltages
greater than 40 volts will result in severe damage. Do not exceed these values
while using the bias networks. Remove the bias networks from the circuit if
greater voltages or currents are required.
The Agilent 11612V Option K11 and K21 bias networks combine the DC
and RF signals and apply them simultaneously to the device under test
(DUT). The bias networks are configured with 2.4 mm DC/RF output
connectors for connection to a DUT, a test fixture, or probe station, as
shown in Figure 21 on page 54.
Installation and User’s Guide21
1Introducing the Agilent 85225F Performance Modeling System
Component Integration
System component integration is performed at the Agilent Technologies
factory. The individual components are placed into the rack, and the
required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional
performance.
The Agilent 85225F performance modeling system includes the following
components, as shown in Figure 2 on page 23:
• Agilent E8364B PNA Series vector network analyzer
Introducing the Agilent 85225F Performance Modeling System1
Figure 6Rear Panel Wiring with Agilent E5260A or E5270B
Installation and User’s Guide31
1Introducing the Agilent 85225F Performance Modeling System
Figure 7DC/RF Cabling Diagram - DC and RF Configuration
Systems with Agilent 4156C
Systems with Agilent E5260A or E5270B
32Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System1
CV, RF, and DC Measurement System Configuration
With the addition of a precision LCR meter, the Agilent 85225F
performance modeling system measures the DC, RF, and CV performance
of active and passive devices. The IC- CAP software then extracts the
device parameters and displays the results.
The Agilent 85225F performance modeling system for CV, RF, and DC
measurement is the integration of rack-mounted RF, DC, and CV
subsystems, bias networks, and a system controller, as shown in Figure 8.
Figure 8System Block Diagram
Installation and User’s Guide33
1Introducing the Agilent 85225F Performance Modeling System
The CV Subsystem
The Agilent 4284A precision LCR meter provides a wide 20 Hz to 1 MHz
test frequency range and superior test- signal performance, allowing CV
testing to the most commonly- used test standards, such as IEC/MIL, and
under conditions that simulate the intended application.
Optionally, the system can be configured with the Agilent E5250A low
leakage switch mainframe. The Agilent E5250A is used for precise
parametric test. It improves measurement efficiency by eliminating the
need to manually change the probe positions on a manual probe station.
The E5250A is used to route signals from the DC and CV subsystems to
the probe card cable, and on to the probe card and probe station.
Component Integration
System component integration is performed at the Agilent Technologies
factory. The individual components are placed into the rack, and the
required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional
performance.
The Agilent 85225F performance modeling system includes the following
components, as shown in Figure 18 on page 49:
• Agilent E8364B PNA Series vector network analyzer
1Introducing the Agilent 85225F Performance Modeling System
The Low Leakage Switch Mainframe
The Agilent E5250A is used for precise parametric test. It improves
measurement efficiency by eliminating the need to manually change the
probe positions on a manual probe station. The E5250A is used to route
signals from the 4156C and the 4284A to the probe card cable, and on to
probe card and probe station.
Table 9Rear Panel Connections, including Low Leakage Switch Mainframe
Component InformationConnection Information
DesignatorModel
Introducing the Agilent 85225F Performance Modeling System1
1/f Noise, CV, RF, and DC Measurement System Configuration
With the addition of a dynamic signal analyzer and a precision LCR meter,
the Agilent 85225F performance modeling system measures the DC, RF,
CV, and 1/f noise performance of active and passive devices. The IC- CAP
software then extracts the device parameters and displays the results.
The Agilent 85225F performance modeling system is the integration of
rack- mounted RF and DC subsystems, a precision LCR meter, a dynamic
signal analyzer, bias networks, and a system controller, as shown in
Figure 16.
Installation and User’s Guide45
1Introducing the Agilent 85225F Performance Modeling System
Figure 16System Block Diagram
The 1/f Noise Subsystem
The Agilent 35670A dynamic signal analyzer (in conjunction with a
customer- furnished Stanford Model SR570 low noise amplifier) measures
the flicker noise (1/f noise) of active devices. Controlled by IC- CAP device
modeling software, the dynamic signal analyzer generates reliable 1/f noise
measurement data, which are analyzed and extracted in IC-CAP. Figure 17
shows the system configuration for 1/f noise measurements.
46Installation and User’s Guide
Figure 171/f Noise Measurement Block Diagram
Introducing the Agilent 85225F Performance Modeling System1
Installation and User’s Guide47
1Introducing the Agilent 85225F Performance Modeling System
Component Integration
System component integration is performed at the Agilent Technologies
factory. The individual components are placed into the rack, and the
required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional
performance.
The Agilent 85225F performance modeling system includes the following
components, as shown in Figure 18 on page 49:
• Agilent E8364B PNA Series vector network analyzer
• Agilent 4156C precision semiconductor parameter analyzer with
1Introducing the Agilent 85225F Performance Modeling System
Instrument Control Interface
Instrument control interface is provided by a General Purpose Interface
Bus (GPIB) or LAN/GPIB gateway. GPIB addresses for programmable
system components are listed in Table 12.
Table 12 GPIB Addresses
ComponentGPIB Address
Agilent 34401A digital multime te r9
Agilent 35670A dynamic signal analyzer10
Agilent 4156C precision semiconductor parameter analyzer19
Agilent 4284 A precision LCR meter24
Agilent E5810A LAN/GPIB gateway21
Agilent E8364B PNA Series vector network analyzer16
* The 4156C default GPIB address 17 is sometimes used by other devices with a GPIB address at 16 (for example, an external display (set
to 17) to display the results generated by an instrument at address 16). Change the 4156C GPIB address to 19 using the procedure
described in step13 of “To switch on power to the system"on page 87 to ensure that IC-CAP can recognize the 4156C.
56Installation and User’s Guide
The LAN/GPIB Gateway
The Agilent E5810A LAN/GPIB gateway provides access to the system’s
GPIB instrumentation over an existing local area network. It allows the
use of SICL- or VISA- based applications designed for GPIB over the LAN
without modifying the application beyond a simple address change.
The gateway is a combination of hardware and SICL/VISA software. It
uses client/server technology to extend the standard remotely over the
LAN, allowing remote control from an alternative, more convenient, or
safer location.
Introducing the Agilent 85225F Performance Modeling System1
Installation and User’s Guide57
1Introducing the Agilent 85225F Performance Modeling System
Figure 23Rear Panel Wiring Diagram for LAN/GPIB Gateway
58Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System1
The System Controller
A customer- furnished UNIX workstation or personal computer running
Agilent IC- CAP software controls the hardware via GPIB while making
device measurements, then stores, simulates, and optimizes device
parameters, using predefined or user- defined device models. Table 13 on
page 59 lists the personal computer requirements. Table 14 on page 59
lists the UNIX workstation requirements.
Table 13 Personal Computer Requirements
ParameterRequirement
Operating sys t emMicrosoft Windows NT® 4.0- SP6a or Windows 2 000 Professional-SP3.
CPUIntel Pentium® class 200 MHz CPU or higher
DisplaySuper VGA 800×600, 15 inch moni tor (1024×728 recommended)
Hard disk space370 MB. It is recommended that you install IC-C AP software on your local drive.
Recommended file systems are FAT32 and NTFS. Novell file servers are not
supported. VFAT/FAT systems are not recommended for full installations.
* Windows 95, 98, and ME are not supported.
† For NT 4.0 only: to avoid potential memory problems, ensure your virtu al memo ry space is always greater than your RAM space.
†
(Increased virtual memory may be required)
Table 14 UNIX Workstation Requirements
ParameterRequirement
HP UNIX WorkstationSunOS Workstation
Operating systemHP-UX 11.i with the following patches:
PHSS_24627 HP aC++, AA Runtime
Libraries (aCC A.03.33), PHSS_25718 LI BCL
Window managerHP VUE or CDE/X-Windows V.X11R5 Motif V.1.1/1.2 Open Windows 3.0, or CDE
RAM128 megabytes (additional RAM will improve software performance)
Swap space200 megabytes (additional swap space will improve software performance)
Hard disk300 megabytes for minimum installation
500 megabytes for complete installation including online documentation and application
examples
SunOS 5.7, 5.8, and 5.9
(Solaris 7.0, 8.0, 9.0)
DisplayHigh resolution color only
Installation and User’s Guide59
1Introducing the Agilent 85225F Performance Modeling System
The Rack Cabinet
The system is housed in a 1.6 meter rack cabinet. The cabinet provides
line power access, ventilation, mobility, and protection to the system
instrumentation.
A rack- mounted work surface is included for maximum flexibility and
convenience in making in- fixture or coaxial measurements. The work
surface can be removed to facilitate on- wafer measurements using a probe
station. The work surface is coated with antistatic material and connected
to chassis ground. Therefore, an antistatic mat is not required. For
installation instructions, see “To install the work surface" on page 74.
60Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System1
Performance Characteristics and Specifications
Supplemental characteristics are not specifications, but are provided in
Table 15 for your convenience.
Table 15 Supplemental System Characteristics
CharacteristicValue
Line voltage115 volts nominal (90 volts to 132 volts) or 220 volts nominal
(210 volts to 250 volts)
Line frequency48 Hz to 66 Hz
Circuit breaker amper a ge rating6 amps (115 volts), 3.5 amps (220 volts)
Rack weight capacity818 kilograms (1800 pounds) maximum load ed
Rack external di m ensions1620 mm high × 600 mm wide × 905 mm deep
Rack footprint (top view)
Interference Standards
The IEC/EN 61326- 1 and CISPR Publication 11 standards define the RFI
and EMI susceptibility of the performance modeling system.
Performance Modeling System Performance Specifications
The Agilent 85225F performance modeling system adheres to the
performance specifications of an Agilent E8364B PNA Series vector
network analyzer. Refer to Appendix G, “Network Analyzer Performance
Specification Summary,” starting on page 141.
Installation and User’s Guide61
1Introducing the Agilent 85225F Performance Modeling System
RF Subsystem Performance Specifications
The overall performance of a network analyzer is dependent on the
individual instruments, system configuration, user- defined operating
conditions, measurement calibration, and cables.
For a specification summary, refer to Appendix G, “Network Analyzer
Performance Specification Summary,” starting on page 141.
In any high- frequency measurement, residual errors contribute
uncertainties to the results.
NOTE
When the system is configured with a probe station, microwave probes, on-wafer
calibration standards, or t est fixtures, additional uncertainties are contributed to the
measurement results. Refer to the manufacturer’s documentation for information on probe
station or test fixture characteristics.
DC Subsystem Specifications
Specifications for the Agilent 4156C precision semiconductor parameter
analyzer are listed in its user’s guide, chapter 7 of Volume 1, “General
Information.”
Specifications for the Agilent E5260A 8- slot high speed measurement
mainframe and Agilent E5270B 8- slot precision parametric measurement
mainframe are listed in its user’s guide, Chapter 2, “Introduction.”
Bias Network Characteristics
Table 20, “11612V Option K11/K21 Bias Network Characteristics,” on
page 139 lists the operational characteristics of the bias networks. For
detailed information, refer to Appendix F, “Understanding the Bias
Networks,” starting on page 137.
62Installation and User’s Guide
Agilent 85225F Perfor mance Model ing System
Installation and User’s Guide
2
Installing the System
To prepare the installation site 64
Environmental Requirements 64
Electrical Requirements 64
To receive the system 65
To unpack the shipment crate containing the rack cabinet 66
To verify the shipment68
To install the work surface74
To ensure your safety while using the system 76
Precautions for Performing Floating-Ground Measurements 78
Precautions for Avoiding Electrostatic Discharge 79
To connect the bias networks 80
To switch on power to the system 87
This completes the inst allat i on pro ce ss. To confirm the functional it y of t he
system, continue to Chapter3, “Verifying System Functionality,”
starting on page 91. 89
Related Topics“Introducing the Agilent 85 22 5F Perfo rma nce Mod el in g Syst em" on
page 17
Use this chapter to learn how to first prepare the installation site, and
then receive, unpack, install, and configure the system. This chapter
includes important information on operational safety, as well as
instruction on preparing the installation site, unpacking the system,
ensuring the completeness of the system shipment, installing the work
surface, performing final system configuration, and powering- on the
system.
Agilent Technologies
63
2Installing the System
To prepare the installation site
CAUTION
Follow these steps to prepare the site for system installation.
To prepare the installation site
StepNotes
1 Ensure that your installation site meets the
2 Ensure that your installation site meets the
Table 16 Environmental Requirements
Environmental ParameterSystem Requirement
Temperature+0°C to +45°C (+32 °F to +113°F)
Relative humidityMaximum 80% for temperatures up to 31°C
This product is designed for indoor use in Installation Category II
and Pollution Degree 2 per IEC 61010-1 and 664 respectively.
environmental requirem ents.
electrical requirements.
• Environmental requirements (temperature,
relative humidity, altitude, and clearance) are
listed in Table 16.
• Electrical requirements are listed in Table 17.
*†
decreasing linearly to 50% at 40°C
AltitudeUp to 3000 meters (approximately 10000 feet)
Clearance (behind an d abo v e rac k)
* Install air conditioning and heating as needed to achieve the required ambient temperature range.
† Accuracy-enhanced measurement performance is specified at an ambient temperature range of
+25°C ±5°C. After calibration, hold the ambient temperature of the measurement environment to
±1°C of the ambient temperature at the time of calibration.
‡ Required to ensure the extractor fans can properly ventilate the system .
‡
15 centimeters (6 inches) minimum
Table 17 Electrical Requirements
Electrical ParameterSystem Requirement
Supply capability100/120 volts, 2000 VA
200/240 volts, 2000 VA
Circuit sharingDo not connect air conditioning or
motor-operated equipment to the sa me ac
circuit supplying line voltage to the system.
64Installation and User’s Guide
To receive the system
Follow these steps to store, inspect, and confirm the system shipment.
To receive the system
StepActionNotes
Installing the System2
1 Store and inspect the
shipment.
2 If the system is damaged,
notify appropriate parties.
a Keep the shipping containers toge ther ,
unopened, located in one area.
b Inspect the s hipping containers for
damage.
• If the shipment is damaged, contin ue to
step 2.
• If the shipment is verif i ed undamaged,
continue to the next section, “To unpack
the shipment crate containing the rack
cabinet" on page 66.
a Report the shipment damage to your
Agilent Technologies sales representative.
b Report the ship ment damage to the
shipping carrier.
c Provide all cartons and packaging material
for inspection by the shipping ca rrier .
• Keep all cartons and packaging
material until the entire shipment has
been verified undamaged and
complete, and the sy stem has passed
visual inspection and functional
verification.
• Agilent Technologies wi ll r ep ai r or
replace damaged equipment without
waiting for a claim settlement fr om the
shipping carrier.
Installation and User’s Guide65
2Installing the System
To unpack the shipment crate containing the rack cabinet
The racked system is shipped upright secured to a pallet. Other system
components are shipped separately. Follow these instructions to unpack
and inspect the rack cabinet and the racked system components.
Required Tools
• 9/16 inch wrench or adjustable end wrench
• Prying tool to remove packaging clamps
WARNING
Always wear safety glasses when removing the clamps and other
packing materials fr om th e cra te .
CAUTION
Be careful not to bend the clamps while removing them from the
shipping crate. You may reuse the clamps when the system is
repacked.
To unpack the shipment crate containing the rack cabinet
StepActionNotes
1 Remove the outer packing
crate.
a Remove the clamps holding the packing
crate top cover in place.
b Remove the top cover and set it aside.
c Remove the clamps holding the first
packing crate wall in place.
d Insure that two other people are available
to hold the last two walls in place as the
last set of clamps is removed.
e Remove the other walls.
f Set the loading ramp panel aside for now.
• Which wall is removed first does not
matter.
• In double- rack crates, the heaviest wal l
is the loading ramp. In single-rack
crates, the loading ramp is shipped
inside the crate, plac ed on to p of th e
rack (it is a hinged asse mbly , shipped in
the folded position).
2 Remove the packaging
materials.
66Installation and User’s Guide
a Remove the foam top cover.
b Remove the plastic wrapping from the
system.
To unpack the shipment crate containing the rack cabinet (continued)
StepActionNotes
3 Unload the systema Remove the two brace bolts attaching the
side brace assembly to the bottom pallet.
b Remove the side brace assembly.
c Lift the hinged slat and remove the ramp
anchor bolt.
d Place one end of the ramp on the p allet
ramp ledge.
e Insert the ramp anchor bolt and fold down
the hinged slat.
f Fold down the ramp’s end flap.
• To secure the ramp, you may place long
wood screws through the ramp and
into the ramp ledge.
Installing the System2
WARNING
A racked system is tall and top-heavy. It is easy to ti p the rack
over while moving it, which could result in injury or death.
Unloading the system safely requires the participation of four
persons exercising care so as not to topp l e the rac k cabi net . Do
not stand in front of the rack as it rolls down the ramp.
g Ensure that the rack cabinet leveling feet
are retracted and that the cabinet casters
are rolling freely.
h Roll the system down the ramp using
extreme care.
i Carefully roll the rack toward its prepared
place within the measurement
environment.
• In case the system must be moved in
the future, retain and reuse these
packing materials. You can also
purchase replacement packing
materials from Agilent Technologies.
Installation and User’s Guide67
2Installing the System
To verify the shipment
Use Table 18 “System Receiving Checklist and Replaceable Parts” to:
• confirm the completeness of the shipment
• provide component part and model numbers required to order
All replacement items are available from Agilent Technologies. Part
numbers for replacement instrument subassemblies are listed in their
individual service manuals.
The majority of the system components are shipped preconfigured in the
system rack cabinet.
The PGUs (pulse generator units) and GNDU (active ground unit) are
factory- installed in the 41501B PGU expander.
With the exception of the bias networks, semi- rigid and SMU triaxial
cables, all other cables are connected at the factory.
replacement parts
NOTE
To verify the shipment
StepActionNotes
1 Verify that the serial
numbers on the rear panel
of the system inst rument s
match the serial numbers
listed in the shipping
documentation.
Other cables and accessories are shipped inside the rack-mounted storage drawer.
Compare the serial numbers liste d in the
shipping documents with the serial numbers
on the instrument’s rear panel serial number
labels.
• If an instrument serial number does not
match the shipping document, report
mismatched serial number to your
Agilent Technologies sales
representative.
• If all instrument serial numbers match
the shipping documents, continue to
step 2.
• For a list of support contacts, see “To
receive additional assistance" on
page107.
68Installation and User’s Guide
To verif y the shipme nt (con ti nu ed)
StepActionNotes
Installing the System2
2 Complete the receiving
checklist.
3 If the system is
incomplete, report
missing item s to your
Agilent Technologies
sales representative.
1
Table 18 Replaceable Parts
Part or Model N umberDescription
10833AGPIB cable, 1 meter
10833BGPIB cable, 2 meter
a Compare the Bill of Materials to the system
components received in the shipment.
b Verify the shipment is complete.
• If the shipment is confi rm ed incomplete,
go to step3.
• If you have confirmed the presence of all
system components, the receiving
process is complete. Proc ee d to the next
section “To ensure your safety while
using the system"on page76.
• Refer to the Bill of Materials included
with the shipment.
• For a list of support contacts, see “To
receive additional assistance" on
page107.
10833CGPIB cable, 4 meters
10833DGPIB cable, 0.5 meter
11612T Option K33Mounting plates, bias networ ks to probe station
11612V Option K11Bias netwo rk, port 1, 45 MHz to 50 GHz, 0.5 A
11612V Option K21Bias netwo rk, port 2, 45 MHz to 50 GHz, 0.5 A
11900AAdapter, 2.4 mm (male-to-male)
11900BAdapter, 2.4 mm (female-to-female)
11900CAdapter , 2.4 mm (male-to-female)
1250-0080Adapter, BNC, 50 ohm (female-female)
1250-1700Adapter, coax
1250-2405CAdapter , BNC coaxial tee
1250-3231Adapter, triaxial BNC (female to male)
16048DTest leads, 4 terminal pair, 1.98 meter
16493J Option 001Interlock cable, 1.5 meter
16493L Option 001G ND U c a bl e, 1.5 meter
Installation and User’s Guide69
2Installing the System
Table 18 Replaceable Parts (continued)
Part or Model N umberDescription
16494A Option 001Triaxial cable, 1.5 meter
16494A Option 002Triaxial cable, 3 meter
16494B Option 001Kelvin triaxial cable, 1.5 meter
16494B Option 002Kelvin triaxial cable, 3 meter
34401ADigital multimeter
35181MStorage drawer
35670ADynamic signal analyzer
35670A Option AX4Rack flange kit
35670A Option AY2Two-input channel configuration
35670A Option AY6Four-input channel configuration
35670A Option 1D4Arbitrary source
41501BSMU/PGU expander with GNDU and cable
41501B Option 410Add 41501B with high power SMU and cables
41501B Option 412Add high power SMU, 2 PGUs, and cables
41501B Option 420Add 2 medium power SMUs and cables
41501B Option 422Add 2 medium power SMUs, 2 PGUs, and cables
41501B Option 902Cable, power, Europe
41501B Option 903Cable, power, US and Canada
4156CPrecision semiconductor parameter analyzer
4156C Option 010Delete all 4156C cables
4156C Option 020Delete Windows controller for pa rameter analysis and
characterization
4156C Option 2001.5 meter interlock, 4 coaxial, 4 triaxial cables
4156C Option 2303.0 meter interlock, 4 coaxial, 4 triaxial cables
4284APrecision LCR meter
4284A Option 001Add DC amplifier
4284A Option 006Add 2 meter/4 meter cable operation
4284A Option 909Rack mount kit
4284A Option ABAEnglish documentation
4284A Option ABJJapanese documentation
5063-9220Rack mount kit with handles, 2-EIA
70Installation and User’s Guide
Table 18 Replaceable Parts (continued)
Part or Model N umberDescription
5063-9221Rack mount kit with handles, 3-EIA
5063-9222Rack mount kit with handles, 4-EIA
5063-9223Rack mount kit with handles, 5-EIA
5063-9224Rack mount and handle kit, 6-EIA
5063-9225Rack mount and handle kit, 7-EIA
8120-1396Line power cord, 220V
8120-1839Cable, BNC, 50 ohm, 24 inch
8120-1405Line power cord, 120V
8120-1840Cable, BNC, 48 inch
8120-2582Cable
8120-5068Cable
Installing the System2
8490D Option 010Attenuator, 2.4 mm coaxial, fixe d 10 dB, DC to 50 GHz
85043-20001Ground stud
85043-20002Shoulder screw
85043-80013Anti-static mat kit
85056APrecision calibration kit, 2.4 mm
85056DEconomy calibration kit, 2.4 mm
85106-60038Work surface, 1 meter
85107-20004Semi-rigid cable, 9 inc h, 2.4 mm (m-m)
85133F2.4 mm flexible test port cable set
85225-90023Agilent 85225F Performance Modeling System Installation and
User’s Guide
C2790ACBallast, 30 pounds
E3661BRack cabinet, 1.6 me ter
E3661B Option AW3Power distributio n unit, 100/120 volts
E3661B Option AW5Power distributio n unit, 220/240 volts
E3663ACRail kit (2 rails per)
E3668BFeedthrough panel
E4470AZExtractor fan, 100 to 120 volts
E4471AZExtractor fan, 200 to 240 volts
E5250ALow leakage switch mainframe
Installation and User’s Guide71
2Installing the System
Table 18 Replaceable Parts (continued)
Part or Model N umberDescription
E5252A10 x 12 switch mat rix
E5260A8-slot high spe ed parametric measurement mainframe
E5260A Option 05050 Hz line power frequency
E5260A Option 06060 Hz line power frequency
E5260A Option ABAEnglish documentation
E5260A Option ABJJapanese documentation
E5290AHigh speed high power source mon i tor unit
E5291AHigh speed medium power source monitor unit
E5270B 8-slot parametric measurement solution
E5270B Option 05050 Hz line power frequency
E5270B Option 06060 Hz line power frequency
E5270B Option ABAEnglish localization
E5270B Option ABJJapanese localization
E5280APrecisio n high power source monitor unit
E5281APrecisio n medium power source monitor unit
E5286AHigh resolution source monitor unit
E5810ALAN/GPIB gateway
E5810A Option 100Rack mount kit
E5810A Option AG6I/O libraries client softw a re for MS Windows
E5810A Option ABJJapanese documentation
E7731AFiller panel, 1-EIA unit
E7732AFiller panel, 2-EIA unit
E7733AFiller panel, 3-EIA unit
E7734AFiller panel, 4-EIA unit
E7735AFiller panel, 5-EIA unit
E7736AFiller panel, 6-EIA unit
E7737AFiller panel, 7-EIA unit
E8364BPNA Series vector network analyzer, 10 MHz to 50 GHz
E8364B Option 010Time domain analysis capability
E8364B Option 014Configurable test set
E8364B Option 016Receiver attenuators
If the system will not be used with a probe station, install the work
surface to facilitate in- fixture or coaxial measurements.
The work surface is designed to fit onto the rack below the network
analyzer. When the following procedure is performed properly, the work
surface maintains a path to chassis ground through the support rails and
the rack cabinet.
Follow these instructions to install the work surface.
Required Tools
• Medium Pozidriv screwdriver
• Small flat- blade screwdriver
To install the work surface
StepActionNotes
1 Prepare to install the work
surface.
a Fully extend the four lock feet at th e bottom
of the rack cabinet.
b Unpack the work surface and the work
surface support rails.
c Lay the rails down so that:
• the ends with the single pemmed hole
are facing the front,
• the ends with the keyhole-shaped cutout
are facing the rear,
• and the rails are facing inward toward
each other.
2 Attach the support rails to
the rack cabinet.
a Pass the large end of the keyhole-shaped
cutout in each rail over the shoulder screw
already mounted inside the rack.
b Slide the rails to the rear of the rack.
c Use one 1/2 inc h long 10 -32 Pozidr iv scre w ,
one split lock washer, and one flat washer
to secure each rail.
d Before tightening th e screws, ensure that
the rails are level.
e Tighten the screws.
• The lock feet prevent the cabinet from
moving on the casters.
• These are the positions of the support
rails when installed in the rack.
• You may need to hold the rails in place
as you tightened the screws.
74Installation and User’s Guide
To install the work surface (continued)
StepActionNotes
Installing the System2
3 Attach the work surface
to the support rails.
a Slide the work surface onto the support
rails.
b Slide the work surface all the way back on
the rails until it comes to rest against the
front of the rack c a binet.
c Use 1/2 inch long 10-32 screws to secure
the work surface to th e rails from the
beneath.
Installation and User’s Guide75
2Installing the System
To ensure your safety while using the system
This product has been designed and tested in accordance with
international standards. Bias current and voltage are supplied to the DUT
from the DC subsystem. This instrument can force dangerous voltages to
the FORCE, SENSE, and GUARD connectors. DC subsystem is connected to
the device through the bias networks and test fixture or probe station.
WARNING
Failure to comply with the following precautionary safety ins truct ions prior to
operating the system could result in serious injury or death.
With some installed options, the Agilent 4156C or Agilent E5260A/70B used in
this system can supply voltages up to ±200 volts DC. Depending on operating
conditions, hazardous voltages can be present at points in the system that
could potentially come in contact with the system operator.
Before operating the system , follow thes e step s to ensur e your saf et y.
To ensure your safety while using the system
StepActionNotes
1 Never operat e the sy st em
without a safety earth
ground.
2 Never attempt to service
the system.
a Ensure that a safety earth ground is
connected between the system power
distribution unit and the line power source.
b If it is likely that the safety earth ground
has been impaired, the system must be
rendered inoperative and secured against
unintended operation .
a Contact Agilent Technologies if service is
required.
• Capacitors within the system
components can remain charged even
after the system is disconnect ed from
its line power source.
• The system may only be serviced,
adjusted, maintained, or repaired by
qualified personnel.
3 Open the DC subsystem
interlock connection
whenever possible.
a Close the DC subsystem INTLK (Interlock)
connection only when voltages greater
than ±42 volts DC are required.
• Depending on installed options, the
SMU output can be as high as
±200 volts DC. As long as the INTLK
connection is open, the voltage is
clamped to ±42 volts DC maximum.
• For instruction on installing an
interlock switch on a sh ielding box, se e
“To Make an Interlock Connection” in
the 4156C user’s guide (volume 1) or
“Connecting the Interlock Terminal” in
chapter 3 of the E5270 user’s guide.
76Installation and User’s Guide
To ensure your safety while using the system (continued)
StepActionNotes
Installing the System2
4 Be aware of potential
shock hazards during
floating-ground
measurements.
5 Before touching the
FORCE, SENSE, or GUARD
connectors, ensure your
safety.
6 Never use replacement
fuses with incorrect
ratings.
a Do not touch any of the DC subsystem
output connectors when the shorting bar is
disconnected and a floating-ground
measurement is in progress.
b Warn others working in the system’s
vicinity of the potential shock hazards .
a Switch off the DC subsystem and discharge
the capacitors.
b If you do not switch off the instruments,
complete ALL of the following
precautionary steps:
• Terminate the DC subsystem
measurement by pressing the Stop key
and confirmi ng t hat t he MEA SUREMEN T
indicator is not lit.
• Deacti vate th e st and by mo de (if use d) b y
pressing the Standby key and con fir ming
that the Standby indicator is not lit.
• Confirm that the HIGH VOLTAGE
indicator is not lit.
• Open the interlock connection.
a After fin d ing the cause of failure, replace
component fuses with fuses of the same
current rating and of the type specified in
the instrument’s product documentation .
• For additional information, see the
following section “Precautions for
Performing Floating-Ground
Measurements.”
• Failure to use correctly rated fuses
could result in a fire hazard and
damage to the equipment .
7 Install the instrument so
that the ON/OFF switch is
readily identifiable and
easily reached by the
operator.
Installation and User’s Guide77
• The ON/OFF switch is the system
disconnecting device. It disconnects
the mains circuit from the mains supply
before other parts of the instrument.
• Alternately, an externally installed
switch or circuit breaker (readily
identifiable and easily reached by the
operator) may be used as a
disconnection device.
2Installing the System
Precautions for Performing Floating-Ground Measurements
IC- CAP measurements can be performed with the device in a
floating- ground configuration. This prevents ground- loop noise and, in the
case of a bipolar junction transistor, damage to the device under test.
A floating- ground configuration is created by removing the shorting bar
that connects the CIRCUIT COMMON and CHASSIS GROUND terminals.
WARNING
A potential shock hazard exists when the shorting bar is disconnected for
floating-ground measurements. Do not touch any of the DC subsystem rear
panel connectors while a floating ground measurement is in progress.
If you are making measurements in a floating- ground configuration, ensure
that the shorting bar is disconnected between the CIRCUIT COMMON and
CHASSIS GROUND terminals.
To perform floating-ground measurements
When floating ground measurements are necessary, remove the rear front
panel shorting bar connecting the CIRCUIT COMMON and CHASSIS
GROUND terminals.
When the shorting bar is removed, you must drive the DUT circuit
common with either an SMU, GNDU, or by connecting directly to the DC
subsystem circuit common. The circuit common can be found at the DUT
ends of the SMU and GNDU cables.
NOTE
The circuit common is not connected through the bias networks.
Read “Precautions for Performing Floating- Ground Measurements" on
page 78, then follow these steps to connect the CIRCUIT COMMON to an
external ground.
To connect an external ground to the circuit common
Step
1 Remove the shorting bar connecting the CIRCUI T COMMON and CHASSIS
GROUND terminals.
2 Connect the external ground to the CIRCUIT COMMON of the DC subsystem.
78Installation and User’s Guide
Precautions for Avoiding Electrostatic Discharge
Never operate the system without taking precautions to avoid electrostatic
discharge that could damage the system or the device under test.
Installing the System2
CAUTION
Even relatively small currents resulting from electrostatic discharge
undetectable to the system operator can damage current-sensitive devices and
system components.
To take precautions against electrostatic discharge
Step
1 Wear an antistatic wrist strap.
2 Connect the wrist strap to chassis ground.
Installation and User’s Guide79
2Installing the System
To connect the bias networks
Follow these steps to connect the bias networks to the system, and the
device under test to the bias networks.
Agilent 4156C Systems
To connect the bias networks
StepActionNotes
1 For Agilent 4156C
systems, refer to the
following figure.
2 Connect the triaxial
cables from the 4156C to
the bias networks.
a Connect the triaxial cables from the 4156C
HRSMU1 FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 1 bias network.
b Connect the triaxial cables from the 4156C
HRSMU2 FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 2 bias network.
• From the rear of the system, route the
cable through the feedthrough panel.
• The port 1 bias network is the 11612V
K11.
• The port 2 bias network is the 11612V
K21.
• Refer to Figure 5 on page 29.
80Installation and User’s Guide
To connect the bias networks (continued)
StepActionNotes
Installing the System2
3 Connect and route the
triaxial cable from 4156C
HRSMU3 FORCE
connector to the port 2
bias network.
4 Connect the Agilent
E8364B test ports to the
bias networks.
a From the rear of the system, connect one
end of the triaxial cable to the HRSMU3
FORCE connector on the 4156C rear panel.
b From the rear of the system, route the cable
through the upper feedthro ug h pa nel.
c Connect the triaxial cable to the port 2 bias
network GNDU conne ct or.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male
adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the
11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male
adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN
connector of the 11612V K21 bias network.
• The triaxial cable model number is
16494A Option 002.
• Leave the GNDU connector on the port
1 bias network open.
• The test port cables model number is
85133F.
• The 2.4 mm female-to-male adapter
model number is 11900C. Without this
adapter, the test port cable will not
properly mate with the bias network
RF IN connector.
• The 2.4 mm male-to-male adapter
model number is 11900A.
5 Connect the bias
networks to the devi ce
under test.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias
network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias
network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias
network.
• The semi-rigid cables part number is
85107-20004.
Installation and User’s Guide81
2Installing the System
Agilent 4156C Systems with Agilent 41501B Expander Box
To connect the bias networks
StepActionNotes
1 For Agilent 4156C with
41501B expander box
systems, refer to the
following figure.
2 Connect the triaxial
cables from the 4156C to
the bias networks.
3 Connect and route the
triaxial cable from 4156C
HRSMU3 FORCE
connector to the port 2
bias network.
a Connect the triaxial cables from the 4156C
HRSMU1 FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 1 bias network.
b Connect the triaxial cables from the 41 501B
HPSMU FORCE and SENSE connectors to
the DC FORCE and DC SENSE connectors
on the port 2 bias network.
a From the rear of the system, connect one
end of the triaxial cable to the GNDU
connector on the 41501B rear panel.
b From the rear of the system, route the cable
through the upper feedthro ug h pa nel.
c Connect the triaxial cable to the port 2 bias
network GNDU conne ct or.
• From the rear of the system, route the
cable through the feedthrough panel.
• The port 1 bias network is the 11612V
K11.
• The port 2 bias network is the 11612V
K21.
• Refer to Figure 20,“Rear Panel Wiring
Diagram,” on page53.
• The triaxial cable model number is
16494A Option 002.
• Leave the GNDU connector on the port
1 bias network open.
• Refer to Figure 3,“Front Panel
Connections with Agilent 4156C,” on
page25.
82Installation and User’s Guide
To connect the bias networks (continued)
StepActionNotes
Installing the System2
4 Connect the Agilent
E8364B test ports to the
bias networks.
5 Connect the bias
networks to the devi ce
under test.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male
adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the
11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male
adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN
connector of the 11612V K21 bias network.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias
network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias
network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias
network.
• The test port cables model number is
85133F.
• The 2.4 mm female-to-male adapter
model number is 11900C. Without this
adapter, the test port cable will not
properly mate with the bias network
RF IN connector.
• The 2.4 mm male-to-male adapter
model number is 11900A.
• The semi-rigid cables part number is
85107-20004.
Installation and User’s Guide83
2Installing the System
Agilent E5260A/70B Systems
To connect the bias networks
StepActionNotes
1 For Agilent E5260A/70B
systems, refer to the
following figure.
2 Connect the triaxial
cables to the E5260A/ 70B
*
SMUs.
a From the front of the system rack, connect
triaxial cables to t he fron t panel h igh powe r
SMU FORCE and SENSE outputs on the
E5260A/70B.
b Connect a triaxial cable to the f r ont panel
medium powe r SMU F O RCE and SENSE
outputs on the E5260A/70B.
c From the front of the system, route the
cable through the upper feedthrough panel.
d From the rear of the system, route the
HPSMU cable through the port 2 hole in the
lower feedthrough pane l.
e From the rear of the system, route the
MPSMU cable throu g h the port 1 hole in
the lower feedthrough panel.
• The triaxial cable model number is
16494A Option 002.
• The high speed high power SMU is the
Agilent E5290A.
• The high speed medium power SMU is
the Agilent E5291A.
• The high power SMU is the Agilent
E5280A.
• The medium power SMU is the Agilent
E5281A.
• Refer to Figure 4 on page 27
84Installation and User’s Guide
To connect the bias networks (continued)
StepActionNotes
Installing the System2
3 Connect the triaxial
cables from the
E5260A/70B SMUs t o the
bias networks.
4 Connect and route the
ground cable fr om
E5260A/70B GNDU to th e
port 2 bias network.
5 Connect the Agilent
E8364B test ports to the
bias networks.
a Connect the triaxial cable from the
E5260A/70B medium powe r SMU to the
DC FORCE and DC SENSE connectors on
the port 1 bias network.
b Connect the triaxial cable from the
E5260A/70B high power SMU to the DC
FORCE and DC SENSE con nectors on the
port 2 bias network.
a From the front of the system , conne ct one
end of the ground triax ial cab le to th e
GNDU connector on the E5260A/70B front
panel.
b From the front of the system, route the
cable through the upper feedthrough panel.
c From the rear of the system, route the
ground triaxial cable through the port 2 hole
in the lower feedthrough panel.
d Connect the ground triaxi al cabl e to the
port 2 bias network GNDU connector.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male
adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the
11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male
adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN
connector of the 11612V K21 bias network.
• The medium power SMU is the Agilent
E5281A.
• The high power SMU is the Agilent
E5280A.
• The port 1 bias network is the 11612V
K11.
• The port 2 bias network is the 11612V
K21.
• The ground triaxial cable model number
is 16493L Option 002.
• Leave the GNDU connector on the port
1 bias network open.
• The test port cables model number is
85133F.
• The 2.4 mm female-to-male adapter
model number is 11900C. Without this
adapter, the test port cable will not
properly mate with the bias network
RF IN connector.
• The 2.4 mm male-to-male adapter
model number is 11900A.
Installation and User’s Guide85
2Installing the System
To connect the bias networks (continued)
StepActionNotes
6 Connect the bias
networks to the devi ce
under test.
* Refer to Figure4, “Front Panel Wiring Diagram with Agilent E5260A or E5270B,” on page27.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias
network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias
network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias
network.
• The semi-rigid cables part number is
85107-20004.
86Installation and User’s Guide
To switch on power to the system
Installing the System2
NOTE
To switch on power to the system
StepNotes
1 Ensure that the individual instru ments are
configured to match the available line power
*
source.
2 Ensure that all component line power
switches are set to the OFF position.
3 Connect the system to line power.
4 Switch the rack cabinet ~ Line switch from
to (from standby to energized).
5 Switch on the low leakage switch mainframe
line power.
6 Switch on the precision LCR meter line
power.
7 If present, on the 4284A SYSTEM CONFIG
page, highlight the value in th e GPIB
ADDRESS field, and press 24 > Enter.
Proper system operation is depe ndent on the sequence in which the
system components are switched on.
• For information regarding line power
settings, refer to the individual instrument
manuals.
• This sets the correct system GPIB address
(24) for the 4284A.
8 If present, on the 35670A front panel, press
Local/GPIB > ANALYZER ADDRESS > 22 >
ENTER.
9 If present, switch on the Agilent 4156C
precision semiconductor parameter analyzer
line power.
10 If present, switch on the 41501B expander
line power.
11 Switch on the Agilent 4156C precision
semiconductor parameter analyzer line
power.
12 On the 4156C, press System >
[MISCELLANEOUS], move the pointer to the
POWER LINE FRE QU E NCY field, and press
[50 Hz] or [60 Hz].
Installation and User’s Guide87
• This sets the correct system GPIB address
(22) for the 35670A.
• The expander must be switched on before
the 4156C.
• Ensure that th e 41 501B has already been
switched on prior to activating the 4156C.
• This ensure s that the 4156C is configured to
match the available line power frequency.
• The value is set to 60 Hz at the factory.
• Use th e front panel arrow keys t o mov e the
cursor.
2Installing the System
To switch on power to the system
StepNotes
13 On the 4156C, press System >
[MISCELLANEOUS], hi ghlight the 4156C
value in the GPIB ADDR ESS field, and press
19 > Enter.
14 On the E5260A/70B, move the cursor to
CONFIG, pres s Enter, move the cursor to
ADDRESS, press Enter. Use the arrow keys
to set the address to 19 and press Ent er.
15 Switch on the Agilent E8364B PNA Series
vector network analyzer line power.
16 On the Agilent E8364B PNA Series v ector
network analyzer, from the Main dialog,
select System > Configure > SICL/GPIB. In
the SICL/GPIB dia log GPI B group b ox, se lect
the Talke r/ L istener radio button and select
16 in the Address scroll list.
17 If present, switch on the LAN/GPIB gateway
line power.
18 Switch on the computer line power.
19 Allow the system to warm up for one hour.
* If the system is to be use d with an autotransformer, ensure that t he common terminal is connected to the
neutral (gr ou nded) side of the powe r source.
• This sets the correct system GPIB address
(19) for the 4156C.
• This sets the correct system GPIB address
(19) for the E5260A/70B.
• This sets the correct system GPIB address
(16) for the E8364B.
Refer to the LAN/GPIB gate way documenta tion
for instruction on installation and configuration.
88Installation and User’s Guide
Installing the System2
To configure the LAN/GPIB gateway for functional verification
If your system includes an Agilent E5810A LAN/GPIB gateway, follow
these steps to configure the LAN/GPIB gateway in order to verify the
functionality of the performance modeling system.
NOTE
This procedure explains how to configure IC-CAP to use the default
LAN/GPIB gateway server IP address in order to verify the
functionality of the performance modeling system.
After functional verification, contact your corporate IT professional and
ask for a permanent server IP address assignment for the LAN/GPIB
gateway. For instructions on changing the server IP address, see the
LAN/GPIB gateway installation and configuration guide.
To configure the LAN/GPIB gateway
StepActionNotes
20 Start the IC-CAP
software.
21 Add the interface to the
IC-CAP Hardware Setup.
*
PC version:
a Select Start > Programs > IC-CAP 2004 >
IC-CAP.
UNIX version:
a Open a UNIX terminal window.
b At the prompt, ty pe iccap.
c Press Enter.
a From the IC-CAP/Main window menu bar,
choose Tools > Hardware Setup... .
b Below the HP-IB Interface group box, click
Add Interface.
c In the Add HP-IB In terface dialog b ox, enter
lan[192.0.0.192]:hpib.
d Click OK.
†
• This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main
windows.
• This opens the IC-CAP/Hardware
Setup window.
• This opens the Add HP-IB dialog box.
• This configures the LAN/GPIB
gateway (with its default se rver
address) as the performance modeling
system GPIB interface.
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model numbe r
85190D.
† If there is an existing IC-C AP i nterface (for example, HP-IB), select the existing interface and click Delete Interface before continuing to
the next action.
This completes the installation process. To confirm the functionality of the
system, continue to Chapter 3, “Verifying System Functionality,” starting
on page 91.
Installation and User’s Guide89
Agilent 85225F Perfor mance Model ing System
Installation and User’s Guide
3
Verifying System Functionality
To choose a verification process 92
Understanding the System Functional Verification Test 94
Performing the System Functional Verification Test 95
Related Topics“To enhance measurement accuracy" on page 112
“Performing a Coaxial System Measurement Calibration" on page 115
“Performing the DC Subsystem Functional Verification Test"on page 121
“Performing the RF Subsystem Functional Verification Test" on page 125
Use the procedures in this chapter to verify the functionality of the
Agilent 85225F performance modeling system. This chapter includes
procedures for choosing varying degrees of functional verification and
performing the required post- installation system functional verification
test.
Agilent Technologies
91
3Verifying System Functionality
To choose a verification process
System functionality can be verified using several different processes,
depending on the level (system or subsystem) of functional verification
required and the available tools.
The procedure provided in “Performing the System Functional Verification
Test" on page 95 verifies that all of the system instruments interface
correctly, and that the system can make software- driven measurements
using a controller running the IC- CAP software.
Manual functional verification procedures for DC, RF, CV, and 1/f noise
subsystem components can be found in:
• Appendix D, “CV Subsystem Functional Verification Test,” starting on
page 129
• Appendix E, “1/f Noise Subsystem Functional Verification Test,” starting
on page 133
NOTE
To choose a system verification process
SituationActionNote
1 If the system has recently
been installed or one of
the DC or RF subsystem
instruments has been
replaced, and you have
IC-CAP software...
These tests do NOT verify that the system instrumentation conform to their individual
performance specifications.
To verify the performance of the individual system components, complete the appropriate
performance tests listed in their individual product documentation.
Follow these steps to choose a system functional verification process based
upon your current situation.
Complete the steps listed in “Performing the
System Functional Verification Test"on
page95.
• Completion of the System Functional
Verification Test is required after
system installation, or whenever an RF
or DC subsystem component has been
serviced or replaced.
• The System Functional Verification Test
verifies the functionality of instruments
in the RF and DC subsystems. To verify
the functionality of other system
components, continue to the
appropriate situation listed in this
table.
92Installation and User’s Guide
To choose a system verification process (continued)
SituationActionNote
Verifying System Functionality3
2 If you would like to
manually check the
functionality of the
Agilent E8364B PNA
Series vector ne twork
analyzer without using
the GPIB interface...
3 If you would like to
manually check the
functionality of the
Agilent 4156C precision
semiconductor parameter
analyzer without using
the GPIB interface...
4 If you would like to
manually check the
functionality of the
Agilent E5260A/70B
without using the GPIB
interface...
5 If you would like to
manually check the
functionality of the
Agilent 428 4A wi th out
using the GPIB interface...
Complete the steps listed in “Performing the
RF Subsystem Functional Verification Test"on
page125.
Complete the steps listed in “Performing the
DC Subsystem Function al V erifica tion Test" on
page121.
Complete the steps listed in “Performing the
DC Subsystem Function al V erifica tion Test" on
page121.
Complete the steps listed in “Performing the
CV Subsystem Functional Verification Test"on
page131.
6 If you would like to
manually check the
functionality of the
Agilent 35670A without
using the GPIB interface...
7 If you would like to
manually check the
functionality of the
Agilent E5250 low leakage
switch mainframe without
using the GPIB interface...
Installation and User’s Guide93
Complete the steps listed in “Performing the
1/f Noise Subsystem Functional Verification
Test" on page135.
Complete the Agilent E5250A self-test found in
Chapter 3 of the low leakage switch
mainframe user’s guide.
3Verifying System Functionality
Understanding the S ystem Functional Verification Test
The system functional verification test is a standard IC-CAP measurement
and simulation procedure. The system performs DC and S- parameter
extraction, optimization, and simulation using a fixed 10 dB attenuator as
the device under test (DUT).
The procedure provided in “Performing the System Functional Verification
Test” confirms:
• the IC- CAP software can communicate with and control the system
instrumentation via GPIB (through the LAN/GPIB gateway, if so
configured)
• the system can make measurements and display the results
• the IC- CAP software can simulate data
• the IC-CAP software can converge the simulated data with the
extracted (measured) data
The system applies forward and reverse current to the attenuator and
monitors voltage at the attenuator’s input and output. IC- CAP then uses
the measured data to model the device- intrinsic resistances and
transmission line delay.
Required Tools
• Agilent 85225F performance modeling system
• A system controller
• Agilent 85190A IC- CAP software
• Test port cables
• Agilent 8490D 10 dB fixed RF attenuator
• BNC tee (2)
• Agilent 11900A, 2.4 mm male- to- male adapter
• Agilent 11900B, 2.4 mm female- to- female adapter
• Agilent 11900C, 2.4 mm female- to- male adapter
• Agilent 85056A 2.4 mm precision calibration kit, or
• Agilent 85056D 2.4 mm economy calibration kit
*
†
‡
* For requirements, see “The System Controller" on page 59 .
† These compon en t s ar e s up p li e d a s part of the system.
‡ For CV subsystem verification only
94Installation and User’s Guide
Performing the System Functional Verification Test
Complete the following steps to verify system functionality using the
supplied Agilent 8490D 10 dB fixed RF attenuator as the device under
test.
To perform the system functional verification test
StepActionNotes
Verifying System Functionality3
1 Switch on power to the
system.
2 Connect the device under
test to the bias networks.
a Complete the steps listed in “To switch on
power to the system"on page87.
a Refer to the following figure.
• Proper system function is dependent
upon the order in which the system
components are switched on.
b Connect one end of the 2.4 mm
female-to-female ada pter to the semi-rigid
cable attached to the RF/DC OUT
connector of the port 1 bias network.
c Connect male end of the attenuator to the
other end of the 2.4 mm female-to-female
adapter.
d Connect the female e nd of the attenuator to
the semi-rigid cable attached to the RF/DC
OUT connector of the port 2 bias netwo rk .
Installation and User’s Guide95
• The 2.4 mm female-to-female adapter
is a 11900B. Use the 11900B provided
in the calibration kit.
• The port 1 bias network is a 11612V
K11.
• The port 2 bias network is a 11612V
K21.
3Verifying System Functionality
To perform the system functional verification test (continued)
StepActionNotes
3 Start the IC-CAP
software.
4 Add the system inte rfa ce
and components to the
IC-CAP Hardware Setup.
5 For 4156C systems,
change the Agilent 4156C
precision semiconductor
parameter analyzer
HRSMU names in IC-CAP.
*
PC version:
a Select Start > Programs > IC-CAP 2004 >
IC-CAP.
UNIX version:
a Open a UNIX terminal window.
b At the prompt, ty pe iccap.
c Press Enter.
a From the IC-CAP/Main window menu bar,
choose Tools > Hardware Setup... .
b If you have not added an interface, click
Add Interface, enter the system interface
(hpib or other), and click OK.
c In the Instrument List, click Rebuild.
d Verify that all system components appear
in the Instrument List.
a In the Instrument List, select HP4156 (hpib,
19) and click Configure... .
b In the Configuration of HP4156 window
Unit Table group box, highlight the
characters in the HRSMU1 ent r y bo x and
type VG.
c In the Unit Table group box, highlight the
characters in the HPSMU entry box and
type VD.
d Click OK.
e Close the Hardware Setup window.
†
• This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main
windows.
• This opens the IC-CAP/Hardware
Setup window.
• This polls the GPIB and adds all
connected and activated system
instruments to the Instrument List.
• Disregard error messages on the
system instrument displays. The errors
are a by-product of the GP IB po l li ng
process.
• This ope ns the Configuration of
HP4156 window where the Unit Table
dialog box is used to change t he names
of the HRSMUs.
• Actions b, c, and d change the name of
HRSMU1 to VG and HPSMU to VD.
• Renaming the SMUs is necessary for
proper execution of the example model
file.
6 For E5260A/70B syst ems,
change the Agilent 4156C
precision semiconductor
parameter analyzer SMU
names in IC-CAP.
96Installation and User’s Guide
a In the Instrument List, select Agilent E5270
(hpib, 19) and click Configure... .
b In the Configuration of E5270 window Unit
Table grou p bo x, hig hligh t th e charact ers in
the MPSMU<slot number> entry box and
type VG.
c In the Unit Table group box, highlight the
characters in the HPSMU<slot number>
entry box and type VD.
d Click OK.
e Close the Hardware Setup window.
‡
• This ope ns the Configuration of E5270
window where the Unit Table dialog
box is used to chan ge th e names of th e
SMUs.
• Actions b, c, and d change the name of
MPSMU<slot number> to VG and
HPSMU<slot number> to VD.
• Renaming the SMUs is necessary for
proper execution of the example model
file.
To perform the system functional verification test (continued)
StepActionNotes
Verifying System Functionality3
7 Open the attenuator test
model in IC-CA P.
8 Set the model variables
for the measur ement in
IC-CAP.
9 Set the network analyzer
instrument options for the
attenuator test model in
IC-CAP.
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../examples/model_files.
c In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../model_files/misc.
d In the Files list of the File Open dialog,
double-clic k on sys_testrf.mdl.
a Click the Model Variables tab folder.
b Highlight the freq_start v ariable and enter
4.5E+07.
c Highlight the freq_stop variable and enter
50E+09.
d Highlight the imax variable and enter 90m.
a In the DUTs-Setups tab folder , click
S_vs_freq in the Select DUT/Setup list.
b Click the Instrument Options tab folder.
c Highlight the Cal Type value and enter H.
d Highlight the Ca l File Name value and ent er
TEST.CST.
• This opens the File Open dialog box.
• This ope ns a list of model files.
• (Scroll th e list , if nee ded.) Th is dis plays
a list of modeling f iles i n the Fil es list o f
the File Open dialog.
• This opens the Atten model window.
• The Atten model window contains tab
folders used to interact with the model
data.
• This ope ns the Model Variables tab
folder.
• This sets the start frequ ency of the
model to 45 MHz.
• This sets the stop frequency of the
model to 50 GHz.
• This se ts the maximum current to
90 milliamps.
• This opens the 5 tab folders used for
the S_vs_freq setup.
• This ope ns the Instrument Options tab
folder.
• Adjustable instrument parameters are
listed in group boxes titled as
instrument model number.bus address.
instrument address (for example,
AgilentPNA.7.16 for the network
analyzer group box).
• Terminate your value entries by
pressing Enter on the controller
keyboard.
Installation and User’s Guide97
3Verifying System Functionality
To perform the system functional verification test (continued)
StepActionNotes
10 Configure the instrument
state for calibration.
11 Calibrate the network
analyzer.
a Disconnect the attenuator from the
semi-rigid cables.
b In the Utility key group on the network
analyzer front panel, press Macro/Local >
Preset.
c In the Channel menu, click Power... .
d In the Power dialog box, scroll the Test Port
Power value to −20 dBm, then click OK.
e In the Sweep menu, point to Number of
Points and click 101.
f In the Sweep menu, select IF Bandwidth. .. .
g In the IF Bandwidth dialog box, scroll to
1.000 kHz, then click OK.
h In the Sweep menu, select Sweep Setup... .
i In the Sweep Setup dialog box, select the
Stepped Sweep check box and click OK.
a In the network analyzer’s Calibrati on menu,
select Calibra tion Wizard... .
b In the Calibration Wizard: B egin C alibratio n
dialog box, select SmartCal (GUIDED
Calibration): Use Mechanical Standards
radio button and click Next.
c Follow the displayed prompts to calibrate
the network analyzer.
• This removes th e instrument from GPIB
control, activates the front panel
interface, and returns the instrument to
its factory preset condition.
• Actions c through h prepare the
network analyzer for a full range 2-port
calibration using the same instrument
state settings used by the IC-CAP
functional verification test macro.
• For detailed instructions, see
“Performing a Coaxial System
Measurement Calibration"on
page115.
• This begins a modified full 2-port
calibration.
• When prompted to connect a standard
to either Port 1 or Port 2, connect the
standard to the semi-rigid cable
attached to the Por t 1 or Port 2 bias
network.
• An isolation calibration is not needed
for this measurement.
12 Save the calibration and
instrument state data to
the C:/Program Files/
Agilent /Network
Analyzer/Documents
folder.
98Installation and User’s Guide
a In the File menu, select Save As... .
b In the Save As di alog box, using the
keyboard or by clicking Edit File Name,
enter TEST.CST then click OK.
• This saves the calibration and
instrument state data in the network
analyzer’s operating system
C:/Program Files/Agilent/
Network Analyz er/ D ocu m ent s folder.
To perform the system functional verification test (continued)
StepActionNotes
Verifying System Functionality3
13 Make the DC and RF
measurements in IC-CAP.
14 Interpret the results.a Observe the displayed plots.
a Reconnect the attenuator.
b In the Atten model window, click the
Macros tab.
c In the Select Macro: list, select Test_atten
and click Execute.
d When prompted to ensure you have
calibrated the network analyzer, enter Y
and click OK.
e When prompted to enter a cal set number
use the default (1) and click OK.
f As IC-CAP performs an optimization,
observe the simulated data trace converge
with the measured data trace.
b In the IC-CAP/Status window, observe the
final DC/RF values parameter and the Final
RMS error.
• This opens the Atten model window’s
Macros tab folder.
• This starts a sequence of prompts and
responses.
• IC-CAP performs measurements of the
attenuator’s DC and RF parameters.
• When complete, IC-CAP displays plots
of the forward and reverse DC voltage
transfer through the attenuator. The
solid line is the measured data, the
dashed line is the simulated data.
• IC-CAP be gins the optimization process
using default simulated data. These
values are optimized to converge with
the measured data.
• When the S-parameter measurement is
complete, IC-CA P displays plots of the
S12/21, 20×log
magnitude, and S21 phase.
of the S21
10
**
• IC-CAP also lists the attenuato r’s
resistance values. The error between
measured and simulated data should
be less than 2%.
• The error between the measured and
simulated S21 phase data should be
less than 2%.
• The value for T1.TD is the transmission
time through the attenuator , modeled
as transmission line delay. Thi s v alue
(typically in the femtosecond to
nanosecond range) de pends on the
length of the attenua tor.
• The S21 magnitude and the S12/S21
plots, of less significance, are included
for your interest.
• The S21 simu lated trace is determined
from the measured DC resistances, and
therefore is not expected to converge
with the measured data over the full
frequency range. However , the
measured S21 data will show a normal
frequency response variation.
• The S12/S21 plot (also displayed on
the network analyzer) confirms that the
system is capable of making forward
and reverse transmission
measurements.
Installation and User’s Guide99
3Verifying System Functionality
To perform the system functional verification test (continued)
StepActionNotes
15 For systems with the
Agilent 4284A precision
LCR meter, open the
junction capacitance
model (juncap.mdl) in
IC-CAP.
16 Select the DUT/Setup.a Select the DUTs-Setups tab.
17 Set the LCR meter
instrument options.
18 Calibrate the LCR meter.a On the 16048D test leads, connect one tee
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../examples/model_files.
c In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../model_files/diode.
d In the Files list of the File Open dialog,
double-clic k on juncap.mdl.
b Click the area > cv setup.
a Select the Instrument Options tab.
b Highlight the Cable Leng th variable and
enter 2.
between the H
b Connect the other tee between the L
L
connectors. Select the
cur
Measure/Simulate tab.
c Select Calibrate.
d Follow the IC-CAP prompts.
pot
and H
connectors.
cur
pot
• This opens the File Open dialog box.
• This ope ns a list of model files.
• (Scroll th e list , if nee ded.) Th is dis plays
a list of modeling f iles i n the Fil es list o f
the File Open dialog.
• This opens the juncap model window.
• The juncap model win dow contai ns tab
folders used to interact with the model
data.
• This opens the capacitance vs. voltage
setup.
• This se ts the cable length in the LCR
meter instrument options.
• This perfor ms a calibration on the LCR
meter.
and
• The BN C tee part number is 1250-2405.
19 Measure the open circuit
capacitance.
20 Observe the resultsa After the calibration is complete, observe
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model numbe r
85190D.
† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
**S12 is identical to S21 because the attenuator is assumed to be symmetrical.
a Select Measure.
the plot.
b Rescale the measured result (displayed in
red) cap.m.
• This measu res the open circuit
capacitance.
• The measured result should be less
than ±10E-15 fara ds .
This completes the functional verification procedure.
100Installation and User’s Guide
If you encounter a problem
✔ Check the system connections and settings:
1 system connections to the DUT
2 system interconnections
3 GPIB cabling
4 GPIB address settings
✔ Perform the DC subsystem self- test in Appendix B, “DC Subsystem
Functional Verification Test,” starting on page 119.
✔ Perform the Agilent E8364B PNA Series vector network analyzer
operator’s check in Appendix C, “RF Subsystem Functional Verification
Test,” starting on page 123.
✔ Perform the Agilent 4284A precision LCR meter self- test in
Appendix D, “CV Subsystem Functional Verification Test,” starting on
page 129.
Verifying System Functionality3
✔ Perform the Agilent 35670A dynamic signal generator self- test in
Appendix E, “1/f Noise Subsystem Functional Verification Test,” starting
on page 133.
Installation and User’s Guide101
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.