Atec Agilent-85133F User Manual

Agilent 85225F
Performance Modeling System
Installation and User’s Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2005 No part of this manual may be reproduced
Acknowledgments
UNIX® is a registered trademark of the Open Group.
Windows NT mark of Microsoft Corporation.
Windows istered trademarks of Microsoft Corpora­tion.
®
is a U.S. registered trade-
®
and MS Windows are U.S. reg-
Manual Part Number
85225-90023
Edition
First edition, April 2005 Printed in USA Agilent Technologies, Inc.
1400 Fountaingrove Parkway Santa Rosa, CA 95403 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, with out notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damage s in connection with the furnishing, use, or perfor­mance of this document or of any information contained he re in. Should Agilent and the user ha v e a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
agency regulation or co ntract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencie s of t h e U .S . Gov ­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987) . U.S . Gov er n­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Technology Licenses
The hardware and/or sof tware describe d in this document are furnished under a license and may be used or copied only in accordance wi t h the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent

Safety and Regulatory Information

Warnings, Cautions, and Not es
This installation and user’s guide utilizes the following safety notations. Familiarize yourself with each notation and its meaning before operating the Agilent 85225F performance modeling system.
WARNING
CAUTION
NOTE
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proce ed beyond a WARNING notic e unt il the indicated conditions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A NOTE calls the user’s attention to an important point or special information within the text. It provides additional information or instructions.
Installation and User’s Guide 3
Safety Symbols and Instrument Markings
Symbols and markings in documentation and on instruments alert you to potential risks, provide information about conditions, and comply with international regulations. Table A defines the safety symbols and Table B on page 5 defines the instrument markings you may find in the documentation or on an instrument.
Table A Safety Symbols
Symbols Definition
Warning: risk of electric shock.
Warning: hot surface.
Caution: refer to instrument documen tat i on.
Laser radiation symbol: marked on pr oducts that have a laser output.
Alternating current. Both direct and alternating current.
Three-phase alternating current. Earth (ground) terminal.
Protective earth (ground) terminal.
Frame or chassis terminal.
Terminal is at earth potential. Used for measurement and control circuits designed to be operated with one terminal at earth potential.
Terminal for neutral conductor on permanently installed equipment.
Terminal for line conductor on permanently installed equipment.
4 Installation and User’s Guide
Table A Safety Symbols (continued)
Symbols Definition
Standby (supply). Units with this s ymb ol are not completely disconnected from AC mains w hen this s witch is in th e stand by position. To completely disconnect the unit from AC mains, either disconnect the power cord, or have a qu alified/licensed electrician install an external switch.
ON (supply). A switch with this symbol closes the instrument’s power supply circuit, connecting it to the mains supply .
OFF (supply). A switc h with th is symbo l open s the instr umen t’s power supply circuit, disconnecting it from the mains supply.
Table B Instrument Markings
Marking Definition
The instruction documentation symbol appears when it is necessary for the user to refer to the instruction in the documentation.
The CE mark is a registered trademark of the European Community.
This product compli es with the WEEE Directiv e (2002/96/EC) marking requirements. The affixed la bel indicates th at you must not discard this electrical/el ectronic product in domestic household waste. To return unwanted products, contact your local Agilent Technologies office, or see www.agilent.com for more information.
The CSA mark is a registered trademark of the CSA-International.
N10149
ISM1-A This text indicates t hat the instrume nt is an In dustrial S cienti fic
ICES/NMB-001 This text indicates product compliance wi th the Canadian
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australian EMC Framework regulations under the terms of the Radio Communications Act of 1992.
and Medical Group 1 Class A product (CISPER 11, Clause 4).
Interference-Causing Equipment Standard (ICES-001).
Operator Safety Requirements
The following general safety precautions must be observed during all phases of operation of this system. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates
Installation and User’s Guide 5
safety standards of design, manufacture, and intended use of the product. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements.
For additional safety precautions, including precautions for making device measurements in a floating ground configuration, see “To ensure your
safety while using the system" on page 76.
WARNING
WARNING
WARNING
This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the mains supply cord). The mains plug shall be inserted only in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used only in a normal condition (in which all means for protection are intact) only.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. Do not operate the instrument in the presence of flammable gases or flames.
WARNING
DO NOT REMOVE THE INSTRUMENT COVER. Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be m a de only by qualified service personnel. Instruments that appear damaged or defective should be made inoperative and se cured against unintended operation until they can be repaired by qualified service personnel.
WARNING
WARNING
WARNING
6 Installation and User’s Guide
Installing additional instrument s may dest abi l ize the rack cabinet.
Installing additional instruments into the ca bine t electrical system could produce excessive leakage current. If the protective earth conductor is interrupted or faulted, the user risks serious inj u ry or death.
Prior to adding any additional instruments, review all wiring and cooling capabilities to verify adeq uate design margins for nor mal and under single fault conditions.
Mains power
CAUTION
Ground the system
WARNING
WARNING
Before applying power
CAUTION
The mains cable shall be permanently connected to the premise circuit breaker or connected using an agency approved twist-lock connector.
To minimize shock hazard, the rack cabinet must be connected to an electrical protective earth ground. The power distribution unit (PDU) must be connected to the AC power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet.
Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety precautions are taken. Before applying power, note the product’s external markings described in
Table A, “Safety Symbols,” on page4 and Table B, “Instrument
Markings,” on page5.
CAUTION
CAUTION
CAUTION
Installation and User’s Guide 7
It is recommended that the premise wiring contain an adequate circuit breaker for system protection.
To remove power from the cabinet, remove the mains supply from the premise electrical supply.
Before switching on this system, make sure that the supply voltage is in the specified range.
CAUTION
Fuses and breakers
The front panel LINE switch disconnects the mains circuit from the mains supply. Ho wev e r, the mains supply to the po wer d is tr ib ut io n u n it remains energized.
WARNING
Before cleaning the system
WARNING
Overcurrent protection
CAUTION
For continued protection against fire hazard, use only fuses with the required rated current, voltage, and specified type (normal blow , time delay). Do not use repaired fuses or short-circuited fuse holders. Replace only with an identical fuse.
There are two resettable thermal bre a ke rs located on the power strips. These are in the “hot” and “neutral” lines.
To prevent electrical shock, disconnect the system from mains before cleaning. Use a dry (or slightly water-dampened) cloth to clean external case parts. Do not atte mpt to clean internally.
If the power outlet strip breaker trips once, reset the breaker. If the breaker trips twice, call a qualified/licensed electrician to service the test system.
Statement of Compliance and Declaration of Conformity
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
The Manufacturer’s Declaration of Conformity is available upon request.
Statement of CAN/CSA Compliance
This product has been designed and tested in accordance with CAN/CSA- C22.2 No. 61010- 1 IEC.
8 Installation and User’s Guide
Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung - 3.GSGV Deutschland).
Acoustic Noise Emission/Geraeuschemission
LpA <70 dB LpA <70 dB Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t.19
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES- 001. Cet appareil ISM est conformé à la norme NMB du Canada.
IEC/EN 61000-4-2 Electrost atic Discharge Immunity Test
This system passes using criterion C where operator intervention may be necessary to restart the measurement software operations.
IEC/EN 61326 Electrostat ic Di sc harg e and Surge Im muni t y Test
This system complies with the Electrostatic Discharge and Surge Immunity requirements in the IEC/EN 61326 standard using Performance Criterion C.
For Technical Assistance
To receive technical assistance, visit the online assistance web site, or call the telephone number listed in Table 19 on page 107 appropriate to the location of modeling system.
Installation and User’s Guide 9

In This Guide...

1 Introducing the Agilent 85225F Performance Modeling System
2 Installing the System
3 Verifying System Functionality
This guide provides instruction on installing, verifying, and servicing the system, as well as an introductory system overview and reference material. This information is presented for use by the customer or an Agilent Technologies field engineer.
This chapter provides a description of the system, its components, integration, and characteristics.
Here you will find instruction on preparing the installation site, receiving and inspecting the system (including a receiving checklist), installing the worksurface, ensuring operator safety, connecting the bias networks, and powering- on the system.
Turn here for instruction on choosing a level of system verification and performing a post-installation functional verification test using a system controller running IC- CAP software.
4 Servicing the System
This chapter includes instruction on troubleshooting the system, removing and replacing system components, ordering replacement parts, and acquiring additional assistance in solving measurement problems.
A Enhancing Measurement Accuracy
See this appendix for instruction on cleaning the system connections, performing a system measurement calibration, and suggested intervals for periodic component calibration.
B DC Subsystem Functional Verificat ion Tests
Turn here to find Agilent 4156C precision semiconductor parameter analyzer and Agilent E5260A/70B high speed/precision parameteric measurement mainframe functional verification tests that do not require the IC- CAP software.
C RF Subsystem Functional Verification Tests
This appendix includes an Agilent E8364B PNA Series vector network analyzer functional verification test that does not require the IC- CAP software.
10 Installation and User’s Guide
D CV Subsystem Functional Verification Tests
This appendix includes an Agilent 4284A precision LCR meter functional verification test that does not require the IC- CAP software.
E Noise Subsystem Functional Verifi cation Tests
This appendix includes an Agilent 35670A dynamic signal analyzer functional verification test that does not require the IC- CAP software.
F Understanding the Bias Networks
Here you will find features, characteristics, a schematic diagram, and operational information on the bias networks.
G Network Analyzer Performance Specification Summary
See this appendix for a summary of the network analyzer’s performance specifications.
For Additional
Information on...
Hardware
Additional information regarding instruments and accessories within the system is provided in the individual instrument or accessory’s documentation.
Software
IC- CAP software operating instructions and tutorials are provided in the Agilent 85190D IC- CAP user’s guide.
Installation and User’s Guide 11
Typeface
Conventions
This guide uses the following typeface conventions to describe various aspects of a particular hardware or software user interface.
Hardware
Interface Examples in Body Text Examples in Procedural Text
and Tables
Front panel hardke ys
Front panel display softkeys
Front or rear panel connectors, instrument markings
Data field entries
Keyboard keys
Press Preset Press Cal
Press [MORE] Press [Return]
RF/DC OUT connector STIMULUS key group
Enter Calset Enter 18
Press Ctrl+8 Press Enter
Press Preset Press Cal
Press [MORE] Press [Return]
RF/DC OUT connector STIMULUS key group
Enter Calset Enter 18
Press Ctrl+8 Press Enter
Software
Interface Examples in Body Te xt Examples in Procedural Text
and Tables
Screen buttons and selections
Menu selections
Click Enter Select Continuous
Choose Format > Small Choose Cal > Full
Click Enter Select Continuous
Choose Format > Small Choose Cal > Full
Command and menu names
Icon and window titles
The Save commands are in the File menu.
The Model icons are in the IC- CAP/Main window.
Program messages Data field entries
Is the device conn ected?
Enter Calset Enter 18
12 Installation and User’s Guide
The Save commands are in the File menu.
The Model icons are in the IC-CAP/Main window.
Is the device connected?
Enter Calset Enter 18

Contents

1 Introducing the Agilent 85225F Performance Modeling System
Performance Modeling System Configuration Overview 18 RF and DC Measurement System Configuration 19
Figure 1. System Block Diagram 19 The RF Subsystem 20 The DC Subsystem 20 The Bias Networks 21 Component Integration 22 Figure 2. System Components 23 Table 3.Front Panel System Connections, with Agilent 4156C 24 Figure 3.Front Panel Connections with Agilent 4156C 25 Table 4.Front Panel System Connections, with Agilent E5260A or E5270B 26 Figure 4.Front Panel Wiring Diagram with Agilent E5260A or E5270B 27 Table 5.Rear Panel System Connections with Agilent 4156C 28 Figure 5.Rear Panel Wiring Diagram with Agilent 4156C 29 Table 6. Rear Panel System Connections with Agilent E5260A or E5270B 30 Figure 6.Rear Panel Wiring with Agilent E5260A or E5270B 31 Figure 7.DC/RF Cabling Diagram - DC and RF Configuration 32
CV, RF, and DC Measurement System Configuration 33
Figure 8. System Block Diagram 33 The CV Subsystem 34 Component Integration 34 Figure 9. System Components 35 Table 7.Front Panel System Connections 36 Figure 10. Front Panel Wiring Diagram 37 Table 8.Rear Panel System Connections 38 Figure 11.Rear Panel Wiring Diagram 39 Figure 12. DC/RF Cabling Diagram - DC and RF Configuration 40 Figure 13.DC/RF Cabling Diagram - Parametric Configuration 41 The Low Leakage Switch Mainframe 42 Table 9.Rear Panel Connections, including Low Leakage Switch Mainframe 42 Figure 14. Rear Panel Wiring Diagram including Low Leakage Switch Mainframe 43 Figure 15.DC/RF Cabling Diagram - Parametric Configuration with Low Leakage Switch
Mainframe 44
Installation and User’s Guide 13
1/f Noise, CV, RF, and DC Measurement System Configuration 45
Figure 16.System Block Diagram 46 Figure 17. 1/f Noise Measurement Block Diagram 47 Component Integration 48 Figure 18. System Components 49 Table 10.Front Panel System Connections 50 Figure 19. Front Panel Wiring Diagram 51 Tab le 11. Rear Panel System Connections 52 Figure 20.Rear Panel Wiring Diagram 53 Figure 21. DC/RF Cabling Diagram - DC and RF Configuration 54 Figure 22.DC/RF Cabling Diagram - Parametric Configuration 55
Instrument Control Interface 56
Tab le 12. GPIB Addresses 56
The LAN/GPIB Gateway 57
Figure 23.Rear Panel Wiring Diagram for LAN/GPIB Gateway 58 The System Controller 59 Table 13. Personal Computer Requirements 59 Table 14. UNIX Workstation Requirements 59 The Rack Cabinet 60
Performance Characteristics and Specifications 61
Table 15.Supplemental System Characteristics 61 Interference Standards 61 Performance Modeling System Performance Specifications 61 RF Subsystem Performance Specifications 62 DC Subsystem Specifications 62 Bias Network Characteristics 62
2 Installing the System
To prepare the installation site 64
Ta b le 16 . Environmen t al Requ ir em e nts 64 Table 17. Electrical Requirements 64
To receive the system 65
To unpack the shipment crate containing the rack cabinet 66 To verify the shipment 68
Tab le 18. Replaceable Parts 69 To install the work surface 74 To ensure your safety while using the system 76
Precautions for Performing Floating-Ground Measurements 78
To perform floating-ground measurements 78
14 Installation and User’s Guide
Precautions for Avoiding Electrostatic Discharge 79 To connect the bias networks 80
Agilent 4156C Systems 80
Agilent 4156C Systems with Agilent 41501B Expander Box 82
Agilent E5260A/70B Systems 84 To switch on power to the system 87 To configure the LAN/GPIB gateway for functional verification 89
3 Verifying System Functionality
To choose a verification process 92 Understanding the System Functional Verification Test 94
Required Tools 94 Performing the System Functional Verification Test 95
If you encounter a problem 101
4 Servicing the System
To troubleshoot the system 104 To remove or replace a system component 105
To order replacement parts 106 To receive additional assistance 107
Ta b le 19. Contacting Agil ent Technologies 107 To package the system for transport 108
A Enhancing Measurement Accuracy
To enhance measurement accuracy 112 Understanding System Measurement Cal ibrati on 114
Required Tools 114 Performing a Coaxial System Measurement Calibration 115
If you encounter a problem 116 Periodic System Component Calibration 117
B DC Subsystem Functional Verification Test
Understanding the DC Subsystem Functional Verification Test 120
Required Tools 120 Performing the DC Subsystem Functional Verification Test 121
If you encounter a problem 121
Installation and User’s Guide 15
If you encounter a problem 122
C RF Subsystem Functional Verification Test
Understanding the RF Subsystem Functional Verification Test 124
Required Tools 124 Performing the RF Subsystem Functional Verification Test 125
If you encounter a problem 127
D CV Subsystem Functional Verification Test
Understanding the CV Subsystem Functional Verification Test 130
Required Tools 130 Performing the CV Subsystem Functional Verification Test 131
If you encounter a problem 132
E 1/f Noise Subsystem Functional Verific ation Test
Understanding the 1/f Noise Subsystem Functional Verification Test 134
Required Tools 134 Performing the 1/f Noise Subsystem Functional Verification Test 135
If you encounter a problem 136
F Understanding the Bias Networks
Features 138 Characteristics 139
Table 20. 11612V Option K11/K21 Bias Network Characteristics 139 Operation 140
Figure 24. Bias Network Schematic 140
G Network Analyzer Performance Specification Summary
Network Analyzer System Performance 142
Maximum Output Power 142
Dynamic Range 142
Measurement Port Charact eristics 143
Measurement Uncertainty 143
Index
16 Installation and User’s Guide
Agilent 85225F Perfor mance Model ing System Installation and User’s Guide
1 Introducing the Agilent 85225F Performance Modeling System
Performance Modeling System Configuration Overview 18 RF and DC Measurement System Configuration 19 CV, RF, and DC Measurement System Configuration 33 1/f Noise, CV, RF, and DC Measurement System Configuration 45 The System Controller 59 Performance Characteristics and Specifications 61
Related Topics “Installing the System" on page 63
“Network Analyzer Performance Specification Summary"on page 141 “Understanding the Bias Networks"on page137
Use this chapter to familiarize yourself with the measurement configurations of the performance modeling system. This chapter introduces the system by describing its operational theory, integration, and performance.
Agilent Technologies
17
1 Introducing the Agilent 85225F Performance Modeling System

Performance Modeling System Configuration Overview

The standard Agilent 85225F performance modeling system measures the DC and RF performance of active and passive devices. You may configure the Agilent 85225F performance modeling system to measure CV and 1/f noise with the addition of optional instrumentation and IC-CAP 1/f noise measurement modules.
For RF and DC performance measurement system configurations, see “RF
and DC Measurement System Configuration" on page 19.
For CV, RF, and DC performance measurement system configurations, see
“CV, RF, and DC Measurement System Configuration" on page 29.
For 1/f noise, CV, RF, and DC performance measurement system configurations, see “1/f Noise, CV, RF, and DC Measurement System
Configuration" on page 45.
18 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1

RF and DC Measurement System Configuration

In conjunction with a compatible controller running 85190-Series IC- CAP software, the Agilent 85225F performance modeling system measures the DC and RF performance of active and passive devices. The IC- CAP software then extracts the device parameters and displays the results.
The Agilent 85225F performance modeling system is the integration of rack- mounted RF and DC subsystems, bias networks, and a system controller

Figure 1 System Block Diagram

*
, as shown in Figure 1†.
* The system controller is not included and must be provided. † This block diagram shows a system with an Agile nt 4156C as the DC subsystem. Other instrumentation may
be used. See “The DC Subsystem"on page 20.
Installation and User’s Guide 19
1 Introducing the Agilent 85225F Performance Modeling System

The RF Subsystem

S- parameter device characterization is provided by the RF subsystem.
The RF subsystem contains the Agilent E8364B PNA Series vector network analyzer.
Its integrated synthesizer supplies a swept or CW RF source signal from 10 MHz
The integrated test set separates the RF source signal into reference and test signals, and provides RF connection via cables and adapters to the external bias networks.
*
to 50 GHz.

The DC Subsystem

Precision DC characterization and bias for the S-parameter measurements are provided by one of the following three DC subsystems.
The DC subsystem may contain one of the following three instruments.
Agilent 4156C Precision Semi conducto r Par a meter Analyzer
The Agilent 4156C precision semiconductor parameter analyzer provides DC force (supply) and sense (measure) capability from its HRSMUs (high resolution source/monitor units).
Optionally, the Agilent 4156C may be configured with a 41501B SMU PGU expander is connected to and controlled by the 4156C via the expander box interface. The 41501B provides a GNDU (active ground unit) and, depending on option configuration, an HPSMU (high- power source/monitor unit), two MPSMUs (medium- power source monitor units), and/or two PGUs (pulse generator units).
The DC signals are routed through feedthrough panels via triaxial cables to the bias networks.
Agilent E5260A 8-Slot High Speed Parametric Measurement Mainframe
The Agilent E5260A provides DC force (supply) and sense (measure) capability from its plug- in source/monitor units.
The Agilent E5290A plug- in high speed high power source/monitor unit provides up to 200 volts of potential and 1 amp of current to the device under test.
The Agilent E5291A plug- in high speed medium power source/monitor unit provides up to 100 volts of potential and 200 milliamps of current to the device under test.
* Due to the minimum operating frequency of the bias networks, the performance modeling system low end
frequency range is 45 MHz.
20 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Agilent E5270B 8-Slot Precision Parametric Measurement Mainframe
The Agilent E5270B provides DC force (supply) and sense (measure) capability from its plug- in source/monitor units.
The Agilent E5280A plug- in high power source/monitor unit provides up to 200 volts of potential and 1 amp of current to the device under test.
The Agilent E5281A plug- in medium power source/monitor unit provides up to 100 volts of potential and 200 milliamps of current to the device under test.
CAUTION

The Bias Networks

Exposing the bias networks to currents greater than 500 milliamps or voltages greater than 40 volts will result in severe damage. Do not exceed these values while using the bias networks. Remove the bias networks from the circuit if greater voltages or currents are required.
The Agilent 11612V Option K11 and K21 bias networks combine the DC and RF signals and apply them simultaneously to the device under test (DUT). The bias networks are configured with 2.4 mm DC/RF output connectors for connection to a DUT, a test fixture, or probe station, as shown in Figure 21 on page 54.
Installation and User’s Guide 21
1 Introducing the Agilent 85225F Performance Modeling System

Component Integration

System component integration is performed at the Agilent Technologies factory. The individual components are placed into the rack, and the required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional performance.
The Agilent 85225F performance modeling system includes the following components, as shown in Figure 2 on page 23:
Agilent E8364B PNA Series vector network analyzer
Agilent 4156C precision semiconductor parameter analyzer (or
optionally Agilent E5260A or E5270B)
Agilent 11612V Option K11 bias network (port 1)
Agilent 11612V Option K21 bias network (port 2)
Agilent 85133F flexible test port cable set
Agilent E3661B 1.6 meter rack cabinet
filler panels, feedthrough panels, work surface, cables, and adapters
For systems with Agilent 4156C, front panel connections are listed in
Table 3 on page 24 and illustrated in Figure 3 on page 25.
For systems with Agilent 4156C, rear panel connections are listed in
Table 5 on page 28 and illustrated in Figure 5 on page 29.
For systems with Agilent E5260A or E5270B, front panel connections are listed in Table 4 on page 26 and illustrated in Figure 4 on page 27.
For systems with Agilent E5260A or E5270B, rear panel connections are listed in Table 6 on page 30 and illustrated in Figure 6 on page 31.
22 Installation and User’s Guide
Figure 2 System Components
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 23
1 Introducing the Agilent 85225F Performance Modeling System
Table 3 Front Panel System Connections, with Agilent 4156C
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16494A
Option 002
6 85133F Flexible test
7 85133F Flexible test
Description Connector
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
port cable
port cable
From
Type
2.4 mm E8364B PORT 1 RF IN 11612V K11
2.4 mm E8364B PORT 2 RF IN 11612V K21
Instrument
HRSMU1
HRSMU1
HRSMU2
HRSMU2
HRSMU3
Connector Labeled
SENSE DC SENSE 11612V K11
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K21
FORCE DC FORCE 11612V K21
FORCE GNDU 11612V K21
T o Connector Labeled
On Instrument
24 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 3 Front Panel Connections with Agilent 4156C
Installation and User’s Guide 25
1 Introducing the Agilent 85225F Performance Modeling System
Table 4 Front Panel System Connections, with Agilent E5260A or E5270B
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16493L
Option 002
6 85133F Flexibl e test
7 85133F Flexibl e test
Description Connector
Triaxial cable Triax BNC E5260A/70B
Triaxial cable Triax BNC E5260A/70B
Triaxial cable Triax BNC E5260A/70B
Triaxial cable Triax BNC E5260A/70B
Triaxial GNDU cable
port cable
port cable
From
Type
Triax BNC E5260A/70B
2.4 mm E8364B PORT 1 RF IN 11612V K11
2.4 mm E8364B PORT 2 RF IN 11612V K21
Instrument
HPSMU1
HPSMU1
MPSMU3
MPSMU3
GNDU
Connector Labeled
SENSE DC SENSE 11612V K21
FORCE DC FORCE 11612V K21
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
GNDU GNDU 11612V K21
To Connector Labeled
On Instrument
26 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 4 Front Panel Wiring Diagram with Agilent E5260A or E5270B
Installation and User’s Guide 27
1 Introducing the Agilent 85225F Performance Modeling System
Table 5 Rear Panel System Connections with Agilent 4156C
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16494A
Option 002 6 10833D GPIB cabl e GPIB 4156C GPIB GPIB E8364B 7 10833C GPIB cable GPIB E8364B GPIB GPIB Controller
Description Connector
Type
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
NOTE
If the system does not include an Agilent 41501B SMU/PGU expander, use the Agile nt 4156C HRSMU3 FORCE as the GND ( ground unit).
From Instrument
HRSMU2
HRSMU2
HRSMU1
HRSMU1
HRSMU3
Connector Labeled
FORCE DC FORCE 11612V K21
SENSE DC SENSE 11612V K21
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
FORCE GNDU 11612V K21
To Con nector Labeled
On Instrument
28 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 5 Rear Panel Wiring Diagram with Agilent 4156C
Installation and User’s Guide 29
1 Introducing the Agilent 85225F Performance Modeling System
CV, RF, and DC Mea surement System Configuration
Table 6 Rear Panel System Connections with Agilent E5260A or E5270B
Component Information Connection Information Designator Model
Number
1 10833A GPIB cable GPIB 4156C GPIB GPIB E8364B 2 10833C GPIB cable GPIB E8364B GPIB GPIB Controller
Description Connector
Type
From Instrument
Connector Labeled
To Connector Labeled
On Instrument
30 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 6 Rear Panel Wiring with Agilent E5260A or E5270B
Installation and User’s Guide 31
1 Introducing the Agilent 85225F Performance Modeling System
Figure 7 DC/RF Cabling Diagram - DC and RF Configuration
Systems with Agilent 4156C
Systems with Agilent E5260A or E5270B
32 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1

CV, RF, and DC Measurement System Configuration

With the addition of a precision LCR meter, the Agilent 85225F performance modeling system measures the DC, RF, and CV performance of active and passive devices. The IC- CAP software then extracts the device parameters and displays the results.
The Agilent 85225F performance modeling system for CV, RF, and DC measurement is the integration of rack-mounted RF, DC, and CV subsystems, bias networks, and a system controller, as shown in Figure 8.

Figure 8 System Block Diagram

Installation and User’s Guide 33
1 Introducing the Agilent 85225F Performance Modeling System

The CV Subsystem

The Agilent 4284A precision LCR meter provides a wide 20 Hz to 1 MHz test frequency range and superior test- signal performance, allowing CV testing to the most commonly- used test standards, such as IEC/MIL, and under conditions that simulate the intended application.
Optionally, the system can be configured with the Agilent E5250A low leakage switch mainframe. The Agilent E5250A is used for precise parametric test. It improves measurement efficiency by eliminating the need to manually change the probe positions on a manual probe station. The E5250A is used to route signals from the DC and CV subsystems to the probe card cable, and on to the probe card and probe station.

Component Integration

System component integration is performed at the Agilent Technologies factory. The individual components are placed into the rack, and the required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional performance.
The Agilent 85225F performance modeling system includes the following components, as shown in Figure 18 on page 49:
Agilent E8364B PNA Series vector network analyzer
Agilent 4156C precision semiconductor parameter analyzer (or
optionally Agilent E5260A or E5270B)
Agilent 11612V Option K11 bias network (port 1)
Agilent 11612V Option K21 bias network (port 2)
Agilent 4284A precision LCR meter
Agilent 85133F flexible test port cable set
Agilent E3661B 1.6 meter rack cabinet
filler panels, feedthrough panels, work surface, cables, and adapters
System front panel connections are listed in Table 10 on page 50 and illustrated in Figure 19 on page 51.
System rear panel connections are listed in Table 11 on page 52 and illustrated in Figure 20 on page 53.
34 Installation and User’s Guide
Figure 9 System Components
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 35
1 Introducing the Agilent 85225F Performance Modeling System
Table 7 Front Panel System Connections
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16494A
Option 002
6 85133F Flexibl e test
7 85133F Flexibl e test
8 16048D LCR meter
Description Connector
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial GNDU cable
port cable
port cable
test cable
From
Type
Triax BNC 4156C
2.4 mm E8364B PORT 1 RF IN 11612V K11
2.4 mm E8364B PORT 2 RF IN 11612V K21
BNC 4284A UNKNOWN Test fixture
Instrument
HRSMU1
HRSMU1
HRSMU2
HRSMU2
HRSMU3
Connector Labeled
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
FORCE DC FORCE 11612V K21
SENSE DC SENSE 11612V K21
FORCE GNDU 11612V K21
To Connector Labeled
On Instrument
or probe station
36 Installation and User’s Guide
Figure 10 Front Panel Wiring Diagram
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 37
1 Introducing the Agilent 85225F Performance Modeling System
Table 8 Rear Panel System Connections
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16494A
Option 002 6 10833D GPIB cable GPIB 4156C GPIB GPIB 4284A 7 10833D GPIB cable GPIB 4284A GPIB GPIB E8364B 8 10833C GPIB cable GPIB E8364B GPIB GPIB Controller
Description Connector
Type
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
From Instrument
HRSMU1
HRSMU1
HRSMU2
HRSMU2
HRSMU3
Connector Labeled
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
FORCE DC FORCE 11612V K21
SENSE DC SENSE 11612V K21
FORCE GNDU 11612V K21
To Connector Labeled
On Instrument
38 Installation and User’s Guide
Figure 11 Rear Panel Wiring Diagram
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 39
1 Introducing the Agilent 85225F Performance Modeling System
Figure 12 DC/RF Cabling Diagram - DC and RF Configuration
40 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 13 DC/RF Cabling Diagram - Parametric Configuration
Installation and User’s Guide 41
1 Introducing the Agilent 85225F Performance Modeling System

The Low Leakage Switch Mainframe

The Agilent E5250A is used for precise parametric test. It improves measurement efficiency by eliminating the need to manually change the probe positions on a manual probe station. The E5250A is used to route signals from the 4156C and the 4284A to the probe card cable, and on to probe card and probe station.
Table 9 Rear Panel Connections, including Low Leakage Switch Mainframe
Component Information Connection Information Designator Model
Number
1 16494A
Option 002 2 16494A
Option 002 3 16494A
Option 002 4 16494A
Option 002 5 16048D LCR meter
6 10833D GPIB cable GPIB 4156C GPIB GPIB 4284A 7 10833D GPIB cable GPIB 4284A GPIB GPIB E8364B 8 10833D GPIB cable GPIB E8364B GPIB GPIB E5250A 9 10833C GPIB cable GPIB E5250A GPIB GPIB Controller T1 1250-2405 BNC tee BNC 4284A HIpot/HIcur CV1 E5250A T2 1250-2405 BNC tee BNC 4284A LOpot/LOcur CV2 E5250A
Description Connector
Type
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
BNC 4284A UNKNOWN T1 & T2 (CV1
test cable
From Instrument
HRSMU1
HRSMU1
HRSMU2
HRSMU2
Connector Labeled
SENSE SMU
FORCE SMU
FORCE SMU
SENSE SMU
To Connector Labeled
INPUT 2
INPUT 1
INPUT 3
INPUT 4
& CV2)
On Instrument
E5250A
E5250A
E5250A
E5250A
E5250A
42 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 14 Rear Panel Wiring Diagram including Low Leakage Switch Mainframe
Installation and User’s Guide 43
1 Introducing the Agilent 85225F Performance Modeling System
Figure 15 DC/RF Cabling Diagram - Parametric Configuration with Low Leakage Switch Mainframe
44 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1

1/f Noise, CV, RF, and DC Measurement System Configuration

With the addition of a dynamic signal analyzer and a precision LCR meter, the Agilent 85225F performance modeling system measures the DC, RF, CV, and 1/f noise performance of active and passive devices. The IC- CAP software then extracts the device parameters and displays the results.
The Agilent 85225F performance modeling system is the integration of rack- mounted RF and DC subsystems, a precision LCR meter, a dynamic signal analyzer, bias networks, and a system controller, as shown in
Figure 16.
Installation and User’s Guide 45
1 Introducing the Agilent 85225F Performance Modeling System

Figure 16 System Block Diagram

The 1/f Noise Subsystem
The Agilent 35670A dynamic signal analyzer (in conjunction with a customer- furnished Stanford Model SR570 low noise amplifier) measures the flicker noise (1/f noise) of active devices. Controlled by IC- CAP device modeling software, the dynamic signal analyzer generates reliable 1/f noise measurement data, which are analyzed and extracted in IC-CAP. Figure 17 shows the system configuration for 1/f noise measurements.
46 Installation and User’s Guide

Figure 17 1/f Noise Measurement Block Diagram

Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 47
1 Introducing the Agilent 85225F Performance Modeling System

Component Integration

System component integration is performed at the Agilent Technologies factory. The individual components are placed into the rack, and the required cabling is connected between the instruments.
After factory integration, the system is tested to verify functional performance.
The Agilent 85225F performance modeling system includes the following components, as shown in Figure 18 on page 49:
Agilent E8364B PNA Series vector network analyzer
Agilent 4156C precision semiconductor parameter analyzer with
optional Agilent 41501B SMU/PGU expander (or optionally Agilent E5260A or E5270B)
Agilent 11612V Option K11 bias network (port 1)
Agilent 11612V Option K21 bias network (port 2)
Agilent 4284A precision LCR meter
Agilent 35670A dynamic signal analyzer
Stanford Research SR 570 low noise current amplifier
Agilent 85133F flexible test port cable set
Agilent E3661B 1.6 meter rack cabinet
filler panels, feedthrough panels, work surface, cables, and adapters
System front panel connections are listed in Table 10 on page 50 and illustrated in Figure 19 on page 51.
System rear panel connections are listed in Table 11 on page 52 and illustrated in Figure 20 on page 53.
*
* Customer supplied, not included with system.
48 Installation and User’s Guide
Figure 18 System Components
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 49
1 Introducing the Agilent 85225F Performance Modeling System
Table 10 Front Panel System Connections
Component Information Connection Information Designator Model
Number
1 16494A
Option 002 2 16494A
Option 002 3 16494A
Option 002 4 16494A
Option 002 5 16493L
Option 002 6 85133F Flexibl e test
7 85133F Flexibl e test
8 16048D LCR meter
Description Connector
Triaxial cable Triax BNC 41501B
Triaxial cable Triax BNC 41501B
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial GNDU cable
port cable
port cable
test cable
From
Type
Triax BNC 41 501B GNDU GNDU 11612V K21
2.4 mm E8364B PORT 1 RF IN 11612V K11
2.4 mm E8364B PORT 2 RF IN 11612V K21
BNC 4284A UNKNOWN Test fixture
Instrument
HPSMU
HPSMU
HRSMU1
HRSMU1
Connector Labeled
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
FORCE DC FORCE 11612V K21
SENSE DC SENSE 11612V K21
To Connector Labeled
On Instrument
or probe station
9 8120-1839 Coaxial cable BNC 35670A CH1 Test fixture
or probe station
50 Installation and User’s Guide
Figure 19 Front Panel Wiring Diagram
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 51
1 Introducing the Agilent 85225F Performance Modeling System
Table 11 Rear Panel System Connections
Component Information Connection Information Designator Model
Number
1 16494A
Option 002
2 16494A
Option 002
3 16494A
Option 002
4 16494A
Option 002
5 16493L
Option 002 6 10833A GPIB cable GPIB 4156C GPIB GPIB 4284A 7 10833A GPIB cable GPIB 4284A GPIB GPIB E8364B 8 10833A GPIB cable GPIB E8364B GPIB GPIB 35670A 9 10833C GPIB cable GPIB 4156C GPIB GPIB Controlle r
Description Connector
Type
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 4156C
Triaxial cable Triax BNC 41501B
Triaxial cable Triax BNC 41501B
GNDU cable Triax BNC 41501B GNDU GNDU 11612V K21
From Instrument
HRSMU1
HRSMU1
HPSMU
HPSMU
Connector Labeled
FORCE DC FORCE 11612V K11
SENSE DC SENSE 11612V K11
FORCE DC FORCE 11612V K21
SENSE DC SENSE 11612V K21
To Con nector Labeled
On Instrument
52 Installation and User’s Guide
Figure 20 Rear Panel Wiring Diagram
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 53
1 Introducing the Agilent 85225F Performance Modeling System
Figure 21 DC/RF Cabling Diagram - DC and RF Configuration
54 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1
Figure 22 DC/RF Cabling Diagram - Parametric Configuration
Installation and User’s Guide 55
1 Introducing the Agilent 85225F Performance Modeling System

Instrument Control Interface

Instrument control interface is provided by a General Purpose Interface Bus (GPIB) or LAN/GPIB gateway. GPIB addresses for programmable system components are listed in Table 12.
Table 12 GPIB Addresses
Component GPIB Address
Agilent 34401A digital multime te r 9 Agilent 35670A dynamic signal analyzer 10 Agilent 4156C precision semiconductor parameter analyzer 19 Agilent 4284 A precision LCR meter 24 Agilent E5810A LAN/GPIB gateway 21 Agilent E8364B PNA Series vector network analyzer 16
*
Agilent E5250A low leakage switch mainframe 22 Agilent E5260A 8-slot high speed parametric measurement mainframe 19 Agilent E5270B 8-slot precision parametric measurement mainframe 19
* The 4156C default GPIB address 17 is sometimes used by other devices with a GPIB address at 16 (for example, an external display (set
to 17) to display the results generated by an instrument at address 16). Change the 4156C GPIB address to 19 using the procedure described in step13 of “To switch on power to the system"on page 87 to ensure that IC-CAP can recognize the 4156C.
56 Installation and User’s Guide

The LAN/GPIB Gateway

The Agilent E5810A LAN/GPIB gateway provides access to the system’s GPIB instrumentation over an existing local area network. It allows the use of SICL- or VISA- based applications designed for GPIB over the LAN without modifying the application beyond a simple address change.
The gateway is a combination of hardware and SICL/VISA software. It uses client/server technology to extend the standard remotely over the LAN, allowing remote control from an alternative, more convenient, or safer location.
Introducing the Agilent 85225F Performance Modeling System 1
Installation and User’s Guide 57
1 Introducing the Agilent 85225F Performance Modeling System

Figure 23 Rear Panel Wiring Diagram for LAN/GPIB Gateway

58 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1

The System Controller

A customer- furnished UNIX workstation or personal computer running Agilent IC- CAP software controls the hardware via GPIB while making device measurements, then stores, simulates, and optimizes device parameters, using predefined or user- defined device models. Table 13 on page 59 lists the personal computer requirements. Table 14 on page 59 lists the UNIX workstation requirements.
Table 13 Personal Computer Requirements
Parameter Requirement
Operating sys t em Microsoft Windows NT® 4.0- SP6a or Windows 2 000 Professional-SP3. CPU Intel Pentium® class 200 MHz CPU or higher Display Super VGA 800×600, 15 inch moni tor (1024×728 recommended) Hard disk space 370 MB. It is recommended that you install IC-C AP software on your local drive.
Recommended file systems are FAT32 and NTFS. Novell file servers are not supported. VFAT/FAT systems are not recommended for full installations.
*
RAM 128 megabytes (additional RAM will improve software perfor mance) Virtual memory 300 megabytes
* Windows 95, 98, and ME are not supported. † For NT 4.0 only: to avoid potential memory problems, ensure your virtu al memo ry space is always greater than your RAM space.
(Increased virtual memory may be required)
Table 14 UNIX Workstation Requirements
Parameter Requirement
HP UNIX Workstation SunOS Workstation
Operating system HP-UX 11.i with the following patches:
PHSS_24627 HP aC++, AA Runtime Libraries (aCC A.03.33), PHSS_25718 LI BCL
Window manager HP VUE or CDE/X-Windows V.X11R5 Motif V.1.1/1.2 Open Windows 3.0, or CDE RAM 128 megabytes (additional RAM will improve software performance) Swap space 200 megabytes (additional swap space will improve software performance) Hard disk 300 megabytes for minimum installation
500 megabytes for complete installation including online documentation and application examples
SunOS 5.7, 5.8, and 5.9 (Solaris 7.0, 8.0, 9.0)
Display High resolution color only
Installation and User’s Guide 59
1 Introducing the Agilent 85225F Performance Modeling System

The Rack Cabinet

The system is housed in a 1.6 meter rack cabinet. The cabinet provides line power access, ventilation, mobility, and protection to the system instrumentation.
A rack- mounted work surface is included for maximum flexibility and convenience in making in- fixture or coaxial measurements. The work surface can be removed to facilitate on- wafer measurements using a probe station. The work surface is coated with antistatic material and connected to chassis ground. Therefore, an antistatic mat is not required. For installation instructions, see “To install the work surface" on page 74.
60 Installation and User’s Guide
Introducing the Agilent 85225F Performance Modeling System 1

Performance Characteristics and Specifications

Supplemental characteristics are not specifications, but are provided in
Table 15 for your convenience.

Table 15 Supplemental System Characteristics

Characteristic Value
Line voltage 115 volts nominal (90 volts to 132 volts) or 220 volts nominal
(210 volts to 250 volts) Line frequency 48 Hz to 66 Hz Circuit breaker amper a ge rating 6 amps (115 volts), 3.5 amps (220 volts) Rack weight capacity 818 kilograms (1800 pounds) maximum load ed Rack external di m ensions 1620 mm high × 600 mm wide × 905 mm deep Rack footprint (top view)

Interference Standards

The IEC/EN 61326- 1 and CISPR Publication 11 standards define the RFI and EMI susceptibility of the performance modeling system.

Performance Modeling System Performance Specifications

The Agilent 85225F performance modeling system adheres to the performance specifications of an Agilent E8364B PNA Series vector network analyzer. Refer to Appendix G, “Network Analyzer Performance Specification Summary,” starting on page 141.
Installation and User’s Guide 61
1 Introducing the Agilent 85225F Performance Modeling System

RF Subsystem Performance Specifications

The overall performance of a network analyzer is dependent on the individual instruments, system configuration, user- defined operating conditions, measurement calibration, and cables.
For a specification summary, refer to Appendix G, “Network Analyzer Performance Specification Summary,” starting on page 141.
In any high- frequency measurement, residual errors contribute uncertainties to the results.
NOTE
When the system is configured with a probe station, microwave probes, on-wafer calibration standards, or t est fixtures, additional uncertainties are contributed to the measurement results. Refer to the manufacturer’s documentation for information on probe station or test fixture characteristics.

DC Subsystem Specifications

Specifications for the Agilent 4156C precision semiconductor parameter analyzer are listed in its user’s guide, chapter 7 of Volume 1, “General Information.”
Specifications for the Agilent E5260A 8- slot high speed measurement mainframe and Agilent E5270B 8- slot precision parametric measurement mainframe are listed in its user’s guide, Chapter 2, “Introduction.”

Bias Network Characteristics

Table 20, “11612V Option K11/K21 Bias Network Characteristics,” on
page 139 lists the operational characteristics of the bias networks. For detailed information, refer to Appendix F, “Understanding the Bias Networks,” starting on page 137.
62 Installation and User’s Guide
Agilent 85225F Perfor mance Model ing System Installation and User’s Guide
2 Installing the System
To prepare the installation site 64 Environmental Requirements 64 Electrical Requirements 64 To receive the system 65 To unpack the shipment crate containing the rack cabinet 66 To verify the shipment 68 To install the work surface 74 To ensure your safety while using the system 76 Precautions for Performing Floating-Ground Measurements 78 Precautions for Avoiding Electrostatic Discharge 79 To connect the bias networks 80 To switch on power to the system 87 This completes the inst allat i on pro ce ss. To confirm the functional it y of t he
system, continue to Chapter3, “Verifying System Functionality,” starting on page 91. 89
Related Topics “Introducing the Agilent 85 22 5F Perfo rma nce Mod el in g Syst em" on
page 17
Use this chapter to learn how to first prepare the installation site, and then receive, unpack, install, and configure the system. This chapter includes important information on operational safety, as well as instruction on preparing the installation site, unpacking the system, ensuring the completeness of the system shipment, installing the work surface, performing final system configuration, and powering- on the system.
Agilent Technologies
63
2 Installing the System

To prepare the installation site

CAUTION
Follow these steps to prepare the site for system installation.
To prepare the installation site
Step Notes 1 Ensure that your installation site meets the
2 Ensure that your installation site meets the

Table 16 Environmental Requirements

Environmental Parameter System Requirement
Temperature +0°C to +45°C (+32 °F to +113°F) Relative humidity Maximum 80% for temperatures up to 31°C
This product is designed for indoor use in Installation Category II and Pollution Degree 2 per IEC 61010-1 and 664 respectively.
environmental requirem ents.
electrical requirements.
Environmental requirements (temperature,
relative humidity, altitude, and clearance) are listed in Table 16.
Electrical requirements are listed in Table 17.
*†
decreasing linearly to 50% at 40°C Altitude Up to 3000 meters (approximately 10000 feet) Clearance (behind an d abo v e rac k)
* Install air conditioning and heating as needed to achieve the required ambient temperature range. † Accuracy-enhanced measurement performance is specified at an ambient temperature range of
+25°C ±5°C. After calibration, hold the ambient temperature of the measurement environment to ±1°C of the ambient temperature at the time of calibration.
‡ Required to ensure the extractor fans can properly ventilate the system .
15 centimeters (6 inches) minimum

Table 17 Electrical Requirements

Electrical Parameter System Requirement
Supply capability 100/120 volts, 2000 VA
200/240 volts, 2000 VA Circuit sharing Do not connect air conditioning or
motor-operated equipment to the sa me ac
circuit supplying line voltage to the system.
64 Installation and User’s Guide

To receive the system

Follow these steps to store, inspect, and confirm the system shipment.
To receive the system
Step Action Notes
Installing the System 2
1 Store and inspect the
shipment.
2 If the system is damaged,
notify appropriate parties.
a Keep the shipping containers toge ther ,
unopened, located in one area.
b Inspect the s hipping containers for
damage.
If the shipment is damaged, contin ue to
step 2.
If the shipment is verif i ed undamaged,
continue to the next section, “To unpack
the shipment crate containing the rack cabinet" on page 66.
a Report the shipment damage to your
Agilent Technologies sales representative.
b Report the ship ment damage to the
shipping carrier.
c Provide all cartons and packaging material
for inspection by the shipping ca rrier .
Keep all cartons and packaging
material until the entire shipment has been verified undamaged and complete, and the sy stem has passed visual inspection and functional verification.
Agilent Technologies wi ll r ep ai r or
replace damaged equipment without waiting for a claim settlement fr om the shipping carrier.
Installation and User’s Guide 65
2 Installing the System

To unpack the shipment crate containing the rack cabinet

The racked system is shipped upright secured to a pallet. Other system components are shipped separately. Follow these instructions to unpack and inspect the rack cabinet and the racked system components.
Required Tools
9/16 inch wrench or adjustable end wrench
Prying tool to remove packaging clamps
WARNING
Always wear safety glasses when removing the clamps and other packing materials fr om th e cra te .
CAUTION
Be careful not to bend the clamps while removing them from the shipping crate. You may reuse the clamps when the system is repacked.
To unpack the shipment crate containing the rack cabinet
Step Action Notes 1 Remove the outer packing
crate.
a Remove the clamps holding the packing
crate top cover in place.
b Remove the top cover and set it aside. c Remove the clamps holding the first
packing crate wall in place.
d Insure that two other people are available
to hold the last two walls in place as the last set of clamps is removed.
e Remove the other walls. f Set the loading ramp panel aside for now.
Which wall is removed first does not
matter.
In double- rack crates, the heaviest wal l
is the loading ramp. In single-rack crates, the loading ramp is shipped inside the crate, plac ed on to p of th e rack (it is a hinged asse mbly , shipped in the folded position).
2 Remove the packaging
materials.
66 Installation and User’s Guide
a Remove the foam top cover. b Remove the plastic wrapping from the
system.
To unpack the shipment crate containing the rack cabinet (continued)
Step Action Notes 3 Unload the system a Remove the two brace bolts attaching the
side brace assembly to the bottom pallet.
b Remove the side brace assembly. c Lift the hinged slat and remove the ramp
anchor bolt.
d Place one end of the ramp on the p allet
ramp ledge.
e Insert the ramp anchor bolt and fold down
the hinged slat.
f Fold down the ramp’s end flap.
To secure the ramp, you may place long
wood screws through the ramp and into the ramp ledge.
Installing the System 2
WARNING
A racked system is tall and top-heavy. It is easy to ti p the rack over while moving it, which could result in injury or death. Unloading the system safely requires the participation of four persons exercising care so as not to topp l e the rac k cabi net . Do not stand in front of the rack as it rolls down the ramp.
g Ensure that the rack cabinet leveling feet
are retracted and that the cabinet casters are rolling freely.
h Roll the system down the ramp using
extreme care.
i Carefully roll the rack toward its prepared
place within the measurement environment.
In case the system must be moved in
the future, retain and reuse these packing materials. You can also purchase replacement packing materials from Agilent Technologies.
Installation and User’s Guide 67
2 Installing the System

To verify the shipment

Use Table 18 “System Receiving Checklist and Replaceable Parts” to:
confirm the completeness of the shipment
provide component part and model numbers required to order
All replacement items are available from Agilent Technologies. Part numbers for replacement instrument subassemblies are listed in their individual service manuals.
The majority of the system components are shipped preconfigured in the system rack cabinet.
The PGUs (pulse generator units) and GNDU (active ground unit) are factory- installed in the 41501B PGU expander.
With the exception of the bias networks, semi- rigid and SMU triaxial cables, all other cables are connected at the factory.
replacement parts
NOTE
To verify the shipment
Step Action Notes 1 Verify that the serial
numbers on the rear panel of the system inst rument s match the serial numbers listed in the shipping documentation.
Other cables and accessories are shipped inside the rack-mounted storage drawer.
Compare the serial numbers liste d in the shipping documents with the serial numbers on the instrument’s rear panel serial number labels.
If an instrument serial number does not
match the shipping document, report mismatched serial number to your Agilent Technologies sales representative.
If all instrument serial numbers match
the shipping documents, continue to
step 2.
For a list of support contacts, see “To
receive additional assistance" on
page107.
68 Installation and User’s Guide
To verif y the shipme nt (con ti nu ed)
Step Action Notes
Installing the System 2
2 Complete the receiving
checklist.
3 If the system is
incomplete, report missing item s to your Agilent Technologies sales representative.
1
Table 18 Replaceable Parts
Part or Model N umber Description
10833A GPIB cable, 1 meter 10833B GPIB cable, 2 meter
a Compare the Bill of Materials to the system
components received in the shipment.
b Verify the shipment is complete.
If the shipment is confi rm ed incomplete,
go to step3.
If you have confirmed the presence of all
system components, the receiving process is complete. Proc ee d to the next section “To ensure your safety while
using the system"on page76.
Refer to the Bill of Materials included
with the shipment.
For a list of support contacts, see “To
receive additional assistance" on
page107.
10833C GPIB cable, 4 meters 10833D GPIB cable, 0.5 meter 11612T Option K33 Mounting plates, bias networ ks to probe station 11612V Option K11 Bias netwo rk, port 1, 45 MHz to 50 GHz, 0.5 A 11612V Option K21 Bias netwo rk, port 2, 45 MHz to 50 GHz, 0.5 A 11900A Adapter, 2.4 mm (male-to-male) 11900B Adapter, 2.4 mm (female-to-female) 11900C Adapter , 2.4 mm (male-to-female) 1250-0080 Adapter, BNC, 50 ohm (female-female) 1250-1700 Adapter, coax 1250-2405C Adapter , BNC coaxial tee 1250-3231 Adapter, triaxial BNC (female to male) 16048D Test leads, 4 terminal pair, 1.98 meter 16493J Option 001 Interlock cable, 1.5 meter 16493L Option 001 G ND U c a bl e, 1.5 meter
Installation and User’s Guide 69
2 Installing the System
Table 18 Replaceable Parts (continued)
Part or Model N umber Description
16494A Option 001 Triaxial cable, 1.5 meter 16494A Option 002 Triaxial cable, 3 meter 16494B Option 001 Kelvin triaxial cable, 1.5 meter 16494B Option 002 Kelvin triaxial cable, 3 meter 34401A Digital multimeter 35181M Storage drawer 35670A Dynamic signal analyzer 35670A Option AX4 Rack flange kit 35670A Option AY2 Two-input channel configuration 35670A Option AY6 Four-input channel configuration 35670A Option 1D4 Arbitrary source 41501B SMU/PGU expander with GNDU and cable 41501B Option 410 Add 41501B with high power SMU and cables 41501B Option 412 Add high power SMU, 2 PGUs, and cables 41501B Option 420 Add 2 medium power SMUs and cables 41501B Option 422 Add 2 medium power SMUs, 2 PGUs, and cables 41501B Option 902 Cable, power, Europe 41501B Option 903 Cable, power, US and Canada 4156C Precision semiconductor parameter analyzer 4156C Option 010 Delete all 4156C cables 4156C Option 020 Delete Windows controller for pa rameter analysis and
characterization 4156C Option 200 1.5 meter interlock, 4 coaxial, 4 triaxial cables 4156C Option 230 3.0 meter interlock, 4 coaxial, 4 triaxial cables 4284A Precision LCR meter 4284A Option 001 Add DC amplifier 4284A Option 006 Add 2 meter/4 meter cable operation 4284A Option 909 Rack mount kit 4284A Option ABA English documentation 4284A Option ABJ Japanese documentation 5063-9220 Rack mount kit with handles, 2-EIA
70 Installation and User’s Guide
Table 18 Replaceable Parts (continued)
Part or Model N umber Description
5063-9221 Rack mount kit with handles, 3-EIA 5063-9222 Rack mount kit with handles, 4-EIA 5063-9223 Rack mount kit with handles, 5-EIA 5063-9224 Rack mount and handle kit, 6-EIA 5063-9225 Rack mount and handle kit, 7-EIA 8120-1396 Line power cord, 220V 8120-1839 Cable, BNC, 50 ohm, 24 inch 8120-1405 Line power cord, 120V 8120-1840 Cable, BNC, 48 inch 8120-2582 Cable 8120-5068 Cable
Installing the System 2
8490D Option 010 Attenuator, 2.4 mm coaxial, fixe d 10 dB, DC to 50 GHz 85043-20001 Ground stud 85043-20002 Shoulder screw 85043-80013 Anti-static mat kit 85056A Precision calibration kit, 2.4 mm 85056D Economy calibration kit, 2.4 mm 85106-60038 Work surface, 1 meter 85107-20004 Semi-rigid cable, 9 inc h, 2.4 mm (m-m) 85133F 2.4 mm flexible test port cable set 85225-90023 Agilent 85225F Performance Modeling System Installation and
User’s Guide
C2790AC Ballast, 30 pounds E3661B Rack cabinet, 1.6 me ter E3661B Option AW3 Power distributio n unit, 100/120 volts E3661B Option AW5 Power distributio n unit, 220/240 volts E3663AC Rail kit (2 rails per) E3668B Feedthrough panel E4470AZ Extractor fan, 100 to 120 volts E4471AZ Extractor fan, 200 to 240 volts E5250A Low leakage switch mainframe
Installation and User’s Guide 71
2 Installing the System
Table 18 Replaceable Parts (continued)
Part or Model N umber Description
E5252A 10 x 12 switch mat rix E5260A 8-slot high spe ed parametric measurement mainframe E5260A Option 050 50 Hz line power frequency E5260A Option 060 60 Hz line power frequency E5260A Option ABA English documentation E5260A Option ABJ Japanese documentation E5290A High speed high power source mon i tor unit E5291A High speed medium power source monitor unit E5270B 8-slot parametric measurement solution E5270B Option 050 50 Hz line power frequency E5270B Option 060 60 Hz line power frequency E5270B Option ABA English localization E5270B Option ABJ Japanese localization E5280A Precisio n high power source monitor unit E5281A Precisio n medium power source monitor unit E5286A High resolution source monitor unit E5810A LAN/GPIB gateway E5810A Option 100 Rack mount kit E5810A Option AG6 I/O libraries client softw a re for MS Windows E5810A Option ABJ Japanese documentation E7731A Filler panel, 1-EIA unit E7732A Filler panel, 2-EIA unit E7733A Filler panel, 3-EIA unit E7734A Filler panel, 4-EIA unit E7735A Filler panel, 5-EIA unit E7736A Filler panel, 6-EIA unit E7737A Filler panel, 7-EIA unit E8364B PNA Series vector network analyzer, 10 MHz to 50 GHz E8364B Option 010 Time domain analysis capability E8364B Option 014 Configurable test set E8364B Option 016 Receiver attenuators
72 Installation and User’s Guide
Table 18 Replaceable Parts (continued)
Part or Model N umber Description
E8364B Option 022 Extended memory E8364B Option 080 Frequency offset E8364B Option 081 Reference receiver switch E8364B Option 083 Frequency converter measur ement application E8364B Option 1CP Rack mount kit with handles E8364B Option H08 Pulsed RF measurement capability E8364B Option H11 IF access E8364B Option UNL Extended power range
Installing the System 2
Installation and User’s Guide 73
2 Installing the System

To install the work surface

If the system will not be used with a probe station, install the work surface to facilitate in- fixture or coaxial measurements.
The work surface is designed to fit onto the rack below the network analyzer. When the following procedure is performed properly, the work surface maintains a path to chassis ground through the support rails and the rack cabinet.
Follow these instructions to install the work surface.
Required Tools
Medium Pozidriv screwdriver
Small flat- blade screwdriver
To install the work surface
Step Action Notes 1 Prepare to install the work
surface.
a Fully extend the four lock feet at th e bottom
of the rack cabinet.
b Unpack the work surface and the work
surface support rails.
c Lay the rails down so that:
the ends with the single pemmed hole
are facing the front,
the ends with the keyhole-shaped cutout
are facing the rear,
and the rails are facing inward toward
each other.
2 Attach the support rails to
the rack cabinet.
a Pass the large end of the keyhole-shaped
cutout in each rail over the shoulder screw already mounted inside the rack.
b Slide the rails to the rear of the rack. c Use one 1/2 inc h long 10 -32 Pozidr iv scre w ,
one split lock washer, and one flat washer to secure each rail.
d Before tightening th e screws, ensure that
the rails are level.
e Tighten the screws.
The lock feet prevent the cabinet from
moving on the casters.
These are the positions of the support
rails when installed in the rack.
You may need to hold the rails in place
as you tightened the screws.
74 Installation and User’s Guide
To install the work surface (continued)
Step Action Notes
Installing the System 2
3 Attach the work surface
to the support rails.
a Slide the work surface onto the support
rails.
b Slide the work surface all the way back on
the rails until it comes to rest against the front of the rack c a binet.
c Use 1/2 inch long 10-32 screws to secure
the work surface to th e rails from the beneath.
Installation and User’s Guide 75
2 Installing the System

To ensure your safety while using the system

This product has been designed and tested in accordance with international standards. Bias current and voltage are supplied to the DUT from the DC subsystem. This instrument can force dangerous voltages to the FORCE, SENSE, and GUARD connectors. DC subsystem is connected to the device through the bias networks and test fixture or probe station.
WARNING
Failure to comply with the following precautionary safety ins truct ions prior to operating the system could result in serious injury or death.
With some installed options, the Agilent 4156C or Agilent E5260A/70B used in this system can supply voltages up to ±200 volts DC. Depending on operating conditions, hazardous voltages can be present at points in the system that could potentially come in contact with the system operator.
Before operating the system , follow thes e step s to ensur e your saf et y.
To ensure your safety while using the system
Step Action Notes 1 Never operat e the sy st em
without a safety earth ground.
2 Never attempt to service
the system.
a Ensure that a safety earth ground is
connected between the system power distribution unit and the line power source.
b If it is likely that the safety earth ground
has been impaired, the system must be rendered inoperative and secured against unintended operation .
a Contact Agilent Technologies if service is
required.
Capacitors within the system
components can remain charged even after the system is disconnect ed from its line power source.
The system may only be serviced,
adjusted, maintained, or repaired by qualified personnel.
3 Open the DC subsystem
interlock connection whenever possible.
a Close the DC subsystem INTLK (Interlock)
connection only when voltages greater than ±42 volts DC are required.
Depending on installed options, the
SMU output can be as high as ±200 volts DC. As long as the INTLK connection is open, the voltage is clamped to ±42 volts DC maximum.
For instruction on installing an
interlock switch on a sh ielding box, se e “To Make an Interlock Connection” in the 4156C user’s guide (volume 1) or “Connecting the Interlock Terminal” in chapter 3 of the E5270 user’s guide.
76 Installation and User’s Guide
To ensure your safety while using the system (continued)
Step Action Notes
Installing the System 2
4 Be aware of potential
shock hazards during floating-ground measurements.
5 Before touching the
FORCE, SENSE, or GUARD connectors, ensure your safety.
6 Never use replacement
fuses with incorrect ratings.
a Do not touch any of the DC subsystem
output connectors when the shorting bar is disconnected and a floating-ground measurement is in progress.
b Warn others working in the system’s
vicinity of the potential shock hazards .
a Switch off the DC subsystem and discharge
the capacitors.
b If you do not switch off the instruments,
complete ALL of the following precautionary steps:
Terminate the DC subsystem
measurement by pressing the Stop key and confirmi ng t hat t he MEA SUREMEN T indicator is not lit.
Deacti vate th e st and by mo de (if use d) b y
pressing the Standby key and con fir ming that the Standby indicator is not lit.
Confirm that the HIGH VOLTAGE
indicator is not lit.
Open the interlock connection.
a After fin d ing the cause of failure, replace
component fuses with fuses of the same current rating and of the type specified in the instrument’s product documentation .
For additional information, see the
following section “Precautions for
Performing Floating-Ground Measurements.”
Failure to use correctly rated fuses
could result in a fire hazard and damage to the equipment .
7 Install the instrument so
that the ON/OFF switch is readily identifiable and easily reached by the operator.
Installation and User’s Guide 77
The ON/OFF switch is the system
disconnecting device. It disconnects the mains circuit from the mains supply before other parts of the instrument.
Alternately, an externally installed
switch or circuit breaker (readily identifiable and easily reached by the operator) may be used as a disconnection device.
2 Installing the System

Precautions for Performing Floating-Ground Measurements

IC- CAP measurements can be performed with the device in a floating- ground configuration. This prevents ground- loop noise and, in the case of a bipolar junction transistor, damage to the device under test.
A floating- ground configuration is created by removing the shorting bar that connects the CIRCUIT COMMON and CHASSIS GROUND terminals.
WARNING
A potential shock hazard exists when the shorting bar is disconnected for floating-ground measurements. Do not touch any of the DC subsystem rear panel connectors while a floating ground measurement is in progress.
If you are making measurements in a floating- ground configuration, ensure that the shorting bar is disconnected between the CIRCUIT COMMON and
CHASSIS GROUND terminals.

To perform floating-ground measurements

When floating ground measurements are necessary, remove the rear front panel shorting bar connecting the CIRCUIT COMMON and CHASSIS GROUND terminals.
When the shorting bar is removed, you must drive the DUT circuit common with either an SMU, GNDU, or by connecting directly to the DC subsystem circuit common. The circuit common can be found at the DUT ends of the SMU and GNDU cables.
NOTE
The circuit common is not connected through the bias networks.
Read “Precautions for Performing Floating- Ground Measurements" on page 78, then follow these steps to connect the CIRCUIT COMMON to an external ground.
To connect an external ground to the circuit common
Step 1 Remove the shorting bar connecting the CIRCUI T COMMON and CHASSIS
GROUND terminals.
2 Connect the external ground to the CIRCUIT COMMON of the DC subsystem.
78 Installation and User’s Guide

Precautions for Avoiding Electrostatic Discharge

Never operate the system without taking precautions to avoid electrostatic discharge that could damage the system or the device under test.
Installing the System 2
CAUTION
Even relatively small currents resulting from electrostatic discharge undetectable to the system operator can damage current-sensitive devices and system components.
To take precautions against electrostatic discharge
Step 1 Wear an antistatic wrist strap. 2 Connect the wrist strap to chassis ground.
Installation and User’s Guide 79
2 Installing the System

To connect the bias networks

Follow these steps to connect the bias networks to the system, and the device under test to the bias networks.

Agilent 4156C Systems

To connect the bias networks
Step Action Notes 1 For Agilent 4156C
systems, refer to the following figure.
2 Connect the triaxial
cables from the 4156C to the bias networks.
a Connect the triaxial cables from the 4156C
HRSMU1 FORCE and SENSE connectors to the DC FORCE and DC SENSE connectors on the port 1 bias network.
b Connect the triaxial cables from the 4156C
HRSMU2 FORCE and SENSE connectors to the DC FORCE and DC SENSE connectors on the port 2 bias network.
From the rear of the system, route the
cable through the feedthrough panel.
The port 1 bias network is the 11612V
K11.
The port 2 bias network is the 11612V
K21.
Refer to Figure 5 on page 29.
80 Installation and User’s Guide
To connect the bias networks (continued)
Step Action Notes
Installing the System 2
3 Connect and route the
triaxial cable from 4156C HRSMU3 FORCE connector to the port 2 bias network.
4 Connect the Agilent
E8364B test ports to the bias networks.
a From the rear of the system, connect one
end of the triaxial cable to the HRSMU3 FORCE connector on the 4156C rear panel.
b From the rear of the system, route the cable
through the upper feedthro ug h pa nel.
c Connect the triaxial cable to the port 2 bias
network GNDU conne ct or.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the 11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN connector of the 11612V K21 bias network.
The triaxial cable model number is
16494A Option 002.
Leave the GNDU connector on the port
1 bias network open.
The test port cables model number is
85133F.
The 2.4 mm female-to-male adapter
model number is 11900C. Without this adapter, the test port cable will not properly mate with the bias network RF IN connector.
The 2.4 mm male-to-male adapter
model number is 11900A.
5 Connect the bias
networks to the devi ce under test.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias network.
The semi-rigid cables part number is
85107-20004.
Installation and User’s Guide 81
2 Installing the System

Agilent 4156C Systems with Agilent 41501B Expander Box

To connect the bias networks
Step Action Notes 1 For Agilent 4156C with
41501B expander box systems, refer to the following figure.
2 Connect the triaxial
cables from the 4156C to the bias networks.
3 Connect and route the
triaxial cable from 4156C HRSMU3 FORCE connector to the port 2 bias network.
a Connect the triaxial cables from the 4156C
HRSMU1 FORCE and SENSE connectors to the DC FORCE and DC SENSE connectors on the port 1 bias network.
b Connect the triaxial cables from the 41 501B
HPSMU FORCE and SENSE connectors to the DC FORCE and DC SENSE connectors on the port 2 bias network.
a From the rear of the system, connect one
end of the triaxial cable to the GNDU connector on the 41501B rear panel.
b From the rear of the system, route the cable
through the upper feedthro ug h pa nel.
c Connect the triaxial cable to the port 2 bias
network GNDU conne ct or.
From the rear of the system, route the
cable through the feedthrough panel.
The port 1 bias network is the 11612V
K11.
The port 2 bias network is the 11612V
K21.
Refer to Figure 20,“Rear Panel Wiring
Diagram,” on page53.
The triaxial cable model number is
16494A Option 002.
Leave the GNDU connector on the port
1 bias network open.
Refer to Figure 3,“Front Panel
Connections with Agilent 4156C,” on page25.
82 Installation and User’s Guide
To connect the bias networks (continued)
Step Action Notes
Installing the System 2
4 Connect the Agilent
E8364B test ports to the bias networks.
5 Connect the bias
networks to the devi ce under test.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the 11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN connector of the 11612V K21 bias network.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias network.
The test port cables model number is
85133F.
The 2.4 mm female-to-male adapter
model number is 11900C. Without this adapter, the test port cable will not properly mate with the bias network RF IN connector.
The 2.4 mm male-to-male adapter
model number is 11900A.
The semi-rigid cables part number is
85107-20004.
Installation and User’s Guide 83
2 Installing the System

Agilent E5260A/70B Systems

To connect the bias networks
Step Action Notes 1 For Agilent E5260A/70B
systems, refer to the following figure.
2 Connect the triaxial
cables to the E5260A/ 70B
*
SMUs.
a From the front of the system rack, connect
triaxial cables to t he fron t panel h igh powe r SMU FORCE and SENSE outputs on the E5260A/70B.
b Connect a triaxial cable to the f r ont panel
medium powe r SMU F O RCE and SENSE outputs on the E5260A/70B.
c From the front of the system, route the
cable through the upper feedthrough panel.
d From the rear of the system, route the
HPSMU cable through the port 2 hole in the lower feedthrough pane l.
e From the rear of the system, route the
MPSMU cable throu g h the port 1 hole in the lower feedthrough panel.
The triaxial cable model number is
16494A Option 002.
The high speed high power SMU is the
Agilent E5290A.
The high speed medium power SMU is
the Agilent E5291A.
The high power SMU is the Agilent
E5280A.
The medium power SMU is the Agilent
E5281A.
Refer to Figure 4 on page 27
84 Installation and User’s Guide
To connect the bias networks (continued)
Step Action Notes
Installing the System 2
3 Connect the triaxial
cables from the E5260A/70B SMUs t o the bias networks.
4 Connect and route the
ground cable fr om E5260A/70B GNDU to th e port 2 bias network.
5 Connect the Agilent
E8364B test ports to the bias networks.
a Connect the triaxial cable from the
E5260A/70B medium powe r SMU to the DC FORCE and DC SENSE connectors on the port 1 bias network.
b Connect the triaxial cable from the
E5260A/70B high power SMU to the DC FORCE and DC SENSE con nectors on the port 2 bias network.
a From the front of the system , conne ct one
end of the ground triax ial cab le to th e GNDU connector on the E5260A/70B front panel.
b From the front of the system, route the
cable through the upper feedthrough panel.
c From the rear of the system, route the
ground triaxial cable through the port 2 hole in the lower feedthrough panel.
d Connect the ground triaxi al cabl e to the
port 2 bias network GNDU connector.
a Connect one end of the port 1 test port
cable to the Agilent E8364B test port 1.
b Connect the ot her end o f the test port 1 te st
port cable to the 2.4 mm female-to-male adapter.
c Connect the 2.4 mm female-to-male
adapter to the RF IN connector on the 11612V Option K11 bias ne twork.
d Connect one end of the port 2 test port
cable to the Agilent E8364B test port 2.
e Connect the othe r end of t he test po rt 2 test
port cable to the 2.4 mm male-to-male adapter.
f Connect the other end of the 2.4 mm
male-to-male adapter to the RF IN connector of the 11612V K21 bias network.
The medium power SMU is the Agilent
E5281A.
The high power SMU is the Agilent
E5280A.
The port 1 bias network is the 11612V
K11.
The port 2 bias network is the 11612V
K21.
The ground triaxial cable model number
is 16493L Option 002.
Leave the GNDU connector on the port
1 bias network open.
The test port cables model number is
85133F.
The 2.4 mm female-to-male adapter
model number is 11900C. Without this adapter, the test port cable will not properly mate with the bias network RF IN connector.
The 2.4 mm male-to-male adapter
model number is 11900A.
Installation and User’s Guide 85
2 Installing the System
To connect the bias networks (continued)
Step Action Notes 6 Connect the bias
networks to the devi ce under test.
* Refer to Figure4, “Front Panel Wiring Diagram with Agilent E5260A or E5270B,” on page27.
a Connect one semi-rigid cable to the RF/D C
OUT connector of the port 1 bias netwo rk .
b Connect the other semi-rigid cable to the
RF/DC OUT connector of the port 2 bias network.
c Connect the device under test to the
semi-rigid cable atta ched to the port 1 bias network.
d Connect the devi ce under test to the
semi-rigid cable atta ched to the port 2 bias network.
The semi-rigid cables part number is
85107-20004.
86 Installation and User’s Guide

To switch on power to the system

Installing the System 2
NOTE
To switch on power to the system
Step Notes 1 Ensure that the individual instru ments are
configured to match the available line power
*
source.
2 Ensure that all component line power
switches are set to the OFF position.
3 Connect the system to line power. 4 Switch the rack cabinet ~ Line switch from
to (from standby to energized).
5 Switch on the low leakage switch mainframe
line power.
6 Switch on the precision LCR meter line
power.
7 If present, on the 4284A SYSTEM CONFIG
page, highlight the value in th e GPIB ADDRESS field, and press 24 > Enter.
Proper system operation is depe ndent on the sequence in which the system components are switched on.
For information regarding line power
settings, refer to the individual instrument manuals.
This sets the correct system GPIB address
(24) for the 4284A.
8 If present, on the 35670A front panel, press
Local/GPIB > ANALYZER ADDRESS > 22 > ENTER.
9 If present, switch on the Agilent 4156C
precision semiconductor parameter analyzer line power.
10 If present, switch on the 41501B expander
line power.
11 Switch on the Agilent 4156C precision
semiconductor parameter analyzer line power.
12 On the 4156C, press System >
[MISCELLANEOUS], move the pointer to the POWER LINE FRE QU E NCY field, and press [50 Hz] or [60 Hz].
Installation and User’s Guide 87
This sets the correct system GPIB address
(22) for the 35670A.
The expander must be switched on before
the 4156C.
Ensure that th e 41 501B has already been
switched on prior to activating the 4156C.
This ensure s that the 4156C is configured to
match the available line power frequency.
The value is set to 60 Hz at the factory.
Use th e front panel arrow keys t o mov e the
cursor.
2 Installing the System
To switch on power to the system
Step Notes 13 On the 4156C, press System >
[MISCELLANEOUS], hi ghlight the 4156C value in the GPIB ADDR ESS field, and press 19 > Enter.
14 On the E5260A/70B, move the cursor to
CONFIG, pres s Enter, move the cursor to ADDRESS, press Enter. Use the arrow keys to set the address to 19 and press Ent er.
15 Switch on the Agilent E8364B PNA Series
vector network analyzer line power.
16 On the Agilent E8364B PNA Series v ector
network analyzer, from the Main dialog, select System > Configure > SICL/GPIB. In the SICL/GPIB dia log GPI B group b ox, se lect the Talke r/ L istener radio button and select 16 in the Address scroll list.
17 If present, switch on the LAN/GPIB gateway
line power.
18 Switch on the computer line power. 19 Allow the system to warm up for one hour.
* If the system is to be use d with an autotransformer, ensure that t he common terminal is connected to the
neutral (gr ou nded) side of the powe r source.
This sets the correct system GPIB address
(19) for the 4156C.
This sets the correct system GPIB address
(19) for the E5260A/70B.
This sets the correct system GPIB address
(16) for the E8364B.
Refer to the LAN/GPIB gate way documenta tion for instruction on installation and configuration.
88 Installation and User’s Guide
Installing the System 2

To configure the LAN/GPIB gateway for functional verification

If your system includes an Agilent E5810A LAN/GPIB gateway, follow these steps to configure the LAN/GPIB gateway in order to verify the functionality of the performance modeling system.
NOTE
This procedure explains how to configure IC-CAP to use the default LAN/GPIB gateway server IP address in order to verify the functionality of the performance modeling system.
After functional verification, contact your corporate IT professional and ask for a permanent server IP address assignment for the LAN/GPIB gateway. For instructions on changing the server IP address, see the LAN/GPIB gateway installation and configuration guide.
To configure the LAN/GPIB gateway
Step Action Notes 20 Start the IC-CAP
software.
21 Add the interface to the
IC-CAP Hardware Setup.
*
PC version:
a Select Start > Programs > IC-CAP 2004 >
IC-CAP.
UNIX version:
a Open a UNIX terminal window. b At the prompt, ty pe iccap. c Press Enter.
a From the IC-CAP/Main window menu bar,
choose Tools > Hardware Setup... .
b Below the HP-IB Interface group box, click
Add Interface.
c In the Add HP-IB In terface dialog b ox, enter
lan[192.0.0.192]:hpib.
d Click OK.
This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main windows.
This opens the IC-CAP/Hardware
Setup window.
This opens the Add HP-IB dialog box.
This configures the LAN/GPIB
gateway (with its default se rver address) as the performance modeling system GPIB interface.
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model numbe r
85190D.
† If there is an existing IC-C AP i nterface (for example, HP-IB), select the existing interface and click Delete Interface before continuing to
the next action.
This completes the installation process. To confirm the functionality of the system, continue to Chapter 3, “Verifying System Functionality,” starting on page 91.
Installation and User’s Guide 89
Agilent 85225F Perfor mance Model ing System Installation and User’s Guide
3 Verifying System Functionality
To choose a verification process 92 Understanding the System Functional Verification Test 94 Performing the System Functional Verification Test 95
Related Topics “To enhance measurement accuracy" on page 112
“Performing a Coaxial System Measurement Calibration" on page 115 “Performing the DC Subsystem Functional Verification Test"on page 121 “Performing the RF Subsystem Functional Verification Test" on page 125
Use the procedures in this chapter to verify the functionality of the Agilent 85225F performance modeling system. This chapter includes procedures for choosing varying degrees of functional verification and performing the required post- installation system functional verification test.
Agilent Technologies
91
3 Verifying System Functionality

To choose a verification process

System functionality can be verified using several different processes, depending on the level (system or subsystem) of functional verification required and the available tools.
The procedure provided in “Performing the System Functional Verification
Test" on page 95 verifies that all of the system instruments interface
correctly, and that the system can make software- driven measurements using a controller running the IC- CAP software.
Manual functional verification procedures for DC, RF, CV, and 1/f noise subsystem components can be found in:
Appendix B, “DC Subsystem Functional Verification Test,”
starting on page 119
Appendix C, “RF Subsystem Functional Verification Test,”
starting on page 123
Appendix D, “CV Subsystem Functional Verification Test,” starting on
page 129
Appendix E, “1/f Noise Subsystem Functional Verification Test,” starting
on page 133
NOTE
To choose a system verification process
Situation Action Note 1 If the system has recently
been installed or one of the DC or RF subsystem instruments has been replaced, and you have IC-CAP software...
These tests do NOT verify that the system instrumentation conform to their individual performance specifications.
To verify the performance of the individual system components, complete the appropriate performance tests listed in their individual product documentation.
Follow these steps to choose a system functional verification process based upon your current situation.
Complete the steps listed in “Performing the
System Functional Verification Test"on
page95.
Completion of the System Functional
Verification Test is required after system installation, or whenever an RF or DC subsystem component has been serviced or replaced.
The System Functional Verification Test
verifies the functionality of instruments in the RF and DC subsystems. To verify the functionality of other system components, continue to the appropriate situation listed in this table.
92 Installation and User’s Guide
To choose a system verification process (continued)
Situation Action Note
Verifying System Functionality 3
2 If you would like to
manually check the functionality of the Agilent E8364B PNA Series vector ne twork analyzer without using the GPIB interface...
3 If you would like to
manually check the functionality of the Agilent 4156C precision semiconductor parameter analyzer without using the GPIB interface...
4 If you would like to
manually check the functionality of the Agilent E5260A/70B without using the GPIB interface...
5 If you would like to
manually check the functionality of the Agilent 428 4A wi th out using the GPIB interface...
Complete the steps listed in “Performing the
RF Subsystem Functional Verification Test"on
page125.
Complete the steps listed in “Performing the
DC Subsystem Function al V erifica tion Test" on
page121.
Complete the steps listed in “Performing the
DC Subsystem Function al V erifica tion Test" on
page121.
Complete the steps listed in “Performing the
CV Subsystem Functional Verification Test"on
page131.
6 If you would like to
manually check the functionality of the Agilent 35670A without using the GPIB interface...
7 If you would like to
manually check the functionality of the Agilent E5250 low leakage switch mainframe without using the GPIB interface...
Installation and User’s Guide 93
Complete the steps listed in “Performing the
1/f Noise Subsystem Functional Verification Test" on page135.
Complete the Agilent E5250A self-test found in Chapter 3 of the low leakage switch mainframe user’s guide.
3 Verifying System Functionality

Understanding the S ystem Functional Verification Test

The system functional verification test is a standard IC-CAP measurement and simulation procedure. The system performs DC and S- parameter extraction, optimization, and simulation using a fixed 10 dB attenuator as the device under test (DUT).
The procedure provided in “Performing the System Functional Verification
Test” confirms:
the IC- CAP software can communicate with and control the system
instrumentation via GPIB (through the LAN/GPIB gateway, if so configured)
the system can make measurements and display the results
the IC- CAP software can simulate data
the IC-CAP software can converge the simulated data with the
extracted (measured) data
The system applies forward and reverse current to the attenuator and monitors voltage at the attenuator’s input and output. IC- CAP then uses the measured data to model the device- intrinsic resistances and transmission line delay.

Required Tools

Agilent 85225F performance modeling system
A system controller
Agilent 85190A IC- CAP software
Test port cables
Agilent 8490D 10 dB fixed RF attenuator
BNC tee (2)
Agilent 11900A, 2.4 mm male- to- male adapter
Agilent 11900B, 2.4 mm female- to- female adapter
Agilent 11900C, 2.4 mm female- to- male adapter
Agilent 85056A 2.4 mm precision calibration kit, or
Agilent 85056D 2.4 mm economy calibration kit
*
* For requirements, see “The System Controller" on page 59 . † These compon en t s ar e s up p li e d a s part of the system. ‡ For CV subsystem verification only
94 Installation and User’s Guide

Performing the System Functional Verification Test

Complete the following steps to verify system functionality using the supplied Agilent 8490D 10 dB fixed RF attenuator as the device under test.
To perform the system functional verification test
Step Action Notes
Verifying System Functionality 3
1 Switch on power to the
system.
2 Connect the device under
test to the bias networks.
a Complete the steps listed in “To switch on
power to the system"on page87.
a Refer to the following figure.
Proper system function is dependent
upon the order in which the system components are switched on.
b Connect one end of the 2.4 mm
female-to-female ada pter to the semi-rigid cable attached to the RF/DC OUT connector of the port 1 bias network.
c Connect male end of the attenuator to the
other end of the 2.4 mm female-to-female adapter.
d Connect the female e nd of the attenuator to
the semi-rigid cable attached to the RF/DC OUT connector of the port 2 bias netwo rk .
Installation and User’s Guide 95
The 2.4 mm female-to-female adapter
is a 11900B. Use the 11900B provided in the calibration kit.
The port 1 bias network is a 11612V
K11.
The port 2 bias network is a 11612V
K21.
3 Verifying System Functionality
To perform the system functional verification test (continued)
Step Action Notes 3 Start the IC-CAP
software.
4 Add the system inte rfa ce
and components to the IC-CAP Hardware Setup.
5 For 4156C systems,
change the Agilent 4156C precision semiconductor parameter analyzer HRSMU names in IC-CAP.
*
PC version:
a Select Start > Programs > IC-CAP 2004 >
IC-CAP.
UNIX version:
a Open a UNIX terminal window. b At the prompt, ty pe iccap. c Press Enter.
a From the IC-CAP/Main window menu bar,
choose Tools > Hardware Setup... .
b If you have not added an interface, click
Add Interface, enter the system interface
(hpib or other), and click OK. c In the Instrument List, click Rebuild. d Verify that all system components appear
in the Instrument List.
a In the Instrument List, select HP4156 (hpib,
19) and click Configure... .
b In the Configuration of HP4156 window
Unit Table group box, highlight the
characters in the HRSMU1 ent r y bo x and
type VG. c In the Unit Table group box, highlight the
characters in the HPSMU entry box and
type VD.
d Click OK. e Close the Hardware Setup window.
This starts IC-CAP and opens the
IC-CAP/Status and IC-CAP/Main windows.
This opens the IC-CAP/Hardware
Setup window.
This polls the GPIB and adds all
connected and activated system instruments to the Instrument List.
Disregard error messages on the
system instrument displays. The errors are a by-product of the GP IB po l li ng process.
This ope ns the Configuration of
HP4156 window where the Unit Table dialog box is used to change t he names of the HRSMUs.
Actions b, c, and d change the name of
HRSMU1 to VG and HPSMU to VD.
Renaming the SMUs is necessary for
proper execution of the example model file.
6 For E5260A/70B syst ems,
change the Agilent 4156C precision semiconductor parameter analyzer SMU names in IC-CAP.
96 Installation and User’s Guide
a In the Instrument List, select Agilent E5270
(hpib, 19) and click Configure... .
b In the Configuration of E5270 window Unit
Table grou p bo x, hig hligh t th e charact ers in
the MPSMU<slot number> entry box and
type VG. c In the Unit Table group box, highlight the
characters in the HPSMU<slot number>
entry box and type VD.
d Click OK. e Close the Hardware Setup window.
This ope ns the Configuration of E5270
window where the Unit Table dialog box is used to chan ge th e names of th e SMUs.
Actions b, c, and d change the name of
MPSMU<slot number> to VG and HPSMU<slot number> to VD.
Renaming the SMUs is necessary for
proper execution of the example model file.
To perform the system functional verification test (continued)
Step Action Notes
Verifying System Functionality 3
7 Open the attenuator test
model in IC-CA P.
8 Set the model variables
for the measur ement in IC-CAP.
9 Set the network analyzer
instrument options for the attenuator test model in IC-CAP.
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../examples/model_files. c In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../model_files/misc. d In the Files list of the File Open dialog,
double-clic k on sys_testrf.mdl.
a Click the Model Variables tab folder. b Highlight the freq_start v ariable and enter
4.5E+07.
c Highlight the freq_stop variable and enter
50E+09.
d Highlight the imax variable and enter 90m.
a In the DUTs-Setups tab folder , click
S_vs_freq in the Select DUT/Setup list. b Click the Instrument Options tab folder. c Highlight the Cal Type value and enter H. d Highlight the Ca l File Name value and ent er
TEST.CST.
This opens the File Open dialog box.
This ope ns a list of model files.
(Scroll th e list , if nee ded.) Th is dis plays
a list of modeling f iles i n the Fil es list o f the File Open dialog.
This opens the Atten model window.
The Atten model window contains tab
folders used to interact with the model data.
This ope ns the Model Variables tab
folder.
This sets the start frequ ency of the
model to 45 MHz.
This sets the stop frequency of the
model to 50 GHz.
This se ts the maximum current to
90 milliamps.
This opens the 5 tab folders used for
the S_vs_freq setup.
This ope ns the Instrument Options tab
folder.
Adjustable instrument parameters are
listed in group boxes titled as
instrument model number.bus address. instrument address (for example, AgilentPNA.7.16 for the network
analyzer group box).
Terminate your value entries by
pressing Enter on the controller keyboard.
Installation and User’s Guide 97
3 Verifying System Functionality
To perform the system functional verification test (continued)
Step Action Notes 10 Configure the instrument
state for calibration.
11 Calibrate the network
analyzer.
a Disconnect the attenuator from the
semi-rigid cables. b In the Utility key group on the network
analyzer front panel, press Macro/Local >
Preset.
c In the Channel menu, click Power... .
d In the Power dialog box, scroll the Test Port
Power value to 20 dBm, then click OK. e In the Sweep menu, point to Number of
Points and click 101.
f In the Sweep menu, select IF Bandwidth. .. .
g In the IF Bandwidth dialog box, scroll to
1.000 kHz, then click OK.
h In the Sweep menu, select Sweep Setup... .
i In the Sweep Setup dialog box, select the
Stepped Sweep check box and click OK.
a In the network analyzer’s Calibrati on menu,
select Calibra tion Wizard... .
b In the Calibration Wizard: B egin C alibratio n
dialog box, select SmartCal (GUIDED Calibration): Use Mechanical Standards radio button and click Next.
c Follow the displayed prompts to calibrate
the network analyzer.
This removes th e instrument from GPIB
control, activates the front panel interface, and returns the instrument to its factory preset condition.
Actions c through h prepare the
network analyzer for a full range 2-port calibration using the same instrument state settings used by the IC-CAP functional verification test macro.
For detailed instructions, see
“Performing a Coaxial System Measurement Calibration"on
page115.
This begins a modified full 2-port
calibration.
When prompted to connect a standard
to either Port 1 or Port 2, connect the standard to the semi-rigid cable attached to the Por t 1 or Port 2 bias network.
An isolation calibration is not needed
for this measurement.
12 Save the calibration and
instrument state data to the C:/Program Files/
Agilent /Network Analyzer/Documents
folder.
98 Installation and User’s Guide
a In the File menu, select Save As... .
b In the Save As di alog box, using the
keyboard or by clicking Edit File Name, enter TEST.CST then click OK.
This saves the calibration and
instrument state data in the network analyzer’s operating system
C:/Program Files/Agilent/ Network Analyz er/ D ocu m ent s folder.
To perform the system functional verification test (continued)
Step Action Notes
Verifying System Functionality 3
13 Make the DC and RF
measurements in IC-CAP.
14 Interpret the results. a Observe the displayed plots.
a Reconnect the attenuator. b In the Atten model window, click the
Macros tab.
c In the Select Macro: list, select Test_atten
and click Execute.
d When prompted to ensure you have
calibrated the network analyzer, enter Y and click OK.
e When prompted to enter a cal set number
use the default (1) and click OK.
f As IC-CAP performs an optimization,
observe the simulated data trace converge with the measured data trace.
b In the IC-CAP/Status window, observe the
final DC/RF values parameter and the Final RMS error.
This opens the Atten model window’s
Macros tab folder.
This starts a sequence of prompts and
responses.
IC-CAP performs measurements of the
attenuator’s DC and RF parameters.
When complete, IC-CAP displays plots
of the forward and reverse DC voltage transfer through the attenuator. The solid line is the measured data, the dashed line is the simulated data.
IC-CAP be gins the optimization process
using default simulated data. These values are optimized to converge with the measured data.
When the S-parameter measurement is
complete, IC-CA P displays plots of the S12/21, 20×log magnitude, and S21 phase.
of the S21
10
**
IC-CAP also lists the attenuato r’s
resistance values. The error between measured and simulated data should be less than 2%.
The error between the measured and
simulated S21 phase data should be less than 2%.
The value for T1.TD is the transmission
time through the attenuator , modeled as transmission line delay. Thi s v alue (typically in the femtosecond to nanosecond range) de pends on the length of the attenua tor.
The S21 magnitude and the S12/S21
plots, of less significance, are included for your interest.
The S21 simu lated trace is determined
from the measured DC resistances, and therefore is not expected to converge with the measured data over the full frequency range. However , the measured S21 data will show a normal frequency response variation.
The S12/S21 plot (also displayed on
the network analyzer) confirms that the system is capable of making forward and reverse transmission measurements.
Installation and User’s Guide 99
3 Verifying System Functionality
To perform the system functional verification test (continued)
Step Action Notes 15 For systems with the
Agilent 4284A precision LCR meter, open the
junction capacitance model (juncap.mdl) in IC-CAP.
16 Select the DUT/Setup. a Select the DUTs-Setups tab.
17 Set the LCR meter
instrument options.
18 Calibrate the LCR meter. a On the 16048D test leads, connect one tee
a From the IC-CAP/Main window menu bar,
choose File > Examples... .
b In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../examples/model_files.
c In the Directories list of the File Open dialog
box, double-cli ck on the directory
.../model_files/diode.
d In the Files list of the File Open dialog,
double-clic k on juncap.mdl.
b Click the area > cv setup. a Select the Instrument Options tab.
b Highlight the Cable Leng th variable and
enter 2.
between the H
b Connect the other tee between the L
L
connectors. Select the
cur
Measure/Simulate tab.
c Select Calibrate. d Follow the IC-CAP prompts.
pot
and H
connectors.
cur
pot
This opens the File Open dialog box.
This ope ns a list of model files.
(Scroll th e list , if nee ded.) Th is dis plays
a list of modeling f iles i n the Fil es list o f the File Open dialog.
This opens the juncap model window.
The juncap model win dow contai ns tab
folders used to interact with the model data.
This opens the capacitance vs. voltage
setup.
This se ts the cable length in the LCR
meter instrument options.
This perfor ms a calibration on the LCR
meter.
and
The BN C tee part number is 1250-2405.
19 Measure the open circuit
capacitance.
20 Observe the results a After the calibration is complete, observe
* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model numbe r
85190D.
† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example
config1.hwd (the file suffix must be .hwd).
**S12 is identical to S21 because the attenuator is assumed to be symmetrical.
a Select Measure.
the plot.
b Rescale the measured result (displayed in
red) cap.m.
This measu res the open circuit
capacitance.
The measured result should be less
than ±10E-15 fara ds .
This completes the functional verification procedure.
100 Installation and User’s Guide

If you encounter a problem

Check the system connections and settings:
1 system connections to the DUT 2 system interconnections 3 GPIB cabling 4 GPIB address settings
Perform the DC subsystem self- test in Appendix B, “DC Subsystem
Functional Verification Test,” starting on page 119.
Perform the Agilent E8364B PNA Series vector network analyzer
operator’s check in Appendix C, “RF Subsystem Functional Verification Test,” starting on page 123.
Perform the Agilent 4284A precision LCR meter self- test in
Appendix D, “CV Subsystem Functional Verification Test,” starting on
page 129.
Verifying System Functionality 3
Perform the Agilent 35670A dynamic signal generator self- test in
Appendix E, “1/f Noise Subsystem Functional Verification Test,” starting
on page 133.
Installation and User’s Guide 101
Loading...