Atec Agilent-85047A User Manual

300 kHz to 6 GHz
New test sets for the Agilent 8753 network analyzer family
Solid-state test port switches are standard for the Agilent 85046A/B and 85047A S-Parameter Test Sets. These new test sets improve the speed, accu­racy and ease of RF component measurements with the 8753.
There are three S-Parameter test sets in the 8753 family. The 85046A is a 50-ohm model that oper­ates from 300 kHz to 3 GHz. The 85046B is a 75-ohm version that covers the frequency range of 300 kHz to 2 GHz. The 85047A is a 50-ohm test set that extends the frequency range to 6 GHz.
Faster, more accurate S-parameter measurements
Improve the speed and accuracy of your S-Parameter measurements with a solid-state switching test set. These new test sets provide faster, more accurate measurements by continuously updating all four S-Parameters when a full two-port error correction is used. Measurement speed is improved by 160 milliseconds each time the trace is updated for sweeps with 101 points or more.
Measure forward and reverse parameters simultaneously
Solid-state switching allows the selection of any two S-Parameters, both continuously updated, on the display. For example, when adjusting the bias of an amplifier, it is often required to simultane­ously monitor the forward gain and output match. Mechanical versions of the 85046A/B and 85047A test sets will update either forward or reverse para­meters continuously but not a combination of both.
Simplify tuning procedures
Solid-state switching is ideal for use in tuning applications where all four S-Parameters are changing. It is no longer necessary to press MEA­SURE RESTART when using a full two-port error correction. All four S-Parameters are being con­tinuously measured.
Improved reliability for automatic measurement systems
Solid-state switches have no mechanical parts, and therefore will not wear out in a continuously meas­uring automatic test system. Mechanical test set switches (Option 009) typically have a lifetime in excess of three million cycles. The solid-state switches will typically extend the lifetime indefinitely.
Agilent 85046A/B and 85047A Solid-State Switching S-Parameter Test Sets
Data Sheet
2
Achieve the highest system accuracy with faster measurement speeds
The specifications listed in the table and the typi­cal measurement uncertainties shown below apply to all Agilent 8753C systems using either solid­state switching (standard) or mechanical (Option
009) versions of the test sets.
Test set specifications
Specifications describe the instrument’s warranted performance over the temperature range 0 to +55˚C (except where noted).
8753C System measurement uncertainty
2
The graphs below show the typical measure­ment uncertainty for the 8753C using full 2-port error correction with the Agilent 85031B 7-mm Calibration Kit.
Transmission measurements
3
Reflection measurements
4
Parameter 85046A (50 Ω) 85046B (75 Ω) 85047A (50 Ω)
Frequency Range: 300 kHz to 3 GHz 300 kHz to 2 GHz 300 kHz to 3 GHz/3 MHz to 6 GHz
Directivity
1
35 dB: 300 kHz to 1.3 GHz 35 dB: 300 kHz to 1.3 GHz 35 dB: 300 kHz to 1.3 GHz
Port 1, 2: 30 dB: 1.3 GHz to 3 GHz 30 dB: 1.3 GHz to 2 GHz 30 dB: 1.3 GHz to 3 GHz
25 dB: 3 GHz to 6 GHz
Effective Source 14 dB: 300 kHz to 2 MHz 14 dB: 300 kHz to 2 MHz 20 dB: 300 kHz to 1.3 GHz Match1: 20 dB: 2 MHz to 1.3 GHz 17 dB: 2 MHz to 1.3 GHz 16 dB: 1.3 GHz to 3 GHz
16 dB: 1.3 GHz to 3 GHz 16 dB: 1.3 GHz to 2 GHz 14 dB: 3 GHz to 6 GHz
Switch Repeatability: ±0.03 dB
1. Can be greatly improved by accuracy enhancement techniques.
2. These measurement uncertainty curves utilize an RSS model for the contribu­tions of random errors, and a worst-case model for the contributions of dynamic accuracy and residual systematic errors.
3. The graphs shown for transmission measurements assume a well-matched device (S11=S22=0).
4. The graphs for reflection measurement uncertainty apply to either a one-port device or a two-port device with more than 6 dB of insertion loss.