Atec Agilent-83557A, Agilent-83558A User Manual

High performance in a small package
Utilize the Agilent Technologies 83557A/83558A mm-wave source modules to obtain high power, high quality signals covering the full wave­guide bands of 50 to 75 GHz (V band) and 75 to 110 GHz (W band). These efficient frequency multipliers trans­late a 50 mW (+17 dBm) microwave signal (12.5 to 18.75 GHz) to a mm­wave signal at a very low cost com­pared to other alternatives. With their reliable, solid state design, these modules are small and lightweight and can be operated remotely from the driving source to better accomo­date your measurement setup needs.
High output power
With >+2 dBm to 75 GHz and >-1 dBm to 110 GHz, these mm-wave source modules can be used as LO’s in mixer measurements and provide additional dynamic range for insertion loss/gain measurements. The output power is leveled to provide level control and improved source match at the device under test. Additionally, the output power level can be read from the front of the source, or on the 8349B ampli­fier, depending on your configuration.
Spectral purity
The 83557A/83558A source modules offer harmonic and sub-harmonic suppression of <-20 dBc when used with low harmonic sources like the 8360 series synthesized sweeper, the 8673C/D synthesized signal generators, or the standard 8350B/83550A sweep oscillator. In many mm-wave measure­ments, low harmonics are important to achieve greater measurement accu­racy and increased dynamic range.
Dedicated sources simplify connections
Use the 83623A synthesized sweeper or the 8350B/83550A sweep oscillator to drive these source modules directly without requiring an external amplifier. With a dedicated source module interface, these sources display both the frequency and the power level at the module’s output. Or, choose from a wide range of sources to drive the source modules using an 8349B microwave amplifier. The 8350B sweep oscillator makes an excellent, low-cost, swept source. For increased frequency accuracy and stability, use the 8360 series synthesized sweepers, the 8340/41 series synthesized sweep­ers, or the 8673 series synthesized signal generators.
Swept V/W band network analysis
With the 8757 scalar network analyzer, or the 8510 vector network analyzer, complete V and W band measurement systems can be configured using the 8350B, 8360 series, or 8340B/8341B microwave sweepers and the mm-wave source modules. Using the 8757, simultaneous scalar reflection and transmission measurements can be made in waveguide. With the 8510, full S-parameter measurements can be made using two microwave sources and the 83557A/83558A mm-wave modules.
Agilent 83557A/83558A 50 to 75 GHz/75 to 110 GHz
mm-Wave Source Modules
Typical maximum leveled output power for V and W band mm-wave source modules
)
4
3
83557A
2
83623A Synthesized Sweeper
8350/83550A Sweep Oscillator
or
8349B Microwave Amplifier
8360 8350B/8359X 8340B/8341B 8673B/C/D/G/H Microwave Sources
83557A mm-Wave Source Module
50-75 GHz
or
83558A mm-Wave Source Module
75-110 GHz
1
Power (dBm)
0
–1
50 75 110
Frequency (GHz
83558A
2
Specifications
Specifications describe the instrument’s warranted performance over the temperature range 0° to 55°C (except where noted). Supplemental characteristics are intended to provide information useful in applying the instrument by giving typical but non-warranted performance parameters. These are denoted as “typical,” “nominal,” or “approximately.”
8350B/83550A or 8350B/83592C/95C 8341B Opt. 003
1
8673C/D
1
Agilent 83557A 50-75 GHz 8349B 83623A or 83624A 8349B 8349B
Leveled Power Range (25 ± 5 °C) –5 to +2 dBm –5 to +3 dBm –5 to +2 dBm –5 to +2 dBm Unleveled (Opt. 001) –5 to +4 dBm –5 to +4 dBm –5 to +4 dBm –5 to +4 dBm
Power Level Accuracy (25 ± 5 °C) ±2.5 dB ±2.0 dB ±2.5 dB ±2.5 dB
Power Flatness (Max Leveled Power) ±2.0 dB ±1.5 dB ±2.0 dB ±2.0 dB
Source Output SWR
Leveled 2.0 2.0 2.0 2.0 Unleveled (Typically) 3.0 3.0 3.0 3.0
Spurious Signals
2
–20 dBc –20 dBc –20 dBc –20 dBc
8350B/83550A or 8350B/83592C/95C 8341B Opt. 003
1
8673C/D
1
Agilent 83558A 75-110 GHz 8349B 83623A or 83624A 8349B 8349B
Leveled Power Range (25 ± 5 °C) –5 to –1 dBm –5 to 0 dBm –5 to –1 dBm –5 to –1 dBm Unleveled (Opt. 001) –5 to +1 dBm –5 to +1 dBm –5 to +1 dBm –5 to +1 dBm
Power Level Accuracy (25 ± 5 °C) ±2.5 dB ±2.0 dB ±2.5 dB ±2.5 dB
Power Flatness (Max Leveled Power) ±2.0 dB ±1.5 dB ±2.0 dB ±2.0 dB
Source Output SWR
Leveled 2.0 2.0 2.0 2.0 Unleveled (Typically) 3.0 3.0 3.0 3.0
Spurious Signals –20 dBc –20 dBc –20 dBc –20 dBc
Common Specifications
Frequency Accuracy, Resoluton, and Stability 4 (83557A) or 6 (83558A) times the frequency accuracy, resolution, and stability of the input signal.
Accuracy is the same as the time base for synthesized sources.
External Pulse Modulation On/Off Ratio >80 dB >80 dB >80 dB >80 dB
(>60 dB for 83550A)
Rise/Fall Time (Typically) 10 ns 50 ns 50 ns 40 ns
(25 ns for 83550A)
Min Leveled RF Pulse Width (Typically) 1 µs 1 µs 1 µs 1 µs
Amplitude Modulation
Rate (3 dB BW) (Typically) DC – 100 kHz DC – 250 kHz DC – 100 kHz DC – 100 kHz Sensitivity (Typically) 1 dB/V (100%/V, for synthesized sources)
1. Specifications apply for low harmonic sources only. The standard 8340B/8341B and the 8673B/G/H also provide the same source capabilities except the spurious output of the modules is at 0 dBc.
2. Expressed in dB relative to the carrier (dBc)
3
General Specifications
Input Frequencies 12.5 to 18.75 GHz Minimum input power level into RF input cable +17 dBm (50 mW) Maximum input power level into RF input cable +27 dBm (0.5 W) Waveguide output connector 83557A EIA size WR-15 waveguide. Mates with
JAN UG 385 flange.
83558 EIA size WR-10 waveguide. Mates with
JAN UG 387 (mod.) flange.
Weight Net, 1.8 kg (4 lb.)
Ordering Information
Agilent 83557A 50 to 75 GHz mm-Wave Source Module Option 001 Deletes Leveling Coupler and Detector Option 910 Extra Manual Option W30 Two Additional Years of Return-to-Agilent Service
Agilent 83558A 75 to 110 GHz mm-Wave Source Module Option 001 Deletes Leveling Coupler and Detector Option 910 Extra Manual Option W30 Two Additional Years of Return-to-Agilent Service
Agilent 8349B 2.0 to 20.0 GHz Microwave Amplifier Option 001 Rear Panel RF Input/Output Option 002 Rear Panel RF Input and Front Panel Output Option 910 Extra Manual Option W30 Two Additional Years of Return-to-Agilent Service
Furnished with each 83557A Part number
Operating and Service Manual 83557-90019 Procedure and Parts for 8340 Series/83590 Series
0.5 V/GHz Modification 83557-90016 RF Cable 5061-5359 Source Module Interface 5061-5391 Module Base Assembly 83557-60010
Furnished with each 83558A Part number
Operating and Service Manual 83558-90019 Procedure and Parts for 8340 Series/83590 Series
0.5 V/GHz Modification 83558-90016 RF Cable 5061-5359 Source Module Interface 5061-5391 Module Base Assembly 83558-60010
Top
210 mm (8.27 in)
Vertical Adjustment Range
35 mm (1.38 in)
80 mm (3.15 in)
80 mm
105 mm (4.13 in)
(3.15 in)
Front
Right Side
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measure­ment capabilities you paid for and obtain the support you need. Our extensive sup­port resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc­tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product informa­tion, including realistic performance spec­ifications and practical recommendations from experienced test engineers. When
you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabil­ities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve prob­lems efficiently and gain a competitive edge by contracting with us for calibration, extra­cost upgrades, out-of-warranty repairs, and on-site education and training, as well as design, system integration, project man­agement, and other professional services. Experienced Agilent engineers and techni­cians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measure­ment accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test and measurement needs.
Online Assistance
www.agilent.com/find/assist
Phone or Fax
United States: (tel) 1 800 452 4844
Canada: (tel) 1 877 894 4414 (fax) (905) 206 4120
Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599
Australia: (tel) 1 800 629 485 (fax) (61 3) 9272 0749
New Zealand: (tel) 0 800 738 378 (fax) (64 4) 495 8950
Asia Pacific: (tel) (852) 3197 7777 (fax) (852) 2506 9284
Product specifications and descriptions in this document subject to change without notice.
Copyright © 1989, 2000 Agilent Technologies Printed in U.S.A. 8/00 5958-0398