Agilent 6620A Series Multiple-Output, 40-105 W DC Power Supplies, GPIB
Data Sheet
Speed and accuracy for test optimization
• Up to four fully isolated power supplies in a 3U package
• Dual-range outputs
• Fast, low-noise outputs
• Built-in measurements and advanced programmable features
• Protection features to ensure DUT safety
Two, three, or four isolated outputs are integrated into one package, conserv­ing rack space and GPIB addresses. Most of the outputs also provide dual ranges, for more current at lower voltage levels. The outputs can be connected in parallel or series to fur­ther increase the flexibility that these products offer the system designer.
Programming is done using industry standard SCPI commands. Test system integration can be further simplified by using the VXIplug&play drivers. These power supplies help reduce test time with fast up and down programming, enhanced by an active downprogrammer which can sink the full rated current.
Specifi cations
Specifications
(at 0 ° to 55 °C unless otherwise specified)
Output power
Low-range: voltage, current 0 to 7 V 0 to 20 V 0 to 7 V 0 to 20 V 0 to 35 V
High range: voltage, current 0 to 20 V 0 to 50 V 0 to 20 V 0 to 50 V
Output combinations
for each model (total number of outputs)
6621A (2) 2 6622A (2) 2 6623A (3) 1 1 1 6624A (4) 2 2 6627A (4) 4 6623A-J03 (3) Special order option 2 1
Programming accuracy at 25 °C ± 5 °C
Voltage 0.06% + 19 mV 50 mV 19 mV 50 mV 35 mV Current 0.16% + 50 mA 20 mA 100 mA 40 mA 30 mA
Readback accuracy at 25 °C ± 5 °C
Voltage 0.05% + 20 mV 50 mV 20 mV 50 mV 35 mV +Current 0.1% + 10 mA 4 mA 20 mA 8 mA 6 mA –Current 0.2% + 25 mA 8 mA 50 mA 20 mA 15 mA
Ripple and noise (peak-to-peak, 20 Hz to 20 MHz; rms, 20 Hz to 10 MHz)
Constant voltage rms 500 μV 500 μV 500 μV 500 μV 500 μV peak-to-peak 3 mV 3 mV 3 mV 3 mV 3 mV Constant current rms 1 mA 1 mA 1 mA 1 mA 1 mA
Load regulation
Voltage 2 mV 2 mV 2 mV 2 mV 2 mV Current 1 mA 0.5 mA 2 mA 1 mA 2 mA
Load cross regulation
Voltage 1 mV 2.5 mV 1 mV 2.5 mV N/A Current 1 mA 0.5 mA 2 mA 1 mA N/A
Line regulation
Voltage 0.1% + 1 mV 1 mV 1 mV 1 mV 1 mV Current 0.06% + 1 mA 01 mA 1 mA 1 mA 1 mA
Transient response time Less than 75 μs for the output to recover to within 75 mV of nominal value following
40 W output 40 W output 80 W output 80 W output 105 W output
0 to 5 A 0 to 2 A 0 to 10 A 0 to 4 A 0 to 3 A
0 to 2 A 0 to 0.8 A 0 to 4 A 0 to 2 A
a load change within specifi cations
Supplemental characteristics
(Non-warranted characteristics determined by design and useful in applying the product)
Average programming resolution
Voltage 6 mV 15 mV 6 mV 15 mV 10.5 mV Current 25 mA 10 mA 50 mA 20 mA 15 mA OVP 100 mV 250 mV 100 mV 250 mV 175 mV
Output programming response time
(time to settle within 0.1% of full scale output, after Vset command has been processed)
2
40 W output 40 W output 80 W output 80 W output 105 W output
2 ms 6 ms 2 ms 6 ms 6 ms