High quality results
• High confidence testing with contact check function
• Remove parasitics with error correction
• Consistent data with 0.6% basic accuracy
• Compensation for handler contact chattering with trigger delay
Versatile measurements
• Select from four test parameters
• Use a variety of test fixtures and accessories
• Perform a charge-measure­discharge sequence with the test sequence program function
• Save and recall up to ten measurement setups
High test throughput: 4349B
• 9.5 ms measurements
• 4-channels for multiple DUTs
• 4-channel simultaneous testing
• Fast contact checking: 2 ms/ measurement
• GPIB and handler interfaces
• Ideal for high volume capacitor testing
Agilent
4339B/4349B High Resistance Meters
Technical Overview
2
Open:
Compensates leakage current and capacitance in test fixtures and cables.
Contact check:
Verifies reliability of capacitor test connections.
Save/Recall:
Stores and retrieves up to 10 measurement states.
Comparator:
Selects values for high, in, and low testing.
Measurement parameter:
Selects the right parameter for the measurement.
Display:
LCD with back-light. Displays measurement values, instrument states, and comparator results.
Test sequence program:
Controls a series of resistance measurements. (Charge-measure-discharge)
Agilent 4349B 4-channel high resistance meter
Key parameters and specifications
Agilent 4339B Agilent 4349B
Test channels 1 2ch, Option 4349B-001
4ch, Option 4349B-700
Test voltage(Vdc) 0.1 to 1000 Requires external power source
1
Measurement parameters R, I, pv, ps R, I Measurement range
(())
103to 1.6x10
16
103to 10
15
Basic accuracy 0.6% 2% Display resolution 3 / 4 / 5 digits 3 / 4 / 5 digits Measurement time 10 ms/30 ms/390 ms 9.5 ms/28 ms/98 ms/397 ms
1. External power source required for resistance measurements. Recommendation for external power source for
measurement for 1 Gsample at 100 Vdc with accuracy = ±10%:
Ripple: 1 mVrms (50/60 Hz)
Wideband noise: ≤ 5 µVrms/Hz (50 Hz)
Switching noise: 50 mVrms (100 kHz)
3
High quality measurements with flexible hardware
• Resolve data to 5 digits (3, 4, or 5 digits selectable)
• Make precise measurements with 0.6% basic accuracy
• Verify DUT performance at the exact voltage rating
• Reliable and safety measurements with Agilent 16339A component test fixture
4339B solutions for high voltage material testing
• Resistivity mathematics built-in: surface and volume
• Agilent 16008B resistivity cell for solid samples
• Easy measurements with test sequence program function (controls charge-measure­discharge sequence)
• Customize your fixture cabling with the Agilent 16117C test leads
System features you need to be successful
• Maximize accuracy with error correction
• Test capacitor contact failure with contact check function
• Automate testing with GPIB interface
• Reduce ground-loops with isolated handler interface
• Pass/Fail testing with comparator function (high/in/low)
Capacitor evaluation with the 4349B
• Optimize capacitor Vdc rating tests
• Increase throughput four times with 4-channels
• Improve reliability with contact check
• Get low noise results with Agilent 16117E test lead
4
Specifications Measurement Accuracy
Agilent 4339B test conditions*:
1. Warm up time: 30 minutes
2. Ambient temperature: 23 °C ± 5 °C
3. Test cable length: 1.5 meters
4. Open error correction performed
5. Long measurement time setting
6. Contact check: off
Accuracy parameters:
Rm: Measured resistance value in W Im: Measured current value in
amperes
Vs: Source voltage in volts Vo: 0.1 V@Vs 200 V, 0.5 V@Vs > 200 V
Accuracy example:
To determine the accuracy of a meas­urement use Figure 1, “Conversion diagram”. For example: determine the accuracy of a 5 MΩ (= 5 x 10
6
) measurement
at 50 Vdc. Rm = 5 x 10
6
Vs = 50 V
The intersection of Rm running paral­lel to the 10
6
diagonal line intersects
the vertical Vs line at the second row from the bottom of the diagram. Mov­ing horizontally across to Table 1, the following equation is found:
100 Vo + (2.5 x 10
-7
x Rm)
0.5 +
Vs
Entering the values for Rm, Vo, and Vs yields an accuracy of ±0.725%.
Figure1. Conversion diagram Table 1. Agilent 4339B measurement accuracy (±% of reading)
Resistance RM (Ω)
Current IM (A)
Resistance measurement
Accuracy
Current measurement
Accuracy
Test voltage setting Vs (V)
Vo= 0.1 V @ Vs≤ 200 V
= 0.5 V @ V
s
> 200
5
Agilent 4349B test conditions
1
:
1. Warm up time: 30 minutes
2. Ambient temperature: 23 °C ± 5 °C
3. Test cable length: 1.5 meters
4. Open error correction performed
5. 30 ms measurement time setting
Accuracy parameters:
Rm: Measured resistance value in ohms Im: Measured current value in amperes
External power supply parameters:
Vs: Source voltage in volts Vo: Source offset voltage in Volts Av: Voltage accuracy
Figure 2. Conversion diagram Table 2. Agilent 4349B measurement accuracy (±% of reading)
Resistance RM (Ω)
Current IM (A)
Resistance measurement
accuracy
Current measurement
accuracy
Test voltage setting Vs (V)
1. Other test condition data available in the operation manual.
6
Other Specifications
Measurement parameters/ranges
Parameter Range Agilent 4339B
R (dc resistance) 103 Ω to 1.6 x
10
16
I (dc current) 60 fA to 100 µA
ps (surface Refer to operation resistivity) manual
pv (volume Refer to operation resistivity) manual
Agilent 4349B
R (dc resistance) 103 Ω to 10
15
I (dc current) 1 pA to 100 µA
Measurement conditions and functions
DC test voltage (4339B): 0 V to 1000 V, 0.1 V steps @ V ≤ 200V, 1.0 V steps @ V > 200 V DC test voltage (Agilent 4349B): None supplied, use external power supplies and voltage data entry for resistance measurements. Maximum of 5000 V input and 5 digit numerical entry. Max current (Agilent 4339B): 10 mA @ 100 V, 5 mA @ 250 V, 2 mA @ 500 V, 1 mA @ 1 kV Number of test channels: 4339B: 1 channel, 4349B: Option 4349B-700:4 ch Option 4349B-001:2 ch
Ranging: Auto and hold Trigger: Internal, manual, and
external Delay time (trigger): 0 ms to 9999 ms in 1 ms steps Test cable lengths: 2 meters maximum Measurement time (typical): 4339B: 10 ms / 30 ms / 390 ms 4349B: 9.5 ms / 28 ms / 98 ms /
397 ms
Other instrument functions
Error correction: Open (removes errors due to parasitics). Comparator: High, in, and low for each of the test parameters. Save/Recall: 10 instrument states from non-volatile memory. Contact check: Detects contact failure for capacitive devices (2 ms). GPIB: Agilent’s implementation of IEEE 488 for control and data. Handler interface: Negative logic and isolated. Signals are high/in/low, no contact, EOM, index, alarm, keylock, ext. trigger.
Physical characteristics
Power: 90 – 132 Vac or 198 – 264 Vac. 47 Hz – 66 Hz. 45 VA (typical)
Operating temperature/humidity:
0 – 45 °C/95% RH @ 40 °C. Dimensions: 320(W) x 100(H) x 450(D) mm. Weight: 6.5 kg (typical).
7
Test Fixtures/Accessories
Agilent 16339A component test fixture
For manual high voltage testing of descrete components. For 4339B only.
Agilent 16117B low noise test leads
Wide jaw clip leads for 4339B. 1 meter cable. Applicable measurement range: 1 x 10
11
Ω (typical). For 4339B only.
Option 16117B-001 adds a pair of pin-type probes. Option 16117B-002 adds a pair of socket adapters for connecting to a custom made fixture. Option 16117B-003 adds a pair of alligator clips.
Agilent 16117C low noise test leads
Interlock, voltage source, and current sensing cables. Terminations are threaded triaxial, standard BNC, and bare interlock pair. Female BNC and triaxial connectors are included. For 4339B only.
Agilent 16118A tweezer test fixture
Tweezer test fixture for easy testing for chip components. Maximum applied voltage: 100 Vdc. Applicable measurement range: ≤ 1 x 10
11
(typical). For 4339B only.
Agilent 16008B resistivity cell
For resistivity measurements of dry sheet samples. Upper electrode is spring loaded to apply pressure. Surface and volume measurements. Installed with 50 mm diameter elec­trode. Option 16008B-001 adds 26 mm/76 mm diameter electrodes. Option 16008B-002 adds 26 mm diameter electrode. For 4339B only. Maximum applied voltage: 1000 Vdc. Sheet thickness range: 10 µm to 100 mm.
Agilent 16117E low noise test lead
Male-triaxial to male-triaxial connec­tors. One meter cable. One female­triaxial connector included. For 4349B only.
Agilent 16064B LED Display/Trigger box
Displays comparator status. Cable length 1.5 meters. Manual external trigger. For 4339B only.
Option 16117B-001
Option 16117B-003
Option 16117B-002
Ordering Information
= Choose ONE and ONLY one
q = Choose any combination
Agilent 4339B High Resistance Meter
1
Furnished accessories: shunt connector
Documentation options
2
q Option 4339B-ABA Add specified quantities of English manual q Option 4339B-ABD Add specified quantities of German manual q Option 4339B-ABJ Add specified quantities of Japanese manual q Option 4339B-ABE Add specified quantities of Spanish manual q Option 4339B-0BW Add service documentation, assembly level
Certification option
q Option 4339B-1A7 ISO 17025 compliant calibration
Agilent 4339B test fixtures and accessories
Agilent 16008B Resistivity cell (50 mm diameter electrode)
q Option 16008B-001 Add 26 mm and 76 mm diameter electrodes q Option 16008B-002 Add 26 mm diameter electrode
Agilent 16117B Low noise test leads
q Option 16117B-001 Add pin probes q Option 16117B-002 Add soldering sockets q Option 16117B-003 Add alligator clips
3
Agilent 16117C Low noise test leads Agilent 16118A Tweezer test fixture Agilent 16064B LED display/trigger box Agilent 16339A Component test fixture
Agilent 4349B High Resistance Meter
4
Test channel options
5
Option 4349B-700 4-channelsOption 4349B-001 2-channels
Agilent 4349B high resistance meter
6
Documentation options
q Option 4349B-ABA Add specified quantities of English manual q Option 4349B-ABJ Add specified quantities of Japanese manual q Option 4349B-0BW Add service documentation, assembly level
Cabinet options
7
q Option 4349B-1CM Rack mount kit q Option 4349B-1CN Front handle kit
Certification option
q Option 4349B-1A7 ISO 17025 compliant calibration
Agilent 4349B Test Fixtures and Accessories
Agilent 16117E Low noise test lead
1. Test fixture is not furnished with the 4339B.
2. Manual is not furnished with the 4339B.
3. The alligator clips are not furnished as standard.
4. External power source required for resistance measurements.
5. 2-channels to 4-channels upgrade not available.
6. Manual is not furnished as standard.
7. Rack flange and handle kit are not compatible.
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assis­tance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integra­tion, project management, and other professional engineer­ing services. Experienced Agilent engineers and techni­cians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instru­ments and systems, and obtain dependable measurement accuracy for the life of those products.
Agilent T&M Software and Connectivity
Agilent’s Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your com­puter with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit
www.agilent.com/find/connectivity
for more information.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online Assistance:
www.agilent.com/find/assist
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc., 1996, 2000, 2002, 2003, 2004 Printed in USA, April 13, 2004 5964-6182E
Agilent Email Updates
Loading...