Foreword
Thank you for purchasing Arjo equipment.
Your MultiClean™ disinfection system is part of a
series of quality products designed especially for
hospitals, nursing homes and other health care use.
Please contact us if you have any questions about the
operation or maintenance of your Arjo equipment.
Please read this Instructions for Use
thoroughly!
Please read this Instructions for Use (IFU) in its entirety
before using your MultiClean disinfection system.
Information in this IFU is crucial to the proper
operation and maintenance of the equipment, and will
help protect your product and ensure that equipment
performs to your satisfaction. Some of the information
in this IFU is important for your safety and must be read
and understood to help prevent possible injury. Arjo
strongly advise and warn that only Arjo designed parts,
which are designed for the purpose, should be used on
equipment and other appliances supplied by Arjo, to
avoid injuries attributable to the use of inadequate parts.
Unauthorized modifications on any Arjo equipment
may affect its safety. Arjo will not be held responsible
for any accidents, incidents or lack of performance that
occur as a result of any unauthorized modification to its
products.
Serious incident
If a serious incident occurs in relation to this medical
device, affecting the user, or the patient then the user or
patient should report the serious incident to the medical
device manufacturer or the distributor. In the European
Union, the user should also report the serious incident to
the Competent Authority in the member state where
they are located.
Contact your local Arjo representative for replacement
parts. Contact information is found on the last page of
the IFU.
Definitions in this IFU
WARNING
Means:
Safety warning. Failure to understand and obey this
warning may result in injury to you or to others.
CAUTION
Means:
Failure to follow these instructions may cause damage
to all or parts of the system or equipment.
NOTE
Means:
This is important information for the correct use of this
system or equipment.
Means:
The name and address of the manufacturer.
Service and Support
A service routine must be performed on yourMultiClean disinfection system every year by Arjo
authorized service personnel to ensure the safety and
daily operation of your product. See section Care and Preventive Maintenance on page 16.
If you require further information, please contact Arjo
for comprehensive support and service programs to
maximize the long-term safety, reliability and value of
the product.
4
Page 5
Intended Use
This equipment is intended for disinfection
and cleaning of bathing and showering
equipment in hospital or care facilities. The
MultiClean disinfection system is designed for
use only with Arjo disinfectant/cleaners such
as Arjo baths, air spas, showering equipment
and other Arjo bathing accessories.
The MultiClean disinfection system is intended to be
used by appropriately trained caregivers with adequate
knowledge of the care environment, its common
practices and procedures, and in accordance with the
guidelines in the Instructions for Use.
The MultiClean disinfection system should only be
used for the purpose specified in the Instructions for
Use. Any other use is prohibited.
Resident Assessment
This equipment is not intended for residents.
Indication for Use
Installation and Service Requirements
The MultiClean disinfection system must be installed by
suitable qualified personnel according to the Assembly
and Installation Instructions which only can be superseded
by local code.
Expected Lifetime
The expected life of the MultiClean disinfection system
is unless otherwise stated ten (10) years, if subject to
preventive maintenance being carried out in accordance
with the instruction found in section Care and Preventive Maintenance on page 16.
The MultiClean disinfection system equipped with
Arjo disinfectant/cleaner is a disinfecting system
intended to process noncritical medical devices and
equipment surfaces located in bathing / showering
rooms and similar environments.
It is suitable to be used on noncritical medical devices
that make only topical contact with intact skin.
Contra Indications
The MultiClean disinfection system is not to be used to
clean / disinfect critical or semicritical medical devices
that are used in surgical or invasive procedures.
5
Page 6
Safety Instructions
WARNING
To avoid eye and skin damage, always use
protective glasses and gloves. If contact
occurs rinse with plenty of water. If eyes or
skin becomes irritated, contact a physician.
Always read the IFU and the Material Safety
Data Sheet (MSDS) of the disinfectant.
WARNING
If disinfectant liquid is ingested, DO NOT
induce vomiting. Give milk or water to
drink. Seek medical advice as necessary.
WARNING
To avoid eye or skin irritation, never
disinfect in the presence of a patient.
WARNING
Never use the MultiClean disinfection
system or disinfectant on people.
WARNING
To avoid electric shock, make sure that the
equipment is connected to:
• Continuously powered supply mains with
protective earth.
• Separate fuse and ground fault circuit
interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance
with local codes and regulations.
NOTE
Modifications made to the equipment without
the express acknowledgement of Arjo will
invalidate supplier’s product liability.
6
Page 7
Intentionally left blank
7
Page 8
Parts Designation
1
10B
10A
10
11
12
13
14
45
3
2
6
9
8
7
1Disinfectant Bottle
3 l (101 fl.oz)
2Viewing Window
3Disinfectant Storage Tray
4Flowmeter
5Isolation Valve (shut off) Control
I = Open
O = Closed
6Bottle Connector
7Access door
8ON/OFF Control
I = On
O = Off
9Door Lock
10 Function Selection Control
A (Yellow) = Disinfect
B (Blue) = Rinse
11 Spray Handle Holder
12 Spray Handle
13 Shower Hose 2 m (79")
3 m (118") available
14 Adjustment Screw
To adjust concentration setting
8
Page 9
Fig. 1
Fig. 2
Product Description/Functions
NOTE
The information in this manual was correct at the time
of printing. However, Arjo reserves the right to
incorporate changes or modifications to the
specifications without prior notice.
The MultiClean disinfection system addresses the concerns over
the effective cleaning and disinfection of bathing and showering
products for hospitals, nursing homes and other health care
environments.
When used with the Arjo range of disinfectants, the MultiClean
disinfection system delivers a solution of diluted disinfectant at
a preset concentration.
A wall mounted unit, the MultiClean disinfection system is
intended for use in bath and shower rooms. It is designed for use
with Arjo baths, air spas, showering equipment and other supplied
bathing accessories. (See Fig. 1)
The MultiClean disinfection system is not to be used as a patient
shower. (See Fig. 2)
The MultiClean disinfection system features a trigger spray
handle with a 2 m (79”) hose. The spray handle stores
conveniently and safely in a holder at the side of the unit. Two
controls on the main panel enable the user to activate the unit and
select between disinfecting and rinsing modes.
A lockable door gives access to the 3 l (101 fl.oz) disinfectant
supply bottle, isolation valve and concentration setting cap.
The MultiClean disinfection system is available in two models:
•Standard (Mains Water Fed)
•Pumped (Electric Pump Operated)
These are supplied according to installation conditions and local
plumbing code requirements.
The Standard system incorporates an in line back flow prevention
device (vacuum breaker DA device - USA/Canada, pipe
interrupter DB device - Europe), with a positive leakage to drain.
The pumped system is protected from the mains water supply by
an integral water tank with type AA air gap.
9
Page 10
Operating Instructions
WARNING
To avoid eye and skin damage, always use
protective glasses and gloves. If contact occurs
rinse with plenty of water. If eyes or skin becomes
irritated, contact a physician. Always read the IFU
and the Material Safety Data Sheet (MSDS) of the
disinfectant.
General Disinfection Instructions for Arjo
Bathing Equipment
Where multiple use of bathing equipment is undertaken, there are
real risks of cross-contamination from one person to the next. By
definition, the users of these products are high risk persons and
challenges from hostile micro-organisms should be avoided. Arjo
bathing systems rely upon a strict cleaning regime using the
MultiClean disinfection system. The following schedule has been
devised to optimise the hygienic cleaning of bathing or showering
products after each occupant has been removed and the water
drained to waste.
Usage
We recommend that bathing, showering equipment and
accessories should be disinfected/cleaned at the following times.
Start of day
Equipment should be disinfected and rinsed thoroughly before the
first use of the day.
After each use
Equipment should be disinfected and rinsed thoroughly after each
use.
End of each day
Equipment should be disinfected, wiped and dried after the last
use of the day. This is of particular importance to ensure that
soiled water/surfaces are not left to stand overnight.
Setting the Disinfectant Concentration
(10 steps)
1Unlock and open the access door and with the bottle
connector secured in place, place a bottle of Arjo
disinfectant/cleaner onto the disinfectant storage tray.
10
Continue with the steps on the next page.
Page 11
2Make sure all services are connected to the MultiClean
disinfection system.
3Open the water supply by turning the isolation valve
control to the open position.
Fig. 1
e.g. ArjoClean setting 30-35
Fig. 2
4Switch the unit on by turning the ON/OFF control to the ON
position.
5Remove the spray handle from its holder and direct it
against the bottom of the bath or an appropriate drainage
point, but always away from the user.
6Turn the function selection control to the disinfect setting.
7Press the trigger on the spray handle.
8Check the liquid value on the flowmeter setting so it is in
accordance with the settings table Disinfectant flowmeter/mix ratio table on page 11. (See Fig. 1)
If necessary, adjust using the adjustment screw on the bottle
connector - clockwise to decrease the value, anti clockwise
to increase. (See Fig. 2)
9Put the spray handle back in its holder. Close and lock the
access door.
10 Switch the unit off by turning the ON/OFF control to the
OFF position.
Contact Times
The contact time of the disinfectant can be varied depending on
the specific environmental circumstances. Arjo recommend that
10 minutes is the optimum contact time to insure best disinfection.
This should be practised at the start and end of the day and after
each time the bath or shower is used.
However, it may be appropriate for the contact time to be a
minimum of 2 minutes where risk of cross-contamination is
considered low.
Disinfectant flowmeter/mix ratio table
DISINFECTANT
FLOWMETER
SETTING
MIX RATIO
ArjoClean (EU)30–3516 ml/1 l–1440 ppm
CenKleen IV (USA)25–3060 ml/4 l2fl.oz/US gallon873 ppm
$UMR Disinfectant Cleaner IV (CA)40–4532 ml/4 l1fl.oz/US gallon868 ppm
KILL
STRENGTH
$UMR General Purpose
Disinfectant (CA)
25–3032 ml/4 l1fl.oz/US gallon868 ppm
11
Page 12
Disinfect and Clean
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
(13 steps)
This instruction follows the yellow workflow illustrated on the
MultiClean disinfection system cover.
According to Fig. 1
1Close the waste outlet of the bath.
2Close the suction if fitted.
3Turn the function selection control to the disinfection
setting.
4Remove the spray handle from the holder and direct the
spray handle into the product and turn the ON/OFF control
to the ON position.
According to Fig. 2
5Secure the spray handle into the injection port.
6Depress the spray handle trigger until disinfectant solution
is flowing from all air-jets/nozzles.
7Continue injecting disinfectant for 30 s before
disconnecting the spray handle.
According to Fig. 3
8Apply disinfectant to all inner surfaces of the product using
the spray handle.
9Make sure all surfaces are wetted by the disinfectant
solution.
10 Similarly, apply disinfectant to all accessories and
attachments, behind and under the seat (where applicable)
by spraying with the spray handle.
According to Fig. 4
11 Scrub the surface of the product with a soft bristle brush or
cloth to remove any build up of bio-film which may have
been deposited.
According to Fig. 5
12 Allow an appropriate contact time for sanitization or
disinfection according to the instructions for the
disinfectant.
13 While waiting, wipe all other contact areas and accessories
with a disposable cloth soaked in disinfectant solution.
12
Page 13
Rinse
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
(10 steps)
This instruction follows the blue workflow illustrated on the front
of the MultiClean disinfection system cover.
According to Fig. 6
1Turn the function selection control to the rinse setting.
2Open the waste outlet and the suction if fitted and drain the
bath of residual disinfectant solution.
According to Fig. 7
3Reconnect the spray handle to the injection port.
4Close the suction if fitted.
5Depress the spray handle trigger until water is seen flowing
from all air-jets/nozzles.
6Repeat the rinsing procedure (steps 1 – 6) until all
disinfectant in the system is carried away.
According to Fig. 8
7Rinse all parts clean.
According to Fig. 9
8Activate and run the air spa (where fitted) for 30 s to purge
the system of any remaining water.
According to Fig. 10
9After the last disinfection of the day, dry the bathing product
and any accessories.
10 Switch the unit off by turning the ON/OFF control to the
OFF position.
NOTE
At the end of the day close the isolation valve to prevent
unauthorized use.
13
Page 14
Disinfectant Replenishment
Follow these 13 steps for proper replenishment.
WARNING
To avoid eye and skin damage, always use
protective glasses and gloves. If contact
occurs rinse with plenty of water. If eyes or
skin becomes irritated, contact a physician.
Always read the IFU and the Material Safety
Data Sheet (MSDS) of the disinfectant.
NOTE
Use only $UMR disinfectant. If you require
further information regarding the $UMR
disinfectant please contact your local
service department.
1Make sure the ON/OFF control is switched to the
OFF position. For location of parts, see section
Parts Designation on page 8.
2Using the key provided, open the access door.
1 1 Check the liquid value on the flowmeter to make
sure it is in accordance with the setting range in
the recommendations. If necessary, adjust the
adjustment screw on the bottle connector clockwise to decrease the value, anti clockwise to
increase. See section Setting the Disinfectant Concentration on page 10.
12 Put the spray handle back in its holder.
13 Switch the unit off by turning the ON/OFF control
to the closed position.
3Partially withdraw the empty bottle from the
disinfectant storage tray. Unscrew and remove
the bottle connector.
4Remove the empty bottle and use the lid from the
new bottle to seal. Dispose of the used container
safely.
5Refit and tighten the bottle connector. Place the
new bottle of ARJO disinfectant onto the
disinfectant storage tray and close and lock the
door.
6Wipe any residual disinfectant from the unit.
7Switch the unit on by turning the ON/OFF control
to the ON position.
8Remove the spray handle from its holder and
direct it against the bottom of the bath or an
appropriate drainage point.
9Turn the function selection control to the
disinfectant setting.
10 Press the trigger on the spray handle.
14
Page 15
Cleaning and Disinfection Instructions
WARNING
To avoid eye and skin damage, always use
protective glasses and gloves. If contact
occurs rinse with plenty of water. If eyes or
skin becomes irritated, contact a physician.
Always read the IFU and the Material Safety
Data Sheet (MSDS) of the disinfectant.
CAUTION
Do not use phenol, citric acid or petroleumbased solvents, trichloroethylene, chlorine
and hydrogen peroxide or similar
disinfectant as these may damage the
plastic material.
Steam, ethylene oxide sterilization or high
pressure wash can not be used either.
WARNING
To avoid cross-contamination be aware
that alcohol has limited disinfection effect
on bacterial and fungal spores.
WARNING
To avoid eye or skin irritation, never
disinfect in the presence of a patient.
Cleaning
Clean the MultiClean disinfection system with a soft
cloth soaked with diluted disinfectant dispensed from
the handset. Alternatively, use a soap or synthetic
detergent and water soaked cloth ensuring that the unit
is dried after cleaning.
15
Page 16
Care and Preventive Maintenance
The MultiClean disinfection system is subject to wear and tear, and the following actions must be performed when
specified to make sure the product remains within its original manufacturing specification.
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow
the recommended maintenance schedule. In some cases due to heavy use of the product and
exposure to aggressive environment more frequent inspections should be carried out. Local
regulations and standards may be more stringent than the recommended maintenance
schedule.
The equipment is a subject for wear and tear and the following maintenance instructions must be performed when
specified to ensure that the equipment remains within its original manufacturing specification.
Preventive Maintenance Schedule: MultiClean Disinfection System
CUSTOMER OBLIGATION Action/CheckEvery DAYEvery WEEKEvery YEAR
CleaningX
Examine for damageX
Check flowmeter settingX
Examine for leaksX
Yearly checks by qualified personnel onlyX
Yearly Checks by Qualified Personnel Only
The MultiClean disinfection system must be serviced once a year in accordance with the Maintenance and Repair
Manual.
Contact your local Arjo service representative to sign for a service agreement.
WARNING
To avoid injury and/or unsafe product, the maintenance activities must be carried out at the
correct frequency by qualified personnel using correct tools, parts and knowledge of
procedures. Qualified personnel must have documented training in maintenance of this
device.
NOTE
All Caregiver Obligations are to be checked when performing the Qualified Personnel Service. For
details, see separate service instructions.
16
Page 17
Troubleshooting
NOTE
Only suitably qualified personnel should be consulted for any plumbing and electrical problems or
faults. For service or additional technical information please contact your local Arjo service
department.
PROBLEMACTION
No fluid from handset when trigger is
pressed.
Lack of disinfectant in fluid mix (the
presence of disinfectant can be gauged
by checking the flowmeter or the state of
the fluid mix - foam, citrus smell).
Reduced flow rate from handset.•Check the strainers and inlet /outlet fittings are not
•Check water and electrical services for signs of
interruption. Reinstate supply where necessary.
•Make sure the isolation valve is set to the open
position.
•Make sure the ON/OFF control is set to the ON
position.
•Check shower hose for damage or kinks.
•Make sure the strainers and inlet / outlet fittings
are not blocked.
•Check the function selector is set to the disinfect
setting.
•Check fluid level in disinfection bottle.
•Check the position of adjustment screw on bottle
connector. Unscrew and recalibrate the flowmeter
as required.
•Check disinfectant supply hose for damage or
kinks.
blocked.
•Make sure the supply pressure/flow rate is high
enough.
Fluid leaks from handset / hose or main
body of unit.
•Tighten any loose fittings
•Seal wear or damage - contact your local Arjo
service department.
17
Page 18
Technical Specifications
MultiClean Disinfection System – Pumped Version
MEASUREMENTS, VOLUME AND WEIGHTS
Unit height578 mm (22 3/4 inches)
Unit width484 mm (19 inches)
Unit depth200 mm (8 inches)
Total weight (excluding ArjoClean Disinfectant/Cleaner)13 kg (29 lbs)
Total weight (including ArjoClean Disinfectant/Cleaner)16 kg (35 1/2 lbs)
CONNECTION DEMANDS
Inlet Coupling (Cold Water)15 mm (1/2 inches)
Overflow19 mm (3/4 inches)
Minimum Flow Rate - Inlet6 l/minute (1.6 US gallon/minute)
Minimum Flow Rate - Outlet5 l/minute (1.3 US gallon/minute)
Minimum Dynamic Pressure0.3 bar (2.9 PSI)
Maximum Static Pressure6 bar (87 PSI)
ENVIRONMENTAL LIMITATIONS
Maximum ambient temperature40 ºC (104 ºF)
Maximum relative humidity (non-condensing)80%
Disinfectant Storage within unitminimum 0° C – maximum 30 °C
(minimum 32 °F – maximum 86 °F)
WARNING
To avoid injury, always follow the allowed combinations listed in this IFU. No other
combinations are allowed.
Allowed Combinations
Arjo Baths
(Connection to bath system option: air spa inlet)
Freedom BathAll models
Parker (500)UK market only
Model
18
Page 19
MultiClean Disinfection System – Standard Version
MEASUREMENTS, VOLUME AND WEIGHTS
Unit height578 mm (22 3/4 inches)
Unit width484 mm (19 inches)
Unit depth200 mm (8 inches)
Total weight (excluding ArjoClean Disinfectant/Cleaner)11 kg (24 1/2 lbs)
Total weight (including ArjoClean Disinfectant/Cleaner)14 kg (31 lbs)
CONNECTION DEMANDS
Inlet Coupling (Cold Water)15 mm (1/2 inches)
Overflow15 mm (1/2 inches)
Minimum Flow Rate - Inlet6 l/minute (1.6 US US gallon/minute)
Minimum Flow Rate -Outlet5 l/minute (1.3 US gallon/minute)
Minimum Dynamic Pressure2 bar (29 PSI)
Maximum Static Pressure6 bar (87 PSI)
ENVIRONMENTAL LIMITATIONS
Maximum ambient temperature40 ºC (104 ºF)
Maximum relative humidity (non-condensing)80%
Disinfectant Storage within unitminimum 0° C – maximum 30 °C
(minimum 32 °F – maximum 86 °F)
Environmental Conditions
Operating
Ambient temperature10 °C – 40 °C (50 °F – 104 °F) Operating
Relative humidity15-100% at 20 °C (68 °F) Operating
Atmospheric pressure800 hPa – 1060 hPa Operating
Transport and Storage Conditions
Ambient temperature-20 °C to 70 °C (-4 °F to 158 °F)
Relative humidity10% - 80% at 20 °C (68 °F)
Atmospheric pressure500-1100 hPa
End of Life Disposal
•Equipment that has electrical and electronic components should be disassembled and recycled
per Waste of Electrical and Electronic Equipment (WEEE) or in accordance with local or national
regulation.
•Disinfection liquids – if you have extra liquid, it’s safe to pour it down the drain with running water.
These products are water soluble and usually mixed with water when used for cleaning, your
municipality will treat these cleaners the same way they would if you used them for cleaning.
•The packaging containers should be recycled according to the national or local regulations.
19
Page 20
Labels
Label Explanation
Data LabelStates technical performance and
requirements, for example
classified safety standard and input
voltage.
Identification
Label
States the product identification;
serial number, year and month of
manufacturing.
Power Supply
Label
Water
Connection
Label
States IP-rate, fuse information
and wiring diagram.
States the minimum/maximum
temperature and pressure
requirement of inlet water.
Symbol Explanation
Read the IFU before use.
Separate electrical and electronic
components for recycling in
accordance with the European
Directive 2012/19/EU (WEEE)
Name and address of the
manufacturer.
Label for equipotential connection
point (bonding lug).
Classified by Underwriters
Laboratories Inc. with respect to
electric shock, fire, mechanical
hazards and other specified hazards
only in accordance with:
UL 2601-1 and CAN/CSA C 22.2 No 601.1-M90.
UL marking
(US-models 120 V– 60 Hz)
Water Regulations Advisory Scheme.
The product, when installed correctly,
complies with the requirements of
United Kingdom Water regulations/
Byelaws (Scotland)
CE marking indicating conformity
with European Community
harmonised legislation
Figures indicate Notified Body
supervision.
Indicates the product is a Medical
Device according to EU Medical
Device Regulation 2017/745
20
Page 21
Power Supply Label
Equipotential Bonding
Label
Water Connection
Label
(Pumped model only)
Identification Label
Data Label
21
Page 22
Parts and Accessories
6303097
Arjo Spray Handle
(Yellow)
PS1557
Oras Spray Handle
(White with Yellow
Shroud)
600.00.00.03
Hose 3 m
(Yellow)
PRB2000
ArjoClean Disinfectant/
Cleaner
PP0998
Hose 2 m
(Grey)
22
Page 23
AUSTRALIA
Arjo Australia Pty Ltd
78, Forsyth Street
O’Connor
AU-6163 Western Australia
Tel: +61 89337 4111
Free: +1 800 072 040
Fax: + 61 89337 9077
HONG KONG
Arjo Hong Kong Limited
Room 411-414, 4/F, Manhattan Centre,
8 Kwai Cheong Road, Kwai Chung, N.T.,
HONG KONG
Tel: +852 2960 7600
Fax: +852 2960 1711
ITALIA
Arjo Italia S.p.A.
Via Giacomo Peroni 400-402
IT-00131 ROMA
Tel: +39 (0) 6 87426211
Fax: +39 (0) 6 87426222
E-mail: [email protected]
MIDDLE EAST
Arjo Middle East FZ-LLC
Of ce 908, 9th Floor,
HQ Building,North Tower,
Dubai Science Park,
Al Barsha South
P.O Box 11488, Dubai,
United Arab Emirates
Direct +971 487 48053
Fax +971 487 48072
Email: [email protected]
NEDERLAND
Arjo BV
Biezenwei 21
4004 MB TIEL
Postbus 6116
4000 HC TIEL
Tel: +31 (0) 344 64 08 00
Fax: +31 (0) 344 64 08 85
E-mail: [email protected]
SUOMI
Arjo Scandinavia AB
Riihitontuntie 7 C
02200 Espoo
Finland
Puh: +358 9 6824 1260
E-mail: Asiakaspalvelu. [email protected]
SVERIGE
Arjo International HQ
Hans Michelsensgatan 10
SE-211 20 MALMÖ
Tel: +46 (0) 10 494 7760
Fax: +46 (0) 10 494 7761
E-mail: [email protected]
UNITED KINGDOM
Arjo UK and Ireland
Houghton Hall Park
Houghton Regis
UK-DUNSTABLE LU5 5XF
Tel: +44 (0) 1582 745 700
Fax: +44 (0) 1582 745 745
E-mail: [email protected]
USA
Arjo Inc.
2349 W Lake Street Suite 250
US-Addison, IL 60101
Tel: +1 630 307 2756
Free: +1 80 0 323 1245 Institutional
Free: +1 800 868 0441 Home Care
Fax: +1 630 307 6195
E-mail: [email protected]
At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges.
With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective
prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise