1. INTRODUCTION
The WILTRON 54000-6WR10 and -6WR15 are compact, int egrated, waveguide, millimeter (mm) wave, multiplied reflect ometers with an output fre­quency range o f 75 to 110 GHz and 50 to 75 GHz, respectively.
2. DESCRIPTION
The RF input (Figure 2) passes through an isola­tor and is then doubled. The doubled signal is then amplified, passed through a tuner and a low­pass filter and multiplied again (X2 fo r -6WR15 and X3 for -6WR10). At this point the 75 to 110 GHz or 50 to 75 GHz signal is passed exter­nally through user-selectable filters and then back into the unit. From there the signal passes through an isolator and dual-directional coupler to the Test Port. Detectors on the coupler’s coupled arms provide the reference and reflection outputs.
During the remainder of this manual, the 54000­6WR10 and -6WR15 will be referred to as Reflec­tometer.
490 JARVIS DRIVE MORGAN HILL, CA 95037-2809
Page 2
+7 dBm
OUTPUT
RF INPUT
REFLECTION
12.5-18.75 GHz
TEST PORT
Figure 2. WILTRON 54000-6WR15 mm Reflectometer
OUTPUT
DETECTOR
DETECTOR
REFERENCE
POWE R IN P UT
REFLECTOMETER
MILLIMETER WAVE MULTIPLIER
50 GHz to 75 GHz
3. POWER CONNECTION
The Reflectomet er requires DC power of +15V ±0.5V at 550 mA. The connector on the unit is a fe­male BNC “twinax” with +15V on the female cen- ter pin. A WILTRON P/N A36599 power cable is in­cluded to connect directly to below-listed WIL­TRON instruments that provide rear panel auxil­iary +15V power.
54xxA/SM4356
541xxA/Option 16
68xxxB (20 GHz)/SM4957
68xxxB (>20 GHz)/SM4954
An adapter (female twinax to dual banana plug), Wiltron P/N SM4816, is available on Special order. This adapter, in conjunction with cable A36599, will allow connection to any power supply that has banana terminals on 0.75-inch (19 mm) centers.
4. OPERATION AS A mm WAVE SOURCE
The following procedure describes how to operate the mm Wave Reflectometer as a st and alone source.
Step 1. Connect DC power to the Reflectometer
POWER INPUT connector as described in paragraph 3 above. (See Figure 3.)
Step 2. Connect an RF source to the Reflectome-
ter RF INPUT connector.
The level should be +7 dBm ±1 dB at the input co nnec tor.
+15 VDC
54000-6WR15
HARMONIC FILTER
Source freque ncy is 12.5–18.75 GHz for WR15 modules and 12.5–18.333 GHz for WR10 modules.
Step 3. The mm wave signal is available at the
TEST PORT.
Step 4. If leveling is desired, and t he source is
capable of external leveling, connect the Reflectometer REFERENCE DETECTOR OUTPUT to the appropriat e connector on the source. Refer to the Operation Manual for the source.
5. OP ER ATION IN S CA L AR m m WAVE MEASUREMENTS
The following procedure describes how to operate the mm wave r eflectometer in a scalar measure­ment syst em.
Step 1. Connect DC power to the Reflectometer
POWER INPUT connector as described in paragraph 3 above.
Step 2. Connect an RF source, 12.5–18.33 GHz
to the Reflectometer RF INPUT connec- tor.
The level should be +7 dBm ±1 dB at the input co nnec tor.
Source freque ncy is 12.5–18.75 GHz for WR15 modules and 12.5–18.333 GHz for WR10 modules.
Page 2 54000-6WRXX
Page 3
>
>
R
BA
DUT
REFLECTOMETER
MILLIMETER WAVE
TRANSMISSION
DETECTOR
+15V DC
WILTRON
54100A SERIES SCALAR
MEASUREMENT SYSTEM
TEST
PORT
Figure 3. WILTRON 54000-6WRXX mm Reflectometer
Step 3. If leveling is desired, and t he source is
capable of external leveling, connect the Reflectometer REFERENCE DETECTOR OUTPUT to the appropriat e connector on the source. Refer to the Operation Manual for the source.
Step 4. If ratioed Transmission measurements
are desired, and the Scalar Network Analyzer has an “R” Channel, connect the Reflectometer REFERENCE DETEC- TOR OUTPUT connector to the Analyzer R input using the 560-10BX-2 cable pro­vided.
NOTE
Leveling and ratioed measurements can­not be done simultane ously.
Step 5. Connect the reflectometer REFLEC-
TION DETECTOR OUTPUT to the Ana­lyzer Channel B input.
Step 6. Connect a WILTRON 54000-7WR10 or -
7WR15 detector to the TEST PORT and the Analyzer Channel A input as appro­priate. Perfo rm a Transmission/Reflec­tion Calibration on the Analyzer. Refer to the Operation Manual for the Ana­lyzer.
Step 7. Remove the detector from the Test Port
and insert the device to be tested.
6. SP EC IF ICATIONS
Performanc e specifications for the WILTRON 54000-MM Series Reflectometers are given in Tables 1 and 2.
7. ACCESSORIES
The following accessories are included with t he re­flectometer:
DC Power Cable P/N A36599
Detector Cable P/N 560-10BX-2 (2 ea)
Filter FL1 75–110 GHz or 50–75 GHz
Filter FL2 75–92 GHz or 50–58 GHz
Filter FL3 89–110 GHz or 57–75 GHz
W av eguide sec tion, WR10 or WR15, 1 inch (25
mm) Flush short
54000-6WRXX Page 3
Page 4
Table 1. Performance Specifications, 54000-6WR10 Module
DC POWER INPUT Current:
Voltage: Connector:
550 mA max
+15 VDC ±0.5V
BNC Twinax femal e connector, +15V on
female pin
RF INPUT Frequency Range:
Level: Connector: N Female
+7 dBm ±1 dB
12.5–18.33 GHz
NOTE
–6 0 dBc harmonics and spurs required to make output harmonic and spur specifica­tions with FL2 and FL3
TEST PORT Frequency Range:
Frequency Accuracy: Source Dependent, 6x Source Frequency Resolution: Source Dependent, 6x Source
Table 2. Performance Specifications, 54000-6WR15 Module
75–110 GHz
Power:( with Fl1 Installed): –5 dBm Minimum, +1 dBm
Typical, ±3 dB Variation Typical
Harmonics And Spurs:
FL1 (75–110 GHz): –20 dBc typical FL2 (75–92 GHz): –55 dBc max FL3 (89–110 GHz): –55 dBc max
Directivity: Source Match: 1.9:1 VSWR max Waveguide: WR-10, compatible with MIL-F-
35 dB min, 40 dB typical
3922/67B-010, UG-387/U flange
Modula t ion:
AM: Not Recommended FM: 6x source deviation Pulse:
DIMENSIONS
Length: 7.47 inches (190 mm) Width: 4.10 inches 104 mm) Heigth: 1.56 inches (40 mm)
OK
DC POWER INPUT Current:
Voltage: Connector:
550 mA max
+15 VDC ±0.5V
BNC Twinax femal e connector, +15V on
female pin
RF INPUT Frequency Range:
Level: Connector: N Female
+7 dBm ±1 dB
12.5–18.75 GHz
NOTE
–6 0 dBc harmonics and spurs required to make output harmonic and spur specifica­tions with FL2 and FL3
TEST PORT Frequency Range:
Frequency Accuracy: Source Dependent, 4x Source Frequency Resolution: Source Dependent, 4x Source
50–75 GHz
Power:( with Fl1 Installed): 0 dBm Minimum, +5 dBm
Typical, ±3 dB Variation Typical
Harmonics And Spurs:
FL1 (50–75 GHz): –20 dBc typical FL2 (50–58 GHz): –55 dBc max FL3 (57–75 GHz): –55 dBc max
Directivity: Source Match: 1.9:1 VSWR max Waveguide: WR-15, compatible with MIL-F-
35 dB min, 40 dB typical
3922/67B-008, UG-385/U flange
Modula t ion:
AM: Not Recommended
4x source deviation
FM: Pulse: OK
DIMENSIONS
Length: 7.47 inches (190 mm) Width 4.10 inches 104 mm) Heigth: 1.56 inches (40 mm)
Page 4 54000-6WRXX
Loading...