ANDERS ELECTRONICS DPM8100-2, DPM8100A-2, DPM8120-2, DPM8120A-2, DPM8140-2 Service Manual

...
digital panel modules user guide
DPM 8000 Series-2 Voltmeters and Ammeters
Models covered by this instruction sheet
DPM8100-2 3.5 DIGIT 200mV DC DPM8100A-2 3.5 DIGIT 2V DC DPM8120-2 3.5 DIGIT 200mV AC DPM8120A-2 3.5 DIGIT 2V AC DPM8140-2 3.5 DIGIT 200mV True RMS AC DPM8400-2 4.5 DIGIT 200mV DC DPM8400A-2 4.5 DIGIT 2V DC DPM8420-2 4.5 DIGIT 200mV AC DPM8420A-2 4.5 DIGIT 2V AC
The DPM8100-2 and DPM8400-2 series can be ranged as voltmeters or ammeters by fitting multiplier or shunt resistors (not include into the labelled positions on the main printed circuit board as shown on the diagram overleaf. Wire links"JP1" may need to be removed for some voltmeter ranges.
Using close tolerance resistors, the minimum of recalibration is required. For accurate setting a calibrated supply should be used. The resistors fitted should be adequately rated for the current or voltage range.
For non standard ranging and calibration, application sheet No. AES040 is available on request.
Over range is indicated by -1 on the display for 3.5 digit meters and by flashing zero's on the 4.5 digit meters.
STEP BY STEP INSTRUCTIONS
Remove instrument from its protective plastic bag and clip off
1
front cover.
Pull off the three green safety terminals from the rear of the
2
meter. The complete panel meter assembly can now be removed from its case by carefully levering the base of the LED display board over the plastic retaining lug while pushing gently on the terminal pins at the rear.
With reference to the Range Specification table and notes
3
overleaf, determine the resistance values to be fitted for the required voltage or current range. Further note whether link "JP1" should be removed. Carefully de-solder and remove link JP1 if necessary and solder the required resistors in place as shown on the main circuit board diagram.
The decimal point should be set as required by either
4
repositioning the PCB jumper link JP2 on the 3.5 digit models or setting a dip switch on the front of the 4.5 digit model, in one of these locations:­DP1 gives 1.999 i.e. . XXX DP2 gives 19.99 i.e. X. XX DP3 gives 199.9 i.e XX. X DP4 gives 1999.9 i.e XXX. X on 4.5 digit version only.
Slide panel meter back into its case along the guide rails
5
taking care to ease bottom of LED display board over protruding lug. Then push the green safety terminals securely back into place on the rear of the meter.
Make electrical connection to the meter with reference to the
6
Terminal Definition table. Note Mains Earth must NOT connected to pin 4. There is no earth connection required with this meter. Note also that Pin 3 is unused.
The instrument may now be tested.
7
For improved accuracy it can be adjusted by turning the "span adj" potentiometer whilst applying full scale signal from a calibrated source until exactly full scale is shown on the display. On the DPM8140 and DPM8400 only, first set the display to read zero for 0V input using the "offset" potentiometer on the front left hand side of the board.
Finally, clip the cover back on and insert the meter into the
8
panel cut-out. "Snap" the mounting clips into the side of the case (refer to installation diagram) and tighten the fixing screws until secure in the panel - do not over tighten!
be
OPERATING SPECIFICATION
Accuracy: 81xx: +/- 0.1% of rdg +/-1 digit (AC models +/- 0.2%)
Line voltage: 115/ 230 volts +10, -20%, 50 - 60 Hz
Temperature: 0 - 50 degrees C
Humidity: below 85% RH
Power: 4VA
Sampling rate: 2.5 / second
84xx: +/- 0.01% of rdg +/-2 digits DC models 84xx: +/- 0.1% of rdg +/-3 digits AC models
TERMINAL DEFINITIONS
SYMBOL DESCRIPTION
1
2
3
4
5
6
7
8
9
IN HI
IN LO
n/c Not connected (unused)
Gnd (sig)
HOLD
(display)
+5V (o/p) internal +5V output (maximum 50mA)
230V
115V AC mains supply
0V (N)
DIMENSIONS
INSTALLATION DIAGRAM
Apply input to pins 1 and 2. If INLO signal is greater than INHI the display will read -ive.
digital 0 Volts (NOT MAINS EARTH)
connect to terminal 6 (+5V) 8100 series connect to terminal 4 (0V) 8400 series
DPM8000-2 Rev 3 10/03
PCB LAYOUT
SPAN
ADJUST
VR1
A to D Converter Chip
LINK
R17
JP1
R18
OFFSET addition.
VR2
JP2
VR2
F1
1 2 3
Decimal point DIP switch on 4.5 digit models
Zero offset adjust (where fitted)
Decimal point select jumper on 3.5 digit models
Power supply input fuse 100mA rated (see Note 6)
NOTES
1
To measure voltages above the meters basic sensitivity, attenuator resistors R17 and R18 need to be installed. The general equation to determine their values is given by: V = V x (1 + R / R ) Where V is the input voltage required and V is the corresponding display voltage. Note, V max is actually 199.9mV for a nominal
IN D
D
200mV basic sensitivity meter and 1.999V for a nominal 2V meter
Examples
VIN VDDisplay
19.99V
20V
1500rpm
10V
For standard input ranges choose the corresponding resistors from the range specification table below.
2
For inputs greater than 200V, R17 must be a high voltage resistor such as Philips VR37 Series (1x2M2, 0.5W, RS164-8316) with R18=2k2 Alternatively, use two carbon film resistorsinseries,such as Multicomp MCF (2x1M2, 0.5W, Farnell 262-195) with R18 as 2K4. See the options for R17 below. The maximum input voltage on this range with correct resistors is 650V.
3
For ranges of 200V or less, for R17,use metal film resistors, 1%, 0.5W as Farnell 545-314. For basic 200mV use only existing link JP1
4
For R18, on all voltage ranges, use metal film resistors, 1%, 0.25W
5
On current ranges, usemetalfilmresistors, 0.5W for R30. The current, using internal shunt resistor R30 must not exceed 2 Amps. Leave link JP1 in place
IN D 17 18
17
R
1.999V
1.5V
1M
1M
R18 110K
176K
6
R30
123
N/c
In
InLo0vdcHold
Hi
!
Warning:
Make connection
no
from pin 4 to
under any
earth
conditions.
45 6
+5V
i/p
dc
mains
789
230
115
Vac
Vac0Vac
Important Observe the mains input connections
The power supply input protection fuse is a 20mm cartridge type rated at 250V / 100mA.
200mV and
current
R17
Link JP1
RANGE SPECIFICATION
MODEL NUMBER R17
RESOLUTION
(of range required)
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
8100A, 8120A 8400A, 8420A
8100A, 8120A 8400A, 8420A
8100A, 8120A 8400A, 8420A
8100A, 8120A 8400A, 8420A
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
8100, 8120, 8140 8400, 8420
100 Vm
10 Vm
1mV
100 Vm
10mV
1mV 1mV
100 Vm
10mV
1mV
100mV
10mV
1V
100mV
0.1 Am
10nA
1Am
0.1 Am 10 Am
1Am
100 Am
10 Am
1mA
100 Am
BASIC
SENSITIVITY
200mV
200mV
200mV
2V
2V
2V
2V
200mV
200mV
200mV
200mV
200mV
RANGE
REQUIRED
0.2V
2V OUT
20V 1M
2V
OVERLOAD
PROTECTION
100V
100V
300V
100V
100V
300V
650V
(see note 2)
700V
200 Am
DO NOT
EXCEED
DISSIPATION
RATING OF
SHUNT
RESISTOR
RC
(SEE NOTE 5)
anders electronics plc Bayham Place, London NW1 OEU UK
Tel: +44 (0)20 7380 8167 Fax: +44 (0)20 7874 1908 http://www.anders.co.uk
Options for Resistor R17
Ranges below
200V
R17 R17
Metal film
resistor
R18 R30 LINK
WW
110K
1M
10K
110K20V
1M
2M2
10K200V
2K2
1M
1K
100R2mA IN
10R20mA IN
1R200mA IN
0R12A IN
Ranges above 200V
R17
Single Hi
Voltage resistor
W
JP1
IN
OUT
IN
OUT
OUT
OUT
IN
Solder
Two carbon
film resistors
INPUT
IMPEDANCE
>I00M
1M
IM
>I00M
IM
IM
2M2
Ik
IOOR
IOR
IR
R1 (0.1)
DP8000-2 Rev3 10/03