Amprobe IRC-120 User manual

Page 1
IRC-120 IRC-120-CN
Thermal Camera with UV and Flashlight
User Manual
5/2018, 6010574 A ©2018 Amprobe All rights reserved. Printed in Taiwan
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for one year from the date of purchase unless local laws require otherwise. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on the behalf of Amprobe. To obtain service during the warranty period, return the product with proof of purchase to an authorized Amprobe Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES
- WHETHER EXPRESS, IMPLIED OR STATUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All Amprobe returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the meter. Non­warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Amprobe.
In-warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty period, any defective test tool can be returned to your Amprobe distributor for an exchange for the same or like product. Please check
**single contact address in EEA Fluke Europe BV
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Thermal Camera with UV and Flashlight
CONTENTS
SYMBOLS ............................................................................................3
SAFETY INFORMATION ......................................................................3
UNPACKING AND INSPECTION ..........................................................5
FEATURES AND APPLICATIONS ..........................................................5
BASIC NAVIGATION ............................................................................6
MEASUREMENT MODE ......................................................................6
MENU MODE ......................................................................................8
Emissivity .........................................................................................8
UV/Flashlight ...................................................................................9
Laser Sighting .................................................................................9
Memory/SD Card .............................................................................10
Hot and Cold Markers ....................................................................10
Center Point Marker .......................................................................11
Color Palette ...................................................................................11
°F / °C ...............................................................................................11
Auto Power OFF ..............................................................................12
Date & Time ....................................................................................12
SPECIFICATIONS ..................................................................................13
MAINTENANCE AND BATTERY REPLACEMENT ................................14
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Thermal Camera with UV and Flashlight
4
5
7
6
1
LC Display
2
Menu button / Flashlight
3
Battery cover
4
SD card cover
5
Micro SD card slot
6
Micro USB connector
7
Tripod mount
13
12 11 10
1
2
3
14
9
8
15
8
Trigger for laser pointer and
image capture
9
Power ON/OFF (Rotary switch)
10
UV light blue LEDs
11
Flashlight LEDs
12
Visual camera
13
Laser aperture
14
Laser sight (Circle/dot/center point)
15
Infrared lens
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
SYMBOLS
W
P
Caution! Refer to the explanation in this manual.
Consult user documentation.
WARNING! LASER RADIATION. Risk of eye damage.
Complies with European Directives.
Conforms to relevant South Korean EMC Standards.
Conforms to relevant Australian standards.
Battery.
Indicates a Class 2 laser. DO NOT STARE INTO BEAM. The following may appear with the symbol on the product label: ”IEC/EN 60825-1:2014. Complies with 21 CFR 1040.10 and
1040.11 except for deviations pursuant to Laser Notice 50. Dated June 24, 2007.” In addition, the following pattern on the label will indicate wavelength and optical power: λ = xxx nm, x.xx mW.
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types
=
in the WEEE Directive Annex I, this product is classed as category 9 “Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste.
SAFETY INFORMATION
A Warning identifies hazardous conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the product or the equipment under test.
Warning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety information before you use the product.
• Carefully read all instructions.
• Use the product only as specified, or the protection supplied by the product can be compromised.
• Do not use the product around explosive gas, vapor, or in damp or wet environments.
• Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
• Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
• Use the product only as specified, or hazardous laser radiation exposure can occur.
• Do not use laser viewing glasses as laser protection glasses. Laser viewing glasses are used only for better visibility of the laser in bright light.
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
• Do not open the product for reasons other than replacing batteries. The laser beam is dangerous to eyes. Have the product repaired only through an approved technical site.
• Verify the product’s operation by measuring on a known temperature source. Do not use the product if it operates incorrectly or abnormally.
• Do not use the product if it is damaged.
• See emissivity information for actual temperatures. Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard.
• Remove the batteries if the product is not used for an extended period of time, or if stored in temperatures above 122 °F (50 °C). If the batteries are not removed, battery leakage can damage the product.
• Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
• Use only AA alkaline batteries and follow all battery care from the manufacturer.
• Do not leave the product on or near objects of high temperature.
• For use by competent persons only.
• Have an approved technician repair the product.
W Caution
To prevent personal injury and for safe operation and maintenance of the product:
• Batteries contain hazardous chemicals that can cause burns or explode. If exposure to chemicals occurs, clean with water and get medical aid.
• Be sure that the battery polarity is correct to prevent battery leakage.
• Do not short the battery terminals together.
• Keep batteries clean and dry.
To avoid damaging the product under measurement, protect it from the following:
• EMF (electro-magnetic fields) from arc welders and induction heaters.
• Static electricity.
• Thermal shock (caused by large or abrupt ambient temperature changes — allow 30 minutes for the product to stabilize before use).
• Do not leave the product on or near objects of high temperature.
W
λ
= 650 nm, < 1 mW
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Laser aperture
Page 7
UNPACKING AND INSPECTION
Your package should include: 1 IRC-120 or IRC-120-CN Thermal Camera 3 1.5 V AA batteries 1 2 G micro SD card (installed) 1 Standard SD card adapter 1 User manual 1 Wrist strap If any of the items are damaged or missing, return the complete package to
the place of purchase for an exchange. Note: Batteries do not come pre-installed. Please refer to the Maintenance
and Battery Replacement section for further instruction. The product will arrive with a thin plastic screen protector over the display.
Please remove before using the product.
FEATURES AND APPLICATIONS Features
• Five-in-one: UV light, flashlight, infrared temperature, thermal image and laser pointer
• Five UV LEDs detect leak detection dyes*
• Laser pointer to indicate the center of UV field
• Bright flashlight to illuminate dark areas
• Memory capability to save thermal images
• Infrared heat map image blending at 0%, 25%, 50%, 75%, and 100%
• Three selectable color palettes (grayscale, hot iron and rainbow)
• Center-point temperature measurement
• Focus free
• IR measurement 20:1 Distance to Spot ratio
• Circle/dot/center point laser sighting
• Adjustable emissivity from 0.10 to 1.00
• Auto off function
• Selectable °F and °C
• Intuitive joystick navigation to on-screen menu and settings
• Hot and cold markers instantly identify hottest and coldest spots
• Tripod mount
* Standard leak detection dyes are not sold by Amprobe.
Applications
• Highlights leak detection dyes with UV light
• Laser sighting helps locate the target area of measurement
• Illuminate dark areas with the flashlight
• Find areas of heat loss and drift
• Locate electrical problems at appliances, machines and installations
• Quickly verify HVAC/R functionality and performance
• Identify temperature related issue for electrical connections and motors
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
BASIC NAVIGATION
Menu Return to measurement
Image blending (measurement mode)
Setting selection (menu mode)
Temperature
(selectable °F and
°C unit)
Hot / Cold marker
Flashlight
(or UV light: UV)
Laser sighting
SD card Battery level
Center point marker
Emissivity
MEASUREMENT MODE Taking Measurements Power ON / OFF
POWER OFF
Figure 1.1
Rotate the wheel around the lens to the left to turn the product on. To turn it off, rotate right.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Image Capture/Laser Pointer
Figure 1.2
The product will be in measurement mode upon power on. Press the trigger to capture an image and to use the laser pointer to help aim on the target surface (laser pointer must be enabled in the menu, see Figure 2.3). Press navigation button to the left
Note: The laser is used for aiming purposes only and is not related to temperature measurement.
Hold the product 12 inches (30 cm) from the target (>59 (6 cm) from the target (<59 OF / 15 OC) for best temperature accuracy.
Note: Higher Distance to Spot ratios require smaller target areas to make accurate measurements. The Distance to Spot ratio of this product is 20:1.
to save the image or press to the right to cancel .
O
F / 15 OC) and 2.4 inches
Image Blending
0% 25% 50% 75% 100%
UP UP UP UP UP
Figure 1.3
Blending may be adjusted for an easier interpretation between infrared and visible images. To change the level of blending between 0 - 100%, press
down or up on the power button while in measurement mode.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
MENU MODE Changing Settings in the Menu
Once the product is powered on, reach the menu by clicking left on the power button. Settings that can be manipulated include emissivity, flashlight, UV light, laser sight, hot and cold markers, center point marker, color palette, °F/°C, timed auto power off, and date & time. Enabled options will appear as blue squares while disabled options will
appear as
black squares.
Emissivity
MENU BACK TO
ENTER
The emissivity of the surface of a material describes its effectiveness in emitting energy as thermal radiation. Quantitatively, emissivity is the ratio of the thermal radiation from a surface to the radiation from an ideal black surface at the same temperature as given by the Stefan–Boltzmann law. Refer to Table 1 for the emissivity adjustment.
In measurement mode, emissivity level is shown on the bottom right of the screen. Emissivity can be changed to values between 0.10-1.00 in menu.
Note: Surfaces with an emissivity <0.60 make reliable and consistent determination of actual temperatures problematic. The lower the emissivity, the more potential error is associated with the temperature measurement calculations of the product, even when emissivity and reflected background adjustments are attempted and performed properly.
Table 1. Nominal surface emissivity for an accurate non-contact infrared temperature measurement.
Material Value Material Value
Default**** 0.95 Glass (plate) 0.85 Aluminum* 0.30 Iron* 0.70 Asbestos 0.95 Lead* 0.50 Asphalt 0.95 Oil 0.94 Brass* 0.50 Paint 0.93 Ceramic 0.95 Plastic** 0.95 Concrete 0.95 Rubber 0.95 Copper* 0.60 Sand 0.90 Food-frozen 0.90 Steel* 0.80 Food-hot 0.93 Water 0.93
*Oxidized **Opaque, over 20 mils ***Natural ****Factory Setting
DOWN UP
Figure 2.1
Wood*** 0.94
8
MEASUREMENT
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
UV/Flashlight
MENU BACK TO
DOWN UP ENTER
Figure 2.2
The flashlight can be turned on the the menu or by holding down the navigator button >2 seconds.
The UV light will illuminate fluorescence in substances and on surfaces. Use it in combination with UV dyes to inspect automotive and HVAC systems for
leaks.
ENTER
MEASUREMENT
Laser Sighting
MENU BACK TO
DOWN UP ENTER
ENTER
MEASUREMENT
Laser sighting provides an accurate visual target for temperature measurement.
*See SAFETY INFORMATION on pages 3 to 4 when using the laser. Once laser sighting is enabled on menu, return to measurement mode by
pressing the navigator button to the right, and press the trigger to activate laser sighting (see page 7: Image Capture/Laser Pointer).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 2.3
9
Page 12
Memory/SD Card
MENU BACK TO
DOWN UP DOWN UPENTER
To view saved images, navigate to the SD card icon in the menu. Most recent images will appear first. Browse images by toggling up or down.
Note: Images cannot be deleted or renamed through the IRC-120 interface. Make these modifications by installing the SD card in an alternate device such as a computer.
The SD card comes installed in the product and will name saved photos in numerical order from 0001-9999. Access the SD card through the movable flap on the side of the product. Gently push on and release the SD card to pop it out. Insert it into the SD card adapter and then into a computer to download saved images. Saved images can also be downloaded with a micro B USB cable (not supplied). Access the USB port below the SD card. The product does not need to be powered on to download pictures through the USB connection.
If you receive this error ( memory is full. If the images are renamed, the SD card may be able to store
more than 9,999 images.
Figure 2.4
), either the SD card is not inserted or SD card
MEASUREMENT
Hot and Cold Markers
MENU BACK TO
MEASUREMENT
DOWN UP ENTER
Hot (red square outline) and cold (blue square outline) markers indicate the locations of hottest and coldest areas within a target area.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 2.5
10
ENTER
Page 13
Center Point Marker
MENU BACK TO
MEASUREMENT
DOWN UP ENTER
Figure 2.6
The center point marker indicates the center of the infrared thermometer measurement and appears in the center of the measurement screen.
ENTER
Color Palette
Hot Iron
MENU BACK TO
DOWN UP ENTER
Figure 2.7
The color palette can be changed depending on the specific application of the measurement. This option allows for different viewings of an infrared image by changing the false-color presentation. Grayscale shows a more equal and linear presentation of colors to enhance detail, while Hot Iron and Rainbow combine high contrast with Grayscale.
GrayscaleRainbow
MEASUREMENT
ENTER
°F / °C
MENU BACK TO
MEASUREMENT
DOWN UP ENTER
Figure 2.8
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
ENTER
Page 14
Auto Power OFF
MENU BACK TO
MEASUREMENT
ENTER
By default, the product will automatically power off after 1 minute. To choose when the product will power off after a period of inactivity, change the settings in the menu.
Available options for the Auto Power Off setting:
• OFF (the camera will not turn off automatically)
• 1 m (1 minute)
• 2 m (2 minutes)
• 5 m (5 minutes)
• 10 m (10 minutes)
DOWN UP
Figure 2.9
Date & Time
MENU BACK TO
DOWN UP ENTER
ENTER
MEASUREMENT
Press the navigator button to edit the date or time setting, toggling up and down until the correct number is selected. Press to the right to exit the setting and go back to measurement mode.
Date format: YYYY/MM/DD Time format: AM/PM and 24h selectable
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 2.10
12
Page 15
SPECIFICATIONS
UV light
Flashlight
Laser sighting
Temperature measurement
Temperature range
IR accuracy (calibration geometry with ambient temperature 23°C ± 2°C)
Display resolution
IR Repeatability
Temperature Coefficient
Distance to spot
Minimum spot size
Response time (95 %)
Spectral response
Emissivity
Visual image with infrared heat map overlay
Visual image resolution
IR detector resolution
Field of view
Thermal sensitivity
Focus system
Image palettes
Hot and cold marker
Center point marker
Display
Data storage
Operating temperature and humidity
Storage temperature
5 blue LEDs
4 LEDs
Circle/dot/center point laser Output <1 mW, wavelength 650 nm
Yes, center point
14 °F to 932 °F (-10 °C to 500 °C)
≥ 32 °F (≥ 0 °C): ± 4 °F (± 2 °C) or ± 2 % of the reading, whichever is greater < 32 °F (< 0 °C): ± 6 °F (± 3 °C)
0.2 °F / 0.1 °C
± 0.8 % of the reading or ± 2 °F (± 1 °C), whichever is greater
0.1 °C/°C or ± 0.1 %/°C of the reading, whichever is greater
20:1
8 mm
< 125 ms
8 μm to 14 μm
Digitally adjustable from 0.10 to 1.00 by 0.01
Five blending modes (0%, 25%, 50%, 75% and 100%)
16,384 pixels (128 x 128 pixels) (Interpolation pixels)
32 x 32 pixels
33 ° x 33 °
150 mK
Focus free
Grayscale (white hot), Hot Iron and Rainbow
Yes
Yes
1.77 in color TFT with 128 x 160 pixels
Stored image size: 124 x 160 pixels Image file size: typical 40 KB Maximum stored images: 9999
32 °F to 122 °F (0 °C to 50 °C) 10 % to 90 % RH non-condensing at 86 °F (30 °C)
-4 °F to 140 °F (-20 °C to 60 °C) without batteries
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
Visual to IR effective image alignment
Laser sighting to center of visual image
Laser sighting to UV field
Operating and storage altitude
Drop proof
Vibration and shock
Power supply
Battery life
Auto power off
Agency approvals
Laser safety compliance
Electromagnetic Compatibility
Size (H x W x L)
Weight
≥ 18 inches (45 cm), optimal for 1 m
≥ 18 in (45 cm) typical
Approx. 18 in (45 cm) typical
< 6561 ft (< 2000 m)
4 ft (1.2 m)
IEC 60068-2-6, 2.5g, 10 to 200 Hz, IEC 60068-2­27, 50g 11ms
Three (3) 1.5 V AA IEC LR6 alkaline batteries
8 hours with display ON (Typical) Power consumption: 220 mA (Typical)
Selectable modes: OFF, 1 minute, 2 minutes, 5 minutes and 10 minutes
P
IEC 60825-1, Class 2
EN 61326-1 Korea (KCC): Class A Equipment (Industrial Broadcasting & Communication Equipment)
[1]
This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
Approx. 7.3 x 2.1 x 4.1 in (185 x 54 x 104 mm)
Approx. 0.64 lb (0.29 kg)
[1]
MAINTENANCE AND BATTERY REPLACEMENT
Do not use abrasives, isopropyl alcohol, or solvents to clean the case or lens/ window. If used and stored properly, the infrared lens on the product should require only occasional cleaning (Figure 3.1).
When necessary, to clean the lens:
1. Use a hand-squeeze air pump to gently blow off any dust or debris from the lens surface.
2. If the lens surface requires additional cleaning, use a clean, fine-fiber cloth, micro-fiber cloth, or cotton swab dampened with a mild, soapy water solution. Gently wipe surface of lens to remove smudges and debris.
3. Dry with an absorbent, clean fine-fiber or micro-fiber cloth.
Note: Minor smudges and dirt should not significantly affect the performance of the product. However, large scratches or the removal of the protective coating on the infrared lens can affect both image quality and the temperature measurement accuracy.
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
To avoid incorrect measurements, replace the batteries when the battery indicator is low. Use only AA alkaline batteries and follow all battery care guidelines from the manufacturer. To remove the batteries, first ensure the product is in OFF position, then unscrew the battery cover. When installing the batteries, be sure that the battery polarity is correct to prevent battery leakage. Secure the batteries by screwing the battery cover back in place (Figure 3.2).
Clean
Battery Replacement
3 x AA batteries
Figure 3.1
Figure 3.2
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...