Amprobe CO2-200 Specifications

Data Sheet
CO2-200
CO2 Wall-Mounted Meter
Instant feedback with CO2 levels, with additional temperature and humidity measurement s. Built in alarm to warn against dangerous CO2 levels. The permanent mounted CO2-200 Air Quality meter is designed to measure air quality compliance in critical environment controlled areas such as laboratories and production.
n Low drift wall mounted meter for monitoring indoor air quality
n CO2 Measurement up to 9999 ppm, Temperature, Relative humidity, Dew Point and Wet Bulb
n Super-Large LCD display with CO2 reading
n High performance low drift sensor designed with NDIR
(Non-Dispersive Infrared) waveguide technology with automatic background calibration
No hassle warranty
No waiting.
No shipping charges.
Our commitment to high-quality products and customer service is demonstrated by our industry exclusive “No Hassle” warranty. In the unlikely event that an Amprobe Test Tool requires warranty service, any of our local dealers are authorized to replace it, on the spot.
(note: $500 MSLP limit)
n Audible alarm threshold setting to setup warning against carbon dioxide level with output drive relay to control air flow system
n Optional PC download capabilities (USB interface)
n Data Hold
n Recording of Maximum, Minimum and Average values with
recall function
Amprobe® Test Tools
www.Amprobe.com
CO2-200 CO2 Wall-Mounted Meter
Specifications
Measuring Range CO2 0 ~ 9999 ppm 2001 ~ 9999 ppm (out of scale range) Resolution 1ppm - CO2 concentration, 0.1°C / °F - Temp, 0.1% RH - Humidity Accuracy ±75ppm±5% of reading (0 ~ 2000 ppm) Others range accuracy are not specified Temperature ±0.6°C / ±0.9°F Temp. Range -10°C to 60°C (14°F to 140°F) RH Range 0% ~ 95% Accuracy ±3% RH (at25°C, 10-90%) others ±5% for other RH range DP (Dew Point Temp.) -20°C to 59.9°C (- 4°F to 139.8°F) WB (Wet Bulb Temp.) -5.0°C to 59.9°C (23°F to 139.8°F) Warm-up Time 30 seconds Response Time About 30 seconds Current Consumption < 50mA average, < 310mA peak current ( ~55ms) Operating Condition 0 ~ 50°C (32°F to 122°F), 0 ~ 95% RH (avoid condensation) for CO2 Dimension 130(L) x 85(W) x 60(H) mm (5.1 x 3.3 x 2.4 in) Storage condition -20°C to 60°C (-4°F to 140°F), 0~99%RH (avoid condensation) Power Supply 12 VDC adaptor Standard packing includes Meter, Manual, Power adapter 100-240vac Weight Approx. 193g (0.43 lb)
Data Sheet
Amprobe® Test Tools
website: www.Amprobe.com email: info@amprobe.com Everett, WA 98203 Tel: 877-AMPROBE
Amprobe® Test Tools Europe
In den Engematten 14 79286 Glottertal, Germany Tel.: +49 (0) 7684 8009 - 0
Amprobe® Test Tools
©2009 Amprob e Test Tools. All rights re serve d.
7/2009 3520597 Rev A
www.Amprobe.com