Data Sheet
ACD-15 TRMS-PRO
2000A Clamp-on Multimeter
Wide range of measuring features built into one, professional
meter. The TRMS version with backlight display, improves
performance and reliability.
True RMS
■
Backlight ■
Measurements: AC/DC Voltage ■
up to 600V, AC Current up to
2000A Resistance, Frequency
and Capacitance
Non-contact Voltage Level ■
Detection
Audible continuity ■
Auto-check feature ■
automatically selects DCV,
ACV or Resistance (Ω)
Auto power off
■
Low battery indication ■
Data hold ■
Large, easy to read LCD display ■
Accommodates conductors up to ■
45mm (1.77”) in diameter
Carrying case, test leads, batteries ■
(installed) and manual included
Voltage overload protection for all ■
functions up to 600V AC/DC
No hassle
warranty
No waiting.
No shipping
charges.
Our commitment to
high-quality products
and customer service
is demonstrated by
our industry exclusive
“No Hassle” warranty.
In the unlikely event
that an Amprobe Test
Tool requires warranty
service, any of our
local dealers are
authorized to replace
it, on the spot.
(note: $500 MSLP limit)
Amprobe® Test Tools
www.Amprobe.com
ACD-15 TRMS-PRO 2000A Digital Clamp-on Multimeters
Data Sheet
Electrical Specifications
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23℃ ±5℃ & less than 75% R.H. True RMS Model ACD-15
TRMS-PRO ACV & ACA clamp-on accuracies are specified from 5% to 100% of range or otherwise specified. Maximum Crest Factor are as
specified below, and with frequency spectrums, besides fundamentals, fall within the meter specified AC bandwidth for non-sinusoidal
waveforms.
Function Range Accuracy
DC Voltage
6.000V 0.5% + 3d
60.00V 1.0% + 5d
600.0V 2.0% + 5d
NMRR : >30dB @ 50/60Hz
CMRR : >100dB @ DC, 50/60Hz, Rs=1kΩ
Hi-Z DCV Input Impedance: 5MΩ, 90pF nominal
AutoCheck™ Lo-Z DCV input impedance: Initially 1.6kΩ, 90pF nominal; Impedance increases significantly as display voltage
increases from 50V (typical). Typical impedances vs display voltages for reference are:
15kΩ @ 100V; 100kΩ @ 300V; 210kΩ @ 600V
AutoCheck™ DCV Threshold: > +1.5VDC or < -1.0VDC nominal
AC Voltage
50Hz / 60Hz 6.000V, 60.00V 1.5% + 5d
600.0V 2.0% + 5d
50Hz ~ 500Hz 6.000V, 60.00V 2.0% + 5d
600.0V 2.5% + 5d
CMRR: >60dB @ DC to 60Hz, Rs=1kΩ
Hi-Z ACV Input Impedance: 5MΩ, 90pF nominal
AutoCheck™ Lo-Z ACV input impedance: Initially 1.6kΩ, 90pF nominal; Impedance increases significantly as display voltage
increases from 50V (typical). Typical impedances vs display voltages for reference are:
5kΩ @ 100V; 100kΩ @ 300V; 210kΩ @ 600V
AutoCheck™ ACV Threshold: > 2VAC (50/60Hz) nominal. True RMS model ACD-15 TRMS-PRO Crest < 1.6 : 1 at full scale
& < 3.3 : 1 at half scale
Resistance
6.000kΩ
60.00kΩ, 600.0kΩ 1.0% + 4d
6.000MΩ 2.0% + 4d
Open Circuit Voltage: 0.4VDC typical
600Ω with Continuity Beeper 600.0Ω 2.0%+8d
Continuity Beeper Response: < 100µs
Open Circuit Voltage: 0.4VDC typical
Audible Threshold: between 10Ω and 300Ω 1) Add 40d to specified accuracy while reading is below
20% of range
Frequency
Range Sensitivity (Sine RMS) Voltage Range
10Hz ~ 30kHz 4V 6.000V
0Hz ~ 1kHz 30V 1 60.00V
10Hz ~ 1kHz 60V 600.0V
Accuracy: 0.5%+4d
Max display: 9999 counts
1)
2)
3)
2)
1.2% + 6d
Cool down interval 2 minutes after over 50V measurements in Auto-VΩ position
Beeper on while reading < 0.025kΩ
Add 40d to specified accuracy while reading is
1) 3)
1)
Amprobe® Test Tools
www.Amprobe.com