Data Sheet
ACD-10 PLUS
600A Clamp-On Multimeter
Amprobe’s ACD-10 PLUS meter offers thinner jaws over standard clamp meters. Allowing access to tight measurement areas and still accommodating conductors up to 25 mm. It also measures Capaci­tance and Frequency. Frequency is measured by either jaws or test leads. Very versatile clamp multimeter.
AC & DC Voltage to 600V AC Current to 600A Thin Jaws, only 10mm (0.4”) thick Resistance to 40M W Continuity Buzzer
Capacitance to 3000 uF Frequency measurement In rush current Hold & Maximum reading functions Accommodates conductors up to 26mm (1.02”) in diameter Auto ranging Auto power off Rubber over-molded case Overload protected
(note: $500 MSLP limit)
Amprobe® Test Tools
www.Amprobe.com
ACD-10 PLUS 600A Clamp-On Multimeter
Data Sheet
General Specifications
Display: 3-3/4 digits 4000 counts LCD display Update Rate: 3 per second nominal Polarity: Automatic Operating Temperature: 0 °C to 40 °C; < 80% RH for temperature up to 31 °C decreasing linearly to 50% RH at 40 °C Altitude: Operating below 2000m; Indoor use Storage Temperature: -20 °C to 60 °C, < 80% RH (with battery removed) Temperature Coefficient: nominal 0.15 x (specified accuracy)/°C @ (0 °C ~ 18 °C or 28 °C ~ 40 °C) Low Battery: Below approx. 2.4V Power Supply: 3V coin battery IEC-CR2032 Power Consumption: 2.8 mA typical except that 3.3 mA typical for ACA function APO Timing: Idle for 30 minutes APO Consumption: 5 μA typical Dimension: 190 x 63 x 32 mm (7.4 x 2.5 x 1.3 in) Weight: 139 gm approx Jaw opening & Conductor diameter: max 26 mm Special Features : 30ms Max Hold; Data Hold; Relative Zero mode Safety: Meets EN61010-2-032, UL61010B-2-032, IEC61010-1 2nd Ed., EN61010-1
2nd Ed., UL61010-1 2nd Ed. CAT III-600 Volts ac & dc; Pollution degree : 2 EMC: Conforms to EN61326-1. This product complies with requirements of the following European Community Directives: 89 / 336/ EEC (Electromagnetic Compatibility)
and 73/ 23/ EEC (Low Voltage) as amended by 93/ 68/ EEC (CE Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit. Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit. Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference.
Electrical Specification Accuracy (23 °C ± 5 °C & < 75% R.H.)
Function Range Accuracy
DC Voltage
400.0 mV ±( 0.3% + 4 digits)
4.000, 40.00, 400.0 V ±( 0.5% + 3 digits) 600 V ±( 1.0% + 4 digits) NMRR: >50 dB @ 50/60Hz CMRR: >120 dB @ DC, 50/60 Hz, Rs =1 kW Input Impedance: 10 MW, 30 pF nominal (1000 MW for 400.0 mV range) Transient protection: 6.5 kV (1.2/50 μs surge)
AC Voltage (50Hz ~ 500Hz)
4.000, 40.00, 400.0 V ±( 1.5% + 5 digits) 600 V ±( 2.0% + 5 digits) CMRR: >60dB @ DC to 60 Hz, Rs=1 kW Maximum Crest Factor: < 1.75 : 1 at full scale & < 3.5 : 1 at half scale limited to
fundamental and harmonics, that fall within the meter specified AC bandwidth for non-sinusoidal waveforms
Input Impedance: 10 MW, 30 pF nominal Transient protection: 6.5 kV (1.2/50μs surge) ACD-10 Plus: Average Sensing
Amprobe® Test Tools
www.Amprobe.com