AMETEK SL User Manual

Elgar Electronics Corporation 9250 Brown Deer Road San Diego, CA 92121-2294
1-800-73ELGAR (1-800-733-5427) Tel: (858) 450-0085 Fax: (858) 458-0267
Email: sales@elgar.com
This document contains information proprietary to Elgar Electronics Corporation. The information contained herein is not to be duplicated or transferred in any manner without prior written permission from Elgar Electronics Corporation.
SLM DC MODULAR/
SLH H
E
©2006 by Elgar Electronics Corporation
SL-SERIES
IGH POWER DC
Calibration Manual
www.elgar.com
July 31, 2006 Document No. M540070-04 Rev A
CONTENTS
SECTION 1 OPERATION DESCRIPTION 1-1
1.1 NAVIGATION AND SETTING 1-2
SECTION 2 CALIBRATION 2-1
2.1 CALIBRATION SETUP 2-1
2.1.1 Equipment Required ......................................................................................... 2-1
2.2 DVM CALIBRATION 2-1
2.2.1 DVM, Range I Offset and Gain, Items 1 and 2.................................................. 2-2
2.2.2 DVM, Range II Offset and Gain, Items 3 and 4................................................. 2-3
2.3 VSENSE CALIBRATION 2-4
2.3.1 VSense, Range I Offset and Gain, Items 5 and 6 ............................................. 2-4
2.3.2 VSense, Range II Offset and Gain, Items 7 and 8 ............................................ 2-5
2.4 DAM AND CC MODE CALIBRATIONS 2-7
2.4.1 DAM, Range I Offset and Gain, Items 9 and 10................................................ 2-7
2.4.2 DAM, Range II Offset and Gain, Items 11 and 12............................................. 2-9
2.4.3 CC Mode, Level Low, Range I Offset and Gain, Items 13 and 14 .......... 2-10
2.4.4 CC Mode, Level High, Range I Offset and Gain, Items 15 and 16 ......... 2-12
2.4.5 CC Mode, Level Low, Range II Offset and Gain, Items 17 and 18......... 2-13
2.4.6 CC Mode, Level High, Range II Offset and Gain, Items 19 and 20 ........ 2-15
2.5 CR MODE CALIBRATION 2-16
2.5.1 CR Mode, Level Low, Range I Offset and Gain, Items 21 and 22 .......... 2-17
2.5.2 CR Mode, Level High, Range I Offset and Gain Items 23 and 24........... 2-19
2.5.3 CR Mode, Level Low, Range II Offset and Gain, Items 25 and 26........ 2-21
2.5.4 CR Mode, Level High, Range II Offset and Gain, Items 27 and 28 ........ 2-23
SLD Calibration Manual i
2.6 CV MODE CALIBRATION 2-25
2.6.1 CV Mode, Level Low, Offset and Gain, Items 29 and 30 ................................ 2-26
2.6.2 CV Mode, Level High, Offset and Gain, Items 31 and 32........................ 2-28
2.7 CP MODE CALIBRATION 2-29
2.7.1 CP Mode, Level Low, Offset and Gain, Items 33 and 34 ................................ 2-30
2.7.2 CP Mode, Level High, Offset and Gain, Items 35 and 36................................ 2-32
2.8 STORE CALIBRATION DATA 2-33
SECTION 3 CALIBRATION DATA RECORD 3-1
LIST of FIGURES
Figure 1-1. SLM Series Front Panel .....................................................................................................1-1
Figure 2-1. Connections for DVM Calibration Items 1-4)......................................................................2-1
Figure 2-2. Connections for Vsense Calibration (Items 5-8). ................................................................2-4
Figure 2-3. Connections for DAM and CC Mode Calibrations..............................................................2-7
Figure 2-4. Connections for CR Mode Calibration..............................................................................2-16
Figure 2-5. CV Mode Calibration Connections ...................................................................................2-25
Figure 2-6. Connections for CP Mode Calibration..............................................................................2-29
LIST of TABLES
Table 2-1. DVM, Range I Offset Adjustment Settings (Item 1).............................................................2-2
Table 2-2. DVM Range I Gain Adjustment Settings (Item 2)................................................................2-2
Table 2-3. DVM, Range II Gain Adjustment Settings (Item 3)..............................................................2-3
Table 2-4. DVM, Range II Gain Adjustment Settings (Item 4)..............................................................2-3
Table 2-5. VSense, Range I Offset Adjustment Settings (Item 5).........................................................2-4
Table 2-6. VSENSE, Range I Gain Adjustment Settings (Item 6).........................................................2-5
Table 2-7. VSENSE, Range II Offset Adjustment Settings (Item 7)......................................................2-5
Table 2-8. VSense, Range II Gain Adjustment Settings (Item 8)..........................................................2-6
Table 2-9. DAM, Range I Offset Adjustment Settings (Item 9).............................................................2-7
Table 2-10. DAM, Range I Gain Adjustment Settings (Item 10)..........................................................2-8
Table 2-11. DAM, Range II Offset Adjustment Settings (Item 11).......................................................2-9
Table 2-12. DAM, Range II Gain Adjustment Settings (Item 12)........................................................2-10
Table 2-13. CC Mode, Level Low, Range I Offset Adjustment Settings (Item 13)..............................2-10
Table 2-14. CC Mode, Level Low, Range I Gain Adjustment Settings (Item 14)................................2-11
Table 2-15. CC Mode, Level High, Range I Offset Adjustment Settings (Item 15).............................2-12
Table 2-16. CC Mode, Level High, Range I Gain Adjustment Settings (Item 16)...............................2-13
Table 2-17. CC Mode, Level Low, Range II Offset Adjustment Settings (Item 17).............................2-13
ii SLM/SLH Calibration Manual
Table 2-18. CC Mode, Level Low, Range II Gain Adjustment Settings (Item 18)...............................2-14
Table 2-19. CC Mode, Level High, Range II Offset Adjustment Settings (Item 19)............................2-15
Table 2-20. CC Mode, Level High, Range II Gain Adjustment Settings (Item 20)..............................2-16
Table 2-21. CR Mode, Level Low, Range I Offset Adjustment Settings (Item 21)...............................2-17
Table 2-22. CR Mode, Level Low, Range I Gain Adjustment Settings (Item 22)................................2-18
Table 2-23. CR Mode, Level High, Range I Offset Adjustment Settings (Item 23).............................2-19
Table 2-24. CR Mode, Level High, Range I Gain Adjustment Settings (Item 24)...............................2-20
Table 2-25. CR Mode, Level Low, Range II Offset Adjustment Settings (Item 25).............................2-21
Table 2-26. CR Mode, Level Low, Range II Gain Adjustment Settings (Item 26)...............................2-22
Table 2-27. CR Mode, Level High, Range II Offset Adjustment Settings (Item 27)............................2-23
Table 2-28. CR Mode, Level High, Range II Gain Adjustment Settings (Item 28)..............................2-24
Table 2-29. CV Mode, Level Low, Offset Adjustment Settings (Item 29)............................................2-26
Table 2-30. CV Mode, Level Low, Gain Adjustment Settings (Item 30)..............................................2-27
Table 2-31. CV Mode, Level High, Offset Adjustment Settings (Item 31)...........................................2-28
Table 2-32. CV Mode, Level High, Gain Adjustment Settings (Item 32).............................................2-29
Table 2-33. CP Mode, Level Low, Offset Adjustment Settings (Item 33)............................................2-30
Table 2-34. CP Mode, Level Low, Gain Adjustment Settings (Item 34)..............................................2-31
Table 2-35. CP Mode, Level High, Offset Adjustment Settings (Item 35)...........................................2-32
Table 2-36. CP Mode, Level High, Gain Adjustment Settings (Item 36).............................................2-33
SLM/SLH Calibration Manual iii
This page intentionally left blank.
iv SLM/SLH Calibration Manual
SECTION 1
OPERATION DESCRIPTION
This section provides a brief orientation for how to enter, navigate, adjust and store settings, and exit the calibration mode.
Figure 1-1. SLM Series Front Panel
SLM/SLH Calibration Manual 1-1
Operation Description Elgar Electronics Corporation

1.1 NAVIGATION AND SETTING

1. Press MODE and LEVEL keys simultaneously to enter calibration mode.
2. Press © (large up arrow) key to move to the next item.
3. Press ª (large down arrow) key to move to the previous item.
4. Press  (smaller arrow) keys to adjust the calibration settings.
5. Press the SHORT key to store the calibration settings.
6. Press the LIMIT key to exit the calibration mode.

1.1.1 CALIBRATION PROCEDURE FLOW CHART

1-2 SLM/SLH Calibration Manual

2.1 CALIBRATION SETUP

Prior to starting the calibration procedure, gather the necessary equipment (Section minutes.

2.1.1 EQUIPMENT REQUIRED

2.1.1), then power up the chassis and allow it to warm up for 15
DC Voltage Standard: Krohn-Hite EDC 521
Current Shunt: Prodigit 7550
SECTION 2
CALIBRATION
Digital Multi-Meter (DMM): Agilent 34401A
DC Power Supply: Sorensen SG Series (Supply depends upon
models being calibrated.)

2.2 DVM CALIBRATION

Connect the SL Series electronic load as shown in Figure 2-1
Figure 2-1. Connections for DVM Calibration Items 1-4)
Press MODE and LEVEL keys simultaneously to enter Calibration Mode and to be at Item 1 to start calibration.
SLM/SLH Calibration Manual 2-1
Calibration Elgar Electronics Corporation

2.2.1 DVM, RANGE I OFFSET AND GAIN, ITEMS 1 AND 2

7. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-1 for the load model being calibrated.
8. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
MODEL MODEL
Input Voltage Input Voltage
SLM-60-30-150 0.200V SLH-60-120-600 0.200V SLM-60-60-300 0.200V SLH-60-120-1200 0.200V SLM-250-10-300 0.200V SLH-60-120-1800 0.200V SLM-500-10-300 0.400V SLH-60-240-1200 0.200V SLM-60-15-75 0.200V SLH-60-240-1800 0.200V
Table 2-1. DVM, Range I Offset Adjustment Settings (Item 1)
9.
Press the © key to go to Item 2.
10. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-2 for the load model being calibrated.
11. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
12. Repeat steps 1 through 5 (press ª to return to Item 1) until no further
adjustments are necessary for the load DVM to match the DMM reading.
MODEL MODEL
Input Voltage Input Voltage
SLH-60-360-1800 0.200V SLH-500-60-1800 0.400V
SLM-60-30-150 15.00V SLH-60-120-600 15.00V SLM-60-60-300 15.00V SLH-60-120-1200 15.00V SLM-250-10-300 30.00V SLH-60-120-1800 15.00V SLM-500-10-300 60.00V SLH-60-240-1200 15.00V SLM-60-15-75 15.00V SLH-60-240-1800 15.00V
SLH-60-360-1800 15.00V SLH-500-60-1800 60.00V
Table 2-2. DVM Range I Gain Adjustment Settings (Item 2)
2-2 SLM/SLH Calibration Manual
Elgar Electronics Corporation Calibration

2.2.2 DVM, RANGE II OFFSET AND GAIN, ITEMS 3 AND 4

1. Press the © key to go to Item 3.
2. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-3 for the model being calibrated.
3. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
Item 3 Item 3
Input Voltage Input Voltage
SLM-60-30-150 0.200V SLH-60-120-600 0.200V SLM-60-60-300 0.200V SLH-60-120-1200 0.200V SLM-250-10-300 0.200V SLH-60-120-1800 0.200V SLM-500-10-300 0.400V SLH-60-240-1200 0.200V SLM-60-15-75 0.200V SLH-60-240-1800 0.200V
Table 2-3. DVM, Range II Gain Adjustment Settings (Item 3)
4.
Press the © key to go to Item 4.
5. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-4 for the model being calibrated.
6. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
7. Repeat steps 1 through 6 (press ª to return to Item 3) until no further
adjustments are necessary for the load DVM to match the DMM reading.
SLH-60-360-1800 0.200V SLH-500-60-1800 0.400V
MODEL Input Voltage MODEL Input Voltage
SLM-60-30-150 60.00V SLH-60-120-600 60.00V SLM-60-60-300 60.00V SLH-60-120-1200 60.00V SLM-250-10-300 250.00V SLH-60-120-1800 60.00V SLM-500-10-300 500.00V SLH-60-240-1200 60.00V SLM-60-15-75 60.00V SLH-60-240-1800 60.00V SLH-60-360-1800 60.00V SLH-500-60-1800 500.00V
Table 2-4. DVM, Range II Gain Adjustment Settings (Item 4)
SLM/SLH Calibration Manual 2-3
Calibration Elgar Electronics Corporation

2.3 VSENSE CALIBRATION

Make the connections for VSENSE calibration as shown in Figure 2-2.
Figure 2-2. Connections for Vsense Calibration (Items 5-8).

2.3.1 VSENSE, RANGE I OFFSET AND GAIN, ITEMS 5 AND 6

1. Press the © key to go to Item 5.
2. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-5 for the model being calibrated.
3. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
MODEL Input Voltage MODEL Input Voltage
SLM-60-30-150 0.200V SLH-60-120-600 0.200V SLM-60-60-300 0.200V SLH-60-120-1200 0.200V SLM-250-10-300 0.200V SLH-60-120-1800 0.200V SLM-500-10-300 0.400V SLH-60-240-1200 0.200V SLM-60-15-75 0.200V SLH-60-240-1800 0.200V
Table 2-5. VSense, Range I Offset Adjustment Settings (Item 5)
4. Press the © key to go to Item 6.
SLH-60-360-1800 0.200V SLH-500-60-1800 0.400V
2-4 SLM/SLH Calibration Manual
Elgar Electronics Corporation Calibration
5. Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-6 for the model being calibrated.
6. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
7. Repeat steps 1 through 6 (press ª to return to Item 5) until no further
adjustments are necessary for the load DVM to match the DMM reading.
MODEL MODEL
SLM-60-30-150 15.00V SLM-60-60-300 15.00V SLM-250-10-300 30.00V SLM-500-10-300 60.00V SLM-60-15-75 15.00V
Table 2-6. VSENSE, Range I Gain Adjustment Settings (Item 6)
Input Voltage Input Voltage
SLH-60-120-600
SLH-60-120-1200 SLH-60-120-1800 SLH-60-240-1200 SLH-60-240-1800 SLH-60-360-1800 SLH-500-60-1800
15.00V
15.00V
15.00V
15.00V
15.00V
15.00V
60.00V

2.3.2 VSENSE, RANGE II OFFSET AND GAIN, ITEMS 7 AND 8

1. Press the © key to go to Item 7.
2. Set the DC Voltage Standard to the Input Voltage value shown in Table
2-7
for the model being calibrated.
3. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
MODEL Input Voltage MODEL Input Voltage
SLM-60-30-150 0.200V SLH-60-120-600 0.200V SLM-60-60-300 0.200V SLH-60-120-1200 0.200V SLM-250-10-300 0.200V SLH-60-120-1800 0.200V SLM-500-10-300 0.400V SLH-60-240-1200 0.200V
SLM-60-15-75 0.200V SLH-60-240-1800 0.200V SLH-60-360-1800 0.200V SLH-500-60-1800 0.400V
Table 2-7. VSENSE, Range II Offset Adjustment Settings (Item 7)
4.
Press the © key to go to Item 8.
SLM/SLH Calibration Manual 2-5
Calibration Elgar Electronics Corporation
5. Set the DC Voltage Standard to the Input Voltage value shown in Table
2-8
for the model being calibrated.
6. Use the  keys to adjust the load DVM display until it matches the
DMM reading.
7. Repeat steps 1 through 6 (press ª to return to Item 7) until no further
adjustments are necessary for the DVM display to match the DMM reading.
MODEL Input Voltage MODEL Input Voltage
SLM-60-30-150 60.00V SLH-60-120-600 60.00V
SLM-60-60-300 60.00V SLH-60-120-1200 60.00V
SLM-250-10-300 250.00V SLH-60-120-1800 60.00V
SLM-500-10-300 500.00V SLH-60-240-1200 60.00V
SLM-60-15-75 60.00V SLH-60-240-1800 60.00V
SLH-60-360-1800 60.00V SLH-500-60-1800 500.00V
Table 2-8. VSense, Range II Gain Adjustment Settings (Item 8)
2-6 SLM/SLH Calibration Manual
Loading...
+ 30 hidden pages