Down to ±0.11°C using the external reference sensor. 4-wire True-Ohm-Measurement
technology is used.
»
Excellent stability to 0.008˚C
»
Wide temperature range
RTC-700 from 33 to 700ºC (91 to 1292ºF)
»
Improved temperature
homogeneity
Unique active triple-zone block
ensures good temperature homo-
geneity in the calibration zone.
NEW
Triple-zone
Best dry-block
on the market!
Reference Temperature Calibrator
RTC-700
Patent
pending!
»
DLC
Dynamic Load
Compensation
Perfect temperature uniformity in the insert,
even when calibrating large sensors or many
sensors at a time.
B and C models only
»
Display indicator
for temperature uniformity
Shows the degree of temperature uniformity in
the insert when using the new DLC technology.
B and C models only
»
Intelligent reference sensors
JOFRA reference sensors are supplied with
intelligent plugs, holding the calibration data
(coefficients) of the reference sensor. This is
a truly plug’n’play calibration system.
»
USB communication
All RTC calibrators communicate via an
easy-to-use USB port.
Patent
»
Time-saving
Super high speed cooling.
»
EURAMET
Best performing dry-block with regard to the
EURAMET/cg-13/v.01 guideline for testing of
dry blocks.
ISO 9001 Manufacturer
pending!
AMETEK continues to develop new techniques to improve performance,
accuracy, convenience and functionality of the well-known JOFRA calibration products. By doing so, we maintain our position as the leading
worldwide manufacturer of temperature dry-block calibrators.
We are proud to introduce our new high temperature model RTC (Reference
Temperature Calibrator), which is no exception to the above and even
more sophisticated than any existing calibrators.
The RTC offers many new fantastic features, such as:
• Patent pending DLC, Dynamic Load Compensation system, for perfect
temperature uniformity in the insert
• Unique intelligent sensors for plug’n’play connection
• USB connector for communication
• Easy-to-read color VGA display with perfect overview of the actual status
• Intuitive, fast and user-friendly navigation
• Lightweight and easy to carry around
• New functional carrying case design with integrated trolley
• New multi-hole insert kits covering all the most used sensor sizes
• High profile design and the well-known long lasting JOFRA quality
The new RTC calibrator comes in three different models – A, B, and C.
• RTC-A reference temperature calibrator
• RTC-B reference temperature calibrator with input for reference sensor,
DLC sensor and sensors-under-test
• RTC-C reference temperature calibrator with input for reference sensor
and DLC sensor
Specification Sheet, SS-RTC-700
NEW
Triple-zone
Unique temperature performance
The RTC series of calibrators provides precision temperature calibration of sensors, whatever the type or format.
This is accomplished through an innovative active triplezone heating technology.
The RTC-700 features our new active triple-zone heating
technology. Each heating zone is independently controlled
for precision temperature calibration. The homogeneity in
the lower part is close to that of a laboratory liquid bath.
The two lower zones ensures optimum heat dissipation
throughout the entire calibration zone. The upper zone
compensates for heat loss from the sensor-under-test and
from the open top. This design also eliminates the need for
extra insulation of sensors-under-test and makes it possible
to calibrate liquid-filled and other mechanical sensors.
DLC - Dynamic Load Compensation
To bring our well documented active dual-zone
Patent
pending!
to perform top calibration specifications without being affected by the actual load e.g. many sensors or very big sensors.
technology to an even higher level, we have
developed the patent pending new DLC system.
This newly developed unique feature makes it possible
Based on the new DLC functionality, the RTC is the best
performing dry-block calibrator on the market, when being
calibrated and tested according to the globally accepted
EURAMET/cg-13/v.01 guideline for calibration and testing
of dry-blocks.
The DLC system comprises a special differential temperature
sensor designed especially for the RTC. The sensor is placed
in the insert and connected to the calibrator. When the DLC
function is enabled, the calibrator will automatically equalize
the temperature homogeneity inside the insert on top of the
normal temperature control and stabilization.
DLC - User advantages
Calibrating with DLC sensor gives the following advantages:
1. Calibration of several sensors simultaneously
2. Calibration of thick sensors
3. Gives TSL (Thermo Sensitive Length) independency. It is
no longer necessary to know the TLS of the sensor
4. Compensates for sensor production tolerances like the
PT100 element being mounted in various positions in
the sensor
5. Trouble free and full acuracy calibration of sensors with
PT100 elements up to 60 mm length
6. The DLC indicator proves that the dual-zone is active an
functioning well
7. Proves that the calibrator is working perfect. The DLC
value will be very close to 0.00 when the calibrator only
is loaded with DLC sensor and an external reference
sensor
8. Together with the stability indication, the DLC indicates
when the calibration values can be read
A new DLC sensor, Dynamic Load Compensation sensor,
has been specially developed for this purpose.
The RTC-700 features the active triple-zone temperature
control which improves the homogeneity in the well by
adjusting the temperature in the top of the well to the same
temperature as in the bottom. The triple-zone keeps this difference at a minimum.
The new DLC sensor
improves the homogeneity even more by controlling
the homogeneity not only in
the well, but also inside the
insert, where the sensorsunder-test are placed during
calibration. The DLC sensor
measures the temperature
homogeneity in the insert
and provides feedback to the
active triple-zone system,
which now compensates the
temperature difference to a
minimum inside the insert.
In this way, the DLC function makes the homogeneity
independent of the different
loads of the insert.
DLC Sensor
Axial temperature curves for a RTC calibrator with and without the DLC functionality activated
NEW!
USB connector for communication
Another new RTC feature is the USB connection that facilitates easy communication with JOFRACAL. The USB connection also supports easy download of future firmware
upgrades.
The new USB connection provides fast and easy access to
all laptops without the need of RS-232 to USB converters.
Future-proof through e.g. a flash capability for easy firmware upgrades as well as already integrated LAN communication, SD-card slot and USB host connectors for future
use.
2 www.ametekcalibration.comReference Temperature Calibrator • RTC-700
NEW!
plug’n’play
Intelligent reference sensors
The JOFRA STS-200 intelligent reference sensors as well
as the new DLC sensor are all containing all individual calibration data regarding the sensor.
Firstly, this means that the time-consuming coefficient downloading sequence with risk of errors is no longer necessary.
Secondly, the user can change the reference sensor and be
up and running immediately. Time saving and convenient.
With the intelligent sensors, AMETEK has eliminated a
source of error and the system is now giving a fail-safe
plug’n’play calibration system.
Unique
reference sensors
The new STS-200 reference sensors and the DLC
sensors have been specially designed. They are
both angled 90˚ and have
been customized to fit the
calibrator so they are only
slightly higher than the top
of the RTC calibrator.
The unique design makes
it possible to calibrate
threaded sensors and sensors with connection heads
without any problems.
New multi-hole insert kits
Two special multi-hole insert kits have been developed to
comply with calibration of almost any sensor diameter without having to buy numerous inserts.
The first kit is a metric insert kit consisting of only four
inserts covering all diameters from 3 to 13mm. The other is
an imperial insert kit consisting of only three inserts covering six different sizes from ⅛” to ½”.
All inserts have holes for both
NEW!
Wide temperature range
The RTC-700 can perform calibration over a very wide temperature range starting from 33°C and up to 700°C (91 to
1292°F). This makes it possible to perform calibration jobs
over a range of 667°C (1232°F) with only one calibrator.
metric and imperial STS reference sensors and DLC sensor.
With this new insert kit in the
carrying case, the user is now
able to calibrate all commonly
known sensor sizes.
These new insert kits are part
of the JOFRA lightweight
strategy.
700˚C
Easy-to-read
color display and user-friendly navigation
The new 5.7” full color VGA display is very easy to read.
The main temperatures, like SET, READ, TRUE and SUT
(Sensor under test), are always displayed at all stages of the
programming or calibration procedure.
Easy to carry
A calibrator is carried from one job to another. Therefore, it
is essential that it is easy to carry.
We have thoroughly included the weight issue in our design
and have developed new design techniques that have
made the RTC calibrator easy to carry around without compromising its quality, durability and functionality.
Long inserts for accurate calibration
The new extended insert length increases the maximum
possible diameter of the sensor under test to 16 mm.
Switch test of long bulb sensor can be carried out with very
high accuracy.
Patent
pending!
Time is money! This is why all the new RTC calibrators
have an increased calibration speed compared to all other
calibrators. The implication is savings in both production
downtime and general calibration costs.
Fast temperature calibration
Based upon our new developed heating block
the RTC-700 is able to perform a unique com-
bination of high speed and high accuracy.
NEW!
The navigation is menu-driven and very logical and intuitive to use. The display shows any important information
needed for the current function in use. The communication
windows pop up and are followed by discrete sound messages.
The display is very bright and the main information can easily
be read from a distance.
The large display contains many detailed information at a
glance, such as:
• Up to five temperature readings simultaneously
• Stability status
• Load compensation status
• Real time clock
• Serial number of reference sensor
• Sensor-under-test status
www.ametekcalibration.com 3RTC-700 • Reference Temperature Calibrator
Special designed carrying case
NEW!
AMETEK has designed an all-in-one-handle carrying case.
We have now made it possible to store both the STS reference sensors and DLC sensors in the carrying case with an
optimum physical protection. There are compartments for
inserts to cover all sensor under test dimensions and compartments for the new integrated support rod set, wires,
manuals, certificates, plugs, insert tools etc.
All compartments are specially designed to hold one of the
above mentioned items. This makes it very easy to keep
track of any accessories.
For optimum protection of the calibrator and the accessories, the compartments are designed to hold the accessories
fixed during transportation.
Integrated support rod
The new integrated support rod is part
of the reduced weight philosophy. It is
lightweight and very easy to mount on
the RTC. Two fixing holes are integrated
in the calibrator where the support rods
can be mounted. Even though weight
has been reduced with 50% the support
rod is very firmly positioned.
NEW!
SET-Follows-TRUE (B & C model only)
Available on B and C models only, the “SET-Follows-TRUE”
makes the instrument tune in until the temperature reading
of the external reference “TRUE” meets the desired “SET”
temperature. This feature is important when it is critical that
the temperature of the calibration zone matches the desired
temperature when measured with accurate external reference sensors.
Reading of sensor-under-test (B model only)
Model B of the RTC is equipped with a built-in accurate measuring circuit for sensor-under-test (input), which
enables measurement of virtually any type of temperature
sensors including: Resistance thermometers (RTD), thermocouples (TC), transmitters, milliamps (mA), voltage (V)
and thermostats
The RTC calibrators can be user-programmed from the
keyboard for fully automatic sensor calibrations. Once the
unit is programmed, the instrument is self-operating and
performs the configured calibration routine. All calibration
data are stored and can be read in the display.
MVI - Secure temperature stability
MVI stands for ’’Mains power Variance Immunity’’.
Unstable mains power is a major contributor to on-site
calibration inaccuracies. Traditional temperature calibrators
often become unstable in production environments where
large electrical motors, heating elements, and other devices
are periodically cycled on or off. The cycling of supply power
can cause the temperature regulator to perform inconsistently, leading to both inaccurate readings and unstable
temperatures.
The RTC calibrators all employ the MVI functionality, thus
avoiding such stability problems.
Highest accuracy (B & C model only)
The RTC series calibrators may be supplied with a built-in
reference thermometer to be used with an external reference sensor only B & C models. This feature allows the
instrument to perform calibrations on-site, while maintaining
a high accuracy.
Switch test (B model only)
Users may perform a thermoswitch test and find “Open”,
“Closed” and the hysteresis (deadband) automatically. The
instrument retains the last twenty test results.
Auto-stepping
Up to 20 different temperature steps may be programmed
including the hold time for each step. Upon completion of
an auto step routine, the user can easily read the results
for the sensor-under-test on the RTC display. Results from
twenty auto-step calibrations are stored.
The “Set temperature” feature allows the user to set the
exact desired temperature with a resolution of 0.001°.
A special 90° angled external reference sensor has been
designed to accommodate calibration of sensors with a
transmitter head, top connector or similar arrangement. The
user can decide whether to read the built-in reference sensor or the more accurate angled reference sensor from the
large, easy-to-read LCD display of the calibrator. The external sensor and the internal sensor readings are independent
of one another.
4 www.ametekcalibration.comReference Temperature Calibrator • RTC-700
Enhanced stability
A stability indicator shows when the RTC calibrator has
reached the desired temperature and is stable. The user
may change the stability criteria for the external reference
and the sensor-under-test quickly and simply. The stability criterion is the user’s security of a correct calibration.
A count-down timer is displayed next to the temperature
read-out.
Instrument setups
The RTC series allows the user to store up to ten (10) complete instrument setups. You may store all sorts of information including temperature units, stability criteria, use of
external reference sensors, resolutions, sensors-under-test
(SUT), conversions to temperature, display contrasts, etc.
The setup may be recalled at any time.
Maximum and minimum temperature
From the setup menu, the user can select the maximum
and minimum temperature limit for the calibrator. This function prevents damage to the sensor-under-test caused by
excessive temperatures and it helps reducing sensor drift
from exposures of too high temperatures. This feature can
be locked with an access code.
Silent Operation
The RTC calibrator can be programmed to run in silent
operation. This function is an advantage if calibrating in
a laboratory or in an office. If used in silent operation the
calibrator is not using its full speed potential.
As found/As left(B model only)
When running a calibration initiated from a work order, the
user can select the calibration as an As Found or an As Left
calibration.
SYNC output
A synchronization output signals when the instrument is
stable and may be used with ancillary devices such as
video recorders, digital cameras or as an input to a data
logging device. The SYNC output may be useful for automating and documenting your calibrations when calibrating
external reading devices.
Calibration of indication devices
When calibrating an indicating device in the work order
mode, users may key in the results during or after the test.
Using the “Calibration info” function, the user may view the
complete calibration task, including the “Scenario” before
the calibration takes place.
Calibration of up to 24 sensors
with JOFRA ASM
Using the RTC series together with the ASM, Advanced
Signal Multi-scanner, offers a great time-saving automatic
solution to calibrate multiple temperature sensors at the
same time. The ASM series is an eight channel scanner
controlled by the JOFRACAL software on a PC. Up to 3
ASM units can be stacked to calibrate up to 24 sensors at
a time. It can handle signals from 2-, 3- and 4 wire RTD’s,
thermocouples, transmitters, temperature switches and
voltage.
JOFRACAL calibration software
JOFRACAL is a highly versatile calibration software that
is supplied together with the RTC calibrators. The software ensures easy calibration of all kind of temperature
sensors, such as RTD´s, thermocouples, transmitters and
thermoswithes. Furthermore, it can be used for pressure
calibration i.e. pressure gauges and pressure switches.
JOFRACAL integrates with all JOFRA calibration instruments. As for temperature calibrators, it is the whole range
of temperature calibrators. Regarding pressure calibrators,
it integrates with DPC-500, HPC and IPI pressure calibrators. JOFRACAL also has full integration with the series of
signal calibrators.
JOFRACAL can also be used for manual calibrations, as it can
be set up to accept manual entry of calibration data together
with other liquid baths, ice points or dry-block heat sources.
The calibration data collected can be stored on a PC for
later recall or analysis. The RTC calibrator stores the calibration procedure and can be taken out to the process site
without bringing a personal computer.
This allows the RTC calibrator to:
• Operate as a stand-alone instrument, using advanced
calibration routines without the assistance of a personal
computer on site. The work order functionality
• Prevent unauthorized changes to a calibration routine.
Personnel who are not authorized to alter a calibration
routine cannot do so
Once all calibrations are completed, the data may be uploaded to the JOFRACAL for printing of certificates. The
data collected may be stored on the personal computer for
later recall or analysis.
JOFRACAL offers extended output formats of the captured calibration data such as PDF file format and ASCII/
semicolon separated text format for further processing and
calculation of data in spreadsheets and word processors.
Hardware requirements
• INTEL® PENTIUM® II 1.4 GHz processor
• 128MB RAM (256MB recommended)
• 80MB free disk space on hard disk
(120MB recommended) prior to installation
• Standard VGA (800x600, 256 colors).
1024x768 recommended
• USB input device for installation of program
• 1 free USB and 1 free RS-232 serial port
• Microsoft Windows® XP, Microsoft Windows® Vista
• System fonts; MS San Serif and Arial
www.ametekcalibration.com 5RTC-700 • Reference Temperature Calibrator
FUNCTIONAL SPECIFICATIONS
Temperature range
@ ambient temp. 0°C/32°F: ........10 to 700°C/32 to 1292°F
@ ambient temp. 23°C/73°F: ......33 to 700°C/91 to 1292°F
@ ambient temp. 40°C/104°F: ..50 to 700°C/122 to 1292°F
INPUT SPECIFICATIONS
All input specifications apply to the dry-block of the calibrator
running at the respective temperature (stable plus an additional
20 minute period).
All input specifications are valid for RTC-700.
Accuracy (model B & C) with external STS ref. sensor
RTC-700 B & C (33 to 660°C) ..................... ±0.11°C/±0.2°F
Relative to reference standard. Specifications by use of the
external JOFRA STS-200 reference sensor
Accuracy with internal reference sensor
RTC-700 A, B & C .................................... *±0.29°C/±0.52°F
* Accuracy 33 to 660ºC is including calibration uncertainty.
Accuracy 660 to 700ºC ± 0.29ºC exclusive calibration uncertainty.
RTC-700 is delivered calibrated in 660 to 700ºC with calibration
uncertainty ±1.4ºC
Stability
@ 33 to 125ºC/91 to 257ºF ................... ±0.008°C/±0.015°F
@ 125 to 425ºC/257 to 797ºF ............... ±0.015ºC/±0.027ºF
@ 425 to 700ºC/797 to 1292ºF ............... ±0.02ºC/±0.036ºF
Measured after the stability indicator has been on for 15 minutes.
Measuring time is 30 minutes.
Axial homogeneity @ 60 mm
@ temp. range 33 to 100°C..................................... ±0.10°C
@ temp. range 100 to 420°C................................... ±0.25°C
@ temp. range 420 to 700°C................................... ±0.40°C
@ temp. range 91 to 212°F ......................................±0.18°F
@ temp. range 212 to 788°F ....................................±0.45°F
@ temp. range 788 to 1292°F ..................................±0.72°F
Resolution (user-selectable)
All temperatures ........................... 1°, 0.1°, 0.01° or 0.001˚
Heating time
33 to 700°C/91 to 1292°F .................................. 45 minutes
Cooling time
700 to 100°C/1292 to 212°F .............................. 43 minutes
100 to 50°C/212 to 122°F .................................. 17 minutes
50 to 33°C/122 to 91°F ...................................... 20 minutes
Carrying case ............................................... 11.0 kg/24.3 lb
Instrument in Carrying case ......................... 24.3 kg/53.6 lb
Instrument in packaging ............................... 16.3 kg/36.0 lb
Instrument in Carrying case & packaging .... 27.3 kg/60.2 lb
Dimensions (LxWxH)
Instrument ............................................... 362x171x421 mm
Carrying case .......................................... 470x350x600 mm
Instr. & Carrying case .............................. 470x350x600 mm
Instr. & packaging ................................... 442x251x481 mm
Instr., Carrying case & packaging ........... 550x430x660 mm
INSERTS
Insert dimensions
RTC-700 outer diameter ............................29.8 mm/1.17 in
RTC-700 inner diameter ............................. 25.6 mm/1.01 in
RTC-700 length ........................................... 210 mm/8.27 in
Weight of non-drilled insert (approx.)
RTC-700 ....................................................... 1060 g/37.3 oz
Use of other inserts may reduce performance of the calibrator. To get the best results out of the calibrator, the
insert dimensions, tolerance and material is critical. We
highly advise using JOFRA inserts, as they guarantee
trouble free operation.
www.ametekcalibration.com 7RTC-700 • Reference Temperature Calibrator
Single Hole Insert
PREDRILLED INSERTS FOR RTC-700
All predrilled inserts have holes for:
4 mm reference sensor • ¼” reference sensor • 4 mm hole for DLC sensor
Spare part no. for predrilled inserts with reference holes
Instrument
Sensor diameterInsert code
3 mm
4 mm
5 mm
6 mm
7 mm
8 mm
9 mm
10 mm
11 mm
12 mm
13 mm
14 mm
15 mm
16 mm
Package of the above inserts
Spare part no. for predrilled inserts with reference holes
Sensor diameterInsert code
1/8 in
3/16 in
1/4 in
5/16 in
3/8 in
7/16 in
1/2 in
9/16 in
5/8 in
Package of the above inserts
Note 1: Use the insert code, when ordering a JOFRA standard insert together with the RTC calibrator
1
RTC-700 A/B/C
003127148
004
005
006
007
008
009
010
011
012
013
014
015
016
SMM12716 2
Instrument
1
125
187
250
312
375
437
500
562
625
SIM12717 3
127149
127150
127151
127152
127153
127154
127155
127156
127157
127158
127159
127160
127161
RTC-700 A/B/C
127164
127165
127166
127167
127168
127169
127170
127171
127172
UNDRILLED INSERTS FOR RTC SERIES
Inserts, undrilled incl. insulation plugs
Instrument
UN1
UN2
UN3
1
RTC-700 A/B/C
127197
127198
127199
UN1
InsertsInsert code
5-pack, undrilled inserts with no holes
5-pack, undrilled inserts with hole for DLC sensor
5-pack, undrilled inserts with 2 holes for STS reference sensors
(4mm & ¼”) and 1 hole for DLC sensor
Note 1: Use the insert code, when ordering a JOFRA standard undrilled insert together with the RTC calibrator
UN2
UN3
8 www.ametekcalibration.comReference Temperature Calibrator • RTC-700
MULTI-HOLE INSERTS FOR RTC-700 - METRIC (mm)
Spare part no. for multi-hole inserts - metric (mm)
Instrument
M01
M02
M03
M04
M07
M08
M09
SM
1
RTC-700 A/B/C
8 mm
127200
127201
127202
127203
127244
127245
127246
1272 52
12 mm
Insert type Insert code
Multi-hole type 1
Multi-hole type 2
Multi-hole type 3
Multi-hole type 4
Multi-hole type 7
Multi-hole type 8
Multi-hole type 9
Set of 4 Metric Multi Inserts, 3mm to 13mm
(Incl. 127203, 127244, 127245, 127246)
Note 1: Use the insert code, when ordering a JOFRA standard multi-hole insert together with the RTC calibrator
3 mm
1/4”
9 mm
6 mm
4 mm
6 mm
4 mm
Multi-hole M01
4 mm
1/4”
4 mm
4 mm
4 mm
4 mm
1/4”
6 mm
4 mm
4 mm
6 mm
Multi-hole M04
7 mm
4 mm
4 mm
Multi-hole M07
Set of Metric multi-hole inserts.
4 mm
5 mm
13 mm
1/4”
4 mm
4 mm
10 mm
4 mm
4 mm
1/4”
Multi-hole M08
11 mm
1/4”
Multi-hole M09
MULTI-HOLE INSERTS FOR RTC-700 - IMPERIAL (INCH)
Spare part no. for multi-hole inserts - imperial (inch)
Instrument
M05
M06
M10
SI
1
RTC-700 A/B/C
127204
127205
127249
127254
Insert code
Multi-hole type 5
Multi-hole type 6
Multi-hole type 10
Set of 3 Imperial Multi Inserts, 1/8 to 1/2 inch
(Incl. 127170, 127205, 127249)
Note 1: Use the insert code, when ordering a JOFRA standard multi-hole insert together with the RTC calibrator
Insert code
4 mm
6 mm
4 mm
4 mm
3 mm
Multi-hole M02
3 mm
3 mm
Multi-hole M03
6 mm
6 mm
1/4”
3 mm
3 mm
6 mm
3 mm
3 mm
4 mm
4 mm
1/4”
Multi-hole M05
Set of Imperial multi-hole inserts.
1/4”
1/4”
1/4”
1/4”
4 mm
4 mm
1/8”
Multi-hole M06
3/8”
1/4”
1/4”
3/16”
www.ametekcalibration.com 9RTC-700 • Reference Temperature Calibrator
5/16”
4 mm
4 mm
7/16”
1/4”
Multi-hole M10
STANDARD DELIVERY
Model A, B and C:
• RTC dry-block calibrator (user specified)
• Mains power cable (user specified)
• Traceable certificate - temperature performance
• Tool for insertion tubes
• Heat shield
• JOFRACAL
• USB cable
• Manual
Model B instruments contain the following extra items:
• Test cables (2 x red, 2 x black)
• Traceable certificate - input performance for reference
sensor and DLC sensor
• Traceable certificate - input performance for sensor-
under-test inputs (RTD, mA, mV, TC)
Model C instruments contain the following extra items:
• Traceable certificate - input performance for reference
sensor and DLC sensor
ACCESSORIES
125066 Extra fixture for sensor grip
125067 Extra sensor grip
122771 Mini-Jack connector for stable relay output
120516 Thermocouple Male Plug - Type J - Black
120517 Thermocouple Male Plug - Type K - Yellow
120514 Thermocouple Male Plug - Type N - Orange
120515 Thermocouple Male Plug - Type T - Blue
120518 Thermocouple Male Plug - Type R / S - Green
120519 Thermocouple Male Plug - Type Cu-Cu - White
FUNCTIONAL COMPARISON
Model AModel BModel C
Triple-zone heating block•••
MVI - Mains Variance Immunity (or similar)•••
Stability indicator•••
Automatic step function•••
USB communication•••
Display resolution 0.001˚•••
Programmable max. temperature•••
SYNC output (for external recording device)•••
Calibration of short sensors in special insert••
External precision reference sensor input••
External precision DLC reference sensor input••
“SET” follows “TRUE”••
Load compensation functionality••
Input for RTD, TC, V, mA •
4-20 mA transmitter input incl. 24 VDC supply•
All inputs scalable to temperature•
Automatic switch test (open, close and hysteresis)•
Download of calibration work orders from PC•
Upload of calibration results (as found & as left)•
Model AModel BModel C
10 www.ametekcalibration.comReference Temperature Calibrator • RTC-700
ORDERING INFORMATION
Order numberDescription
Base model number
RTC700RTC-700 Series, 33˚C to 700˚C (91˚F to 1292˚F)
FSTraceable System Calibration Certificate (B & C model only)
HSAccredited System Calibration Certificate (B & C model only)
EASFull EURAMET System Calibration Certificate (B & C model only)
EASDFull EURAMET System Calibration Certificate with DLC (B & C model only)
Accessories
CT Solid Protective Carrying case with trolley
SR Support rod set
TRSolid Protective Carrying case with trolley & Support rod set
Sample order number
RTC700 B 230 ASMDLC R4EACT
JOFRA RTC-700 B with 230VAC, EU power cord, set of metric inserts, DLC, STS-200 ref. sensor,
full EA temp. calibration certificate, and carrying case with trolley.
www.ametekcalibration.com 11RTC-700 • Reference Temperature Calibrator
OPTIONS
NEW!
Carrying Case - Option CT
With our new special designed carrying case it is now possible to store
both reference and DLC sensor and 5 inserts in the case with an optimum physical protection. With improved integrated trolley system for
easy and safe transportation.
Support rod set - Option SR
Support rod for sensors to be mounted on
all JOFRA RTC dry-block calibrators. Holds
the sensor under test in its position, while
calibrating. Includes 2 sensor grips and 2
fixtures for sensor grips.
A business unit of AMETEK Measurement & Calibration
AMETEK Test & Calibration Instruments
Technologies Division offering the following industry
leading brands for test and calibration instrumentation.
and liquid baths. Temperature sensors for industrial and
marine use.
Pressure Calibrators
Convenient electronic systems ranging from -25 mbar to
1000 bar - fully temperature-compensated for problem-
free and accurate field use.
Signal Instruments
Process signal measurement and simulation for easy
control loop calibration and measurement tasks.
M&G Dead Weight Testers & Pumps
Pneumatic floating-ball or hydraulic piston dead weight
testers with accuracies to 0.015% of reading. Pressure
generators delivering up to 1,000 bar.
Crystal Pressure
Digital pressure gauges and calibrators that are
accurate, easy-to-use and reliable. Designed for use in
the harshest environments; most products carry an IS,
IP67 and DNV rating.
Lloyd Materials Testing
Materials testing machines and software that
guarantees expert materials testing solutions. Also
covering Texture Analysers to perform rapid, general
food testing and detailed texture analysis on a diverse
range of foods and cosmetics.
Davenport Polymer Test Equipment
Allows measurement and characterization of moisture-
sensitive PET polymers and polymer density.
Chatillon Force Measurement
The hand held force gauges and motorized testers
have earned their reputation for quality, reliabilit y and
accuracy and they represent the de facto standard for
force measurement.
Newage Hardness Testing
Hardness testers, durometers, optical systems and
software for data acquisition and analysis.
United Kingdom
Tel +44 (0)1243 833 302
jofra@ametek.co.uk
France
Tel +33 (0)1 30 68 89 40
general.lloyd-instruments@ametek.fr
Germany
Tel +49 (0)2159 9136 510
info.mct-de@ametek.de
Denmark
Tel +45 4816 8000
jofra@ametek.com
12 www.ametekcalibration.com
www.ametekcalibration.com
USA
Florida
Tel +1 (800) 527 9999
cal.info@ametek.com
California
Tel +1 (800) 444 1850
sales@crystalengineering.net