Ametek RTC-156, RTC-157 Specifications

Page 1
temperature
»
High accuracy
Down to ±0.04°C using the external refer­ence sensor. 4-wire True-Ohm-Measurement technology is used.
»
»
Wide temperature range
RTC-156 from -30 to 155°C (-22 to 311°F) RTC-157 from -45 to 155°C (-49 to 311°F)
»
Improved temperature
homogeneity
Unique active dual-zone block ensures good temperature homogeneity in the calibration zone
Best dry-blocks
on the market!
Reference Temperature Calibrator
Model RTC-156 & RTC-157
»
DLC
Dynamic Load
Patent
pending!
Compensation
Perfect temperature uniformity in the insert, even when calibrating large sen­sors or many sensors at a time. B and C models only
»
Display indicator
for temperature uniformity
Shows the degree of temperature uniformity in the insert when using the new DLC technology. B and C models only
»
Intelligent reference sensors
JOFRA reference sensors are supplied with intelligent plugs, holding the calibration data (coefficients) of the reference sensor. This is a truly plug’n’play calibration system
»
USB communication
All RTC calibrators communicate via an easy-to-use USB port
»
EURAMET
Best performing dry-block with regard to the EURAMET/cg-13/v.01 guideline for testing of dry blocks
ISO 9001 Manufacturer
Specification Sheet, SS-RTC-156/157
DLC
AMETEK continues to develop new techniques to improve performance, accuracy, convenience and functionality of the well-known JOFRA cali­bration products. By doing so, we maintain our position as the leading worldwide manufacturers of temperature dry-block calibrators.
We are proud to introduce our top model RTC (Reference Temperature Calibrator), which is no exception to the above and even more sophisti­cated than any existing calibrators.
The RTC offers many new fantastic features, such as:
Patent pending DLC, Dynamic Load Compensation system, for perfect
temperature uniformity in the insert
Unique intelligent sensors for plug’n’play connection
USB connector for communication
Easy-to-read color VGA display with perfect overview of the actual status
Intuitive, fast and user-friendly navigation
Lightweight and easy to carry around
New functional carrying case design
New multi-hole insert kits covering all the most used sensor sizes
High profile design and the well-known long lasting JOFRA quality
The new RTC calibrator comes in three different models – A, B, and C.
RTC-A reference temperature calibrator
RTC-B reference temperature calibrator with input for reference sensor,
DLC sensor and sensors-under-test
RTC-C reference temperature calibrator with input for reference sensor
and DLC sensor
Page 2
Unique temperature performance
The RTC series of calibrators provides precision tempera­ture calibration of sensors, whatever the type or format. This is accomplished through an innovative active dual­zone heating technology.
The JOFRA RTC-series features our well-known active dual­zone heating technology. Each heating zone is indepen­dently controlled for precision temperature calibration. The homogeneity in the lower part is close to that of a laboratory liquid bath. The lower zone ensures optimum heat dissipa­tion throughout the entire calibration zone. The upper zone compensates for heat loss from the sensor-under-test and from the open top. This design also eliminates the need for extra insulation of sensors-under-test and makes it possible to calibrate liquid-filled and other mechanical sensors.
DLC - Dynamic Load Compensation
To bring our well documented active dual-
Patent
pending!
zone technology to an even higher level, we have developed the patent pending new DLC
system.
tionality, the RTC is the best performing dry-block calibrator on the market, when being calibrated and tested according to the globally accepted EURAMET/cg-13/v.01 guideline for calibration and testing of dry-blocks.
The DLC system comprises a special differential temperature sensor designed especially for the RTC. The sensor is placed in the insert and connected to the calibrator. When the DLC function is enabled, the calibrator will automatically equalize the temperature homogeneity inside the insert on top of the normal temperature control and stabilization.
DLC - User advantages
Calibrating with DLC sensor gives the following advantages:
1. Calibration of several sensors simultaneously
2. Calibration of thick sensors
3. Gives TSL (Thermo Sensitive Length) independency. It is no longer necessary to know the TLS of the sensor
4. Compensates for sensor production tolerances like the PT100 element being mounted in various positions in the sensor
5. Trouble free calibration of sensors with PT100 elements up to 60 mm length
6. The DLC indicator proves that the dual-zone is active an functioning well
7. Proves that the calibrator is working perfect. The DLC value should be very close to 0.00 when the calibrator is loaded with DLC sensor and an external reference sensor
8. Together with the stability indication, the CLD indicates when the calibration values can be read
This newly developed unique feature makes it possible to perform top calibration specifications without being affect­ed by the actual load e.g. many sensors or very big sensors.
A new DLC sensor, Dynamic Load Compensation sensor, has been specially developed for this purpose.
The RTC calibrator features the active dual zone tempera­ture control which improves the homogeneity in the well by adjusting the temperature in the top of the well to the same temperature as in the bottom. The dual zone keeps this dif­ference at a minimum.
The new DLC sensor improves the homogene­ity even more by controlling the homogeneity not only in the well, but also inside the insert, where the sensors­under-test are placed during calibration. The DLC sensor measures the temperature homogeneity in the insert and provides feedback to the active dual-zone system, which now compensates the temperature difference to a minimum inside the insert. In this way, the DLC func­tion makes the homogeneity independent of the different loads of the insert. Based on the new DLC func-
DLC Sensor
Axial temperature curves for a RTC calibrator with and with­out the DLC functionality activated
USB connector for communication
Another new RTC feature is the USB connection that facili­tates easy communication with JOFRACAL. The USB con­nection also supports easy download of future firmware upgrades.
The new USB connection provides fast and easy access to all laptops without the need of RS-232 to USB converters.
Future-proof through e.g. a flash capability for easy firm­ware upgrades as well as already integrated LAN commu­nication, SD-card slot and USB host connectors for future use.
NEW!
2 www.ametekcalibration.com Reference Temperature Calibrator • RTC-156/157
Page 3
NEW!
plug’n’play
Intelligent reference sensors
The JOFRA STS-200 intelligent reference sensors as well as the new DLC sensor are all containing all individual cali­bration data regarding the sensor.
Firstly, this means that the time-consuming coefficient downloading sequence with risk of errors is no longer nec­essary. Secondly, the user can change the reference sensor and be up and running immediately.
With the intelligent sensors, AMETEK has eliminated a source of error and the system is now giving a fail-safe plug’n’play calibration system.
Unique reference sensors
The new STS-200 refer­ence sensors and the DLC sensors have been spe­cially designed. They are both angled 90˚ and have been customized to fit the calibrator so they are only slightly higher than the top of the RTC calibrator.
The unique design makes it possible to calibrate threaded sensors and sen­sors with connection heads without any problems.
Only
10.5 kg
Easy to carry
A calibrator is carried from one job to another. Therefore, it is essential that the weight of the calibrator is as low as possible.
New multi-hole insert kits
Two special multi-hole insert kits have been developed to comply with calibration of almost any sensor diameter with­out having to buy numerous inserts.
The first kit is a metric insert kit consisting of only four inserts covering all diameters from 3 to 13mm. The other is an imperial insert kit consisting of only three inserts cover­ing six different sizes from ” to ½”.
All inserts have holes for
NEW!
Wide temperature range
The RTC-series can perform calibration over a very wide temperature range starting from -45°C and up to 155°C (-49 to 311°F). This makes it possible to perform calibration jobs over a range of 200°C (360°F) with only one calibrator.
both STS reference sen­sors and DLC sensors.
With this new insert kit in the carrying case, the user is now able to calibrate all commonly known sensor sizes.
These new insert kits are part of the JOFRA light­weight strategy.
-45˚C
Easy-to-read color display and user-friendly navigation
The new 5.7” full color VGA display is very easy to read. The main temperatures, like SET, READ, TRUE and SUT (Sensor under test), are always displayed at all stages of the programming or calibration procedure.
NEW!
We have thoroughly included the weight issue in our design and have developed new design techniques that have made the RTC calibrator lightweight and easy to carry around without compromising its quality, durability and functionality.
The purpose of minimizing the weight of the RTC calibrator is to protect, especially frequent users from overload.
Same size of inserts
The new cooling calibrator RTC-156/157 are using the same insert dimensions (30x150mm) as some of our exist­ing calibrators. This makes it possible to reuse inserts from other calibrators.
-24˚C
@15 min.
Fast temperature calibration
Time is money! This is why all the new RTC calibrators have an increased heating and cooling speed compared to all other calibrators. Heating and cooling speed has been increased by up to 20%. The implication is savings in both production downtime and general calibration costs.
The navigation is menu-driven and very logical to use and the display shows any important information needed for the current function in use. The communication windows pop up and are followed by discrete sound messages. The display is very bright and the main information can eas­ily be read from a distance.
The large display contains more detailed information at a glance, such as:
Stability status
Load compensation status
Real time clock
Serial number of reference sensor
Sensor-under-test status
www.ametekcalibration.com 3RTC-156/157 • Reference Temperature Calibrator
Page 4
Special designed carrying case
NEW!
AMETEK has designed an all-in-one-handle carrying case.
We have now made it possible to store both the STS refer­ence sensors and DLC sensors in the carrying case with an optimum physical protection. There is room for inserts and insulation plugs to cover all dimensions and compartments for the new integrated support rod set, wires, manuals, cer­tificates, plugs, insert tools etc.
All rooms are specially designed to hold one of the above mentioned items. This makes it very easy to keep track of any accessories.
For optimum protection of the cali-brator and the accesso­ries, the compartments are designed to hold the accessories fixed during transportation.
Integrated support rod
The new integrated support rod is part of the reduced weight philosophy. It is lightweight and very easy to mount on the RTC. Two fixing holes are integrated in the calibrator where the support rods can be mounted.
NEW!
SET-Follows-TRUE (model B & C only)
Available on B and C models only, the “SET-Follows-TRUE” makes the instrument tune in until the temperature reading of the external reference “TRUE” meets the desired “SET” temperature. This feature is important when it is critical that the temperature of the calibration zone matches the desired temperature when measured with accurate external refer­ence sensors.
Reading of sensor-under-test (model B only)
Model B of the RTC is equipped with a built-in accu­rate measuring circuit for sensor-under-test (input), which enables measurement of virtually any type of temperature sensors including: Resistance thermometers (RTD), ther­mocouples (TC), transmitters, milliamps (mA), voltage (V) and thermostats
The RTC calibrators can be user-programmed from the keyboard for fully automatic sensor calibrations. Once the unit is programmed, the instrument is self-operating and performs the configured calibration routine. All calibration data are stored and can be read in the display.
MVI - Secure temperature stability
MVI stands for ’’Mains power Variance Immunity’’. Unstable mains power is a major contributor to on-site calibration inaccuracies. Traditional temperature calibra­tors often become unstable in production environments where large electrical motors, heating elements, and other devices are periodically cycled on or off. The cycling of sup­ply power can cause the temperature regulator to perform inconsistently, leading to both inaccurate readings and unstable temperatures.
The JOFRA RTC calibrators all employ the MVI functional­ity, thus avoiding such stability problems. The MVI func­tionality is obtained by running the calibrator on stabilized DC voltage.
Highest accuracy (model B & C only)
The RTC series calibrators may be supplied with a built-in reference thermometer to be used with an external refer­ence sensor. This feature allows the instrument to perform calibrations on-site, while maintaining a high accuracy.
Switch test (model B only)
Users may perform a thermoswitch test and find “Open”, “Closed” and the hysteresis (deadband) automatically. The instrument retains the last twenty test results.
Auto-stepping
Up to 20 different temperature steps may be programmed including the hold time for each step. Upon completion of an auto step routine, the user can easily read the results for the sensor-under-test on the RTC display. Results from twenty auto-step calibrations are stored.
The “Set temperature” feature allows the user to set the exact desired temperature with a resolution of 0.001°.
A special 90° angled external reference sensor has been designed to accommodate calibration of sensors with a transmitter head, top connector or similar arrangement. The user can decide whether to read the built-in reference sensor or the more accurate angled reference sensor from the large, easy-to-read LCD display of the calibrator. The external sensor and the internal sensor readings are inde­pendent of one another.
4 www.ametekcalibration.com Reference Temperature Calibrator • RTC-156/157
Page 5
Enhanced stability
A stability indicator shows when the RTC calibrator has reached the desired temperature and is stable. The user may change the stability criteria for the external reference and the sensor-under-test quickly and simply. The stabil­ity criterion is the user’s security of a correct calibration. A count-down timer is displayed next to the temperature read-out.
Instrument setups
The RTC series allows the user to store up to ten (10) com­plete instrument setups. You may store all sorts of informa­tion including temperature units, stability criteria, use of external reference sensors, resolutions, sensors-under-test (SUT), conversions to temperature, display contrasts, etc. The setup may be recalled at any time.
Maximum and minimum temperature
From the setup menu, the user can select the maximum and minimum temperature limit for the calibrator. This func­tion prevents damage to the sensor-under-test caused by excessive temperatures and it helps reducing sensor drift from exposures of too high temperatures. This feature can be locked with an access code.
Silent Operation
The RTC calibrator can be programmed to run in silent operation. This function is an advantage if calibrating in a laboratorie or in an office. If used in silent operation the calibrator is not using its full speed potential.
As found/As left (model B only)
When running a calibration initiated from a work order, the user can select the calibration as an As Found or an As Left calibration.
SYNC output
A synchronization output signals when the instrument is stable and may be used with ancillary devices such as video recorders, digital cameras or as an input to a data logging device. The SYNC output may be useful for auto­mating and documenting your calibrations when calibrating external reading devices.
JOFRACAL calibration software
JOFRACAL is a highly versatile calibration software that is supplied together with the RTC calibrators. The software ensures easy calibration of all kind of temperature sen­sors, such as RTD´s, thermocouples, transmitters and thermoswithes. Furthermore, it can be used for pressure calibration i.e. pressure gauges and pressure switches. JOFRACAL integrates with JOFRA calibration instruments. As for temperature calibrators, it is the whole range of temperature calibrators. Regarding pressure calibrators, it integrates with DPC-500, HPC and IPI pressure calibrators. JOFRACAL also has full integration with the series of signal calibrators.
Free
download
www.jofra.com
JOFRACAL may also be used for manual calibrations, as it can be set up to accept manual entry of calibration data together with other liquid baths, ice points or dry-block heat sources.
The calibration data collected can be stored on a PC for later recall or analysis. The RTC calibrator stores the cali­bration procedure and may be taken out to the process site without bringing a personal computer.
This allows the RTC calibrator to:
Operate as a stand-alone instrument, using advanced
calibration routines without the assistance of a personal computer on site. The work order functionality
Prevent unauthorized changes to a calibration routine.
Personnel who are not authorized to alter a calibration routine cannot do so
Once all calibrations are completed, the data may be up­loaded to the JOFRACAL for printing of certificates. The data collected may be stored on the personal computer for later recall or analysis.
Calibration of indication devices
When calibrating an indicating device in the work order mode, users may key in the results during or after the test. Using the “Calibration info” function, the user may view the complete calibration task, including the “Scenario” before the calibration takes place.
Hardware requirements
INTEL
(PENTIUMTM 800 MHz recommended)
32 MB RAM (64 MB recommended)
80 MB free disk space on hard disk prior to installation
Standard VGA (800 x 600, 16 colors) compatible screen
(1024 x 786, 256 colors recommended)
TM
486 processor
JOFRACAL offers extended output formats of the cap­tured calibration data such as PDF file format and ASCII/ semicolon separated text format for further processing and calculation of data in spreadsheets and word processors.
Calibration of up to 24 sensors with JOFRA ASM
Using the JOFRA RTC series together with the ASM, Advanced Signal Multi-scanner, offers a great time-saving automatic solution to calibrate multiple temperature sen­sors at the same time. The ASM series is an eight channel scanner controlled by the JOFRACAL software on a PC. Up to 3 ASM units can be stacked to calibrate up to 24 sen­sors at a time. It can handle signals from 2-, 3- and 4 wire RTD’s, thermocouples, transmitters, temperature switches and voltage.
www.ametekcalibration.com 5RTC-156/157 • Reference Temperature Calibrator
Page 6
FUNCTIONAL SPECIFICATIONS
Temperature range
RTC-156
@ ambient temp. 0°C/32°F ........-46 to 155°C/-51 to 311°F
@ ambient temp. 23°C/73°F ......-30 to 155°C/-22 to 311°F
@ ambient temp. 40°C/104°F .... -15 to 155°C/ 5 to 311°F
RTC-157
@ ambient temp. 0°C/32°F ........-57 to 155°C/-71 to 311°F
@ ambient temp. 23°C/73°F ......-45 to 155°C/-49 to 311°F
@ ambient temp. 40°C/104°F ....-31 to 155°C/-24 to 311°F
INPUT SPECIFICATIONS
All input specifications apply to the dry-block of the calibrator running at the respective temperature (stable plus an additional 20 minute period).
All input specifications are valid for RTC-156 and RTC-157.
RTD reference input (B & C models only)
Type .................. 4-wire RTD with true ohm measurements
F.S. (Full Scale) ...................................................... 400 ohm
Accuracy (12 months) ......
±(0.0012% rdg. + 0.0005% F.S.)
1)
Accuracy (model B & C) with external STS ref. sensor
RTC-156 B & C ......................................... ±0.04°C/±0.07°F
RTC-157 B & C ......................................... ±0.04°C/±0.07°F
12-month period. Relative to reference standard. Specifications by use of the external JOFRA STS-200 reference sensor
Accuracy with internal reference sensor
RTC-156 A, B & C ..................................... ±0.10°C/±0.18°F
RTC-157 A, B & C ..................................... ±0.10°C/±0.18°F
Stability
RTC-156 ...............................................±0.005°C/±0.009°F
RTC-157 ...............................................±0.005°C/±0.009°F
Measured after the stability indicator has been on for 15 minutes. Measuring time is 30 minutes.
Radial homogeneity (difference between holes)
RTC-156 .........................................................0.01°C/0.02°F
RTC-157 .........................................................0.01°C/0.02°F
Resolution (user-selectable)
All temperatures ......................1° or 0.1° or 0.01° or 0.001˚
Heating time
RTC-156 -30 to 23°C/-22 to 73°F ................. 4 minutes
23 to 155°C/73 to 311°F .............. 15 minutes
RTC-157 -45 to 23°C/-49 to 73°F ................. 7 minutes
23 to 155°C/73 to 311°F ................ 8 minutes
Cooling time
RTC-156 155 to 100°C/311 to 212°F ............ 4 minutes
100 to 23°C/212 to 73°F ................ 8 minutes
23 to -24°C/73 to -11°F ............... 15 minutes
-24 to -30°C/-11 to -22°F ............ 10 minutes
RTC-157 155 to 100°C/311 to 212°F ............ 5 minutes
100 to 23°C/212 to 73°F .............. 10 minutes
23 to -30°C/73 to -22°F ............... 17 minutes
-30 to -45°C/-22 to -49°F ............ 25 minutes
RTD Type Temperature 12 months
°C °F °C °F
Pt100 reference
Note 1: True ohm measurement is an effective method to eliminate errors from induced thermoelectrical voltage
-50 -58 ±0.008 ±0.015 0 32 ±0.008 ±0.015
155 311 ±0.011 ±0.019
DLC sensor input (B & C models only)
Type Temperature 12 months
1)
1)
DLC 155
°C °F °C °F
-50 -58 ±0.014 ±0.025 0 32 ±0.010 ±0.018
155 311 ±0.010 ±0.018
RTD sensor under test input (B model only)
F.S. (range) ............................................................. 400 ohm
Accuracy (12 months).................. ±(0.002% Rdg.+0.002% F.S.)
F.S. (range) ........................................................... 4000 ohm
Accuracy (12 months) ........ ±(0.005% Rdg. + 0.005% F.S.)
2-wire ............................................................ add 50 mOhm
RTD Type Temperature 12 months
°C °F °C °F
Pt1000
Pt500
Pt100
-50 -58 ±0.064 ±0.115 0 32 ±0.073 ±0.131
155 311 ±0.076 ±0.136
-50 -58 ±0.115 ±0.191 0 32 ±0.127 ±0.228
155 311 ±0.142 ±0.255
-50 -58 ±0.026 ±0.046 0 32 ±0.026 ±0.046
155 311 ±0.030 ±0.054
Time to stability (approx.)
RTC-156 ............................................................. 10 minutes
RTC-157 ............................................................. 10 minutes
Immersion depth
RTC-156 ...................................................... . 160 mm/6.3 in
RTC-157 ...................................................... . 160 mm/6.3 in
6 www.ametekcalibration.com Reference Temperature Calibrator • RTC-156/157
Page 7
Thermocouple input
Range ..................................................................... ±78 mV
F.S. (Full Scale) ......................................................... 78 mV
Accuracy (12 months) ........ ±(0.005% Rdg. + 0.005% F.S.)
TC Type Temperature 12 months*
°C °F °C °F
E
J
.
K
T
R
S
N
XK
(only in Russian versions)
U
* Excl. CJC accuracy ±0.3˚C / ±0.54˚F
-50 -58 ±0.09 ±0.17 0 32 ±0.06 ±0.11
155 311 ±0.06 ±0.11
-50 -58 ±0.10 ±0.18 0 32 ±0.08 ±0.14
155 311 ±0.09 ±0.16
-50 -58 ±0.14 ±0.24 0 32 ±0.10 ±0.19
155 311 ±0.11 ±0.20
-50 -58 ±0.15 ±0.26 0 32 ±0.10 ±0.18
155 311 ±0.08 ±0.15
-50 -58 ±1.30 ±2.35 0 32 ±0.78 ±1.40
155 311 ±0.47 ±0.84
-50 -58 ±0.98 ±1.76 0 32 ±0.78 ±1.40
155 311 ±0.49 ±0.89
-50 -58 ±0.20 ±0.35 0 32 ±0.15 ±0.27
155 311 ±0.13 ±0.23
-50 -58 ±0.09 ±0.15 0 32 ±0.06 ±0.11
155 311 ±0.06 ±0.11
-50 -58 ±0.13 ±0.24 0 32 ±0.10 ±0.18
155 311 ±0.08 ±0.14
Transmitter supply
Output voltage ................................................ 24VDC ±10%
Output current .......................................... Maximum 28 mA
Communication interface
Serial data interface .............................USB 2.0 device port
Serial data interface ...................USB 2.0 host double port*
LAN ...................................... Ethernet MAC 10/100 Base-T*
SD ............................................................................ SD slot*
* for future expansion
Miscellaneous
Operating temperature ......................0 to 40°C/32 to 104°F
Storage temperature ....................... -20 to 50˚C/-4 to 122˚F
Humidity ......................................................... 0 to 90% RH
Protection class ..........................................................IP-10
PHYSICAL SPECIFICATIONS
Weight and instrument size (L x W x H)
RTC-156/157 ................................................ 10.5 kg/23.2 lb
RTC-156/157 .......362 x 171 x 363 mm/14.3 x 6.7 x 14.3 in
Shipping (without carrying case)
RTC-156/157 ................................................ 14.8 kg/32.6 lb
Size ......................570 x 235 x 440 mm/22.4 x 9.3 x 17.3 in
Shipping (including optional carrying case)
RTC-156/157 ................................................ 20.5 kg/45.1 lb
Size ....................650 x 380 x 500 mm/25.5 x 14.9 x 19.7 in
Shipping (carrying case only)
Weight ............................................................ 8.0 kg/13.2 lb
Size ....................650 x 380 x 500 mm/25.5 x 14.9 x 19.7 in
INSERTS
Transmitter input mA (B model only)
Range ................................................................. 0 to 24 mA
Accuracy (12 months) ..........±(0.005% Rdg. +0.010% F.S.)
Voltage input VDC (B model only)
Range: .............................................................. 0 to 12 VDC
Accuracy (12 months) ..........±(0.005% Rdg. +0.010% F.S.)
Switch input (B model only)
Switch dry contacts
Test voltage ............................................... Maximum 5 VDC
Test current ..............................................Maximum 2.5 mA
Mains specifications
Voltage ...............................115V (90-127) / 230V (180-254)
Frequency, non US deliveries ...............50 Hz ±5, 60 Hz ±5
Frequency, US deliveries .......................................60 Hz ±5
Power consumption (max.) ........................................400 W
Insert dimensions
RTC-156/157 outer diameter ..................... 29.7 mm/1.17 in
RTC-156/157 inner diameter (multi hole) ... 25.6 mm/1.01 in RTC-156/157 inner diameter (single hole) .22.0 mm/0.87 in
RTC-156/157 length .................................... 150 mm/5.91 in
Weight of non-drilled insert (approx.)
RTC-156/157 .................................................. 290 g/10.2 oz
Use of other inserts may reduce performance of the cali­brator. To get the best results out of the calibrator, the insert dimensions, tolerance and material is critical. We highly advise using JOFRA inserts, as they guarantee trouble free operation.
www.ametekcalibration.com 7RTC-156/157 • Reference Temperature Calibrator
Page 8
UN1
PREDRILLED INSERTS FOR RTC-156/157
All predrilled inserts have holes for: 4 mm reference sensor • ¼” reference sensor • 3 mm hole for DLC sensor
All inserts are supplied with an insulation plug drilled with the necessary holes
Spare part no. for predrilled inserts with reference holes
Instrument
RTC-156/157
Sensor diameter Insert code
3 mm
4 mm
5 mm
6 mm
7 mm
8 mm
9 mm
10 mm
11 mm
12 mm
13 mm
14 mm
15 mm
16 mm
Package of the above inserts
Spare part no. for predrilled inserts with reference holes
Sensor diameter Insert code
1/8 in
3/16 in
1/4 in
5/16 in
3/8 in
7/16 in
1/2 in
9/16 in
5/8 in
Package of the above inserts
Note 1: Use the insert code, when ordering a JOFRA standard insert together with the RTC calibrator
1
003 127312
004
005
006
007
008
009
010
011
012
013
014
015
016
SMM 1273 36
Instrument
1
125
187
250
312
375
437
500
562
625
SIM 1273 35
A/B/C
127313
127314
127315
127316
127317
127318
127319
127320
127321
127322
127323
127324
127325
RTC-156/157
A/B/C
127302
127303
127304
127305
127306
127307
127308
127309
127310
UNDRILLED INSERTS FOR RTC SERIES
Inserts, undrilled incl. insulation plugs
Instrument
UN1
UN2
UN3
1
RTC-156 A/B/C
127299
127300
127301
122 7 81
UN2
Inserts Insert code
5-pack, undrilled inserts with no holes
5-pack, undrilled inserts with hole for DLC sensor
5-pack, undrilled inserts with 2 holes for STS reference sensors (4mm & ¼”) and 1 hole for DLC sensor
Undrilled insulation plug
Note 1: Use the insert code, when ordering a JOFRA standard undrilled insert together with the RTC calibrator
UN3
APPLICATION KIT FOR CALIBRATION OF SANITARY SENSORS
Picture 1 shows a custom-made insert and our STS-102 A cable reference sensor placed in a RTC-156 calibrator. Picture 2 shows the sanitary sensor when fitted into the insert, ready for calibration. Note that the design makes room for the reference sensor cable. Complete STS102 application kit for RTC-156 model B and C including STS102A030EH, recalibration tube, 5-Pack
Picture 2Picture 1
8 www.ametekcalibration.com Reference Temperature Calibrator • RTC-156/157
undrilled flange insertion tubes with cable groove, and carrying case. Order No. 127279
Page 9
MULTI-HOLE INSERTS FOR RTC-156/157 - METRIC (MM)
All inserts are supplied with an insulation plug drilled with the necessary holes
Spare part no. for multi-hole inserts - metric (mm)
Instrument
RTC-156/157
M01
M02
M03
M04
M07
M08
M09
SM
1
8 mm
A/B/C
127329
127330
127331
127332
127241
127242
127243
1273 26
12 mm
Insert type Insert code
Multi-hole type 1
Multi-hole type 2
Multi-hole type 3
Multi-hole type 4
Multi-hole type 7
Multi-hole type 8
Multi-hole type 9
Set of 4 Metric Multi Inserts, 3mm to 16mm
(Incl. 127332, 127241, 127242 and 127243)
Note 1: Use the insert code, when ordering a JOFRA standard multi-hole insert together with the RTC calibrator
3 mm
1/4”
9 mm
6 mm
4 mm
6 mm
4 mm
Multi-hole M01
3 mm
1/4”
3 mm
4 mm
4 mm
4 mm
1/4”
6 mm
4 mm
3 mm
7 mm
4 mm
3 mm
4 mm
5 mm
6 mm
Multi-hole M04
13 mm
1/4”
Multi-hole M07
4 mm
3 mm
10 mm
4 mm
3 mm
1/4”
Multi-hole M08
11 mm
1/4”
Multi-hole M09
MULTI-HOLE INSERTS FOR RTC-156/157 - IMPERIAL (INCH)
All inserts are supplied with an insulation plug drilled with the necessary holes
Spare part no. for multi-hole inserts - imperial (inch)
Instrument
RTC-156/157
M05
M06
M10
SI
1
A/B/C
127327
127328
127247
127311
Insert code
Multi-hole type 5
Multi-hole type 6
Multi-hole type 10
Set of 3 Imperial Multi Inserts, 1/8 to 1/2 inch
(Incl. 127308, 127328 and 127247)
Note 1: Use the insert code, when ordering a JOFRA standard multi-hole insert together with the RTC calibrator
Insert code
4 mm
6 mm
3 mm
4 mm
3 mm
Multi-hole M02
3 mm
3 mm
Multi-hole M03
6 mm
6 mm
1/4”
3 mm
3 mm
6 mm
3 mm
3 mm
4 mm
3 mm
1/4”
Multi-hole M05
1/4”
1/4”
1/4”
1/4”
4 mm
3 mm
1/8”
Multi-hole M06
3/8”
1/4”
1/4”
3/16”
www.ametekcalibration.com 9RTC-156/157 • Reference Temperature Calibrator
5/16”
4 mm
3 mm
7/16”
1/4”
Multi-hole M10
Page 10
STANDARD DELIVERY
Model A, B and C:
RTC dry-block calibrator (user specified)
Mains power cable (user specified)
Traceable certificate - temperature performance
Tool for insertion tubes
JOFRACAL
USB cable
Set of rubber cones for insulation plugs
Manual
Model B instruments contain the following extra items:
Test cables (2 x red, 2 x black)
Traceable certificate - input performance for reference
sensor and DLC sensor
Traceable certificate - input performance for sensor-
under-test inputs
OPTIONS
NEW!
Carrying Case - Option CT
With our new special designed carrying case it is now possible to store all your sensors in the case with an opti­mum physical protection. With improved integrated trolley system for easy and safe transportation.
Model C instruments contain the following extra items:
Traceable certificate - input performance for reference
sensor and DLC sensor
ACCESSORIES
125066 Extra fixture for sensor grib 125067 Extra sensor grib 122771 Mini-Jack connector for stable relay output 120516 Thermocouple Male Plug - Type J - Black 120517 Thermocouple Male Plug - Type K - Yellow 120514 Thermocouple Male Plug - Type N - Orange 120515 Thermocouple Male Plug - Type T - Blue 120518 Thermocouple Male Plug - Type R / S - Green 120519 Thermocouple Male Plug - Type Cu-Cu - White
Model A Model B Model C
Support rod set - Option SR
Support rod for sensors to be mount­ed on all JOFRA RTC dry-block cali­brators. Holds the sensor under test in its position, while calibrating. Includes 2 sensor grips and 2 fixtures for sen­sor gribs.
FUNCTIONAL COMPARISON
Model A Model B Model C Dual-zone heating/cooling block MVI - Mains Variance Immunity (or similar) Stability indicator Automatic step function USB communication Display resolution 0.001˚ Programmable max. temperature SYNC output (for external recording device) Calibration of short sensors in special insert External precision reference sensor input External precision DLC reference sensor input “SET” follows “TRUE” Load compensation functionality Input for RTD, TC, V, mA 4-20 mA transmitter input incl. 24 VDC supply All inputs scalable to temperature Automatic switch test (open, close and hysteresis) Download of calibration work orders from PC Upload of calibration results (as found & as left)
10 www.ametekcalibration.com Reference Temperature Calibrator • RTC-156/157
Page 11
ORDERING INFORMATION
Order number Description
Base model number
RTC156 RTC-156 Series, -30˚C to 155˚C (-22˚F to 311˚F)
RTC157 RTC-157 series, -45˚C to 155˚C ( -49˚F to 311˚F)
Model version
A Basic model, without input
B Full model, incl. DLC sensor input, Reference sensor input, Sensor-under-test input
C Middle model, incl. DLC sensor input , Reference sensor input,
Power supply (US deliveries 60 Hz only)
115 115V AC
230 230VAC
Mains power cable
A European, 230V
B USA/Canada, 115V
C UK, 240V
D South Africa, 220V
E Italy, 220V
F Australia, 240V
G Denmark, 230V
H Switzerland, 220V
I Israel, 230V
Insert type and size
NON No insert selected (standard)
UNX 1 x Undrilled Insert (Please see Insert selection for code)
XXX 1 x Single hole insert (Please see Insert selection for code)
MXX 1 x Multi hole insert (Please see Insert selection for code)
SIX Set of 3 Imperial multi hole inserts. Covering holes from 1/8” to 1/2”
SMX Set of 4 Metric multi hole inserts. Covering holes from 3mm to 13mm
SIM Set of 9 Imperial inserts. Covering holes from 1/8” to 5/8”
SMM Set of 14 Metric inserts. Covering holes from 3mm to 16mm
Dynamic L oad Compensation (B & C models only, optional)
DLC DLC sensor
STS Reference sensor (B & C models only, optional)
R1 STS-102 Ref. sensor. Dia. 4mm. Length 30mm (STS102A030EH)
R2 STS-200 Ref. sensor. Dia. 4mm. Length 161mm (STS200A915EH)
R3 STS-20 0 Ref. sensor. Dia. 1/4”. Length 161mm (STS200B915EH)
Calibration Certificate
F Traceable Callibration Certificate. (standard)
H Accredited Calibration Certificate
EA Full EURAMET Calibration Certificate
FS Traceable System Calibration Certificate (B & C model only)
HS Accredited System Calibration Certificate (B & C model only)
EAS Full EURAMET System Calibration Certificate (B & C model only)
EASD Full EURAMET System Calibration Certificate with DLC (B & C model only)
Accessories
CT Solid Protective Carrying case with trolley
SR Support rod set
TR Solid Protective Carrying case with trolley & Support rod set
Sample order number
RTC156 B 230 A SM DLC R2 EA CT
JOFRA RTC-156 B with 230VAC, EU power cord, set of metric inserts, DLC, STS-20 0 ref. sensor, full EA temp. calibration certificate, and carrying case with trolley.
www.ametekcalibration.com 11RTC-156/157 • Reference Temperature Calibrator
Page 12
A business unit of AMETEK Measurement & Calibration
leading brands for test and calibration instrumentation.
Portable dry-block calibrators, precision thermometers
and liquid baths. Temperature sensors for industrial and
Convenient electronic systems ranging from -25 mbar to 1000 bar - fully temperature-compensated for problem-
Process signal measurement and simulation for easy
Pneumatic floating-ball or hydraulic piston dead weight
testers with accuracies to 0.015% of reading. Pressure
accurate, easy-to-use and reliable. Designed for use in the harshest environments; most products carry an IS,
covering Texture Analysers to perform rapid, general
food testing and detailed texture analysis on a diverse
Allows measurement and characterization of moisture-
have earned their reputation for quality, reliability and
accuracy and they represent the de facto standard for
AMETEK Test & Calibration Instruments
Technologies Division offering the following industry
JOFRA Calibration Instruments
Temperature Calibrators
marine use.
Pressure Calibrators
free and accurate field use.
Signal Instruments
control loop calibration and measurement tasks.
M&G Dead Weight Testers & Pumps
generators delivering up to 1,000 bar.
Crystal Pressure
Digital pressure gauges and calibrators that are
IP67 and DNV rating.
Lloyd Materials Testing
Materials testing machines and software that
guarantees exper t materials testing solutions. Also
range of foods and cosmetics.
Davenport Polymer Test Equipment
sensitive PET polymers and polymer density.
Chatillon Force Measurement
The hand held force gauges and motorized testers
force measurement.
Newage Hardness Testing
Hardness testers, durometers, optical systems and
software for data acquisition and analysis.
United Kingdom
Tel +44 (0)1243 833 302 jofra@ametek.co.uk
France
Tel +33 (0)1 30 68 89 40 general.lloyd-instruments@ametek.fr
Germany
Tel +49 (0)2159 9136 510 info.mct-de@ametek.de
Denmark
Tel +45 4816 8000 jofra@ametek.com
12 www.ametekcalibration.com
www.ametekcalibration.com
USA
Florida
Tel +1 (800) 527 9999 cal.info@ametek.com
California
Tel +1 (800) 444 1850 sales@crystalengineering.net
India
Tel +91 22 2836 4750 jofra@ametek.com
Information in this document is subject to change without notice. ©2013 by AMETEK, Inc., www.ametek.com. All rights reserved.
Singapore
Tel +65 6484 2388 jofra@ametek.com
China
Shanghai
Tel +86 21 5868 5111
Beijing
Tel +86 10 8526 2111
Guangzhou
Tel +86 20 8363 4768 jofra.sales@ametek.com.cn
Pub code SS-RTC-156/157 Issue 1302
Loading...