Ametek APC010C, APC030C, APC200C, APC500C, APC01KG Specifications

...
p r e s s u r e
Specification Sheet
SS-CP-2181-US
NEW SYSTEM A + B PUMPS
Model APC
Advanced Pressure Calibrator
Pressure range
APC010C Vacuum to 0.7 bar / 10 psi APC030C Vacuum to 2 bar / 30 psi APC200C Vacuum to 15 bar / 200 psi APC500C Vacuum to 35 bar / 500 psi APC01KG 0 to 70 bar / 1,000 psi APC03KG 0 to 200 bar / 3,000 psi APC05KG 0 to 350 bar / 5,000 psi APC10KG 0 to 700 bar / 10,000 psi
High accuracy
±0.04% of reading ±0.01% F.S. with no low range degradation in accuracy
True field calibrator
Fully temperature compensated - same high accuracy in the lab and in the field
Transmitter scaling
Calculation of % error of the device-under­test with user-defined scaling
Easy pressure switch test
Automatically records opening and closing point values on the display
Record min and max readings
Capture the min and max pressure readings for safety valve applications
External Pressure Modules
More than 60 different pressure modules, accuracy up to 0.01% F.S.
Documentation made easy
RS232 communication and JOFRACAL calibration software are included in the standard delivery
Complete marine program
Part of a complete program of marine approved temperature, pressure and signal calibrators; including temperature sensors
PRODUCT DESCRIPTION
This remarkable calibrator fea­tures deadweight tester accuracy in a modern digital package. The APC is available as an calibrator or in one of 6 test-ready systems that are complete and equipped to meet your pressure calibration needs.
The APC may be used in appllia­tions from safety valve verification to pressure transmitter.
Features
The JOFRA APC calibrators are the latest in the new generation of AMETEK handheld pressure calibrators. This series was designed to meet high accuracy pressure calibration applications and to make your tasks easier. The APC offers features such as dual information display, 17 different pressure units, transmitter supply, scalable mA input for % error calculation, serial communications, and external pressure module capability. The accuracy of the APC calibrators is deadweight caliber throughout the range and is temperature compensated for operation in controlled or in process applications. This is a truly superior pressure calibrator.
The JOFRA APM series of pressure modules extends the application base for the APC calibrators by allowing for calibrations in ranges other than those of the calibrator; and at a lower cost than purchasing another APC. The APM modules are also compatible with several other JOFRA calibrators.
Calibration of the instrument may be performed locally without returning the APC unit to the manufacturer. Everything needed is an accurate pres­sure reference. If a factory calibration is require, the pump and calibrator are independent and only the calibrator needs to be returned.
ISO 9001 Manufacturer
Model APC Advanced Pressure Calibrator
Specification Sheet
SS-CP-2181-US
Clear dual line display
Large digits display the pressure, icons show current status and mode, and the smaller digits show device­under-test conditions.
ON / OFF
Auto shut-off to extend battery life (user programmable).
mA / %
Save time and effort in calculating the error between the APC and the device- under-test by entering the pressure range corresponding to 4­20mA. This gives immediate read-out of the error in %. You can also read mA or % of 4-20mA in the display.
Switch test
Per form semi-automati c pressure switch tests. The APC records the “close” and “open” pressure values and calculates the hysteresis (deadband). You change the pressure and the APC does the rest.
Electrical connections
The APC calibrator have inputs for mA readings for pressure switches and transmitters, including 24 VDC supply, and for external JOFRA APM pressure modules.
MAX / MIN
Reads the highest and lowest pressure value since you last reset the APC. Monitor a safety valve or look for pressure spikes in the process.
Pressure connections
1/8 in NPT female 15 bar (200 psi) and below. 1/8 in NPT male for 35 bar (500 psi) and above.
2 www.jofra.com
Units
Easy ac cess to 17 d i f fe r e n t p re s s u re engineering units.
Serial interface
Ser i a l in t e r f a c e to communicate with a PC (termin a l ) . Also used if adjustment of the unit is necessary.
Zero
Keep the high accuracy: ZERO the instrument before every test.
Model APC Advanced Pressure Calibrator
Specification Sheet
SS-CP-2181-US
JOFRACAL CALIBRATION SOFTWARE
JOFRACAL calibration software ensures easy calibration of RTD´s, thermocouples, transmitters, thermoswithes, pressure gauges and pressure switches. JOFRACAL can be used with JOFRA DPC-500, APC, CPC and IPI pres­sure calibrators, all JOFRA temperature calibrators, as well as JOFRA AMC910, ASC300 multi signal calibrator and ASM-800 signal multi scanner. When used with JOFRA ASM-800 signal multi scanner, JOFRACAL can perform a simultaneous semi automatic calibration on up to 24 pres­sure and/or temperature devices under test in any combi­nation.
JOFRACAL software controls the complete calibration procedure, stores the results and provides a calibration audit trail through hard-copy certificates. All calibration data are stored for each sensor to monitor drift and opti­mise recalibration intervals. A scheduler feature allows planning of future calibrations.
Please find more information about JOFRACAL calibration software in specification sheet SS-CP-2510, which may be found at www.jofra.com
JOFRA APM pressure module
The JOFRA APM external pressure modules extend the range of the JOFRA APC. There are more than 60 models available with gauge, absolute, differential, and vacuum pressure references and in metric and imperial engineering units.
The modules are engineered for in-plant, field, or labora­tory use. They are ready-to-use with the JOFRA APC and the protocol allows for immediate recognition and use of the module once plugged into the calibrator.
Please find more information about the APM series in specification sheet SS-CP-2190 at www.jofra.com
REQUIREMENTS JOFRACAL
Minimum hardware requirements:
Intel® Pentium® lI 1.4 GHz processor.
64MB RAM (128MB recommended)
80MB free disk space on hard disk (120MB
recommended) prior to installation
Standard VGA (800x600, 256 colours). 1024x768
recommended.
CD-ROM drive for installation of program
1 or 2 free RS-232 serial ports, depending on
configuration
Minimum software equirements:
Microsoft Windows® 98,Microsoft Windows® NT 4.0,
Microsoft Windows® 2000, Microsoft Windows® ME, Microsoft Windows® XP, Vista.
System fonts: MS Sans Serif and Arial
www.jofra.com 3
Model APC Advanced Pressure Calibrator
Specification Sheet
SS-CP-2181-US
FUNCTIONAL SPECIFICATIONS
Pressure range
psi ............................................. vacuum to 10, 30, 200, 500
bar ..................................................vacuum to 0.7, 2, 15, 35
psi ...................................... 0 to 1,000, 3,000, 5,000, 10,000
bar .....................................................0 to 70, 200, 350, 700
Engineering units (built-in)
User selectable ........................................................17 units
(psi, inH2O@4°C, inH2 O@20°C, inH2 O@60°F, inHg@0°C, ftH2 O@4°C, ftH2O@20°C, ftH2O@60°F, bar, mbar, kPa, kg /cm cmH2O@20°C, mH2O@4°C, mH2 O@20°C, mmHg@0°C )
2
, cmH
O@4°C,
2
Pressure accuracy
Pressure APC030C to APC10KG ±0.04% rdg + 0.01% F.S.
Pressure APC010C .............................................0.05% F.S.
Full temperature compensation: .... 0 to 50°C / 32 to 122°F
Vacuum (10 psi / 0.7 bar calibrator) .................... ±0.1% F.S.
Vacuum (30, 200 psi / 2, 15 bar calibrator) ...... ±0.05% F.S.
Full temperature compensation .....10 to 45°C / 50 to 113°F
F.S. (full scale) is the numerical value of the postive pressure range.
Display
LCD .......................................................Dual line plus icons
Display resolution ............ 5 digits (see table on back page)
Display update ..........................................twice per second
Input / output
mA range ............................................................. 0 to 24 mA
mA accuracy ......................................... ±0.015% rdg +2 µA
Transmitter supply ....................................................24 VDC
Peak hold capture ....................................... 50 milliseconds
Switch test input .................................... supply max. 9 VDC
ACCESSORIES
Complete aluminum carrying case
The APC is available as an cali­brator or in one of 6 test-ready systems that are complete and equipped to meet your pressure calibration needs.
The complete ready-to-use sys­tem is delivered in a practical aluminum carrying case.
RS232 communication interface
Connector ........................................................... Stereo jack
Serial ...........0-5 VDC, 9600 baud, 8 data, no parity, 1 stop
Protocol ........................................ ASCll command languag
Media compatibility
........................................Non corrosive gasses and liquids.
Environmetal
Operating temperature .......... -10°C to 50°C / 14°F to 122°F
(Temp. comp. 0 to 50°C)
Storage temperature ............. -20°C to 60°C / -4°F to 140°F
Pressure connection
Calibrators 200 psi / 15 bar and below.......1/8” NPT female
Calibrators 500 psi / 35 bar and above .........1/8” NPT male
Adapters to 1/4 NPT and BSP female are included as standard
Pressure overload
Overload .............................................................. +50% F.S.
Overload alarm ........................ “OL” in display at +20% F.S.
Power supply
Battery ................................................... 1 x 9 VDC Alkaline
Battery life ............................................. Minimum 20 hours
Battery life with APM module connected Minimum 10 hours
Low battery calibrator .............................................at 7 VDC
Instrument dimensions (L x W x H)
Calibrator .................... 200 x 82 x 54 mm / 7.9 x 3.2 x 2.1 in
Calibrator weight (including battery) ................ 470 g / 17 oz
Calibrator, shipping .333 x 888 x 444 mm / 45 x 567 x 45 in
..............................................................................................
Calibrator weight, shipping ............................. 1.1 kg / 2.4 lb
Miscellaneous
CE Conformity ............................................. EN61326: 1998
The JOFRA APC is type approved by Det Norske Veritas. Find the certificate at our web-page www.jofra.com
Approval, Certificate no. ..........................................A-10549
Useful soft case
The APC calibrator is supplied in a handy softcase designed for easy vertical operation during the calibration. The soft case have a pocket for the test leads and an opening in the top andbot­tom provide easy access to the termination block and the pressure module connector. A shoulder strap ensures convenient transportation when climbing ladders, etc. and the handy strap at the back makes it possible to hang the instrument on a pipe, ladder etc, while performing the calibration, test, or service task.
4 www.jofra.com
Loading...