When using the padded carrying case for the AM5 or AM5e Classic
(P/N 87-0070C), DO NOT PLACE the power cord around the side of the
case in the manner demonstrated in Figure 1.
Instead, be certain to WIND THE CORD UNDER THE CASING in the
Area between the feet, as shown in Figure 2.
Failure to follow this procedure could cause deterioration of the cord and
Increase the possibility of electric shock.
SECTION 1 —OVERVIEW/UNPACKING describes the models of the AM5 and
AM5e Classic that are available and provides guidance for proper
unpacking and inspection of the unit.
SECTION 2 —POWER CONSIDERATIONS describes the commercial power
requirements, precautions, battery options, and power ON and
OFF.
SECTION 3 —PHYSICAL AND FUNCTIONAL DESCRIPTION describes the
physical characteristics; front-panel switches, indicators, displays,
and connectors; rear-panel power connector, voltage-select switch,
fuse location, method of identifying units, and permanent
connections. Also described in this section are the Factoryinstalled options and accessories.
SECTION 4 —SELF TEST OPERATIONS describes the operations which should
be performed after the unit is turned on for the first time. Default
settings are described and SEND and MEASURE tests that can be
performed by connecting the TX/2W to the RX jacks are described.
This section also describes restoration of settings before shut down
and a full-calibration procedure.
SECTION 5 —LINE TEST CONNECTIONS AND CONFIGURATIONS describes
the cables that may be used to connect the unit to the equipment
under test. Also covered are schematic diagrams of the line
interfaces for the various configurations that can be selected by an
operator through the use of the TX/2W. RX, LINE, and DIAL
sections of the front panel.
SECTION 6 —MEASURE FUNCTIONS AND PARAMETERS describes each
measurement that may be selected by the MEASURE key on the
front panel. Included in the description are all parameters that
may be changed for each type of measurement. Some typical test
configurations are covered.
SECTION 7 —FILTERS FOR AM5 Classic describes the Noise-Weighting filters
that are provided to IEEE standards.The Filters are switched by
the FILTER key into the measurement circuitry for noise tests.
Included in these sections are frequency response graphs for each
filter type.
SECTION 8 —FILTERS FOR AM5e Classic describes the Noise-Weighting
filters that are provided to CCITT standards.The Filters are
switched by the FILTER key into the measurement circuitry for
noise tests. Included in these sections are frequency response
graphs for each filter type.
SECTION 9 —SEND FUNCTIONS AND PARAMETERS describes each signal
generation function that may be selected by the SEND key on the
front panel. Included in the description are all parameters that
may be changed for each signaling requirement.
SECTION 10 —AM5/AM5e CLASSIC TECHNICAL SPECIFICATIONS provides a
tabular listing of the specifications for the AM5 Classic and the
AM5e Classic.
SECTION 11 —WARRANTY AND SERVICE POLICY provides information on
warranty and on returning unit to Ameritec for service.
iiiNovember 19, 1997
Page 7
AM5(e) Classic Instruction Manual(18-0047)TABLE OF CONTENTS
This Instruction Manual describes the operation of the Ameritec Model AM5
and AM5e Classic Voice/Data Transmission Test Sets.
The AM5 Classic and AM5e Classic
make Transmission Impairment Measurements (“TIMS”) on 2- and 4-wire circuits.
Models are available for IEEE Standards and CCITT Recommendations as follows:
AM5CLASSIC-120 - - - IEEE Standard 743-1984 (Bell Standard 41009)
AM5CLASSIC-200 - - - Same as AM5 Classic-120 with extended range to 200 kHz
AM5e CLASSIC-120 - - - CCITT International Recommendations
AM5e CLASSIC-200 - - - Same as AM5e Classic-120 with extended range to 200 kHz
Both the AM5 Classic and AM5e Classic are hereafter referred to as “AM5(e) Classic”,
and except where specified “AM5” or “AM5e” all information applies to both types of
Classics.
The AM5(e) Classic also contains a separate full-function test tone generator which
produces operator-selected and automatically-stepped (sweep) tones from 20 Hz to 120
kHz (or 200 kHz) required for tests.
The AM5(e)
separate DC hold circuits, and a built-in speaker monitor.
Optional features include a Signaling Adapter with Ring Generator, “Siemens”-type
Line Test Connections, Rechargeable batteries for cordless portable operation, a snapon cover, and a rack-mounting kit for permanent installations.
Classic provides five (5) impedance selections for RX and TX lines,
are microprocessor-based test instruments used to
Classic
The 200 kHz Model of both the AM5 Classic and AM5e Classic extends the maximum
frequency response measurement and generation from 120 kHz to 200 kHz.
1.2UNPACKING
The AM5(e) Classic was thoroughly tested and carefully packed before shipment, and
was in good condition when given to the carrier for shipment.
Upon receipt, thoroughly inspect the outside of the shipping container for damage. If
damage is noted, immediately contact the carrier. The name of the carrier will be noted
on the packing slip which is attached to the outside of the shipping container.
Open the container carefully and compare the contents with the packing slip. Note any
damage or shortages. Notify the carrier in the event of damage. Notify Ameritec in the
event of shortages.
Save the shipping container for future use in case the AM5(e) Classic must be returned
to the factory.
CAUTION: Before connecting the AM5(e) Classic to AC Power , be sure
that the AC power selector switch is in the correct position. For more
details refer to Section 3, sub-sections 3.3.1. and 3.4.1.
All models of the AM5(e) Classic
50/60 Hz power, and have a power consumption of 26 Watts (VA). A rear-panel
selector switch selects 115 V AC or 230 V AC.
A detachable 3-wire power cord is furnished which mates with a CCE standard V-type
connector on the rear panel. The other end of the power cord has a standard threeconductor AC plug. A replaceable 1/2 Ampere fuse is also accessible on the rear panel.
2.2OPTIONAL BATTERY PACK
The AM5(e) Classic may be equipped with an internal battery pack to allow for full
cordless (no commercial power) operation.
The battery pack consists of two sealed lead-acid batteries and associated charging
circuitry. When fully charged, the batteries will power the AM5(e) Classic for about
five to eight (5-8) hours.
A low battery is indicated by blinking decimal points on the front panel display.
Blinking decimal points indicate about one-half hour of remaining battery power
before recharging is required.
The AM5(e) Classic may be operated while the batteries are recharging. Recharging
occurs continuously whenever the AM5(e) Classic is plugged into AC power whether
the power switch is turned ON or OFF. However, charging will be accomplished faster
when the power switch is turned OFF. Commercial power interruption will not cause
unit to stop operating as long as the batteries remain charged.
2.3POWER ON AND OFF
are powered from commercial 115 V AC or 230 V AC
The power ON-OFF switch is located on the upper right side of the front panel (refer to
Section 3 for illustration). It is an alternate-action push button. When pushed-in, it will
lock in that position and the power is ON. Push it again, the button will pop out, and
the power is OFF. When the power is turned ON, the AM5(e) Classic will go through a
self-test procedure, then will show factory-set test selections referred to as “defaults”.
The operator usually will change these settings for a particular test configuration.
The AM5(e) Classic is equipped with a non-volatile memory for test setups. Although
loss of power will cause operation to cease, current test parameter settings will be
stored in memory. When the AM5(e)
re-established instead of the defaults. A procedure for restoring these settings when
power is turned ON is described in Section 4 of this Manual.
This section describes the general physical characteristics of the AM5(e) Classic, the
Front and Rear Panel controls, indicators, and connectors, and the optional equipment.
3.1PHYSICAL CHARACTERISTICS
3.1.1Weight: The AM5(e) Classic weighs six (6) pounds (2.7 kilograms) without the
optional battery pack, and 10 pounds (4.5 kilograms) with the battery pack.
3.1.2 Portable Dimensions: The AM5(e) Classic in its portable case measures 8.3 in.
(210 mm) wide, 3.5 in. (89 mm) high, and 12.1 in. (307 mm) deep. With the
optional snap-on cover in place, the Classic is 14 in. (356 mm) deep.
3.1.3Rack Mounted Dimensions: A 19 in. Rack Mount kit (P/N 850076) can be
ordered for fixed installations. When mounted in the Rack Mount kit, the
AM5(e) Classic measures 19 in (483 mm) wide, 3.5 in. (89 mm) high, and 12.1 in.
(307 mm) deep. A Rack Mount kit shelf is also available (P/N 850233). For more
detail refer to sub-section 3.6.5.
3.1.4 Construction: The front panel (with switch and indicators labels appropriate for
the AM5 Classic or AM5e Classic ) and a black, steel rear panel are attached to a
sturdy aluminum case. Within the case, electronic components are mounted on
three plug-in printed circuit boards, interconnected with ribbon cables. The AC
power supply and optional battery pack with charging circuitry are also
contained within the case.
CAUTION: None of the electronic components within the case are userserviceable, and opening the case will void the manufacturer’s warranty.
3.1.5Rubber Feet: There are four (4) rubber feet on the bottom of the case, and two
rear-panel support brackets for positioning the AM5(e) Classic horizontally or
vertically during operation. There are also four (4) rubber feet on the left side for
setting it down after carrying it with the carrying handle.
3.1.6Carrying Handle: The carrying handle is made of thick, soft, flexible
plastic. It is located on the right side of the case. It extends for use and retracts
against the case when not in use.
NOTE: Neither the rubber feet nor the carrying handle are provided when the AM5(e)
Classic is installed in the 19 inch Rack Mount kit (P/N 850076).
3.2FRONT PANEL DESCRIPTION
The front panel measures approximately 3 in. (75 mm) high by 8 in. (205 mm) wide.
Both panels are illustrated in Figures 3-1 and 3-2. The panels are similar in appearance
with the exception of the label nomenclature for filters, certain noise measurements,
and DIAL terminal identification.
For the AM5(e) Classic models that have extended frequency response, the following
designation plate appears on the lower-left side of the panel:
The front panel contains push-button switches, LED indicators, and LED displays for
setting test conditions and displaying measurements. A speaker volume control knob
is also located on the front panel. Bantam jacks are provided for line connections
(duplicated on the rear panel as terminals) and two terminals are provided for a user’s
“Butt Set” to allow dialing and talking by the user.
All push button keys are non-membrane type, and offer good visual, tactile, and
audible feedback. All indicators, including the display, are LEDs and should never
need replacement.
The remaining sub-sections in Section 3 describe the controls, indicators, and displays
on the front panel. The numbering of the sub-sections corresponds with the
identifications of sub-sections shown on Figures 3-1 and 3-2.
Further descriptions are contained in the other Sections of this manual which describe :
•Self Test OperationsSection 4
•Line Test Connections and ConfigurationsSection 5
CAUTION: Before connecting the AM5(e) Classic to AC Power , be sure
that the selector switch is in the correct position. For more details refer
to 3.4.1.
Push the button IN to turn power ON. Push the button again and the button
will pop out to turn power OFF. The front panel illustrates the power positions
as shown below:
Button in Power ON
Button outPower OFF
When the power is turned ON, The Red LED indicators will come on to show
the default conditions. The two display windows also may show red characters.
For more information on default settings, refer to Section 4.
NOTE: The functions selected for the TX/2W jack described in the next three
sub-sections are duplicated by the T and R terminals on the rear panel.
3.3.2TX/2W bantam jack for 2-wire circuits (2W LED on): The Classic both transmits
and receives signals over this jack through a built-in hybrid circuit. Connect a 2wire circuit to this jack.
3.3.3TX/2W bantam jack for 4-wire circuits (4W LED on): The Classic transmits
signals over this jack. Connect the send (transmit, or TX) pair of a four-wire
circuit to this jack.
3.3.4TX/2W bantam jack for REVERSE selection (4W REV LED on): This jack
becomes a receive (RX) jack as shown on the panel. The Classic receives signals
over this jack (as described in 3.3.5). Connections are per test requirements.
NOTE: The functions selected for the RX jack described in the next two sub-sections
are duplicated by the T1 and R1 terminals on the rear panel.
3.3.5RX bantam jack for 4-wire circuits (4W LED on): The Classic receives signals
over this jack. Connect the receive (or RX) pair of a four-wire circuit to this jack.
3.3.6RX bantam jack for 4-wire circuits for REVERSE selection (4W REV LED on):
The jack becomes a transmit (TX) jack as shown on the panel. The Classic
transmits signals over this jack (as described in 3.3.3). Connections are per test
requirements.
3.3.7MONITOR (TX, MEAS, RX) switch: This three position switch selects the
circuit point where the internal amplifier/speaker is connected:
3.3.8Speaker Volume Control: After the MONITOR switch has been used to select
the monitor point, the Volume control can be turned clockwise to increase the
speaker volume and counter-clockwise to decrease the volume.
NOTE: The speaker is located at the top of the unit near the front.
3.3.9DISPLAY Key: This key selects whether the data shown in the left-hand and
right-hand display windows represent test signals sent to a circuit under test
(SEND) or test data measurements on the circuit under test (MEAS). Each time
the key is pressed, the SEND or MEAS(ure) LED will come on to indicate which
function has been selected.
3.3.10 Left-Hand 7-Segment Display: This is a red LED display containing four 7segment characters with floating decimal point. These characters will indicate
Levels for the SEND or MEASURE Selections in accordance with the LED
indicators to the right of the display (except for Impulse Noise measurements
where they identify the value being displayed in the right-hand 7-Segment
Display).
After the PARAM SEL key is pressed, the characters identify a parameter that
may be set in the right-hand LED display (for those functions that have
parameters). Parameters for MEAS and SEND functions are described in
Sections 6 and 9, respectively.
3.3.11 Left-Hand LED’s (dB, dBm, dBrn, mSEC, HOLD TONE): These LEDs provide
units of measurement and other information about the current SEND or
MEAS(ure) DISPLAY selection. Refer to the table on the next page:
3.1.7Right-Hand 7-Segment Display: This is a red LED display containing four 7-segment
characters with floating decimal point display. These characters will indicate values for
the SEND or MEASURE frequency, time, or impulse noise count selections in
accordance with the LED indicators to the right of the display (Refer to 3.3.13). When
setting parameters, the numeric value of the parameter identified in the left-hand display
is shown.
3.1.8Units of Measurement LED’s (kHz, SEC, MIN, CNT): These LEDs provide the units
of measurement that apply to the data displayed in the right-hand 7-Segment Display for
the SEND or MEAS(ure) DISPLAY selection.
3.1.9PARAM SEL keys (Parameter Select and arrow keys): The PARAM SEL key is
used to step through parameters which may be pre-set for certain of the SEND or
MEAS(ure) functions (Refer to Sections 6 and 9). The table that follows explains the
functions of these keys.
NOTE: The key labelled START STOP, and the LED labelled IMPULSE
NOISE RUNNING are not related to parameter selection or entry (see
3.3.15START/STOP Key (Impulse Noise Measurement): This key is pressed to start
Impulse Noise measurements after IMP NOISE has been selected in the
MEASURE area of the panel. (see 3.3.16).
When Impulse Noise measurements are running the IMPULSE NOISE RUNNING
indicator beside the switch flashes. Impulse Noise measurements may be set
(through parameters) to have a time duration, so after the test is started it will stop
after the time has elapsed. To stop Impulse Noise measurements while the
indicator is flashing, press the START/STOP key again. Refer to Section 6 for a
description of Impulse Noise measurement.
3.1.10MEASURE Key: Eight types of measurement may be selected by this key.
Selection is by rotation and is indicated by the lighting of the LED adjacent to the
abbreviation for the measurement. Measurements are always made from a signal
source coming into the AM5(e) Classic (RX for 4-wire lines, or TX/2W for 2-wire
lines). The following table identifies the measurements, indicates units of
measurement, and whether there are parameters. MEASURE functions and
parameters are described in Section 6.
3.3.17 FILTER (left key): This key can select six filters and RMS detection for use
with the MEASURE NOISE functions of an AM5 Classic (as shown in Figure 3-
1) and five filters and RMS or Quasi-Peak detectors for an AM5e Classic (shown
in Figure 3-2). Filters are selected in rotation each time the key is pressed, and
the selection is indicated by the lighting of the LED adjacent to the abbreviation
for the measurement.
NOTE: When MEASURE selection is LEVEL/FREQ or L/F 15 kHz on an AM5
Classic (LEVEL/FREQ or L/F UNWTD for AM5e Classic), none of the above filters
can be selected. However, the L/F 15 kHz (L/F UNWTD) selection places a low pass
filter into the measurement circuits to eliminate high frequencies which are not of
interest or when high-frequency signals (in addition to Voice Frequencies) on the
line could give false readings. Filters are described in detail in Section 7 for the AM5
Classic, and in Section 8 for the AM5e Classic.
3.3.18 FILTER (right key): This key selects a 60 Hz filter; the 60 Hz LED is lighted
when the filter is selected. The 60 Hz filter can be selected in addition to any other
filters, and is useful when making measurements on circuits that have power-line
interference. This filter cuts off frequencies below 60 Hz, and so also applies to 50
Hz or 25 Hz power-line interference. If the LEVEL or FREQuency measurement
display is not steady, there may be an improvement when this function is ON.
3.3.19SEND (left key): This key allows the user to select and activate, in rotation,
•QUIET no output
•1004 Hz steady tone
•VAR Hz Variable frequency tone (manually adjustable over the
entire frequency range of the generator)
•SWEEP Sweep frequency generator which may be adjusted by
parameters for lower limit (STrT), upper limit (SToP),
frequency increment (STEP), time on each frequency
(rATE) and dwell time on the first and last frequencies
(dLAY).
When a new function is selected, the LED is lighted immediately, the
previously selected SEND function stops, and the new function starts.
•QUIET is the default.
3.3.20 SEND (middle key): The middle key activates SF SKIP, and the LED is lighted
when this function is turned ON. SF skip prevents any tones from being sent in
the SF signaling range (2450 - 2750 Hz for AM5 Classic; 2130 - 2430 Hz for
AM5e Classic). This skip prevents interference with other equipment while
making tests.
•ON (SF SKIP lighted) is the default.
3.3.21 SEND (right key): This key selects and sends one of four frequencies (each of
which may be manually adjusted from the nominal frequencies stamped on the
panel by parameters).
To select a frequency, press the key momentarily (less than one second) to step,
in rotation, from F1 to F4. The LED adjacent to the frequency flashes on and off
at 1/2 second intervals to show the frequency is selected.
To send the frequency, press and hold the key for approximately three seconds
until the LED goes steady, then release the key. The tone will be sent until the
key is pressed again (for less than one second), or if a frequency is selected from
the SEND (left key) function.
The single-tone over-rides any other SEND function. If a SWEEP frequency is
being generated, the SWEEP is reset to its starting frequency after the tone is
stopped.
The single-tone may also be over-ridden by pressing any of the left SEND keys.
For example, if an F4 tone is being output, shifting from QUIET to 1004 Hz will
automatically stop the F4 tone and send the 1004 Hz tone.
•F1 is the default.
NOTE: Send Functions and Parameters are described in Section 9.
3.3.22 TX/2W (left key): This key selects, in rotation, the impedance of the output of the
AM5(e) Classic SEND functions at the TX/2W jack (and rear-panel T and R
terminals) when the LINE selector (3.3.27) is set for any position but 4W REV(erse).
When in the LINE selector is in 4W REVerse position, the selected impedance is for
output at the TX REVERSE jack (and rear-panel T1 and R1 terminals).
•600 ohm is the default.
3.3.23 TX/2W (right key): This key is an alternate-action key which connects a holding coil
equivalent (DC path of about 200Ω) to the TX/2W circuit while maintaining an AC
impedance of greater than 50 KΩ.. When OFF HOOK, the holding coil is across the
TX/2W tip and ring terminals. When ON HOOK, the holding coil is disconnected.
•ON HOOK (OFF HOOK not lighted) is the default.
3.3.24RX (left key): This key selects, in rotation, the impedance of the input of the AM5(e)
Classic for MEASURE functions at the RX (and rear-panel T1 and R1 terminals)
when the LINE selector (3.3.27) is set to 4W.
When the LINE selector is in 4W REVerse position, the selected impedance is for
input at the RX REVERSE jack (and rear-panel T and R terminals).
When the LINE selector is in either 2W position, the measure section presents highimpedance (> 50 KΩ) to the TX/2W jack.
•600 ohm is the default.
The TERMinating function must be active for any impedance setting to apply (see
sub-section 3.3.26).
3.3.25 RX (middle key): This key is an alternate-action key which connects a holding coil
equivalent (DC path of about 200Ω) to the RX circuit while maintaining an AC
impedance of greater than 50 KΩ.. When OFF HOOK, the holding coil is across the
RX tip and ring terminals. When ON HOOK, the holding coil is disconnected.
•ON HOOK (OFF HOOK not lighted) is the default.
3.3.26RX right key): This key selects, in rotation, a termination with the selected
impedance (TERMinate lighted) or a high-impedance (> 50 KΩ) bridging impedance
(BRDG lighted). This key has no effect on RX functions when the LINE selector is in
a 2W position.
3.3.27 LINE Key: This key selects, in rotation, the line connections for the TX/2W and
RX jacks. Selections are shown in the following table. Descriptions of the physical
connections to the jacks are provided in Section 5.
3.3.28 DIAL key: This key selects, in rotation, the DIAL and MEAS(ure) functions. When
DIAL is selected, the TX/2W jack is directly connected to the DIAL terminals on
the front panel and all internal AM5(e) Classic circuitry is disconnected.
When MEAS(ure) is selected, the DIAL terminals are disconnected and the internal
AM5(e) Classic circuitry is connected.
•MEAS(ure) is the default.
3.3.29 DIAL Terminals: The DIAL terminals permit you to connect a Lineman’s
Handset (also called a “butt set”) to the terminals labelled “TIP” and “RING” on the
AM5 Classic or “A” and ”B” on the AM5(e) Classic.
When the DIAL function has been selected, you can use the Lineman’s Handset to
dial out on the TX/2W connection (either DTMF or MF) to access a test line,
responder, or other location. For pulse dialing, you must set the unit to either 4W (or
4W REVerse with change to the RX connection). This applies to units with software
version 103A and before (refer to 4.6.3, step 3, to determine software used in the
unit.).
After the connection is established, select MEAS(ure) to conduct the test.
The rear panel of the AM5(e) Classic is shown in Figure 3-4. The AC power supply
assembly is mounted on the right side of the rear panel. It has a CCE standard V-type
connector for the power cord, a fuse holder for the replaceable 1/2 Ampere fuse, and
a selector switch for 115 V AC or 230 V AC.operation.
CAUTION: Before connecting the Classic to AC Power be sure that the AC Power
selector switch is in the correct position. For more details refer to 3.4.1.
There are five terminals on the lower left side of the rear panel for “permanent”
connections to a line (especially when the AM5(e) Classic is rack mounted).
NOTE: the REVERSE switch on the front panel reverses the functions of both the
front-panel jacks and the rear-panel terminals.
3.4.1 AC Power Selector Switch: This is a red slide switch that is located on the top of
the Power supply. It is used to select between 115 Volt and 230 Volt operation.
Before plugging in the power cord, set this switch to match the power source to
which the power cord is to be connected.
CAUTION: Observe the CAUTION notice printed on the back of the AM5(e)
Classic. When planning to connect the power cord to a 230 volt AC source, be
sure to set this switch to the 230 V position before turning on the power switch.
Damage to the unit will result if 230 V AC is applied with this switch set to 115 V
AC.
3.4.2 Fuse Holder: This holder has a slotted cover which can be unscrewed with a
common blade screwdriver so the fuse may be inspected/replaced. The fuse is 1/2
Ampere 250 V.
3.4.3 Power Plug: This is a 3-pronged male CCE plug for 115V or 230V AC operation.
The mating power cord is supplied. Be sure AC Power Selector switch (3.4.1) is in
the correct position.
3.4.4 Identification Label: This label is attached to the rear of the power supply. The six-
digit Serial Number is coded as follows:
•First two digits: 00 - 52Week of manufacture
•Second two digits:00 - 99Year of manufacture
•Third two digits:00 - 99Identification number
For example: Serial Numbers 519207 and 519211 are for two different units
built in the 51st week of 1992, while 529212 would be for a unit built in the
52nd week.
3.4.5 Screw Terminal, Ground:This is the same connection as the Sleeve contacts on
the front-panel bantam jacks. Ground (connected at either location)is required for
the reference in the NTG (Noise to Ground) Test. (NTG test is described in Section
6.)
3.4.6 Screw Terminals, T1 and R1: These are the same connections as the Tip and Ring
of the right front-panel bantam jack (RX).
3.4.7 Screw Terminals, T and R: These are the same connections as the Tip and Ring of
This sub-section describes the features that must be ordered for the AM5(e) Classic before
shipment. Other accessories that can be ordered at any time are described in sub-section
3.6.
3.5.1Sealed Lead-Acid Batteries and Integral Charger (Option 24-0017): These
batteries are located inside the case and provide five to eight hours of operation
without any external power connection. The charger permits the batteries to be
charged while the unit is in operation or turned OFF. The battery pack adds
approximately four pounds (1.8 kg) to the weight of the AM5(e) Classic .
3.5.2Siemens type “banana” input adapter for AM5e Classic only (Option 25-
0041): The “banana” input adapter brings the line connections to two in-line
socket connectors that are designed to mate with lines that are connected with a 3pronged Siemens-type Banana connector. This is a standard in some European
countries. The adapter is mounted on the bottom of the AM5e Classic , flush with
the front panel. Refer to the front cover of this manual for a view of this unit
mounted underneath the AM5e Classic.
3.5.3Model 30-0033XT Signaling Adapter with Ring Generator: This is an adapter
unit which is attached to the clips on the left- and right-front sides of the AM5(e)
Classic .
The Signaling Adapter provides Loop Start, Ground Start, DID, and Types I, II,
and III E & M signaling. The adapter also incorporates a ring generator with
manual cadence selection and automatic ring trip.
The adapter supplies both 24 and 48 Volt DC loop supply voltage with a switch
to allow for simulation of a 1240 ohm loop resistance.
LED indicators are provided to indicate signaling status.
Bantam connectors are used for access to E & M, SB/SG, Line Tip and Ring,
and the connection to the AM5(e) Classic.
The Signaling Adapter is separately powered by 115 Volts AC.
For more information refer to the "AM5XT Signaling Adapter Models 30-0033
and 30-0033XT Instruction Manuals" (P/N 180023) or contact Ameritec
Corporation Sales Representative.
Note: "Adapter" is alternate spelling on front panel of this particular unit.
This sub-section describes the accessories for the AM5(e) Classic . These items are not
automatically supplied with the AM5(e) Classic . They must be added to the order by
customer request; however, they can be added in the field if desired.
3.6.1 Line Cables: Three types of Line Cables may be ordered for a front-panel line
connection to the bantam jacks (TX/2W and RX). The cables are six-feet in length
and differ in the terminations to the equipment. Cables are as follows (two must be
ordered for four-wire operation):
• 48-0047Bantam (M) to Bantam (M) Input Cable
• 48-0048Bantam (M) to 310 (M) Input Cable
• 48-0062Bantam (M) to mini-clip Input Cable
Refer to Section 5, Line Connections, for illustrations.
3.6.2Protective front panel (85-0078): This is a plastic cover which is designed to fit
over the front panel of the AM5(e) Classic . It clips to the slots on either side of the
front panel bezel and protects the front panel from damage.
Note : This cover is not needed if the Model 30-0033XT Signaling Adapter option is ordered.
3.6.3Padded Carrying Case (87-0070): This case is intended to protect and carry the
AM5(e) Classic to field-testing locations. The AM5(e) Classic can be operated
without removing it from the case. The case is constructed of padded green/grey
heavy canvas and has pouches to hold Instruction Manual, Cables, small tools, and
other items.
3.6.419” Rack Mounting Kits (85-0076): This kit (usually ordered before shipment)
provides the brackets for permanent mounting of an AM5(e) Classic in a standard 19
inch rack. Mounting feet and carrying handle are not installed on the AM5(e) Case
when this item is ordered.
3.6.519” Rack Mounting Shelf (85-0233): This shelf is similar in purpose to the Rack
Mounting Kit. However, it is specifically designed to be used with an AM5e Classic
with the Siemens-type “banana” input adapter.
(END OF SECTION 3)
November 19, 1997Page 3-18
Page 34
AM5(e) Classic Instruction Manual (18-0047)SELF TEST and CALIBRATION
4. SELF TEST AND CALIBRATION
This section describes procedures to verify that the AM5(e) Classic is working properly.
It also identifies the “factory default” settings and provides a calibration procedure.
4.1SELF TEST CONFIGURATION
Self tests are performed with the AM5(e) Classic looped back on itself and not
connected to any external equipment.
Signals sent through the transmit (TX/2W) jack are routed directly back to the Receive
(RX) jack through a Bantam Cable (48-0047) plugged into these jacks. If this cable is not
available, a pair of wires may be connected to the rear panel screw terminals (T to T1
and R to R1).
4.2POWER ON SELF-TEST
To begin a self-test, press the POWER switch in (it will lock). All LEDs will light, all
segments and decimal points will light in the display (“all 8s”), then the words TEST
PASS will be displayed on the display windows on the front panel. These displays will
last for about three (3) seconds. The AM5(e) will then assume its default configuration.
NOTE: If the message “Full Cal ?” appears, it indicates that the user should
perform a calibration check for absolute assurance that the unit will measure
within specifications. This indication should not appear during power turn-on
except at the factory.
4.3DEFAULT SETTINGS
4.3.1Function LEDs at Power Turn-on: The following LEDs will be lighted after the
“all 8s”and TEST PASS message has been displayed. These are the “factory
default” settings which place the AM5(e) Classic in its basic mode of operation.
Refer to Figure 3-1 or 3-2 for the locations of the indicators. The following is a list
of LEDs that will be lighted (starting with DISPLAY MEAS and moving counterclockwise on the front panel):
FunctionLED
DISPLAYMEAS
MEASURELEVEL
FREQ
SENDQUIET, SF SKIP, and F1 (flashing)
TX/2W600Ω
RX600Ω andTERM
LINE4W
DIALMEAS
November 19, 1997Page 4-1
Page 35
AM5(e) Classic Instruction Manual (18-0047)SELF TEST and CALIBRATION
4.3.2Parameter Settings at Power Turn-on: Many of the functions have parameters
that may be changed for a specific test setup. More detail on the settings of
parameters and measurements are described in Sections 6 and 9 that describe
MEASURE and SEND, respectively. A brief summary follows (Value Range
column shows the minimum <default> and maximum values for parameters):
Function keyFunction LEDMeas*/ParameterValue Range
MEASURE
Use up and down
arrow keys to
step through
measurements.
Use PARAM SEL
key to step
through
parameters.
IMP N
W/T
(Impulse Noise
with tone)
or
IMP
NOISE (without
tone)
*Note: Time,
impulse counters ,
and noise are
measurements
listed in order of
display as up arrow
key is pressed).
TiME
iHi (High,
iMid Middle,
iLoLow
Impulse
Counters)
noiS (Noise)
dur(duration)
iTHd (Impulse
Threshold)
idLT (Impulse
Delta)
bLnc (Blanking
Interval)
0.0 MINutes
0 CNT
0 CNT
0 CNT
(Above data
depend on last test)
uNdr
000.1 <015.0>
999.9 MINutes
20.0 <68.0> 110.0
dBrn (AM5) or
–70 <–-22> +20
dBm (AM5e)
2, 3, <4>, 6 dB
0001 <0125> 0255
mSEC
SENDQUIET
NOTE: Any LEVL
setting affects all
SEND levels.
All Frequency
ranges are 0.020
through 120.0
kHz (or 200.0
LEVL (all SEND
Levels)
F1
F2
F3
F4
-50.0 <00.0> 13.0
dBm
AM5 [AM5e]
0.404 [0.304] kHz
1.004 [1.004] kHz
2.804 [2.004] kHz
2.713 [3.004] kHz
kHz). See Value
Range column
VERsion (typical)
101A
defaults
SEND1004 HzLEVL
FrEQuency
LEVL
SENDVAR Hz
FrEQuency
LEVL
SENDSWEEP
STrT Frequency
SToP Frequency
STEP Size
rATE
dLAY (delay)
November 19, 1997Page 4-2
(all SEND)
1.004 kHz
(all SEND)
1.004 kHz
(all SEND)
0.304 kHz
3.004 kHz
0.100 kHz
000.1 <003.0>
999.9 SEConds
<000.0> 999.9
SEC
Page 36
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
2.Press the DISPLAY key so that MEAS is lighted to measure the signal from the
equipment or line under test.
or
Press the DISPLAY key so that SEND is lighted to display the output of the
signal generator to the equipment or line under test.
3.Except for IMPulse NOISE measurements, the left display will be a level in dB,
dBm, or dBrn (AM5 Classic only). The right display will be a frequency in kHz.
NOTE: If the level is too low (or nothing is connected) “UndR” will be displayed
in the left window, and the right window will be blank..
4.Press the PARAM SEL key once
5.The first parameter will be displayed. The left window will identify the
parameter, and the right window will show the value. One of the LEDs beside
the displays will be lighted to show the units.
6.Press the PARAM SEL key again. If there is another parameter, it will be
displayed. Otherwise; the unit will return to displaying the SEND or MEAS
function.
4.3.3 Restoring Previous Settings: Whenever the AM5(e) Classic is turned OFF, the
current test settings are stored in its memory. When the unit is turned back ON
the default settings are restored. If the operator wants to return to the settings
just before the power was turned off, follow this procedure:
1.Press the SEND key until QUIET is lighted.
2.Press the DISPLAY key so SEND is lighted.
3.Hold the
key for more than one (1) second.
4.The display will give the power down? parameters prompt.
5.Press the PARAM SEL key to recall all configuration settings and parameters
before the last power down.
NOTE: To escape from the sequence, press the SEND key instead of PARAM SEL/ETR.
6.After parameters have been reset, set the
desired configuration for tests and continue
operations.
(Power)
(Down?)
November 19, 1997Page 4-3
Page 37
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
4.5LOOPBACK TESTS
This sub-section contains some basic procedures for verifying the operation of the
AM5(e) Classic. In these procedures, those steps identified as “(df)” do not have to be
performed if the factory defaults are in effect, as indicated by the LED display
described in the step. If a different display is shown on the panel, or you want to verify
the operation of the key, then perform the step. Press keys momentarily for less than
1 second, until LED indication occurs, unless otherwise directed.
4.5.1 Set up
1.If not already on, turn on the AM5(e) by pressing the power switch to the
ON (locked-in) position. Observe all lighted LEDs, “8.8.8.8. 8.8.8.8” then
“TEST PASS” display, then default configuration (see 4.5.2).
2.Verify that there is a bantam cord connected between the TX/2W jack and
the RX jack, or (alternatively) there is a T <—> T1 and R <—> R1 connection
at the rear-panel terminals.
4.5.2 QUIET Test: This test checks the AM5(e) Classic default settings, and verifies
QUIET mode displays.
1.(df) Press DISPLAY key until MEAS is lighted.
2.(df) Press MEASURE key until LEVEL FREQ is lighted.
3.Press MEASURE key till NOISE is lighted. Verify that FILTER CMSG
(PSHO on AM5e) is lighted, then continue to press MEASURE until
LEVEL FREQ is lighted again.
(df) Press right FILTER key until 60 Hz is
not lighted.
4.(df) Press left SEND key until QUIET is lighted.
Press middle SEND key until SF SKIP is
(df) Press right SEND key until F1 404Hz (F1 304 on AM5e) is flashing.
5.(df) Press left TX/2W key until 600Ω is lighted.
(df) Press right TX/2W key until OFF HOOK is not lighted.
6.(df) Press left RX key until 600Ω is lighted.
(df) Press middle RX key until OFF HOOK is
(df) Press right RX key until TERM is lighted.
7.(df) Press LINE key until 4W is lighted.
8.(df) Press DIAL key until MEAS is lighted.
November 19, 1997Page 4-4
not lighted.
not lighted.
Page 38
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
9.Observe indication on display:
•Left display reads “Undr” with dBm lighted.
•Right display is blank, and no LED is lighted.
10. Press DISPLAY key until SEND is lighted and observe
• Dashes are displayed in both displays and no LEDs are lighted.
11. Press PARAM SEL key repeatedly and observe that the displays step
through the SEND QUIET parameters (LEVL, F1, F2, F3, F4 and VEr as
shown in the parameters table under 4.5.1). Stop when the display of step
10 is again reached.
12. Set MONITOR switch to the TX position and turn the volume control all the
way counter-clockwise, then turn back about 1/8 turn.
13. Press the right SEND key briefly until F2 is flashing. Then hold key for over
three seconds (then release) until readings below are checked. Press the key
again briefly to stop the tone when done.
•F2 is lighted, steady.
•Left display reads 00.0 with dBm lighted.
•Right display reads 1.004 with kHz lighted.
•Tone is heard from speaker (stops when key is released)
•Displays go back to those of step 10 (when key is released).
14. Press MEASURE key until NOISE is lighted.
15. Press DISPLAY key until MEAS is lighted.
•Left display reads “undr” with
dBrn lighted (dBm on AM5e Classic)
•Right display is blank and no LEDs are lighted.
16. Set the MONITOR switch to the MEAS position, and turn volume control
knob clockwise slowly until any sound from speaker is at a comfortable
level. The most that should be heard is hiss from the internal electronics of
the AM5(e) Classic since SEND is set to QUIET. Turn volume control back
to about 1/8 turn from full counter-clockwise limit before starting the next
test.
NOTE: When the MONITOR switch is set to MEAS position, the speaker is
connected to an automatic-gain-controlled amplifier used in the NOISE and
auto-ranging features. Even though no signal is connected to the RX jacks,
there will be feed through of tones from the SEND function. If the signal is
looped back, the amplifier will compensate for the input, and the tone will
sound about the same as it will when MONITOR switch is in the TX or RX
position.
November 19, 1997Page 4-5
Page 39
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
4.5.3 SEND 1004 Hz Test: This test activates 1.004 kHz continuous test tone, checks
MEASURE functions, and verifies operation of HOLD TONE indicator and
notch filter for noise tests. Unless otherwise indicated, a test cable must be
connected from TX/2W jack to the RX jack (or between rear terminals).
1.Set MONITOR switch to the TX position and turn volume control to about
1/8 turn clockwise (from full counter-clockwise).
2.Press the left SEND key until 1004 Hz is lighted.
•Tone will be heard from speaker.
3.Press MEASURE until LEVEL FREQ is lighted.
4.Press DISPLAY until SEND is lighted
• Left display will read 00.0 ±1 (1.0 through -1.0) with dBm lighted.
• Right display will read 1.004 ±.002 (1.002 through 1.006) with kHz
lighted.
5.Press DISPLAY until MEAS is lighted.
•Displays will read the same as step 4 (assumes TX/2W jack is connected
to RX jack).
6.Set MONITOR switch to RX position.
•Tone will be heard from speaker
7.Press MEASURE key until NOISE is lighted
•For AM5 Classic — Left display will read 90.0 ±1.0 with
•For AM5e Classic – Left display will read 00.0 ±1.0 with dBm lighted.
•Right display will be blank. (Frequency measurement is disabled
during NOISE measurement.)
8.Press MEASURE key until NOISE W/T (Noise with Tone) is lighted.
•Green HOLD TONE LED will be lighted beside left display.
•For AM5 — Left display will read less than 40.0 with dBrn lighted.
•For AM5e – Left display will read less than –50.0 with dBm lighted.
•Right display will read 1.004 kHz
•Tone will be heard from speaker. (Tone will be fainter if MONITOR
switch is set to the MEAS position.)
NOTE: NOISE with Tone incorporates a sharply-tuned notch filter at 1.004 kHz.
This filter keeps the HOLD tone from the measuring circuits. In some cases, this
level will be low enough that “undr” is shown in the left display. In this case, the
right display will be blank.
November 19, 1997Page 4-6
dBrn lighted.
Page 40
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
9.Press MEASURE key until S/N RATIO is lighted.
•Green HOLD TONE LED will be lighted beside left display.
•Left display will read greater than 50.0 or “oVEr” with dB lighted (this
indicates a very quiet “line”).
•Right display will read 1.004 ±.002 with kHz lighted.
10. Disconnect test connections from TX/2W to RX. Turn off power then turn it
back on to re-establish the default settings.
NOTE: If a problem is found in any of the checkouts or the Calibration Procedure
contact Ameritec Customer service. See Section 11, Warranty.
4.6CALIBRATION PROCEDURE
The AM5(e) Classic was calibrated at the Factory for proper measurement accuracy by
loading preset values into its internal RAM. RAM is maintained by a small internal
battery.
During the power-on self-test, the unit will automatically diagnose itself and, if
necessary, provide a display to inform the operator of the need for recalibration.
4.6.1 External Test Equipment: A high-accuracy digital voltmeter must be used to
calibrate the AM5(e) Classic. The voltmeter must measure AC volts RMS, and be
accurate at 1000 Hz within .4%. Minimum
input impedance of the voltmeter should be
200 kΩ..
(Full)
(Calibration?)
A test cable such as the 48-0062 (Refer to
Section 5) may be used to connect the voltmeter leads to the front panel jack.
4.6.2 Calibration Setup
1. Disconnect all self-test or other input or output connections at the AM5(e)
front panel or at the rear panel.
2.Connect leads of the volmeter to RX Jack at the front panel (using test cable)
or make the connection at the T1 and R1 terminals at the rear panel.
November 19, 1997Page 4-7
Page 41
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
4.6.3 Calibration Procedure: This procedure may be performed at any time to bring
up the calibration function. If power has just been turned on, the display at step
4 will already have appeared.
1.Press the SEND key until QUIET is lighted.
2. (df) Press the DISPLAY key so that SEND is lighted.
3.Press PARAM SEL/ETR key repeatedly
to step through the QUIET parameter
list. Stop at the Version display.
(VErsion)
(101A – typical)
4.Hold the START/STOP key (adjacent to
IMP NOISE RUNNING LED) for more
than one (1) second until the full
(Full)
(Calibration?)
calibration prompt is displayed.
5.To skip the calibration, press SEND key (QUIET will remain lighted).
or
To go into the full calibration mode of
operation, press the
key.
Selections will go to the defaults, the
1004Hz LED (and tone) will come on, and the display will prompt for the
high calibration (V1).
6.Check the reading on the external voltmeter. The 1.900 display represents
the expected AC RMS value, the actual reading on the voltmeter varies
from unit to unit. If the display is greater than 1.900, turn the power OFF
and restart calibration.
7.Use the
and keys to change the displayed value to agree with the
voltage measured by the voltmeter. This voltage should be larger than 1.200
and smaller than 2.200. If the voltage is outside of these limits, the unit
must be repaired.
8.When the display matches the voltmeter
reading, press
key to accept the
reading and to display the prompt for
low calibration (V2). (The level of the tone will drop).
9.As with V1, the 0.350 represents an expected value. Use the and keys to
change the displayed value to agree with the voltage measured by the
voltmeter. This voltage should be larger than 0.250 and smaller than 0.450.
If the voltage is outside of these limits, the unit must be repaired.
However, if the display is greater than 0.350, turn the power OFF and
restart the calibration.
November 19, 1997Page 4-8
Page 42
AM5(e) Classic Instruction Manual(18-0047)SELF TEST and CALIBRATION
10. Press
reading and display the self calibration
prompt.
11. Disconnect the voltmeter, then press the key to start the selfcalibration process.
•MEASURE will automatically go to L/F 15 kHz (L/F UNWTD on
AM5e Classic ) and the appropriate LED will be lighted.
•There will be a SEND display for
the first calibration tone
•The self-calibration process takes about four (4) minutes. During this
time various tone levels will be displayed in the left window, many of
them preceded by a minus sign. (see step 12 for possible error).
•The conclusion of the calibration will occur when the L/F UNWTD LED
will go off, MEASURE will return to LEVEL FREQ lighted, and the
DISPLAY selection will automatically go to MEAS lighted.
• The DISPLAY selection will
automatically go to MEAS and
(since nothing is connected to the
RX jack) will give an under-value reading.
12. ERROR CONDITION: If the unit detects
and error during the calibration process,
an error reading may occur in the right
display window (see example to the right). If this happens turn the unit off
then back on so that the calibration can be performed again. An error may
signify a defective unit, so Ameritec Customer Service should also be
notified for further instructions.
This section describes the Line Functions which set up the 2-Wire
and 4-Wire interfaces to the AM5(e) Classic. It includes a
functional block diagrams to illustrate LINE termination and
signal routing within the unit, drawings of the test connector
cables, and a description of each LINE selection. All line
connections are made to the front panel jacks) or to corresponding
TX/2W RX
RX TX
REVERSE
RT1
T
R1
terminals on the rear panel (shown to the right). Typical test
configurations are also described in this section.
(TX/2W) (RX)
(Sleeves
The AM5(e) Classic line connections are not polarity sensitive. Both the TX/2W T
and R connections and the RX T1 and R1 connections can be reversed without
affecting any measured values.
5.1LINE CABLES
Figure 5-1 shows the line cables which may be ordered as accessories for the AM5(e)
Classic. The bantam plug on any of the cables plugs into either the TX/2W or RX jack.
The bantam or the PL310 plug on the other end of the 48-0047 or 48-0048 cables can be
plugged into jacks at the Central Office or other connectorized equipment.
The 48-0062 Line Cable has clip leads used to connect to instruments, terminal blocks,
or other non-connectorized circuit points.
BANTAM PLUG
CABLE 48-0047, 6 FT.
BANTAM PLUG
BANTAM PLUG310 PLUG
NOTE: Cables manufactured before 5/7/92
are GREEN (Ring—B) and RED (Tip—A)
and will not have this color-code label.
CABLE 48-0048, 6 FT.
RED (Ring — B)
ringsleeve
tip
BANTAM PLUG
CABLE 48-0062, 6 FT.
Figure 5-1 — Line Cable Accessories for AM5(e) Classic
Circuitry shown for LINE 4W Position. In 4W REVerse position T and R are exchanged with T1 and R1.
RX
135
600
450
300
75
67.5
600
450
300
75
67.5
150
1200Ω
900Ω
600Ω
150Ω
135Ω
900
600
TERM
BRDG
TX/2W
Impedance
Selections
RX
1200
200 K
Frequency,
Level,
Noise
MEASURE
Selections
200 K
NTG
MEASURE
Signal
Generator
SEND
Selections
QUIET
SEND
3/1/93
Figure 5–2 — 4-Wire Line Circuit Block Diagram
5.2 FUNCTIONAL DIAGRAMS OF AM5(e) CLASSIC
Figures 5-2 and 5-3 are Block Diagrams of the 4-Wire and 2-Wire line terminating
circuitry. The following sub-sections describe the LINE circuit functions in terms of
front-panel keys and the LEDs that show their selection.
5.2.1 LINE 4W selection: In the 4-Wire configuration (Figure 5-2),
signal flow is from the SEND functions out of the unit via the
TX/2W jack (or rear-panel T and R terminals). Signal flow
from the tested circuit is into the RX jack (or rear-pane T1 and
R1 terminals) and into the MEASURE functions.
5.2.2 LINE 4W REVerse selection: When in the 4-Wire REVerse
4W
REV
4W
2W
2.16µf
2W
configuration, the MEASURE and SEND functions are
switched to the TX/2W and RX jacks, respectively (as
indicated in note on Figure 5-2). The 4W REVerse setting is
also useful for making receive-only measurements on a
TX/2W line. Refer to NOTE at 5.2.11.
Circuit shown for LINE selection in 2W or position.
2W
2.16 µf
Figure 5–3 — 2–Wire Line Circuit Block Diagram
5.2.3 LINE 2W selection: When in the 2-Wire configuration (Figure 5–3), the
MEASURE and SEND functions are both connected to the TX/2W jacks.
However, the RX OFF HOOK, and TERMinating selections are inoperative for
the MEASURE functions. The impedance and loading coil selections are still
reflected at the RX jacks.
NOTE: dbm and dBrn calculation continue to be made based on the RX impedance
selection (see 5.2.7 and 5.2.10).
5.2.4LINE 2W 2.16µf selection: This is the same as described above, except for the
insertion of a series capacitance of 2.16µf between the TX/2W jacks and the
terminating TX/2W impedances. This selection is typically used with SEND
QUIET mode for return loss measurements from the distant end of the circuit
5.2.5 DIAL terminal DIAL selection: completely
disconnects the TX/2W jack from the internal circuitry
of the unit, and routes the tip and ring leads directly to
the front-panel DIAL terminals. (If 4W REVerse is
selected, the RX jack is routed to the DIAL terminals).
DIAL
MEAS
TIPRING
A circuit connected to the jack is only terminated by an
instrument connected to the DIAL terminals.
Terminals are labelled “TIP” and “RING” for an AM5
DIAL
Classic (shown at right) and “A” and “B” for an AM5e
Classic.
The SEND impedances are selected by the left
TX/2W front-panel switch, and the corresponding
LEDs are lighted for each impedance selection.
5.2.7 TX/2W OFF HOOK selection: The electronic
equivalent of a loading coil (OFF HOOK) is selected
by the right TX/2W switch.
900
600
150
135
1200
OFF
HOOK
5.2.8 RX 135 , 150 , 600 , 900 , 1200 selections: The
MEASURE impedances are selected by the left RX
front-panel switch, if TERM has been selected by the
right RX switch. If BRDG has been selected, the
impedance selector is disconnected and the
impedance is 50 KΩ or greater. Calculation of dBm
and dBrn is based on the impedance selected (see
5.2.10).
TX/2W
900
600
150
135
1200
OFF
HOOK
TERM
BRDG
5.2.9 RX OFF HOOK selection: The electronic equivalent
of a loading coil (OFF HOOK) is selected by the
middle RX switch. The holding coil presents a DC
resistance of 200 Ω or less while maintaining a high-
RX
impedance to AC.
5.2.10 RX TERM or BRDG selection: The MEASURE impedances are selected by the
left RX front-panel switch, only if TERMinate has been selected by the right RX
switch. If BRDG (Bridging) has been selected, the impedance selector is
disconnected and the impedance at the RX jack is 50 KΩ or greater.
NOTE: dBm and dBrn readings displayed are calculated on the setting of the RX impedance
switch, even though the network has been disconnected by the BRDG selection or because one of
the LINE 2W selections has been made. The formula used is:
5.2.11 SEND QUIET selection: The signal generator is
disconnected from the line when QUIET has been
selected by the left SEND switch; however the TX/2W
line remains terminated by the selected TX/2W
impedance.
NOTE: For an “open circuit” configuration refer to
the description in sub-section 9.7.
5.2.12 SEND SF SKIP selection: This is a lock-out which
prevents the SEND circuits from sending any
frequencies in the SF signaling band. It has no effect on
the impedance presented to the TX/2W line.
5.2.13 FILTER selections: Noise-Weighting Filters are
selected by the left FILTER key, and are provided to
limit certain parts of the signal spectrum being applied
at the RX jack (or TX/2W in either 2W mode) from
reaching the measure circuits. The filters have no effect
on the impedance presented to the line, but they will
affect the levels measured in accordance with the filter
characteristics. Refer to Section 7 for filters supplied
with the AM5 Classic and Section 8 for filters supplied
with the AM5e Classic.
NOTE: These filters are only switched into the circuitry
when noise measurements are being made. Refer to Section 6.
SWEEP
VAR Hz
1004 Hz
QUIET
SEND
(AM5 Classic)
SWEEP
Hz
1004 Hz
QUIET
SEND
(AM5e Classic)
15kHz
3kHz
PGM
CMSG
SF
SKIP
SF
SKIP
3004Hz
2004Hz
1004Hz
304Hz
WIDE
50kBit
60 Hz
F4
2713Hz
F3
2804Hz
F2
1004Hz
F1
404Hz
F4
F3
F2
F1
5.2.14 FILTER 60 Hz selection: The 60 Hz filter, selected by
the right FILTER key produces a sharp roll-off in low
frequency response below 60 Hz. It takes effect
independently of other FILTER selections. Its primary
purpose is to limit power-line interference from
affecting readings. This filter has no effect on the
impedance presented to the RX line.
5.2.15 MEASURE NTG selection: When the NTG (Noise to
Ground) function has been selected by the MEASURE
switch, a balanced connection from both sides of the
RX line (or TX/2W line if in a 2-Wire configuration) is
made through 200 KΩ resistors to a high-gain amplifier
which then is connected to the level measurement
circuitry.
The frequency measurement circuitry is disconnected,
and only noise in dBrn (for AM5 Classic) or dBm (for
AM5e Classic) is displayed. (The right display reads
“NTG–”). The ground connection must be made to the
equipment ground for NTG measurements.
NOTE: The sleeve or the rear-panel ground terminal must be
connected to the ground of the equipment being measured.
There is no continuity between the AM5(e) Classic ground
and the chassis or the AC grounding lead.
5.2.16 MONITOR (TX, MEAS, RX) Selections: The monitor
switch (shown with its associated volume control) has
no effect on the line connections. It allows the user to
connect the built-in speaker to the TX/2W circuit, the
RX circuits, or the internal automatic-gain-control
amplifier which is used for RX measurements. The
MEAS selection permits the user to hear low-level
noise on the RX circuit which would be inaudible in the
RX position.
NOISE
W/T
NOISE
L/F
15kHz
LEVEL
FREQ
IMP N
W/T
IMP
NOISE
S/N
RATIO
NTG
MEASURE
(AM5 Classic)
NOISE
W/T
NOISE
L/F
UNWTD
LEVEL
FREQ
IMP N
W/T
IMP
NOISE
S/N
RATIO
NTG
MEASURE
(AM5e Classic)
MONITOR
MEAS
RXTX
5.2.17 DISPLAY (SEND, MEAS) Selections: Like the
MONITOR switch, the DISPLAY selections have no
effect on line connections. The selections of the
DISPLAY key connect the internal level and frequency
measurement circuits to either the TX/2W circuits or
the RX circuits.
NOTE: The MEASURE parameters for IMPulse NOISE (with or without tone)
must be set in the MEAS position (Refer to sub-section 6.7.2). All other
parameters are set in the SEND position (Refer to Section 9).
For correct level measurements and to ensure that the tested equipment operates
properly, the correct terminating impedance must be selected. The AM5(e) Classic can
terminate TX/2W lines with 135, 150, 600, 900, or 1200 Ω impedance (see 5.2.6).
RX lines may be similarly terminated. In addition for a 4-Wire line, the Bridging position
provides a high-impedance (greater than 50,000 Ω). Bridging is useful where
measurements are made on a circuit between two interconnected pieces of equipment
which are already properly terminated to one another (see 5.2.8).
The terminations provided by the AM5(e) Classic are resistive for AC and open-circuit
for DC. However, for some types of 2-Wire lines, a series capacitance may be inserted
into the circuit through the LINE 2W 2µf selection.
For TX/2W or RX circuits which require a DC path between tip and ring (A and B), the
OFF HOOK selection provides the electronic equivalent of a loading coil. OFF HOOK
provides a DC path of 200Ω or less for DC, while maintaining the AC impedance that
has been selected.
The general guidelines for terminating lines correctly are:
1. For 4-Wire circuits
•Select proper impedance for incoming (RX) line unless the line is already
terminated and the AM5(e) Classic is being used to monitor the line. If
monitoring, set BRDGing and then set the impedance to correspond to the
impedance of the circuit being monitored.
NOTE: Impedance must be set to obtain proper level measurements. This is
necessary because RX level calculation is based on the RX impedance setting. For
more information refer to 5.2.10
•Select proper impedance for outgoing (TX) line.
•If a DC path is required, select OFF HOOK.
2.For 2-Wire circuits
•Select proper impedance for the TX/2W line. To obtain proper level
measurement calculations, set the RX impedance to match since level
measurements are based on the RX impedance settings (see NOTE above).
•If a low-value series capacitance is required, select LINE 2W 2.16µf.
•If a DC path is required, select OFF HOOK.
NOTE: An external instrument connected to the DIAL terminals should also be
set to match the impedance of the TX/2W line that is connected to the TX/2W jack.
No internal AM5(e) Classic circuitry is connected when the DIAL is selected.
5.4TEST CONFIGURATIONS
There are three basic AM5(e) Classic configurations for testing 2-Wire and 4-Wire
circuits.
•End-to-end which requires two AM5(e) Classic units
•Loopback
•Tests with remotely-located responders
The remainder of this sub-section describes these configurations, and their advantages
and disadvantages.
NOTE: End-to-End and responder testing may be performed on either 2-wire or 4-wire lines.
Loopback testing requires a 4-wire line.
5.4.1End-to-End Testing: Measurements on telephone transmission lines are usually
made by applying an appropriate signal (single tone or a sweep of tones) at one end of
the transmission line and measuring the level at the other end of the circuit.
End-to-end testing is the most reliable method for testing near-to-far impairments and
any differences between near-to-far and far-to-near transmission paths. The major
disadvantage is that it requires a test set and an operator at each end.
Figure 5-4 shows an example end-to-end test situation. Somewhere in a telephone
network, a four-wire circuit carries full-duplex communications between a computer
(CPU) and a terminal device (DTE). MODEMs are used at each end to convert the DC
voltage levels that are required by the CPU and the DTE to tones which are transmitted
over the circuits.
If the circuit does not meet the transmission requirement of the MODEMs due to
inadequate frequency response, signal loss, or excessive noise, there will be errors in the
transmission of data. When this occurs, the MODEMs are disconnected and an AM5(e)
Classic is substituted.
Measurements can then be made in either direction by having the AM5(e) Classic at the
CPU-end SEND while having the AM5(e) Classic at the DTE-end MEASURE. After
checking level, frequency response, and noise, the process is then reversed with the
AM5(e) Classic at the DTE-end SENDing.
5.4.2 Loopback Testing: The distant end of a 4-Wire transmission circuit can be
looped back manually by changing cross-connects or by patch cords at a Central Office
or through the use of a remotely-controlled loop-back device such as a Western Electric
Model 829 or an Ameritec AM3-4A Responder. (Responders can do additional tests as
described in 5.4.3).
The AM5(e) Classic at the near end sends (from the TX/2W jack) a signal into the
SND pair which travels to the RCV pair at the far end. At the far end the
loopback connection joins the incoming RCV pair to the outgoing SND pair and
the signal is returned to the near end.
The AM5(e) Classic receives the signal (into the RX jack) from the near-end RCV
pair and measurements are made on the incoming signal. By comparing SEND
levels with MEASured levels, line degradation can be measured.
NOTE: A loopback test cannot be done on 2-Wire circuits because signals can not
be sent and measured simultaneously on the same pair.
The advantage of loopback testing is that only one AM5(e) Classic and one
operator is needed at the test site.
Though loopback testing is more convenient and less costly than end-to-end
testing, it is also less reliable. The disadvantage is that it does not identify
transmission impairments in each direction. In fact, a satisfactory loopback
measurement could actually hide impairments in each direction which cancel
each other out.
To perform a loopback test with a Western Electric Model 829 at the far end and
an AM5 Classic at the near end test site:
1. Connect the TX/2W jack to the SND pair of a 4-Wire circuit, and connect the
RX jack to the RCV pair. Ground the sleeves (or ground terminal) to the CO
ground if Noise-to-Ground tests are required.
2. Turn off SF SKIP, if lighted, then send a brief 2713 Hz tone from the AM5
Classic. (If a different loopback frequency is required, set the QUIET mode
parameter for F4 to that frequency before sending the tone.) This will activate
loopback in the distant Model 829.
(To send the tone, repeatedly press the right SEND key until F4 is flashing.
Then press and hold the right SEND key for over three seconds until F4 is
lighted steady. After about five seconds, press the key again briefly to stop the
tone.)
3. As required, SEND the required test signals and make the desired
measurements on the looped back test signal. (If frequency sweeps are made,
be sure SF SKIP is ON to avoid false commands to the Model 829.)
4. To end the test, repeat step 2 to deactivate loopback in the distant Model 829.
5.4.3 4-Wire Testing with Responder: Most of the advantages of both end-to-end and
loopback testing can be achieved if multifunction responders, such as Ameritec’s
AM3-4A (4-Wire DTMF Commandable V.F Responder) or AM3-2 (2-Wire DTMF
Commandable V.F. Responder), are located at the far end of circuits which may
require testing.
Like 4-Wire loopback testing, the use of a responder eliminates the need for
another AM5(e) Classic and operator at the far end. In addition to loopback tests,
the responder can generate tones which can be measured by the AM5(e) Classic;
therefore, one-way impairments from far-to-near and near-to-far end can also be
measured. Figure 5-5 shows a typical 4-wire test configuration using an AM3-4A
1. Connect the TX/2W jack to the SND pair of a 4-Wire circuit, and connect the
RX jack to the RCV pair. Ground the sleeves (or ground terminal) to the CO
ground if Noise-to-Ground tests are required.
2. Connect a handset (capable of generating DTMF signals) to the DIAL
terminals on the front panel and press the key beside the terminals until DIAL
is lighted. (this disconnects all internal AM5(e) Classic circuitry).
3. Send the required DTMF digits to command the responder into the “milliwatt
mode”; it will send 1004 Hz at 0 dBm. (For this and other responder
commands refer to the AM3-4A Responder Instruction Manual — 18-0025).
4. Press the key beside the DIAL terminals so that MEAS is lighted, then select
MEASURE LEVEL FREQ)..
5. Select the MEASure Display and note the Level and Frequency displayed in
the LED display windows.
6. Repeat steps 2 and 3 with the DTMF digits to command the responder into
loopback mode.
7. Select SEND 1004 Hz (level should be 00.0 dBm) then perform step 5 again to
determine the loopback received Level and Frequency. Compare with the
one-way test to determine impairment in the near-to-far end direction.
8. When finished with the test, repeat steps 2 and 3 with the DTMF digits to
command the responder to “reset” (go into “THRU” data mode).
• LOOPBACK
• QUIET
• SEND TONES
• AMPUTATE
• THRU (Reset)
T R
<SND>
<RCV>
T1 R1
AM3-4A RESPONDER
DT DR
MODEM
DT1 DR1
C
P
U
Digital
Send/Rcv Signals
to DIAL
HANDSET
AM5(E) CLASSIC
MODEM
RX
TX/2W
Modem
Disconnected
Telephone
Network
<RCV>
<SND>
Figure 5-5 – 4-Wire Testing with Responder
5.4.4 2-Wire Testing with Responder: On a 2-Wire network, dial access responders are
very useful in measuring far-to-near impairments. An Ameritec AM3-2C
(Automatic Milliwatt Test Line) or AM3-2A (DTMF Commandable 2-Wire
Responder) can be dialed up over the network by connecting an external handset
to the DIAL terminals of the AM5(e) Classic and selecting the DIAL connection.
Figure 5-6 (shows a 2-Wire network with responders and AM5(e) Classic units at
strategic locations. This configuration allows end-to-end testing (one direction at
a time) between the two AM5(e) Classic units, and allows testing with any
responder in the network from either of the AM5(e) Classic units.
Figure 5-6 — 2-Wire Testing with Responders
A 2-Wire network is typical of a business or residential environment where a user dials
up a distant location (switched access line) through a PBX and/or a Central Office.
The responders are connected to distant telephone numbers just like any other endinstrument. Like any end-instrument, a responder remains on-hook until a ringing
signal is sent to it. It then goes off-hook automatically, (within 2.5 seconds) and applies
a tone (usually 1.004 kHz at -10.0 dBm) to the line. The responder then goes quiet and
waits for DTMF digits which will command it into its various modes of operation or
cause it to go on-hook.
Different DTMF digits can be sent to the responder to command it into its test modes.
The AM3-2C can be used to perform far-to-near tests to determine far-to-near level and
frequency, idle channel noise, and noise with tone.
In addition to the AM3-2C functions, the AM3-2A can also be used to perform far-tonear gain slope and near-to-far level tests. In near-to-far tests the responder receives a
tone from the AM5(e) Classic. Results are then transmitted as a frequency deviation of a
returned signal. This can be observed at the AM5(e) Classic.
Both responders are described in one Instruction Manual, “Model AM3-2A DTMF
Commandable 2 Wire Responder and Model AM3-2C Automatic Milliwatt Test Line
Instruction Manual” (18-0006).
The AM5(e) Classic contains highly versatile detectors and filters which are connected
to the receive line under test. The term MEASURE refers to the detection mode that has
been selected by the MEASURE key to detect and display the characteristics of the
received signal. The labels for the LEDs above the MEASURE key are different for the
AM5 Classic and the AM5e Classic to accommodate the characteristics of IEEE
Standard or CCITT Recommendations. The front panel areas for MEASURE functions
are shown below:
AM5 ClassicAM5e Classic
This section describes all the MEASURE functions in the order they are selected by each
press of the MEASURE key (bottom to top, left to right). Unless otherwise specified, the
descriptions apply to both the AM5 Classic and the AM5e Classic. Views of the LED
indicators and displays showing typical readings are provided. A lighted LED is shown
by shading, as shown by above defaults.
If the LINE selection is 4W , the signal being measured must be connected to the RX
jack (or T1 and R1 terminals).
If the LINE selection is 4W REVerse, 2W, or 2W+2.16µf, the signal being measured
must be connected to the TX/2W jack (or T and R terminals).
For NTG (Noise-to-Ground) measurements, a reference ground must be connected to
the ground terminal on the rear panel or to the sleeve terminal of the RX (or TX/2W)
bantam jack.
Impulse Noise (IMP NOISE) and Impulse Noise with Tone (IMP N W/T) are the only
MEASURE functions that have parameters.
To measure level and frequency, an auto-ranging amplifier, average detector, and
frequency counter are connected to the signal. A 120 kHz low-pass filter is used for full
bandwidth operation. (In the -200 kHz models a 200 kHz filter is used). The detected
average voltage is converted to dBm, based on the impedance selected for the RX line.
To display the measured level and frequency, select MEAS with the DISPLAY key. If
the incoming level is 1004 Hz at a level of –10 dBm, the display will appear as shown
below.
If there is power line interference present on the circuit that could interfere with the
measurement, select the 60 Hz filter by pressing the right FILTER key so the 60 Hz LED
is lighted.
NOTE: None of the filters, selectable by the left FILTER key, apply to the Level
and Frequency measurements; their LEDs are not lighted.
If high frequencies are of no interest, use the narrow band
measurement function described in the next sub-section.
6.2MEASURE LEVEL AND FREQUENCY,
NARROW BAND (L/F 15 kHz or L/F UNWTD)
This is a level and frequency measurement which includes a low-pass filter in the
measurement circuits. It is useful for making measurements on voice-band circuits
where high-frequency components are of no interest or could interfere with
measurements. The filter has the following characteristics:
•AM5 Classic15 kHz low-pass
•AM5e ClassicLow Frequency (Sound) Unweighted 30 Hz to 20 kHz
6.3 MEASURE IDLE LINE NOISE (NOISE)
The Noise selection is a high-gain measurement of level only. It is designed to make
“Idle Line” measurements. Level is displayed in the left display window, and the right
window is unlighted. Noise measurement range and accuracy are as follows:
•AM5 Classic10 to 20 dBrn at ±2 dBrn; 20 to 100 dBrn at ±1 dBrn with CMSG
filter selected.
•AM5e Classic80 to -70 dBm at ± 2 dBm; -70 to +10 dBm at ±1 dBm with
PSHO Filter selected.
6.3.1 Units of Noise Measurement: For noise measurements, the voltage of the
received signal is measured with an RMS detector on an AM5 Classic. For an
AM5e Classic both RMS detection and Quasi-Peak (Q PEAK) detection is
available. Q PEAK detection (and appropriate filtering) is selected with the
FILTER key.
The voltage is first converted to dBm. The RX impedance selected determines the
voltage to dBm conversion.
In an AM5e Classic the dBm measurements are displayed in the left display. For
example, if the noise level were -40 dBm, the display would appear as follows:
In an AM5 Classic, the reading is further converted to dBrn (decibels relative
noise). The dBrn measurement is designed to present a 0 reading for an
optimally quiet line (-90 dBm) and is determined by the formula dBrn = dBm +
90. A –40 dBm noise level would read 50 dBrn on an AM5 Classic as below:
A comparison of dBm and dBrn units of measurement is shown in Figure 3-3
and below:
HOLD TONE
MEAS
HOLD TONE
MEAS
SEND
DISPLAY
dB
dBm
mSEC
SEND
DISPLAY
dB
dBm
d
Brn
mSEC
Relative zero reference for dB readings can be
established at any point. Example below.
0
+50 dB
Intensity of 1 milliwatt of power
-90 -80 -70 -60 -50 -40 -30 -20 -10 0 +10 dBm
KHz
SEC
MIN
CNT
KHz
SEC
MIN
CNT
0 +10 +20 +30 +40 +50 +60 +70 +80 +90 +100 dBrn
-90 dBm for zero reference point of d B rn noise measurements.
6.3.2 Test Setup for Idle Line Noise: For a typical noise measurement, there must be
no signal on the received line. For end-to-end tests using AM5(e) Classic units,
the far-end unit SEND function should be set to QUIET while the near-end unit
MEASURE function is set to NOISE. (Refer to Figures 5-3, 5-4, and 5-5 for
typical locations of equipment).
Where a responder is installed at the far-end, an external handset should be
connected to the DIAL terminals, the DIAL function selected, and the
responder’s number called. When the responder answers, send the appropriate
code to put it into the QUIET mode.
For loopback tests, one AM5(e) Classic is used. The SEND selection is QUIET
and the MEASURE selection is NOISE.
6.3.3 Noise Weighting Filters: The default filter selection is usually used for idleline noise measurements (CMSG for the AM5 Classic; PSHO for the AM5e
Classic). Other filters may be selected by the left FILTER key to “weight” the
measurement over the band of frequencies that is important for the circuit. Noise
Weighting Filters for the AM5 Classic and AM5e Classic are described in detail
in Sections 7 and 8, respectively. A summary of filters will also be found in subsection 3.3.17.
6.4MEASURE NOTCH NOISE (NOISE W/T)
The Noise with Tone test (also referred to as Notch Noise test) measures noise on a
circuit which has a 1004 Hz tone applied to it. (On many types of circuits a signal is
required to hold the circuit active.) The tone is removed in the measuring circuit of the
AM5(e) Classic through a 1010 Hz notch filter, then the residual noise is measured.
Both level and frequency displays operate when this function is selected. (See 6.3 for
units of noise measurement).
6.4.1 Validity of Notch Noise: In order for a Notch Noise measurement to be valid,
the AM5(e) Classic must be receiving a tone as indicated by HOLD TONE being
lighted (green). When HOLD TONE is lighted it indicates that the received
signal is at a frequency of 995 Hz to 1025 Hz at a level greater than –40 dBm.
(individual units may vary slightly in this range). If there is measurable noise,
the noise will be displayed in the left display and the frequency of the hold tone
will show in the right display.
The UNDR (under-range) indication will be displayed, if the noise is below the
measurement range, and the frequency will not be shown:
6.4.2 Test Setup for Notch Noise: For a typical notch noise measurement, there must
be a hold-tone signal on the measured circuit. For end-to-end tests using AM5(e)
Classic units, the far-end unit SEND function should be set to 1004 Hz at a level
sufficient to ensure a received level of greater than -40 dBm, while the near-end
MEASURE function is set to NOISE W/T. (Refer to Figures 5-3, 5-4, and 5-5 for
typical locations of equipment).
Where a responder is installed at the far-end, an external handset should be
connected to the DIAL terminals, the DIAL function selected, and the
responder’s number called. When the responder answers, send the appropriate
code to put it into the “send 1004 Hz” mode.
For loopback tests, one AM5(e) Classic is used. The SEND selection is 1004 Hz
and the MEASURE selection is NOISE W/T.
6.4.3 Noise Weighting Filters: The default filter selection is usually used for NotchNoise measurements (CMSG for the AM5; PSHO for the AM5e Classic). Other
filters may be selected by the left FILTER key to “weight” the measurement over
the band of frequencies that is important for the circuit. Filters are described in
detail in Sections 7 and 8. A summary of filter types will also be found in subsection 3.3.17.
6.5MEASURE NOISE TO GROUND (NTG)
Noise-to-Ground is measured by internally summing the signals on the Tip and Ring
wires on the measured circuit within the AM5(e) Classic, then measuring the noise
through a selected filter, with reference to ground.
The ground reference must be connected through the ground screw terminal on the
rear panel (Figure 3-4) or through the sleeve contact of the RX (or TX/2W if in a 2-Wire
LINE selection) bantam jack.
6.5.1 Noise to Ground Measurement Range: The range of the noise to ground
measurements differs from those shown in sub-section 6.3 in that the range is
shifted toward the high-level end as follows:
•AM5 Classic 40 to 129 dBrn at ±1.5 dBm
•AM5e Classic–50 to +39 dBm at ±1.5 dBm
The NTG level is displayed in the left display window, while the right display
window is not used. If the NTG level is below those shown above, “undr” is
displayed. (Refer to example under 6.4.1). The following shows a 100 dBrn
measurement (on an AM5 Classic). The AM5e Classic would read 10 dBm (not
shown) for the same noise level.
6.5.2 Test Setup for NTG: A noise-to-ground test does not require any external
equipment, but the correct impedance for the line being measured should be set
by the RX key. Though signals may be applied between the Tip and Ring leads
of the RX line, these will not affect the test results unless the line is unbalanced.
6.5.3 Noise Weighting Filters: The default filter selection is usually used for NTG
measurements (CMSG for the AM5; PSHO for the AM5e Classic). Other filters
may be selected by the left FILTER key to “weight” the measurement over the
band of frequencies that is important for the circuit. Filters are described in
detail in Sections 7 and 8. A summary of filter types will also be found in subsection 3.3.17.
6.5.4 NTG and SEND functions: Selection of NTG affects the functions controlled by
the SEND key. For NTG readings, QUIET must be selected, and both SEND
displays will show dashes. QUIET parameters may be changed while in the
NTG position.
If any other left SEND key functions are selected, “nTG-” is displayed in the
right window, and 1004 Hz, VAR Hz, and SWEEP functions are disabled
(regardless of their LED indicator status), and their parameters cannot be
changed.
HOLD TONE
MEAS
S/N
RATIO
DISPLAY
SEND
dB
dBm
d
Brn
mSEC
KHz
SEC
MIN
CNT
The single-tones selected by the right SEND key can still be used to send signals
out of the TX/2W jack. When a single tone is sent, Level and Frequency are
displayed.
6.6SIGNAL-TO-NOISE RATIO (S/N RATIO)
The Signal-to-Noise Ratio measurement is similar to the Noise with Tone measurement.
A 1004 Hz test tone is expected on the measured circuit. The valid signal range is –40 to
+10 dBm.
The receive circuit contains an average detector, a notch filter (previously described
under 6.4), a noise weighting filter, and an RMS detector. The AM5e Classic provides
Quasi-Peak (Q-PEAK) or RMS Detectors which may be switched (along with associated
filters) by the FILTER key.
6.6.1 Signal-to-Noise Ratio Measurements: The range of the S/N Ratio
measurements is 10 to 50 dB with accuracy as follows:
•10 to 40 dB at ±1 dB
•40 to 45 dB at ±2 dB
•45 to 50 dB at ±3 dB
The measurement technique for S/N Ratio incorporates the following steps
which are automatically repeated while the measurement is being made:
1.The average detector measures the amplitude of a received holding tone
(1004 Hz) and keeps this value in memory. This is the “signal” reading.
The incoming signal range is -40 to +10 dB.
2.The 1004 Hz signal is notched out, and the RMS (or Q-PEAK) detector
measures the signal again. This is the “noise” reading.
NOTE: The “noise” reading, could be 0 to 40 dBrn (–90 to -50 dBm) for a –40
dBm holding tone, up to 40 to 80 dBrn (–50 to 0 dBm) for a +10 dBm holding
tone. If the “noise” reading is less than 20 dBrn (–70 dBm), the accuracy of the
computed S/N ratio is reduced to ±2 dBm unless CMSG or PSHO filter is used.
3.A computation occurs which determines the difference, in dB between the
“signal” (step 1) and the “noise” (step 2). The result is the signal-to-noise
ratio.
4.This value is displayed in the left display window as a range of 10 to 50
dB. The holding tone frequency is displayed in the right display window.
For example, a measured S/N ratio of 30 dB and 1004 Hz hold tone
would be displayed as follows:
HOLD TONE
MEAS
HOLD TONE
MEAS
SEND
DISPLAY
For a line that is so quiet that the S/N ratio exceeds 50 dB, the left
window will show “oVEr” as illustrated below:
SEND
DISPLAY
A very noisy line (worse than 10 dB), would show “undr” and the right
window would be blank. The HOLD TONE indicator will be lighted but
kHz is not. Tones outside the notch band are treated as noise and may be
filtered.
6.6.2 Test Setup for S/N Ratio: For S/N Ratio measurement there must be a hold-tone
signal on the measured circuit. For end-to-end tests using AM5(e) Classic units,
the far-end unit SEND function should be set to 1004 Hz at a level of greater than
-40 dBm, while the near-end unit MEASURE function is set to S/N RATIO
(Refer to Figures 5-3, 5-4, and 5-5 for typical locations of equipment).
Where a responder is installed at the far-end, an external handset should be
connected to the DIAL terminals, the DIAL function selected, and the
responder’s number called. When the responder answers, send the appropriate
code to put it into the “send 1004 Hz” mode.
For loopback tests, one AM5(e) Classic is used. The SEND selection is 1004 Hz
and the MEASURE selection is NOISE W/T.
6.6.3 Noise Weighting Filters: The default filter selection for S/N Ratio
measurements (CMSG for the AM5 Classic and PSHO for the AM5e Classic) are
normally used. Other filters may be selected by the left FILTER key to “weight”
the measurement over the band of frequencies that is important for the circuit.
On an AM5e Classic, two of the filter selections also select the Q-Peak detector.
The remaining filters provide RMS detection. Filters are described in detail in
Sections 7 and 8. A summary of filter types will also be found in sub-section
3.3.17.
6.7IMPULSE NOISE WITHOUT TONE (IMP NOISE) or
IMPULSE NOISE WITH TONE (IMP N W/T)
6.7.1 General Description of Impulse Noise Measurement: Impulse noise testing is a
timed study which counts and stores the number of noise pulses that exceed
each of three levels (thresholds). The use of a holding tone is optional for
impulse noise tests, and this description applies to either selection.
NOTE: Information on setup for noise tests and filters is provided in sub-sections
6.7.4 through 6.7.6.
The time over which the test will be run is set by a duration parameter (“dur”).
The lowest level noise threshold is set by the impulse threshold (“iTHd”)
parameter.
The two higher-level thresholds are set by the delta threshold (“idLT”)
parameter. The delta determines the level interval between the thresholds and
can be set at 2, 4, or 6 dB. To control the way that the pulses are counted, a
blanking interval parameter (“bLnC”) is also provided. This parameter is used
by the blanking interval counter.
When a pulse first exceeds a threshold, it is counted. Then the blanking interval
counter starts. While the blanking interval is in effect, no further pulses will be
counted. The purpose of the blanking interval is to minimize the effect of ringing
on the count by providing a time for the ringing to die down.
Without a blanking interval, several counts could be made immediately
after the first pulse, due to secondary pulses caused by ringing. (For more
explanation and a graphic representation, refer to 6.7.3).
After the parameters are set, the measurement display must be selected.
Then the impulse noise test is started by pressing the START/STOP key.
When the key is pressed, the IMPULSE NOISE RUNNING indicator
flashes and a running count is kept of the number of pulses that exceed
each threshold.
NOTE: While the Impulse Noise test is running, the MEASURE key is
inoperative.
While the test is running, the elapsed time (“TiME”), can be displayed
and will count up by .1 minute (6 second) steps. The three impulse noise
counters (“iLo”, “iMid”, and “iHi”) and noise on the circuit (“noiS”) may
be displayed while the test is in progress. Counter and time
measurements are saved after the test is stopped.
The test stops automatically when the time set by the duration parameter
is reached. The test also may be stopped while the test is running by
pressing the START/STOP key.
6.7.2 Impulse Noise Measurements, Parameters, and Procedure: The
following procedure displays the impulse noise measurements (Figure 6-
1) and displays/or sets the impulse noise parameters (Figure 6-2). Finally,
it runs the test:
1.Press the MEASURE key until the IMP NOISE or IMP N W/T
indicator is lighted.
2.Be sure that the DISPLAY key is in the MEASure position.
3.The left display should identify one of the measurements; the
appropriate indicator will also be lighted.
DisplayMeasurementRangeLED
TiMEMinutes 0.1 — 999.9 MIN
iHi High threshold impulse count 0 — 9999 CNT
iMidMid threshold impulse count 0 — 9999 CNT
iLoLow threshold impulse count0 — 9999 CNT
noiSNoise on line (AM5)30.0 — 110.0dBrn
(AM5e)–60.0 — 10.0dBm
NOTE: Time of 0.0 and counts of 0 will be displayed if the test has never been run before.
If tests have been run before, the measurements from the last test will be shown until
reset by starting another impulse noise test. Noise is always displayed as received (even
4The right display will display a time, impulse counter, or a noise
measurement. In Figure 6-1, the example 90 dBrn reading (0 dBm on an
AM5e Classic) would occur if a 1004 Hz “noise” signal at 0 dBm were
being received for a MEASURE selection of IMP NOISE.
That same signal with a MEASURE selection of IMP N W/T would
probably display “undr” if the line was very quiet.
5. To display additional measurements press the up- or down-arrow key.
The identification and the current value will be displayed (refer to step 3
and Figure 6-1).
6. To display the first parameter (duration), press the PARAMeter SELect
7.The left display will identify the parameter and the right display will
show the value. Each time the PARAMeter SELect key is pressed,
additional parameters are displayed as shown below:
KHz
SEC
MIN
CNT
10/31/94
DisplayParameterRange <default>LED
durduration0.1 <15.0> 999.9 MIN
iTHdimpulse (low-level) (AM5)20 <68.0> 110 dBrn
threshold(AM5e )–70 <–22> 20 dBm
idLTimpulse delta 2, <4>, 6 dB
(interval between low, medium
and high thresholds)
bLnCblanking interval 0.0 <125> 255mSEC
8.After all parameters have been displayed, the last selected measurement
will be displayed (Figure 6-1). (See steps 3 and 4).
9.To set a different parameter press the PARAMeter SELect key until the
parameter is identified in the left window.
10.To set a displayed parameter value, use the arrow keys, as needed.
11.Repeat steps 8 through 10 until all values desired have been set. Then
select measurements (step 8).
12.To run the impulse noise test, press the START/STOP key. The IMPULSE
NOISE RUNNING indicator will flash while the test is being run.
NOTE: For Impulse Noise with Tone test, “no TonE” will be displayed and test
will not start if hold tone is not being received by the AM5(e) Classic
13.To observe the measurements while the test is running, press up- or
down-arrow key until the desired measurement is displayed.
14.The test will stop when the duration has been reached, or the test may be
stopped manually by pressing the START/STOP key so the IMPULSE
NOISE RUNNING indicator is not lighted.
15.After the test is stopped, the final measurements should be reviewed by
performing steps 2 – 5.
6.7.3Evaluation of Measurement Data: Although the noise, and counters can
be observed while the test is in progress, the final evaluation should be
made after the test has stopped.
Figure 6-3 graphically shows an impulse noise test. Default parameters
are shown in bold boxes and type, while example final measurements are
shown in light boxes and type.
1.While the test is running, impulse spikes that exceed one of the three
thresholds set the counters as shown by sequence number following each
spike.
2.The blanking interval, shown by cross-hatch, prevents any counts from
being registered for a short period of time after a pulse is received. This is
illustrated in Figure 6-3 by the uncounted spikes within the cross-hatched
areas.
3.Noise is constantly measured. For an AM5 Classic the measurement will
reflect the average RMS measurement in dBrn.
For an AM5e Classic, the noise measurement will reflect the average RMS
noise unless either UNWTD Q PEAK or SWTD Q PEAK filter was
selected for Q PEAK detection. The Q PEAK measurement will be for
peak noise. All AM5e Classic measurements are in dBm.
6.7.4 Test Setup for Impulse Noise without Tone: For a typical impulse noise
without tone measurement, there should be no signal on the measured circuit.
For end-to-end tests using AM5(e) Classic units, the far-end unit SEND function
should be set to QUIET while the near-end unit MEASURE function is set to IMP
NOISE. (Refer to Figures 5-3, 5-4, and 5-5 for typical locations of equipment).
Where a responder is installed at the far-end, an external handset should be
connected to the DIAL terminals, the DIAL function selected, and the
responder’s number called (refer to sub-section 3.3.29). When the responder
answers, send the appropriate code to put it into the QUIET mode.
For loopback tests, one AM5(e) Classic is used. The SEND selection is QUIET
and the MEASURE selection is IMP NOISE.
6.7.5 Test Setup for Impulse Noise with Tone: For Impulse Noise with Tone
measurement there must be a hold-tone signal on the measured circuit.
For end-to-end tests using AM5(e) Classic units, the far-end unit SEND function
should be set to 1004 Hz (at a level sufficient to ensure a received level of greater
than -40 dBm so that the HOLD indicator is lighted).
The near-end unit MEASURE function is set to IMP N W/T (Refer to Figures 5-3,
5-4, and 5-5 for typical locations of equipment).
Where a responder is installed at the far-end, an external handset should be
connected to the DIAL terminals, the DIAL function selected, and the
responder’s number called (refer to sub-section 3.3.29). When the responder
answers, send the appropriate code to put it into the “send 1004 Hz” mode.
For loopback tests, one AM5(e) Classic is used. The SEND selection is 1004 Hz
and the MEASURE selection is IMP NOISE W/T.
6.7.6 Noise Weighting Filters: The default filter selection for impulse noise
measurements (CMSG for the AM5 Classic and PSHO for the AM5e Classic) are
normally used. Other filters may be selected by the left FILTER key to “weight”
the measurement over the band of frequencies that is important for the circuit.
On an AM5e Classic, two of the filter selections also select the Q-Peak detector.
The remaining filters provide RMS detection.
Filters are described in detail in Sections 7 and 8. A summary of filter types will
also be found in sub-section 3.3.17.
This Section describes the filters that are supplied with the AM5
Classic. Except for the 60 Hz filter, which is common to the AM5
and the AM5e Classic, the filters described comply with the
IEEE 743-1984 Standards and are used for noise measurements
described in Section 6 (sub-sections 6.3 through 6.7). The front
15kHz
3kHz
PGM
CMSG
WIDE
50kBit
60 Hz
panel area for FILTER is shown to the right. Default selections
are shown by shading.
NOTE: The default indication for CMSG does not apply to the
default MEASURE LEVEL FREQ selection. The CMSG
default will come on when any MEASURE noise function is
FILTER
selected, then go out when MEASURE LEVEL FREQ or L/F 15 kHz are
selected. No Noise weighting filter can be selected while LEVEL FREQ or L/F 15
kHz are selected.
A low-pass filter (L/F 15 kHz) can be selected by the MEASURE key for
level/frequency measurements only. The response of this filter rolls off frequencies
above 15 kHz (refer to 7.3).
A 995 Hz to 1025 Hz notch filter is automatically switched in when measuring noise
with (1004 Hz) tone, impulse noise with tone, or signal-to-noise ratio (NOISE W/T,
IMP NOISE W/T, S/N RATIO). This filter is not illustrated.
7.1 NOISE WEIGHTING FILTERS
Noise weighting filters are selected by each press of the left FILTER key (bottom to top,
left to right). Other than the lighted LED, there are no displays or parameters
associated with the FILTER selections.
Only one noise weighting filter can be selected at a time, but the 60 Hz filter (refer to
7.2) may be selected in addition to the noise weighting filter to eliminate AC power
interference. The noise weighting selections are summarized below, and the curves for
the filters are shown in the referenced figures.
FILTERDESCRIPTIONIEEE/Bell STANDARDCURVE
CMSGC-Message bandpass743-1984/41009Figure 7-1
PGMProgram bandpass“Figure 7-2
3 kHz3 kHz low pass (3 dB down
The 60 Hz high-pass filter is provided to eliminate frequencies of 60 Hz or lower.
Power line interference due to 60 Hz, 50 Hz or lower components may cause erratic
readings on the display. Connect this filter to eliminate intererence as follows:
1.Press right filter key to connect the 60 Hz filter. 60 Hz LED is lighted.
2.Press right filter key again to disconnect the filter. 60 Hz LED is not
lighted.
7.3L/F 15 kHz MEASURE FILTER
When the MEASURE L/F 15kHz function is selected, a low-pass filter is connected to
the front end of the measurement circuitry, effectively eliminating high-frequency
components. The characteristics of this filter are the same as those shown in Figure
7–4. For more details on MEASURE selections refer to Section 6.
NOTE: As is the case with the MEASURE LEVEL FREQ function, average
detection is used with the MEASURE L/F 15 kHz function.
None of the Noise Weighting Filters, previously described and illustrated in subsection 7.1, will be active (even if previously selected) when L/F 15 kHz or LEVEL
FREQ functions are being used.
This Section describes the filters that are supplied with the AM5e
Classic. Except for the 60 Hz filter, which is common to the AM5
and the AM5e Classic, the filters described comply with CCITT
Recommendations and are used for noise measurements
described in Section 6 (sub-sections 6.3 through 6.7). The front
UNWTD
RMS
SWTD
Q PEAK
UNWTD
Q PEAK
WIDE
2kHz
SWTD
RMS
60 HzPSHO
panel area for FILTER is shown to the right. Default selections are
shown by shading.
NOTE: The default indication for PSHO does not apply to the default
FILTER
MEASURE LEVEL FREQ selection. The PSHO default will come on when any MEASURE
noise function is selected, then go out when MEASURE LEVEL FREQ or L/F UNWTD are
selected. No noise weighting filter can be selected while LEVEL FREQ or L/F UNWTD are
selected.
A low-pass filter (L/F UNWTD can be selected by the MEASURE key for
level/frequency measurements only. The response of this filter rolls off frequencies
above 15 kHz (refer to 8.3).
A 995 Hz to 1025 Hz notch filter is automatically switched in when measuring noise
with (1004 Hz) tone, impulse noise with tone, or signal-to-noise ratio (NOISE W/T,
IMP NOISE W/T, S/N RATIO). This filter is not illustrated.
8.1 NOISE WEIGHTING FILTERS
Noise weighting filters are selected by each press of the left FILTER key (bottom to top,
left to right). Other than the lighted LED, there are no displays or parameters
associated with the FILTER selections.
Only one noise weighting filter can be selected at a time, but the 60 Hz filter (8.2) may
be selected in addition to the noise weighting filter to eliminate power interference.
The noise weighting selections are summarized below, and the curves for the filters are
shown in the referenced figures. The UNWTD and SWTD filters permit selection of
RMS or Quasi-Peak (Q-PEAK) detectors. The other filters use RMS detection.
The Frequency Response for the Unweig hted Filter is within the limits shown in
the above figure. The AM5e Classic implements the low-frequency slope (shown
by the heavy lines) at 18 dB/octave from 20 Hz. The upper-frequency slope is 24
dB/octave from 20 kHz. Response from 20 to 20 kHz is flat within ± 0,5 dB.
The 60 Hz high-pass filter is provided to eliminate frequencies of 60 Hz or lower.
Power line interference due to 60 Hz, 50 Hz or lower components may cause erratic
readings on the display. Connect this filter to eliminate interference as follows:
1.Press right filter key so 60 Hz LED is lighted to connect the filter.
2.Press right filter key again to disconnect the filter; 60 Hz LED is not
lighted.
8.3L/F UNWTD MEASURE FILTER
When the MEASURE L/F UNWTD function is selected, a low-pass filter is connected
to the front end of the measurement circuitry, effectively eliminating high-frequency
components. The characteristics of this filter are the same as those shown in Figure 8-2.
NOTE: As is the case with the MEASURE LEVEL FREQ function, average
detection is used with the MEASURE L/F UNWTD function.
None of the Noise Weighting Filters, previously described and illustrated in subsection 8.1, can be selected when L/F UNWTD or LEVEL FREQ functions are being
used.
The AM5(e) Classic contains a highly-flexible signal generator that is capable of
generating a variety of test tones. The labels for the LEDs above the right SEND key are
different for the AM5 Classic and the AM5e Classic. The front panel areas for SEND
functions are shown below:
N
D
N
D
SWEEP
VAR Hz
1004 Hz
QUIET
SE
SF
SKIP
F4
2713Hz
F3
2804Hz
F2
1004Hz
F1
404Hz
AM5 ClassicAM5e Classic
SWEEP
Hz
1004 Hz
QUIET
SE
SF
SKIP
F4
3004Hz
F3
2004Hz
F2
1004Hz
F1
304Hz
This section describes all the send functions and parameters in the order they are
selected by each press of the SEND keys. Unless otherwise specified, the descriptions
apply to both the AM5 Classic and the AM5e Classic. Views of the LED indicators and
displays showing typical readings are provided. A lighted LED is shown by shading,
as shown by above defaults.
If the LINE selection is 4W, 2W, or 2W +2.16µf, the signal generator will output from
the TX/2W jack (or T and R terminals). If the LINE selection is 4W REVerse, the signal
generator will output from the RX jack (or T1 and R1 terminals). For any LINE
selection, the SEND impedance is set by the left TX/2W key. (Refer to sub-sections 5.2.1
through 5.2.7 and 5.3 for more details)
The DIAL selection (sub-section 5.2.5) will disconnect the signal generator function and
permit the operator to substitute a handset for dialing a remote number, talking, or
listening. DIAL may also be used to provide an“Open” function (sub-section 9.7).
SEND functions are displayed by pressing the DISPLAY key so that the SEND indicator
is lighted (sub-section 3.3.9).
The loudspeaker is connected to the SEND functions when the MONITOR switch is in
the TX position (sub-sections 3.3.7 and 3.3.8).
The left SEND key is a toggle on/off selection of one of the functions above it (subsections 9.1 through 9.4).
The middle SEND key toggles on/off the Signalling Frequency lockout (SF SKIP) to
prevent transmission of frequencies in the SF SKIP range (sub-section 9.5) by any of the
left key functions.
The right SEND key selects or sends four fixed frequencies that are set by default to the
frequencies labelled on the panel. These frequencies may be changed (sub-section 9.6).
The basic QUIET function is no output from the signal generator. Internally, the
generator is disconnected and the line is quiet-terminated by the TX/2W impedance
selection described in sub-sections 5.2.6 and 5.3. While QUIET is selected, the Level
(“LEVL”) parameter of all signal generator functions may be set, and the FREQuencies
for F1, F2, F3 and F4 may be changed.
CAUTION: Level may be changed when any SEND function is selected;
however, this is for convenience to permit level change while a signal is
being output. The LEVL parameter will apply to all SEND functions.
9.1.1 QUIET Data: The display of QUIET data is all dashes, and the SEND LED is
lighted as shown in Figure 9-1.
does not appear on AM5e Classic panel
dB
dBm
dB r n
mSEC
HOLD TONE
DISPLAY
SEND
MEAS
Figure 9–1 — Send QUIET Data Display
KHz
SEC
MIN
CNT
9.1.2 QUIET Parameters: The QUIET parameter displays and default values (for an
AM5 Classic) are shown in Figure 9–2. To select the parameters, press the
PARAMeter SELect key. Each press of the key will display another parameter.
The quiet parameters and their abbreviations, range and default values follow:
Parameter
Description
Level for all SEND
Display
Abbreviation Low <default> HighLED
LEVL- 50.0<00.0> 13.0dBm
functions
Frequencies for
right SEND key
selections.
1. Press PARAM SEL to select and step through QUIET parameters.
PARAM
SEL
2. While an identification
and value are displayed,
use up and down arrow
keys to increase or
decrease value.
3. If needed, press left
arrow key to select
flashing mode and
individual digit or
decimal point.
DO NOT TRY TO
CHANGE PARAMETERS
WHILE NTG OR IMPULSE
NOISE TEST IS RUNNING.
Parameter Identification
Last selection returns to QUIET mode
(all dashes) displays. See Figure 9-1.
dB
dBm
dB r n
mSEC
HOLD TONE
Parameter Value Examples
for AM5 Classic.
VErsion display; no LED's lighted.
KHz
SEC
MIN
CNT
KHz
SEC
MIN
CNT
KHz
LED
lighted
for
frequencies
10/28/94
Figure 9–2 — Send QUIET Parameter Displays
Use the left-arrow key to select a digit or decimal point and/or use the up and down arrow key to
increase or decrease a digit or move the decimal point (see 3.3.14 for general information on
arrow keys). For more details on setting frequency refer to 9.3.2.
9.21004 Hz TONE (1004 Hz)
To select a continuous 1004 Hz sine wave output, press the left SEND key so that 1004
Hz indicator is lighted. The signal generator then outputs a continuous 1004 Hz sine
wave. Figure 9-3 shows the 1004Hz data display and the level parameter. Level is
adjustable after being selected with the PARAMeter SELect key.
This function is useful for tests which require that a hold tone be sent to a remotelylocated AM5(e) Classic. The level should be set so that at least –40 dBm is received at
the remote-end.
The AM5(e) receive circuits provide a notch filter and light the HOLD TONE indicator
when one of the MEASURE Functions that requires a received 1004 Hz tone is selected
(S/N RATIO, IMP N W/T, NOISE W/T). For details refer to sub-sections 6.4, 6.6, and
9.2.11004 Hz Data Display: Figure 9-3 shows a –20 dBm level and 1004 Hz
frequency displayed in the left and right display windows. This figure also
shows the level (LEVL) parameter which may be set with the PARAMeter
SELect key.
9.2.21004 Hz Parameter Display/Adjustment: Figure 9-3 also shows the level
parameter. The level is common to all SEND functions and is shown at a
setting of –20 dBm. The display of level permits changes to be made while the
signal is being sent. Level changes affect all SEND functions.
9.3VARIABLE FREQUENCY GENERATOR (VAR Hz)
The Variable Frequency generator can be selected to send out any frequency within the
range of the AM5(e) unit. This generator is very flexible and may be adjusted while a
signal is being sent. Both level (LEVL) and FREQuency may be set.
9.3.1VAR Hz Data Display: Figure 9-4 shows a 10 dBm level and 1.004 kHz
frequency display. The figure also shows the level and frequency parameter
displays. Both level (LEVL) and FREQuency may be set after they have been
displayed with the PARAMeter SELect key.
9.3.2VAR Hz Parameter Display/Adjustment: The level and frequency parameters
(Figure 9-4) are displayed in sequence as the PARAMeter SELect/EnTeR key is
pressed, then the data is displayed again.
While either parameter is displayed, it may be adjusted with the arrow keys.
The change in generator level or frequency occurs immediately; it is not
necessary to step to the next parameter or data.
The FREQuency change with the up- or down-arrow keys (without using the
flashing-digit mode) and the display depend on the range being set, and
whether the key is pressed momentarily or pressed-and-held for greater than
one (1) second. A value from each FREQuency range is illustrated in Figure 9-
4. The following table shows the fixed leading– zeroes, decimal point, and the
digit (d) which can be changed for each range. To change any single digit of
frequency, press the left-arrow key then use the up- or down-arrow keys in the
flashing-digit mode.
Frequency RangePress OncePress-and-Hold
20 Hz to 99 Hz10 Hz (0.0dx)100 Hz (into next range)
100 Hz to 999 Hz10 Hz (0.xdx)100 Hz (0.dxx)
1000 Hz to 9999 Hz10 Hz (x.xdx) 100 Hz (x.dxx)
10 kHz to 99.99 kHz10 Hz (xx.xd)100 Hz (xx.dx)
100 kHz to 120 kHz
(200 kHz for –200 models)
100 Hz (xxx.d)1 KHz (xxd.x)
NOTE: If a value is set outside the range of the unit, it will be automatically set to the closest
limit after PARAMeter SELect is pressed to display the data.
9.4SWEEP GENERATOR (SWEEP)10/94 /28 (Fri) 17:03 to here.
When SWEEP is selected, the signal generator steps through a series of frequencies. The
programmed defaults starts the sweep at 304 Hz and increases in 100 Hz steps at a
three (3) -second rate until 3.004 kHz. When the 3.004 kHz stop frequency is reached,
the sweep immediately begins again, unless the zero delay is changed.
To provide an audible reference to the beginning and end of the sweep, a delay may be
programmed so that the first and last tone will have a longer duration than the other
tones.
The SWEEP function is useful for performing step-frequency tests to measure the
response/bandwidth of a circuit. A slow sweep rate within the voice-frequency range
allows an operator at the far end to note the levels as the tones are received.
9.4.1SWEEP Data Display: Figure 9-5 shows a 0 dBm level and 1.904 kHz frequency
display. The frequency will be increasing in steps while the data is displayed.
The figure also shows the default parameter displays.
If the SF SKIP function (see 9.5) is turned ON, frequencies in the rejection band
of the SF SKIP cannot be displayed or output if they are in the range of a
SWEEP.
Any of the single tones (F1 through F4) may be selected or sent during the
sweep (see 9.6). The sweep is interrupted while the single tone is sent. When
the tone is stopped, the sweep begins again at the start frequency.
9.4.2SWEEP Parameter Display/Adjustment: The level, start frequency, stop
frequency, step frequency, and delay parameters (Figure 9-5) are displayed in
sequence as the PARAMeter SELect/EnTeR key is pressed, then the data is
displayed again. The following table shows the abbreviations, range, and
default values for the SWEEP parameters:
Parameter
Description
Display
AbbreviationLow <default> HighLED
Level for all SEND
functions
LEVL- 50.0<00.0> 13.0dBm
Start Frequency
Stop Frequency
Step Frequency
(High-limit is 200.0
kHz for AM5(e)
Classic-200)
STrT
SToP
STEP
0.020<0.304>120.0
0.020<3.004>120.0
0.020<0.100>120.0
200.0
kHz
Sweep RaterATE000.1<003.0>999.9SEC
Delay at start and stop
Frequency
dLAY
000.0<000.0>999.9
SEC
NOTE: The finest resolution of sweep rate and delay is .24 seconds (based on change over
frequency) Due to this fact, the AM5(e) Classic will round the entered value to the nearest 0.24
second
Parameter changes take effect as soon as they are made. If a start or stop frequency is
set within the SF SKIP rejection band (see 9.5), the start or stop frequency will be moved
to the lower or upper frequency limit of the rejection band (depending on which is
closest) when SF SKIP is turned on. (Frequencies are not readjusted after SF SKIP is
turnedoff.).
Frequency steps that fall within the SF SKIP rejection band are never sent if SF SKIP is
on. They will be sent as soon as SF SKIP is turned off.
Press to select and step
through parameters
PARAM
SEL
DISPLAY
Selects flashing mode.
Increase or
decrease
digit value.
To change any
single digit of
frequency or time,
press the leftarrow key then use
the up- or downarrow keys in the
flashing-digit
mode.
STrT and SToP
will be set to
reject band
limits when SF
SKIP is turned
ON, if they are
in reject band.
Actual value of
rATE and dLAY
(over 000.0) is
rounded to the
nearest 0.24
seconds of
set/displayed
value.
SEND
MEAS
Display of VAR Hz level and frequency
unless a VAR Hz parameter is selected.
dB
dBm
dB r n
mSEC
HOLD TONE
dBm
LEVeL affects all SEND functions
dBm
(FREQUENCY steps up)
KHz
SEC
MIN
CNT
KHz
KHz
KHz
SEC
10/31/94
Figure 9–5 — Send SWEEP Data and Parameter Display
The SF SKIP is used to avoid sending tones in the “signalling frequency” band that
could interfere with other equipment. SF SKIP is an ON/OFF toggle that is affects the
VAR Hz or SWEEP functions.
When SF SKIP is ON (LED lighted), the VAR Hz generator cannot be set to send
frequencies in the rejection band. If a frequency is set in the band while SF is OFF, the
frequency will be reset to the upper or lower limit of the rejection band when SF SKIP is
turned ON. The frequency will not be reset when SF SKIP is turned OFF.
When SF SKIP is ON, the SWEEP generator start or stop frequencies cannot be set (or
remain) within the rejection band. If the SWEEP range includes the SF SKIP rejection
band, those frequencies will not be output. The rejection band is different for the AM5
and AM5e Classic.
QUIET Frequencies (F1, F2, F3, or F4) can be set within the rejection band when SF
SKIP is ON. This allows the user to momentarily overide the SF SKIP band to send
control tones.
9.5.1SF SKIP Rejection for AM5 Classic: The rejection band for the
AM5 Classic is 2450 Hz through 2750 Hz to conform to
IEEE standards.
9.5.2SF SKIP Rejection for AM5e Classic: The rejection band
for the AM5e Classic is 2130 Hz through
2430Hz to conform to CCITT standards. AM5 Classic
9.6 SINGLE FREQUENCY TONES (F1, F2, F3, F4)
9.6.1General Information: The single-frequency tones are
selected by pressing the right SEND key until the LED
for the tone is flashing. Once the LED is flashing, the tone
may be output by pressing and holding the right SEND
key for three seconds until the LED is steady. The tone is
stopped by pressing and releasing the SEND key.
The single-tone over-rides any other SEND function. If a
SWEEP frequency is being generated, the SWEEP is reset to its
starting frequency after the tone is stopped.AM5eClassic
9.6.2 The single-tone may also be over-ridden by pressing any of the left SEND keys.
For example, if an F4 tone is being output, shifting from QUIET to 1004Hz will
automatically stop the F4 tone and send the 1004 Hz tone.
The frequency labels on the front panel are the defaults that have been chosen
for various control or test purposes and are different for the AM5 and AM5e
Classic. Frequencies and level may be changed when QUIET is selected, refer to
9.1.
One use of a single-frequency tone is to command a remote device into loopback
mode. This is described in the next sub-section
This Section consists of the specifications for all models of the AM5 Classic. It is in a
tabular format (beginning on Page 10-2) , and should be consulted for an overview of
each model’s capabilities. A left arrow (“ <———”) indicates specifications identical to
previous columns.
Signal Range–40 to +10dBm<———<———
Noise Range10 to 70dBrn
(20 to 70dBrn
@ 135/150 Ω
–80 to –20dBm
(–70 to –20dBm
@ 135/150 Ω)
<———
Ratio Range10 to 50dB<———<———
Accuracy± 1dB @ 10 to 40dB
± 2dB @ 40 to 45dB
<———
<———
± 3dB @ 45 to 50dB
Noise below –70dBm
reduces accuracy to
±2dB except when
usin
PSHO filter.
<———
IMPULSE NOISE (3 LEVEL)
PER IEEE 743 AND CCITT 0.71
<———<———
TransmitterQUIET termination
Receiver
Minimum
Threshold
30dBrn
–60dBm
<———
Threshold
Difference2, 3, 4 or 6dB<———<———
Threshold
Accuracy± 1dB<———<———
Timer.1 minute to
999.9 minutes or
<———<———
Continuous
Maximum
Count
(3 Counters)
Dead Time1 to 255ms
high9999
mid 9999
low9999
<———<———
<———<———
(125ms preset)
November 19, 1997Page 10-5
Page 96
AM5(e) Cl
l(18-0047)SPECIFICATIONS
panel. Optional front panel “Siemens” jacks.
assic Instruction Manua
Technical Specifications (Sheet 5 of 5)
GENERAL SPECIFICATIONS
AM5/AM5e
Classic-200
Specification
AM5 Classic-120AM5e Classic-120
Models
Input4-Wire
4-Wire Reverse
<———<———
2-Wire
2-Wire with 2.16 µf
Termination
Receive
Impedance
135 Ω, 150 Ω,
600 Ω, 900 Ω,
1200 Ω<———<———
High
Impedance
Bridging
Transmit
Source
Impedance
50,000 Ω
Bridging Loss is less
<———<———
than . 2dB
135 Ω, 150 Ω,
600 Ω, 900 Ω,
1200 Ω<———<———
DC Blocking 200 VDC<———<———
Balance> 90dB @ 50Hz –
120Hz. Decreases
<———<———
6dB/octave above
120Hz
Hold Circuit
Independent
Transmit and
Receive
DC = 200 Ω
AC = > 20,000 Ω
<———<———
Monitor
Built-in
Speaker/Volume
Control Selection
Power115 VAC or
Transmit
Receive
Measure
<———<———
<———<———
230 VAC 50/60Hz
Weight
with batteries
Optional Internal Sealed Lead-Acid Battery with Recharger for all
6 lbs
10 lbs
2.72 kg
4.54 kg
Dimensions
Width
Height
Depth
Operating
Temperature
8.3 in.
3.5 in.
12.1 in.
0 to 50 Degrees
Celsius
210 mm
89 mm
307 mm
<———<———
Line
Connections
Dual Miniature Phone Jacks (mate with ADC PJ777 or
Switchcraft TT253) on front panel. Screw Terminals on rear
November 19, 1997Page 10-6
<———
<———
Page 97
AM5 (e) Classic Instruction Manual (18-0047)WARRANTY, SERVICE AND CALIBRATION
November 19, 1997 Page 11-1
11.WARRANTY, SERVICE AND CALIBRATION
11.1 WARRANTY
AMERITEC Corporation warrants that its electronic instrument products are manufactured to the
highest commercial standards and are free from any defects in material or workmanship.
For a period of one (1) year from shipment, AMERITEC will repair, without charge to the original
purchaser, any unit which upon inspection by AMERITEC proves to be defective.
This warranty is the sole warranty offered by AMERITEC and is in lieu of all other obligations or
liabilities, including claims of consequential damage; however, an EXTENDED WARRANTYPLAN may be purchased. For information contact an Ameritec Sales Representative.
11.2 SERVICE POLICY
AMERITEC products are designed with plug-in printed circuit boards and modular
assemblies. Once a problem is localized, service is accomplished by PC board (or module)
replacement.
11.3 CALIBRATION POLICY
All AMERITEC products are manufactured to commercial standards and are calibrated with
equipment traceable to NIST (National Institute of Standards and Technology). With the
exception of component failures or abuse, AMERITEC instruments are designed to maintain
compliance with their published specifications throughout their service life.
While periodic calibration verification is normally not required, in critical applications it is
recommended that verification be accomplished annually.
Calibration verification is most efficiently accomplished by return of the equipment to the
AMERITEC factory where specialized test equipment is used. Field calibration verification is not
supported by AMERITEC.
Page 98
AM5 (e) Classic Instruction Manual (18-0047)WARRANTY, SERVICE AND CALIBRATION
November 19, 1997 Page 11-2
11.4 RETURN OF UNIT
In the event of a malfunction, call or write to the AMERITEC factory and obtain a return
authorization number. Return the unit to AMERITEC freight prepaid with a note (in-warranty
repair) or a Purchase Order for the repair (out-of-warranty repair) listing the following
information:
•Return authorization number from AMERITEC.
•Return shipment address of purchaser.
• Name and telephone number of person at purchaser's location familiar with the
problem.
• Brief description of problem (include any printouts that may have a bearing on the
problem, if possible).
•Terms of payment for repair costs (out-of-warranty unit).
The unit will be repaired and returned freight-prepaid for units in warranty and freight-collect for
units out-of-warranty. As stated above, a Purchase Order to cover the cost of repair must
accompany any out-of-warranty return of the unit to AMERITEC.
Field Calibration and validation procedures AM5-C-120, and AM5-EC-120, and
AM5-C-200 with software versions 105A or later. Refer to the AM5-Classic
Instruction Manual Section 4.6, Calibration Procedure, to help identify the software
version the unit currently has installed.
A.1.2. EQUIPMENT REQUIRED
1. A highly accurate AC voltmeter (up to 4 decimal places, +/- 0.1% at 1 KHz)
2. Bantam to Minigator and/or Bantam to Bantam Cables.
3. A 600 ohm load resistor, accurate to 1% tolerance (1/4 watt).
A.1.3. CALIBRATION PROCEDURE
TRANSMIT AND RECEIVE CALIBRATION
1. Remove all input leads.
2. Connect T1 and R1 leads of the AM5- Classic of either the front right bantam
jack or rear terminal screws to the AC Voltmeter.
3. Set the voltmeter to measure AC voltage.
4. Power up the unit to be calibrated and verify it is in [SEND] and [QUIET]
modes.
5. Press [PARAM SEL] key until the Software Version number appears.
6. Press and hold down [START/STOP] key for 3 seconds until “FULL CAL?”
appears on the screen.
7. Press the [ENTER] key (left arrow pointing key) to start full calibration.
Note: To abort “FULL CAL?” press any key other than ENTER].
8. The display will read V1 on the left and 3850 on the right. 1004 Hz LED
will come on.
9. Enter the total voltage in millivolts, by using the UP/DOWN arrow key to
increas/decrease the V1 value (or hold down the key to go faster). Voltage
should be larger than 3431 mV and smaller than 4320 mV. Press the
[ENTER] key. Any voltage entered outside the lower and upper ranges will
display “DATA EROR” and any key pressed will abort calibration.