Alaris 7130, 7230 Service manual

Page 1
Models 7130/7131 and 7230/7231
Technical Service Manual
Signature Edition®GOLD Infusion System
A
SEC
PRI
Clear
Enter
OPT PRI HLD SEC KVO
1
RUN
HOLD
POWER
OPTIONS
23
4
56
7
8
9
.
0
ml/hr
RUN
HOLD
A B
HLD PRI SECKVO OPT
ml/hr
1
4
7
PRI
PRI SEC HLD
SEC
2
5
8
0
Clear
3
6
9
RUN
HOLD
OPTIONS
Enter
ml/hr
OPT KVO
B
Page 2
This Technical Service Manual is subject to change without notification.

GENERAL CONTACT INFORMATION

ALARIS Medical Systems, Inc. 10221 Wateridge Circle San Diego, California 92121
Customer Advocacy - North America
Clinical and technical feedback.
Phone: (800) 854-7128, Ext. 7812
E-Mail: CustomerFeedback@alarismed.com
Technical Support - North America
Maintenance and service information support; troubleshooting.
United States:
Phone:
(858) 458-6003 (800) 854-7128, Ext. 6003
Canada:
Phone:
Eastern: (800) 908-9918 Western: (800) 908-9919
Customer Care - North America
Instrument return, service assistance, and order placement.
United States:
Phone:
(800) 482-4822
Canada:
Phone:
(800) 387-8309
Technical Support and Customer Care - UK
Maintenance and service information support.
Instrument return, service assistance, and order placement.
Customer Service:
Freephone: 0800 917 8776 Fax: 01256 330 860
Technical Support:
Freephone: 0800 389 6972
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 3
Chapter 1 - General Information
1.1 Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Precaution Definitions
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4 Accessories
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-2
1.4.1 Nurse Call (7130/7230 only)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4.2 Learn/Teach RS-232 Cable
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4.3 Flow Sensor
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.5 Alarms, Errors, Messages
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-7
1.5.1 Silencing Alarms
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6 Battery Management System
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.1 Fan
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.2 Battery and Charging Process
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.3 Refresh Cycle
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.6.4 Battery Gauge
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
1.6.5 Power On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
1.6.6 Lower LCD Display
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
1.6.7 Clock
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
1.6.8 Battery Maintenance
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
1.7 NiCad Battery Capacity Information
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
1.8 Dynamic Monitoring
®
System (DMS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
1.9 Data Communications Function
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
1.10 Trumpet and Start-Up Curves
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Chapter 2A - Checkout and Configuration (Software Versions 4.06 and 4.08)
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-1
2.2 New Instrument Checkout
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-1
2.3 Configurable Options and Defaults
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-1
2.4 Configuration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-2
2.4.1 Entering Configuration Mode
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-3
2.4.2 Setting to Defaults
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-3
2.4.3 Regional Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-4
2.4.4 Setting Air-in-Line Threshold
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-4
2.4.5 Profiles
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-5
2.4.6 Setting Maximum Rate
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-6
2.4.7 Setting Computer Link
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-6
2.4.8 Setting Optional Modes
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-7

TABLE OF CONTENTS

i
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 4
Chapter 2A - Checkout and Configuration (Continued)
2.4.9 Setting Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-8
2.4.10 Setting KVO Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-9
2.4.11 Setting Dynamic Monitoring
®
System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-10
2.4.12 Setting Audio Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-12
2.4.13 Setting Configuration Name (Instrument ID Label)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-13
2.4.14 Resistance Options
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-14
2.4.15 Pressure Options
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-15
2.4.16 Manual Baseline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-16
2.5 Transferring Settings to Another Instrument
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-17
2.5.1 Learn/Teach Instrument Procedure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-18
2.5.2 Pop-Up Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2A-19
Chapter 2B - Checkout and Configuration (Software Version 2.78)
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-1
2.2 New Instrument Checkout
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-1
2.3 Configurable Options and Defaults
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-1
2.4 Configuration Procedure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-2
2.4.1 Entering Configuration Mode
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-3
2.4.2 Setting to Defaults
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-3
2.4.3 Setting Language
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-4
2.4.4 Setting Air-in-Line Threshold
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-4
2.4.5 Setting Dose Rate Drugs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-5
2.4.6 Setting Maximum Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-7
2.4.7 Setting Computer Link
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-8
2.4.8 Setting Optional Modes
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-9
2.4.9 Setting Optional Features
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-10
2.4.10 Setting KVO Rate
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-10
2.4.11 Setting Dynamic Monitoring
®
System Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-11
2.4.12 Setting Audio Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-14
2.4.13 Setting Configuration Name (Instrument ID Label) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-14
2.5 Transferring Settings to Another Instrument
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-15
2.5.1 Learn/Teach Instrument Procedure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-15
2.5.2 Pop-Up Displays
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-16
2.6 Resistance Options
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-17
2.7 Pressure Options
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2B-18
ii
TABLE OF CONTENTS
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 5
TABLE OF CONTENTS
Chapter 3 - Preventive Maintenance
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Preventive Maintenance Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2.1 Regular Inspection
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2.2 Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.2.3 Flow Stop Test
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.2.4 Rate Calibration Procedure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.2.5 Post Calibration Rate Accuracy Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.6 Pressure Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.2.7 Ground Current Leakage Test
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.8 Ground Resistance Test
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.9 Battery Refresh Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.10 Reset Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.2.11 Reset PM Due
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.3 Storage and Cleaning
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.3.1 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Chapter 4 - Principles of Operation
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.3 Overview
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.4 Main PCB
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.4.1 CPU Kernel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.4.2 Combo IC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.4.3 EEPROM
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.4.4 RAM
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.4.5 Flash EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.4.6 RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.5 Power System
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.5.1 Battery Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.5.2 AC Off-Line Switcher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4.5.3 Battery Charge Regulator
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4.5.4 Refresh Cycle Load
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.5.5 VAO Shutdown
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.5.6 AC Line Sense
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.5.7 System Power Source Select
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.5.8 Battery Voltage Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
iii
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 6
Chapter 4 - Principles of Operation (Continued)
4.5.9 VMEAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.5.10 Voltage Reference 4.1V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.5.11 System Current Monitor
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.5.12 Always On Supply (+5VAO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.5.13 System Switching Supplies
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.5.14 VRAM Supply
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.5.15 VPOS Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.5.16 Battery Temperature Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.5.17 System Watchdog
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4.5.18 Power Switch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
4.5.19 System Reset/Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4.5.20 Lower LCD Display Backlight Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4.6 Motor Drive/Sensors
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.6.1 Motor Drive
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.6.2 Air-in-Line Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
4.6.3 Transducer
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
4.7 User Interface
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
4.7.1 Main Speaker Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
4.7.2 Backup Audio Buzzer and Test Circuit
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
4.8 LED Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
4.9 Lower LCD Display
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4.10 Main LCD Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4.10.1 Main LCD Backlight
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.10.2 Graphic LCD Contrast
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.11 Nurse Call Circuit (7130/7230 only)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-25
4.12 Panel Lock Switch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.13 ECD Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Chapter 5 - Corrective Maintenance
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Disassembly/Reassembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.1 Removing Battery
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.2 Separating Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.2.3 Removing Power Supply Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
5.2.4 Removing ECD and RS-232 Board Assemblies
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
iv
TABLE OF CONTENTS
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 7
TABLE OF CONTENTS
Chapter 5 - Corrective Maintenance (Continued)
5.2.5 Removing Line Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
5.2.6 Removing Speaker Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.2.7 Removing Pole Clamp
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
5.2.8 Disconnecting Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5.2.9 Removing Main Board
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5.2.10 Removing LED and LCD Modules
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-26
5.2.11 Removing Mechanism
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-27
5.2.12 Removing AIL Transmitter (Arm)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
5.2.13 Removing AIL Receiver (Button) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5.2.14 Removing Seal Clip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
5.2.15 Removing Mechanism Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
5.2.16 Removing Keypad Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
5.2.17 Routing and Connecting Cables
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
5.3 Test and Calibration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-38
5.3.1 Power-On Self Test
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-38
5.3.2 Mechanism Visual Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-38
5.3.3 Mechanical Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-42
5.3.4 Pressure Verification and Calibration Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-43
5.3.5 Set Sensor Check
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-43
5.3.6 Test Run Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-44
5.3.7 Hard Pressure Cal Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-45
5.3.8 Checking Pressure Calibration Set
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-46
5.4 Level of Testing Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-47
Chapter 6 - Troubleshooting
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Diagnostics Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.2.1 Entering Diagnostics Mode
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.2.2 Setting Preventive Maintenance Interval
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
6.2.3 Viewing Alarm or Event History (Event Log)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.2.4 Setting Time (and Date)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.2.5 Viewing Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.2.6 Changing Rated Capacity of Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
6.2.7 Viewing DC Voltages
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.2.8 Setting ID Number
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.2.9 Viewing Battery and Total Run Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.2.10 Viewing Self-Check Timer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
v
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 8
Chapter 6 - Troubleshooting (Continued)
6.2.11 Testing Channel Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6.2.12 Viewing/Changing Rate Calibration Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6.2.13 Testing Main LCD
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.2.14 Testing Aux (Lower) LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.2.15 Testing Switches
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.2.16 Changing Main LCD Contrast
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.2.17 Calibrating Channel Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.2.18 DAC Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
6.2.19 Configuring Pressure System Auto Zero
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Chapter 7 - Illustrated Part Breakdown
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.3 Parts List
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.4 Ordering Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
List of Figures
1-1 Battery Refresh Cycle
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1-2 Lower LCD Display Layout
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
1-3 Resistance Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
1-4 Pressure and Resistance Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
2-1 Map of Configuration Screens (Software versions 4.06 and higher) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22A
2-2 Map of Configuration Screens (Software version 2.78)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23B
3-1 Rate Accuracy Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-2 Pressure Test Setup
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
4-1 Main Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-2 COMBO IC Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3 Electrical Partitioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4-4 Battery Manager Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-5 Battery Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4-6 Main Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4-7 System Watchdog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4-8 System Reset/Power On
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
4-9 Motor Drive Circuit, Phase 1(A)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
4-10 Motor and Mechanism Sensors Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
4-11 Air-in-Line Detector Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
4-12 Pressure Sensor Interface Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
vi
TABLE OF CONTENTS
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 9
TABLE OF CONTENTS
List of Figures (Continued)
4-13 User Interface Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
4-14 Main Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
4-15 Backup Audio
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-16 Lower LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-17 Flow Sensor Interface Block Diagram
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
5-1 Instrument Assembly Organization
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-2 View From Back of Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-3 Alternate View From Back of Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5-4 Cap Handle Screws
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5-5 Hidden Case Screw in Battery Compartment
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5-6 Case Screw and Pole Clamp Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-7 Separating Case Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5-8 Locking Bar
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5-9a Torque Sequence (Single)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5-9b Torque Sequence (Dual) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5-10a Removing Power Supply Board Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-11
5-10b Disconnecting Cable
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5-11 Rear Case ECD Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5-12 RS-232 Cover
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5-13 RS-232 Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
5-14 Line Filter Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
5-15 Speaker and Fan Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5-16 Pole Clamp Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
5-17 Retaining Ring Installation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
5-18a Rear Case Cable Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5-18b Front Case Cable Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5-18c Front Case Cable Routing
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5-19 Removing Main Board
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
5-20 LED / LCD Modules and Snap Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5-21 Mechanism Latch in Middle Position
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5-22 Remove Mechanism
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5-23 Unlocking AIL Gear Tab
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5-24 AIL Transmitter in Open Position
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5-25 AIL Transmitter in Closed Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5-26 Flat Edge of AIL Receiver
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5-27 Faceplate Hook
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5-28 Mechanism Latch in Middle Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
vii
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 10
List of Figures (Continued)
5-29 Apply RTV Under Edges of Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
5-30a Front Case Cable Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-35
5-30b Front Case Cable Routing
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36
5-31 Rear Case Cable Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37
5-32 Mechanism Assembly Spring Location:
Between motor plate and clamp arm
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40
5-33 Mechanism Assembly Spring Location:
Between back guide and clamp arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40
5-34 Detail of Proper Mechanism Spring Position
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-41
5-35 Spring Installation Tool Position
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-41
5-36 Leak Test Setup
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-42
5-37 Run-In Set
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-44
5-38 Transducer Pot
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-45
6-1 Map of Diagnostic Screens
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
7-1 Power Cord Wrap Kit (Item 403)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-2a Case Assembly, Single Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
7-2b Case Assembly, Dual Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
7-3a Front Case Assembly, Single Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7-3b Front Case Assembly, Dual Channel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
7-4 Mechanism Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
7-5a Rear Case Assembly, Single Channel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
7-5b Rear Case Assembly, Dual Channel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7-6 Pole Clamp Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
7-7 Label / Literature Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
List of T
ables
1-1 Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-2 Battery Trip Points
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1-3 Abbreviations, Acronyms, Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
2-1 Record of Configured Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20A
2-2 Drug List (7130/7230 only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6B
2-3 Record of Configured Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20B
2-4 Drug List Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22B
3-1 PM Inspections
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-11
4-1 Motor Control Signals
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
5-1 Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
viii
TABLE OF CONTENTS
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 11
TABLE OF CONTENTS
List of Tables (Continued)
5-2 Level of Testing Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-47
6-1 Technical Troubleshooting Guide
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 Error Messages
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6-3 Battery Manager Error Codes (not in Alarm History) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
7-1 Other Parts
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7-2 Case Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7-3 Front Case Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
7-4 Mechanism Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-11
7-5 Rear Case Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
7-6 Pole Clamp Assembly
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15
7-7 Label / Literature
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
7-8 Packing Materials
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
ix
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 12
x
TABLE OF CONTENTS
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 13
1
GENERAL INFORMATION
Page 14
1.1 INTRODUCTION
This manual covers Signature Edition
®
GOLD Infusion Systems, with software versions 2.78, 4.06 and 4.08. It is used in conjunction with an applicable Signature Edition
®
GOLD Infusion System Directions
for Use (DFU).
This manual contains instructions for maintenance, repair, and configuration of the instrument. It is intended for personnel experienced in the analysis, troubleshooting, and repair of analog/digital microprocessor-based electronic equipment.
If the instrument requires service while under warranty, it is to be serviced only by ALARIS Medical Systems authorized service personnel. Refer to the “Service Information” and “Warranty” sections of the applicable Signature Edition
®
GOLD
Infusion System DFU.
The Signature Edition
®
GOLD Infusion
System includes:
Single-Channel, Models 7130/7131
Dual-Channel, Models 7230/7231
AccuSlide
®
Flow Regulator
administration sets
The Signature Edition
®
GOLD Infusion System series includes the following configurations:
7130B, 7130D, 7130E
7131A, 7131B
7230B, 7230D, 7230E
7231A, 7231B
Chapter 1 — GENERAL INFORMATION
1-1
Do not use sharp objects (pens, pencils, etc.) to activate switches, as this will damage the keypad.
Any attempt to service an ALARIS Medical Systems instrument by anyone other than an authorized ALARIS Medical Systems Service Representative while the instrument is under warranty may invalidate the warranty.
CAUTION
CAUTION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 15
1.1 INTRODUCTION (Continued)
7130/7131 and 7230/7231 Series:
is a 100-240 VAC, 50/60 Hz instruments family that supports both single and dual channel fluid delivery
features user-interactive software
displays prompts, alarms and alert messages, and troubleshooting information on main LCD display
can be configured to specific operational requirements
allows upgrades for future instrument enhancements
has been designed to interface with accessory equipment.
7131/7231 Series key differences from 7130/7230 Series are:
HARDWARE
The 7131/7231 Series is labeled for 220V with two power cord options and has isolated RS-232 Board, potential equalization (PE) connector, and drop sensor board installed.
There is no Nurse Call option.
Keypads, other than English, have symbols instead of words.
SOFTWARE
The 7131/7231 Series has a drug list only if the software version is 4.08 or higher and profiles are enabled. Some defaults are different in configuration mode and there are several languages to choose from in version 2.78.
Refer to the applicable Signature Edition
®
GOLD Infusion System DFU for complete setup and operation information.
1.2 PRECAUTION DEFINITIONS
A is an alert to a potential hazard which could result in serious personal injury and/or instrument damage if proper procedures are not followed.
A is an alert to a potential hazard which could result in minor
personal injury and/or instrument damage if proper procedures are not followed.
1.3 SPECIFICATIONS
Refer to Table 1-1.
1.4 ACCESSORIES
Accessory items are available for use with the instrument. These items are described in the following paragraphs.
1.4.1 Nurse Call (7130/7230 only)
All instruments are equipped with the nurse call feature. Alarms and some alerts from the instrument will be relayed to the facility’s existing nurse call system. No operating features of the instrument are changed and it will alarm with or without the nurse call. The only additional item needed is a cable with a 9-pin to mono phone jack (ALARIS P/N 136111).
1.4.2 Learn/Teach RS-232 Cable
This is a standard commercially available 9-pin Null Modem RS-232 cable (ALARIS P/N 133450). The Learn/Teach RS-232 cable is used to connect two instruments for the purpose of transferring (downloading) configuration data from/to another instrument.
NOTE: Guardrails®Safety Software data sets cannot be transferred via the Learn/Teach function. They must be downloaded directly from a PC.
GENERAL INFORMATION
1-2
WARNING
CAUTION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 16
GENERAL INFORMATION
1-3
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Administration Sets:
Use only ALARIS Medical Systems®72 Series administration sets. All disposable IV set and IV set accessory models are defined on a separate card.
Air-in-Line Accuracy:
Alarms:
Accumulated Air in Line
Air in Line
Battery Depleted
Channel Malfunction
• Computer Link Failure
Flow Sensor Unplugged
• Hold Time Exceeded
Instrument Malfunction
Key Stuck
Latch Open
No Upstream Flow Detected
Occlusion Downstream
Occlusion Upstream
Primary Flow Detection During Secondary
Set Out
Set Up Time Exceeded
Altitude:
Operating Altitude: -500 ft. (-150m) to 7,500 ft. (2285m)
Battery:
Rechargeable nickel-cadmium (NiCad). Use only NiCad 12V, 1.8 ampere-hour (Ah) (minimum) batteries. A single channel instrument will operate for 4 hours nominal and a dual channel instrument will operate for 3 hours nominal, under the following conditions:
• new, fully-charged battery
• ambient room temperature 73 ± 7°F (23 ± 4°C)
• resistance monitoring modes
• rate: 100 mL/h on a single channel instrument and 50 mL/h on each channel of a dual channel instrument.
Battery run time is affected by operating mode, rate, monitoring options, and back pressure.
Table 1-1. Specifications
Configured Threshold Air Volume Detection Range
50µL 15µL - 85µL 100µL 35µL - 140µL 200µL 100µL - 235µL 500µL 275µL - 565µL
Page 17
GENERAL INFORMATION
1-4
Table 1-1. Specifications (Continued)
Bolus Volume Limits:
NOTE: When the occlusion alarm pressure limit is set to the maximum
threshold setting, the maximum infusion pressure generated into a hard occlusion at 25 mL/h is 11.6±3.9 psi. Testing performed using IV set Model 72003, at 68±4°F (20±2°C).
Case:
Impact and flame resistant plastic.
Critical Volume:
Maximum incremental volume in case of single point failure will not exceed 1.0 mL at 999.9 mL/h.
Dimensions (Nominal):
7130/7131 7230/7231
Width Height Depth § Weight §§ Power Cord
§ without pole clamp
§§ without power cord
7.6 in/19.3 cm
8.6 in/21.8 cm
5.0 in/12.7 cm
6.6 lbs/3.0 kg 10 ft/3 m
10.7 in/26.7 cm
8.6 in/21.8 cm
5.0 in/12.7 cm
8.4 lbs/3.8 kg 10 ft/3 m
Environmental Conditions:
Operating Storage/Transport
Temperature Range
Relative Humidity
Atmospheric Pressure
5 to 40°C
(41 to 104°F)
20 to 90%
Non-condensing
700 to 1060 hPa
-40 to 60°C
(-40 to 104°F)
5 to 95%
Non-condensing
500 to 1060 hPa
Fluid Ingress Rating:
Drip proof IPX1
Time Bolus Volume Released Upon Correcting Downstream Occlusion (mL)
Monitoring Options
Pressure
Resistance and
High Resistance
Threshold Settings
25
mmHg
600
mmHg
100%
25 mmHg
100%
600 mmHg
1 mL/h
Maximum 0.5 0.5 0.5 0.5
Typical <0.1 0.3 <0.1 <0.1
25 mL/h
Maximum 0.5 0.5 0.5 0.5
Typical <0.1 0.3 <0.1 <0.1
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 18
GENERAL INFORMATION
1-5
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Table 1-1. Specifications (Continued)
Ground Current Leakage:
Electrical leakage current, enclosure: <100 microamperes Electrical leakage current, patient: <10 microamperes
Log Capacity:
Software version 2.78:
2000 Event History Log
Software version 4.06/4.08:
300 Continuous Quality Improvement (CQI) 1500 Event History Log
Maximum Infusion Pressure:
16 psi
NOTE: Testing performed per proposed standard IEC 601-2-24 using IVAC®IV sets.
Maximum Time to Alarm:
NOTE: When the occlusion alarm pressure limit is set to the
maximum threshold setting, the maximum infusion pressure generated into a hard occlusion at 25 mL/h is 11.6±3.9 psi. Testing was performed using IV set Model 72003, at 68±4°F (20±2°C).
Memory: 4.06 and higher: Maintains infusion perimeters indefinitely until the
instrument is reprogrammed.
2.78: Interrupted secondary or advanced operating modes retain special program settings up to 6 hours. Resistance/pressure trending information is retained for 6 hours.
Mode of Operation: Continuous
Number of Instruments per Pole: 3 (713X), 2 (723X), single pole, weighted base, 5 legs.
Time to Detect
Downstream
Occlusion (minutes)
Monitoring Options
Pressure
Resistance and
High Resistance
Threshold Settings
25
mmHg
600
mmHg
100%
25 mmHg
100%
25 mmHg
1 mL/h
Maximum 2 75 2 7
Typical 0.6 30 0.6 4
25 mL/h
Maximum 1 25 1 3
Typical 0.1 1 1.0 1
Page 19
GENERAL INFORMATION
1-6
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Occlusion Alarm Pressure:
12 ± 4 psi maximum
Parts per mL:
Precision Flow 3600 steps to deliver 1 mL Pressure Mode (rate < 50 mL/h). All other occlusion monitoring options 1150 steps to deliver 1 mL (Cal # = 0.0) steps will vary with Cal #.
Power Requirements:
100-240 VAC, 50-60 Hz (40 W), 3-wire grounded system. Class 1 with internal power source.
Rate Accuracy:
For rates greater than 1 mL/h, up to 999.9 mL/h: ±5%, 95% of the time with 95% confidence, under the conditions listed below.
For rates equal to or less than 1 mL/h: ±6.5%, 95% of the time with 95% confidence, under the conditions listed below.
Variations of head height, back pressure, time, monitoring mode option, instrument tilt, or any combination of these, may affect rate accuracy. Factors that can influence head height and back pressure are: IV set configuration, IV solution viscosity and IV solution temperature. Back pressure may also be affected by type of catheter.
Rate Range:
0.1 to 999.9 mL/h in 0.1 mL/h increments (primary)
0.1 to 270.0 mL/h in 0.1 mL/h increments (secondary)
RFI:
Tolerance < 10 V/m across frequency range.
Temperature:
Operating above 30°C, for extended periods will reduce battery life.
Volume to Be Infused Range:
0.1 to 9999.9 mL in 0.1 mL increments (primary and dose rate)
0.1 to 999.9 mL in 0.1 mL increments (secondary and loading dose)
0.1 to 999.9 mL per step in 0.1 mL increments (multi-step)
0.1 to 999.9 mL per dose in 0.1 mL increments (multi-dose)
WARNINGS (Alerts):
• Battery Low • Load Dose Complete
• Checking Line • Multi-Step Complete
• Complete Entry • Resistance Alert
• Computer Control Released • Secondary Complete
• Dose Complete • VTBI = 0
• Infusion in KVO
CAUTION
Table 1-1. Specifications (Continued)
NOTE: The Signature Edition®GOLD Infusion System has been assessed and complies with the following
Technical Standards: IEC 60601-1 / BS 5724, including amendments A1 and A2; IEC 60601-2-24; CISPR 11, Group 1, Class B Emissions; IEC 60601-1-2.
Infusion rate range:
0.1 to 999.9 mL/h
Head height: 24 ± 1 in.
(61 ± 2.5 cm)
Environment temperature:
68±8°F (20±4°C)
Back pressure: 0 psi.
Test solution: distilled water Set Model: 72003 Needle: 18 gauge Minimum collection volume: 6 mL
Page 20
GENERAL INFORMATION
1-7
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.4 ACCESSORIES (Continued)
1.4.3 Flow Sensor
Flow sensor capability is available with an upgrade kit for the 7130/7230 (refer to the “Illustrated Parts Breakdown” chapter). For 7131/7231, all that is needed is a flow sensor.
The flow sensor attaches to the administration set's drip chamber. It detects an empty solution container and verifies fluid flow. When installed, it will allow VTBI to be turned off. The flow sensor will not see drops falling if the drip chamber is tilted more than 24°.
1.5 ALARMS, ERRORS, MESSAGES
Alarm messages are displayed on the Main Display. Refer to the applicable Signature Edition
®
GOLD Infusion System DFU for
detailed information.
1.5.1 Silencing Alarms
All alarms can be temporarily silenced by pressing the Silence Key.
1.6 BATTERY MANAGEMENT SYSTEM
This section contains general information on the battery management system. Included is information on how the Battery Manager monitors and maintains the battery, controls the power on/off for the rest of the instrument, and provides support functions for the main processor. Refer to the “Principles of Operation” chapter for more detailed functional descriptions.
The battery management system consists of the Battery Manager IC and various sensors and signal processing circuits. The Battery Manager IC (Rev. 3.06) is a custom­programmed microcontroller that performs the following functions:
Controls battery charger
Provides a battery status "battery gauge"
Monitors voltage and temperature of battery
Controls instrument power source (on/off function)
Drives Lower LCD Display (refer to Figure 1-2)
Includes a relative-time clock
The Battery Manager communicates with the main processor via a serial data channel. The main processor issues commands to the Battery Manager which then responds with status information and data using this channel.
1.6.1 Fan
The internal fan is used for cooling, mainly to help prolong battery life. It is a ball­bearing, brushless DC fan. The fan is always on when the battery is charging with "Fast" or "Top-up" charge. The fan will go on any time battery temperature is over 22°C.
1.6.2 Battery and Charging Process
The battery is a ten-cell (1.2V per cell), high capacity nickel-cadmium type rated at 12 volts and 1.8 amp-hours (with a minimum of 500 charge cycles).
The battery pack (10 to 18V) has a built-in temperature sensor which allows the Battery Manager to monitor the temperature of the battery. The pack also includes a temperature-limiting thermostat which opens the circuit if the battery temperature gets too hot and closes again when the temperature returns to normal.
Page 21
GENERAL INFORMATION
1-8
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.2 Battery and Charging Process
(Continued)
The battery charge circuit charges the battery with a constant current of 1 ampere whenever the Battery Manager turns the charger on. The Battery Manager regulates average charge current by turning the charger on and off with the appropriate duty ratio. The battery charge cycle consists of four modes; fast charge, top-up charge, float charge, and hot charge.
Fast Charge
: Fast charge is initiated whenever the battery is less than 36°C, and has been discharged by more than 200 Ampere-seconds through actual use or self discharge. Leaving the instrument unplugged for a day would cause about 200 Ampere-seconds of self discharge. The charge current is a continuous 1 Ampere. The end of a fast charge is detected when the temperature of the battery rises 7°C above its temperature at start of charge and is at least 30°C, or when the battery voltage declines by 192mV below its peak value, or total charge time exceeds 3.2 hours. Refer to “Battery Charge Regulator” section in the “Principles of Operation” chapter for further details.
•T
op-Up Charge: The top-up charge phase begins at the end of the fast charge phase and finishes adding the last few percent of charge to the battery and balances individual cell charges. This phase charges at an average rate of 180 mA (1 A for 0.9 seconds every 5 seconds) for 180 minutes. At that time, the instrument will go into float charge mode.
The charger will suspend top-up if the battery temperature exceeds 37°C. The time spent to cool down to below 37°C is in addition to the 180 minutes top-up charge time. If top-up cool down time exceeds 5 1/2 hours, the instrument will go into float charge mode.
Float Charge
: The float charge phase begins at the end of the top-up phase and helps maintain a fully charged battery. This phase charges at an average rate of 40 mA (1A for 0.2 seconds every 5 seconds). The fan remains on or turns on when battery temperature exceeds 22°C.
Hot Charge
: The hot charge mode occurs when the instrument determines that the battery is >36°C (normally due to ambient temperature being >27°C) to allow a charge after waiting 3 hours for it to cool down. Hot charge mode charges at an average rate of 180 mA (1A for 0.9 seconds every 5 seconds) for a total charge time of 18 hours. If the battery temperature exceeds 43°C, the charging is turned off until the temperature falls below 43°C.
Note that the cool down time is in addition to the 18-hour charge time. The float charge cycle begins at the end of the hot charge cycle.
1.6.3 Refresh Cycle
A battery refresh cycle performs a full charge, discharge, and recharge to condition and measure the capacity of the battery. This refresh results in a new "Measured Capacity" in the battery diagnostics and can be used to judge the condition of the battery.
Page 22
Battery Voltage Instrument Response
12.0V (Single)
12.1V (Dual)
• One tick mark left on gauge
• Instrument continues to function
• Warning tone activated
• Low battery warning
11.45V • Instrument does not pump
• Constant alarm
• Low battery alarm (depletion)
10.25V • 1 minute or longer (nominal 5 min.) after low battery alarm
• Backup speaker activated
• Instrument shutdown (5 min. after alarm)
9.75V • No AC power applied
• Battery disconnected from circuit by shutdown signal
GENERAL INFORMATION
1-9
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.3 Refresh Cycle (Continued)
The refresh cycle must be initiated manually, either by disconnecting/ reconnecting the battery or by loading 0.0 Ah as the rated capacity in the battery diagnostics page. After the zero rated capacity is loaded and the OK soft key is pressed, the original rated capacity must be reloaded to preserve the battery gauge. The empty ("E") icon will flash during the discharge part of the refresh cycle (see Figure 1-1). Important: The AC power must remain connected and uninterrupted during the discharge cycle. If the AC power is removed during this cycle, the discharge cycle will be terminated and another refresh cycle will have to be initiated. The time for complete refresh is dependent upon battery
Figure 1-1. Battery Refresh Cycle
Table 1-2. Battery Trip Points
Refresh cycle starts
“E” blinks
Cycle ends
11.45 Volts
Fast Fast
Cool* Cool*
Top-Up*Top-Up*
FloatDischarge
* Cool and Top-Up portions of cycle will vary with temperature. The complete cycle could last 24 hours if instrument turned on with the Fan running most of the time.
Page 23
GENERAL INFORMATION
1-10
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.3 Refresh Cycle (Continued)
temperature and takes an average of 20 hours to complete (refer to Figure 1-1).
Two ways to initiate a manual refresh cycle are:
Disconnect from AC, unplug battery, and then press power switch. Lower LCD will go blank. Reconnect battery and then reconnect instrument to AC.
Enter Diagnostics Mode, go to page D2, Select Battery Status, and enter 0.0 Ah for battery-rated capacity. Press Enter and ok. Turn instrument off and back on, and then reconnect to AC. Once the fan turns on (indicating start of a refresh cycle), return to page D2 and reset battery rated capacity to 1.3 Ah.
1.6.4 Battery Gauge
The battery gauge provides an indication of the approximate amount of charge remaining on the battery. It will usually indicate less charge than is actually available with a new battery. This is displayed in a bar graph format on the lower LCD Display and is active as long as the battery is connected to the instrument. (Refer to Figure 1-2, Lower LCD Display Layout.)
The battery gauge circuitry measures the current flow into and out of the battery and maintains a record of the state of charge of the battery. This record is reset each time the instrument completes a full refresh cycle. The battery gauge uses this record, together with a measurement of the present power requirements of the instrument, to
estimate the charge available on battery power at the current infusion rate.
NOTES:
The instrument label and battery gauge are always displayed, even when the instrument is turned off; however, the battery gauge does not represent the battery charge remaining when the instrument is turned off.
To ensure a more accurate battery gauge reading, review the battery gauge five minutes after starting an infusion. The gauge updates for each program change while infusing. Battery run time may be affected by the operating mode, rate, monitoring options and back pressure.
1.6.5 Power On/Off
The Battery Manager provides the interface between the power on/off switch(es) and the main processor. When the instrument is off, the Battery Manager interprets either power switch as a turn on command and applies power to the rest of the instrument, informing the main processor which switch was pressed. Once power is on, further presses of a power switch are passed on to the main processor which determines the appropriate response under the existing conditions. If the response is to turn the power off, the main processor requests that the Battery Manager remove power from the rest of the instrument.
If an error has been detected which causes the watchdog to be in alarm, pushing the power switch (or for dual channel instruments, either power switch) immediately causes the power to be turned off, without intervention by the Battery Manager.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 24
GENERAL INFORMATION
1-11
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.6 Lower LCD Display
The Battery Manager also contains the driver for the Lower LCD Display. In addition to the battery gauge, this display contains a four-character alphanumeric "configuration" display and several icons. The information for these other displays is controlled by the main processor and is communicated to the Battery Manager through the serial channel. The Battery Manager also uses the four­character display to indicate errors detected in the Battery Manager system itself.
1.6.7 Clock
The Battery Manager provides a "relative time" clock which the main processor can set and read. This clock consists of a 32-bit counter which is incremented once a second under all conditions. The main processor uses this counter as a means of determining elapsed time even when power has been turned off. The clock is used to compensate for normal battery capacity degradation over time.
1.6.8 Battery Maintenance
Several features have been included in the Battery Manager to help properly maintain the battery.
A battery capacity measurement is available in diagnostic mode.
A special circuit removes all load from battery when voltage falls too low, preventing damage from over discharge due to long-term storage.
NiCad batteries can be stored indefinitely with no load but will self-discharge from a charged state in about 100 days. This does not damage the battery as it would if it were a lead acid-type battery. Connect instrument to AC to recharge batteries.
Figure 1-2. Lower LCD Display Layout
Use only batteries approved by ALARIS Medical Systems, due to Battery Manager requirements and the thermostat contained in the battery assembly. If the instrument has been in storage, connect it to AC power before turning it on. Usually one refresh cycle is sufficient to restore battery capacity. If necessary, repeat the procedure at 24-hour intervals, 2 or 3 times, to increase capacity.
Battery replacement should be performed by qualified service personnel while instrument is not in use. DO NOT open, incinerate or short circuit the battery. Worn-out batteries must be disposed of properly, according to local regulations.
CAUTIONS
GOLD
MNTR
Page 25
GENERAL INFORMATION
1-12
1.6 BATTERY MANAGEMENT SYSTEM
(Continued)
1.6.8 Battery Maintenance (Continued)
NOTE: If under load (inside instrument), the
maximum storage time would be about 130 days before electrolyte extrusion occurs past the battery seal.
If the battery exhibits short run times, a reconditioning procedure can be used:
1. Disconnect battery and AC
2. Press On/Off switch and verify lower LCD goes blank.
3. Reconnect battery.
4. Reconnect instrument to AC. A refresh cycle will be initiated.
5. Repeat this procedure at 24-hour intervals, 2 or 3 times, to increase capacity or use a battery conditioner.
1.7 NICAD BATTERY CAPACITY INFORMATION
All batteries have specific conditions under which they are guaranteed to meet their published specifications. Deviations from these conditions typically result in a reduction of available capacity. Manufacturers of NiCad batteries rate capacities, usually expressed in Ah, based on a specified "ideal" charge and discharge condition, as well as the use of a "new" battery. Battery manufacturers do not use the calendar year for battery date codes. The battery date code year starts on September 1. The first two digits of the date code are for the year and the last two digits are for the week of the year (for example, date code 0411 = the year 2004 and the eleventh week from September 1).
An ideal charge cycle starts with a fully discharged battery charged at C/10 (C is the rated capacity in Ah) constant current for 15 hours while at room temperature. For instance, a 1.8Ah battery would be charged for 15 hours at 180mA constant current with a room temperature of 23°C.
The ideal discharge starts with a fully charged battery under a C/5 constant current load at room temperature, discharging to a cell voltage of 0.9V. The rated capacity is then calculated as the time to discharge divided by 5. Again, a 1.8Ah cell would be discharged at 360mA constant current and not reach 0.9V for at least 5 hours. Note that a given battery type has different capacities based on the load. For instance, a battery rated at 1.8Ah at a 360mA load may have only 1.6Ah at a 1600mA load.
As can be seen from the preceding ideal conditions, there are many conditions which can affect the battery capacity. The following conditions have the most practical impact on battery capacity delivered in this instrument.
•T
emperature During Charge: As the effective ambient temperature of the battery increases, the amount of charge that the battery will accept is decreased. At an ambient temperature of 35°C, an enclosed battery will temporarily accept only about 90% of the charge it would otherwise accept at 23°C. Because the batteries are internal to the instrument case and the instrument itself generates heat, the batteries will be exposed to temperatures above room temperature. Some of the ways the instrument limits the temperatures that the battery sees include forcing air across it (an internal fan) and turning off the charger when the battery temperature gets too high.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 26
GENERAL INFORMATION
1-13
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.7 NICAD BATTERY CAPACITY INFORMATION (Continued)
Cycle Life and
Aging: As batteries get older and go through many charge/discharge cycles, batteries "wear out," in that the chemicals and materials used to construct the cell break down. The way the instrument deals with this is to assume that a battery will continually reduce capacity at a rate equivalent to 30% over 4 years, and continually reduce capacity at a rate equivalent to 30% per 200 full charge cycles. These calculated values are used to reduce the displayed runtime on the battery gauge.
Partial Discharge/Charge: When a battery is partially discharged, then charged for less than the full time, differences between individual cell capacities result in cells completing charge at different times. If the full charge sequence is not then completed, the cell "mismatch" becomes progressively greater. This is viewed by the user as low apparent runtimes and premature low battery warning and alarms.
The problem is cumulative in that the mismatch increases for every partial cycle. The lowered capacity is not permanent, but may require 2-3 full charge cycles to recover. The way the instrument deals with this is to reduce the displayed runtime based on a limited history of partial cycles.
Charge Rate: The ideal charge rate requires 15 hours to get to full charge, which is undesirable from the user's perspective. The instrument provides a multiphase charge cycle which results in about 80% capacity
in the first 2 hours after fast charge. The next charge phase, top-up, is designed to finish the charge and to bring all individual cells to the fully charged state, essentially rematching them. Refer to “Battery and Charging Process” section in this chapter for “Fast Charge” and “Top-up Charge” information. If the top–up charge is not completed, then the cell mismatch is not reduced and the cumulative capacity reduction occurs. Top-up is a 3-hour charge, but the elapsed time to complete it may be over 5 hours, as the charger is turned on and off to keep the battery cool during that time.
Battery
Alarm Voltage: The battery alarm voltage is the voltage at which the instrument stops operating and generates an alarm indicating the instrument needs to be connected to AC power. As noted in the ideal discharge condition, the end of discharge is determined by 0.9V/cell.
Under perfect conditions, a battery of 10 cells connected in series would reach the end of discharge at 9.0V; however, cells are not perfectly matched so some will reach 0.9V before others. The problem occurs when a cell in series with other cells can go below 0.9V and actually can go into cell reversal, which permanently damages the particular cell. On the other hand, increasing the alarm voltage to compensate for imperfectly matched cells results in reduced runtimes with available capacity.
Page 27
1.7 NICAD BATTERY CAPACITY INFORMATION (Continued)
The user sees this as premature low battery warnings and alarms. The way the instrument deals with this is to increase the alarm voltage to guarantee the battery is not damaged and to reduce the assumed capacity to below that printed on the battery. The battery gauge is intended to show the minimum run time left on the battery taking all these factors into account.
NOTE: In the event ALARIS Medical Systems provides different battery packs in the future, the replacement battery may have a different rated capacity. Therefore, the Battery Manager of the instrument needs to know that a new battery has been added and the rated capacity has changed. Refer to the “Removing Battery” section of “Corrective Maintenance” chapter and the “Changing Rated Capacity of Battery” section of “Troubleshooting” chapter.
1.8 DYNAMIC MONITORING®SYSTEM (DMS)
The following is general information regarding the Dynamic Monitoring
®
System.
In order for fluid to move through the administration set, a pressure difference (gradient) must exist. In a gravity setup, this is done by head height. In a pump, the instrument will develop pressure to overcome downstream effects on fluid flow.
The fundamental concept behind the Dynamic Monitoring
®
System is that the resistance to fluid flow from the mechanism to the patient's infusion site can be measured. This is done by intentionally varying the flow rate while monitoring the resulting changes in fluid pressure (refer to Figure 1-3 "Resistance Graph"). Signal
processing of the pressure and flow data can then produce the fluid impedance value. Such measurements can be made continuously at short intervals and be independent of the selected rate. Head height and resulting pressure variations, likewise, will not affect the measurement.
When a complete occlusion occurs, the resistive part of the fluid impedance is very large (theoretically infinite). Elevated resistances due to clotting, clogged filters, partial occlusions or infiltrations can be measured.
The Dynamic Monitoring
®
System provides a means to measure the mechanical properties of the downstream flow path.
Figure 1-3. Resistance Graph
Pressure =
Effect when a force is against a restriction.
Resistance =
Caused when impediment to fluid flow occurs.
Resistance = Change in Pressure (
P)
Change in Flow ( F)
Refer to Figure 1-4 "Pressure and Resistance Graph".
GENERAL INFORMATION
1-14
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
W
RATE HI
SET RATE
P
RATE LO
PRESSURE (result)
0
FLOW (cause)
F
F Nominal
Page 28
GENERAL INFORMATION
1-15
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1.8 DYNAMIC MONITORING®SYSTEM (DMS) (Continued)
Features include:
Alarm setting is based on dynamic system impedance (pressure changes, not absolute pressure).
Detection of complete occlusions.
Reduced nuisance alarms by minimizing artifact effects such as head height and patient movement.
AutoRestartPlus feature allows instrument to automatically continue operation if an occlusion is cleared within self-check period (40 seconds). A warning tone and “Checking Line” message will occur for up to 40 seconds. The feature can be turned off (set restarts to zero), or number of restarts may be set from 1 to 9.
NOTES:
• The restart counter is reset whenever Run/Hold is pressed, the instrument or channel is turned off or an alarm occurs.
• When infusion is started, the resistance may be other than 0%, depending on solution viscosity, catheter/tubing size and filters.
Resistance Alert provides an early warning of slow or gradual changes in resistance of IV line/site. Resistance Alert marker can be set from 5 to 100%. It allows resistance to be monitored and provides a tone every 30 seconds if percent resistance exceeds resistance alert mark.
Resistance display may be turned off. If off, system continues to monitor downstream resistance and alarm appropriately.
Figure 1-4. Pressure and Resistance Graph
Pressure rise is quite small with site complication, especially with low flow rates.
Resistance rises dramatically with site complication.
600
300
IN PRESSURE
PRESSURE mmHg
0
OCCLUSION
100%
75%
50%
FLUID OHMS
mmHg/Liter/Hr
0
OCCLUSION
IN PRESSURE
TIME
TIME
Page 29
GENERAL INFORMATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
1-16
1.9 DATA COMMUNICATIONS FUNCTION
The instrument has built-in remote monitoring capability. This allows features and their data to be monitored by a computer, providing a means to create advanced clinical systems. A separate manual on data communications is available and organized to support technical personnel with a wide range of experience and needs. The separate manual includes:
General Information: Includes phone number for technical support.
Operation: Instrument's communications modes, controls, indicators, and procedures from user's point of view.
Instrument Setup: How to set the baud rate, enable or disable computer monitoring, and other parameters.
Electrical Interfacing: RS-232 background information, connectors and recommended wiring for common computers.
Communications Protocol: Inquiry, response, and command codes, data formats, message sequences, and error detection.
1.10 TRUMPET AND START-UP CURVES
Refer to the applicable Signature Edition
®
GOLD Infusion System DFU.
Page 30
GENERAL INFORMATION
Signature Edition®GOLD, Models 7130/7131 and 7230/7231
Technical Service Manual
1-17
Table 1-3. Abbreviations, Acronyms, Symbols
Various abbreviations, acronyms and symbols are used throughout this manual. The following are those that are not commonly known or easily recognized.
* "active Low" logic signal VAO voltage alternating oscillator
AlL air-in-line VBKUP voltage backup
DFU directions for use VNEG negative voltage
DS display VMEAS voltage measured
ECD empty container detection VMOTOR voltage to motor
F fuse VPOS positive voltage
FB ferrite bead VRAW voltage raw (unregulated voltage)
H hexadecimal VTBI volume to be infused
KVO keep vein open VTHRES voltage threashold
PM preventive maintenance WD watchdog
PR power regulator WDI watchdog input
PRI/SEC primary to secondary WDO watchdog output
SCU serial control unit
Page 31
GENERAL INFORMATION
Signature Edition®GOLD, Models 7130/7131 and 7230/7231
Technical Service Manual
1-18
THIS PAGE
INTENTIONALLY
LEFT BLANK
Page 32
2A CHECKOUT & CONFIGURATION
Page 33
CHAPTER 2A — CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-1
2.1 INTRODUCTION
This chapter describes the initial setup and configuration of an instrument with software versions 4.06 and 4.08.
2.2 NEW INSTRUMENT CHECKOUT
Refer to the instrument's Directions For Use (DFU) for instructions regarding unpacking and setting up the instrument for the first time.
When turning on the instrument, verify the instrument beeps and that all display LED segments flash. This confirms that the instrument has performed its self test and is operating correctly. During operation, the instrument continually performs a self test, and will alarm and display a message if it detects an internal malfunction.
Contact ALARIS Medical Systems authorized service personnel if the instrument has physical damage, fails to satisfactorily pass startup sequence, fails a self test or continues to alarm.
For new instrument checkout refer to the applicable Signature Edition
®
GOLD
Infusion System DFU.
2.3 CONFIGURABLE OPTIONS AND DEFAULTS
A hospital/facility biomedical technician has the capability to set all configuration parameters to their startup defaults in a single operation. The terms "configuration parameters" and "programmable features" are interchangeable and have the same meaning. Refer to the applicable Signature Edition
®
GOLD Infusion System DFU for
more information.
Keep latch closed when instrument is not in use.
CAUTION
Should an instrument be jarred severely or dropped, remove it from use immediately. It should be thoroughly tested and inspected by qualified service personnel to ensure proper function prior to reuse.
CAUTION
Page 34
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-2
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
When an instrument’s configuration is changed, the configuration name should also be changed in order to document the new parameter settings. The intent of the configuration name is to have only one set of parameters for each alpha-numeric code. Refer to “Pop-Up Displays” section.
Powering down in configuration mode during an alarm or error will NOT save any configuration changes.
The configuration procedure is for use by qualified service personnel only. The configuration mode is intended for programming the technical and clinical features in accordance with current procedures and practices.
Table 2-1, “Record of Configured Instruments”, can be reproduced and used to record and track instrument configuration settings.
A map of all the configuration screens is located at the end of this chapter (Figure 2-1).
Software version 2.78 is not Guardrails
®
Safety Software compatible and the configuration mode is different. Refer to Chapter 2 for “Software Version 2.78” for detailed calibration information.
The configuration mode for versions 4.06 and 4.08 are identical except when Profiles is enabled. When Profiles is enabled, the only configuration settings that can be changed in version 4.08 are Regional Settings, Profiles, Computer Link
WARNINGS
and Optional Features, because the Guardrails
®
Safety Software profile overrides every other setting in the configuration mode.
NOTES:
Pressing a soft key at the side of the Main Display the first time, selects it for editing. Some features are edited by subsequent presses of the soft key to cycle through available options. Other features are edited by means of the numeric keypad entry.
Pressing undo undoes any edits made to that page, and stays on the page.
• Pressing ok accepts all information on the page, and returns to menu page.
• Pressing the POWER switch after editing a configurable item evokes an invalid key tone and a message to "ok entry". ok must be selected to accept the edit before the instrument can be powered off.
• The page number is located in the upper­right corner of the display.
The instrument’s configuration information is not lost when disconnected from AC power. However, error history and infusion program settings may be lost. If this information is to be saved, be sure to record it before disconnecting power.
CAUTION
Page 35
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.1 Entering Configuration Mode
The instrument must be off (both channels must be off for
dual channel).
The procedure for the single and dual channel is the same. Any configuration in the dual channel sets the same value for both channels.
1. Press and hold left-bottom display soft key.
2. Press POWER switch. Continue to press display soft key until configuration mode display appears, then release.
Page C1 of configuration mode is displayed. This is a read-only display.
3. Press page to advance to page to be configured.
NOTE: For 7131/7231, the default code will be 5B44 for instruments set to factory defaults (instead of 2D15).
2.4.2 Setting to Defaults
The Set To Defaults mode programs all configuration items to their default values.
It is strongly recommended that the Configurable Options be reviewed for a complete list of defaults before using this feature.
1. Advance to C2 page.
2. Press Set to Defaults soft key.
3. Press ok to accept change and return to beginning of C2 page. Pressing cancel leaves all items set to their previous values and returns to beginning of C2 page.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-3
CAUTION
B
A
OPTIONS
B
A
Set To Defaults
Regional Settings Air In Line Profiles
page
OPTIONS
C
2
CONFIGURATION MODE
ID No. : 001234567
SW Rev.: 04.08
Config.: 2D15 GOLD
C
1
page
B
A
cancel
Set all
Configuration
items to their
default values
OPTIONS
C
2
ok
Page 36
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.3 Regional Settings
1. From C2 page, press Regional Settings
soft key.
2. Press Region soft key to select for
editing. Press again to cycle through region choices (N. America, European).
3. Press Language soft key to select for
editing. Press again to cycle through language choices .
NOTES:
If there is no tick mark next to
Language, English is the only option.
Instruments configured in European Regional Settings (with profiles OFF) do not have a drug list. When Dose Rate Calculator is chosen (from OPTIONS menu), followed by Enter New Program, the instrument defaults to generic drug dose rate calculation.
4. Press ok to accept change and return to
beginning of C2 page.
2.4.4 Setting Air-in-Line Threshold
The air-in-line threshold sets the bubble size sensitivity. The air-in-line reset allows the clinician to respond to an AIR IN LINE alarm, assess its clinical significance, and choose whether or not to continue the infusion without removing the air. The reset feature allows only the current bubble to proceed without tripping an alarm. The air in line threshold value choices are 50, 100, 200 and 500 microliters.
The Accumulator setting, when on, looks for 10% to 15% of the downstream path to be air before giving an ACCUMULATED AIR IN LINE alarm. The amount of air that causes the alarm will vary with the threshold setting and rate.
An AIR IN LINE alarm is the result of the bubble size exceeding the Threshold setting. An ACCUMULATED AIR IN LINE alarm is the result when 10 to 15% of the downstream path is filled with air, exceeding the accepted level looked for when the Accumulator is set to On.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-4
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
B
A
Set To Defaults Regional Settings
Air In Line Profiles
OPTIONS
C
2
page
B
A
Regional Settings
Region= N. America
Language= English
undo
OPTIONS
ok
C 2
C
Page 37
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.4 Setting Air-in-Line Threshold
(Continued)
1. From C2 page, press Air In Line soft
key.
2. Press Threshold soft key to select for editing. Press again to cycle through 50
mcL, 100 mcL, 200 mcL and 500 mcL.
NOTE: Use a 50 microliters setting on
microbore tubing. The other three settings may be used on macrobore tubing.
3. Press Reset soft key to select for editing. Press again to cycle between On and Off.
4. Press Accumulator soft key to select for editing. Press again to cycle between On and Off.
5. Press ok to accept change and return to beginning of C2 page.
2.4.5 Profiles
1. From C2 page, press Profiles soft key.
2. Press Off or On soft key to select for
editing. Press again to cycle between
On and Off.
NOTES:
The data set for Guardrails
®
Safety
Software must be loaded to set Profiles to On. For the Guardrails
®
Safety
Software to function, Profiles and Dose Rate must be On.
• When Profiles is turned on, the configuration mode will have limited access. Access will be limited to Regional Settings, Profiles, Computer Link and Optional Features, due to the data set overriding the configuration settings as part of the Guardrails
®
Safety Software.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-5
Threshold=
Air In Line
Reset= Off
Accumulator= On
2
undo
ok
100 mcL
OPTIONS
C
A
B
B
A
Set To Defaults
Regional Settings Air In Line Profiles
OPTIONS
page
C
2
B
A
Set To Defaults
Regional Settings Air In Line Profiles
page
B
A
Profiles
Off
undo
OPTIONS
C
2
OPTIONS
C
2
ok
Page 38
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.6 Setting Maximum Rate
This sets the maximum selectable rate. The range for Maximum Rate is 0.1 to
999.9 mL/h.
NOTES:
• The maximum rate setting applies to all infusion modes.
• Setting the Maximum Rate below the preset KVO Rate will lower the KVO Rate. The KVO Rate will not exceed the Maximum Rate.
1. Advance to C3 page.
2. Press Maximum Rate soft key.
3. Press the soft key to select the value for editing.
4. Use numeric keypad to enter maximum
rate. Press Enter.
5. Press ok to accept change and return to
beginning of C3 page.
2.4.7 Setting Computer Link
The Computer Link feature allows a hospital/facility computer to interact with the instrument and programs the level of computer control available. If the feature is off, the computer cannot communicate with the instrument.
Monitor Mode allows computer to only receive information from instrument. Enabling of monitoring mode automatically places it in options menu.
Off Mode does not allow any communication between instrument and a computer.
1. On C3 page, press Computer Link soft
key.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-6
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Clear
Enter
1
23
4
56
7
8
9
.
0
Maximum Rate =
C
3
ok
OPTIONS
A
B
999.9 mL/hr
B
A
Maximum Rate
Computer Link Optional Modes
Optional Features
OPTIONS
C
3
page
B
A
Maximum Rate Computer Link
Optional Modes Optional Features
OPTIONS
C
3
page
Page 39
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-7
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.7 Setting Computer Link (Continued)
2. Press Mode soft key to select for editing.
Press again to cycle through Off and Mntr.
3. Press Baud Rate soft key to select for
editing. Press again to cycle through choices (300, 600, 1200, 1800, 2400, 4800 and 9600).
4. Press Parity soft key to select for editing.
Press again to cycle through Even, Odd, and None.
5. Press ok to accept changes and return
to beginning of C3 page.
2.4.8 Setting Optional Modes
The Optional Modes feature allows the options menu to be configured to how it will appear in normal mode. Enabling of these modes will automatically place them in the option menu.
When Loading Dose, Multi-Step, and Multi-Dose are on, they appear in the menu when options key is pressed in normal mode.
For Dose Rate to appear in the options menu, it must be turned on.
Universal Rate: If the DRC mode is discontinued, the instrument's primary setting will now be the same as the last rate and VTBI used in the DRC mode.
NOTES:
• Dose Rate must be on in order to use Guardrails
®
Safety Software drug selection
capability.
• Instruments configured in European Regional Settings (with profiles OFF) do not have a drug list. When Dose Rate Calculator is chosen (from OPTIONS menu), followed by Enter New Program, the instrument defaults to generic drug dose rate calculation.
1. On C3 page, press Optional Modes soft
key.
2. Press Loading Dose soft key to select
for editing. Press again to cycle between
On and Off.
NOTE: This feature allows an initial
infusion rate to be set up for a specific volume, automatically followed by a maintenance rate from the same container.
Mode = Mntr Off Baud Rate = 9600
Parity = Even
C
3
ok
OPTIONS
A
B
undo
A
B
Maximum Rate Computer Link
Optional Modes Optional Features
OPTIONS
C
3
page
Page 40
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-8
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.8 Setting Optional Modes (Continued)
3. Press Dose Rate soft key to select for editing. Press again to cycle between On and Off.
NOTE: This feature allows dose
parameters to be programmed so that the instrument then calculates the volumetric rate.
4. Press Multi-Step soft key to select for editing. Press again to cycle between On and Off.
NOTE: This feature allows a sequential program to deliver up to nine steps; fluid volumes and delivery rates may be programmed for each step.
5. Press Multi-Dose soft key to select for editing. Press again to cycle between On and Off.
NOTE: This feature allows multiple infusions to be programmed over a period of up to 24 hours; the fluid volume and delivery rate is repeated for each delivery.
6. Press ok to accept changes and return to beginning of C3 page.
2.4.9 Setting Optional Features
The Panel Lock feature allows the front panel to be locked and unlocked to help prevent tampering.
When VTBI (volume to be infused) is on, a volume to be infused must be entered; otherwise, the last remaining VTBI or last entered VTBI, depending on the channel's last usage, will be in effect. When VTBI is off, there is no VTBI line capability.
When the Multi-Dose Alert feature is on, the option is available to set an alert at the end of every dose when in the Multi-Dose mode.
1. On C3 page, press Optional Features
soft key.
2. Press Panel Lock soft key to select for
editing. Press again to cycle between On and Off.
B
A
Maximum Rate Computer Link
Optional Modes
Optional Features
OPTIONS
page
C
3
B
A
Loading Dose On Dose Rate= On
Multi-Step= Off Multi-Dose= Off
undo
ok
OPTIONS
C
3
B
A
Panel Lock= On VTBI= On
Multi-Dose Alert=Off
undo
OPTIONS
C
3
ok
Page 41
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-9
2.4 CONFIGURATION PROCEDURE (Continued)
2.4.9 Setting Optional Features (Continued)
3. Verify VTBI is On.
NOTE: The flow sensor option (empty
container detector) needs to be installed to be able to turn VTBI off.
4. Press Multi-Dose Alert soft key to select
for editing. Press again to cycle between On and Off.
5. Press ok to accept changes and return to
beginning of C3 page.
2.4.10 Setting KVO Rate
The instrument will automatically operate at the KVO rate (or current rate, whichever is less) once the primary VTBI has counted down to zero. The KVO rate range is 0.1 to
20.0 mL/h.
1. Advance to C4 page.
2. Press KVO Rate soft key.
3. Press soft key next to rate value once to select for editing.
4. Use numeric keypad to enter KVO rate. Press Enter.
5. Press ok to accept change and return to beginning of C4 page.
NOTE: The KVO Rate will NOT exceed the present Maximum Rate.
Clear
Enter
1
23
4
56
7
8
9
.
0
B
A
KVO Rate =
5.0 mL/hr
undo
OPTIONS
C
4
ok
B
A
KVO Rate Monitoring Options
Audio Configuration Name
page
OPTIONS
C
4
B
A
KVO Rate =
1.0 mL/hr
undo
OPTIONS
C
4
ok
Page 42
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.11 Setting Dynamic Monitoring
®
System
Features
The Signature Edition
®
Infusion System
incorporates the Dynamic Monitoring
®
System, a versatile intravenous site monitoring system for detection of both full and partial occlusions of the fluid pathway.
This system includes both precision pressure sensing and continuously computed Flow Resistance.
Flow Resistance measurement dramatically improves the ability to detect partial and full occlusions, particularly at low flow rates. Resistance monitoring eliminates the impact of patient elevation and flow rate, to provide the most direct assessment of patient.
Both the Pressure and Flow Resistance monitoring modes feature numeric and bar graph displays of current values plus easily selectable trend graphs, displaying up to twelve hours of monitored values.
The Pressure Mode provides 1 mmHg display resolution, with manual baseline setting, and a user-adjustable alarm limit.
The Resistance Mode provides two ranges and user-alert limit adjustment, to meet neonatal to adult patient needs.
Many monitoring features and functions are configurable, by qualified technical personnel, to meet specific needs by allowing customization.
Monitoring Mode Options
IV lines, catheters, and applications create various levels of resistance to flow. Monitoring mode options are available to meet each clinical application, as follows:
Monitoring Options: To select IV
line/site monitoring modes of resistance, high resistance, and adjustable pressure.
Pressure: Highest occlusion method, when set to a default of 600 mmHg, when Resistance and High Resistance modes are not sufficient.
Resistance: Designed to monitor IV line/site resistance, providing optimum sensitivity. Used for larger catheter sizes (up to 24 gauge).
High Resistance: Designed to monitor IV line/site resistance, with optimum sensitivity where smaller (higher resistance) catheters (26 or 28 gauge) are used.
AutoRest
artPlus
The AutoRestartPlus feature provides the ability to automatically continue an infusion if downstream resistance or pressure measurements indicate that an occlusion condition has cleared with a 40-second Checking Line period (excluding High Resistance monitoring mode). If the condition is not cleared, the OCCLUSION DOWNSTREAM alarm occurs and infusion is stopped until manually restarted.
The Checking Line message and tone are presented whenever a resistance or pressure measurement exceeds its alarm threshold.
In resistance monitoring modes, the Checking Line period is caused by a measured resistance of 100% or a pressure level exceeding a configured threshold. This pressure threshold is separate from the Pressure mode threshold.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-10
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 43
2.4 CONFIGURATION PROCEDURE
(continued)
2.4.11 Setting Dynamic Monitoring
®
System
Features (Continued)
If resistance measurements initiate the Checking Line condition, the channel will continue infusing in order to determine if the measured flow resistance has changed. In Resistance monitoring mode, if the measured flow resistance falls to any value below 100%, the channel will resume normal operating conditions automatically (excluding High Resistance monitoring mode).
In resistance monitoring modes, pressure measurements initiate the Checking Line period when the pressure exceeds the limit. If the pressure falls to less than one third of the configured limit with 40 seconds, normal flow resumes.
In pressure monitoring mode, the
Checking Line period is caused by pressure exceeding the alarm limit. If the pressure falls to less than one third of the alarm limit within 40 seconds, normal flow resumes.
Qualified service personnel can turn off this feature (set restarts to zero), or program from 1 to 9 restarts.
In High Resistance monitoring mode, restarts do not occur for resistance measurements.
After the programmed number of restarts has occurred, the channel will immediately alarm OCCLUSION DOWNSTREAM, if pressure or flow resistance conditions indicate an occlusion. The programmed number of restarts become available again when RUN/HOLD or the run soft key is pressed.
NOTE: The restart counter is reset whenever RUN/HOLD is pressed, the instrument or channel is turned off or an alarm occurs.
Trend Graphs
Trend graphs display downstream pressure or flow resistance over time (up to 12 hours).
In Pressure, Resistance, and High Resistance monitoring modes, the trend graph displays the flow pattern over time. Downstream occlusions, which occur in the
Pressure or Resistance modes, will be indicated by a vertical tick mark at the top of the trend screen.
Trend graphs of fifteen minutes, one hour, four hours and twelve hours are available during normal operation.
1. On C4 page, press Monitoring Options
soft key.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-11
B
A
OPTIONS
KVO Rate Monitoring Options
Audio Configuration Name
C
4
page
B
A
OPTIONS
C
Mode= Pressure Restarts= 3 Trends= On
Undo
4
ok
Page 44
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.11 Setting Dynamic Monitoring
®
System
Features (Continued)
2. Press Mode soft key to select for editing. Press again to cycle between Hi Resist, Resist and Pressure.
3. Press Restarts soft key. Use numeric keypad to enter number of restarts, from "0" (which turns feature off) to "9", then press Enter.
4. Press Trends soft key to select for editing. Press again to cycle between On and Off.
In Pressure mode, graph displays in
mmHg.
Trend data is lost when:
graph information is clearedinstrument is off for more than
12 hours
Trend data is maintained when:
rate is changedbattery is depleted
Trend data is scaled (by 3 times or 1/3, as applicable) when cycling between Resistance and Hi Resistance modes.
5. Press ok to accept changes and return to beginning of C4 page.
6. Precision Flow: Selecting Pressure Mode and running below 50 mL/h will place instrument into Precision Flow Mode.
Precision Flow addresses both flow resolution (parts or pulses per mL) and flow continuity (time between pulses). The instrument will divide each mL into 3600 parts and equally space these parts into a virtually continuous delivery. With Precision Flow, the longest time between a part/pulse at 0.1 mL/h is 10 seconds. As the rate increases, this time will become shorter.
NOTE: Resistance measurement is restarted at 0% when:
• RUN/HOLD is pressed to put on hold and again to start.
• Dose ends in Multi-Dose mode.
Checking Line message appears.
Checking Line alert applies to pressure, resistance and upstream occlusion, with one tone at the beginning and a flashing popup display (on for 4 seconds, off for 6 seconds).
2.4.12 Setting Audio Volume
The volume settings determine which range of audio volume is available. For example, "Low" may be too low, therefore "Med Hi" would be chosen. A transition tone, if enabled, will sound upon completion of a secondary VTBI, step in multi-step mode, dose beginning and ending in multi-dose mode, and completion of a loading dose in loading-dose mode.
The speaker volumes are approximately: Low = 60 dB, Med = 65 dB, and Hi = 70 dB at one meter.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-12
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Clear
Enter
1
23
4
56
7
8
9
.
0
Page 45
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.12 Setting Audio Volume (Continued)
1. On C4 menu, press Audio soft key.
2. Press Volumes soft key to select for
editing. Press again to cycle between Low, Med, Hi, Med Hi, or Hi.
3. Press Trans. Tone soft key to select for editing. Press again to cycle between On and Off.
4. Press ok to accept changes and return to beginning of C4 page.
2.4.13 Setting Configuration Name (Instrument ID Label)
The characters entered here will be shown in the lower LCD display. This electronic label is normally displayed, even when the instrument is off.
The configuration name can be used to uniquely identify the instrument's configuration, facility location, or reference number. It is a 4-character alpha-numeric name; for example, PICU (Pediatric Intensive Care Unit), ICU (Intensive Care Unit), or 2400.
NOTE: A 4-digit alpha-numeric code can be seen next to the configuration name upon entering Configuration Mode (refer to “Entering Configuration Mode” section in this chapter). This code is only a hexadecimal reflection of the instrument’s configuration.
1. On C4 menu, press Configuration
Name soft key.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-13
B
A
KVO Rate Monitoring Options
Audio Configuration Name
OPTIONS
C
4
page
B
A
Audio
Volumes= Low Med Hi
Trans. Tones= On
OPTIONS
C
4
B
A
OPTIONS
Undo
ok
KVO Rate Monitoring Options
Audio Configuration Name
C
4
page
Page 46
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.13 Setting Configuration Name (Instrument ID Label) (Continued)
2. Press Config. Name soft key to position
highlight on character to be changed.
3. Press character soft keys aligned with rows A to M, N to Z or O to "_" to highlight needed character.
4. Press Enter. Up to 4 characters can be programmed in this way. Repeat steps 2, 3, 4 as necessary.
5. Press ok to accept changes and return to beginning of C4 page.
NOTES:
• If the configuration has been changed but not the name, the instrument will display the option to rename before turning off or proceeding to the Teach Mode. Refer to “Pop-Up Displays” section for further explanation.
• There must be 4 characters in the configuration name. Use the space symbol ( _ ) to fill in any open characters.
2.4.14 Resistance Options
Resistance Display: Provides a bar graph
and indication of resistance in the line while monitoring a single channel.
Resistance Alert: Provides an early warning of increases in downstream flow resistance by allowing the level to be adjusted in 5% increments during operation.
Default Alert: Provides a starting point for occlusion detection if a different level has not been chosen. The selected level will remain in effect while the instrument is on, until another option is selected from the Monitoring Options page.
Alarm: Provides the option to lower the pressure occlusion setting that is still active, even in resistance modes.
1. Advance to C5 page.
2. Press Resistance Options soft key.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Resist Display= On Resist Alert = On
Default Alert = 100% Alarm = 600 mmHg
undo
ok
5
OPTIONS
C
A
B
2A-14
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
4
AB
OPTIONS
C
NOPQRSTUVWXYZ
0123456789_
ABCDEF H
JKLMG I
Config. Name= OLDG
undo ok
enter
B
A
Resistance Options Pressure Options
Manual Baseline
OPTIONS
C
5
page
Page 47
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.14 Resistance Options (Continued)
3. Press Resist Display soft key to select
for editing. Press again to cycle between
On and Off.
NOTE: Setting Resist Display to Off will
clear Default Alert from the screen and change Resist Alert to Off.
4. Press Resist Alert soft key to select for
editing. Press again to cycle between
On and Off.
NOTES:
Setting Resist Alert to Off will clear Default Alert from the screen.
Resistance alert will give an alarm tone every thirty seconds, with a popup display (on for 4 seconds, off for 6 seconds). This occurs when the resistance measurement is above the alert threshold or at 100%, even when checking the line. If the display is off, the alert feature will automatically be turned off.
5. Press Default Alert soft key to select for
editing. Enter a number using keypad, then press Enter. Instrument will round up or down to nearest 5% increment.
NOTE: To set the Default Alert level, both Resist Display and Resist Alert need to be On.
6. Press Alarm soft key to select for editing.
Enter a number using keypad, then press Enter.
7. Press ok to accept change and return to
beginning of C5 page.
2.4.15 Pressure Options
Pressure Display: Provides a bar graph
and indication of pressure in the line while monitoring a single channel.
Pressure Alarm: Allows the choice of a fixed pressure mode or an adjustable pressure mode.
Adjustable Pressure: Provides an early warning of increases in downstream pressure. Designed to monitor IV line/site based on pressure, and provide user-adjustable pressure alarm limits. Used for the smallest of catheter sizes and special applications; such as, infusion through transducers, into dialysis systems, and for the highest resistance catheters (including some neonatal PICC catheters).
Fixed Pressure: Designed to monitor IV line/site pressure and alarm, based on a fixed pressure limit of 600 mmHg. Used when adjustable pressure is not wanted. Used for the smallest of catheter sizes and special applications; such as, infusion through transducers, into dialysis systems, and for the highest resistance catheters (including some neonatal PICC catheters).
Maximum Pressure: Provides the maximum pressure alarm limit if the Pressure Alarm mode is set to Adjustable. Any value between 25 mmHg and 600 mmHg (in increments of 25 mmHg) inclusively can be entered.
If Max Press is set to a value less than Def Alarm, the Def Alarm threshold will automatically be set to the Max Press value and Default Alarm Must Be Less Than or Equal to XXX mmHg (which is Max Press level) will be displayed.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-15
Page 48
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.15 Pressure Options (Continued)
Default Alarm: Provides an adjustable
pressure setting in 25 mmHg increments (25-600 range). This value can not exceed the Maximum Pressure that was set above. If an attempt is made to enter a Def Alarm value greater than Max Press, the entry will be rejected with a popup screen and an audible tone.
In addition, the Def Alarm edit field will be highlighted and set to the Max Press value. Pressure Alarm = Adj must be selected to change from 600 mmHg.
1. Advance to C5 page.
2. Press Pressure Options soft key.
3. Press Pressure Dsply soft key to select
for editing. Press again to cycle between On and Off.
4. Press Pressure Alarm soft key to select
for editing. Press again to cycle between Adj and Fix.
5. Press Max Press soft key to select for
editing. Enter a number using keypad, then press Enter. Instrument will round up or down to nearest 25 mmHg increment.
6. Press Def Alarm soft key to select for
editing. Enter a number using keypad, then press Enter. Instrument will round up or down to nearest 25 mmHg increment.
NOTE: Enter the Def Alarm value that is less than or equal to the Max Press value.
7. Press ok to accept change and return to
beginning of C5 page.
2.4.16 Manual Baseline
Auto Baseline: The instrument reads a
baseline at startup and then adds the alarm limit, to determine the point of alarm.
For example, if an alarm limit of 300 mmHg was selected, the instrument would alarm at baseline + 300 mmHg. Auto Zeroing occurs the first time RUN/HOLD is pressed; this will be the maximum baseline. Subsequent presses of RUN/HOLD will only lower (not raise) the baseline.
Manual Baseline: Sets a fixed baseline and overrides the Auto Zero level until the instrument is turned off, the latch is opened, the set is reloaded, or the pressure baseline function is performed again. This allows the instrument to display the actual pressure required for an IV to infuse. To get this real time readout, Manual Baseline must be
On.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-16
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
B
A
Resistance Options Pressure Options
Manual Baseline
OPTIONS
C
5
page
B
A
Pressure Dsply= On Pressure Alarm= Adj
Max Press= 600 mmHg Def Alarm= 600 mmHg
undo
OPTIONS
ok
C
5
Page 49
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.16 Manual Baseline (Continued)
To activate this mode with the instrument on hold, or at startup, press the OPTIONS key and then select Set Pressure Baseline and press ok.
When RUN/HOLD is pressed, the bar graph will show the alarm point with a tic mark on the bar graph, and the actual pressure in the line will be displayed below the bar graph.
1. Advance to C5 page
2. Press Manual Baseline soft key to select
for editing.
3. Press On or Off soft key to cycle between On and Off.
4. Press ok to accept change and return to beginning of C5 page.
2.5 TRANSFERRING SETTINGS TO ANOTHER INSTRUMENT
Once an instrument has been programmed to meet technical and clinical needs, the settings can be transferred to other instruments. The programmed instrument is referred to as the "Teacher" and the other instrument is referred to as the "Learner".
NOTE: Only the configuration settings will be transferred. The instrument ID number, periodic maintenance settings and other settings from the diagnostics mode will not be transferred. The Guardrails
®
Safety Software
data sets will NOT be transferred.
Connect a standard 9-pin Null Modem RS-232 cable (ALARIS P/N 133450) to the RS-232 ports on the instruments.
NOTE: Instrument software versions 4.06 and 4.08 use same "Rev.01.24" and can be used to learn/teach one instrument to the next.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-17
Manual Baseline=
On
5
C
ok
A
B
OPTIONS
C
undo
B
A
Resistance Options
Pressure Options Manual Baseline
OPTIONS
C
5
page
Page 50
2.5 TRANSFERRING SETTINGS TO ANOTHER INSTRUMENT (Continued)
2.5.1 Learn/Teach Instrument Procedure
NOTE: The Learn/Teach function will not
work if the version level for Learn/Teach is not the same on both instruments. It is not recommended to Learn/Teach different model families; such as, using a Model 7130/7230 to Learn/Teach a Model 7131/7231.
1. Access Configuration Mode for both
"teacher" and "learner" instruments and advance to C6 page.
2. Press Teach soft key of "teacher" instrument and Press Learn soft key of "learner" instrument.
3. Press start soft key of "teacher" instrument.
Downloading displays until transfer is complete, and then display indicates if transfer was successful or unsuccessful.
NOTES:
Repeat allows the operator to reattempt
teaching the current instrument or to teach the next instrument.
• Reattach the cover, and/or cover and screws, over the RS-232 port after disconnecting the cable.
• When the profile/data sets are used they will override (replace) the selection in the Configuration Mode. Learn/Teach will need to be done after the profile/data set is loaded to ensure proper use (Profiles On).
• Learn/Teach can be used for transferring the Configuration Mode settings, but not the Guardrails
®
Safety Software data set
(flashed from computer only).
When a Learn/Teach operation is unsuccessful due to incompatible versions, the “teacher” or “learner” instrument will display the message "Learn/Teach Failed Software Mismatch".
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-18
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
AB
Learn: Rev. 01.24 Teach: Rev. 01.24
OPTIONS
C
6
page
AB
Teach Status
OPTIONS
C
6
Awaiting Start
2D15 GOLD
cancel start
AB
Teach Status
OPTIONS
C
6
Learn/Teach Failed Software Mismatch
cancel repeat
Page 51
2.5.2 Pop-Up Displays
The pop-up screens appear when an attempt is made to turn off an instrument or execute the teach mode after changing the configuration (or accessing a configured item), but not changing the configuration name. If the instrument's configuration is changed and not the configuration name, instruments with the same name may have different configurations. The pop-up menus ask if the configuration should be renamed.
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-19
AB
OPTIONS
Configuration has
been changed
Press POWER key
to keep name GOLD
cancel rename
AB
OPTIONS
Configuration has
been changed
Press teach key
to keep name GOLD
cancel rename
teach
C
C
Page 52
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-20
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
#1 #2 #3 #4 #5
§
Instrument ID/Serial #
Config Name (Instr Label)
Location
Regional Settings
Air-In-Line:
Threshold
Reset
Accumulator
µl µl µl µl µl
Profiles
Dose Rate Calculator Generic Dose Rate Calculator only
Maximum Rate mL/h mL/h mL/h mL/h mL/h mL/h
Computer Link:
Mode
Baud Rate
Parity
Optional Modes:
Loading Dose
Dose Rate
Multi-Step
Multi-Dose
Optional Features
Panel Lock
VTBI
Multi-Dose Alert
KVO Rate mL/h mL/h mL/h mL/h mL/h mL/h
Table 2-1. Record of Configured Instruments
--- Table Continued on Next Page ---
Page 53
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-21
#1 #2 #3 #4 #5
Monitoring Options:
Mode
Restarts
Trends
Audio:
Volume
Trans. Tone
Resistance Options:
Resist Display
Resist Alarm
Default Alarm
Alarm
%
mmHg
%
mmHg
%
mmHg
%
mmHg
%
mmHg
Pressure Options:
Pressure Display
Pressure Alarm
Maximum Pressure
Default Alarm
mmHg mmHg
mmHg mmHg
mmHg mmHg
mmHg mmHg
mmHg mmHg
Manual Baseline
§
Instrument ID/Serial # is accessed through Diagnostic Mode. Refer to the “Entering Diagnostic
Mode” section in the “Troubleshooting” chapter.
Table 2-1. Record of Configured Instruments (Continued)
Page 54
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-22
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Figure 2-1. Map of Configuration Screens
Set To Defaults Regional Settings
Air In Line Profiles
C
2
page
OPTIONS
A
B
Threshold=
Air In Line
Reset= Off
Accumulator= On
2
undo
ok
100 mcL
OPTIONS
C
A
B
Profiles Off
C
2
OPTIONS
A
B
ok
undo
3
Maximum Rate
Computer Link Optional Modes
Optional Features
page
OPTIONS
C
A
B
Maximum Rate =
C
3
ok
OPTIONS
A
B
999.9 mL/hr
B
A
CONFIGURATION MODE
ID No. : 001234567
SW Rev.: 04.08
Config.: 2D15 GOLD
page
B
A
OPTIONS
C
1
OPTIONS
cancel
A
Regional Settings
Region= N. America
Language= English
undo
Set all
Configuration
C
2
items to their
default values
ok
B
OPTIONS
C
C 2
ok
Page 55
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2A-23
Figure 2-1. Map of Configuration Screens (Continued)
Mode = Mntr Off Baud Rate = 9600
Parity = Even
C
3
ok
OPTIONS
A
B
undo
3
Loading Dose On
Dose Rate= On Multi-Step= Off
Multi-Dose= Off
OPTIONS
C
A
B
ok
undo
KVO Rate
Monitoring Options Audio
Configuration Name
page
4
OPTIONS
C
A
B
4
5.0 mL/hr
KVO Rate =
ok
OPTIONS
C
undo
A
B
4
Mode= Pressure Restarts= 3 Trends= On
ok
OPTIONS
C
Undo
A
B
4
Volumes= Low Med Hi Trans. Tones= On
Audio
ok
OPTIONS
C
Undo
A
B
4
AB
OPTIONS
C
NOPQRSTUVWXYZ
0123456789_
ABCDEF H
JKLMG I
Config. Name= OLDG
undo ok
enter
B
A
Panel Lock= On VTBI= On
Multi-Dose Alert=Off
undo
OPTIONS
ok
C
3
Page 56
CHECKOUT AND CONFIGURATION
Software Versions 4.06 and 4.08
2A-24
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Manual Baseline=
On
5
C
ok
A
B
OPTIONS
C
undo
Figure 2-1. Map of Configuration Screens
(Continued)
B
A
Resistance Options Pressure Options
Manual Baseline
page
B
A
Resist Display= On
Resist Alert = On Default Alert = 100% Alarm = 600 mmHg
undo
B
A
OPTIONS
OPTIONS
ok
OPTIONS
OPTIONS
C
6
C
5
AB
Learn: Rev. 01.24 Teach: Rev. 01.24
page
AB
C
Learn Status
5
OPTIONS
C
6
Awaiting Download
cancel start
AB
OPTIONS
Pressure Dsply= On
Max Press= 600 mmHg Def Alarm= 600 mmHg
C
Pressure Alarm= Adj
undo
5
ok
Teach Status
Awaiting Start
2D15 GOLD
cancel start
C
6
Page 57
2B
CHECKOUT & CONFIGURATION
Page 58
Chapter 2B — CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-1
2.1 INTRODUCTION
This chapter describes the initial setup and configuration of an instrument with software version 2.78.
2.2 NEW INSTRUMENT CHECKOUT
Refer to the instrument's Directions For Use (DFU) for instructions regarding unpacking and setting up the instrument for the first time.
When turning on the instrument, verify the instrument beeps and that all display LED segments flash. This confirms that the instrument has performed its self test and is operating correctly. During operation, the instrument continually performs a self test, and will alarm and display a message if it detects an internal malfunction.
Contact ALARIS Medical Systems authorized service personnel if the instrument has physical damage, fails to satisfactorily pass startup sequence, fails a self test or continues to alarm.
For new instrument checkout refer to the applicable Signature Edition
®
GOLD Infusion
System DFU.
2.3 CONFIGURABLE OPTIONS AND DEFAULTS
A hospital/facility biomedical technician has the capability to set all configuration parameters to their startup defaults in a single operation. The terms "configuration parameters" and "programmable features" are interchangeable and have the same meaning. Refer to the applicable Signature Edition
®
GOLD Infusion System DFU for
more information.
Keep latch closed when instrument is not in use.
CAUTION
Should an instrument be jarred severely or dropped, remove it from use immediately. It should be thoroughly tested and inspected by qualified service personnel to ensure proper function prior to reuse.
CAUTION
Page 59
CHECKOUT AND CONFIGURATION
2B-2
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
When an instrument’s configuration is
changed, the configuration name should also be changed in order to document the new parameter settings. The intent of the configuration name is to have only one set of parameters for each alpha-numeric code. Refer to “Pop-Up Displays” section.
Powering down in configuration mode during an alarm or error will NOT save any configuration changes.
The configuration procedure is for use by qualified service personnel only. The configuration mode is intended for programming the technical and clinical features in accordance with current procedures and practices.
Table 2-3, “Record of Configured Instruments,” can be reproduced and used to record and track instrument configuration settings.
A map of all the configuration screens is located at the end of this chapter (Figure 2-2).
SEC
PRI
Clear
Enter
OPT PRI HLD SEC KVO
1
RUN
HOLD
POWER
OPTIONS
23
4
56
7
8
9
.
0
ml/hr
WARNINGS
NOTES:
Pressing a soft key at the side of the Main Display the first time, selects it for editing. Some features are edited by subsequent presses of the soft key to cycle through available options. Other features are edited by means of the numeric keypad entry.
Pressing undo undoes any edits made to that page, and stays on the page.
Pressing ok accepts all information on the page, and returns to menu page.
Pressing the POWER switch after editing a configurable item evokes an invalid key tone and a message to "ok entry". ok must be selected to accept the edit before the instrument can be powered off.
The page number is located in the upper­right corner of the display.
Software Version 2.78
Page 60
2.4 CONFIGURATION PROCEDURE
(Continued)
The instrument’s configuration information is not lost when disconnected from AC power. However, error history and infusion program settings may be lost. If this information is to be saved, be sure to record it before disconnecting power.
2.4.1 Entering Configuration Mode
The instrument must be off (both channels must be off for dual-channel).
The procedure for the single and dual channel instrument is the same. Any configuration in the dual channel sets the same value for both channels.
1. Press and hold left-bottom display soft key.
2. Press POWER switch. Continue to press display soft key until configuration mode display appears, then release.
Page C1 of configuration mode is displayed. This is a read-only display.
3. Press page to advance to page to be configured.
NOTE: For 7131/7231, the default code will be 4EF3 (instead of BFFF) for instruments set to factory defaults.
2.4.2 Setting to Defaults
The Set To Defaults mode programs all configuration items to their default values.
It is strongly recommended that the Configurable Options be reviewed for a complete list of defaults before using this feature.
1. Advance to C2 page.
2. Press Set to Defaults soft key.
3. Press ok to accept change and return to beginning of C2 page. Pressing cancel leaves all items set to their previous values and returns to beginning of C2 page.
CHECKOUT AND CONFIGURATION
Software Version 2.78
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-3
CAUTION
CAUTION
B
A
CONFIGURATION MODE
ID No. : 001234567
SW Rev.: 02.78
Config.: BFFF IVAC
OPTIONS
page
C
1
B
A
Set To Defaults Language
Air In Line
Dose Rate Drugs
page
OPTIONS
B
A
Set All
configuration
items to their
default values
cancel
C
2
OPTIONS
C
2
ok
Page 61
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.3 Setting Language
Language choices for Models 7130/7230 are English and Canadian French. Language choices for Models 7131/7231 are Dutch, English, French, German, Italian, Spanish and Swedish. All displays will be in language selected, except diagnostic and configuration modes, which remain in English.
1. From C2 page, press Language soft key.
2. Press soft key to select for editing. Press again to cycle through language choices.
NOTE: Dutch and German are choices in the language options, but labels and DFUs are not available in these languages for this software version.
3. Press ok to accept change and return to beginning of C2 page.
2.4.4 Setting Air-in-Line Threshold
The air-in-line threshold sets the bubble size sensitivity. The air-in-line reset allows the clinician to respond to an “Air-in-Line” alarm, assess the clinical significance of the air, and choose whether or not to continue the infusion without removing the air. The reset feature allows only the current bubble to proceed without tripping an alarm. The air in line threshold value choices are 50, 100, 200 and 500 microliters.
The Accumulator setting, when on, looks for 10% to 15% of the downstream path to be air before giving an “Accumulated Air-in­Line” alarm. The amount of air that causes the alarm will vary with the threshold setting and rate.
An “Air-in-Line” alarm is the result of the bubble size exceeding the Threshold
setting. An “Accumulated Air-in-Line” alarm is the result when 10 to 15% of the downstream path is filled with air, exceeding the accepted level looked for when Accumulator is set to On.
1. From C2 page, press Air In Line soft
key.
2. Press Threshold soft key to select for editing. Press again to cycle through 50
mcL, 100 mcL, 200 mcL and 500 mcL.
CHECKOUT AND CONFIGURATION
2B-4
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Software Version 2.78
B
A
Set To Defaults Language
Air In Line
Dose Rate Drugs
OPTIONS
page
A
B
OPTIONS
C
2
Language=
English
undo
ok
C
C
B
2
A
Set To Defaults
Language Air In Line Dose Rate Drugs
OPTIONS
C
2
page
Page 62
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.4 Setting Air-in-Line Threshold
(Continued)
NOTE: Use a 50 microliters setting on microbore tubing. All other settings may be used on macrobore tubing.
3. Press Reset soft key to select for editing. Press again to cycle between On and Off.
4. Press Accumulator soft key to select for editing. Press again to cycle between On and Off.
5. Press ok to accept change and return to beginning of C2 page.
2.4.5 Setting Dose Rate Drugs
NOTE: In Models 7131/7231, Drug? is used
instead of a drug list option. There is no tick mark next to Dose Rate.
The Dose Rate Drugs feature allows selection of a drug name to program a dose rate calculated infusion while in normal mode.
1. From C2 page, press Dose Rate Drugs
soft key.
Introduction screen appears and displays legend used when selecting a drug.
2. Page forward to select drug or select
done to go to Summary Page.
Use short list symbol “s” to select drugs that appear in short list. This is a convenient way to display frequently used drugs immediately in normal drug selection process (startup mode, press options key).
Use extended list symbol “e” when selecting drugs to appear in extended list. Extended list, if configured, provides a secondary list of drug names not normally used.
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-5
Software Version 2.78
B
A
OPTIONS
B
A
Air In Line
Threshold=
Reset= Off
Accumulator= On
undo
100 mcL
OPTIONS
ok
Set To Defaults
Language Air In Line Dose Rate Drugs
C
2
page
C
2
B
A
Dose Rate Drugs
(s) = short list
(e) = extended list
( ) = not displayed
done
OPTIONS
C
2
page
Page 63
CHECKOUT AND CONFIGURATION
2B-6
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Alfentanil (Alfent) Methylprednisolone
Alprostadil (Prostin) (PGE-1)
Milrinone
Alteplase (Activase) Morphene
Aminophylline Nitroglucerin (Tridil)
Amrione (Inocar) Nitroprusside (Nipride)
Atracurium (Tracurium) Norepinephrine (Levophed)
Bretylium (Bretylol) Oxytocin
Cimetidine (Tagamet) Phenylephrine (Neo-Synephrine)
Diltiazem (Cardizem) Potassium Chlor
Dobutamine (Dobutrex) Procainamide
Dopamine (Intropin) Propofol (Diprivan)
Esmilol (Brevibloc) Ranitidine (Zantac)
Heparin Streptokinase (Streptase)
Isoproterenol (Isuprel) Succinylcholine (Anectine)
Labetalo (Normodyne) (Trandate)
Theophylline
Lidocaine (Xylocaine) Urokinase (Abbokinase)
Magnesium Sulfate Vecuronium (Norcuron)
Table 2-2. Drug List (7130/7230 only)
Software Version 2.78
Page 64
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.5 Setting Dose Rate Drugs (continued)
3. Press done at any time to display Summary Page. Press page to advance to next page.
4. Press Extended List soft key to cycle between “s” and “blank”. If “blank” is selected, extended list will not be available.
5. Press Drug? soft key to cycle between “s”, “e” and “blank”. Use “s” for commonly used drugs.
6. Press a soft key next to a drug to cycle between “s”, “e” and “blank”. Press Page to continue viewing drug list. Refer to Table 2-2 “Drug List” to view full list of drugs.
7. Press done to display Summary Page.
8. Press review to return to introduction screen or press ok to accept selections and return to C2 page.
2.4.6 Setting Maximum Rate
This sets the maximum selectable rate. The range for Maximum Rate is 0.1 to
999.9 mL/h.
NOTES:
• The maximum rate setting applies to all infusion modes.
• Setting the Maximum Rate below the preset KVO Rate will lower the KVO Rate. The KVO Rate will not exceed the Maximum Rate.
1. Advance to C3 page.
2. Press Maximum Rate soft key.
3. Press the soft key to select the value for editing.
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-7
Software Version 2.78
C
3
OPTIONS
Maximum Rate =
999.9 mL/hr
ok
A
B
B
A
EXTENDED LIST s-Drug?
-Alfentanil
-Alprostadil
done
page
OPTIONS
C
2
B
A
Maximum Rate Computer Link
Optional Modes
Optional Features
OPTIONS
page
C
3
B
A
Summary
14 drugs selected
EXT D LIST selected
Drug? Selected
review
OPTIONS
C
2
ok
Page 65
CHECKOUT AND CONFIGURATION
2B-8
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.6 Setting Maximum Rate (Continued)
4. Use numeric keypad to enter maximum
rate. Press Enter.
5. Press ok to accept change and return to
beginning of C3 page.
2.4.7 Setting Computer Link
The Computer Link feature allows a hospital/facility computer to interact with the instrument and programs the level of computer control available. The computer cannot start or stop the instrument, set the rate, or make any change in status. If the feature is off, the computer cannot communicate with the instrument.
Control Mode allows computer to send information to instrument’s display.
Monitor Mode allows computer to only receive information from instrument. Enabling of monitoring mode automatically places it in options menu.
Off Mode does not allow any communication between instrument and a computer.
1. On C3 page, press Computer Link soft
key.
2. Press Mode soft key to select for editing.
Press again to cycle through Off, Mntr Off and Ctrl Mntr Off.
3. Press Baud Rate soft key to select for
editing. Press again to cycle through choices (300, 600, 1200, 1800, 2400, 4800 and 9600).
4. Press Parity soft key to select for editing.
Press again to cycle through Even, Odd, and None.
5. Press ok to accept changes and return to
beginning of C3 page.
Software Version 2.78
Clear
Enter
1
23
4
56
7
8
9
.
0
B
A
Maximum Rate
Computer Link
Optional Modes
Optional Features
OPTIONS
C
3
page
B
A
Mode = Off Baud Rate = 9600
Parity = None
OPTIONS
C
3
okundo
Page 66
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-9
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.8 Setting Optional Modes
The Optional Modes feature allows the options menu to be configured to how it will appear in normal mode. Enabling of these modes automatically places them in the option menu.
When Loading Dose, Multi-Step, and Multi-Dose are on, they appear in the menu when the OPTIONS key is pressed in normal mode.
For Dose Rate to appear in the options menu, it must be turned on.
1. On C3 page, press Optional Modes soft
key.
2. Press Loading Dose soft key to select
for editing. Press again to cycle between
On and Off.
NOTE: This feature allows an initial
infusion rate to be set up for a specific volume, automatically followed by a maintenance rate from the same container.
3. Press Dose Rate soft key to select for editing. Press again to cycle between On and Off.
NOTE: This feature allows dose parameters to be programmed so that the instrument then calculates the volumetric rate.
4. Press Multi-Step soft key to select for editing. Press again to cycle between On and Off.
NOTE: This feature allows a sequential program to deliver up to nine steps; fluid volumes and delivery rates may be programmed for each step.
5. Press Multi-Dose soft key to select for editing. Press again to cycle between
On and Off.
NOTE: This feature allows multiple
infusions to be programmed over a period of up to 24 hours; the fluid volume and delivery rate is repeated for each delivery.
6. Press ok to accept changes and return to beginning of C3 page.
Software Version 2.78
B
A
Maximum Rate Computer Link
Optional Modes
Optional Features
OPTIONS
C
3
page
B
A
Loading Dose On
Dose Rate = On Multi-Step = Off Multi-Dose = Off
undo
OPTIONS
C
3
ok
B
A
Loading Dose On Dose Rate= On
Multi-Step= Off
Multi-Dose= Off
undo
OPTIONS
C
3
ok
Page 67
CHECKOUT AND CONFIGURATION
2B-10
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.9 Setting Optional Features
The Panel Lock feature allows the front panel to be locked and unlocked to help prevent tampering.
When VTBI (volume to be infused) is on, a volume to be infused must be entered; otherwise, the last remaining VTBI or last entered VTBI, depending on the channel's last usage, will be in effect. When VTBI is off, there is no VTBI line capability.
When the Multi-Dose Alert feature is on, the option is available to set an alert at the end of every dose when in the Multi-Dose mode.
1. On C3 page, press Optional Features
soft key.
2. Press Panel Lock soft key to select for
editing. Press again to cycle between On and Off.
3. Verify VTBI is On.
NOTE: The flow sensor option (empty container detector) needs to be installed to be able to turn VTBI off.
4. Press Multi-Dose Alert soft key to select
for editing. Press again to cycle between On and Off.
5. Press ok to accept changes and return
to beginning of C3 page.
2.4.10 Setting KVO Rate
The instrument will automatically operate at the KVO rate (or current rate, whichever is less) once the primary VTBI has counted down to zero. The KVO rate range is 0.1 to
20.0 mL/h.
1. Advance to C4 page.
2. Press KVO Rate soft key.
Software Version 2.78
B
A
Maximum Rate
Computer Link Optional Modes
Optional Features
OPTIONS
C
3
page
B
A
KVO Rate Monitoring Options
Audio Configuration Name
OPTIONS
C
4
B
A
Panel Lock= On VTBI= On
Multi-Dose Alert=Off
undo
OPTIONS
ok
C
3
page
B
A
KVO Rate =
5.0 mL/hr
undo
OPTIONS
C
4
ok
Page 68
CHECKOUT AND CONFIGURATION
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.10 Setting KVO Rate (Continued)
3. Press soft key next to rate value once to select for editing.
4. Use the numeric keypad to enter KVO rate. Press Enter.
5. Press ok to accept change and return to beginning of C4 page.
NOTE: The KVO Rate will NOT exceed the present Maximum Rate.
2.4.11 Setting Dynamic Monitoring®System Features
The Signature Edition
®
Infusion System
incorporates the Dynamic Monitoring
®
System, a versatile intravenous site monitoring system for detection of both full and partial occlusions of the fluid pathway.
This system includes both precision pressure sensing and continuously computed Flow Resistance.
Flow Resistance measurement dramatically improves the ability to detect partial and full occlusions, particularly at low flow rates. Resistance monitoring eliminates the impact of patient elevation and flow rate, to provide the most direct assessment of patency.
Both the Pressure and Flow Resistance monitoring modes feature numeric and bar graph displays of current values plus easily selectable trend graphs, displaying up to twelve hours of monitored values.
The Pressure Mode provides 1 mmHg display resolution, either manual or auto baseline setting, and a user-adjustable alarm limit.
The Resistance Mode provides two ranges and user-alert limit adjustment, to meet neonatal to adult patients’ needs.
Many monitoring features and functions are configurable, by qualified technical personnel, to meet specific needs by allowing customization.
Monitoring Mode Options
IV lines, catheters, and applications create various levels of resistance to flow. Monitoring mode options are available to meet each clinical application, as follows:
Monitoring Options: to select IV line/site
monitoring modes of resistance, high resistance, and adjustable pressure.
Pressure: highest occlusion method, when set to a default of 600 mmHg, when Resistance and High Resistance modes are not sufficient.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-11
Software Version 2.78
Clear
Enter
1
23
4
56
7
8
9
.
0
B
A
KVO Rate =
1.0 mL/hr
undo
OPTIONS
ok
C
4
Page 69
CHECKOUT AND CONFIGURATION
2B-12
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.11 Setting Dynamic Monitoring
®
System Features (Continued)
Resistance: designed to monitor IV
line/site resistance, providing optimum sensitivity. Used for larger catheter sizes (up to 24 gauge).
High Resistance: designed to monitor IV line/site resistance, with optimum sensitivity where smaller (higher resistance) catheters (26 or 28 gauge) are used.
AutoRest
artPlus
The AutoRestartPlus feature provides the ability to automatically continue an infusion if downstream resistance or pressure measurements indicate that an occlusion condition has cleared with a 40-second Checking Line period (excluding High Resistance monitoring mode). If the condition is not cleared, the OCCLUSION DOWNSTREAM alarm occurs and infusion is stopped until manually restarted.
The Checking Line message and tone are presented whenever a resistance or pressure measurement exceeds its alarm threshold.
In resistance monitoring modes, the
Checking Line period is caused by a measured resistance of 100% or a pressure level exceeding a configured threshold. This pressure threshold is separate from the Pressure mode threshold.
If resistance measurements initiate the Checking Line condition, the channel will continue infusing in order to determine if the measured flow resistance has changed. In Resistance monitoring mode, if the measured flow resistance falls to any value below 100%, the channel will resume normal operating conditions automatically (excluding High Resistance monitoring mode).
In resistance monitoring modes, pressure measurements initiate the Checking Line period when the pressure exceeds the limit. If the pressure falls to less than one third of the configured limit within 40 seconds, normal flow resumes.
In Pressure monitoring mode, the Checking Line period is caused by pressure exceeding the alarm limit. If the pressure falls to less than one third of the alarm limit within 40 seconds, normal flow resumes.
Qualified service personnel can turn off this feature (set restarts to zero), or program from 1 to 9 restarts.
In High Resistance monitoring mode,
restarts do not occur for resistance measurements.
After the programmed number of restarts has occurred the channel will immediately alarm OCCLUSION DOWNSTREAM, if pressure or flow resistance conditions indicate an occlusion. The programmed number of restarts become available again when RUN/HOLD or the run soft key is pressed.
NOTE: The restart counter is reset whenever RUN/HOLD is pressed, the instrument or channel is turned off or an alarm occurs.
Software Version 2.78
Page 70
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-13
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.11 Setting Dynamic Monitoring
®
System Features (Continued)
T
rend Graphs
Trend graphs display downstream pressure or flow resistance over time (up to 12 hours).
In Pressure, Resistance, and High Resistance monitoring modes, the trend graph displays the flow pattern over time. Downstream occlusions, which occur in the Pressure or Resistance modes, will be indicated by a vertical tick mark at the top of the trend screen.
Trend graphs of fifteen minutes, one hour, four hours, and twelve hours are available during normal operation.
1. On C4 page, press Monitoring Options
soft key.
2. Press Mode soft key to select for editing. Press again to cycle between Hi Resist, Resist, and Pressure.
3. Press Restarts soft key. Use numeric keypad to enter number of restarts, from "0" (which turns feature off) to "9", then press Enter.
4. Press Trends soft key to select for editing. Press again to cycle between On and Off.
In Pressure mode, graph displays in
mmHg.
Trend data is lost when:
graph information is clearedinstrument is off for more than
6 hours
Trend data is maintained when:
rate is changedinstrument is off for less than
6 hours
Trend data is scaled (by 3 times or by
1/3, as applicable) when cycling between Resistance and Hi Resistance modes.
5. Press ok to accept changes and return to beginning of C4 page.
6. Precision Flow: Selecting Pressure Mode and running below 50 mL/h will place instrument into Precision Flow Mode.
Precision Flow addresses both flow resolution (parts or pulses per mL) and flow continuity (time between pulses). The instrument will divide each mL into 3600 parts and equally space these parts into a virtually continuous delivery. With Precision Flow the longest time between a part/pulse at 0.1 mL/h is 10 seconds. As the rate increases, this time becomes shorter.
Software Version 2.78
B
A
KVO Rate Monitoring Options
Audio Configuration Name
OPTIONS
C
4
page
B
A
Monitoring Options
Mode= Pressure Restarts= 3 Trends= On
undo
OPTIONS
C
4
ok
Page 71
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.11 Setting Dynamic Monitoring
®
System Features (Continued)
NOTE: Resistance measurement is
restarted at 0% when:
RUN/HOLD is pressed to put on hold and again to start.
Dose ends in Multi-Dose mode.
Checking Line message appears.
Checking Line alert applies to pressure, resistance and upstream occlusion, with one tone at the beginning and a flashing popup display (on for 4 seconds, off for 6 seconds). No nurse call activation.
2.4.12 Setting Audio Volume
The volume settings determine which range of audio volume is available. For example, "Low" may be too low, so "Med Hi" would be chosen. A transition tone, if enabled, will sound upon completion of a secondary VTBI, step in multi-step mode, dose beginning and ending in multi-dose mode, and completion of a loading dose in loading­dose mode.
The speaker volumes are approximately: Low = 60 dB, Med = 65 dB, and Hi = 70 dB at one meter.
1. On C4 menu, press Audio soft key.
2. Press Volumes soft key to select for
editing. Press again to cycle between Low, Med, Hi, Med/Hi, or Hi.
3. Press Trans. Tone soft key to select for editing. Press again to cycle between On and Off
4. Press ok to accept changes and return to beginning of C4 page.
2.4.13 Setting Configuration Name (Instrument ID Label)
The characters entered here will be shown in the lower LCD display. This electronic label is normally displayed, even when the pump is off.
The configuration name can be used to uniquely identify the instrument's configuration, facility location, or reference number. It is a 4 character alpha-numeric name. For example, PICU (Pediatric Intensive Care Unit), ICU (Intensive Care Unit), or 2400.
NOTE: A 4-digit alpha-numeric code can be seen next to the configuration name upon entering Configuration Mode (refer to “Entering Configuration Mode” section in this chapter). This code is only a hexadecimal reflection of the instrument’s configuration.
CHECKOUT AND CONFIGURATION
2B-14
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
KVO Rate Monitoring Options
Audio Configuration Name
page
4
OPTIONS
C
A
B
Software Version 2.78
B
A
Audio
Volumes = Low Med Hi Trans. Tones = On
undo
OPTIONS
ok
C
4
Page 72
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-15
2.4 CONFIGURATION PROCEDURE
(Continued)
2.4.13 Setting Configuration Name (Instrument ID Label) (Continued)
1. On C4 menu, press Configuration Name soft key.
2. Press Config. Name soft key to position highlight on character to be changed.
3. Press character soft keys aligned with rows A to M, N to Z or O to "_" to highlight desired character.
4. Press Enter. Up to 4 characters can be programmed in this way. Repeat steps 2, 3, 4 as necessary.
5. Press ok to accept changes and return to beginning of C4 page.
NOTES:
• If the configuration has been changed, but not the name, the instrument will display the option to rename before turning off or proceeding to the Teach Mode. Refer to “Pop-Up Displays” section for further explanation.
• There must be 4 characters in the configuration name. Use the space symbol ( _ ) to fill any open characters.
2.5 TRANSFERRING SETTINGS TO ANOTHER INSTRUMENT
Once an instrument has been programmed to meet technical and clinical needs, the settings can be transferred to other instruments. The programmed instrument is referred to as the "Teacher" and the other instrument is referred to as the "Learner".
NOTE: Only the configuration settings will be transferred. The instrument ID number, periodic maintenance settings, and other settings from the diagnostics mode will not be transferred.
Connect a standard 9-pin Null Modem RS-232 cable (ALARIS P/N 133450) to the RS-232 ports on the instruments.
2.5.1 Learn/Teach Instrument Procedure
NOTE: The Learn/Teach function will not
work if the revision level for Learn/Teach is not the same on both instruments. It is not recommended to Learn/Teach different model families; such as, using a Model 7130/7230 to Learn/Teach a Model 7131/7231.
Software Version 2.78
B
A
KVO Rate
Monitoring Options Audio
Configuration Name
OPTIONS
C
4
page
AB
Config. Name= IVAC
A B C D E F G H I J K L M
N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 _
undo ok
enter
OPTIONS
C
4
Page 73
CHECKOUT AND CONFIGURATION
2B-16
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.5 TRANSFERRING SETTINGS TO ANOTHER INSTRUMENT (Continued)
2.5.1 Learn/Teach Instrument Procedure (Continued)
1. Access Configuration Mode for both
"teacher" and "learner" instruments and advance to page C5.
2. Press Teach soft key of "teacher" instrument and Press the Learn soft key of "learner" instrument.
3. Press start soft key on "teacher" instrument.
Downloading displays until transfer is complete, and then displays if transfer was successful or unsuccessful.
NOTES:
Repeat allows the operator to
reattempt teaching the current instrument or to teach the next instrument.
• Reattach the cover, and/or cover and screws, over RS-232 port after disconnecting the cable.
2.5.2 Pop-Up Displays
The pop-up screens appear when an attempt is made to turn off an instrument or execute the teach mode after changing the configuration (or accessing a configured item) but not changing the configuration name. If the instrument's configuration is changed and not the configuration name, instruments with the same name may have different configurations. The pop-up menus ask if the configuration should be renamed.
Software Version 2.78
AB
Learn: Rev. 01.20 Teach: Rev. 01.20
OPTIONS
C
5
Resistance Options Pressure Options
page
AB
OPTIONS
Teach Status
Awaiting Start
BFFF IVAC
cancel start
AB
C
5
Configuration has
been changed.
Press POWER key
OPTIONS
C
to keep name IVAC.
cancel rename
AB
OPTIONS
Configuration has
been changed.
Press teach key
to keep name IVAC.
cancel rename
teach
C
Page 74
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-17
2.6 RESISTANCE OPTIONS
Resistance Display: Provides a bar graph
and indication of resistance in the line while monitoring a single channel.
Resistance Alert: Provides an early warning of increases in downstream flow resistance by allowing the level to be adjusted in 5% increments during operation.
Default Alert: Provides a starting point for occlusion detection if a different level has not been chosen. The selected level will remain in effect while the instrument is on, until another option is selected from the Monitoring Options page. The channel will retain the selected option up to six hours after the instrument is turned off. After six hours the channel will return to the default setting.
Alarm: Provides the option to lower the pressure occlusion setting that is still active, even in resistance modes.
1. Advance to C5 page.
2. Press Resist Display soft key to select
for editing. Press again to cycle between
On and Off.
NOTE: Setting Resist Display to Off will
clear Default Alert from the screen and change Resist Alert to Off.
3. Press Resist Alert soft key to select for
editing. Press again to cycle between
On and Off.
NOTES:
Setting Resist Alert to Off will clear Default Alert from the screen.
Resistance alert will give an alarm tone every thirty seconds, with a popup display (on for 4 seconds, off for 6 seconds). Nurse call activated. This occurs when the resistance measurement is above the alert threshold or at 100%, even when checking the line. If the display is off, the alert feature will automatically be turned off.
4. Press Default Alert soft key to select for
editing. Enter a number using keypad, then press Enter. The instrument will round up or down to the nearest 5% increment.
NOTE: To set the Default Alert level, both Resist Display and Resist Alert need to be On.
5. Press Alarm soft key to select for editing.
Enter a number using keypad, then press Enter.
6. Press ok to accept change and return to
beginning of C5 page.
Software Version 2.78
AB
OPTIONS
Learn: Rev. 01.20 Teach: Rev. 01.20
Resistance Options Pressure Options
C
5
B
A
Resist Display= On Resist Alert = On
Default Alert = 100% Alarm = 600 mmHg
undo
page
OPTIONS
C
5
ok
Page 75
CHECKOUT AND CONFIGURATION
2B-18
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2.7 PRESSURE OPTIONS
Pressure Display: Provides a bar graph
and indication of pressure in the line while monitoring a single channel.
Pressure Alarm: Allows the choice of a fixed pressure mode or an adjustable pressure mode.
Adjustable Pressure: Provides an early warning of increases in downstream pressure. Designed to monitor IV line/site based on pressure, and provide user- adjustable pressure alarm limits. Used for the smallest of catheter sizes and special applications; such as infusion through transducers, into dialysis systems, and for the highest resistance catheters (including some neonatal PICC catheters).
Fixed Pressure: Designed to monitor IV line/site pressure and alarm, based on a fixed pressure limit of 600 mmHg. Used when adjustable pressure is not wanted. Used for the smallest of catheter sizes and special applications; such as, infusion through transducers, into dialysis systems, and for the highest resistance catheters (including some neonatal PICC catheters).
Alarm: Provides an adjustable pressure setting in 25 mmHg increments (25-600 range). This acts as the default setting after the instrument has been off for more than 6 hours and the previous setting was cleared. Pressure Alarm = Adj must be selected to change from 600 mmHg.
Auto Baseline: The instrument reads a baseline at startup and then adds the alarm limit, to determine the point of alarm. For example, if an alarm limit of 300 mmHg was selected, the instrument would alarm at baseline + 300 mmHg. Auto Zeroing occurs
the first time RUN/HOLD is pressed; this will be the maximum baseline. Subsequent presses of RUN/HOLD will only lower (not raise) the baseline.
Manual Baseline: Sets a fixed baseline and overrides the Auto Zero level until the instrument is turned off, the latch is opened, the set is reloaded, or the pressure baseline function is performed again. This allows the instrument to display the actual pressure required for an IV to infuse. To get this real time readout, Manual Baseline must be On.
To activate this mode with the instrument on hold, or at startup, press the Options key and then select Set Pressure Baseline and press ok.
When RUN/HOLD is pressed, the bar graph will show the alarm point and the actual pressure in the line will be displayed below the bar graph.
1. Advance to C5 page.
Software Version 2.78
AB
Learn: Rev. 01.20 Teach: Rev. 01.20
OPTIONS
C
5
Resistance Options Pressure Options
page
B
A
Pressure Dsply= On
Pressure Alarm= Adj Alarm = 600 mmHg Manual Baseline=On
undo
OPTIONS
C
5
ok
Page 76
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-19
2.7 PRESSURE OPTIONS (Continued)
2. Press Pressure Dsply soft key to select
for editing. Press again to cycle between On and Off.
3. Press Pressure Alarm soft key to select
for editing. Press again to cycle between Adj and Fix.
4. Press Alarm soft key to select for editing.
Enter a number using keypad, then press Enter. Instrument will round up or down to nearest 25 mmHg increment.
5. Press Manual Baseline soft key to select
for editing. Press again to cycle between On and Off.
6. Press ok to accept change and return to
beginning of C5 page.
Software Version 2.78
Page 77
CHECKOUT AND CONFIGURATION
2B-20
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
#1 #2 #3 #4 #5
§
Instrument ID/Serial #
Config Name (Instr Label)
Location
Language
Air-In-Line:
Threshold
Reset
Accumulator
µl µl µl µl µl
Dose Rate Drugs (See Table 2-4 “Drug List “)
Maximum Rate mL/h mL/h mL/h mL/h mL/h mL/h
Computer Link:
Mode
Baud Rate
Parity
Optional Modes:
Loading Dose
Dose Rate
Multi-Step
Multi-Dose
Optional Features
Panel Lock
VTBI
Multi-Dose Alert
KVO Rate mL/h mL/h mL/h mL/h mL/h mL/h
Table 2-3. Record of Configured Instruments
--- Table Continued on Next Page ---
Software Version 2.78
Page 78
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-21
#1 #2 #3 #4 #5
Monitoring Options:
Mode
Restarts
Trends
Audio:
Volume
Trans. Tone
Resistance Options:
Resist Display
Resist Alarm
Default Alarm
Alarm
%
mmHg
%
mmHg
%
mmHg
%
mmHg
%
mmHg
Pressure Options:
Pressure Display
Pressure Alarm
Alarm
Manual Baseline
mmHg mmHg mmHg mmHg mmHg
§
Instrument ID/Serial # is accessed through Diagnostic Mode. Refer to the “Entering Diagnostic
Mode” section in the “Troubleshooting” chapter.
Table 2-3. Record of Configured Instruments (Continued)
Software Version 2.78
Page 79
CHECKOUT AND CONFIGURATION
2B-22
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
DRUG NAME/OPTION TRADE NAME NOT SELECTED SHORT LIST EXTENDED LIST
Extended List Access
DRUG?
Alfentanil Alfenta
Alprostadil Prostin (PGE-1)
Alteplase Activase
AminoPhylline
Amrinone Inocar
Atracurium Tracurium
Bretylium Bretylol
Cimetidine Tagamet
Diltiazem Cardizem
Dobutamine Dobutrex
Dopamine Intropin
Esmolol Brevibloc
Heparin
Isoproterenol Isuprel
Labetalol Normodyne /Trandate
Lidocaine Xylocaine
Magnesium Sulfate
Methylprednisolone
Milrinone
Morphine
Nytroglycerin Tridil
Nitroprusside Nipride
Norepinephrine Levophed
Ocytocin
Phenlyphrine Neo-Synephrine
Potassium Chlor
Procainamide
Propofol Dipravan
Ranitidine Zantac
Streptokinase Streptase
Succinylcholine Anectine
Theophylline
Urokinase Abbokinase
Vecuronium Norcuron
Facility: Unit:
Table 2-4. Drug List Configuration
Software Version 2.78
Page 80
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-23
Figure 2-2. Map of Configuration Screens
Set To Defaults Language
Air In Line
Dose Rate Drugs
C
2
page
OPTIONS
A
B
Language=
English
2
ok
OPTIONS
C
undo
C
A
B
Threshold=
Air In Line
Reset= Off
Accumulator= On
2
undo
ok
100 mcL
OPTIONS
C
A
B
OPTIONS
Dose Rate Drugs
( ) = not displayed
C
page
2
(s) = short list
(e) = extended list
A
B
done
Software Version 2.78
B
A
CONFIGURATION MODE
ID No. : 001234567
SW Rev.: 02.78
Config.: BFFF IVAC
page
B
A
OPTIONS
C
1
OPTIONS
B
A
OPTIONS
cancel
Set All
configuration
items to their
default values
ok
C
2
Maximum Rate Computer Link
Optional Modes
Optional Features
page
B
A
Maximum Rate =
999.9 mL/hr
OPTIONS
ok
C
3
C
3
Page 81
CHECKOUT AND CONFIGURATION
2B-24
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Figure 2-2. Map of Configuration Screens (Continued)
Software Version 2.78
B
A
Mode = Off Baud Rate = 9600
Parity = None
B
A
Loading Dose On
Dose Rate = On Multi-Step = Off Multi-Dose = Off
undo
B
A
Panel Lock= On VTBI= On
Multi-Dose Alert=Off
OPTIONS
okundo
OPTIONS
ok
OPTIONS
C
3
C
3
C
3
B
A
KVO Rate =
5.0 mL/hr
undo
B
A
Monitoring Options
Mode= Pressure Restarts= 3 Trends= On
undo
B
A
Audio
Volumes = Low Med Hi Trans. Tones = On
OPTIONS
C
4
ok
OPTIONS
C
4
ok
OPTIONS
C
4
undo
B
A
KVO Rate Monitoring Options
Audio
Configuration Name
ok
OPTIONS
C
4
page
undo
AB
Config. Name= IVAC
A B C D E F G H I J K L M
N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 _
undo ok
enter
ok
OPTIONS
C
4
Page 82
CHECKOUT AND CONFIGURATION
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-25
Figure 2-2. Map of Configuration Screens (Continued)
5
Teach Status
Awaiting Start
BFFF IVAC
cancel start
AB
OPTIONS
C
Software Version 2.78
AB
Learn: Rev. 01.20 Teach: Rev. 01.20
Resistance Options Pressure Options
page
B
A
Learn Status
Awaiting Download
cancel
OPTIONS
C
5
OPTIONS
C
5
B
A
Resistance Display = On
Resistance Alert = On Default Alert = 100% Alarm = 600 mmHg
undo
B
A
Pressure Display = On
Pressure Alarm = Adj Alarm = 600 mmHg Manual Baseline = On
undo
OPTIONS
C
5
ok
OPTIONS
C
5
ok
Page 83
CHECKOUT AND CONFIGURATION
2B-26
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
THIS PAGE
INTENTIONALLY
LEFT BLANK
Software Version 2.78
Page 84
3
PREVENTIVE MAINTENANCE
Page 85
Chapter 3 — PREVENTIVE MAINTENANCE
3-1
3.1 INTRODUCTION
To ensure the Signature Edition
®
GOLD Infusion System remains in good operating condition, regular and preventive maintenance inspections are required. Regular inspections must be performed by hospital/facility before each use.
Preventive maintenance inspections should be performed once a year in accordance with ALARIS Medical Systems requirements and guidelines. Amaintenance reminder will occur after 52 weeks, unless the feature has been changed to select a different time interval or has been disabled. These inspections are also intended to complement the intent of JCAHO requirements.
Use Table 3-1, "PM Inspections," to record the completion of preventive maintenance inspections.
3.2 PREVENTIVE MAINTENANCE INSPECTIONS
A message can be set through the diagnostics mode which automatically reminds the user when preventive maintenance inspections are due. Refer to "Setting Preventive Maintenance Interval" section in the “Troubleshooting” chapter.
3.2.1 Regular Inspection
Regular inspections consist of a visual inspection for damage and cleanliness, and performing the procedure described in the “Start-Up” section of the Directions for Use (DFU), before each usage of the instrument.
Failure to perform regular and preventive maintenance inspections may result in improper instrument operation.
WARNING
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 86
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.1 Regular Inspection (Continued)
Case
Examine the instrument for overall condition. The case should be clean and free from IV solution residue, especially near moving parts. Also check for dried solution deposits on accessible areas of air­in-line sensor, pressure transducer, and latch mechanism. Check that labels and markings are legible.
Mounting Bracket
Pole mounting bracket should be secure and functioning. If the instrument is mounted on a pole or stand, examine the condition of the mount. Also, examine the pole and stand.
Power Cord Assembly
Examine the power cord assembly for:
Signs of damage, cuts or deformities in the cord. If damaged, replace the entire cord.
Integrity of hospital grade power plug. Attempt to wiggle blades to ensure they are secure. If any damage is suspected, replace entire cord.
Appropriate tension and connection. Applicable only if IV pole has electrical receptacles for accessories.
Strain reliefs. Examine strain reliefs at both ends of line cord. Be sure they hold cord securely.
Keypad
Check membrane switches for damage, such as from fingernails and pens. During the course of the inspection, be sure to
check that each switch performs its proper function. Refer to "Testing Switches" section of the “Troubleshooting” chapter.
Mechanism
Clean any surfaces where solution or obstructions have accumulated. Verify:
Mechanism seal is not torn or worn.
Cam followers are not broken or cracked and are free of foreign matter.
Proper operation of latching mechanism. Cam followers should retract and extend smoothly.
Air-in-line arm moves smoothly from opened to closed position.
Fluid Control Actuator rotates 180°.
3.2.2 Functional Test
1. Turn instrument on without set installed. Verify it "beeps" and red alarm light flashes.
2. Set infusion rate to 460 mL/h and VTBI to 100 mL.
3. Press RUN/HOLD switch with latch closed, and rate and VTBI
0 to cause
"set out" and "air in line" messages.
4. Open latch.
5. Install primed administration set with latch open.
6. Verify instrument displays "air in line" and "latch open" messages.
7. Close latch and verify display returns to setup page.
8. Perform upstream occlusion test as follows:
a. Verify infusion rate is set to 460 mL/h.
PREVENTIVE MAINTENANCE
3-2
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 87
PREVENTIVE MAINTENANCE
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.2 Functional Test (Continued)
b. With instrument on hold, or at start-up,
verify primary VTBI is set to 100.
c. Press RUN/HOLD switch to begin
infusion.
d. Clamp off IV line just above
instrument to simulate an upstream occlusion. Verify instrument stops running, alarms, and displays OCCLUSION UPSTREAM within 60 seconds.
e. Press RUN/HOLD switch to silence
alarm and put instrument on hold. f. Remove or open clamp on line. g. Press RUN/HOLD switch to resume
infusion. Alarm should not reoccur.
9. Perform downstream occlusion test as follows:
a. Continue infusing from above step. b. Verify rate is set to 460 mL/h. Clamp
off set just below instrument.
c. Allow instrument to run until it alarms
OCCLUSION DOWNSTREAM within 60 seconds.
d. Press RUN/HOLD switch to silence
alarm and put instrument on hold. e. Release or open clamp. f. Press RUN/HOLD switch to resume
infusion. Alarm should not reoccur.
3.2.3 Flow Stop Test
1. Turn power off with administration set primed and loaded in instrument.
2. With all tubing clamps open and fluid container 2 or more feet above instrument, verify no fluid flows out of set.
3. Remove set. Verify no fluid flows out of set.
3.2.4 Rate Calibration Procedure
To maintain system accuracy, the rate calibration should be done first, followed by a verification rate test, when doing Preventive Maintenance or post repair testing.
1. Change Cal # to "0.0" to run rate calibration and calculate a new calibration number.
NOTES:
• Once the Rate Cal # is set to "0.0" and accepted, the instrument will need to be run for at least two seconds before the Rate Cal # can be changed to a non­zero value. If not, the instrument will display Do Rate Accuracy Test at 0%. The instrument will not allow one non­zero value to be changed to another non-zero value.
• Rate Calibration is run at the nominal value (0.0%) so that the percentage can be directly entered in the instrument without another calculation.
2. Run rate calibration (using an 80VCS set) at 400 mL/h, with a VTBI of 40 mL and VI reset to zero. Follow procedure in “Post Calibration Rate Accuracy Verification” section, steps 1-23, and then determine rate calibration number (in %), as shown below.
3-3
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 88
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.4 Rate Calibration Procedure (Continued)
Calculate volume accuracy as follows:
Volumetric Volume Accuracy Error Computation
Vcollected =
volume in burette in milliliters
Vexpected =
characterized volume printed on 80VCS set insert
Step 1:
A = Vcollected / Vexpected
Step 2:
B = Ax 100
Step 3:
% Error (Round % Error to nearest tenth of a percent.) = B - 100
NOTES:
The range of the percent error can be
from -5.3 to +15.6%, based on mechanism to mechanism differences and performing the initial run for calibration at 0.0%.
In addition to performing this process
during Preventive Maintenance, this process would also apply when replacing a mechanism or installing a new main board assembly.
3. Do not remove 80VCS set from instrument until one of following is determined:
Instrument has passed Post
Calibration Rate Accuracy Verification and calibration is not needed.
Rate calibration number was
changed and instrument now passes Post Calibration Rate Accuracy Verification.
Mechanism replacement is required.
4. Reverse sign (+/- ) of % Error value from Rate Calibration results in step 3.
Example: Result is 4% high (+4%). Reverse the
sign (to get -4%). Number to enter for Rate Cal # is -4%.
NOTES:
In the example above, the new Rate Cal # tells the instrument to count more volume per revolution of the mechanism, so that the output will be less due to fewer revolutions.
The limits for the rate calibration entry are -15.6 to +5.3%, to adjust for differences from mechanism to mechanism. If the Rate Cal # is outside this range, then the mechanism needs to be replaced.
5. Enter new Rate Cal # for applicable channel using keypad and bottom center soft key (+/- ) for sign. Press ENTER.
NOTES:
Make sure the "+/- " sign is used with
the percent change when doing rate calibration.
• To change the rate calibration number, refer to “Viewing/Changing Rate Calibration Information” section in “Troubleshooting” chapter.
PREVENTIVE MAINTENANCE
3-4
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 89
Water source
PREVENTIVE MAINTENANCE
3-5
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.4 Rate Calibration Procedure (Continued)
6. Press ok soft key to accept change and return to beginning of D4 page.
NOTE: When the main board is changed, the Rate Cal # defaults to 100% and 0.0 mL/rev. Reset the Rate Cal # to "0.0%" or previous Cal #, to clear an Instrument Malfunction message.
3.2.5 Post Calibration Rate Accuracy Verification
Perform the following steps without removing the 80VCS set or turning the instrument off.
Due to the Dynamic Monitoring®Feature, the rate is varied during operation. For this reason, ALARIS Medical Systems does not recommend using automatic testers to check rate accuracy. Generally, these devices collect small samples and may cause results to be incorrect, even though the instrument is accurate.
It is recommended, with the initial use of the 80VCS set on each instrument, to perform rate calibration first, to save time.
Do not
use the Model 80VCS Calibration Set for more than 15 rate calibration and post calibration rate accuracy verifications (count one use for both calibration and verification of the same instrument).
CAUTION
Figure 3-1. Rate Accuracy Test Setup
Table or
Bench
Equipment
Stand
Burette
Clamp
Burette
(50 mL)
Tubing
IV Tubing
Used Fluid Receptacle
Three­Way Stopcock
80VCS Rate Cal Set
Instrument Stand
30 +
2 inches
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 90
PREVENTIVE MAINTENANCE
3-6
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.5 Post Calibration Rate Accuracy
Verification (Continued)
Keep track of the number of times the set is used by recording each use on the 80VCS insert or on a separate record.
1. Fill solution container with clean tap water. Close AccuSlide®Flow Regulator clamp on 80VCS Calibration Set and then insert spike into solution container.
2. Open AccuSlide
®
Flow Regulator clamp and prime set. Pay particular attention to ensure all air is expelled from set. Close AccuSlide
®
Flow Regulator clamp.
3. Connect output of set to one side of three-way stopcock.
4. Load set into instrument.
5. Close latch.
6. Verify there is no fluid flow or drops falling in drip chamber.
7. Plug instrument into a properly grounded AC outlet.
8. Set stopcock to output into a class A or B burette.
9. Press POWER to turn channel on.
10. Set primary infusion rate to 400 mL/h.
11. Set VTBI to 20 mL.
12. Ensure instrument (both channels if dual channel) is set to Pressure mode.
NOTE: The factory default for the Monitoring Options mode is Pressure.
13. Press RUN/HOLD to start primary infusion. Infuse until tubing and burette are fully primed (approximately 1 minute).
14. Press RUN/HOLD to stop infusion.
15. Adjust height of instrument and/or fluid container to attain a head height of 30 ±1 inches / 76.2 ±2.5 centimeters between middle of pumping mechanism and fluid level in either the:
NOTE: A 30" head height was used in the initial qualification of this process and is the recommended head height. Based on observed field use, a 24" head height was also tested and verified for the Rate Accuracy Specification.
bag or vented bottle (vent closed on
administration set) or
drip chamber (unvented bottle with
vent open on administration set).
16. Adjust fluid level in burette until meniscus is level with zero mark on burette.
NOTE: The instrument may need to be run to prime the line to the zero level of the burette (step 13).
17. Verify primary infusion rate is 400 mL/h.
18. Reset VTBI to 40 mL and clear volume infused.
19. Press RUN/HOLD to start primary infusion.
20. Instrument will run approximately 360 seconds (6 minutes) to complete delivery and then go into KVO mode. Stop instrument within 1 second of its entering KVO mode.
21. Make a note of volume collected in burette.
22. Note expected volume, as identified on 80VCS calibration set insert.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 91
PREVENTIVE MAINTENANCE
3-7
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.5 Post Calibration Rate Accuracy
Verification (Continued)
23. Do not remove 80VCS set from instrument until one of following is determined:
Instrument has passed rate
verification and calibration is not needed.
Rate calibration number was changed
and instrument now passes verification.
Mechanism replacement is required.
24. Calculate volume accuracy as follows:
Volumetric Volume Accuracy Error Computation
Vcollected = volume in burette in mL Vexpected = characterized volume
printed on 80VCS set insert
Step 1: A = Vcollected ' Vexpected Step 2: B = Ax 100 Step 3: % Error (Round % Error to
nearest tenth of a percent.) = B - 100
25. Result should be 0.0±1%.
26. If volume accuracy does not fall within required range of ±1% from expected volume and test results were:
inside a range of -5.5 to +7.0% from
expected volume, perform "Rate Calibration Procedure". Set rate calibration number to 0.0% before running rate test, to determine a new calibration number.
• outside a range of -5.5 to +7.0% from expected volume, return instrument to ALARIS Medical Systems for repair or replace mechanism.
27. Set stopcock to drain fluid in burette to zero level, in preparation for next test.
NOTE: If additional low rate (5 to 20 mL/h) testing is desired, use an 80VCS set and collect at least 6 mL of fluid. The results should be ±5% of the expected output. At lower rates (less than 5 mL/h), evaporation may need to be prevented or accounted for in the results. Calibration must be performed at 400 mL/h and collecting 40 mL of fluid.
3.2.6 Pressure Calibration
1. Place instrument on bench or other flat surface and connect to AC power.
2. Connect pressure meter, pressure source, and reservoir to pressure calibration set. Refer to Figure 3-2 "Pressure Test Setup".
3. Install a pressure cal set (70ISS) into instrument.
4. Enter Diagnostics Mode by pressing and holding top soft key, then turn instrument on and release soft key when diagnostics display appears. Refer to "Entering Diagnostics Mode" section in “Troubleshooting” chapter.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 92
3.2 PREVENTIVE MAINTENANCE INSPECTIONS (Continued)
3.2.6 Pressure Calibration (Continued)
5. Advance to D6 page by pressing page soft
key 5 times.
On D6 page, press Cal A Pressure or Cal B Pressure (dual channel). D6A or D6B will be displayed, depending on which channel was selected.
6. Adjust pressure to "0 mmHg" from test fixture. Press and release 0 mmHg soft key. If readings are in a valid range, it will display Pass.
7. Apply 500 mmHg (±2 mmHg) from test fixture. Press and release 500 mmHg soft key. If readings are in a valid range, it will display Pass.
8. Remove 500 mmHg pressure applied to instrument, then remove set.
9. Press ok soft key to accept calibration and return to main D6 page.
10. Set sensor check/calibration verification:
a. Press Cal A Pressure or Cal B Pressure
(dual) soft key to re-enter the same Cal Pressure screen.
b. Verify both 0 mmHg and 500 mmHg
readings display Pass.
c. Install a standard set and close latch.
Verify reading is over 170.
PREVENTIVE MAINTENANCE
3-8
PSI
mmHg
0 775
Figure 3-2. Pressure Test Setup
Air Pressure Source
0 - 500 mmHg
Instrument
Stand
70ISS Pressure Calibration Set
Reservoir
Pressure
Gauge
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
AB
Cal A Pressure DAC A Settings
OPTIONS
D
6
Cal B Pressure DAC B Settings
page
Page 93
3.2 PREVENTIVE MAINTENANCE INSPECTIONS (Continued)
3.2.6 Pressure Calibration (Continued)
d. Remove set. Verify Sensor = reading is
in -80 to +30 mmHg range without set installed. If instrument will not soft cal, perform Hard Pressure Cal Procedure (“Corrective Maintenance” chapter).
3.2.7 Ground Current Leakage Test
Refer to the applicable Signature Edition
®
GOLD Infusion System DFU.
3.2.8 Ground Resistance Test
Refer to the applicable Signature Edition
®
GOLD Infusion System DFU.
3.2.9 Battery Refresh Cycle
METHOD ONE
1. Connect instrument to AC power.
2. Enter Diagnostics mode by holding top left soft key while powering instrument on.
3. Press page soft key (bottom right) to advance to D2 page.
4. Press lower left soft key to access Battery Status screen.
5. Press lower left soft key to highlight Rated Cap value.
6. Using numeric keypad, enter 0.0 (in Rated Cap value).
7. Press the ok soft key (bottom right).
8. Repeat step 4 to access 'Battery Status'.
9. Repeat step 5 to highlight Rated Cap value.
10. Using numeric keypad, enter 1.3 in Rated Cap.
11. Press ok soft key.
12. Press and hold POWER key for a second to power instrument down.
Battery Refresh procedure has been initiated and will continue to run until complete, whether instrument is on or off, generally 12 to 24 hours. Instrument needs to be connected to AC power during this period. Disconnecting instrument from AC power will stop refresh cycle.
METHOD TWO
1. Disconnect instrument from AC power.
2. Disconnect battery from instrument.
3. Press and hold POWER key for
5 seconds.
4. Reconnect battery and connect instrument to AC power.
Battery Refresh procedure has been initiated and will continue to run until complete, whether instrument is on or off, generally 12 to 24 hours. Instrument needs to be connected to AC power during this period. Disconnecting instrument from AC power will stop refresh cycle.
PREVENTIVE MAINTENANCE
3-9
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
AB
OPTIONS
Calibrate Pressure
0 mmHg=1100 Pass 500 mmHg=2400 Pass Sensor=+001 Pass
ok
D
6
A
Page 94
PREVENTIVE MAINTENANCE
3-10
3.2 PREVENTIVE MAINTENANCE
INSPECTIONS (Continued)
3.2.10 Reset Time
1. Enter Diagnostic Mode and advance to D2 page.
2. Reset hours and minutes as needed from time reference. Refer to “Setting Time (and Date)” in the “Troubleshooting” chapter.
NOTE: Clock will lose about 3 minutes per month since it is not a true real-time clock. Once reset, the previous loop will not be affected or adjusted.
3.2.11 Reset PM Due
Enter Diagnostic Mode and go to D2 page. Access PM Setup and reset PM Due by pressing lower-left soft key.
3.3 STORAGE AND CLEANING
Refer to the applicable Signature Edition
®
GOLD Infusion System DFU.
3.3.1 Storage
The instrument may be stored without connection to AC power. It will automatically disconnect the battery when the voltage gets too low. To reuse the instrument after storage, connect it to AC power for a minimum of three (3) hours before placing it back into service. When temporarily taking the instrument out of service, connect it to AC power to ensure a fully charged battery when needed.
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 95
PREVENTIVE MAINTENANCE
3-11
Table 3-1. PM Inspections
I.D. Number Instrument Serial Number
Ref.
Section
Frequency Date Completed Date Completed
Regular Inspection
(record every 12 months)
3.2.1 Every Use
Functional Test
3.2.2 12 Months
Flow Stop Test
3.2.3 12 Months
Rate Calibration Procedure
3.2.4 12 Months
Post Calibration Rate Accuracy Verification
3.2.5 12 Months
Pressure Calibration
3.2.6 12 Months
Ground Current Leakage Test
3.2.7 12 Months
Ground Resistance Test
3.2.8 12 Months
Battery Refresh Cycle
3.2.9 12 Months
Reset Time
3.2.10 12 Months
I.D. Number Instrument Serial Number
Ref.
Section
Frequency Date Completed Date Completed
Regular Inspection
(record every 12 months)
3.2.1 Every Use
Functional Test
3.2.2 12 Months
Flow Stop Test
3.2.3 12 Months
Rate Calibration Procedure
3.2.4 12 Months
Post Calibration Rate Accuracy Verification
3.2.5 12 Months
Pressure Calibration
3.2.6 12 Months
Ground Current Leakage Test
3.2.7 12 Months
Ground Resistance Test
3.2.8 12 Months
Battery Refresh Cycle
3.2.9 12 Months
Reset Time
3.2.10 12 Months
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 96
PREVENTIVE MAINTENANCE
3-12
THIS PAGE
INTENTIONALLY
LEFT BLANK
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 97
4
PRINCIPLES OF OPERATION
Page 98
4.1 INTRODUCTION
This chapter describes the mechanical and electrical systems that comprise the Signature Edition
®
GOLD Infusion System.
4.2 GENERAL INFORMATION
The main PCB for both the single and dual channel instrument is a double-sided multi­layered Surface Mount Technology (SMT) board. If a board is determined to have failed, it can be replaced or the instrument can be returned to ALARIS Medical Systems for repair. ALARIS Medical Systems does not provide replacement components for repair of SMT boards nor does ALARIS Medical Systems recommend attempting field service of the instrument's SMT circuit boards.
Full board assembly schematics are not included with this service manual.
The AC off-line switcher and RS-232 boards are replaced as an assembly. If a board is determined to have failed, it is replaced with a new board (for part number refer to “Illustrated Parts Breakdown” chapter). ALARIS Medical Systems does not provide replacement components for repair of these boards.
Both single and dual channel instruments function in the same manner; however, they use two different main PCBs. The component reference designations are therefore different for each board. To help distinguish between the single and dual channel instrument reference designations in this chapter, the dual channel instrument will be represented in parenthesis; for example, (U13).
Chapter 4 — PRINCIPLES OF OPERATION
4-1
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Page 99
4.2 GENERAL INFORMATION (Continued)
The instrument contains a peristaltic pumping mechanism and support circuitry to ensure controlled flow. The peristaltic mechanism consists of a linear array of 12 cam followers which travel perpendicular to the administration set. These cam followers act like "fingers" kneading the membrane. When the fluid-filled disposable is placed against the array of cam followers, the coordinated, sinusoidal motion of the cams causes a peristaltic wave of fluid displacement in the instrumenting segment of the disposable.
The instrument will alarm at signs of internal problems and at preset thresholds for external problems (for example, when battery charge falls below a critical level, or instrument output pressure exceeds a programmed limit). All alarms provide visual and auditory signals to alert the operator.
Accuracy of fluid delivery is a function of the microprocessor-controlled rotation cycle of
the camshaft, and the administration set section compressed by the cam followers.
4.3 OVERVIEW
The instrument contains one main PCB and several modules that interface to it. The interfacing modules are as follows:
LED module
Graphic LCD module (MAIN)
Lower LCD module
Battery
AC off-line switcher
Keypad
Nurse Call / RS-232 board
• Motor
Air-in-line sensors
Pressure module
Motor rotation sensor and mechanism latch detector (optocouplers)
• ECD board
PRINCIPLES OF OPERATION
4-2
Figure 4-1. Main Block Diagram
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
AC
OFFLINE
SWITCHER
BATTERY
CHARGER
DC/DC
POWER
SUPPLY
LOWER
LCD
WATCHDOG
MOTOR/
CONTROL
MECH
TAMP
SW
KEY PAD
LED
MAIN
LCD
SPKR
TEST
MAIN
SPKR
SPKR TEST
BKUP SPKR
AUDIO
AMP
VBKUP
1.0F
5.5V
SUPER CAP
PRESS
SENSOR
ADC
VCC ADREF
RS232
AIL
COMBO
IC
32 Mhz
Y1
CPU / EPROM
EEPROM / RAM
CPU KERNEL
BATTERY
MANAGER
L
L
+
ECD
Page 100
PRINCIPLES OF OPERATION
4.3 OVERVIEW (Continued)
The instrument power is supplied through the AC off-line switcher module and the battery.
The main PCB contains all the control circuitry required for the instrument. The board can be broken down into four main sections, as follows:
CPU Kernel
The microprocessor, RAM, ROM, data communication, and COMBO IC make up the heart of the system. These are collectively referred to as the CPU kernel. The CPU kernel is responsible for controlling the motor actuation, sensing and responding to user input, monitoring various system sensors, and performing start-up and ongoing system operational testing.
Power System
The power system is responsible for charging the battery, generating the DC power, displaying battery status and performing watchdog (clock sync checks) functions. The power system includes the Battery Manager custom IC.
Motor Drive and Sensor Control
The motor drive and sensor control circuitry drives the motors, the air-in-line sensors, the mechanism latch sensors, and the rotation sensors. The circuit is also responsible for monitoring the pressure sensors, the power supply voltages, the motor current, and the air­in-line sensor outputs.
User Interface Circuitry
The user interface circuitry connects to the keypad, LED modules, and LCD modules to the kernel circuitry for monitoring and control. This circuitry also contains the audio interface, and audio test.
4.4 MAIN PCB
4.4.1 CPU Kernel
The CPU kernel is responsible for controlling the motor actuation, sensing and responding to user input, monitoring various system sensors, and performing start-up and on-going system operational testing. The kernel is based on a 16-bit 80C188 microprocessor U11 (U15), 1Mbytes of Flash EEPROM program storage, and 128K bytes of battery backed up RAM data storage. In addition, the kernel has 2K bits of EEPROM memory and a selectable baud rate for serial communications interface.
The COMBO IC U10 (U14) is a custom ASIC (Application Specific Integrated Circuit) which incorporates timing, address decoding, digital I/O, and other system "glue" functions. The Combo IC has a 16­bit CRC generator which is used to periodically test the Flash EEPROM data. The COMBO IC also contains the local serial interface control logic used to interface to serially-accessed peripherals, such as the A/D, EEPROM, LED Module(s), and Battery Manager. Additional information can be found in the section below "COMBO IC".
4-3
Signature Edition®GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
Loading...