Air-Shields C100, C200 MAINTENANCE

PREVENTIVE
100/200
MAINTENANCE
1.
RELEASE
slightly down,
back
THE
in
INNER
the
direction
of
by
pushing
hood
slopes,
then
2.
Release
ing rear then
front
of
inner
lower.
edge
OF
of
inner
wall,
INNER
pull
wall
on
toward
your
you
while
arm;
to
release,
rest-
lift
*
Double
3.
LOWER
AS
IT
WILL
GO
only.
OF
INNER
FIGURE
4.3
AS
FAR
REMOVAL
OF
4-3
4.
REMOVE
rection
INNER
INNER
shown.
WALL*
by
lowering
in
di-
C100/200
PREVENTIVE
MAINTENANCE
6.
REMOVE Main
FIGURE
DECK.
and
4.4
Rotate
ACCESS
the
Gasket.
PANEL
INNER
Retainer
WALL
(Figure
REMOVAL*
parallel
the
slot,
out
the
*
Double
FIGURE
4.5
REMOVAL
OF
MAIN
DECK
7.
REMOVEAIRINTAKE the
Incubator
until
the
TUBE.
end
of
Grasp
the
the
Air
clears
Intake
the
gasket.
(Figure
Remove
4.6),
twist
the
PREVENTIVE
and
MAINTENANCE
it
toward
the
C100/200
the
front
assembly.
of
8.
9.
10.
11.
12.
4.2.2
1.
2.*
*
Model
REMOVE
the
outside;
REMOVE REMOVE
REMOVE
discard
REMOVE
CLEANING
mental
described
Cleanser
After
the
CLEANING
Protection
removing
PROBE.
paper
FIGURE
DISPOSABLE
discard
ACCESS TUBING
DISPOSABLE
sleeves.
THE
AGENTS.
in
may
towel. only.
the
ACCESS
AIR
INTAKE
Agency
paragraph
be
used.
all
solid
Use a disinfectant—detergent
4.6
ACCESS
cuffs.
GASKET
PORTS
ENTRY
MICROFILTER
An
iodophor
should
4.2.1. A cleanser
using
wastes
and
REMOVAL
or
be
contaminants
COVER
quaternary used,
but
any
cleaning
to
OF
AIR
Access
of
SLEEVES
by
disinfectant
after
as
Air-Shields
agent
the
thoroughly
45
INTAKE
Access
by
the
by
by
pulling
loosening
the
follow
disassembled
clean
detergent Incubator
Kleenaseptic®
the
all
surfaces,
TUBE
Gasket
pulling
pulling
Sleeve
the
two
registered
is
empty
manufacturer's
parts,
by
it
off
free.
off
thumbscrews.
by
and
Germicidal
clean
dry
pulling
the
the disassembled
directions
with a clean
the
retainer
as
it
off
outside.
Environ-
Surface
for follows:
rings;
as
cloth
100/200
PREVENTIVE
3.
HUMIDITY ACCESS gent, Place solution.
MAINTENANCE
CHAMBER
PORTS,
remove
the
Air
AND
the
Intake
AND
W-shaped
Tube,
Access
PIPE,
AIR
GASKET.
Baffle
Gaskets, Tubing
INTAKE
the
chamber
TUBE,
the
humidity
Access
and
ACCESS
chamber
dry
it
with a clean
Ports,
and
GASKETS,
with a disinfectant-deter-
cloth
TUBING
or
paper
Gasket
towel.
the
*
NOTE: Humidity
Allow completely all
if
necessary, a larger
Chamber
indentations,
is
necessary
the
the
and
to
as
recommended
with a clean
dry the
to
remove
thumbscrew
Assembly
the
sleeve
container
be
cleaned
cloth
the
may
be
separately.
by
the
cleaning
or
paper
towel.
chamber
secures
becomes a loose
for
by
rotating
and
cleaning,
the
used,
but
if
the
chamber
solution
Drain
the
Humidity
(inside
rotate
Bracket,
counterclockwise
manufacturer,
and
the
and
the
is
Chamber,
with a clean
Assembly
rotate
as
shown
not
used,
scrub
about
the
bracket
in
Figure
Assembly
the
remove
it
thoroughly
cloth
or
1/4
1/4
Clean
is
unscrewed.
and
including
paper
to
the
to
the
and
dry
towel.
left. left.
-
4.
CONTROLLER. the a
front
cloth
FIGURE
panel
and
dampened
4.7
REMOVAL
The
portions
the
with a disinfectant-detergent.
of
the
bottom, and
OF
HUMIDITY
Controller
two
external
sides
46
of
to
the
chassis.
FILL
the
controlled
These
PIPE
portions
ASSEMBLY
incubator
environment
may
be
wiped
include
clean
PREVENTIVE
-
C100/200
MAINTENANCE
CAUTION: mita trical
CAUTION: temperature
A.
B.
NOTE: A disinfecting ofthe gent,
5.
6.
chemical
build-up
components.
The face: the
Remove air
Clean
Controller.
MATTRESS
surfaces
MATTRESS
WASTES tergent
nism. A cleanser any
of
portions
included
Controller
allowing
cleaning
of
Failure
control
any
sensor.
these
surfaces
The
thoroughly,
and
registered
are the to
to
it
TRAY,
TILT
agent
cleaning
or
which
Do
not
spray
the
Controller
air
temperature
which
these
clean
could
and
cause
build-up,
pay
with a disinfectant-detergent,
is
available
as
recommended
AIR
CURTAIN
dry
CONTROL,
contaminants
the
as
Kleenaseptic®
follow
the
agents
may
cleaning
are
components
result
particular
as
surface
with a clean
may
be
conductive.
solutions
within
probe,
in
sufficient
oxygen
attention
an
accessory
is
COVER,
immersed
C100/200
the
Mattress
Environmental
Germicidal
manufacturer's
be
conductive
Do
not
any
the
Incubator's
the fan
are
concentrations.
by
cloth
QT
impeller,
mounted.
build-up
to
the
fan
Air-Shields
the
manufacturer
MAIN
DECK.
or a paper
(SERIES
Control.
Protection
Surface
directions
impeller,
dry
the
—02)
Agency
for
and/or
permit
of
controlled
the
with a clean
Use a disinfectant-detergent
towel.
An
Cleanser
leave a residue
cleaning
these
surfaces.
environment
heater,
to
after
of
the
reduce
heater,
cloth
to
facilitate filling
the
cleaning
gaskets,
air
air
it
INCUBATORS.
iodophor
or
should
may
quaternary
be
which
agents
cleaning
with a disinfectant-deter-
to
are
and
flow,
which
temperature
or
paper
solution.
REMOVE
disinfectant-de-
to
clean
be
used.
may
contact
on
the
the
surface
probe,
towel.
the
to
ALL
the
elec-
rear sur-
of
affect
and
surface
clean
all
SOLID
mecha-
using
WARNING:
in
an
7.
-
CAUTION:
Alcohol
Do
not
Hood.
8.
**
Model
DO
NOT
lubricate
enriched
expose
these
AIR than 3 months,
cover
WARNING: build-up. tions.
oxygen
AND
including etc.,
can
cause
the
sources
INTAKE
with a disinfectant-detergent.
Be
only.
environment.
CABINET
crazing
can
MICROFILTER.
it
A
dirty
(Optional
STAND.
the
inner
dry
with a clean
of
assembly
cause
should
Filter
the
filter
be
elevator
Use a disinfectant-detergent
and
cloth
the
clear
to
cracking
Do
not
replaced.
may
affect
is
checked
in
C200).
mechanism
access
or
paper
acrylic
direct
of
attempt
Before
oxygen
.
shield*.
towel.
Hood.
radiation
gaskets,
on a routine
Do
fading
to
clean
installing a new
concentration
oil
Clean
not
or
reverse
basis
or
other
to
clean
the
exposed
use
alcohol
germicidal
of
paint,
and
the
filter,
clean
and/or
commensurate
potentially
all
to
clean
surfaces
for
lamps.
crazing
microfilter.
cause carbon
flammable
surfaces
all
of
cleaning.
Ultraviolet
of
If
visibly
the
Microfilter
of
the
holes,
indentations,
the
cabinet
the
clear
dirty, chamber
dioxide
local
material
thor-
stand.
radiation
acrylic
or
older
and
condi-
4-7
100/200 PREVENTIVE
MAINTENANCE
4.2.3
IMPORTANT:
not
proper
After
REASSEMBLY
interchangeable
identification
cleaning INSTALL
1.
ure
INSERT
2. INSTALL
3.
Rotate
Specific
between
of
all
parts
THE
AIR
HUMIDITY
THE
the
AFTER
Assemblies
components
and
CLEANING
the
C100/200
assemblies
INTAKE
CHAMBER
deck
one
conditioning
WRONG
placed
side,
as
shown
CORRECT
deck
AND
Retainer
(inner
Wails,
Series
prior
to
assembly.
as
described
(into
the
BAFFLE
POSITION-Main
up on
permitting
by
POSITION-Main
placed
chamber
{Figure
gasket
air
dashed
down
Main
-01
the
on
line.
on
Deck,
Mattress
and
the
in
paragraph
Assembly)
Humidity
GASKET
to
secure
Tilt
C100/200
4.2.2,
by
reversing
Chamber.
the
the
deck.
Mechanisms,
QT
Series
reassemble
the
Assembly
Oval
—02
Incubators.
as
described
procedures
as
shown
Gaskets)
Verify
below.
shown
in
in
Figure
are
FIGURE
INSTALL
4. to
NOTE:
CAUTION: curely.
5.
There
INSTALL
WARNING:
for
slightly
the
INSTALL
tress,
INSTALL in
THE
Figure
4.2.)
is a right-hand
Be
THE
the
lip
the
cover
proper
mattress
DISPOSABLE
then
DISPOSABLE
Figure
4.8
INSTALLATION
MATTRESS
the
two
AIR
CURTAIN
The
Air
Curtain
on
the
bottom
toward
operation
the
rails,
place
of
Access
the
mattress
TILT
MECHANISMS
and
left-hand
thumbscrews
COVER
Cover
of
the
Air
the
front
of
the
the
Air
Curtain
Panel
is
opened.
lowering
MATTRESS
straight
onto
ENTRY
OF
MAIN
mechanism.
AND
be
Curtain
Incubator
Cover.
Install
down.
COVER.
the
tray.
DECK
C100/200
the
MATTRESS
properly
the rod
to
the
The
Air
the
Mattress
Place a new
SLEEVES.
AND
GT
mechanism
TRAY.
installed
at
the
position.
Curtain
disposable
Install a new
HOOD
SERİES
for
correct
of
Close
Cover
by
positioning
SEAT
—02
on
to
the
temperature
the
the
is
operating
Mattress
Entry
the
Assembly
it a few
GASKET
Main
Deck.
are
tightened
Assembly,
correctly
inches
Cover
Sleeve
(Refer
se-
control.
and
check
if it
rises
above
the
as
shown
NOTE:
If
the
Incubator
is
to
be
gas
sterilized,
wait
until
48
after
sterilization
to
install
new
sleeves.
PREVENTIVE
C100/200
MAINTENANCE
1.
Install a
new
housing
the
smaller
sleeve
over
diameter
the
inner
3.
Rotate sieeve
elastic
ring
of
outer
ring
will
open
band
the
again
of
port
to
close,
if
properly
if
rotation
2.
Fold outer
installed,
is
reversed.
back
ring
and
of
the
slip
the
larger
pot
housing.
elastic
band
over
the
FIGURE
4.9
INSTALLATION
OF
IRIS
49
ENTRY
PORT
SLEEVE
C100/200 PREVENTIVE
8.
INSTALL A TUBING
or
torn.
9.
INSTALL
MAINTENANCE
AN
ACCESS
ACCESS
DOOR
GASKET
behind
the
front
each
Access
of
Door,
of
as
the
Hood.
shown
Replace
in
Figure
if
distorted
4.10.
ACCESS
CUFF
GROOVE
ACCESS
GASKET
DOOR
DOOR
——
„|
a
ACRYLIC
É
INSIDE
INCUBATOR
HOOD
OF
FIGURE
10.
INSTALL A NEW ameter
atits depressed.
NOTE:
11.
INSTALL
12.
INSTALL
13.
INSTALL A NEW and
elastic
center.
If
the
tighten
Incubator
THE
the
cover.
IMPORTANT: A complete unit
to
service.
GAS
STERILIZATION
Prior
to
gas
sterilization,
section.
removed stalled
All
used
disposable
and
discarded
after
sterilization.
The
ACCESS
Access
is
INNER
AIR
two
4.10
INSTALLATION
the
groove
Door
should
to
be
gas
sterilized,
SHIELD
INTAKE
thumbscrews.
operational
the
entire
elements
as
described
incubator
in
the
gasket,
latch
with
by
reversing
the
Access
MICROFILTER
ifa
new
filter
checkout
such
in
the
(paragraph
should
as
iris
sleeves,
cleaning
OF
ACCESS
Access
installed
slight
pressure,
until
after
sterilization
the
procedure
Panel.
if
necessary.
is
instatled,
be
thoroughly
access
instructions.
4-10
correctly,
and
shown
Replace
indicate
:
should
cleaned
door
cuffs,
New
the
DOOR
Gasket
should
to
in
be
disposable
GASKET
by
the
open
install
Figure
the
Air
on
performed
as
described
mattress
stretching
has a small
when
the
new
sleeves.
Intake
Microfilter
the
place
provided
before
elsewhere
covers,
elements
the
larger
opening
latch
button
returning
etc.,
shouldbe
should
di-
is
Cover on
the
the
in
be
in-
PREVENTIVE
C100/200
MAINTENANCE
Release Access
NOTE:
the
Doors
Gas
CAUTION:
Standard Sterilizers
should
period
Gas
and
completion
be
properly
of
aeration
disposable
IMPORTANT:
turning
44
It repairs
the
CALIBRATION
is
recommended
Controller
should
sterilization
Sterilization
latches
be
temperature
sterilization
Wilmot
Castle
of
gas
sterilization,
secured
at a temperature
cover
should
A
complete
to
service.
SCHEDULE
the
made.
and
slide
open.
The
not
eliminate
procedures
are
satisfactory
an
in
place
and
of
be
placed
operational
Controller
Calibration
and
the
Air
Intake
the
should
as
programmed
aeration
the
32
to
35
on
the
checkout
of
the
Incubator
out
about
Microfilter
for
not
exceed
by
as
these
do
period
of
Incubator
°C.
After
aeration,
Incubator.
procedure
be
procedures
1/4”.
The
may
routine
replacement
130 ° F
automatic
not
normally
16
to
24
should
(paragraph
tested
are
provided
Access
be
(54.5
equipment
exceed
hours
should
be
operated
if
the
2.4)
and
calibrated
in
Panel
in
place.
of
the
be
allowed.
in a dry
is
not
to
should
at
Section 5 of
may
be
Air
Intake
as
°F
(64.5
condition
be
be
performed
least
annually
closed,
but
Microfilter.
by
American
The
Control
for
the
entire
immediately,
before
and
after
manual.
the
a
re-
4-11
100/200
PREVENTIVE
MAINTENANCE
{This
is
intentionally
blank.)
4-12
C100/200
SERVICE
5.1
GENERAL
section
and
IMPORTANT:
ably
in
this
5.2
CALIBRATION
5.2.1 This
paragraph
Unless
1.
The
2.
The
3.
Primary
GENERAL
otherwise
provides
Incubators.
It
should
manual.
WARNING: frame
provides
indicated,
Gontroller
Controller
is
is
power
calibration,
be
noted
PROCEDURES
The Motor
may
be
at
calibration
all
calibration
removed
connected
applied
to
the
SECTION
SERVICE
troubleshooting,
the
terms
Assembly
mains
potential.
procedures
procedures
the
Incubator
to a primary
Controller
and
is
for
source
for a minimum
5
removal
TEMP”
not
grounded.
the
Controllers and
are
performed
and
the
cover
of
the
correct
of
and
replacement
and
Under
under
is
removed.
voltage and
ten
minutes
POINT"
fault
tests
the
before
instructions
are
conditions,
for
the
Incubator.
following
frequency.
calibrating.
for
the
interchange-
the
metal
conditions:
Model
5.2.2 The
tests. .
ο
. .
.
EQUIPMENT
equipment
Equivalent
Probe
Simulator,
Logic
Probe capable
Variable
Digital Oxygen Flowmeter,
Leakage
Transformer,
Analyzer,
Current
REQUIRED
listed
below
equipment
No.
of
General
Fluke
Model
Ametek,
Victor Model
Tester,
Bio—Tek
is
required
may
68
900
+/—
12
Vdc
Radio
8060A
Thermox
1099-0025
for
calibration
be
substituted.
80
Model
Instrument
501
of
WSM
Division,
(Change
the
Controller
T3AW
16)
Model
and
performing
S-3A1,
Oxygen
Pittsburgh,
PA.
Concentra-
5-1
100/200 SERVICE
Connect
1.
Controller. +/—
2.
Onthe
to
3.
On
location
PCB2—
SETUP
2V~
AIR.
the
PROCEDURE
1.
Refer
2.
Adjust
3.
Onthe SKIN.
4.
Adjust
5.
Monitor
6.
Models);
6a.
On
the
MODEL
the
Simulator
Connect
(110/120V
Controller,
Simulator,
of
points
to
Figure
R35
Simulator,
Connect
R50
off
the TP15
the
simulator,
5.1
on
the
for a reading
POWER
on.
80
+/-
to
the
the
ac
Models);
set the
2V- (100V
set
AIR
set the
set
Set
and
adjustments.
and
connect
for a reading
the
DVM
between of
switch
the
logic
the
variac
the
PATIENT
through a variac
240
+/-
and
switch
the
of
switch
+/-
on
the
Controller.
probe.
as
follows:
Models).
switch
PROBE
2V~
set
DVM
50
On
to
to
(+) mV
and
(220/240V
Temp.
between
PCB2,
AIR
+/-
and
on
95
switches
and
50
on
TP22.
the
Connect
+/-
AUXILIARY
and
Models);
to
the
(+)
mV
on
the
the
Controller,
the
2V~
(110/120V
fully
PROBE
jacks
on
the
unit.
Set
100
+/-
2V-
°C.
Setthe
switch
and
logic
to
36
on
set
the
CONTROL
probe
between
Models);
counterclockwise.
on
the
the
variac
(100V
CONTROL
°C.
Refer
PCB2.
+12
190
as
follows:
Models).
to
Figure
Vdc
+/-
2V~
Panel
of
switch
5.1 for
Switch
and
ground.
(220/240V
the
120
to
7.
On the
7a.
Set Vac (100V
Set
(100V
approximately Turn
the
Controller,
logic
probe
the
variac
Models).
the
variac
Models).
off
the
set
as
follows:
as
follows:
The
two
Power.
the
AIR
stays
off
130
The
light
120
fight
should
seconds
Remove
SET
constantly
Vac
on
the
logic
+/~2V~
(Heater
the
(110/41
go
off
Off).
logic
probe.
Switch
(full
heater
(110/120V
probe
should
20V
Models);
for
approximately
to
37
°C.
power).
Models);
be on
Slowly
240
260+/- 2 Vac
for
approximately
+/-
2V~
seconds
clockwise
(220/240V
50
(220/240V
(Heater
until
Models);
percent
Models);
On)
and
the
108+/—2
of
the
100
light
on
time.
+/-2V-
on
for
5-2
=
>
o
=
Eni ba
a]
P
ae
B
Oto
EC
PS
Bdl
P
EL
0
S
Lo
=>
г
20040
STI
a DPédl
E
v
ARE
eg
a
=)
SEV
Q
mo
o Sat
8
O
gidin
+
эп
sn
<
O
3
5
5
5
CC
mee
om
tetta
п)
suy
a
an
06012
ee!
a
120,
5
¢
be
buy
069
MBE
seu
o
iz
21
mm
ČI
úd
poi
Ve. ©
GF
E
cu
Pau
7
58
O DO
o
O 8001
3
$ Gigi
B
E
EİN,
S
e
è
zou
o
E
o
0 *
=
"L
nana)
| 6
čl
o
mi
2
EE.
Ler
°С
$
em
gın
Zin
Doro
-C
EA
O
ιο
ОНИ
sora
0
9bdl
L
o
00
58
SE
ANY SLNIOd
LS31
osu
JO
|
|
NOLLY901
SIN
WLSPQV
‘(0019
SALYA
o
LABS
+
S oy
10018
μμ
©
го ©
μη
ozdi
are
orali
8
Ze
aziona
à
s
ita
or
M.R.
O
ade
1300W)
6800
WS
+
a
——
“44
15€
89
зая
ON ASSY
da
|
HnOH
10
o
|
seu
C100/200
SERVICE
5.2.4
PCB1 - MODEL
SETUP
Remove
1.
Panel. POWER
2.
Gonnect the
Simulator
points
PROCEDURE
On
PCBI,
2.
Set
the
3.
Repeat
4.
Set
the
5.
On
6.
Set
the
Repeat
7.
ihe
CONTROL
Set
and
switches.
the
Simulator
to
36°C.
and
adjustments.
adjust
R7
until
Simulator
steps 1 and 2 to
Simulator
adjust
Simulator
to
to
R18
to
20
AIR
20
steps 5 and 6 to
to
the
Set
the
the
°C.
ensure
until
°C.
ensure
knob.
vertically
PATIENT
Controller
display
Adjust R2
proper
and
the
right
On
proper
Remove
behind
PROBE
and
(SKIN)
on
settings.
°C.
display
adjust
settings.
the
the
brackets
and
AUXILIARY
Simulator
reads
until
the
Readjust
(AIR)
reads
R13
until
Readjust
screws
to
if
36
+/-
the
right
if
which
located
behind
PROBE
mode.
Refer
+/-
0.1
°C.
display
necessary.
0.1
(SKIN)
°C.
display
necessary.
the
jacks
to
reads
(AIR)
and
to
SILENCE/RESET
on
the
Figure
for
+/-
reads
20+/-
the
Front
Panel.
location
0.1
°C.
0,1
°C.
and
Set
of
5.2.5 TEST
1.
PCB2-
SETUP
Connect ac
els);
240
On
the
Controller,
On
the
Simulator,
location
MODEL
the
Simulator
through a variac
+/-2V~
(220/240V
set
set
of
points
to
the
AUXILIARY
and
Models);
the
SET
the
and
adjustments.
TEMP
switch
on
the
100
°C
to
PROBE
unit.
+/-
switch
AIR and
on
the
Set
the
variac
2V~ (100V
to
36.0
°C.
the
Panel
as
follows:
Models).
switch
to
of
120
36
the
+/-2V~
°C.
Controller.
Connect
(110/120V
Refer
to
Figure
5.3
the
for
5-4
ㄷㄷ
teu
100/200
SERVICE
ssa
Luv
sn
|
o
>
pa
ES
X)
PE üm
o,
TE 45H
Do
5
ο
asd
686s
95
+
oo
913
S
PD
QNV
SINIOd
1S
I O
NOILVOO1
__ SLNAWLSNrav
(0019
(x)
eus
P
sug
oa
вв >=
©
few
8150
215а
9150
t+
zs
OHIO]
|
I
1
Et
+
O
{ORTI
pa
]
1
et
ou.
X
var
+
cul
sw
„=
|
I
©
i
TICON)
¿us
eua
me
ou
oo
=
-
In
3
>
ow
==
Oo
~
Es
O
v
159)
6067
Eh
2
во
75
3)
lgod
z9 UNO
24
wl
©
Sisa
visa
esa
eu
sa
ot
E
O
[ENO
cu
7
a
ui
L
©
j
(X)
~~"
©
:109689
en
5013IHS-U1Y
on assv
vi >
ev
vei
£
E нЕ
se
ni
ーー
za
CT
‘sa
.
55а
>
050
S)
(x
τες
[7]
a
Mo
DZ
0000
5-5
100/200
SERVICE
PROCEDURE
1.
Refer
2.
Adjust
3.
Monitor
4.
Turnthe
Models);
5.
Onthe the
5a.
Setthe Vac
6.
Setthevariacas Vac approximately
7.
5.2.6 TEST
1.
2.
PCB1—
SETUP
Remove behind
Connect the
PROCEDURE
to
Figure
R35
off
the
Controller,
logic
variac
(100V Models).
(100V
off
the
Simulator
on
POWER
on,
80
+/— 2 Vac
probe
as
Models).
the
power.
MODEL
the
brackets
the
Simulator
to
5.3
Set
set
follows:
follows:
two
screws
AIR
and
connect
for a
reading
switch
the logic
the
variac
(100V
the
AIR
stays
off
130
The
light
120
The
light
seconds
Remove
which
located
to
the
mode.
the
DVM
of
on
the
Controller.
probe.
as
follows:
Models).
SET
constantly
Vae
on
the
logic
Vac
should
(Heater
the
behind
AUXILIARY Refer
logic
to
go
Off).
the
Figure
.
between
+/—
Connect
95
+/— 2 Vac
On
Switch
(full
(110/120V
(110/120V
off for
probe.
SILENCE/RESET
PROBE
to
heater
Models);
probe
should
Models);
approximately
and
5.4
for
(+)
on
the
the
logic
(110/120V
37°C.
Slowly
power).
260
be on
240
to
the
Front
and
on
the
location
and
on
probe
between
Models);
fully
counterclockwise.
+/- 2 Vac approximately
+/- 2 Vac
seconds
Panel.
POWER
Panel.
of
(220/240V
(220/240V
Set
switches.
Set
points
(Heater
and
PCB2.
+12
Vde
190
+/— 2 Vac
clockwise
Models);
50
percent
Models);
On)
and
the
Simulator
adjustments.
and
uniil
the
of
and
to
ground.
(220/240V
light
on
108
+/—
2
the
time.
100
+/-
2
on
for
vertically
36
°C.
Set
1.
2.
3.
4
On
adjust
Set
the
Simulator
Repeat
Reassemble
steps 1 and 2 to
R18
to
the
Controller
until
20
*C.
the
display
On
ensure
and
reads
adjust
proper
disconnect
settings.
36
+/-0.1°C.
R13
until
Readjust
the
Simulator.
the
display
if
reads
necessary.
20
+/-
0.1
°C.
on
en
mo
TT
Teo
Sy
Gum
BoA
WSdWOP
TLV
INSNE
Tadı
Zu в
bpm Bovnešy
OZd
al
NEZ
OT
T
Looru
ou
ALI
6001
202
Vera
i
o ©
=
129469
sido
E
+
|
=
e
=
Sad
©
di
a
on
te
Fo
o
su sn
en
ein
58
o
zın
sd
sn
io
addi
cs
FER
sidi
よら
E
Es 3
E
E
JA:
u,
к
ne
ψ
sino
ее
m
ar
5
이은
er
i
sou
an
sin
pin
or
calo
-
o
ες)
ir
so
SP
893
SO
+
o
ozn
SED
εις
EO
£
sin sin
TET:
210
(Doro
=
ОНИ
В
и
sou
96801
α
+
σσ
EN
I
aN9 1190
ze
,
\
C100/200
SERVICE
QNV
SINIOd
1S
1
O NOLVOOT
SLNAWLSNraVv
(0029
TAdOW)
Sidi
240d
£'S
AHNOI
5-7
pe
34
156 89
S0331HS-
ON
ASSV
819
AZ IN
seu
C100/200 SERVICE
esa
giù
ELY
1
eeu
=
©
CINV
©
pedali
3
6H
8
+
ad
ll
st
SINIOd
a
SIGG
isa
AA!
SST
>
ul
=
mu
O
US)
P
στὴ
oy
so
4
=
20800000
a
TE
06
6
|
==
SS
65
©
LS3L 40
NOILV907
ANTIALSAPAY
(0029
TAGOW)
a
=
О
DI
полон
=
S0731HS-HIv
ss
sey
©
西西
а
290
ir
o
Eo)
LAL)
SET
EST
T
|
LEOd #°9
anel
=
|
5-8
C100/200 SERVICE
5.2.7 TEST
1.
2.
LEAKAGE
SETUP
Connect is
ungrounded,
The
leakage
tance
of
be
PROCEDURE
1.
Use
the ground micro-amps
2.
Reverse
3.
Perform
5.2.8 TEST
2.
OXYGEN
SETUP
1.
Place a calibrated Apply
PROCEDURE
1.
the
2.
After
40
3.
Increase centration
CURRENT
the
Controller
current
ohms.
adjusted
Leakage
as in
the
steps 1 and 2 with
to
the
If
the
to
provide
Current
the
ground
120
Vac
and repeat
CONCENTRATION
oxygen
oxygen
at a flow
ON.
minutes
the
Oxygen
level
is
of
operation,
between
TESTS
the
primary
POWER standards
Leakage
it.
Tester
connection
Units
or
the
Controller
analyzer
of 8 LPM
verify
to
12
65%
power
switch
provided Tester
to
measure
500
micro-amps
1.
TESTS
on
to
After
and
source
on.
in
the
being
between
of a wall
POWER
.
the
maitress
the
O2
the
oxygen
another
through
procedure
receptacle.
in
nipple
40
an
not
the
chassis
The
220/240
switch
in
the
Incubator.
on
the
concentration
minutes
ungrounded
below
assume
provide
Vac
Filter
of
of
the
leakage
Units.
Cover.
level
operation,
adapter
current
is
plug,
leakage
resistance,
under
between
verify
so that
the
through a resis-
the
set-up
and a known
not
37%
the
exceed
and
oxygen
100
|
5-9
C100/200
SERVICE
5.3
TROUBLESHOOTING
WARNING:
frame
5.3.1
Troubleshooting form
5.3.2 The
equipment
GENERAL
guides
of
flowcharts.
condition.
EQUIPMENT
equipment
may
be
e
Digital
. + « © Variable
9 e
Voltmeter,
Logic
Probe capable Oscilloscope,Tektronix Probe
Simulator,
Transformer, Resistor, Resistor,
3.6
4.3
The
may
be
at
for
the
lt
is
assumed
REQUIRED
listed
below
substituted.
Fluke
Model
of
Model
No.
General
K,
1/4
K,
1/4
Watt,
Motor
Assembly
mains
Incubators
is
|
12
68
5% 5%
poteniial.
an
attempt
required
8060A
Vdc
465
900
80
Radio
is
not
and
the
Controllers
has
for
troubleshooting
Model
WSM
grounded.
are
to
the
Coniroller
Under
fault
presented
calibrate
the
conditions,
under
and
of
the
Incubator.
the
paragraph
all
Equivalent
metal
5.3.3
cables
in
the
are
in
9
Resistor,
Resistor,
9
Oxygen
5.3.3
ап ac
TROUBLESHOOTING
Unless
procedures Controller
otherwise
power
400
100
Analyzer
specified
source
described
is in
the
Ohms, Ohms,
Incubator.
1/4 1/4
Flowmeter
on
and
the
in
paragraph
the
POWER
Watt,
5%
Watt,
5%
PROCEDURES
troubleshooting
switch
5.2
is
on.
charts,
all
checks
The
troubleshooting
attempted.
Flowcharts
are
charts
the
Controller
assume
and
assume
plugged
the
calibration
the
5-10
Loading...
+ 46 hidden pages